28
Publication No. AC3451 2016-04-29 Supersedes: 2016-02-10 Specifications for the CARESTREAM DRX Plus Detectors Confidential Important Qualified service personnel must perform these procedures. See the Carestream Service Portal for possible updates before you begin. When doing the procedures in this document, you must use safe work practices and wear the correct personal protective equipment (for example, safety eyewear) according to your company’s standard operating procedures. When servicing this equipment, be sure to read and follow the Manufacturer’s Safety Data Sheets for any chemical, cleaner, solvent, paint, lubricant, adhesive, or replacement battery to understand proper use and disposal precautions.

Specifications for the CARESTREAM DRX Plus Detectorsalbatm.com/wp-content/uploads/2017/06/DRXPlusSpecifications.pdf · Specifications for the CARESTREAM DRX Plus Detectors Confidential

  • Upload
    others

  • View
    63

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Specifications for the CARESTREAM DRX Plus Detectorsalbatm.com/wp-content/uploads/2017/06/DRXPlusSpecifications.pdf · Specifications for the CARESTREAM DRX Plus Detectors Confidential

Publication No. AC34512016-04-29

Supersedes: 2016-02-10

Specifications for the CARESTREAM DRX Plus Detectors

Confidential

Important

• Qualified service personnel must perform these procedures.

• See the Carestream Service Portal for possible updates before you begin.

• When doing the procedures in this document, you must use safe work practices and wear the correct personal protective equipment (for example, safety eyewear) according to your company’s standard operating procedures.

• When servicing this equipment, be sure to read and follow the Manufacturer’s Safety Data Sheets for any chemical, cleaner, solvent, paint, lubricant, adhesive, or replacement battery to understand proper use and disposal precautions.

Page 2: Specifications for the CARESTREAM DRX Plus Detectorsalbatm.com/wp-content/uploads/2017/06/DRXPlusSpecifications.pdf · Specifications for the CARESTREAM DRX Plus Detectors Confidential

All rights reserved. No part of this manual may be reproduced or copied in any form by any mean graphic, electronic, or mechanical, including photocopying, typing, or information retrieval systems without written permission.

Page 3: Specifications for the CARESTREAM DRX Plus Detectorsalbatm.com/wp-content/uploads/2017/06/DRXPlusSpecifications.pdf · Specifications for the CARESTREAM DRX Plus Detectors Confidential

AC3451 | 2016-04-29 1–1

1 DRX Plus 3543 Detector

Detector

Imaging Specifications

Detector Type Amorphous Silicon on glass—no tiling

Conversion Screen Attached Gd202 S:Tb

Pixel Size (μm) Resolution Limit (cyc/mm)

139 x 1393.6

Usable Pixel Area (cm) (inch)(pixels)

35.4 x 42.514 x 16.72544 x 3056

MTF (%) RQA-3 Beam RQA-5 Beam

0.056 cyc/mm, Typical0.5 cyc/mm, Typical1.0 cyc/mm, Typical1.5 cyc/mm, Typical2.0 cyc/mm, Typical2.5 cyc/mm, Typical3.0 cyc/mm, Typical3.5 cyc/mm, Typical

9983614228191310

987954362316107

DQE (%) ±2 sigma < 10 %

IEC 62220-1-1 Edition 1.0, 2015

RQA-3 Beam RQA-5 Beam

0.97 μGy

3.1μGy

9.92 μGy

0.97 μGy

3.1 μGy

9.92 μGy

0.056 cyc/mm, Typical0.5 cyc/mm, Typical1.0 cyc/mm, Typical1.5 cyc/mm, Typical2.0 cyc/mm, Typical2.5 cyc/mm, Typical3.0 cyc/mm, Typical3.5 cyc/mm, Typical

3532272114953

38343025191385

373430252015106

3430251813953

35312520151074

35302520151174

Energy Range 40–150 kVp

Page 4: Specifications for the CARESTREAM DRX Plus Detectorsalbatm.com/wp-content/uploads/2017/06/DRXPlusSpecifications.pdf · Specifications for the CARESTREAM DRX Plus Detectors Confidential

DRX Plus 3543 Detector

1–2 AC3451 | 2016-04-29

Scan Method Progressive

A/D Conversion 16 bits

Imaging Specifications

Supported Data Interfaces

Wireless 802.11 A, or N, or G

Tethered 10/100/1000 Ethernet

Environmental

Shock High shock tolerance

Temperature Range

Operating +15 to +30° C (+59 to +86° F)

Ambient storage and shipping -10 to +66° C (+14 to +150° F)

Relative Humidity, Non-Condensing

Operating 10 to 86 %

Storage 10 to 86 %

Ingress Protection IP57

Mechanical

Size (cm) 35 x 43 Cassette (ISO 4090)38.35 x 45.95 x 1.47 cm

Weight 3.06 kg (6.75 lb)

