Soundness Cement

Embed Size (px)

Citation preview

  • 8/13/2019 Soundness Cement

    1/4

    Soundness

    It is essential that the cement paste, once it has set, does not undergo a large change in volume.

    One restriction is that there must be no appreciable expansion, which under conditions of

    restraint could result in disruption of the hardened cement paste. Such expansion may occur due

    to reactions of free lime, magnesia and calcium sulfate, and cements exhibiting this type of

    expansion are classified as unsound. Free lime is present in the clinker and is intercrystallized

    with other compounds; consequently, it hydrates very slowly occupying a larger volume than the

    original free calcium oxide. Free lime cannot be determined by chemical analysis of cement

    because it is not possible to distinguish between unreacted CaO and Ca(OH)2 produced by a

    partial hydration of the silicates when the cement is exposed to the atmosphere. Magnesia reacts

    with water in a manner similar to CaO, but only the crystalline form is deleteriously reactive so

    that unsoundness occurs. Calcium sulfate is the third compound liable to cause expansion

    through the formation of calcium sulfoaluminate (ettringite) from excess gypsum (not used up by

    C3A during setting). Le Chatelier's accelerated test is prescribed by BS EN 196-3: 1995 for

    detecting unsoundness due to free lime only. Essentially, the test is as follows. Cement paste of

    standard consistence is stored in water for 24 hours. The expansion is determined after increasing

    the temperature and boiling for 1 hour, followed by cooling to the original temperature. If the

    expansion exceeds a specified value, a further test is made after the cement has been spread and

    aerated for 7 days. At the end of this period, lime may have hydrated or carbonated, so that asecond expansion test should fall within 50 per cent of the original specified value. Cement

    which fails to satisfy at least one of these tests should not be used. In practice, unsoundness due

    to free lime is very rare. Magnesia is rarely present in large quantities in the raw materials used

    for making cement in the UK, but in the US this is not the case. For this reason, ASTM C 151-05

    specifies the autoclave test which is sensitive to both free magnesia and free lime. Here, a neat

    cement paste specimen of known length is cured in humid air for 24 hours and then heated by

    highpressure steam (2 MPa (295 psi)) for about 1 hour so that a temperature of 216 degrees

    celcius ( 420 F) is attained. After maintaining that temperature and pressure for a further 3

    hours, the autoclave is cooled so that the pressure falls within 1.5 hours and the specimen is

    cooled in water to 23degrees celcius (73 F) in 15 min. After a further 15 min, the length of the

    specimen is measured: the expansion due to autoclaving must not exceed 0.8 per cent of the

    original length. This accelerated test gives no more than a broad indication of the risk of long-

  • 8/13/2019 Soundness Cement

    2/4

    term expansion in practice. No test is available for the detection of unsoundness due to an excess

    of calcium sulfate, but its content can be easily determined by chemical analysis.

    Reference: Concrete technology 2nd

    edition, A.M. Neville

    ---------------------------------------------------------------------------------------------------------------------

    Unsoundness:

    Unsoundness in cement paste results from excessive volume change after setting. If there is any

    appreciable expansion, however slow, cracking and failure of the concrete will result.

    Unsoundness in cement is caused by the slow hydration of MgO or free lime, and by the

    reactions of gypsum with C3A (see Section 3.1). These expansive reactions take place very

    slowly, and so unsoundness will only appear after many months, or even years. Therefore, it is

    necessary to use some form of accelerated test, so that tendencies toward unsoundness can be

    detected as a quality control measure. Although a number of soundness tests have been

    developed over the years, only one is in common use. The autoclave expansion test (ASTM C

    151) will detect unsoundness due to both excess CaO and excess MgO, and is the test designated

    in ASTM C 150. A cement of normal consistency is molded and cured normally for 24 h. The

    specimens are then removed from the molds, measured, and placed in an autoclave. The

    temperature is raised so that the steam pressure in the autoclave reaches 2 MPa (295lb/in.2) in 45

    to 75 min; this pressure is maintained for 3 h, and the autoclave cooled so that the pressure is

    relieved in 1% h. The specimen is then cooled in water, to 23C in 15 min. Its length is again

    measured after a further 15 min; the expansion must be less than 0.80% if the cement is to be

    acceptable. These accelerated tests clearly do not simulate service conditions, and the amount of

    expansion may be affected by a number of factors, such as the fineness of the cement and the

    C3A content. Thus, these tests can only serve as a guide and cannot provide an absolute

    indication of a tendency toward unsoundness for given cement. However, they provide a

    valuable quality control measure, since changes in the expansion tests would indicate some

    anomaly in the cement manufacturing process.

    Reference: Mindess

    =====================================================================

  • 8/13/2019 Soundness Cement

    3/4

  • 8/13/2019 Soundness Cement

    4/4