Upload
agnes-reeves
View
265
Download
1
Embed Size (px)
Citation preview
Spectrophotometer
Prof.Dr. Moustafa M. MohamedVice Dean
Faculty of Allied Medical SciencePharos University in Alexandria,
EGYPT
Absorption Spectroscopy
Sample
M
MM
M
M
MM
M
M
M
reflection scattering
h
in out
h
****
Absorption SpectroscopyGeneric Instrument
DetectorLightSource
Mono-chromator
Sample
P0 P
P0 = intensity of light into sampleP = intensity of light out of sample
Absorption Spectroscopy
• The more photons the sample absorbs, the lower the intensity (transmission) at the detector.
• Transmittance (T)
• 0 T 1• T is independent of P0• %T = T x 100
0P
PT
Absorption Spectroscopy
P0 = 1.00 P = 0.50 P = 0.25 P = 0.125
• sample with T = 50%
Absorption Spectroscopy
0
0.5
1
Concentration or Path Length
T
Absorption Spectroscopy
• Absorbance (A)
TlogT
1log
P
PlogA 0
TAT
T %log2 ;100
%
0 A
Absorption Spectroscopy
Concentration or Path Length
A
Beer’s Law
A = bcBeer’s Law
Light sources
• Different light sources for different regions of the spectrum
• UV/Vis– Tungsten Lamp 320-2,500 nm - run an
electrical current through a wire in vacuum– Deuterium arc lamp, 200-400 nm - electrical
discharge in D2
– Laser source
Monochromators• Mono - one; chromatic - color• Prisms and gratings - disperse (spread out)
light according to wavelength
Prism
h
Monochromators• n = d(sin I + sin r)
I = constant; therefore r
Ir
d
Sample Holders• Cuvettes
– flat surface best - better reproducibility– avoid fingerprints, dust, etc. on surface– must be transparent in region of interest
Sample Holders
• UV - quartz• Visible - glass, quartz• IR - NaCl
Detectors
• Detector converts incident light to an electrical signal that we can measure and process
Detectors
• Ideal Characteristics– sensitive – linear– flat response v. – stable– fast
Detectors - photo tube
• Good for UV and Visibleh
e-
-Vdynode
amplifier
Detectors - photo multiplier• Characteristics
– sensitive (single photons)– linear– flat response v. within limitations– stable w/ time (sensitive decreases over
time, weeks to months)– fast
Detectors - Array• Channel plate (similar to multiplier)• photodiode array (less sensitive than
multiplier)• charge coupled device (CCD) (more
sensitive than multiplier)• Do not need a monochromator
Colorimeter
• An optical electronic device that measures the color concentration of a substance in solution.
• Optical color filters are used to select a narrow wavelength.
• Basic colorimeter analysis involves the precise measurement of light intensity.
• Transmittance is defined as:I1= T * I0
I0 = Initial light intensity and I1= attenuated light intensity
• The results are displayed in percent optical color transmittance or absorption to indicate hemoglobin concentration.
percentxI
IT
o
1001
Single Beam Instrument
DetectorLightSource
(TungstenLamp)
Filter, Mono-
Chromator, Grating
Sample
QuartzCuvette
Photo-multiplier
Slit
Double Beam Instrument
Slit
Beam Chopper
Reference
Mirror Mirror
Semi-transparentMirror
TungstenLamp
Grating Photo-multiplier
Quartz Cuvette
Sample
Mirror
Colorimeter- filter photometer
• The concentration of the unknown solution can be found from the following relationship:
Cu= Cs Au/As
where
Cu= unknown concentration
Cs= standard concentration (for calibration)
Au= unknown absorbance
As= standard absorbance
• Light passes through an optical color filter, is focused by lenses on the reference and sample cuvettes and falls on the reference and sample photodetectors.
• The difference in voltage between the two detectors is increased by a dc amplifier and applied to a meter.
A calibration procedure is as follows:
1- Ground the amplifier input (V1) and adjust potentiometer (R1) for 0 volt
2- Remove the ground and place reference concentration in cuvettes 1 an 2
3- Adjust potentiometer R1 for 0 V4- Leave the reference concentration in cuvette
1 and replace cuvette 2 with a cuvette containing the sample
5-Read the unbalanced voltage on the meter in percent transmittance or absorbance units.