Housing Material Aluminum

Sensor Protection Material Carbon fiber and aluminum plate

Load Limit Applied to a single 4 cm (1.6 in.) point: 114 kg (250 lb)Distributed evenly over the detector area: 170 kg (375 lb)

Detector Battery

Technology • Lithium-polymer Technology

• “Smart” battery technology prevents overcharge

Size 21.0 x 15.0 x 0.67 cm

Page 5: Specifications for the CARESTREAM DRX Plus Detectorsalbatm.com/wp-content/uploads/2017/06/DRXPlusSpecifications.pdf · Specifications for the CARESTREAM DRX Plus Detectors Confidential

DRX Plus 3543 Detector

AC3451 | 2016-04-29 1–3

Wireless System

Weight 0.4 kg (12.4 oz.)

Voltage / Energy 14.6 to 14.8V DC, 2.1 to 2.5Ah capacity

Battery "Hot Swap" Capability Yes

Charge Capability • 340 maximum images per charge in Direct Connect Mode.

• 14 days per charge in undisturbed Sleep Mode.

Expected Life 500 charge / discharge cycles results in ~80% full charge energy

Medical Safety IEC 62133 —Safety requirements for secondary cells and batter-ies containing alkaline or other non-acid electrolytes

Electromagnetic Compatibility IEC 60601-1-2 Ed. 2.1-Medical Electrical Equipment Electromag-netic Compatibility Requirements and tests, including CISPR 11:1999 + A2:02 emissions Group 1, Class A)

Detector Battery

Technical Specifications

Network Protocol TCP/IP, IPv4, IPv6

Network Type Isolated private wireless LAN (WLAN)

Wireless Protocol 802.11 A, or N, or G

Antenna:Frequency Band Available Channels (fixed at installation)

5 and 2.4 GHz 1, 5, 9, 13, 36, 40, 44, 48, 52, 56, 60, 64, 100, 104, 108, 112, 116, 120, 124, 128, 132, 136, 140, 149, 153, 157, 161, and 165

Maximum Power of Detector Radio 50 mW

Number of Antennas on Detector

2

IP Addressing Static private IP addresses for detectors and access point (AP)

Agency approvals FCC Part 15

Typical Data Size One 15 MB file per image

Dual Homed PC (two NIC cards) Hospital network connection, private network connection

Page 6: Specifications for the CARESTREAM DRX Plus Detectorsalbatm.com/wp-content/uploads/2017/06/DRXPlusSpecifications.pdf · Specifications for the CARESTREAM DRX Plus Detectors Confidential

DRX Plus 3543 Detector

1–4 AC3451 | 2016-04-29

Security

WPA2-PSK AES Factory and user loaded keys

SSID Broadcast

Private Patient Identification Data No patient ID data exchanged with detector

User name and Password Non-default user name and password

Enterprise Wireless

IP Addressing Static or DHCP

Authentication The detector supports authentication with most radius servers using the following protocols: EAP-TLS, EAP-PEAP-MSCHAPV2, EAP-FAST, EAP-PEAP-TLS, EAP-TTLS, EAP-TTLS-MSCHAPV2

Encryption Methods AES-CCMP, WPA2-AES

Certificates supported .pern, .crt, .cer, .der, .p7b, .pfx, .p12 extensions

Keys Private keys with .key extensions

QOS Control signals and data path can be separately configured.

Page 7: Specifications for the CARESTREAM DRX Plus Detectorsalbatm.com/wp-content/uploads/2017/06/DRXPlusSpecifications.pdf · Specifications for the CARESTREAM DRX Plus Detectors Confidential

AC3451 | 2016-04-29 2–1

2 DRX Plus 3543C Detector

Detector

Imaging Specifications

Detector Type Amorphous Silicon on glass—no tiling

Conversion Screen Deposited CsI (TI)

Pixel Size (μm) Resolution Limit (cyc/mm)

139 x 1393.6

Usable Pixel Area (cm) (inch)(pixels)

35.0 x 42.313.8 x 16.62520 x 3032

MTF (%) RQA-3 Beam RQA-5 Beam

0.05 cyc/mm, Typical0.5 cyc/mm, Typical1.0 cyc/mm, Typical1.5 cyc/mm, Typical2.0 cyc/mm, Typical2.5 cyc/mm, Typical3.0 cyc/mm, Typical3.5 cyc/mm, Typical

9982594129211511

9983614432231713

DQE (%) ±2 sigma < 10 % RQA-3 Beam RQA-5 Beam

0.78 μGy

2.5μGy

8.0 μGy

0.78 μGy

2.5 μGy

8.0 μGy

0.056 cyc/mm, Typical0.5 cyc/mm, Typical1.0 cyc/mm, Typical1.5 cyc/mm, Typical2.0 cyc/mm, Typical2.5 cyc/mm, Typical3.0 cyc/mm, Typical3.5 cyc/mm, Typical

