DESCRIPTION FEATURES SIGNAL CONDITIONERS MULTISENSOR INPUT configurable Basic Series BS-1500 126.48 The BS-1500 series are configurable converters with 0...4-20 mA selectable analog output, DIN guide mounting. Thermocouple, RTD, mV, mA input can be selected. These low price BS-1500 transmitters are developed to cover the mass-market segment for signal conditioners in industrial processes. FULLY CONFIGURABLE SERIES PROGRAMMABLE USING BS- Win KIT T/C, RTD, mV, mA USER-SELECTABLE INPUT ALL RANGES LINEARIZED NO RECALIBRATION FOR INPUT CHANGES OPTIONAL V/A ac/dc, RMS, Hz, Ohms DIRECT INPUT CONFIGURABLE INSULATED ANALOG OUTPUT SELF-CALIBRATION BY PC DIN GUIDE MOUNTING 24 Vdc AUX. POWER SUPPLY AS STANDARD LOW COST SERIES The BS-1500 is a new generation of configurable signal conditioners developed to cover the low-cost transmitter niche placed under our LS-3000 advanced series. This market sector achieves with this series the best available quality/price relationship without renouncing to foremost specifications of medium range models. The available configurable functions are as follows: Input type, units, decimal point, type of output, etc. Optionally, can be measuring electric variables as Voltage or Current ac/dc or RMS values, Ohms, Frequency Hz, etc. The analog output are generated by a state-of-the-art D/A converter, that includes 1000 V galvanic insulation and selectable direct or reverse output. Both 0-20 mA or 4-20 mA range types can be selected. The BS-1500 series can be configured through a dedicated port and a proprietary interface using a simple program- tool from PC (under Windows ® environment) getting the configuring task on BS-1500 instruments be easily and quickly done. The safety functions in BS-1500 series have taken a large design time of our engineers. In this way this new series includes an improved Watchdog function that joined to a compact design with smaller power consumption it offers a wide margin of security as result. In what concerns to transmitting security it provides a wide range of sophisticated functions like digital filters, etc. assuring the highest grade of security available. APPLICATIONS: Analog transmission of analog variables in industrial processes. Allowing direct input for temperature sensors and all kind of signal transmitters in applications as: - data acquisition - PLC data input - unity re-range - galvanic isulation DIN guide

SIGNAL CONDITIONERS BS-1500 MULTISENSOR INPUT configurable Basic Seriesdesin.es/pdf/HT-BS-1500-en-0231-H126-48-3.pdf · 2017-05-15 · MULTISENSOR INPUT configurable Basic Series

  • Upload
    buitu

  • View
    215

  • Download
    1

Embed Size (px)

Citation preview

Page 1: SIGNAL CONDITIONERS BS-1500 MULTISENSOR INPUT configurable Basic Seriesdesin.es/pdf/HT-BS-1500-en-0231-H126-48-3.pdf · 2017-05-15 · MULTISENSOR INPUT configurable Basic Series

DESCRIPTION

FEATURES

SIGNAL CONDITIONERSMULTISENSOR INPUTconfigurable Basic Series

BS-1500

126.48

The BS-1500 series are configurable converters with 0...4-20 mAselectable analog output, DIN guide mounting.Thermocouple, RTD, mV, mA input can be selected.These low price BS-1500 transmitters are developed to coverthe mass-market segment for signal conditioners in industrialprocesses.

FULLY CONFIGURABLE SERIES PROGRAMMABLE USING BS-Win KIT T/C, RTD, mV, mA USER-SELECTABLE INPUT ALL RANGES LINEARIZED NO RECALIBRATION FOR INPUT CHANGES OPTIONAL V/A ac/dc, RMS, Hz, Ohms DIRECT INPUT CONFIGURABLE INSULATED ANALOG OUTPUT SELF-CALIBRATION BY PC DIN GUIDE MOUNTING 24 Vdc AUX. POWER SUPPLY AS STANDARD LOW COST SERIES

The BS-1500 is a new generation of configurable signalconditioners developed to cover the low-cost transmitterniche placed under our LS-3000 advanced series.

This market sector achieves with this series the bestavailable quality/price relationship without renouncingto foremost specifications of medium range models.

The available configurable functions are as follows:

Input type, units, decimal point, type of output, etc.

Optionally, can be measuring electric variables as Voltageor Current ac/dc or RMS values, Ohms, Frequency Hz,etc.

The analog output are generated by a state-of-the-artD/A converter, that includes 1000 V galvanic insulationand selectable direct or reverse output.

Both 0-20 mA or 4-20 mA range types can be selected.

The BS-1500 series can be configured through a dedicatedport and a proprietary interface using a simple program-tool from PC (under Windows® environment) getting theconfiguring task on BS-1500 instruments be easily andquickly done.

The safety functions in BS-1500 series have taken a largedesign time of our engineers. In this way this new seriesincludes an improved Watchdog function that joined to acompact design with smaller power consumption it offers awide margin of security as result. In what concerns totransmitting security it provides a wide range of sophisticatedfunctions like digital filters, etc. assuring the highest grade ofsecurity available.

