47
SPIRE SUBJECT: Calibration Requirements Document PREPARED BY: Bruce Swinyard DOCUMENT No: SPIRE-RAL-PRJ-001064 ISSUE: DRAFT for Comment Date: 3 January 2002 CHECKED BY: Matt Griffin Date: APPROVED BY: Walter Gear Jean-Paul Baluteau Date:

SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

  • Upload
    others

  • View
    5

  • Download
    0

Embed Size (px)

Citation preview

Page 1: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

SPIRESUBJECT: Calibration Requirements Document

PREPARED BY: Bruce Swinyard

DOCUMENT No: SPIRE-RAL-PRJ-001064

ISSUE: DRAFT for Comment Date: 3 January 2002

CHECKED BY: Matt Griffin Date:

APPROVED BY: Walter Gear

Jean-Paul Baluteau

Date:

Page 2: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 2 of 47

Project Document

SPIRE Bruce Swinyard

Distribution

Matt Griffin Principal Investigator (Cardiff University)Laurent Vigroux Co-Principal Investigator (CEA)Walter Gear Project Scientist (Cardiff University)Jean-Paul Baluteau Project Scientist (LAM)Ken King Project Manager (RAL)Bruce Swinyard Instrument Scientist (RAL)Jamie Bock Co-I (JPL/Caltech)Jason Glenn Assoc. Scientist (University of Colorado)Peter Hargrave Cal Source Scientist (Cardiff University)Gary Davis Co-I (USK)David Smith AIV Manager (RAL)Gustav Ollofsson Co-I (Stockholm)Poalo Saraceno Co-I (IFSI)Sergio Molinari Assoc. Scientist (IFSI)Gillian Wright Co-I (ATC)Sunil Sidher ICC (RAL)Tanya Lim ICC (RAL)Matt Fox ICC (IC)Seb Oliver ICC (Sussex University)Marc Sauvage ICC (CEA)Christophe Morriset ICC (LAM)

Page 3: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 3 of 47

Project Document

SPIRE Bruce Swinyard

CChhaannggee RReeccoorrdd

ISSUE DATEDraft for comment 3 January 2002 First reasonable version sent out for comment to interested

parties

Page 4: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 4 of 47

Project Document

SPIRE Bruce Swinyard

Table of Contents1. Scope....................................................................................................................................82. Derivation of the Calibration Requirements..............................................................................82.1 Scientific Requirements..........................................................................................................82.2 Additional Scientific Requirements ........................................................................................133. Baseline Calibration and Performance Verification Requirements ...........................................153.1 Ground Calibration...............................................................................................................153.1.1 Basic test facility ..............................................................................................................153.1.2 SRD-R1/SRD-R2 Ultimate sensitivity in photometer mode ..................................................153.1.3 SRD-R1/SRD-R2/SRD-R5/SRD-R15 Measurement of the Spatial Impulse Response Functionand Instrument Throughput.............................................................................................................203.1.4 SRD-R6/SRD-R8 Position of the Pixels on the Sky.............................................................243.1.5 SRD-R8/SRD-13/SRD-14 Crosstalk Map; Detector Frequency Response and Linearity ......263.1.6 SRD-R10/CRD-SR1 Spatial Relative Responsivity and Pixel NEP Map...............................293.1.7 CRD-SR2/CRD-SR3/CRD-SR4 Relative Spectral Responsivity; Channel Central Wavelengthand Bandpass; Out of Band Rejection .............................................................................................303.1.8 SRD-R13 Temporal Relative Responsivity .........................................................................323.1.9 SRD-R12 Photometer Absolute Radiometric Accuracy.......................................................343.1.10 CRD-SR5 Rejection of Radiation from Outside the Field of View........................................353.1.11 SRD-R17 FTS Sensitivity – determination of Zero Path Difference Position.........................363.1.12 SRD-R17/SRD-R20/SRD-R22 Velocity Stability; Minimum and Maximum AchievableResolution .....................................................................................................................................373.1.13 SRD-R21 Spectrometer Spectral Response Function ..........................................................403.1.14 SRD-R19 Spectrometer Absolute Radiometric Accuracy....................................................413.2 In orbit calibration ................................................................................................................423.3 AOT Verification.................................................................................................................444. Ground calibration facility outline requirements.......................................................................45

Page 5: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 5 of 47

Project Document

SPIRE Bruce Swinyard

Glossary

ADC Analogue to Digital ConverterAIV Assembly, Integration and VerificationAME Absolute Measurement ErrorAME Absolute Measurement ErrorAOCS Attitude and Orbit Control SystemAPARTAPE Absolute Pointing ErrorASAP Advanced Systems Analysis ProgramAVM Avionics ModelBDA Bolometer Detector ArrayBFL Back Focal LengthBRO Breault Research OrganizationBSM Beam Steering MirrorCDMS Command and Data Management SystemCDMU Command and Data Management UnitCDR Critical Design ReviewCMOS Complimentary Metal Oxide SiliconCPU Central Processing UnitCVV Cryostat Vacuum VesselDAQ Data AcquisitionDCU Detector Control Unit = HSDCUDPU Digital Processing Unit = HSDPUDQE Detective Quantum EfficiencyEGSE Electronic Ground Support EquipmentEGSE Electrical Ground Support EquipmentEMC Electro-magnetic CompatibilityEMI Electro-magnetic InterferenceESA European Space AgencyFCU FCU Control Unit = HSFCUFIR Far InfraredFIRST Far Infra-Red and Submillimetre TelescopeFOV Field of ViewF-P Fabry-PerotFPU Focal Plane UnitFTS Fourier Transform SpectrometerFWHM Full Width Half maximumGSFC Goddard Space Flight CenterHK House KeepingHOB Herschel Optical BenchHPDU Herschel Power Distribution UnitHSDCU Herschel-SPIRE Detector Control UnitHSDPU Herschel-SPIRE Digital Processing UnitHSFCU Herschel-SPIRE FPU Control UnitHSO Herschel Space ObservatoryIF InterfaceIID-A Instrument Interface Document - Part A

Page 6: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 6 of 47

Project Document

SPIRE Bruce Swinyard

IID-B Instrument Interface Document - Part BIMF Initial Mass FunctionIR InfraredIRD Instrument Requirements DocumentIRTS Infrared Telescope in SpaceISM Interstellar MediumJFET Junction Field Effect TransistorLCL Latching Current LimiterLIA Lock-In AmplifierLVDT Linear Variable Differential TransformerMAC Multi Axis ControllerMCU Mechanism Control Unit = HSMCUMDMM-P Martin-PuplettNEP Noise Equivalent PowerNTD Neutron Transmutation DopedOBS On-Board SoftwareOMD Observing Modes DocumentOPD Optical Path DifferencePACS Photodetector Array Camera and SpectrometerPCAL Photometer Calibration sourcePLW Photometer, Long WavelengthPMW Photometer, Medium WavelengthPOF Photometer Observatory FunctionPROM Programmable Read Only MemoryPSW Photometer, Short WavelengthPUS Packet Utilisation StandardSCAL Spectrometer Calibration SourceSCUBA Submillimetre Common User Bolometer ArraySED Spectral Energy DistributionSMEC Spectrometer MechanicsSMPS Switch Mode Power SupplySOF Spectrometer Observatory FunctionSPIRE Spectral and Photometric Imaging ReceiverSRAM Static Random Access MemorySSSD SubSystem Specification DocumentSTP Standard Temperature and PressureSVM Service ModuleTBC To Be ConfirmedTBD To Be DeterminedTC TelecommandURD User Requirements DocumentUV Ultra VioletWE Warm Electronics

Page 7: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 7 of 47

Project Document

SPIRE Bruce Swinyard

References

Applicable Documents

AD1 SPIRE Scientific Requirements SPIRE-UCF-PRJ-000064AD2 SPIRE Instrument Qualification Requirements SPIRE-RAL-PRJ-000592

Reference Documents

RD1 SPIRE Design Description DocumentRD2 “Long Wave Optics”, Derek Martin and John Bowen, IEEE Trans. Microwave Theory

and Techniques 41 Oct 1993, pp1676-1690

Page 8: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 8 of 47

Project Document

SPIRE Bruce Swinyard

1. SCOPE

In this document the calibration requirements for the Herschel SPIRE instrument for both ground and in-orbit testing are laid out. This document will evolve during the instrument definition and subsequentphases as the instrument operation and performance is better understood. It will also be updated in linewith the Herschel calibration requirements as and when they are established.

2. DERIVATION OF THE CALIBRATION REQUIREMENTS

The primary scientific aim of the Herschel SPIRE instrument is to provide high accuracy three bandphotometry of point like objects for deep surveys in the 200 to 670 µm band. A secondary scientificprogramme is the medium resolution spectroscopy and low resolution spectrophotometery of pointsources. To these ends the instrument has been designed with two channels (RD1):

- A three band photometer with three fixed pass bands and three arrays employing single modefeedhorns to couple the radiation from the Herchel telescope and SPIRE optics onto NTDGe bolometer detectors – this is referred to as the photometer throughout this document.

- A two band Fourier Transform Spectrometer using the amplitude beam splitters and the samedetectors technology as the photometer – this is referred to as the spectrometer throughoutthis document.

The Science Requirements Document (AD1 hereafter referred to as the SRD) sets out the top levelperformance and calibration requirements that the instrument must meet in order that it will fulfil thescientific programme of the Herschel mission.

