Upload
augusto-vieira
View
293
Download
18
Tags:
Embed Size (px)
Citation preview
Maintenance Manual
Volumed µVP5000
0123
Swiss Made
ARCOMED AG
8105 Regensdorf - Zürich Switzerland
(an ISO 9001 company)
Important
This manual is exclusively intended for authorized personnel who have beeninstructed by Arcomed AG in the maintenance and repair of the Infusion Pumpindicated above.
Arcomed AG shall assume no liability for tampering by unauthorized persons.
Caution: The manufacturer reserves the right to improve the specifications ofthis product without prior notice.
Edition 11/01 -VA-TM-5000-E
Front view µVP5000 -UK
ALARM
RATE
KVO
DEFECT!
2 MINON
OFF
START
STOP
ML/H TOTAL ML
ML INFUSED
INFCOMPL
26
11 10 12 13
14
18
191517
16
7
88
9
5 6
3
4
Rear view µVP5000 -UK
23 24
2
20
2125
22
1
Index numbers
1 Knurled screw
2 Bottle / Plastic container support
3 Drip chamber
4 Empty container detector
5 Pump door
6 Door latch
7 Stop-flow lever
8 Tube guides
9 Air detector
10 Rate display
11 Rate keys
12 Volume display
13 Volume keys
14 Infused display
15 Information window
16 Alarm window
17 ON/OFF key
18 START/STOP-key
19 Alarm tone-mute key
20 Mains supply socket
21 Fuses
22 Nurse call
23 Option RS232
24 Input external drop detector
25 Earth potential plug
26 Pressure detector
27 Infusion pole clamp knob
Contents
Page
Index numbers 0
1. Introduction 1
2. Specifications 3
3. Operation 5
3.12 Parallel / Multiple Infusions 7
4. Alarm Supervision System 8
5. Special key inputs for auxiliary displays 11
6. Warranty 12
6.1 Design Changes 12
7. Inspection and Maintenance Intervals 13
8. Significance of trumpet curves for practical use 14
9. Technical Description 17
10. Trouble Shooting 21
11. Replacement of Parts 22
12. List of spares 5000 25
Appendix A Drawings 24
Appendix B Schematics 31
Appendix C Component Layout 45
Appendix D Performance and Trumpet Curves 49
1. Introduction
1.0 General
The Volumed® µVP5000 Volumetric Infusion Pump has been developed using thelatest state-of-the-art technology.
The Volumed µVP5000 meets the Medical Device Directory (MDD) requirementsCE 93/42 EWG (MDD) and is labeled CE 0123 (TÜV PS Munich, Germany).The manufacturer according to MDD is Arcomed AG, Althardstrasse 146, CH-8105Regensdorf, Switzerland. The person responsible for the EC is Mr Mark Hillson,arcomedical infusion Ltd, West Horndon, Essex CM13 3XS, UK.
The Volumed® µVP5000 meets the Medical Device Directive (MDD) requirements of the EC Guideline 93/42 EEC and is marked CE 0123 (TUV PS Munich, Germany).
The manufacturer according to MDD is Arcomed AG, Althardstrasse 146, CH 8105 Regensdorf, Zurich, Switzerland. The person responsible for the EC is Mr. Mark Hillson, Arcomedical Infusion Ltd., West Horndon, Essex CM13 3XS, UK.
The Volumed® µVP5000 may be operated only on mains power installed to DIN 57107 VDE 0107 or the appropriate national standards. If the integrity of the mains power supply protective earth system is in doubt, the pump should be operated on battery power. Mobile telephones should not be used anywhere near this equipment.
Every effort has been made to ensure this manual‘s completeness. If there are anyqueries with regard to the product‘s use or any ambiguity, you are advised tocontact our Customer Service Department in or the official distributor in yourcountry.