7063504132231610

7265544537302113

7263534538322315

7064554840322416

7063565044362818

6761534842352818

Energy Range (kVp) 40–150

Page 8: Specifications for the CARESTREAM DRX Plus Detectorsalbatm.com/wp-content/uploads/2017/06/DRXPlusSpecifications.pdf · Specifications for the CARESTREAM DRX Plus Detectors Confidential

DRX Plus 3543C Detector

2–2 AC3451 | 2016-04-29

Scan Method Progressive

A/D Conversion (bits) 16

Imaging Specifications

Supported Data Interfaces

Wireless 802.11 A, or N, or G

Tethered 10/100/1000 Ethernet

Environmental

Shock High shock tolerance

Temperature Range

Operating +15 to +30° C (+59 to +86° F)

Ambient storage and shipping -10 to +66° C (+14 to +150° F)

Relative Humidity, Non-Condensing %

Operating 10 to 86

Storage 10 to 86

Ingress Protection IP57

Mechanical

Size (cm) 35 x 43 Cassette (ISO 4090)38.35 x 45.95 x 1.47 cm

Weight 3.17 kg (6.99 lb)

Housing Material Aluminum

Sensor Protection Material Carbon fiber and aluminum plate

Weight Limit Applied to a single 4 cm (1.6 in.) point: 114 kg (250 lb)Distributed evenly over the detector area: 170 kg (375 lb)

Detector Battery

Technology • Lithium-polymer Technology

• “Smart” battery technology prevents overcharge

Size 21 x 15 x 0.67 cm

Page 9: Specifications for the CARESTREAM DRX Plus Detectorsalbatm.com/wp-content/uploads/2017/06/DRXPlusSpecifications.pdf · Specifications for the CARESTREAM DRX Plus Detectors Confidential

DRX Plus 3543C Detector

AC3451 | 2016-04-29 2–3

Wireless System

Weight 0.4 kg (12.4 oz.)

Voltage / Energy 14.6 to 14.8V DC, 2.1 to 2.5Ah capacity

Battery "Hot Swap" Capability Yes

Charge Capability • 340 maximum images per charge in Direct Connect Mode.

• 14 days per charge in undisturbed Sleep Mode.

Expected Life 500 charge / discharge cycles results in ~80% full charge energy

Medical Safety IEC 62133 -Safety requirements for secondary cells andbatteries containing alkaline or other non-acid electrolytes

Electromagnetic Compatibility IEC 60601-1-2 Ed. 2.1, 2004-Medical Electrical Equipment Electromagnetic Compatibility Requirements and tests, including CISPR 11:1999 + A2:02 emissions Group 1, Class A)

Detector Battery

Technical Specifications

Network Protocol TCP/IP, IPv4/IPv6

Network Type Isolated private wireless LAN (WLAN) Enterprise Wireless

Wireless Protocol 802.11 A, or N, or G

Antenna:Frequency BandAvailable Channels (fixed at installation)

5 and 2.4 GHz1, 5, 9, 13, 36, 40, 44, 48, 52, 56, 60, 64, 100, 104, 108, 112, 116, 120, 124, 128, 132, 136, 140, 149, 153, 157, 161, and 165

Maximum Power of Detector Radio 50 mW

Number of Antennas on Detector

2

IP Addressing Static private IP addresses for detectors and access point (AP)

Agency approvals FCC Part 15

Typical Data Size One 15 MB file per image

Dual Homed PC (two NIC cards) Hospital network connection, private network connection

Page 10: Specifications for the CARESTREAM DRX Plus Detectorsalbatm.com/wp-content/uploads/2017/06/DRXPlusSpecifications.pdf · Specifications for the CARESTREAM DRX Plus Detectors Confidential

DRX Plus 3543C Detector

2–4 AC3451 | 2016-04-29

Security

WPA2-PSK AES Factory and user loaded keys

SSID Broadcast

Private Patient Identification Data No patient ID data exchanged with detector

User name and Password Non-default user name and password

Enterprise Wireless

IP Addressing Static or DHCP

Authentication The detector supports authentication with most radius servers using the following protocols: EAP-TLS, EAP-PEAP-MSCHAPV2, EAP-FAST, EAP-PEAP-TLS, EAP-TTLS, EAP-TTLS-MSCHAPV2

Encryption Methods AES-CCMP, WPA2-AES

Certificates supported .pern, .crt, .cer, .der, .p7b, .pfx, .p12 extensions

Keys Private keys with .key extensions

QOS Control signals and data path can be separately configured.