In this section we take each of the scientific requirements and indicate those which require some aspectof the instrument performance to be measured or characterised in order to verify that the scientificrequirement is met. The starting point for this is AD1 and these are discussed in section 2.1. However,some additional top-level requirements are needed in order to verify compliance with those in AD1 –these are described in section 2.2.

2.1 Scientific RequirementsThe scientific requirements in the SRD can be broken down into two main categories: those that onlydrive the basic design of the instrument and those that specify the accuracy to which the instrumentperformance must be characterised. There is of course some blurring between the two but here wewish to distinguish those that will determine what calibration measurements need to be carried out and towhat accuracy. In table 1 the scientific requirements are repeated from the SRD in short form andcomment made on whether they require calibration measurements to be made.

Page 9: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 9 of 47

Project Document

SPIRE Bruce Swinyard

SRD-ID

Requirement Comment Instrumentlevelcalibration?

SRD-R1

The photometer shall be capableof a 60 sq. deg extragalacticsurvey to 1-σ limit of 3 mJy inall bands in 6 months or less

This defines the basicperformance requirements for theinstrument. All aspects of theinstrument performance must becompatible with this requirement

Yes

SRD-R2

The photometer shall be capableof a 1sq. deg galactic survey toa 1-σ limit of 3mJy at 250 µm in1 month or less

The basic performancerequirement here is easily met ifSRD-R1 is satisfied with theexception perhaps of the ability toobserve a crowded field of brightpoint and extended sources.

Yes

SRD-R3

Maximising the “mappingspeed” at which the confusionlimit is reached…. Is the primaryscience driver.

This is a statement about thepriority of the quality of the PSFover the size of the FOV of theinstrument and was used to drivethe design

No

SRD-R4

The photometer observingmodes should provide amechanism for telemeteringundifferenced samples to theground

This is a statement about theproblems of source identificationfrom chopped observations andwas used to define the instrumentdesign.

No

SRD-R5

The photometer shall have anobserving mode the permitsaccurate measurement of thepoint spread function.

The qualifying text associated withthis requirement states that the psfshall be measured “to very highaccuracy”. This requirement istherefore associated both with thedesign of the instrument and thecalibration of the main beam ofthe instrument. Some qualificationof “very high” is requiredhowever!

Yes

SRD-R6

Optical field distortion<10% This is set out as a designrequirement. The requirement onthe accuracy of the actualpositions of the pixels is notexplicitly detailed, although somelimit is set by SRD-R15. We willinterpret this as a requirement tomap the pixel positions as acalibration measurement.

Yes

SRD-R7

Photometer FOV at least 4x4arcminutes

A design driver only No

Page 10: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 10 of 47

Project Document

SPIRE Bruce Swinyard

SRD-ID

Requirement Comment Instrumentlevelcalibration?

SRD-R8

Crosstalk <1% (0.5% goal) foradjacent detectors and 0.1%(0.05% goal) non nearestneighbours.

A complex requirement involvingeverything from the optical inputto the multiplexing and the ADC!An end-to-end calibration isrequired to produce a cross talkmap.

Yes

SRD-R9

Maximum chop throw 4 arcminminimum 10 arcsec

A design requirement for theBSM. No requirement is set onthe repeatability or stability of thechopping mirror. These are infact set as requirements on theBSM and will be dealt with as asub-system issue – see AD2.

No

SRD-R10

The rms detector NEP variationacross any photometer array<20%

A complex requirement involvingeverything from the optical inputthrough to the stability and noiseof the amplifiers. An end-to-endcalibration is required to produce aresponsivity and noise map.

Yes

SRD-R11

The photometer dynamic rangefor astronomical signals shallbe 12 bits or higher

Note the use of a number of bitshere is incorrect – the dynamicrange requirement is 4000. Howthis is encoded into bits is animplementation issueThis is a design driver for theinstrument. The calibration isdealt with under SRD-R14.

No

SRD-R12

SPIRE absolute photometryaccuracy shall be 15% or betterat all wavelengths with a goal of10%

Defines the end-to-end calibrationaccuracy required for both pointsources and extended emission.Both sub-system (filters anddetectors) and instrument levelcalibration will be required toverify this.

Yes

SRD-R13

The relative photometricaccuracy should be shall be10% or better at allwavelengths, with a goal of 5%

This refers to temporalrepeatability of the measurementof a known source. This willrequire the characterisation of thetemporal stability of the system –especially the thermal stability.

Yes

Page 11: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 11 of 47

Project Document

SPIRE Bruce Swinyard

SRD-ID

Requirement Comment Instrumentlevelcalibration?

SRD-R14

SPIRE photometricmeasurements shall be linear to5% over a dynamic range of4000 for astronomical signals.

This defines the usable dynamicrange of the system and isassumed to go from the confusionlimited sensitivity (let us say a S/Nof 20 in a 1 hour observation~0.02 mJy) to a signal 4000xhigher (about 80 Jy on thisdefinition). The instrumentphotometric calibration shallencompass at least this range ofsignal inputs.

Yes

SRD-R15

…the overlapping sets of…detectors…should be co-aligned to within 2 arcsec onthe sky (goal 1 arcsec)

This is a design requirement thatwill need verifying duringinstrument alignment. Theultimate test will be done using thedetectors themselves and a broadband FIR source.

Yes

SRD-R16

Spectrometer optimised for pointsources but with an imagingcapability…

Design driver – no calibrationrequired

No

SRD-R17

The FTS sensitivity shall belimited only by the photon noisefrom the telescope

Note this requirement should bechanged to include the photonnoise from the calibrator –otherwise it is impossible to meet!This looks superficially like adesign driver – however it is acomplex matter and actuallyrequires a good deal of calibrationto ensure that the requirement ismet.

Yes

SRD-R18

The spectrometer dynamic rangefor astronomical signals shallbe 12 bits or higher

Note the use of a number of bitshere is incorrect – the dynamicrange requirement is 4000. Howthis is encoded into bits is animplementation issueThis is a design driver for theinstrument. There is no linearityrequirement for the spectrometer– is it assumed to be the same asfor the photometer? ThereforeSRD-14 pertains?

No

Page 12: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 12 of 47

Project Document

SPIRE Bruce Swinyard

SRD-ID

Requirement Comment Instrumentlevelcalibration?

SRD-R19

The FTS absolute photometry atthe required spectral resolutionshall be 15% or better at allwavelengths with a goal of10%.

Defines the end-to-end calibrationaccuracy required for both pointsources and extended emission.Both sub-system (filters anddetectors) and instrument levelcalibration will be required toverify this.

Yes

SRD-R20

The FTS shall be capable ofmaking spectrophotometricmeasurements with a resolutionof 2 cm-1 with a goal of 4 cm-1

The minimum resolutionachievable is a complex functionof a variety of instrumentperformance factors. An end toend test will be required tocharacterise the achievable photonlimited minimum resolution (seeSRD-17 as well)

Yes

SRD-R21

The width of the FTS instrumentresponse function at therequired resolution shall beuniform to within 10% acrossthe FOV

The line width could be affectedby vignetting of the off-axis pixelsand increased beam shear at offaxis positions. This is both adesign driver and needsverification at instrument level.

Yes

SRD-R22

The maximum spectralresolution of the FTS shall be atleast 0.4 cm-1 with a goal of 0.04cm-1

Note that the spectral resolution isdefined as the FWHM of theresponse function assuming linear(triangular) apodisation of the rawinterferogram

The maximum resolution isgoverned both by the physicalmovement of the SMEC and byany loss in fringe contrast due tobeam shear and vignetting. Thiscan only be characterised atinstrument level.

Yes

SRD-R23

The SPIRE photometer shallhave a an observing modecapable of implementing a 64-point jiggle map…

A design driver – there will be ageneral requirement tocharacterise all SPIRE observingmodes but this isn’t a specificdriver on the instrumentcalibration.

No

SRD-R24

The photometer observingmodes shall include provisionfor 5-point or 7 point jigglemaps…

A design driver – there will be ageneral requirement tocharacterise all SPIRE observingmodes but this isn’t a specificdriver on the instrumentcalibration.

No

Page 13: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 13 of 47

Project Document

SPIRE Bruce Swinyard

SRD-ID

Requirement Comment Instrumentlevelcalibration?

SRD-R25

The photometer shall have apeak-up mode…

A design driver – there will be ageneral requirement tocharacterise all SPIRE observingmodes but this isn’t a specificdriver on the instrumentcalibration.

No

2.2 Additional Scientific RequirementsSome requirements have not been specifically identified in the SRD but will also need calibration. Theseare given in the table below and are given CRDS-R# identifications as new scientific requirements to betested during instrument calibration.

ID Requirement Comment Instrumentlevelcalibration?

CRD-SR1

The photometer and spectrometerrelative response across anindividual array shall be known towithin TBC% with respect to anygiven pixel on the array.

This is the relative calibrationbetween two detectors on a singlearray. It is extremely importantthat this is very well established inorder to allow co-addition of thesignal seen from a source as it isscanned or chopped across thearrays.

Yes

CRD-SR2

The relative response of thespectrometer at any wavelengthwithin the instrument passbandshall be known to within TBC%with respect to the response atany given wavelength.

