Upload
marcos-loredo
View
217
Download
0
Embed Size (px)
Citation preview
8/13/2019 Single+Laboratory+Validation
1/39
8/13/2019 Single+Laboratory+Validation
2/39
Definitions
Validation is the processof demonstratingor confirming the performancecharacteristicsof a method of analysis.
A process of evaluating methodperformance and demonstrating that itmeets a particular requirement.
Validation applies to a specific operator,laboratory, and equipmentutilizing themethod over a reasonable concentration
rangeand period of time.
8/13/2019 Single+Laboratory+Validation
3/39
Why Method Validation?
To minimize analytical and instrumental errors
To give reliable and reproducible results in
accordance with the given specifications of the
test method
To ensure the quality of the test results
To meet accreditation requirement
Objective evidence for defense againstchallenges
To be assured of the correctness of results
8/13/2019 Single+Laboratory+Validation
4/39
When Method Validation Not
Required? Standard methods on condition that
used within their scope of applicability(e.g. matrices, ranges, etc)
without modifications (including QA planand reporting)
Otherwise, required
8/13/2019 Single+Laboratory+Validation
5/39
Ask yourself
What analytes should be detected?
What are the expected concentration levels? What are the sample matrices? Are there interfering substances expected, and, if so, should they be
detected and quantified?
Are there any specific legislative or regulatory requirements? Should information be qualitative or quantitative? What are the required detection and quantitation limits? What is the expected concentration range? What precision and accuracy is expected? How robust should the method be? Which type of equipment should be used? Is the method for one specific
instrument, or should it be used by all instruments of the same type? Will the method be used in one specific laboratory or should it be
applicable in all laboratories at one side or around the globe?
What skills do the anticipated users of the method have?
Before start to validate a method
8/13/2019 Single+Laboratory+Validation
6/39
Parameters for Method Validation
Accuracy
Precision
Specificity Limit of detection
Limit of quantitation
Linearity and range Ruggedness
Robustness
8/13/2019 Single+Laboratory+Validation
7/39
Strategy for Method Validation
1. Develop a validation protocol, an operatingprocedure or a validation master plan for thevalidation.
2. For a specific validation project define owners andresponsibilities.
3. Develop a validation project plan.
4. Define the application, purpose and scope of themethod.
5. Define the performance parameters and acceptance
criteria.6. Define validation experiments.
Source: LabCompliance (2007). Validation of Analytical Methods and Procedures:
Tutorial. http://www.labcompliance.com/tutorial/methods/default.aspx?sm=d_d
8/13/2019 Single+Laboratory+Validation
8/39
8/13/2019 Single+Laboratory+Validation
9/39
Strategy for Method Validation
Major
compounds
Major
compounds
Traces Traces
quantitative qualitative quantitative qualitative
Limit of
detection
Limit of
quantitation
Linearity Range Precision Accuracy Specificity Ruggedness
8/13/2019 Single+Laboratory+Validation
10/39
Method validation tools
Linear-data plotter can be used with data from a linearity experiment to
assess the reportable range of a method.
provides a plot of the average of a group of replicatetest results on the y-axis versus the assigned value(in % or concentration units) on the x-axis.
SD Calculator
can be with data from a replication experiment tocalculate the mean, standard deviation (SD orsmeas), and coefficient of variation (CV).
a histogram display of the data is also available.
8/13/2019 Single+Laboratory+Validation
11/39
Method validation tools
Paired-data Calculator
can be used with data from a comparison of methodsexperiment to calculate linear regression statistics(slope, y-intercept, and standard deviation about theregression line, sy/x), and the correlation coefficient(r, Pearson product moment correlation coefficient); t-test statistics (average difference between twomethods or biasmeas; SDdiff, standard deviation ofthe differences between the two methods).
can also be used to provide a "comparison plot" thatshows the test method results on the y-axis versusthe comparative method results on the x-axis, as wellas a "difference plot" that displays the differencebetween the test minus comparative results on the y-axis versus the comparative method result on the x-axis.