Page 11: Specifications for the CARESTREAM DRX Plus Detectorsalbatm.com/wp-content/uploads/2017/06/DRXPlusSpecifications.pdf · Specifications for the CARESTREAM DRX Plus Detectors Confidential

AC3451 | 2016-04-29 3–1

3 DRX Plus 4343 Detector

Detector

Imaging Specifications

Receptor Type Amorphous Silicon on glass—no tiling

Conversion Screen Attached Gd202 S:Tb

Pixel Size (μm) Resolution Limit (cyc/mm)

139 x 1393.6

Usable Pixel Area (cm) (inch)(pixels)

42.5 x 42.516.7 x 16.73056 x 3056

MTF (%) RQA-3 Beam RQA-5 Beam

0.056 cyc/mm, Typical0.5 cyc/mm, Typical1.0 cyc/mm, Typical1.5 cyc/mm, Typical2.0 cyc/mm, Typical2.5 cyc/mm, Typical3.0 cyc/mm, Typical3.5 cyc/mm, Typical

9983614228191310

987954362316107

DQE (%) ±2 sigma < 10 % RQA-3 Beam RQA-5 Beam

IEC 62220-1-1, Edition 1.0, 2015 0.97 μGy

3.1μGy

9.92 μGy

0.97 μGy

3.1 μGy

9.92 μGy

0.056 cyc/mm, Typical0.5 cyc/mm, Typical1.0 cyc/mm, Typical1.5 cyc/mm, Typical2.0 cyc/mm, Typical2.5 cyc/mm, Typical3.0 cyc/mm, Typical3.5 cyc/mm, Typical

3532272114953

38343025191385

373430252015106

3430251813953

35312520151074

35302520151174

Energy Range 40–150 kVp

Page 12: Specifications for the CARESTREAM DRX Plus Detectorsalbatm.com/wp-content/uploads/2017/06/DRXPlusSpecifications.pdf · Specifications for the CARESTREAM DRX Plus Detectors Confidential

DRX Plus 4343 Detector

3–2 AC3451 | 2016-04-29

Scan Method Progressive

A/D Conversion 16 bits

Imaging Specifications

Supported Data Interfaces

Wireless 802.11 A, or N, or G

Tethered 10/100/1000 Ethernet

Environmental

Shock High shock tolerance

Temperature Range

Operating +15 to +30° C (+59 to +86° F)

Ambient storage and shipping -10 to +66° C (+14 to +150° F)

Relative Humidity, Non-Condensing

Operating 10 to 86 %

Storage 10 to 86 %

Ingress Protection IP57

Mechanical

Size (cm) 43 x 43 Cassette (ISO 4090)45.95 x 45.95 x 1.47 cm

Weight 3.63 kg (8.0 lb)

Housing Material Aluminum

Sensor Protection Material Carbon fiber and aluminum plate

Load Limit Applied to a single 4 cm (1.6 in.) point: 114 kg (250 lb)Distributed evenly over the detector area: 170 kg (375 lb)

Detector Battery

Technology • Lithium-polymer Technology

• “Smart” battery technology prevents overcharge

Size 21.0 x 15.0 x 0.67 cm

Page 13: Specifications for the CARESTREAM DRX Plus Detectorsalbatm.com/wp-content/uploads/2017/06/DRXPlusSpecifications.pdf · Specifications for the CARESTREAM DRX Plus Detectors Confidential

DRX Plus 4343 Detector

AC3451 | 2016-04-29 3–3

Wireless System

Weight 0.4 kg (12.4 oz.)

Voltage / Energy 14.6 to 14.8V DC, 2.1 to 2.5Ah capacity

Battery "Hot Swap" Capability Yes

Charge Capability • 340 maximum images per charge in Direct Connect Mode.

• 14 days per charge in undisturbed Sleep Mode.

Expected Life 500 charge / discharge cycles results in ~80% full charge energy

Medical Safety IEC 62133 —Safety requirements for secondary cells andbatter-ies containing alkaline or other non-acid electrolytes

Electromagnetic Compatibility IEC 60601–1–2 Ed. 2.1 —Medical Electrical EquipmentElectro-magnetic Compatibility Requirements and tests,including CISPR 11:1999 + A2:02 emissions Group 1, ClassA)

Detector Battery

Technical Specifications

Network Protocol TCP/IP, IPv4, IPv6

Network Type Isolated private wireless LAN (WLAN)Enterprise Wireless

Wireless Protocol 802.11 A, or N, or G

Antenna:Frequency Band Available Channels (fixed at installation)

5 and 2.4 GHz 1, 5, 9, 13, 36, 40, 44, 48, 52, 56, 60, 64, 100, 104, 108, 112, 116, 120, 124, 128, 132, 136, 140, 149, 153, 157, 161, and 165

Maximum Power of Detector Radio 50 mW

Number of Antennas on Detector

2

IP Addressing Static private IP addresses for detectors and access point (AP)

Agency approvals FCC Part 15

Typical Data Size One 18.5 MB file per image

Dual Homed PC (two NIC cards) Hospital network connection, private network connection