One of the most importantscientific projects for the FTS isthe ratio of intensities of lines atdifferent wavelengths. Therelative calibration from point topoint in the spectral range,including between detector bands,is extremely important.

Yes

CRD--SR3

The relative response of thephotometer as a function ofwavelength shall be known towithin TBC% with respect to theresponse at any given wavelength.

The actual response vswavelength functions for each ofthe photometer bands are requiredfor the colour correction of thesource intensity. The equivalentwidth and bandpass centroid canbe derived from these functionsand are also used for determiningthe sensitivity of the instrument.

Yes

Page 14: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 14 of 47

Project Document

SPIRE Bruce Swinyard

ID Requirement Comment Instrumentlevelcalibration?

CRD-SR4

The spectral out of band rejectionof the filtering on the instrumentshall be such as to preventcontamination of the survey databy any non-legitimate sourceemission.

The out of band rejection of thefilters will be tested for eachindividual filter but will not be fullyevaluated for the whole opticaltrain except at instrument level. Itis important that there are nospectral leaks in the MIR and NIRbands as bright stars etc willotherwise appear as legitimatesources in the surveys.

Yes

CRD-SR5

The instrument shall be designedso as to reduce the straylight fromsources outside the field of viewof the instrument to less than 5%of the legitimate background fromthe Herschel telescope.

This is the instrument part of thestraylight budget defined in theIID-A. The control of straylightonto the detectors is done byoptical filtering (see CRD-SR4);the use of “clean” optical stopsand baffles and by keeping theinstrument itself cold. Someassessment of the instrument’sability to reject radiation fromsources outside the nominal fieldof view must be made to ensureits compatibility with thisrequirement.

Page 15: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 15 of 47

Project Document

SPIRE Bruce Swinyard

3. BASELINE CALIBRATION AND PERFORMANCE VERIFICATION

REQUIREMENTS

We can now take those SRD requirements identified as needing calibration or performance verificationand break the basic requirement into the parts of the instrument performance that require verificationand calibration on the ground and in flight.

This section is in two parts: in the first the calibration requirements are discussed in some more detailand the method of verification and calibration on the ground is detailed. The parameters required forcalibration of the instrument data are introduced and given a nomenclature that will be used throughoutthis and other documentation. In the second section those calibration measurements that will need to berepeated in flight are discussed together with possible methods for carrying them out.

3.1 Ground Calibration

3.1.1 Basic test facility

We can start with the basic assumption that the ground calibration and test facility for the instrument willconsist of a cryostat that will allow the instrument to be operated at temperatures and backgroundloadings representative of flight. We can also assume that some form of optical simulator is present thatpresents the instrument with an optical beam representative of the Herschel telescope. Therequirements on the ground calibration facility are discussed further in section 4.

3.1.2 SRD-R1/SRD-R2 Ultimate sensitivity in photometer mode

The sensitivity of the instrument is characterised by the Noise Equivalent Power (NEP) usually quotedin W Hz-1/2. The NEP is given in terms that will be measured in practice by:

NEP = RV >δ< W Hz-1/2 (1)

Where <δV> is the voltage noise on the signal referred to the input of the JFET amplifiers measured ina bandwidth of 1 Hz – equivalent to an integration time of 0.5 seconds – and R is the responsivity of thesystem measured in V/W, again referred to the input of the JFET amplifiers (i.e. all electronics gainsremoved). Immediately we have to define at which point in the detection system we define thisresponsivity, and therefore the NEP. The only measure of astronomical interest is to refer this to theinput of the instrument, which can then be referred to the input of the telescope as the Noise EquivalentFlux Density (NEFD) knowing the telescope area and spectral bandwidth of the instrument.

NEFD = λ∆tel

instrument

ANEP W cm-2 µm-1 Hz-1/2 ♦ (2)

where Atel is the telescope geometric area, if we assume that all coupling efficiencies are factored intothe instrument NEP, and ∆λ is the equivalent width of the detector bandpass. This NEFD is then the

♦ Conversion to Janskys from W cm-2 µm-1 is given by I(Jy) = I(sensible units)*λ2/3e-16

Page 16: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 16 of 47

Project Document

SPIRE Bruce Swinyard

minimum detectable flux per beam in a 0.5 second observation for a source that uniformly fills the fieldof view of a detector.

In flight the instrument will be background limited by the flux of photons from the telescope, eachdetector will receive power from the telescope given by:

Pdet = AΩinst εBλ(T) ∆λ ηoptηdet Watts (3)

Here AΩinst is the throughput (also called the etendue or grasp) of the system; εΒλ(T) is the blackbodypower emitted by the telescope at temperature T and emissivity ε ; ηopt is the transmission efficiency ofthe instrument from the input aperture to the front of the feedhorn and ηdet the detector quantumefficiency referred to the front of the feedhorn. The power received at the entrance to the instrument isgiven by:

Pinst = AΩinst εBλ(T) ∆λ Watts (4)

The NEP referred to the instrument aperture due to the photon shot noise is (ignoring the Bose-Einsteincorrection)

NEPpinst = (2Pinsthc/λ)1/2 W Hz-1/2 (5)

And the total NEP will be given by

NEP instrument = (NEPdinst2 + NEPpinst

2)1/2 W Hz-1/2 (6)

Where NEPdinst is the intrinsic NEP of the detector in the absence of any photon background butmeasured under the same operating conditions and referred to the input of the instrument.

We can measure the NEP of the detectors with a blanked off instrument – or equivalently with ablanked off detector – this can be referred to the input of the instrument by removing the instrumentefficiency:

NEPdinst = opt

detNEPη

W Hz-1/2 (7)

and, knowing the absolute power falling on the instrument, we can derive NEP instrument. This can becompared to the direct measurement of the NEP using the measured noise and the responsivity referredto the instrument aperture. This can be derived from the responsivity measured at the detector unit leveland knowledge of the instrument efficiency:

Rinst = ηoptRdet V/W (8)

Two basic methods of deriving the instrument sensitivity are open to us.

1. If we can measure the responsivity, and therefore the NEP, of the detectors at unit level underthe same operating conditions (temperatures and background loading) and using the samespectral input, and we can know the efficiency of the instrument from the input aperture to the

Page 17: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 17 of 47

Project Document

SPIRE Bruce Swinyard

feedhorns by modelling or direct independent measurement, then we can derive the instrumentresponsivity. Measurement of the voltage noise on the DC output of the system with inputpower somewhere representative of flight then gives us the instrument level NEP directly.

We then need only measure the spectral bandpass to obtain the NEFD.

If we also know the absolute power falling on the instrument aperture (see next point) we cancompare the estimated total NEP in equation (6) with that directly measured.

2. If we have an absolutely calibrated blackbody in the beam of the instrument and can measurethe instrument throughput and spectral bandpass we can know absolutely the power falling onthe instrument aperture. We can derive the responsivity of the instrument by varying thetemperature of the blackbody (and therefore the power) and measuring the voltage response ofthe detectors (using V-I curves for instance). Measurement of the voltage noise on the DCoutput of the system with the input power set to the in flight level then gives the instrumentNEP.

In practice both of these methods are likely to be used and both have their advantages and difficulties.In particular the independent measurement of the instrument efficiency, ηopt, is problematic. It can beobtained by comparison of the responsivity measured at detector unit level and at instrument level butthis is a) non-independent and b) pre-supposes that the measurement conditions were equivalent at unitand instrument level. Alternatively we can measure the characteristics of the component parts of theoptical train and combine these into an optical model of the instrument. We should attempt to measureall the relevant parameters (see table below) and cross check the end-to-end derivation against theproduct of the unit level derived parameters.

The Science Requirement actually refers to the detection of point sources to a 1-σ limit. Therefore weare interested in knowing the signal to noise ratio of the background limited performance of theinstrument to the detected flux from a point source. The signal to noise is given by:

σ =instrument

sig

NEP

t2P(10)

and the signal from a point source with a flux spectrum of Sλ W cm-2 um-1 is given by

Psig = Sλ ∆λΑtel ηoptηdet ηpoint Watts (11)

Where ηpoint is the coupling efficiency between the instrument and the beam pattern from the telescope.To fulfil the verification of the requirement we therefore need to measure the relative response of theinstrument to a point and an extended source or, knowing the illumination pattern of the instrument onthe telescope pupil, use optical modelling to calculate ηpoint.

It is pertinent to note here that we do not in practice need to know the absolute detector quantumefficiency ηdet because we always measure the signal in terms of volts – this is converted to power atthe front of the feedhorn by using the detector responsivity, which already includes the detector quantumefficiency.

Page 18: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 18 of 47

Project Document

SPIRE Bruce Swinyard

Parameter/notation/units Description and use Outline GroundMeasurement Method

CRD-PAR-1Detector AbsoluteResponsivity(Rdet – V/W)

Responsivity of an individualdetector referred to the input tothe feedhorn. This is used tocross check the measuredinstrument responsivity againstthat predicted from the unit levelcalibrations.

Measured at unit level using ablackbody placed in the front ofthe feedhorn and any associatedfeed optics. Measurement ofvoltage response vs input powerdone by V-I curve method withdifferent temperatures of theblackbody.