8/13/2019 Single+Laboratory+Validation
12/39
Sampling
Sample preparation
Analysis
Calibration
Data evaluation
Reporting
8/13/2019 Single+Laboratory+Validation
13/39
Sampling
Analysis starts with sampling
For trace analysis: sampling becomes amajor source of error
Differs from matrix to matrix
8/13/2019 Single+Laboratory+Validation
14/39
Sampling Method
When, where and howto collect samples
Sample transportationto laboratory
Samplingequipment Sample containers
Sample-treatment procedures(drying,
mixing, etc. prior to measurements) Sub-sampling procedures
Storageduring sampling
8/13/2019 Single+Laboratory+Validation
15/39
Sampling Method
To obtain a representative sample
Types of sampling
car or bin sampling
stratified sampling
random sampling
2 steps sampling interval sampling
8/13/2019 Single+Laboratory+Validation
16/39
Problems during sampling and storage
of samples
Losses from volatilization
Decomposition by means of:
temperature
UV irradiation
microbial activity chemical reactions with oxygen, samplecontainer, etc.
8/13/2019 Single+Laboratory+Validation
17/39
ISTA Validation Process
ISTA method validation is a five-stepprocess:
1. Test method selection and/or development.
2. Validation through either multi-laboratorycharacterization of the test method performance, peerverification of the test method, or verification ofperformance claims for the test method.
3. Review of data.
4. Approval of the test method by the relevant ISTATechnical Committee, publication in ISTA MethodValidation Reports and preparation of a Rules proposal
for the test method.5. Final acceptance by the ISTA voting members andpublication of the test method in the ISTA Rules.
Source: Hampton, J. (2005), ISTA Method Validation, Issues of Technical Common Interest,
Seed Testing International No. 130, p. 22, October 2005
http://www.seedtest.org/upload/cms/user/130_p22.pdf
8/13/2019 Single+Laboratory+Validation
18/39
Accuracy
To determine the closeness of the testresults obtained to the true value of the
standard used
To measure the systemic error of the
analysis
8/13/2019 Single+Laboratory+Validation
19/39
Accuracy
Method used:
Purchase standardfrom the supplier witha known tolerance given or standard
recommended by a particular reference
method. Working standard can be used as a
standard addition(spiked analysis).
Perform the method with more than 2determinationsat low, middle and high
concentrations.
8/13/2019 Single+Laboratory+Validation
20/39
Accuracy
Method used:
For methods using an instrument with a knownstandard calibration curve, selection of the rangecan be made.
For methods without the above, the range canbe selected from a knowledge of theconcentration of samples analyzed.
Based on the results obtained, calculate theaverage of the standard values obtained andcompare them with recommended values(Table1) or
Compare the result obtained with certifiedreference material(CRM).
8/13/2019 Single+Laboratory+Validation
21/39
Accuracy
Analyte (%) Unit Mean Recovery (%)
100 100% 98-102
10 10% 98-102
1 1% 97-103
0.1 0.1% 95-105
0.01 100 ppm 90-107
0.001 10 ppm 80-110
0.0001 1 ppm 80-110
0.00001 100 ppb 80-110
0.000001 10 ppb 60-115
0.0000001 1 ppb 40-120
Table 1: Acceptable Recovery Percentages
Source: AOAC (2002). AOAC Requirements for Single Laboratory Validation of
Chemical Methods. DRAFT 2002-11-07,\AOACI\eCam\Single-
Lab_Validation_47.doc. http://www.aoac.org/Ag_Materials/additives/aoac_slv.pdf.
http://www.aoac.org/Ag_Materials/additives/aoac_slv.pdfhttp://www.aoac.org/Ag_Materials/additives/aoac_slv.pdf8/13/2019 Single+Laboratory+Validation
22/39
Precision
To determine the degree of agreementwithinthe test results for a particular sample.
This is a measurement of the random errorsofan analysis.