Page 14: Specifications for the CARESTREAM DRX Plus Detectorsalbatm.com/wp-content/uploads/2017/06/DRXPlusSpecifications.pdf · Specifications for the CARESTREAM DRX Plus Detectors Confidential

DRX Plus 4343 Detector

3–4 AC3451 | 2016-04-29

Security

WPA2-PSK AES Factory and user loaded keys

SSID Broadcast

Private Patient Identification Data No patient ID data exchanged with detector

User name and Password Non-default user name and password

Enterprise Wireless

IP Addressing Static or DHCP

Authentication The detector supports authentication with most radius servers using the following protocols: EAP-TLS, EAP-PEAP-MSCHAPV2, EAP-FAST, EAP-PEAP-TLS, EAP-TTLS, EAP-TTLS-MSCHAPV2

Encryption Methods AES-CCMP, WPA2-AES

Certificates supported .pern, .crt, .cer, .der, .p7b, .pfx, .p12 extensions

Keys Private keys with .key extensions

QOS Control signals and data path can be separately configured.

Page 15: Specifications for the CARESTREAM DRX Plus Detectorsalbatm.com/wp-content/uploads/2017/06/DRXPlusSpecifications.pdf · Specifications for the CARESTREAM DRX Plus Detectors Confidential

AC3451 | 2016-04-29 4–1

4 DRX Plus 4343C Detector

Detector

Imaging Specifications

Receptor Type Amorphous Silicon on glass—no tiling

Conversion Screen Deposited CsI (Tl)

Pixel Size (μm) Resolution Limit (cyc/mm)

139 x 1393.6

Usable Pixel Area (cm) (inch)(pixels)

42.3 x 42.316.6 x 16.63032 x 3032

MTF (%) RQA-3 Beam RQA-5 Beam

0.056 cyc/mm, Typical0.5 cyc/mm, Typical1.0 cyc/mm, Typical1.5 cyc/mm, Typical2.0 cyc/mm, Typical2.5 cyc/mm, Typical3.0 cyc/mm, Typical3.5 cyc/mm, Typical

9982594129211511

9983614432231713

DQE (%) ±2 sigma < 10 % RQA-3 Beam RQA-5 Beam

IEC 62220-1-1, Edition 1.0, 2015 0.78 μGy

2.5μGy

8.0 μGy

0.78 μGy

2.5 μGy

8.0 μGy

0.056 cyc/mm, Typical0.5 cyc/mm, Typical1.0 cyc/mm, Typical1.5 cyc/mm, Typical2.0 cyc/mm, Typical2.5 cyc/mm, Typical3.0 cyc/mm, Typical3.5 cyc/mm, Typical

7063504132231610

7265544537302113

7263534538322315

7064554840322416

7063565044362818

6761534842352818

Energy Range 40–150 kVp

Page 16: Specifications for the CARESTREAM DRX Plus Detectorsalbatm.com/wp-content/uploads/2017/06/DRXPlusSpecifications.pdf · Specifications for the CARESTREAM DRX Plus Detectors Confidential

DRX Plus 4343C Detector

4–2 AC3451 | 2016-04-29

Scan Method Progressive

A/D Conversion 16 bits

Imaging Specifications

Supported Data Interfaces

Wireless 802.11A, or N, or G

Tethered 10/100/1000 Ethernet

Environmental

Shock High shock tolerance

Temperature Range

Operating +15 to +30° C (+59 to +86° F)

Ambient storage and shipping -10 to +66° C (+14 to +150° F)

Relative Humidity, Non-Condensing

Operating 10 to 86 %

Storage 10 to 86 %

Ingress Protection IP57

Mechanical

Size (cm) 43 x 43 Cassette (ISO 4090)45.95 x 45.95 x 1.47 cm

Weight 3.74 kg (8.25 lb)

Housing Material Aluminum

Sensor Protection Material Carbon fiber and aluminum plate

Load Limit Applied to a single 4 cm (1.6 in.) point: 114 kg (250 lb)Distributed evenly over the detector area: 170 kg (375 lb)

Detector Battery

Technology • Lithium-polymer Technology

• “Smart” battery technology prevents overcharge

Size 21.0 x 15.0 x 0.67 cm

Page 17: Specifications for the CARESTREAM DRX Plus Detectorsalbatm.com/wp-content/uploads/2017/06/DRXPlusSpecifications.pdf · Specifications for the CARESTREAM DRX Plus Detectors Confidential

DRX Plus 4343C Detector

AC3451 | 2016-04-29 4–3

Wireless System

Weight 0.4 kg (12.4 oz.)

Voltage / Energy 14.6 to 14.8V DC, 2.1 to 2.5Ah capacity

Battery "Hot Swap" Capability Yes

Charge Capability • 340 maximum images per charge in Direct Connect Mode.