CRD-PAR-2Instrument AbsoluteResponsivity(Rinst – V/W)

DC responsivity of each detectorreferred to the input to theinstrument. This is the basiccorrection between Watts at theentrance of the instrument andvolts referred to the input of thewarm amplification chain. It is afunction of the operatingconditions (temperature; bias;background loading etc) of thedetectors. It must be determinedunder a range of conditions tocreate a matrix that will allow usto predict what it will be duringflight and to pick the optimumconditions for instrumentoperation.

Measured at instrument using ablackbody placed in the beam ofthe instrument beam.Measurement of voltageresponse vs input power done byV-I curve method with differenttemperatures of the blackbody.Under each loading the voltagenoise is also measured to givethe DC NEP as a function ofposition in the field of view.All detectors in all arrays can beread out simultaneously thusgiving both the absoluteresponsivity and the relativeresponsivity of each pixel.The same measurement can bemade with the instrument indifferent operating conditions.

CRD-PAR-3Instrument Efficiency(ηopt – no units)

End-to-end transmissionefficiency of the instrumentoptical train – allows referenceof detected signal to the powerreceived at the input aperture ofthe instrument

Can be deduced from absoluteefficiency measurements over agiven waveband of individualcomponents and/or as a singleend-to-end efficiencymeasurement at instrument levelwith known input power anddetector responsivity.

CRD-PAR-4Channel EquivalentBandpass(∆λ − microns)

The effective spectral width ofthe end-to-end optics; filters anddetectors on each of the threephotometer channels – used toconvert from detected power topower per unit frequency(wavelength) interval.

Can be deduced from spectralrelative efficiency measurementsof individual components and/oras a single end-to-end efficiencymeasurement at instrument level– see later.

Page 19: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 19 of 47

Project Document

SPIRE Bruce Swinyard

Parameter/notation/units Description and use Outline GroundMeasurement Method

CRD-PAR-5Pixel Nominal Load NoiseMap(δVnom – Volts)

Measured voltage noise of thedetection system for eachdetector in each array referredto the input of the warmamplifier chain. This is used toverify the measured NEP undernominal background loading. Itis a function of the operatingconditions (temperature; bias;background loading etc) of thedetectors. It must be determinedunder a range of conditions tocreate a matrix that will allow usto predict what it will be duringflight and to pick the optimumconditions for instrumentoperation.

The background into theinstrument is controlled eitherusing the shutter or by staring ata black surface of well knowntemperature that fills the FOV ofthe instrument. The noise ismeasured for all detectors in allarrays simultaneously.The same measurement can bemade with the instrument indifferent operating conditions.Initially this test will done using awell calibrated externalblackbody; the shutter emissionwill be calibrated against thisblackbody in order to provide thesame or similar conditions duringsystem level testing.

CRD-PAR-6Instrument Throughput(AΩinst − cm2 steradian)

Etendue or grasp of theinstrument which defines howthe radiation from an extendedsource is coupled into thebolometers. It is required toconvert from power detected topower per unit area per unit solidangle.

Difficult! In principle (DerekMartin (RD2)) the A and the Ωcan be measured separately bymeasuring the beam pattern atthe pupil and at field. Seediscussion in section 3.1.3.

CRD-PAR-7Point Source CouplingEfficiency.(ηpoint – no units)

This is the relative amount ofenergy from the telescopeamplitude pattern that, when thetelescope is illuminated by aparallel beam, is detected by thebolometer. It is required toconvert from detected powerfrom a point source to actualinput power into the instrument.

One method (perhaps thesimplest?) is to compare thedetected signal from an extendedand a point source of the sameemitted power per unit area andper unit solid angle. This couldbe achieved using a variableaperture in front of a hotblackbody placed at the inputfocal plane of the telescopesimulator. Alternatively this isalso given (in principle) bymeasurement of the relativeillumination pattern of theinstrument on the telescopesecondary – see below.

Page 20: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 20 of 47

Project Document

SPIRE Bruce Swinyard

3.1.3 SRD-R1/SRD-R2/SRD-R5/SRD-R15 Measurement of the Spatial Impulse ResponseFunction and Instrument Throughput

The Spatial Impulse Response Function describes the shape of the angular response of an individualpixel in the bolometer arrays to a delta function or point source. The most obvious method ofdetermining the impulse response function is to scan a point source across the field of view of anindividual detector at the input focal plane of the telescope simulator. However the interpretation of thismeasurement will suffer from the need to have, inevitably, a source of finite extent and from anydistortion of the field due the systematic behaviour of the telescope simulator. The use of a fullycoherent source with a small angular spread and fully coherent phase is desirable as this can bemodelled to a very high degree of fidelity and therefore de-convolution of the source from the measuredresponse is very much easier.

Another way to determine the point spread function is to measure the response of the instrument to apoint source scanned across the image of the instrument cold stop as projected outside the instrument.The Fourier transform of this response then gives the impulse response function of the instrument on thesky. Making the measurement in this fashion has the advantage that a source of reasonably extent canbe used as it only has to be small compared with the size of the pupil image and the beams of alldetectors can be measured simultaneously. The disadvantage of this method is that no phaseinformation is found and so aberrations in the optics are not detected and the source will need to becarefully designed to avoid problems with straylight.

To measure the throughput of the instrument we need to measure the area of the beam at one of theconjugate planes (focal plane or pupil plane) and the angular extent of the beam at the other. Toillustrate how this can be done by measurement of the response of the instrument we can derive thethroughput for a single Gaussian mode as follows.

The radius of the waist (defined as the Half Width Half Maximum of the intensity pattern) of a singlemode diffraction limited system at the focal plane is given by (Goldsmith):

wo = π

λ)/#f(2 mm (12)

where f/# is the focal ratio of the final optics and the wavelength, λ, defines the units (millimetres hereby convention). The opening angle of the beam from a single mode feed horn – i.e. the angular spreadof the illumination in the far field (at the pupil of the input optics for instance) – is given by

θ ≈ )/#f(2

1 radians (13)

If we assume that the beam is circular in the far field then the solid angle subtended is given by:

Ω = 2π (1 – cos(θ)) (14)

≈ 2π (1 – (1-2

2θ )) = πθ2

Page 21: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 21 of 47

Project Document

SPIRE Bruce Swinyard

= 2)/#f(4

π steradians (15)

The area of the beam at the focal plane waist is given by:

A = πwo2

= π

λ22)/#f(4 mm2 (16)

The throughput is therefore given by:

AΩ ≈λ2 mm2 sr (17)

and we have reproduced the well known result for diffraction limited throughput.

In principle measurement of the point spread function will give both the spatial extent of the intensitypattern at the focal plane – the equivalent waist radius, and, by Fourier transform, the angular spread ofthe beam in the far field. However, because we only measure the intensity it is not always possible tounambiguously disentangle the effects of any multi-mode response of the feedhorns and aberrationscaused by misalignment or other systematic errors in the telescope simulator and/or instrument optics. Itis highly desirable therefore to make a direct measurement of the angular extent of the beam directly atthe instrument pupil image. This measurement also has the advantage of giving the beam filling factorcompared to the full aperture of the telescope and is therefore a direct measurement of the point sourcecoupling efficiency of the instrument.

One other science requirement that is verified during measurement of the point spread function is thaton the co-alignment of the arrays. The position of the image at the output focal plane of the telescopesimulator of a source placed at the input will be absolutely calibrated during the commissioning of thetelescope simulator. The beam conditioning optics of the source used at the input focal plane of thetelescope simulator will also be accurately calibrated. Therefore there will be absolute knowledge of theposition of the input source whatever its wavelength characteristics and both the alignment of the arrayswith respect to each other and their absolute alignment with respect to the telescope boresight can beverified at FIR wavelengths.

Page 22: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 22 of 47

Project Document

SPIRE Bruce Swinyard

Parameter/notation/units Description and use Outline GroundMeasurement Method

CRD-PAR-8Instrument Spatial ImpulseResponse(SIRF - no units)

Relative response of a pixel as afunction of near beam position inthe focal plane of the telescopesimulator. This defines theimpulse response of theinstrument to a point source andis also used in determination ofthe instrument throughput.

The favoured method is to usean FIR laser focussed at theinput focal plane of the telescopesimulator. This gives a welldefined illumination pattern onthe telescope simulator pupil witha flat phase front. A blackbodysource can also be used but thiswill be more difficult to interpretdue to partial coherence andbrightness problems. The imageof the source is finely steppedover the position of the targetpixel preferably in raster patternto determine the 2-D impulseresponse.In principle this measurementshould be made for each pixel ineach array. This may prove totake too long and a minimumsub-set of pixels for which thismeasurement will be carried outwill be identified – see alsooptical distortion measurementbelow. It is also possible thatshorter cross pattern is used forsome pixels rather than the fullraster. Both types ofmeasurement should beavailable.

Page 23: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 23 of 47

Project Document

SPIRE Bruce Swinyard

Parameter/notation/units Description and use Outline GroundMeasurement Method

CRD-PAR-9Telescope Pupil IlluminationFunction(PIF – no units)

Relative response of a pixel as afunction of position in the pupilplane of the telescope simulator.This defines the illuminationpattern of the instrument on theprimary mirror of the Herscheltelescope and is used fordetermination of the instrumentthroughput; the determination ofthe point source couplingefficiency and can be used toverify the measurement of thespatial impulse response.

