Short report on MPPC gain monitoring system S. Uozumi Jun-27 2008 ScECAL meeting L BT, having the gain monitoring system for all the channel is cruci suring the MPPC gain ct drift of the MPPC gain occurred by some reasons. o indispensable for saturation correction. ew photoelectrons are necessary in each channel. ics inter-calibration to measure gains of electronics in low and high gain mode. o necessary for the saturation correction. ge amount of light (~100p.e., at least >10p.e.) is necessary to be d

Short report on MPPC gain monitoring system S. Uozumi Jun-27 2008 ScECAL meeting At the FNAL BT, having the gain monitoring system for all the channel

Embed Size (px)

DESCRIPTION

Revision of the system Look over whole system and correct any problematic points. – Optical connections – status of fiber – Fixture of fiber – And so on … Change LED to very bright one. Use CMB to issue LED driving pulses.

Citation preview

Page 1: Short report on MPPC gain monitoring system S. Uozumi Jun-27 2008 ScECAL meeting At the FNAL BT, having the gain monitoring system for all the channel

Short report onMPPC gain monitoring system

S. UozumiJun-27 2008 ScECAL meeting

At the FNAL BT, having the gain monitoring system for all the channel is crucial for :• keep measuring the MPPC gain … Correct drift of the MPPC gain occurred by some reasons. Also indispensable for saturation correction. A few photoelectrons are necessary in each channel.• Electronics inter-calibration … need to measure gains of electronics in low and high gain mode. Also necessary for the saturation correction. Large amount of light (~100p.e., at least >10p.e.) is necessary to be distributed.

Page 2: Short report on MPPC gain monitoring system S. Uozumi Jun-27 2008 ScECAL meeting At the FNAL BT, having the gain monitoring system for all the channel

So far, not so good…(Ikeda-san`s talk on Jun-13)

●:acrylic plate

●:side emitting fiber 1

●:side emitting fiber 2

●:optical fiber with

notches

Dark noise level (0.02 p.e.)

Page 3: Short report on MPPC gain monitoring system S. Uozumi Jun-27 2008 ScECAL meeting At the FNAL BT, having the gain monitoring system for all the channel

Revision of the system

• Look over whole system and correct any problematic points.– Optical connections– status of fiber– Fixture of fiber– And so on …

• Change LED to very bright one.• Use CMB to issue LED driving pulses.

Page 4: Short report on MPPC gain monitoring system S. Uozumi Jun-27 2008 ScECAL meeting At the FNAL BT, having the gain monitoring system for all the channel

Some remarks on light distribution• Connection between different fiber/material causes a lot of light loss.• We should just use clear fibers, rather than acrylic bars or side-emitting

fibers with clear fiber bundles.• After realizing this, only the choice is clear fiber with notches. We will adopt it for the FNAL BT.

Acrylic bar

Clear fiberAlmost of lights are lost here

Clear or side-emitting fiber

Clear fiber with v-shape notches

Page 5: Short report on MPPC gain monitoring system S. Uozumi Jun-27 2008 ScECAL meeting At the FNAL BT, having the gain monitoring system for all the channel

Revised bench-test system

CMB

Cable (~5cm)

LEDFiber bundle(20 fibers ganged)

1mm clear fiber

Scintillator-layers(Kuraray fiber mega-stripused at 2007 DESY BT)

V-shape notches

Page 6: Short report on MPPC gain monitoring system S. Uozumi Jun-27 2008 ScECAL meeting At the FNAL BT, having the gain monitoring system for all the channel

CMB• Originally developed for AHCAL calibration system by Prague guys.• Can drive up to 12 LEDs.• Provide 8-50 ns pulse into LED.• Currently bench test is ongoing with borrowed CMB, but 2 CMBs will be

delivered soon from Prague.• Control through CANbus is recently succeeded by Daniel.

CANbusline

Power

TCALIB (timing signal)VCALIB (amplitude signal)

12 LEDs

Page 7: Short report on MPPC gain monitoring system S. Uozumi Jun-27 2008 ScECAL meeting At the FNAL BT, having the gain monitoring system for all the channel

LED

Emission angleBlig

htne

ss (a

rbitr

ary

unit)

Blig

htne

ss (a

rbitr

ary

unit)

Wavelength (mm)

PDE of the MPPC

• Adopt Nichia NSPW500GS-K1 W-rank white LED• One of the most efficient LED currently available.• Brightness : 31000 - 44000 mcd @ IF=20mA (~x3 blighter than LED previous used)• Forward voltage / current : 3.2-3.5V / 30mA (100mA possible with pulsing mode)• Emission spectrum more-or-less matches to MPPC PDE spectrum.

Page 8: Short report on MPPC gain monitoring system S. Uozumi Jun-27 2008 ScECAL meeting At the FNAL BT, having the gain monitoring system for all the channel

Clear fiber bundle• Bundle of 20 clear fibers, made for 2000 HCAL beam test @ FNAL.• Diameter of the fiber is 1mm. (One concern : does it fit to 1 mm gap in ScECAL?)• Fiber length ~ 75cm.•White disk is attached for fixture purpose.

Page 9: Short report on MPPC gain monitoring system S. Uozumi Jun-27 2008 ScECAL meeting At the FNAL BT, having the gain monitoring system for all the channel

Notches on the clear fiber• Making many uniform notches is important.• To achieve that, melting method by heating tool would be better

than cutting or gliding.• =0.8mm wire is used as a space to control the depth of notches.

• With the notched fiber, stronger light is emitted to opposite side of the notches.

LEDNotches

Page 10: Short report on MPPC gain monitoring system S. Uozumi Jun-27 2008 ScECAL meeting At the FNAL BT, having the gain monitoring system for all the channel

Fiber arrangement on the mega-strips

Mega-strips (Kuraray fiber readout, used at DESY BT)

• 3mm-wide slits are made on reflector and black sheet. (2mm hole will be enough for actual system at FNAL)• Notched fiber is put on the slit (notches come upside) and fixed by black tape.

Notchedfiber

To bundle

Holes for MPPCs

Page 11: Short report on MPPC gain monitoring system S. Uozumi Jun-27 2008 ScECAL meeting At the FNAL BT, having the gain monitoring system for all the channel

Result of quick test

• CMB DAC count can be set to 0-255 counts. (it is proportional to amplitude of the LED driving pulse.)• LED starts flashing around DAC~100 counts.• With DAC~120, we saw >10 photons at most far-side strip.• With higher DAC setting (>150 DAC count), all the strips get

too much light which causes ADC overflow.• Plots and numbers will follow soon, but notched fiber seems

to be distributing large amount of lights more than enough!

Page 12: Short report on MPPC gain monitoring system S. Uozumi Jun-27 2008 ScECAL meeting At the FNAL BT, having the gain monitoring system for all the channel

Summary & Plans

• Revised system with CMB, new LED and notched fibers works well for the MPPC gain monitoring.

• At the first quick test, we saw large amount of light with this system.

• Plots, numbers and some more tests will follow soon.• At the same time, we will design the actual layout of the

system at the FNAL beamline.