• 14 days per charge in undisturbed Sleep Mode.

Expected Life 500 charge / discharge cycles results in ~80% full charge energy

Medical Safety IEC 62133:—Safety requirements for secondary cells and batter-ies containing alkaline or other non-acid electrolytes

Electromagnetic Compatibility IEC 60601-1-2 Ed. 2.1, 2004-Medical Electrical Equipment Elec-tromagnetic Compatibility Requirements and tests, including CISPR 11:1999 + A2:02 emissions Group 1, Class A)

Detector Battery

Technical Specifications

Network Protocol TCP/IP, IPv4, IPv6

Network Type Isolated private wireless LAN (WLAN)Enterprise Wireless

Wireless Protocol 802.11 A, or N, or G

Antenna:Frequency Band Available Channels (fixed at installation)

5 and 2.4 GHz 1, 5, 9, 13, 36, 40, 44, 48, 52, 56, 60, 64, 100, 104, 108, 112, 116, 120, 124, 128, 132, 136, 140, 149, 153, 157, 161, and 165

Maximum Power of Detector Radio 50 mW

Number of Antennas on Detector

2

IP Addressing Static private IP addresses for detectors and access point (AP)

Agency approvals FCC Part 15

Typical Data Size One 18.5 MB file per image

Dual Homed PC (two NIC cards) Hospital network connection, private network connection

Page 18: Specifications for the CARESTREAM DRX Plus Detectorsalbatm.com/wp-content/uploads/2017/06/DRXPlusSpecifications.pdf · Specifications for the CARESTREAM DRX Plus Detectors Confidential

DRX Plus 4343C Detector

4–4 AC3451 | 2016-04-29

Security

WPA2-PSK AES Factory and user loaded keys

SSID Broadcast

Private Patient Identification Data No patient ID data exchanged with detector

User name and Password Non-default user name and password

Enterprise Wireless

IP Addressing Static or DHCP

Authentication The detector supports authentication with most radius servers using the following protocols: EAP-TLS, EAP-PEAP-MSCHAPV2, EAP-FAST, EAP-PEAP-TLS, EAP-TTLS, EAP-TTLS-MSCHAPV2

Encryption Methods AES-CCMP, WPA2-AES

Certificates supported .pern, .crt, .cer, .der, .p7b, .pfx, .p12 extensions

Keys Private keys with .key extensions

QOS Control signals and data path can be separately configured.

Page 19: Specifications for the CARESTREAM DRX Plus Detectorsalbatm.com/wp-content/uploads/2017/06/DRXPlusSpecifications.pdf · Specifications for the CARESTREAM DRX Plus Detectors Confidential

AC3451 | 2016-04-29 5–1

5 DRX Plus 4343F Detector

Imaging Specifications

Receptor Type Amorphous Silicon on glass—no tiling

Conversion Screen Attached Gd202 S:Tb

Pixel Size (μm) Resolution Limit (cyc/mm)

139 x 1393.6

Usable Pixel Area (cm) (inch)(pixels)

42.5 x 42.516.7 x 16.73056 x 3056

MTF (%) RQA-3 Beam RQA-5 Beam

0.056 cyc/mm, Typical0.5 cyc/mm, Typical1.0 cyc/mm, Typical1.5 cyc/mm, Typical2.0 cyc/mm, Typical2.5 cyc/mm, Typical3.0 cyc/mm, Typical3.5 cyc/mm, Typical

9983614228191310

987954362316107

DQE (%) ±2 sigma < 10 % RQA-3 Beam RQA-5 Beam

IEC 62220-1-1, Edition 1.0, 2015 0.97 μGy

3.1μGy

9.92 μGy

0.97 μGy

3.1 μGy

9.92 μGy

0.056 cyc/mm, Typical0.5 cyc/mm, Typical1.0 cyc/mm, Typical1.5 cyc/mm, Typical2.0 cyc/mm, Typical2.5 cyc/mm, Typical3.0 cyc/mm, Typical3.5 cyc/mm, Typical

3532272114953

38343025191385

373430252015106

3430251813953

35312520151074

35302520151174

Energy Range 40–150 kVp

Page 20: Specifications for the CARESTREAM DRX Plus Detectorsalbatm.com/wp-content/uploads/2017/06/DRXPlusSpecifications.pdf · Specifications for the CARESTREAM DRX Plus Detectors Confidential

DRX Plus 4343F Detector

5–2 AC3451 | 2016-04-29

Scan Method Progressive

A/D Conversion 16 bits

Imaging Specifications

Supported Data Interfaces

Tethered 10/100/1000 Ethernet

Environmental

Shock High shock tolerance

Temperature Range

Operating +15 to +30° C (+59 to +86° F)

Ambient storage and shipping -10 to +66° C (+14 to +150° F)