A small source is scannedacross the location of the imageof the instrument cold stop. Themethod of realising thismeasurement using the telescopesimulator is described in RD2.All pixels are measured with asingle measurement – thepreferred source to use here isthe FIR laser however thisrestricts the measurement to asingle channel per scan and thecomparison between the laserand a black body should be done.Although it is preferred that thefull 2-D area of the pupilresponse is explored this may beimpossible due to practicaldifficulties with implementation.At least a cross patternmeasurement should be possible.

Page 24: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 24 of 47

Project Document

SPIRE Bruce Swinyard

3.1.4 SRD-R6/SRD-R8 Position of the Pixels on the Sky

The image of the detector arrays on the sky may not be perfect due to distortions introduced by theinstrument optics. The arrays may also not be perfectly aligned with the telescope co-ordinate system.The FIR images of the arrays on the sky can be determined by finding the centres of the responsefunctions of pixels at strategic locations in the field of view. In principle this measurement is alreadycovered by the measurement of the impulse response function of these pixels as described above.However here the requirement on the knowledge of the input source characteristics are much less thanrequired for the SIRF measurement because we are only interested in the centroids of the responsefunctions. It will be sufficient to scan the image of a larger than point source black body across theoutput focal plane of the telescope simulator as long as the same source is used for each pixel position tobe measured. Once the positions of the centroids of a subset of the pixels within an array have beendetermined, the positions of the others should be found from optical modelling and alignmentmeasurements on the arrays and the instrument optics.

If the output of the source is stable over time and repeated operations, then this measurement can alsobe used to determine the relative efficiency of each of the pixels. This is useful in determining therelative efficiency of the instrument as a function of position in the instrument field of view (see alsosection on Relative Efficiency)

Page 25: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 25 of 47

Project Document

SPIRE Bruce Swinyard

Parameter/notation/units Description and use Outline GroundMeasurement Method

CRD-PAR-10Pixel Relative Position Map((αdet,δdet) - arcsec)

Relative position of the centre ofthe response of each pixel ineach array. The position isquoted in arcsec with respect tothe boresight of the instrument.This is used to reconstruct thepointing of each pixel on the skyfor the construction of sky mapsand the correct pointing of thetelescope in flight.

A point like blackbody source isplaced at the input focal plane ofthe telescope simulator and theimage scanned is over theposition of each pixel of interest.In principle this measurementshould be made for each pixel ineach array. This may prove totake too long and a minimumsub-set of pixels for which thismeasurement will be carried outwill be identified.It is desirable that the sourcepower be stable and repeatable– this will allow thismeasurement to be used fordetermination of the instrumentspatial relative response function– see below.

Page 26: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 26 of 47

Project Document

SPIRE Bruce Swinyard

3.1.5 SRD-R8/SRD-13/SRD-14 Crosstalk Map; Detector Frequency Response andLinearity

The detector frequency response and linearity should be characterised at unit level ahead of integrationinto the instrument. Indeed we will depend on this knowledge as all measurements on the PSF etc willlikely need to be done with a chopped signal in order to overcome the large and potentially unstablebackground flux♣ . However, the frequency response will not have been verified with the flight likeelectronics and harnesses and a repeat of the unit level characterisation will be necessary on at least afew pixels. This forms part of the verification of the instrument relative response requirement being therelative frequency response of the instrument.

The method is straightforward: a black body is placed at the input to the telescope simulator and theimage centred on the detector of choice – one of the overlapping detectors between the three arrays. Achopper is placed in front of the black body aperture and the signal recorded as the chop frequency isvaried. The relative response of the detector versus frequency is thus established. From this it is asimple matter to determine the detector time constant. The responsivity in the frequency domain isgiven by:

Rinst(ν) = Rinst (1 + (2πν)2τdet2)-1/2 (18)

where τdet is given by

τdet = db32

1πν

seconds (19)

and ν3db is the frequency at which the responsivity of the instrument falls to half of the DC responsivityRinst.

While doing this test we can take advantage of the set up to test two other aspects of the instrumentperformance. Because we are recording data from all the pixels we will automatically have the datasetrequired to establish the degree of cross talk present between the detectors on an individual array. Ifwe use instead a monochromatic source (an FIR laser) then we can also establish the degree of crosstalk between different arrays. Again using a laser as a source would mean that this test could also beextended to test the linearity of the detector by inserting filters into the optical path of the laser we canattenuate it by an accurately known amount. Both the linearity and cross talk characteristics of thedetection system can then be established over a wide range of signal power.

How many pixels will need to be tested in this way is not yet clear; it is probable that all of the pixels tobe used for chopped observations will need to be fully characterised and, as time allows, a sub-set of theother detectors starting with the overlapping detectors across the arrays.

♣ Especially true because we have baselined a warm telescope simulator so the instrument will see a diluted roombackground

Page 27: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 27 of 47

Project Document

SPIRE Bruce Swinyard

Parameter/notation/units Description and use Outline GroundMeasurement Method

CRD-PAR-11Cross Talk Map( cix,ciy - no units)

For each of i detectors, therelative response of thedetectors within the same arrayat position (x,y). Anarray-to-array cross talk mapmight also be necessary so eachpixel tested would then havethree such matrices associatedwith it. Used for removingspurious signals from choppedobservations and ghosting frommapping observations.

A point like blackbody or lasersource is placed at the inputfocal plane of the telescopesimulator and a chopper isplaced in the optical beam. Thedata from all detectors isrecorded and analysed for thepresence of signals at the knownfrequency – these are comparedto the signal recorded at thetarget detector to construct therelative response matrix.In principle this measurementshould be made for each pixel ineach array. This may prove totake too long and a minimumsub-set of pixels for which thismeasurement will be carried outwill be identified.It is desirable that the sourcepower be variable over a widerange in order to detect crosstalk to a level <0.05%.

CRD-PAR-12Detector Response VersusInput Power(Rlinear – no units)

For each detector this is thecorrection to be applied to thedetected signal to refer it to thestandard calibration signal. Thiscorrection removes any non-linear response of the detectionsystem (detector pluselectronics) as a function ofpower at the input to theinstrument. The correction maybe in the form of coefficients ora look up table. Severalcorrections may be necessary asthe non-linearity of the systemcould be a function of chopfrequency and/or operatingconditions.

A point like blackbody or lasersource is placed at the inputfocal plane of the telescopesimulator and a chopper isplaced in the optical beam. Thepower in the beam can be variedusing filters with predeterminedattenuation. The samemeasurement can be used withthe instrument in differentoperating conditions and withdifferent chop frequencies tocomplete the correction matrix.

Page 28: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 28 of 47

Project Document

SPIRE Bruce Swinyard

Parameter/notation/units Description and use Outline GroundMeasurement Method

CRD-PAR-13Detector Time Constant(τdet – seconds)

For each detector this is the timeconstant used to correct theresponsivity to the DC value fortime varying signals of a knownfrequency. Its use is predicatedon the assumption that thedetection system behaves as asingle pole RC filter. If thisproves not to be the case then asingle time constant per detectormay not be enough to describethe frequency response of theinstrument.The time constant will depend onthe operating conditions(temperature; bias; backgroundloading etc) of the detectors. Itmust be determined under arange of conditions to create amatrix that will allow us topredict what it will be duringflight and to pick the optimumconditions for instrumentoperation.

A point like blackbody is placedat the input focal plane of thetelescope simulator and achopper is placed in the opticalbeam. The chopper frequency isvaried from near DC to severaltimes the known or estimated 3dB frequency of the targetdetector with the signal recordedat each step in frequency. Thesame measurement can be usedwith the instrument in differentoperating conditions to determinethe relationship betweenoperating condition andfrequency response.In principle this measurementshould be made for each pixel ineach array. This may prove totake too long and a minimumsub-set of pixels for which thismeasurement will be carried outwill be identified.

Page 29: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 29 of 47

Project Document

SPIRE Bruce Swinyard

3.1.6 SRD-R10/CRD-SR1 Spatial Relative Responsivity and Pixel NEP Map

Once the absolute instrument DC responsivity has been established for a given set of operatingconditions the responsivity of each pixel in each array can be determined relative to the nominaloperating point. Variation in the instrument responsivity as a function of position might be due to anumber of effects – detector responsivity; vignetting; change in the instrument bandpass as a function ofposition; change in the optical efficiency as a function of position etc. Whilst the measurement of therelative DC responsivity is established by the load curve measurement using a source that fills the fieldof view as described above, many of the other possible effects can only be tested using a correctlyconditioned input beam. The measurement used for the determination of the pixel positions on the skywill also give a relative response versus position in the focal plane if the source used is sufficiently stableand repeatable.

To determine the NEP as a function of position in the focal plane we only need then to measure thenoise under the same set of operating conditions as used for the determination of the DC responsivity.The Pixel Nominal Load Noise Map (see above) will also be used to cross check the measured NEP.

Parameter/notation/units Description and use Outline GroundMeasurement Method

CRD-PAR-14Spatial Relative Responsivity(rxy – no units)

Relative response of each pixelin each array with respect to anominal pixel position and set ofoperating conditions. Whilst thismay seem superfluous as wealready have the absoluteinstrument responsivity as afunction of position in the FOV,the determination of the absoluteresponsivity will be done, bynecessity, using a source thatfills the FOV uniformly. Thespatial relative responsivity is therelative point source response ofthe instrument as a function ofpixel position in the FOV. Thisis used to flat field the instrumentimage frames to allowco-addition across differentpixels.