3 types of precision measurement(1) Repeatability(one single operator, single laboratory,
short time span)
(2) Intermediate precision(internal reproducibility:between operators, single laboratory)
(3) Reproducibility (proficiency testing/collaborative
studies between laboratories)
8/13/2019 Single+Laboratory+Validation
23/39
Precision
Repeatability (one single operator, single laboratory,
short time span).
Refers to the degree of agreement of results whenconditions are maintained as constant as possible.
Select one sample for every matrix and performreplicationfor each matrix within the same day.
If resources permit, run the same sample for 3different days (within 7 days).
Calculate the standard deviation (SD) and relativestandard deviation(RSD) for each matrix.
Compare the precision results with the RSD% inTable 2.
8/13/2019 Single+Laboratory+Validation
24/39
Precision
Intermediate precision (internal reproducibility: between
operators, single laboratory). To evaluate the degree of agreement between different
operatorsfor a particular sample.
Measure the random errorsinherent when different
analysts perform the same analysis with the samesample.
Use 2-tailed F-test to determine whether there is anysignificant difference in the results between the
precisions of two operators. Calculate the sample variance for each operator.
Calculate the precision for each operator according tothe types of samples and compare the RSD(%).
8/13/2019 Single+Laboratory+Validation
25/39
Precision
Reproducibility (proficiency testing/collaborative
studies between laboratories). Only applicable when resourcesare available. The blind sample(s) will be analyzed at least
three timesby using the routine method and/orother standard methods.
Calculate the average value of the replicatesand SDif more than three replicates areperformed.
Compare the resultsobtained from otherlaboratoriesin terms of the z-score, mean valeuand median once these results are madeavailable.
8/13/2019 Single+Laboratory+Validation
26/39
Analyte (%) Unit RSDr(%) RSDR(%)
100 100% 1 2
10 10% 1.5 3
1 1% 2 4
0.1 0.1% 3 6
0.01 100 ppm 4 8
0.001 10 ppm (g/g) 6 11
0.0001 1 ppm 8 16
0.000001 10 ppb (g/kg) 15 32
Table 1: Acceptable Recovery Percentages
AOAC (2002). AOAC Requirements for Single Laboratory Validation of Chemical
Methods. DRAFT 2002-11-07,\AOACI\eCam\Single-Lab_Validation_47.doc.
http://www.aoac.org/Ag_Materials/additives/aoac_slv.pdf.
Precision
8/13/2019 Single+Laboratory+Validation
27/39
The value of RSD% can be calculatedfrom HORRAT formula:
RSDr= C0.15(Repeatability)
RSDR= 2C0.15
(Reproducibility)
HORRAT = RSD (found)/RSD (calculated)
Acceptable values for this ratio aretypically 0.5 to 2.
Precision
8/13/2019 Single+Laboratory+Validation
28/39
Limit of Detection
It is the smallest amountor concentration of an analytethat can be estimatedwith acceptable reliability.
An alternative definition of the limit of detection and limitof determination is based upon the variability of the blank.
The blank value plus three times the standard deviationof
the blank is taken as the detection limit and the blankvalue plus 10 times the standard deviationof the blank istaken as the determination limit.
The detection limit is only useful for control of undesirableimpuritiesthat are specified as not more than a specifiedlow level and for low-level contaminants.
Limits of detection and determination are unnecessary forcomposition specificationsalthough the statistical problemof whether or not a limit is violated is the same near zero
as it is at a finite value.
8/13/2019 Single+Laboratory+Validation
29/39
Instrument Detection Limit
To determine the lowest concentrationof analytein a sample that can be detected. It is important for trace analysis. Perform several determinations and calculate
the SD of the blank LOD = x + 3 SD(x = average blank reading) or Perform several determinations at the lowest
acceptable concentration using a standard
sample (fortified blank sample) and calculate theSD.
