Radon Eye RD200 & RD200M Smart Radon Detector & Module Type : pulsed ion chamber 200cc First reliable data out : < 60min Data interval : 10min update (60min moving average) Sensitivity : 0.5cpm/pCi/l at 10pCi/l (30cph/pCi/l) Operating range : 10~40℃, RH < 90% Range : 0.1 ~ 99.99 pCi/l Precision : < 10% at 10pCi/l Accuracy : < ±10% (min. error <±0.5pCi/l ) Power : DC 12 ± 0.1V, 65mA (12V DC adapter) Size : RD200 : Φ80(mm) x 120(mm) , 240g RD200M : Φ60(mm) x 98(mm) Data communication : Bluetooth LE (Android/iOS) UART (RD200M) Data log : max 1year (1h step) Display : 0.96 inch OLED DESCRIPTIONS Applications RD200 is a real time smart radon detector for home owner which has the high sensitivity 0.5cpm/pCi/l, about 20~30 times more than conventional radon detector by FTLAB’s high stable circuit technology. Its first reliable data out time is below 60min from measurement start. Also the accuracy is <10% at 10pCi/l. (The accuracy and precision spec were tested by the KTL(Korea Testing Laboratory) administrated by KOREAN government) RD200M is a radon sensing module for OEM supply. RD200= RD200M+ display+ logger+ Bluetooth+ etc * NET certification from Republic of South Korea (2015.12) Ordering information contact point : [email protected] All products are made in KOREA RD200 RD200M indoor real time radon monitoring IoT Radon sensor Automatic ventilation system radon suppressed system Air purifier <Smart phone App> <Typical measurements data > FEATURES & SPECIPICATION of RD200 0 500 1000 1500 2000 2500 3000 0.0 0.5 1.0 1.5 2.0 ventilation SN15 SN20 SN18 SN25 SN27 SN28 Multi sample test during 48hour in Background pCi/L Time(min) www.radonftlab.com [email protected] V1.4

Smart Radon Detector & Moduleradonftlab.com/.../2015/data_sheet/datasheet_RD200_v1.4_eng.pdf · Theory of Operations Radon Measurement by pulsed ion chamber Fig1. Pulsed Ion Chamber

  • Upload
    vohanh

  • View
    221

  • Download
    0

Embed Size (px)

Citation preview

Radon Eye RD200 & RD200MSmart Radon Detector & Module

• Type : pulsed ion chamber 200cc

• First reliable data out : < 60min

• Data interval : 10min update (60min moving average)

• Sensitivity : 0.5cpm/pCi/l at 10pCi/l (30cph/pCi/l)

• Operating range : 10~40℃, RH < 90%

• Range : 0.1 ~ 99.99 pCi/l

• Precision : < 10% at 10pCi/l

• Accuracy : < ±10% (min. error <±0.5pCi/l )

• Power : DC 12 ± 0.1V, 65mA (12V DC adapter)

• Size : RD200 : Φ80(mm) x 120(mm) , 240g

RD200M : Φ60(mm) x 98(mm)

• Data communication : Bluetooth LE (Android/iOS)

UART (RD200M)

• Data log : max 1year (1h step)

• Display : 0.96 inch OLED

DESCRIPTIONS

Applications

RD200 is a real time smart radon detector for home owner

which has the high sensitivity 0.5cpm/pCi/l, about 20~30

times more than conventional radon detector by FTLAB’s high

stable circuit technology.

Its first reliable data out time is below 60min from

measurement start. Also the accuracy is <10% at 10pCi/l.

(The accuracy and precision spec were tested by the KTL(Korea

Testing Laboratory) administrated by KOREAN government)

RD200M is a radon sensing module for OEM supply.

RD200= RD200M+ display+ logger+ Bluetooth+ etc

* NET certification from Republic of South Korea (2015.12)

Ordering information

• contact point : [email protected]

• All products are made in KOREA

RD200 RD200M

• indoor real time radon monitoring• IoT Radon sensor• Automatic ventilation system• radon suppressed system• Air purifier

<Smart phone App>

<Typical measurements data >

FEATURES & SPECIPICATION of RD200

0 500 1000 1500 2000 2500 3000

0.0

0.5

1.0

1.5

2.0ventilation

SN15

SN20

SN18

SN25

SN27

SN28

Multi sample test during 48hour

in Background

pC

i/L

Time(min)

www.radonftlab.com [email protected]

