Back to BasicsElectronics & Filters
In PSG Systems
Will Eckhardt, BS,RPSGT, CRT
2
Conflicts
• I work for Respironics, Inc.
• I do not believe this has any influence on the content of this presentation.
3
What Will Be Answered in this Presentation
• Where do EEG potentials originate?• We record various physiological signals what
means of processing do we utilize ?• What are the analog components of signal
processing?• What are the digital components of signal
processing?• An understanding some basic underlying
principles of the technology by looking at the analog signal output.
4
Pyramidal Neuron
EEG is derived from thousands of synchronized pyramidal cell postsynaptic potentials.
Volume Conduction is the process of current flow through the tissues between the electrical generator and the electrode.
5
Amplitude of the Recorded Potentials
• Depends on:– Intensity of the Electrical Source
– Distance of Source to Recording Electrodes
– Spatial Orientation• Orientation of the electrical generator to the electrode
– Electrical Resistance• Scalp-electrode interface• Recording electrodes
6
7
Electrode Placement
• Exploring electrode
• Reference electrode
• Referential derivation
• Bipolar derivation
8
Components of Polysomnography Equipment
and their relationship to signal processing.
9
Path of Signals from the Patient to the Tracing
10
Recording Electrodes
• Electrode types• Gold, Silver Chloride, Tin, Platinum
• Placement• 10-20 system
• Application• Electrode impedance
11
Headbox• The headbox is an intermediate
connection conducting the electrical signal from the electrode to the amplifier.
• Enables multi-sensor output to one low impedance cable.
• The cable then connects to:– electrode selector panel (analog)– amplifier
• Note: the headbox may include the systems
amplifiers.
12
Wave Form - Input & Output
• Voltage input to output using a known input.– 50µV– Square wave
• Gain setting (amount of amplification)• Filters (high, low, notch)• Sensitivity (amplifier output voltage to pen
deflection)
• Verify signal integrity– Like channels give the same output
13
Types of Calibration
• Machine Cal – All channels respond the same
• Montage Cal – Channels respond as set
• Bio-Cal – Channels respond appropriately for the
maneuver/sensor– Should be performed at the beginning & end
of the study.
14
Amplifiers
• Function: Signal modification– Amplification (↑ voltage of potential to drive pens or
card)– Discrimination (reveal differences in inputs 1 & 2)
• Inputs: Input 1 (G1) – Input 2 (G2)G1 exploring G2 reference
• Types: AC & DC Amplifiers• Digital and Analog Amplifiers
15
Differential Amplifier
• Amplifies only the difference in voltage between two inputs.
• Does not amplify equal voltages.
• Has the ability to reject external interference.
• The amount that an amplifier will reject a common mode signal is called “common mode rejection ratio (CMRR).
• 10,000/1- Common mode/output voltageIdentical signals
16
Signal Polarity
• Designed such that (-) voltage to input G1 of a differential amplifier [with respect to G2] will result in a upward deflection.
• When G1 becomes more (+) than G2 this will result in a downward deflection.
17
AC Amplifier
• This type of amplifier is used for signals that vary frequently from positive to negative.
Examples: EEG, EMG, EOG
18
DC Amplifier
• This type of amplifier is used for signals that slowly increase or decrease.
• Use DC amplifiers: – Used for signals i.e. oximetry, respiration, temp, PH.– Used to generate an output that is linear and can
scale (i.e. 0-100%).
19
Display Features
• Controls that change the output from the amplifiers. – Amplification– Filters– Sampling rate– Monitor settings– Analog features
20
Amplification
• Increases voltage difference between inputs.
• Small potentials° Voltage to drive pens (not often anymore)° Voltage for analog to digital converter (ADC)
• Boost voltage of biological signal µV to V• Characterized by:
° Gain° Sensitivity
21
Gain
• Gain is a ratio: amplifier output voltage to input voltage.– Output of 10 V for input 10 µV = gain of 1
million
• Gain is the degree of magnification of the amplifier signal.
• We do not directly measured gain.
22
Sensitivity• Relates to Ohm’s Law I=E/R or in PSG D=V/S
– Current (I) = Pen Deflection (D
– Voltage (E) = Input voltage (V)
– Resistance (R) = Sensitivity (S)
• Sensitivity is a ratio: of input voltage to deflection produced. Sensitivity = Voltage/Deflection
• Sensitivity is a measure of the response of an instrument to an incoming signal.– A sensitive instrument will produce a large response
to a small incoming signal.– A low sensitivity setting produces a large signal
change (deflection). • µV/mm (voltage per unit of pen deflection)• Exam may express as µV/cm
23
Video Sensitivity
Sensitivity 2 2008_05_04_10_30_38.avi
24
Computing the Voltage
• Measure the amplitude of the wave– Peak to trough
• The vertical distance of the rise or deflection (in mm) is multiplied by the sensitivity of the amplifier.– Analog sensitivity is on a dial
• Gives us the voltage of the wave.
25
Sensitivity Settings
•The higher the sensitivity setting the smaller the wave (deflection).•V/S=D •The sensitivity setting affects how the wave is displayed, it does not change the actual voltage.