Switzerland: ARCOMED AG, Althardstr. 146, CH-8105 Regensdorf/Zurich,Switzerland.Tel: +41 1 840 47 40, Fax: +41 1 840 06 49
United Kingdom: Arcomedical Infusion Ltd., 5g West Horndon Industrial Estate,West Horndon, Essex CM13 3XS, England.Tel: +44 1 277 810432, Fax: +44 1 277 811967
The Volumed® µVP5000 is designed for use with mains power supply, and hasrechargeable battery back-up in the event of mains supply failure.
1.1 Device set-up and installation requirements
Unpack and check the Volumed® µVP5000 and accessories. Should there be anyphysical damage, the equipment must not be used.
The standard accessories are the drop detector, integral infusion pole, this userinstructions and a mains cord.
The unit when used with on mains, must only be installed with the supplied cord-setfrom Arcomed or approved to BS 1363A by a recognized Test House.
Page 1
Warnings:- do not use this equipment on mains power if the integrity of the protective earthsystem is suspect.- to prevent a possible explosion hazard do not use this equipment in presence ofinflammable anesthetic gases.- do not use this equipment in close proximity to equipment such as surgicaldiathermy with generates high levels of RFI.- do not permit the use of portable telephones in wards where this equipments is inuse.- do not permit this equipment to used if standing in a pool of liquid.
1.2 Routine and maintenance procedure
It is advisable to keep the pump clean and to carry out routine maintenanceperiodically. A mandatory safety and performance test must be performed every 24months or after 10’000 hours of use. (see chapter 7). Maintenance carried outannually ensures optimal performance and prolongs the life of the product.
Should the pump be dropped or damaged in any way, it should be examined by asuitably qualified service technician or returned to the manufacturer for inspectionand test.
1.2.1 Cleaning, disinfection and storage
CAUTION: The pump must be switched off and disconnected from the mains powersupply before cleaning and disinfecting.
The pump must be kept clean and dry. Remove any spillage immediately. Thepump must not be placed in an autoclave.The unit is disinfected by wiping over with a cloth which has been damped slightlywith an alcohol-based disinfectant. Take care when cleaning that no liquid entersthe inside of the pump case. Wait at least 30 seconds after disinfecting beforeswitching the pump on.
1.2.2 Routine maintenance
The following schedule is suggested:1) Clean the pump after each infusion.2) Every 4-8 months (according to usage) check if all alarm systems are operative,inspect mains cord and mains connections and perform routine self test,calibration*and battery check.3) Please refer to the chapter 8 for further details.
*Use an established test procedure or refer to the maintenance manual for advice.Circuits, calibration instructions and component list etc. will be made available toappropriately qualified technical personnel.
Page 2
1.3 Description of Symbols
external empty container detector
nurse call
COM interface RS232
IPX 1
cardiac floating (CF) - type unit - equipment witha floating applied part, providing an adequate degree of protection against electric shock.
Attention: consult accompanying documents.
drip-proof - equipment provided with sufficient protection to prevent fluid ingress.
equipotential point - terminal connected to earthed conductive parts within the device. For connection to a potential equalization conductor, used in medical environments where potential equalization is required.
!
