Radon Eye RD200 & RD200MSmart Radon Detector & Module
• Type : pulsed ion chamber 200cc
• First reliable data out : < 60min
• Data interval : 10min update (60min moving average)
• Sensitivity : 0.5cpm/pCi/l at 10pCi/l (30cph/pCi/l)
• Operating range : 10~40℃, RH < 90%
• Range : 0.1 ~ 99.99 pCi/l
• Precision : < 10% at 10pCi/l
• Accuracy : < ±10% (min. error <±0.5pCi/l )
• Power : DC 12 ± 0.1V, 65mA (12V DC adapter)
• Size : RD200 : Φ80(mm) x 120(mm) , 240g
RD200M : Φ60(mm) x 98(mm)
• Data communication : Bluetooth LE (Android/iOS)
UART (RD200M)
• Data log : max 1year (1h step)
• Display : 0.96 inch OLED
DESCRIPTIONS
Applications
RD200 is a real time smart radon detector for home owner
which has the high sensitivity 0.5cpm/pCi/l, about 20~30
times more than conventional radon detector by FTLAB’s high
stable circuit technology.
Its first reliable data out time is below 60min from
measurement start. Also the accuracy is <10% at 10pCi/l.
(The accuracy and precision spec were tested by the KTL(Korea
Testing Laboratory) administrated by KOREAN government)
RD200M is a radon sensing module for OEM supply.
RD200= RD200M+ display+ logger+ Bluetooth+ etc
* NET certification from Republic of South Korea (2015.12)
Ordering information
• contact point : [email protected]
• All products are made in KOREA
RD200 RD200M
• indoor real time radon monitoring• IoT Radon sensor• Automatic ventilation system• radon suppressed system• Air purifier
<Smart phone App>
<Typical measurements data >
FEATURES & SPECIPICATION of RD200
0 500 1000 1500 2000 2500 3000
0.0
0.5
1.0
1.5
2.0ventilation
SN15
SN20
SN18
SN25
SN27
SN28
Multi sample test during 48hour
in Background
pC
i/L
Time(min)
www.radonftlab.com [email protected]
V1.4
Theory of Operations
Radon Measurement by pulsed ion chamber
Fig1. Pulsed Ion Chamber theory
Fig 2. L75 x D60, 200cc class ion chamber
Figure 3. Improved detecting signal by FTLAB’s circuit technology
α-particle that are generated during the radioactive decay of
radon and radon progeny create the thousand of ± charge
particles by collision with air in Figure 1. These electric
charges signal is usually very weak and noisy, so it is difficult
to amplify the signal using a circuit having a high input
impedance. RD200 detects effectively these secondary
charges using the built-in 200cc class pulsed ionization
chamber in Figure 2 and a special amplifier circuit developed
by FTLAB’s own technology. The improved detection signal is
shown in Figure 3. Thus, from the statistical method based
on frequency, pulse counts and interval of these alpha decay,
it can be measured the average radon concentrations indoor.
www.radonftlab.com [email protected]
Compensation technic in RD200
Figure 4. Compensation by MCU using waveform analysis & sensor data
RD200 has been applied to a special signal waveform analysis technology. Based on the very stable
signal waveform, it can be introduced a algorithm to distinguish normal alpha signal waveform and
the noise waveform in internal MCU.
Also it was overcome the shortcomings of the ion-chamber system by utilizing a various smart sensor.
Waveform analysis by MCU & smart sensor
www.radonftlab.com [email protected]
Comparison of RD200 with other devices
RD200 is the first real time Radon detector in the low price market . The performance
of RD200, sensitivity or response time is closely up to the inspection/research devices.
However the Price of RD200 is nearly same with other conventional cheap devices on
public domain.
No longer need to wait a long time to find the indoor radon concentration. No longer
need to buy a expensive device to check the accurate value. You can just enjoy the
excellence performance of RD200 as a Best price.
The Best Price Performance in the world
www.radonftlab.com [email protected]
Reference test result of RD200 with RAD7
Experimental setup at radon safety lab of Yonsei University in KOREA (2015. 6. 9)
time temp (℃) humi (%) total Count Count CPM(30min) RAD7 data (pCl/L) 비고
12:00 26.0 42 0 0 0.0 0우라늄 원석가루 1그릇, 라돈가스 추가 삽입,
제습제 1개 배치, 측정시작
12:30 26.3 40.6 488 488 16.3 28.4±3.19
1:00 26.4 39.5 1024 536 17.9 32.7±3.41
1:30 26.7 38 1698 674 22.5 37.4±3.65
2:00 26.7 37.5 2166 468 15.6 39.2±3.72
2:30 26.7 37 2751 585 19.5 39.1±3.70
3:00 26.8 36.6 3319 568 18.9 37.8±3.66
3:30 27.0 36.2 3912 593 19.8 36.8±3.63 humidity growing
4:00 26.8 72.5 4475 563 18.8 38.5±3.71
4:30 25.9 99.3 5021 546 18.2 39.2±3.74
RD200 data RAD7 data
Negligible humidity effect (below error range)
stable results after 1hour
www.radonftlab.com [email protected]
Comparison of RD200 with RTM1688-2 at KRISS(KRISS : Korea Research Institute Standards and Science)
Radon source
1m3 test chamber
RTM1688-2
Experimental setup at Radon test chamber of KRISS (2015. 7. 29)
www.radonftlab.com [email protected]
RD200 multi-sample test
0 500 1000 1500 2000 2500 3000
0.0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
4.0
Multi sample test during 48hour
in Background
SN27
SN26
SN25
SN20
SN28pC
i/L
Time(min)
0 500 1000 1500 2000 2500 3000
0.0
0.5
1.0
1.5
2.0ventilation
SN15
SN20
SN18
SN25
SN27
SN28
Multi sample test during 48hour
in Background
pC
i/L
Time(min)
www.radonftlab.com [email protected]
Block diagram of RD200M application in air purifier
MCU
DC powersupply RD200M
Radon sensor module
UART
DC 12V
Application device (ex : air purifier)
Operation sequence : After power ON, 1st data out at10min. And then data update continuously with 10min interval. (1hour moving average).
Application example of RD200M
Sta
rt
1h
2h
3h
www.radonftlab.com [email protected]
* size can be changed slightly by design
** RD200M size : : Φ60(mm ) x H:98(mm)
Size of RD200M
www.radonftlab.com [email protected]
Connector Pin map1 : TX (3.3V)2: RX (3.3V)3: 12V4: GND
NET(New Excellent Technology) certification from Republic of South Korea