Upload
sharat-chandra
View
220
Download
0
Embed Size (px)
Citation preview
7/27/2019 SMT-04-1001 Rev A1 (App 13-10 Calirbrating Sniffer)
1/2
b
ff lk Dc
ope
his application note addresses the calibration o snifer leak detectors.
BaKGrounD
common leak detection method utilizes a snifer leak detector to locate leaks in a part or system that is pressurized with
tracer gas (usually helium). While very useul in pinpointing the location o a leak, snifer detectors are not as reliable in
uantiying leaks compared to other methods. Furthermore, sensitivity is much less than with hard vacuum helium leaketection, and varies among diferent types o detectors. However, with proper calibration, the snifer leak detector can be
sed as an efective tool or both production and maintenance leak location applications.
Deription
nifer leak detectors draw small amount o air through the snifer probe by means o a pump and measure the concentrati
tracer gas in the drawn sample. Calibration o the snifer can give the operator condence by providing a relative compar
the displayed leak rate to an actual leak value, and by proving that the detector is sensitive enough to detect a particular
te. It can also be helpul to illustrate correct techniques or maximizing sensitivity and success in locating existing leaks (i
ow close the probe needs to be to the leak, at what speed the probe should pass over a leak in order to detect it, the dete
sponse time).
alibration o a snifer leak detector is acilitated by a calibrated leak standard, which is a device that produces a measured
ow o gas that simulates an actual leak. Normally the leak rate value o the calibrated leak standard is at or near the minim
esired sensitivity or reject level. To perorm a snifer calibration, the background tracer gas concentration should be measu
nd veried to be below the leak rate value o the calibrated leak. This is done by taking a reading o the ambient air. I am
acer gas contamination is suspected, samples should be taken in another location. I it is determined that the test area is
ontaminated with tracer gas, then the area should be purged prior to calibration. Following verication o the tracer gas
ackground, the snifer probe is inserted into the outlet o the leak standard, and let there until the reading rom the snife
abilizes. The displayed reading is compared to the leak rate value o the leak standard, rom which a quantiying relations
an be made or uture readings (the leak detector may or may not display a value with any proximity to the value o the
ctual leak that it is reading). Some instruments allow the user to enter in a calibration actor to that will adjust the display atch the reading rom the calibrated leak standard. Additionally, calibrating will determine whether the snifer is capable
etecting the size o leak simulated by the leak standard.
any snifer leak detectors display readings in units o concentration (parts per million, or example), whereas most leak
andards are specied by leak or ow rate (typically atmcc/sec). Concentration can be correlated to leak/ow rate i the
umping speed o the snifer line o the detector is known. To correlate parts per million (ppm) to atmcc/sec:
mcc/sc = m x 106 x
where S is the pumping speed o the detector in cc/sec
note #13-10
SMT-04-1001 Rev
7/27/2019 SMT-04-1001 Rev A1 (App 13-10 Calirbrating Sniffer)
2/2
Deription (ontinueD)
alibration will be most successul when the amount o gas rom the leak standard that enters the probe is maximized, whi
ot impeding normal air ow into the probe. For this reason, an appropriate outlet on the leak standard is important. An
utlet that allows leaked gas to escape beyond the probe or does not allow proper air ow will result in alse readings maki
he calibration less efective. Also, the snifer probe should be very near the actual gas exit point o the leak standard so tha
he tracer gas does not become too diluted by the surrounding air. Other actors that could negatively inuence a snifer
alibration are turbulent air surroundings (such as in ront o a an), or surrounding air that is heavily contaminated with the
acer gas. Manuacturer-recommended practices that are specied in the instrument user manual (such as zeroing the bas
ading) should also be observed.alibration should be perormed regularly during any leak detection process to veriy the integrity o the process. A
commend practice is to calibrate the instrument prior to testing a system or at the beginning o the production shit, then
eriying calibration when complete.
relateD proDut
CalMasterTM Leak Standards
Gas-Check Leak Detector
miniTracerTM Portable Snifer
Mass Spec Helium Leak Detectors
referene
Application Note 07-05 Helium Sning Leak Detection
CalMasterTM Calibrated Leak Standards Manual
ASTM Standard E 499: Standard Methods o Testing or Leaks Using the Mass Spectrometer Leak Detector in the Detecto
Probe Mode.