2. Specifications µVP5000
Classification IIbCE certificate number G5 00 11 13006 008
Flow rate range (Micro-pump) 1 - 999 ml/h (0.1 - 99.9 ml/h)Resolution (Micro-pump) 1 ml/h (0.1 ml/h)
Deviation in flow-rate with8101P series administration set typ. +/- 3%
Max. overinfusion (Mech./electr. Defect) 0.5 ml„KVO“ vein open rate (Micro-pump) 3 ml/h (0.3 ml/h) or set rate if less
Infusion pressure min. 60 kPa / 450 mmHg / 0.6 barInfusion pressure max. 120-150 kPa / 900-1100 mmHg / 1.2-1.5 bar
Pressure alarm-limit (Option) adjustable 0-999 mbar / mmHgAir detection ultrasonicSensitivity typ. 100 µl, adjustable from 50 to 250 µl
Battery operation, charged (1.2Ah) > 3hUnit charging time OFF/ON 15 h / 20 hSupply connection 230/240 VAC +10%-15%, 50-60 HzInput current 80 mAMains fuses (F1,F2) 250 mAT/IEC127/III/ SEV 1064
Page 3
Classification of electrical safety Class I / Type CFLeakage current < 40 µAProtection against ingress of liquids IPX1, drip-proofEquipotential bolt DIN 42801Nurse callPotential-free contact switch 24 V/ 0,2 A
Dimensions with ext. Empty Bag Detector (WxHxD) 160x240x210 mmDimensions with int. Empty Bag Detector (WxHxD) 190x240x210 mmCasing High impact ABS plasticWeight 3,9 kgs
Temperature range Operating/Storage 15°C - 35°C / 0°C- 40°C Permits. rel. humidity max. 85%, no vapor depositMax. storage time 3 months without charging
Safety testing IEC 60601-1IEC 60601-2-24IEC 60601-1-2 EMC
1.5 bar 500 mbar 100 mbar
1 ml/h >60 min. 45 min. 8 min.
20 ml/h 6 min. 100 sec 20 sec
100 ml/h 75 sec 25 sec 5 sec
999 ml/h 7.5 sec 2.5 sec 0.5 sec
Bolus 1.3 ml 0.45 ml 0.09 ml
PressureRate
Time to alarm after occlusion(8101P)
with opt. pressure sensor
Page 4
3. OperationAn illustration of the front panel of the Volumed µVP5000 together with a fullexplanation of reference numbers is to be found at the beginning of this manual.
Caution: Use only Arcomed IV administration sets or equivalent(20 drops/ml, Silicone-insert: type 3/4mm, length 135mm, Shore-hardness Shore A,60)
The performance data depend on the IV-line / pump system. The use of a IV-linehas to be in accordance with the technical adjustment of the pump.
If other IV administration sets are used, the operating safety of the pump can nolonger be guaranteed. Patient safety may, as a result, be compromised.In accordance with the norm IEC 60601-2-24, the IV-set should be replaced every24 hours. The IV-sets are for single use only (BS 5724).
3.1. Preparation / Insertion of the IV-seta) If the unit is mounted on a floor stand, care must be taken to ensure that it is fixed
at a height not exceeding 1.2 m above floor level, in order to ensure stability. Theunit can also be placed on a flat surface.
b) With the aid of the knurled screw (1) on the back of the unit, the bottle/containerholder (2) can be adjusted so that the drip chamber (3) is held firm by slotting itinto the empty bag detector (4). Make sure that there are no large ribs or joints inthe passage of the empty bag detector and that fallen drops are detected by theempty bag-detector‘s light-barrier.
c) Carefully purge the infusion set, without allowing any air bubbles to enter, untilthe drip chamber (3) is 1/4 to 1/3 full. If air has entered, repeat purge-procedure.
d) Close the tubing roller clamp.
e) Open the pump chamber door (5) by raising the door latch (6).
f) Press back the red handled „stop-flow“ lever (7) until it is inside the recess.
g) Starting with the left side, insert the silicone section of the IV set into the tubeguides (8) so that it is slightly tensioned. Ensure that the flow direction of thepump from the left to the right is respected. Insert the tubing carefully in the airdetector (9) from top to bottom in the direction of the arrow. Ensure there is a loopbetween the pump door outlet and the air detector.
h) Close pump door (5).
i) Open tubing roller clamp. Check that there is no „free-flow“.Plug the power supply cable into the back of the pump and connect it to an ACpower supply outlet unless battery operation is required. The AC operationpower indicator (15) will light up and battery charging will commence. If not,check power cable and both fuses (24).