26
27
Sensitivity Equations
• S=V/D 5=50µV/10mm
• V=SxD 50µV=5x10mm
• D=V/S 10mm=50µV/5
Settings:Adult 5-7 µV/mmChildren 10 µV/mm
28
Filters
• Filters give us the ability to focus on only the signal frequencies that we wish to see.• Bandwidth
• They attenuate unwanted signals.• Each channel can be optimized, using
filters, to allow only signals in the desired frequency range.
29
Bandwidth
30
Low Frequency Filter (LFF)
• The low frequency filter is also known as the “High Pass” filter.
• This means that the LFF allows higher frequencies to pass unchanged while lower frequencies are attenuated.
31
Low Frequency Filtering on EEG Channels
Note: Increasing the filter setting attenuates slow waves.
32
Low Frequency Filters
Examples of typical LFF settings by derivation:
• EEG = 0.3 Hz (C4-C3-O1-O2)• EOG = 0.3 Hz (LOC-ROC)• EMG = 10 Hz (Chin, Leg, Intercostals)• EKG = 0.3 Hz
The AASM Manual for the Scoring of Sleep and Associated Events Rules, Terminology and Technical Specifications, 2007
33
High Frequency Filter (HFF)
• The high frequency filter is also known as the “Low Pass” filter.
• This filter lets slower waves through and attenuates higher frequencies.
• Examples– HFF eliminates muscle artifact or external electrical
artifact in the EEG channels.– Caution: This filter can also attenuate desired high
frequencies such as arousals
34
High Frequency Filters
Note: Increasing the filter setting allows more fast waves to be seen.
35
High Frequency Filters
• EEG = 35 Hz
• E0G = 35 Hz
• EMG = 100 Hz
• ECG = 70 Hz
The AASM Manual for the Scoring of Sleep and Associated Events Rules, Terminology and Technical Specifications, 2007
Examples of typical HFF settings by derivation:
36
• The time constant is a measure of how a signal is displayed and is determined by the voltage of the signal and the low and high frequency filters used. – Time constant = time (seconds) for the signal
to change a specified amount. – Two time constants come into play, the rise
time constant and the fall time constant.
Time Constants
37
Time Constants
• Fall time constant = the time it takes (in seconds) for a square wave to decay to 37% of its maximum amplitude.– The fall time constant is usually what the term “time
constant” refers to in polysomnography.– The LFF determines the fall time constant. – ↑ the LFF setting will ↓ the time constant.
• Rise time constant = the time it takes (in seconds) for the square wave to reach 63% of its maximum amplitude. – The HFF determines the rise time constant.
38
Time Constants
Time Constant
Fall time constant = time for the signal to decay from its peak to 37%
39
Video Time Constant
Time Constant 3 2008_05_04_14_40_05.avi
40
LFF - Fall Time Constant
LFF (HZ) TC (sec)
0.1
0.3
1
5
1
0.4
0.12
0.05
41
60 Hertz / Notch Filter
• 60 Hz artifact is a high frequency artifact caused by:– Poor impedance – Interference from surrounding electrical
equipment – Poor application of electrodes
• Notch filters attenuate specific frequencies.• Do not use unless you must!
42
Sampling RateDigital Systems
• The shape of the wave form on the display is determined by how frequently the signal is sampled. – The higher the sampling rate, the more frequently the
signal is sampled. – The more frequent the sampling, the more accurate
the shape. • The minimum acceptable sampling rate is 2.5
times greater than the highest high frequency filter.
• Nyquist theorem: – 2x the fastest freq. = The Nyquist Rate
43
sampling rate
Low sampling rate
Input signalLow sampling rate
High sampling rate
Samplling Rates
Signal Desirable Minimal
EEG 500 Hz 200 Hz
EOG 500 Hz 200 Hz
EMG 500 Hz 200 Hz
ECG 500 Hz 200 Hz
Airflow/Pr 100 Hz 25 Hz
Oximetry 25 Hz 10 Hz
Effort 100 Hz 25 Hz
44
45
Waveform DisplayDigital Systems
• The resolution capabilities:– Computer -bits– Monitor – Resolution
• bits – refer to the digital amplitude resolution by the ADC. – E.g. 8-bit ADC has 28 or 256 amplitude levels
• The resolution of the monitor is expressed by the number of pixels, or points of light.– The higher the number of pixels, the greater the
resolution of the monitor. – The number of pixels is stated for the horizontal and
vertical axis of the monitor (must be >:1600 x 1200). The AASM Manual for the Scoring of Sleep and Associated Events Rules, Terminology and Technical Specifications, 2007
46
Dynamic Range
• Analog & Digital systems
47
Troubleshooting from the Patient to the Tracing
48
Questions?
49
References• EEG Primer Basic Principles of Digital and Analog
EEG, B Fisch• Fundamentals of EEG Technology Basic Concepts
and Methods, F Tyner, J Knott• Principles and Practice of Sleep Medicine, M Kryger,
T Roth, W Dement• Sleep Disorders Medicine Basic Science, Technical
Considerations, and Clinical Aspects, S Chokroverty• Sleep Medicine, T Lee-Chiong, M Sateia, M
Carskadon• The AASM Manual for the Scoring of Sleep and
Associated Events - Rules, Terminology and Technical Specifications, 2007