A point like blackbody source isplaced at the input focal plane ofthe telescope simulator and theimage placed at the position ofeach pixel of interest and thesignal recorded.In principle this measurementshould be made for each pixel ineach array. This may prove totake too long and a minimumsub-set of pixels for which thismeasurement will be carried outwill be identified.The source power must bestable and repeatable. Thismeasurement could be combinedwith the measurement of thepixel relative position map.

Page 30: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 30 of 47

Project Document

SPIRE Bruce Swinyard

3.1.7 CRD-SR2/CRD-SR3/CRD-SR4 Relative Spectral Responsivity; Channel CentralWavelength and Bandpass; Out of Band Rejection

The spectral response of each of the instrument channels is the product of the spectral response of thedetectors; the filter spectral transmission profile and the spectral transmission efficiency of the optics.Whilst the first two of these can be, more or less, determined at unit level, the optics will not be verifiedin the FIR until the instrument is complete. Therefore, the combination, and any effects caused by sucha combination, of all three must also be characterised at instrument level. The measurement of thespectral relative efficiency will be done by the use of an external spectrometer with a black body inputsource. The output of the spectrometer will be focussed at the input focal plane of the telescopesimulator and the image placed at the position of each pixel to be measured in turn. At least a subset ofpixels will be measured to the highest possible spectral resolution and good signal to noise to verify thatthere are no problems with channel fringing or other unaccounted for spectral features present. Otherpixels may also be characterised at a lower resolution and/or with lower signal to noise as a confidencecheck.

Once the relative spectral response of the instrument, finst(λ), has been determined the bandpass orequivalent width of each channel is given by:

∆λ = ∫λ

λ

λλu

linst d)(f (20)

And the central wavelength is given by

λc =

∫λ

λ

λ

λ

λλ

λλλ

u

linst

u

linst

d)(f

d)(f

(21)

Here λl and λu are the lower and upper wavelength limits over which the external spectrometeroperates. It is highly desirable that the spectrometer; telescope simulator and any filtering in the groundtest system are able to operate at wavelengths well below and above the nominal bandpass of theinstrument in order to verify the cleanliness of the instrument spectral bandpass. It is possible that theFIR laser could also be used to check the out of band rejection of the instrument in a semi-quantitativemanner. Another method of checking the out of band rejection is to use a bright optical/NIR sourcewith a suitable highpass (short wavelength) optical filter.

The resolution required for the test spectrometer should be such as to find any possible sources ofchannel fringing within the instrument. We do not expect efficient interference to occur in theinstrument at equivalent resolving powers greater than a few hundred and the narrowest featurespresent in the filters will be at much lower resolution than this. Therefore an external spectrometer witha resolution of several hundred at the longest wavelength of interest should be sufficient.

Page 31: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 31 of 47

Project Document

SPIRE Bruce Swinyard

Parameter/notation/units Description and use Outline GroundMeasurement Method

CRD-PAR-15Spectral RelativeResponsivity(finst(λ) – no units)

Relative response of each pixelin each array as a function ofwavelength. This is used tocolour correct the measuredfluxes of objects with known orassumed continuum spectra. Itis also used to determine theequivalent width and centralwavelength of each channel. Inthe spectrometer channels thespectral relative responsivity isused to spectrally “flat field” themeasured source spectra. It isespecially important that thisparameter is accuratelydetermined for eachspectrometer pixel as all areused for taking astronomicalspectra.

A test spectrometer is used tofeed the input focal plane of thetelescope simulator. The outputof the spectrometer should be asnear to a point source aspossible. The image of the testspectrometer output is placed atthe position of a pixel and thewavelength (or position if anFTS) scanned. Thismeasurement should be done athigh (R~few hundred-thousand)resolution and with good s/n(~1000 per spectral bin) in orderto accurately determine therelative spectral responsefunction of the instrument. Eachpixel in the spectrometer channelmust be measured to highaccuracy; only the overlap pixelsin the photometer arrays need bemeasured with high accuracy.Other pixels maybe measured atlow resolution if time allows.

Page 32: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 32 of 47

Project Document

SPIRE Bruce Swinyard

3.1.8 SRD-R13 Temporal Relative Responsivity

The final aspect of the relative responsivity requirement that requires verification is the requirement tobe able to make repeated observations of the same source and achieve the same answer in terms of theflux measured. In order to achieve this the detection system must be intrinsically temporally stable andwe must have a method of determining its stability both on ground and in flight. One method of doingthis is to measure the signal from a known and stable source with the instrument in a known operatingcondition at intervals that sample a wide frequency range. This will show whether there are changes inthe instrument responsivity over time and their characteristic frequencies. On the ground this might beachieved by placing the instrument into a given stable operating condition – photometer scan mode say –and using a stable external source (it doesn’t need to be of known output, just very stable andrepeatable) to make measurements of the instrument relative response at intervals of a few minutes upto, say, 24 hours. This measurement will be a verification of the intrinsic stability rather than acalibration and no calibration parameter is derived.

In flight this is more difficult as it may not always be possible to acquire a calibration target atconvenient times. Therefore, we have incorporated a calibration source into the design of theinstrument that will illuminate all pixels in both the spectrometer and photometer channels. We requirethat this calibration source be characterised during ground testing so that its performance can be verifiedand its output calibrated against a known external reference source. This can be done by operating thecalibrator during the calibration of the DC responsivity of the instrument to give a small amount of extrainput power onto the detectors. The delta signal seen on top of the known load gives a correlationbetween equivalent power seen at the entrance to the instrument and the set point for the calibrator.The response at each detector may be different, as the calibrator may not be seen equally well by alldetectors in all arrays. The stability of the calibrator output is therefore important and this must betested at the same time as the detection system stability.

Page 33: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 33 of 47

Project Document

SPIRE Bruce Swinyard

Parameter/notation/units Description and use Outline GroundMeasurement Method

CRD-PAR-16Calibrator Equivalent Power (CR – W per step)

Calibration between theoperating set point of the internalcalibrator and the signal seen onthe detectors under nominaloperating conditions. This canbe converted to an equivalentflux referred to the inputaperture of the instrument.The calibrator is used for testingthe stability of the detectionsystem and this calibration willbe used to determine theoptimum operation of thecalibrator in flight. Several typesof calibration may be necessaryto give maximum flexibility to themethod of in flight calibration.

In order to absolutely calibratethe equivalent flux from a smallstep in the calibrator output ablackbody can be placed in thebeam of the instrument and setto give the same flux as used forthe DC responsivity calibration.The instrument is set to itsnominal operating condition(bais; temperature etc) and thecalibrator cycled between on andoff with a small output step Toprovide a stable output thecalibrator may have to be drivenat the upper end of itstemperature range, therefore thisstep may be done on top of aDC pedestal. The net DC signalwill be arranged to be the sameby reducing the load from theexternal blackbody.This can be repeated fordifferent flux levels andcalibrator and detector operatingconditions.

Page 34: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 34 of 47

Project Document

SPIRE Bruce Swinyard

3.1.9 SRD-R12 Photometer Absolute Radiometric Accuracy

The ultimate photometric accuracy in the photometer channels depends on all the parameters discussedabove and the errors generated in their measurement. The statistical errors can be minimised using longintegration times during ground testing and systematic errors minimised by design. However, systematicerrors cannot be fully eliminated and we should not suppose that the ground test has generated anunambiguous absolute calibration of the instrument. Additionally, some aspects of the Herschel systemlevel performance (pointing accuracy and reconstruction; telescope effective area etc) will effect theachievable absolute photometric accuracy and these will not be known or fully tested until Herschel is inorbit. Therefore, while there will be a verification of the potential for absolute photometry with SPIREduring ground testing, the ultimate test will be done in flight using a number of calibration targets withknown spectral characteristics (see later).

The one aspect of instrument level performance that may place a fundamental limit on the photometricaccuracy is the stability and repeatability of the instrument responsivity. This in turn is a complexfunction of the temperature stability of the various stages and the electronics performance. Atinstrument level we can characterise the impact of temperature drifts in the FPU and the electronics butwe will only be able to qualitatively assess the absolute system stability. See also the discussion on thecalibrator stability; this too is important as it will be used to monitor and correct any drifts in internalresponsivity.

Page 35: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 35 of 47

Project Document

SPIRE Bruce Swinyard

3.1.10 CRD-SR5 Rejection of Radiation from Outside the Field of View

The optical design of the instrument is intended to prevent radiation from outside the field of view of thetwo sub-instruments from reaching the detectors directly or via any of the mirrors in the optical chain.This design must be verified and an attempt made to identify if there are any straylight paths within theinstrument for sources outside the instrument FOV.

Testing this will be very difficult

Page 36: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 36 of 47

Project Document

SPIRE Bruce Swinyard

3.1.11 SRD-R17 FTS Sensitivity – determination of Zero Path Difference Position

The photometric sensitivity of the FTS when it is stationary and at zero path difference (ZPD) isgoverned by the same parameters as discussed for the photometer. These parameters will be measuredin the same way with the additional step of determining the position of the scanning mirror where ZPDoccurs. This can be done using the internal spectrometer calibrator in one entrance port whilst viewinga cold black surface in the other (either the shutter or viewing an external black surface that fills to fieldof view of the spectrometer). The Spectrometer Mirror Mechanism (SMEC) will be stepped across thecentral maximum of the interferogram and the position at which the signal reaches a maximum is thenthe ZPD position. Once the ZPD position has been determined the SMEC will be set to that positionduring all photometric calibration measurements described above.