APPLICATIONS: Analog transmission of analog variables in industrial

processes. Allowing direct input for temperature sensorsand all kind of signal transmitters in applications as:

- data acquisition

- PLC data input

- unity re-range

- galvanic isulation

DIN guide

Page 2: SIGNAL CONDITIONERS BS-1500 MULTISENSOR INPUT configurable Basic Seriesdesin.es/pdf/HT-BS-1500-en-0231-H126-48-3.pdf · 2017-05-15 · MULTISENSOR INPUT configurable Basic Series

SPECIFICATIONS

The

co

nte

nts

of t

his

sh

ee

t a

re s

ub

jec

t to

ch

an

ge

wit

ho

ut

no

tic

e

WIRING DIAGRAM

HOW TO ORDER

Av. Frederic Rahola, 49 - 08032 BARCELONA (Spain)Tel. (+34) 93 358 6011 - Fax (+34) 93 357 6850*

e-mail:[email protected] - http://www.desin.com

DIMENSIONS

0231

H12

6.48

-3

ANALOG INPUT:- Type: ............................................................... T/C, RTD, mV, mA- Input impedance: ............................................ > 0,3 MΩ for mV

& external Shunt 3.74 Ω for mA- T/C Cold Junction Compensation : ....... by sensor in terminals- CJC range: .......................................... 30 to 140 °F 0 to 60 °C- Standards: ................................. IEC584 (T/C) & IEC751 (Pt 100)- Input types and configurable maximum ranges:

T/C T -300/750 °F -200/400 °CT/C E -300/1800 °F -200/1000 °CT/C J -300/2100 °F -200/1200 °CT/C K -300/2300 °F -200/1300 °CT/C N 0/2300 °F 0/1300 °CT/C R 0/3000 °F 0/1700 °CT/C S 0/3000 °F 0/1700 °CT/C B 1200/3200 °F 600/1800 °CPt 100 -199.9/999.9 °F -199,9/800,0 °CVoltage: 0 / 75 mVCurrent: 0...4-20 mA (shunt 3.74 Ω not-included)

- Accuracy at 25 °C: ......................................................... ± 0.1 %- A/D converter resolution: ...................................... 40000 points- Sampling rate: ................................................................. 500 ms

ANALOG OUTPUT:- Type: ................................ mA (V w/ external 500 ohms shunt)- Function: ................................................. measuring transmitter- Mínimum configurable range : ................. 2 % of input range- Output range: ............................................. 0-20 mA & 4-20 mA- Current loop max. load: ............................................ 500 Ohms- Accuracy at 25 °C: .......................................................... ± 0.1 %- D/A converter resolution: ................................................. 12 bits- Update time: .................................................................... 250 ms- Insulation: .......................................................................... 1000 V

CONFIGURABLE FUNCTIONS:- Input type, Range, Units, Decimal point, Bias

TRANSMITTER POWER SUPPLY:- Type: ....................................................... Linear (non stabilized)- Nominal output: ............................................................... 24 Vdc

- Current output: ........................................................ 25 mA max.

ENVIRONMENTAL LIMITS:- Operating ambient temperature: ............................ 0 to 50 °C- Storage temperature: ............................................. -20 to 70 °C- Relative Humidity: ................. máx. 85% RH (non-condensing)

CE STANDARDS:- EN 50081 Emission, EN 50082 Immunity, EN 61010 Electric safe

POWER:- Nominal operating voltage: ................. 220 Vac and 110 Vac

.....................................................(optional 24 and 48 Vac/dc)- Tolerance: ........................................................................ ± 15 %- Frequency: ................................................................. 48 to 60 Hz- Power consumption: ............................................. max. 2.5 VA

Factory configured version:BS-1500 Signal conditioner 4-20 mA output

Input: T/C or RTD

Note: This model is only configured by BS-WIN software tool

This version is user configurable nevertheless it can be supplied, underrequest, configured in factory.In this case specify:

- Application- Input, range and engineering units

126.

48BS-1500

Symetric DIN guide mounting: EN 50022 (DIN 46277-3Weight: 11.3 oz (320 g); with packing 13.4 oz (380 g)

123

89

Power supply 220 Vac and 110 Vac 48/60 Hz.Optional 24 and 48 Vac or Vdc

Auxiliary power supply24 Vdc, 25 mA max.

DO NOT CONNECT. Internal use terminals

0...4-20 mA, 0...1/5 ..2/10 V, Analog OutputMeasuring Transmitter function

DO NOT CONNECT. Internal use terminals

RTD Pt 100 input (3 wire)

Input signal T/C: T, E, J, K, N, R, S or B

0 / 75 mV voltage input. 0...4-20 mA currentinput (w/ shunt 3.74 Ω not included)

141516

1516

1516 _

+_+

10111213

xxxx

_+

67

xx

45 _

+