Note: The power supply must include a reliable earth connection for safety. Anearth terminal is provided on the back of the unit (25) for connection to anindependent protective earth system if necessary. If the integrity of the protectiveearth system is in doubt, always operate the pump from its internal batterysupply.
k) Press ON/OFF key ( ) (17). The audible alarm beeps once together withthe indication < ! > in the alarm window (16). The software version number(µVP5000, rx.xx) and the configuration of the pump (µVP5000, c.xxx) light upbriefly.
Page 5
3.2. Rate (ML/H)
Set the desired rate with the Up/Down keys (11) in the rate display(10). Keys witharrow up increase the rate, those with arrow down decrease the rate in single steps.
3.3. Total volume (TOTAL ML)
After setting the rate, use the Up/Down keys (13) in the total ml display (12) to setthe total volume to be infused. When the Volumed® µVP-5000 reaches the presettotal volume the pump goes into infusion complete (indication window), gives anaudible alarm and the rate will be switched to the KVO rate.If no total volume is desired, the pump can be started directly after setting the rate.
3.4. Alternative operationIf you want to use one of the following variants, please contact our CustomerService Department or contact the approved ARCOMED distributor for your country(see also point 1.1).
a) Reset the Ml InfusedReset of the Ml Infused after every Start/Stop.
b) SBS (Step by step) or titrationIf the volume is increased after reaching the preset total volume, only thedifference between the new and old values is infused when the pump isrestarted. Volume infused can be either accumulated or not.
c) Storage of the last set ml/h rate and total ml.
d) Display of time infused and or drops infused.
e) Adjustment of the audible alarm volume level.
3.5. Accessories and consumablesAccessories, expendable parts and single-use items may only be used if theycomply with the appropriate international standard and national approvals.Syringes, filters and extension sets must be CE marked.
The Instructions for Use and the mains power supply cable are included asstandard equipment with the Volumed µVP5000.
3.6. START/STOPOnce the rate is set, the pump can be started with the START/STOP key (18). Atevery fallen drop the drop lamp in the indication window (15) lights up.
3.7. Infused display (ML INFUSED)In the Ml Infused display (14) the actual volume is indicated. This indication is notinfluenced by starting/stopping the pump, however it is reset after the total volume isreached and the pump is restarted again. Also turning off/on the pump resets thevolume infused.
3.8. Alarm tone-mute key
The audible alarm can be suppressed for 2 minutes by using the alarm tone-mutekey (19). After two minutes the audible alarm will automatically restart if the pump isstill in alarm condition.
Page 6
3.9. ON/OFF ( ) -KeyTo turn On/Off the pump the ON/Off key (17) is used. This resets the rate ml/h, totalml and infused ml values to zero. The ON/OFF key has to be held down for morethan 1 second to turn off the pump. This prevents the pump from being turned offaccidentally.This switch is not a mains isolating switch, and the battery charging circuit is alwaysconnected to the mains (stand-by).
3.10. IV container exchangeWhen changing the plastic container or bottle, infusion can be interrupted at anytime by means of the 'Start/Stop' key (18) without affecting the set or displayedvalues. In this state, handling operations such as changing the container or IV setand rate changes can be implemented without activating the alarm. In the stopmode, 'KVO' operation is automatically activated.If the pump remains in the stop mode for more than 4 minutes, the audible reminderalarm will sound.
3.11. Keep-Vein-Open (KVO)In certain conditions the Volumed® µVP5000 automatically switches to the KVO-rate (Keep-Vein-Open rate). The KVO-rate is set to 3 ml/h (0.3ml/h Micro Pumpmode). However, if the rate set by the user is lower than 3 ml/h, the KVO-rate isequal to this rate.
3.12. Use in parallel or multiple infusionsIf several pumps are linked together for multiple or parallel infusions, infusion of air,reverse infusion or deviation of the flow rate might occur. If in doubt, please refer tothe matrix in the VDE 0753 part 5 or equivalent specifications for the correct set-upof your system.