Parameter/notation/units Description and use Outline GroundMeasurement Method

CRD-PAR-17Zero Path Difference Position(ZPDP – commanded positionin ADU)

This is the position in the SMECscan range at which the pathlength in the two spectrometerarms is exactly the same. Itshould be the same in bothspectrometer channels (let’shope). It is used to determinewhere to set the SMEC for allphotometric calibrationmeasurements and so that thecorrect scan length can be inputfor each desired resolution.

A cold black body is placed inthe sky port of the FTS thatcomplete fills the spectrometerFOV (either external black bodyor the shutter). Thespectrometer calibrator is set tomaximum output and allowed tostabilise. The SMEC is steppedfrom a known position (homewould be a reasonable start) tosome other towards the knownZPD direction and back againthen to the next position andback and so on until the ZPDhas definitely been transited.The data from the arrays iscollected continuously at a highrate (16 Hz) during thisoperation. The steppingfrequency should be at least 1Hz.The measurement should berepeated with the calibrationsource cold and the externalblack body set to a knowntemperature to ensure that theZPD position is the same in bothpaths through the spectrometer.

Page 37: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 37 of 47

Project Document

SPIRE Bruce Swinyard

3.1.12 SRD-R17/SRD-R20/SRD-R22 Velocity Stability; Minimum and MaximumAchievable Resolution

The stationary responsivity of the bolometer arrays is only one aspect of the FTS sensitivity. Thebaseline operating mode for the FTS will be to scan the SMEC continuously whilst taking data from thedetector arrays. The velocity stability of the SMEC during scanning will be important in determiningboth the ultimate sensitivity of the FTS and the minimum achievable resolution of the FTS in scan mode.

One method of determining the velocity stability – and the basic test of the FTS spectroscopicperformance – is to feed the FTS with a laser line and to scan the SMEC over its full range ofmovement. Because the spectral input is a delta function the measured signal should be a pure cosine.The expected signal can be well modelled and any deviations from the expected signal due to velocityinstabilities or other problems can be relatively easily identified. This test requires that the laser sourceshall be stable over short and medium timescales (0.1-25 Hz is the nominal detection band of thedetectors for instance). The fringe contrast of the cosine as a function of SMEC position will showwhether the maximum specified resolution is actually achievable and whether problems are occurringdue to beam shear etc within the instrument.

Another test of the velocity stability of the FTS is to measure the spectrum of an accurately knownblack body in the absence of the nulling from the spectrometer calibrator. The ability to reconstruct theinput spectrum under these circumstances is critically dependent on the velocity stability of the mirrormovement. The test can be repeated with the spectrometer calibrator set to various levels of output.This directly calibrates the calibrator output and determines the ability of the calibrator to null thetelescope background in flight.

Both these tests should be carried out under a variety of operating conditions for both the SMEC and thedetectors. The spectrometer calibrator may be capable of providing the correct background loading onthe detectors during tests with the laser. This is important as the frequency response of the detectors isone of the determining factors in the minimum achievable resolution.

Page 38: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 38 of 47

Project Document

SPIRE Bruce Swinyard

Parameter/notation/units Description and use Outline GroundMeasurement Method

CRD-PAR-18SMEC Velocity Stability(δV - µm/s/rt(Hz)

This is not a direct calibrationparameter but rather verificationthat the FTS is capable ofachieving the required sensitivityand minimum resolution.

In one test a laser line isfocussed onto one of thespectrometer pixels and theSMEC is scanned over its fullrange of movement. Severalwavelength lines are required tocover the range of the twospectrometer channels. Thelaser must be stable enough thatthe velocity instability can bedistinguished from any sourceinstability.A second test that will determinestability of the velocity is tomeasure the interferogram of ablack body source placed in thesky port of the FTS thatcompletely fills the spectrometerFOV (either external black bodyor the shutter). In this test thespectrometer calibrator is set tozero output.

CRD-PAR-19Spectrometer CalibratorEquivalent Power (SCR – W per step)

Calibration between theoperating set point of thespectrometer internal calibratorsand the flux entering theaperture of the instrument. Thisis used to absolutely calibrate themeasured spectra in flight.

In order to absolutely calibratethe equivalent flux from thecalibrator output a blackbody canbe placed in the sky port of theFTS. The SMEC is scannedover a short range about theZPD. Each of the spectrometercalibrators is stepped over its fullrange of operating points. Whenthe spectrum is nulled the inputpower from the calibrator andexternal source are matched.This test can be repeated fordifferent temperatures of theexternal blackbody and differentmirror scan velocities.

Page 39: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 39 of 47

Project Document

SPIRE Bruce Swinyard

Parameter/notation/units Description and use Outline GroundMeasurement Method

CRD-PAR-20Fringe Contrast Map(FCM - no units)

The fringe contrast on a deltafunction input spectrum mayvary over the scan range of themirror mechanism. A map ofthe variation in contrast will beused to correct the measuredinterferograms.In principle this must bedetermined for each pixel in thespectrometer arrays.

Same test as for velocity stabilityusing a laser line focussed onto asingle pixel. Different lines willbe used to check for anyvariation with wavelength. Thisalso tests the maximumachievable spectral resolution.This test will be repeated for asmany pixels in the spectrometerarrays as practicable.

Page 40: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 40 of 47

Project Document

SPIRE Bruce Swinyard

3.1.13 SRD-R21 Spectrometer Spectral Response Function

The spectral response function of an FTS spectrometer depends critically on the apodisation techniqueused in the treatment of the interferogram before conversion to the spectral domain. We have assumedthat the requirement refers to the FWHM of a linearly apodised interferogram of a spectral deltafunction input after conversion to the spectral domain. To determine how this varies with position in theinstrument FOV the laser will be focussed onto each pixel in the spectrometer array in turn and theSMEC scanned over its full range of movement. No extra measurement is required as the fringecontrast map measurement provides the data set to evaluate the spectral response function.

Parameter/notation/units Description and use Outline GroundMeasurement Method

CRD-PAR-21Spectral Response Function(SRF - no units)

The relative response of eachpixel in the spectrometer to aspectral delta function input inthe spectral domain. This isused to determine the sensitivityof the instrument to spectral linesand will be used to deconvolvethe instrument function frommeasured spectra.

Same data as obtained for theFringe Contrast Maps.

Page 41: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 41 of 47

Project Document

SPIRE Bruce Swinyard

3.1.14 SRD-R19 Spectrometer Absolute Radiometric Accuracy

The same arguments about the validity of ground testing the absolute radiometric accuracy apply to thespectrometer as were discussed above for the photometer. Only the in-flight calibration will determinethe real absolute photometric accuracy achievable.

Two aspects of the instrument performance that will affect the absolute photometric accuracy, and thatwill be capable of characterisation on the ground, are the stability of the calibrator and the repeatabilityof the SMEC velocity control. The spectrometer calibrator is designed both to null the black bodyspectrum from the telescope and to provide a reference comparator that has been accurately calibratedon the ground. The repeatability and stability of the calibrator output will therefore largely determine theaccuracy to which an astronomical signal will be calibrated in flight.

The repeatability of the mirror scan velocity will also have an impact on the ultimate photometricaccuracy of the low resolution spectra from the spectrometer. Whilst we can know what the velocity ofthe mirrors actually was during a scan, the effect of the induced phase changes from velocity instabilitywithin a scan or from scan to scan are a highly complex matter. Our ability to remove the effects ofvelocity instability and non-repeatability will depend critically on an accurate knowledge of the transferfunction of the instrument in frequency space. This can be assessed and characterised on the groundbut it, too, is a complex function of the detector operating conditions; optical performance and stability ofthe detector electronics chains. Only when we are in flight will we be able to characterise theseparameters in a meaningful way.

While there are no new parameters to be measured for this calibration requirement, it does imply thatthe calibrator output must be characterised to ensure its stability – i.e. the measurement of the calibratorequivalent power must be a) carried out over a contiguous typical operating time period to look at driftand b) repeated over a long time base to characterise the calibrator output repeatability.

Similarly, whilst the SMEC velocity stability will be assessed directly, we also need to make anassessment of the repeatability of the velocity over a number of operations.

Page 42: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 42 of 47

Project Document

SPIRE Bruce Swinyard

3.2 In orbit calibration

Single PixelThis lot could be done for both channels with SMEC set to ZPD

Beam profile – for key pixels – full 2-D map using bright point source. Neptune or bright quasar (10-100 Jy)All pixels x-y scans – could be done in scan mode to get relative impulse response and gain map andpixel position.Long dark noise measurement to characterise the 1/f noise spectrum of the system – repeat withchopper on to test for microphonic input – repeat in scan mode across clean area of sky.All long measurements – 10’s of minutes to an hour to look for 1/f knee.

Point Source PhotometryAgain for both channels with SMEC set to ZPD

V-I curve to check loading from telescope – tells us background – need to set optimum bias.Verify bias by doing photometry on bright source and varying biasThen repeat on a weak source to check this is also the best bias in all circumstancesAgain could do this in scan mode for efficiency – need to repeat at various scan speeds at each biassetting to set up the optimum operating point for the scan mode mapping.