LOD = x + 3 SD (x = average sample blankreading)
8/13/2019 Single+Laboratory+Validation
30/39
Method Detection Limit
It is based on a sample, which have gone through theentire sample preparationprior to analysis.
Method detection limit is approximately 4 instrumentdetection limit.
Perform several determinationseach at low, middle andhigh range concentrations using a sample.
Calculate SD at each limit.
Plot SD against concentration of the analytes.
Extrapolate the graph until the y-axis is reached. At zero concentration, record the SD0.
Method detection limit = 3 SD0 It is unique for a particle sample matrix.
8/13/2019 Single+Laboratory+Validation
31/39
Signal to Noise Ratio
Source: LabCompliance (2007). Validation of Analytical Methods and Procedures:Tutorial. http://www.labcompliance.com/tutorial/methods/default.aspx?sm=d_d
8/13/2019 Single+Laboratory+Validation
32/39
Source: International Conference on Harmonization (ICH) of Technical
Requirements for the Registration of Pharmaceuticals for Human Use,
Validation of analytical procedures: Methodology, adopted in 1996, Geneva
8/13/2019 Single+Laboratory+Validation
33/39
Specificity/Selectivity
The terms selectivity and specificity are oftenused interchangeably.
To determine the purity of the peak areain
chromatogram. Difficult to ascertain whether the peaks within a
sample chromatogram are pureor consist ofmore than one compound.
The analyst should know how many compoundsare in the sample or whether procedures fordetecting impure peaks should be used.
8/13/2019 Single+Laboratory+Validation
34/39
Examples of pure and impure HPLC peaks
Source: LabCompliance (2007). Validation of Analytical Methods and Procedures:Tutorial. http://www.labcompliance.com/tutorial/methods/default.aspx?sm=d_d
8/13/2019 Single+Laboratory+Validation
35/39
Linearity
To determine the degree of varianceabout thecalibration curve
Perform several determinations using astandard sample at low, middle and high range.
Calculate the SD of each concentration group.
Calculate the best-fit line of the calibration curveand the correlation coefficient(r2) of the curve.
Use the F-test to determine the significantdifference in variance of the curve for each
concentration group.
8/13/2019 Single+Laboratory+Validation
36/39
Ruggedness
Ruggedness is defined by U.S. Pharmacoepia as thedegree of reproducibility of resultsobtained under avariety of conditions, such as different laboratories,analysts, instruments, environmental conditions,
operators and materials. Ruggedness is a measure of reproducibilityof test resultsunder normal, expected operational conditions fromlaboratory to laboratory and from analyst to analyst.
Ruggedness is determined by the analysis of aliquotsfrom homogeneous lots in different laboratories.
Refer to the examples of Ruggedness Trial in the AOACRequirements for Single Laboratory Validation ofChemical Methods (DRAFT 2002-11-07).
8/13/2019 Single+Laboratory+Validation
37/39
Robustness Robustness examine the effect that operational
parametershave on the analysis results. These parameters are pH, flow rate, column
temperature, injection volume, detection wavelength ormobile phase composition.
If the influence of the parameter is within a previouslyspecified tolerance, the parameter is said to be withinthe methods robustness range.
Obtaining dataon these effects helps to assesswhether a method needs to be revalidatedwhen one
or more parameters are changed, for example, tocompensate for column performance over time.
It is recommended to consider the evaluation of amethods robustness during the development phase,and any results that are critical for the method should
be documented.
8/13/2019 Single+Laboratory+Validation
38/39
Sources
ISTA guide:
http://www.seedtest.org/upload/cms/user/130_p22.pdf
EURACHEM guide:http://www.farmacia.ufmg.br/lato/eurachem%20guide%
20validation%20(1).pdf
AOAC guide:http://www.aoac.org/Ag_Materials/additives/aoac_slv.p
df
Other guide:http://www.labcompliance.de/documents/FDA/FDA-
Others/Laboratory/f-505-method-validation-draft.pdf
8/13/2019 Single+Laboratory+Validation
39/39