V1.4

Theory of Operations

Radon Measurement by pulsed ion chamber

Fig1. Pulsed Ion Chamber theory

Fig 2. L75 x D60, 200cc class ion chamber

Figure 3. Improved detecting signal by FTLAB’s circuit technology

α-particle that are generated during the radioactive decay of

radon and radon progeny create the thousand of ± charge

particles by collision with air in Figure 1. These electric

charges signal is usually very weak and noisy, so it is difficult

to amplify the signal using a circuit having a high input

impedance. RD200 detects effectively these secondary

charges using the built-in 200cc class pulsed ionization

chamber in Figure 2 and a special amplifier circuit developed

by FTLAB’s own technology. The improved detection signal is

shown in Figure 3. Thus, from the statistical method based

on frequency, pulse counts and interval of these alpha decay,

it can be measured the average radon concentrations indoor.

www.radonftlab.com [email protected]

Compensation technic in RD200

Figure 4. Compensation by MCU using waveform analysis & sensor data

RD200 has been applied to a special signal waveform analysis technology. Based on the very stable

signal waveform, it can be introduced a algorithm to distinguish normal alpha signal waveform and

the noise waveform in internal MCU.

Also it was overcome the shortcomings of the ion-chamber system by utilizing a various smart sensor.

Waveform analysis by MCU & smart sensor

www.radonftlab.com [email protected]

Comparison of RD200 with other devices

RD200 is the first real time Radon detector in the low price market . The performance

of RD200, sensitivity or response time is closely up to the inspection/research devices.

However the Price of RD200 is nearly same with other conventional cheap devices on

public domain.

No longer need to wait a long time to find the indoor radon concentration. No longer

need to buy a expensive device to check the accurate value. You can just enjoy the

excellence performance of RD200 as a Best price.

The Best Price Performance in the world

www.radonftlab.com [email protected]

Reference test result of RD200 with RAD7

Experimental setup at radon safety lab of Yonsei University in KOREA (2015. 6. 9)

time temp (℃) humi (%) total Count Count CPM(30min) RAD7 data (pCl/L) 비고

12:00 26.0 42 0 0 0.0 0우라늄 원석가루 1그릇, 라돈가스 추가 삽입,

제습제 1개 배치, 측정시작

12:30 26.3 40.6 488 488 16.3 28.4±3.19

1:00 26.4 39.5 1024 536 17.9 32.7±3.41

1:30 26.7 38 1698 674 22.5 37.4±3.65

2:00 26.7 37.5 2166 468 15.6 39.2±3.72

2:30 26.7 37 2751 585 19.5 39.1±3.70

3:00 26.8 36.6 3319 568 18.9 37.8±3.66

3:30 27.0 36.2 3912 593 19.8 36.8±3.63 humidity growing

4:00 26.8 72.5 4475 563 18.8 38.5±3.71

4:30 25.9 99.3 5021 546 18.2 39.2±3.74

RD200 data RAD7 data

Negligible humidity effect (below error range)

stable results after 1hour

www.radonftlab.com [email protected]

Comparison of RD200 with RTM1688-2 at KRISS(KRISS : Korea Research Institute Standards and Science)

Radon source

1m3 test chamber

RTM1688-2

Experimental setup at Radon test chamber of KRISS (2015. 7. 29)

www.radonftlab.com [email protected]

RD200 multi-sample test

0 500 1000 1500 2000 2500 3000

0.0

0.5

1.0

1.5

2.0

2.5

3.0

3.5

4.0

Multi sample test during 48hour

in Background

SN27

SN26

SN25

SN20

SN28pC

i/L

Time(min)

0 500 1000 1500 2000 2500 3000

0.0

0.5

1.0

1.5

2.0ventilation

SN15

SN20

SN18

SN25

SN27

SN28

Multi sample test during 48hour

in Background

pC

i/L

Time(min)

www.radonftlab.com [email protected]

Block diagram of RD200M application in air purifier

MCU

DC powersupply RD200M

Radon sensor module

UART

DC 12V

Application device (ex : air purifier)

Operation sequence : After power ON, 1st data out at10min. And then data update continuously with 10min interval. (1hour moving average).

Application example of RD200M

Sta

rt

1h

2h

3h

www.radonftlab.com [email protected]

* size can be changed slightly by design

** RD200M size : : Φ60(mm ) x H:98(mm)

Size of RD200M

www.radonftlab.com [email protected]

Connector Pin map1 : TX (3.3V)2: RX (3.3V)3: 12V4: GND

Test Report by KTL (Korea Testing Laboratory)

www.radonftlab.com [email protected]

NET(New Excellent Technology) certification from Republic of South Korea

www.radonftlab.com [email protected]