Chopped measurement with correct offset and throw on moderately bright point source with variablefrequency from 0.5 to 5 Hz (TBC)

Check of pointing with respect to telescope co-ordinate system – scanned measurement for crude test –then chopped seven point jiggle – tests co-alignment at the same time.

Use a chopped raster measurement to determine the optimum position for chopped observations withdifferent chop throws – finds best chop amplitude.

Do the same measurement in jiggle – slower chopping – and chop between aligned rows…

Routine operation of the calibrators to test for responsivity drifts psuedo random intervals through orbit.

Flux calibration done per modeWill be done with astronomical objects – planets; asteroids; stars etcAll done in seven point mode as primary calibration.

Then use these calibrators to calibrate scan mode as a separate exercise. Scan at different rates; lookat how accurately the telescope can be repointed; accuracy of pointing during scanning.

64-point jiggle done on a point source – repeated with source placed in different parts of the field tocheck the ability to recover the flux from any position in the field.

Make a nodded measurement on moderate point source.

Page 43: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 43 of 47

Project Document

SPIRE Bruce Swinyard

Spectrometer CalibrationAs implied above need initial set up measurement of where the ZPD is – can do this in step and lookmode most accurately (?) with calibrator off looking at bright continuum source and chop on/off thesource as the SMEC is stepped across the maximum in the interferogram or can do it in scan mode withjust the telescope background or partially nulled telescope background. Doing the measurement in scanmode on the telescope should be done anyway as all pixels will respond in this measurement and theposition of the ZPD for each of the pixels can be verified.

Once the ZPD is found we can set this position and carry out all the photometric and beammeasurements listed above. Can we do all detectors together – i.e. run both sets of detectors using thefull data capacity we have available? Check this, if not then we would do it serially – check how longthis takes.

Next we need to look at the SMEC performance for velocity stability and beam shear/vignetting – thiscan only be done on a source with strong lines – need to identify suitable source(s). Run though the setof velocities established on the ground to find the optimum operating point.

We may also need to look at operating the SMEC with a different set of control loop parameters inflight. How do we find the best operating point for these? Could be just done with the engineeringmode and confirmed on the strong line source – if so it is covered by the verification requirements notcalibration.

Once a line source(s) has been identified and the interferogram taken all other performance parametersshould fall out.

We will want to place the line source in each of the pixels in the FOV to establish the off axisperformance of the instrument in flight.

Page 44: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 44 of 47

Project Document

SPIRE Bruce Swinyard

3.3 AOT VerificationHaving discussed the ground and flight calibration needs we have to go on here to look at the AOTsdescribed in the OMD and discuss what calibration measurements are needed to verify these on theground and in flight (some overlap with previous section?). This will place requirements on the groundcal facility.

Page 45: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 45 of 47

Project Document

SPIRE Bruce Swinyard

4. GROUND CALIBRATION FACILITY OUTLINE REQUIREMENTS

Evaluation of the ground test requirements laid out in section 3 leads to the following requirements onthe ground calibration and test facility.

Req. ID Description Source/reasonCRD-GT-R1. A cryostat is required that will house the instrument

and provide the same mechanical; electrical andthermal interfaces as presented by the Herschelcryostat

Needed to provide correct operationalenvironment – no direct calibrationrequirement

CRD-GT-R2. A telescope simulator is required to present theinstrument with a beam that is optically identical tothat of the Herschel telescope.

CRD-PAR-6 – Instrument ThroughputCRD-PAR-8 – Instrument SpatialImpulse Response

CRD-GT-R3. A continuum or monochromatic point like sourceplaced at the input focus of the telescope shall bewell focussed at an image plane at the same relativeposition to SPIRE as will be provided by theHerschel telescope.

CRD-PAR-7 – Point Source CouplingEfficiency

CRD-GT-R4. The telescope beam should be capable of beingplaced and focussed correctly on each pixel withinthe SPIRE FOV. It is not required that the beamfrom the telescope simulator shall simultaneously fillthe SPIRE FOV – i.e. it may be steered from pixelto pixel.

CRD-PAR-6 – Instrument ThroughputCRD-PAR-10 – Pixel Relative PositionMapCRD-PAR-14 – Spatial RelativeResponsivityCRD-PAR-20 – Fringe Contrast Map

CRD-GT-R5. It is desirable that the input FOV of the telescopesimulator is large enough that an extended sourcethat instantaneously fills the beam of a single pixelmay be used.

CRD-PAR-7 – Point Source CouplingEfficiency

CRD-GT-R6. The telescope simulator must be capable ofscanning stepwise the image of a point like coherentsource across the beam of any pixel in the SPIREFOV. The stepsize shall be small enough to providean accurate measurement of the spatial response ofthe instrument – i.e. at least 1/10 of the FWHM at200 µm.

CRD-PAR-6 – Instrument ThroughputCRD-PAR-8 – Instrument SpatialImpulse ResponseCRD-PAR-10 – Pixel Relative PositionMap

CRD-GT-R7. It is desirable that the beam from the telescope canbe scanned continuously across the beam of singlepixel in the SPIRE FOV.

Needed to replicate scan mode AOTs –set a requirement to do this

CRD-GT-R8. The photon background falling onto the entranceaperture of the instrument shall be no greater thanthat expected from an 80-K 4% emissive surfaceduring any measurement set up – i.e. it shall be atleast representative of that from the Herscheltelescope in flight

CRD-PAR-5 – Pixel Nominal LoadNoise MapCRD-PAR-12 – Detector ResponseVersus Input PowerCRD-PAR-13 – Detector TimeConstant

CRD-GT-R9. It shall be possible to have the instrument operate in CRD-PAR-17 – Zero Path Difference

Page 46: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 46 of 47

Project Document

SPIRE Bruce Swinyard

Req. ID Description Source/reasona low background environment – i.e. dark. Position

Also required to trouble shootdetector/microphonic/JFET performance

CRD-GT-R10. A source shall be provided that instantaneouslyilluminates the entire SPIRE FOV. This sourceshall be highly emissive in the FIR and Sub-mm andshall be of known, controllable temperature, andknown emissivity. It is not necessary to feed thissource to the instrument through the telescopesimulator.

CRD-PAR-2 – Instrument AbsoluteResponsivityCRD-PAR-3 – Instrument EfficiencyCRD-PAR-5 – Pixel Nominal LoadNoise MapCRD-PAR-16 – Calibrator EquivalentPowerCRD-PAR-17 – Zero Path DifferencePositionCRD-PAR-18 – SMEC VelocityStabilityCRD-PAR-19 Spectrometer CalibratorEquivalent Power

CRD-GT-R11. An image of the entrance pupil of the telescopesimulator shall be made accessible to allow a pointlike source to be scanned across the pupil locationto make measurements of the instrument beamfalling on the telescope secondary mirror.

CRD-PAR-6 – Instrument ThroughputCRD-PAR-7 – Point Source CouplingEfficiencyCRD-PAR-9 – Telescope PupilIllumination Function

CRD-GT-R12. A spectrometer shall be provided external to theinstrument and fed to the instrument through thetelescope simulator. This spectrometer shall have aresolution of at least 500 at 250 µm; a wavelengthcoverage of at least 200 to 700 µm and shall have acontinuum source at its input.

CRD-PAR-4 – Channel EquivalentBandpassCRD-PAR-15 – Spectral RelativeResponsivity

CRD-GT-R13. A method of measuring the out of band rejection ofthe instrument shall be provided.

CRD-SR4 – Out of Band Rejection

CRD-GT-R14. A method of evaluating the off axis straylightrejection of the instrument shall be provided.

CRD-SR5 – Rejection of Radiation fromOutside the FOV

CRD-GT-R15. A mono-chromatic source with a line width verymuch narrower than 0.04 cm-1 (at least 1/100) shallbe provided that shall be fed to the instrumentthrough the telescope simulator. This source shallhave lines from 100 to 700 µm.

CRD-PAR-6 – Instrument ThroughputCRD-PAR-7 – Point Source CouplingEfficiencyCRD-PAR-8 – Instrument SpatialImpulse ResponseCRD-PAR-18 – SMEC VelocityStabilityCRD-PAR-20 – Fringe Contrast MapCRD-PAR-21 – Spectral ResponseFunction

CRD-GT-R16. A chopper shall be available that can modulate thesignal entering the telescope simulator. Thefrequency of the chopper must be accurately known(within 1%) and variable from 0.5 to 5 Hz. It is

CRD-PAR-11 – Cross Talk MapCRD-PAR-12 – Detector ResponseVersus Input PowerCRD-PAR-13 – Detector Time

Page 47: SPIRE Calibration Requirements Document D1research.uleth.ca/spire/documents/pdf/Calibration_Requirements_Document.pdfR4 The photometer observing modes should provide a mechanism for

Ref: SPIRE-RAL-PRJ-1064

Issue: Draft for CommentDate: 3 January 2002Page: 47 of 47

Project Document

SPIRE Bruce Swinyard

Req. ID Description Source/reasondesirable to have the ability to chop between twosources of different temperature (i.e. hot ~fewhundred K and cold LN2 77 K)

Constant

CRD-GT-R17. A method of commanding and receiving data fromthe instrument that accurately mimics the in-flightcommand and data management system shall beprovided. It shall be capable of simulating actual in-flight operations and allowing astronomicalobservations to be simulated in the ground testfacility.

Required to simulate in-flight operationsand test AOTs etc – add this asrequirement.