3.13. Use of Disposable IV administration setsAlternatives to our recommended 8101P/LP sets can only be used, if their use is notin discrepancy with the security of the system and if the equipment has beencertified by a qualified laboratory or institute.
3.14. External interconnectionsWhen interfacing via the connector (22) and (24), only equipment can be usedwhich meets the specifications of the IEC 60601-1-1:1992 or when it has beenproved by a certified body that there is no influence on the safety of the system.- Use only the external drop- detector Nr. 98502 (24)- Use the cable 94070 for the nurse call (23)- For the RS 232C (optional 20mA passive interface), please contact our Servicedepartment (see 1.1).
3.15. Environment
Used IV set are dangerous and must be discharged according to the localregulations. Used batteries must be recycled or returned to the supplier. A list of theused materials can be obtained form our service department.
Page 7
4. Alarm supervision system
4.1. Alarm causes
During the operation of the unit, the built-in supervision system continually checksthat the pump is operating correctly. In the event of a malfunction, the infusion isimmediately suspended and the alarm is activated. This is signaled by the lightingup of the 'ALARM' display (15) with the corresponding alarm symbols continuouslyilluminated in red and by an intermittent audible alarm. At the same time, the nursecall system will be activated. If there is no 'Occlusion' or 'Air alarm', the KVO rate isactivated.
The Volumed® µVP5000 cannot be started:
- if the infusion tubing is not inserted or is incorrectly inserted in the air detector (4).- if the rate is not set (0 ml/h).- if there is an occlusion or the door is open (alarm through optional pressure
sensor).
During operation, the Volumed 5000 switches on the audible alarm and 'KVO'operation if:
- handling operations are performed on the rate or total input keys (11 and 13)- the Start/Stop (18) key is actuated- the infused quantity has reached the preset total volume value. During operation
the Volumed® µVP 5000 emits an audible alarm and goes into the 'Stop' mode if:
- a deficit/excess total number of mls are detected in relation to the ml/h rate- the liquid level in the drip chamber is too high- the tubing roller clamp has not been opened- the infusion bottle or plastic container is empty- the charge state of the battery can no longer guarantee continuous infusion- the pressure in the tubing exceeds the maximum permissible value- the infusion tubing contains air bubbles- the infusion PVC tubing is not fitted correctly in the air detector- the door is open.
4.2. Cancelling the alarm condition
After rectifying the cause of the alarm or acknowledging the rate change, the alarmcondition is cancelled and infusion resumed by pressing the START/STOP - key(18).
4.3. Air bubble detection
The µVP5000 goes in the following cases in air in line alarm:a) If there is a total of 500µl of air within 15 minutes. Induvidual small air bubbles
(>50µl) will be integrated during a time window of 15 minutes.b) If there is an air bubble greater than 50 - 250µl adjustable.
The air bubble alarm will light up and the audible alarm starts. If the air in line hasbeen solved, the air bubble alarm blinks and the pump can be started.
Page 8
4.4. Battery Alarm
The built-in battery enables the Volumed® µVP5000 to be used independently of amains supply. In the event of a mains failure, the pump automatically passes intobattery operation, without interrupting the infusion. If the battery is completelydischarged, the pump goes into battery alert. The battery alarm symbol in the alarmwindow lights up and an intermittent audible alarm starts. However, the pump runsnormally for at least 3 minutes before stopping. If the mains supply is connected inthe meantime, the alarm is automatically cleared.The Volumed® µVP5000 can also be programmed as to have a prealarm about 30minutes before being stopped, which is indicated when the green battery symbol isflashing. In this mode the battery also starts flashing when the mains supply isdisconnected. In both cases the audible alarm can be muted by pressing the alarmtone-mute key (19).Charging the battery takes about 20 hours while the pump is in operation, 15 hoursif the pump is turned off.