Relative Humidity, Non-Condensing

Operating 10 to 86 %

Storage 10 to 86 %

Ingress Protection IPX0

Mechanical

Size (cm) 43 x 43 Cassette (ISO 4090)45.95 x 45.95 x 1.47 cm

Weight 3.53 kg (7.8 lb)

Housing Material Aluminum

Sensor Protection Material Carbon fiber and aluminum plate

Load Limit Applied to a single 4 cm (1.6 in.) point: 114 kg (250 lb)Distributed evenly over the detector area: 170 kg (375 lb)

Security

Private Patient Identification Data No patient ID data exchanged with detector

User name and Password Non-default user name and password

Page 21: Specifications for the CARESTREAM DRX Plus Detectorsalbatm.com/wp-content/uploads/2017/06/DRXPlusSpecifications.pdf · Specifications for the CARESTREAM DRX Plus Detectors Confidential

AC3451 | 2016-04-29 6–1

6 DRX Plus 4343FC Detector

Imaging Specifications

Receptor Type Amorphous Silicon on glass—no tiling

Conversion Screen Deposited CsI (Tl)

Pixel Size (μm) Resolution Limit (cyc/mm)

139 x 1393.6

Usable Pixel Area (cm) (inch)(pixels)

42.3 x 42.316.6 x 16.63032 x 3032

MTF (%) RQA-3 Beam RQA-5 Beam

0.056 cyc/mm, Typical0.5 cyc/mm, Typical1.0 cyc/mm, Typical1.5 cyc/mm, Typical2.0 cyc/mm, Typical2.5 cyc/mm, Typical3.0 cyc/mm, Typical3.5 cyc/mm, Typical

9982594129211511

9983614432231713

DQE (%) ±2 sigma < 10 % RQA-3 Beam RQA-5 Beam

IEC 62220-1-1, Edition 1.0, 2015 0.78 μGy

2.5μGy

8.0 μGy

0.78 μGy

2.5 μGy

8.0 μGy

0.056 cyc/mm, Typical0.5 cyc/mm, Typical1.0 cyc/mm, Typical1.5 cyc/mm, Typical2.0 cyc/mm, Typical2.5 cyc/mm, Typical3.0 cyc/mm, Typical3.5 cyc/mm, Typical

7063504132231610

7265544537302113

7263534538322315

7064554840322416

7063565044362818

6761534842352818

Energy Range 40–150 kVp

Page 22: Specifications for the CARESTREAM DRX Plus Detectorsalbatm.com/wp-content/uploads/2017/06/DRXPlusSpecifications.pdf · Specifications for the CARESTREAM DRX Plus Detectors Confidential

DRX Plus 4343FC Detector

6–2 AC3451 | 2016-04-29

Scan Method Progressive

A/D Conversion 16 bits

Imaging Specifications

Supported Data Interfaces

Tethered 10/100/1000 Ethernet

Environmental

Shock High shock tolerance

Temperature Range

Operating +15 to +30° C (+59 to +86° F)

Ambient storage and shipping -10 to +66° C (+14 to +150° F)

Relative Humidity, Non-Condensing

Operating 10 to 86 %

Storage 10 to 86 %

Ingress Protection IPX0

Mechanical

Size (cm) 43 x 43 Cassette (ISO 4090)45.95 x 45.95 x 1.47 cm

Weight 3.63 kg (8.0 lb)

Housing Material Aluminum

Sensor Protection Material Carbon fiber and aluminum plate

Load Limit Applied to a single 4 cm (1.6 in.) point: 114 kg (250 lb)Distributed evenly over the detector area: 170 kg (375 lb)

Security

Private Patient Identification Data No patient ID data exchanged with detector

User name and Password Non-default user name and password

Page 23: Specifications for the CARESTREAM DRX Plus Detectorsalbatm.com/wp-content/uploads/2017/06/DRXPlusSpecifications.pdf · Specifications for the CARESTREAM DRX Plus Detectors Confidential

AC3451 | 2016-04-29 7–1

7 DRX Plus Statements Related to Systems Integration

Detector Safety

The detector has been approved to IEC 60601-1 by TUV Rheinland and has a cTUVus and CE safety certification as a Medical Electrical Equipment device.

Recycling

The detector should be returned to Carestream Health at the end of life. The components will be recycled appropriately.

The Lithium-polymer battery is WEEE compliant (as shown by the Wheelie Dustbin logo) and should be recycled at local recycling stations. Carestream Health pays taxes in the EU Battery Directive countries to cover the recycling costs.

Line Noise Correction

The Image Correction Preference is used for the following purpose:

• Normal—Reduces correlated line noise due to readout ICs. This is a very subtle correction, limited to near zero signal levels.

• Enhanced—Reduces correlated line noise due to external EMI sources, such as motor drives and power supplies. This level should only be used if there is detectable line noise in the images.

Grids

• Moving Grids: Moving grids should not exhibit grid aliasing. If the exposure time is very short, the grid may alias with any digital detector. This is normally not a problem, since grids are used on large body parts, which generally require exposure times longer than 20 ms.

• Low Resolution Stationary Grids: 30-to-50 Lines/cm (76-to-127 Lines/inch) Carestream image processing software will automatically remove the visible lines from the full resolution image.

• High Resolution Stationary Grids: ≥ 75 Lines/cm (>191 Lines/inch)—Grid resolutions of 75 Lines/cm or greater will not be visible on images.

• Medium Resolution Stationary Grids: 50-to-75 Lines/cm (127-to-191 Lines/inch)—Stationary grids in this range may cause aliasing and are therefore not recommended. However, medium resolution moving grids should not exhibit aliasing.

Page 24: Specifications for the CARESTREAM DRX Plus Detectorsalbatm.com/wp-content/uploads/2017/06/DRXPlusSpecifications.pdf · Specifications for the CARESTREAM DRX Plus Detectors Confidential

DRX Plus Statements Related to Systems Integration

7–2 AC3451 | 2016-04-29

Durability

The detector has been designed to be a robust device and function properly in normal handling. The instrument may be damaged if it receives a strong jolt from a drop or strike. The warranty does not cover damage as a result of accident or abuse, including broken detector glass and other obvious damage, however a drop rider can be purchased to protect the investment. The detector has been equipped with internal accelerometers that will register when the detector has incurred a significant shock. The drop height that results in a registered event depends on the orientation of the detector during the drop, and the stiffness of the surface it strikes.

Detector Life

The detector system is specified to last 10 years under normal use of 56,000 images per year. That is ~560K images. This statement relates to the life of the physical aspects of the product, including hardware, normal physical handling and image acquisition

No Preventive Maintenance (PM) is required on the system.

Detector Quality Assurance

The detector must undergo gain calibration on an annual basis. A 360 day reminder is built into the software.

Express Calibration restores Gain Calibration and Defect Calibration and does a defect quality check. Full Calibration is the same as Express Calibration and adds quality checks for Signal Uniformity, Noise Uniformity and Dark Noise

Permanent pixel defects are identified in the factory based on outlier performance in dark noise, X-ray uniformity response, and time lag response. Additional pixel defects are identified during Express Calibration and Full Calibration

Quality Limits

• No 4 x 4 defect clusters.

• Signal Uniformity = 2 %

• Noise Uniformity = 40 % (lowest exposure), 31% (1/3 gain calibration exposure), 28 % (gain calibration exposure), 28 % (2 times gain calibration exposure)

• Two Dark Noise = 9.2 ADC

Wireless FAQs

What impact will the system AP have on the Hospital/Clinic WLAN?

The system AP will have the channel fixed and the transmit power reduced to the minimum level needed for reliable operation of the detector moving around the room in which the AP is installed. Because of the intermittent bursts of data lasting only a few seconds, and the low transmission power, the impact of the system WLAN on the Hospital mobile devices is minimized.

Page 25: Specifications for the CARESTREAM DRX Plus Detectorsalbatm.com/wp-content/uploads/2017/06/DRXPlusSpecifications.pdf · Specifications for the CARESTREAM DRX Plus Detectors Confidential

DRX Plus Statements Related to Systems Integration

AC3451 | 2016-04-29 7–3

What is the range of the wireless communication?

The wireless communication should have plenty of range for any single room installation. The connection range is typically more than 30 m, depending on the obstructions between the access point and the detector

Page 26: Specifications for the CARESTREAM DRX Plus Detectorsalbatm.com/wp-content/uploads/2017/06/DRXPlusSpecifications.pdf · Specifications for the CARESTREAM DRX Plus Detectors Confidential

DRX Plus Statements Related to Systems Integration

7–4 AC3451 | 2016-04-29

Page 27: Specifications for the CARESTREAM DRX Plus Detectorsalbatm.com/wp-content/uploads/2017/06/DRXPlusSpecifications.pdf · Specifications for the CARESTREAM DRX Plus Detectors Confidential

AC3451 | 2016-04-29 I

Publication History

Version Date Changes

A 2015-06-16 First release

B 2016-02-10 Updated the specifications for MTF and DQE.

C 2016-04-29 Added information for 4343 detectors.

Page 28: Specifications for the CARESTREAM DRX Plus Detectorsalbatm.com/wp-content/uploads/2017/06/DRXPlusSpecifications.pdf · Specifications for the CARESTREAM DRX Plus Detectors Confidential

Carestream Health

150 Verona Street

Rochester, NY 14608

USA

© Carestream Health, Inc., 2016

Created in the United States

CARESTREAM is a trademark of Carestream Health.

Pub. No. AC3451

Rev. C