4.5. Nurse call alarmThe unit can be connected to an external call system by means of the connectionsocket (22) on the rear of the unit. For each alarm, the nurse call is thus transferredto the call station. This does not affect either the optical alarm display of the pumpor the audible alarm system.
4.6. Alarm tone-mute switchBy pressing the alarm tone-mute key (19) the audible alarm can be muted for aperiod of 2 minutes. After the muting time has elapsed, the alarm again sounds untilthe cause has been eliminated.
4.7. Alarm indications and their meaningIn order to be able to determine quickly the cause of the alarm, the alarm situationsare indicated by illuminated pictograms in the alarm window (16):
ALARM
RATE
KVO
DEFECT!
Container empty or over infusion
OcclusionBattery discharged Rate changed
General alarmHardware defect alarm
Air bubble alarm Keep vein open
4.8. General indications and their meaningIn the indication window (15) further information is indicated as:
INFCOMPL
Mains supply on
Battery supply on
Infusion complete
Drops falling
Page 9
4.9. Technical description
The microprocessor controlled peristaltic pump Volumed® µVP5000 is suppliedwith a step motor drive and comprehensive software monitoring. The operatingrange of the pump allows rates from 0.1-999 ml/h to be selected. An integrated,rechargeable battery permits mains-independent operation for emergencies or forambulatory use.A mechanical stop flow is installed behind the door to prevent the infusion solutionfrom flowing freely in the event of the pump door being opened inadvertently. Theperistaltic system is driven by a step motor via a positive-engagement toothed belt,with the individual blade movements being controlled by an eccentric camshaft.These blades are fully protected by a rubber cover.All important operating parameters are clearly reproduced on an LED display. Thedesired values are entered via a keypad. The unit incorporates state-of-the-art SMDcircuit technology.
Page 10
5. Special key inputs for auxiliary displaysDisplay test and recall of the last values:Hold the Start/Stop key (18) while turning on the pump. Check whether all displayslight up in the correct sequence (..6789).In this Display test mode, last ml/h, total ml, ml infused and the pressure limit (optionpressure-transducer) of the last infusion will be recalled and displayed.
Display of the software version number:Hold the „Total 1000ml down“ key while turning on the pump. The software versionnumber (µVP5000, r1.xx) is displayed. After about 10 seconds the pump switches tonormal operation.
Display of the configuration:Hold the „Total 100ml down“ key while turning on the pump. The configuration(µVP5000, c.xxx) is displayed. After about 10 seconds the pump switches to normaloperation.
Display of the pump serial number:Hold the „Total 10ml down“ key while turning on the pump. The pump serial number(P.nr. xx.xx xxxx) is displayed. After about 10 seconds the pump switches to normaloperation.
Display of the hospital inventory number:Hold the „Total 1ml down“ key while turning on the pump. The hospital inventorynumber (no xxxx) is displayed. After about 10 seconds the pump switches to normaloperation.
Micro-pumping-mode:Hold the „Rate1ml down“ key while turning on the pump. The pump switches to themicro pump mode. Rates can be set from 0.1 to 99.9 ml/h and the total volume to beinfused from 0.1 to 999.9 ml. Carefully watch the decimal point in the 3 displays.Otherwise the pump operates exactly the same as the standard version 1 to 999ml/h.Attention: When turned off in the micro pump mode, the µVP5000 automaticallyreverts to its standard version when switched on again, unless permanentlyconfigured through the software code access.
Pressure display mode (option):Press audible alarm 2 min. suppression key (19) once while the pump is running.Line pressure in mbar or mmHg will be displayed for 20 sec in the ml infuseddisplay (14). By pressing the same key twice in quick succession, not only will thepressure in mbar or mmHg be displayed in the ml infused display (14), but thepressure level will also be displayed in the total ml display (12). During this 20 sectime period, the pressure level can be adjusted below or above the preset 800 mbaror mmHg to a maximum of 999 or a minimum to suit the displayed in line pressure.Remember the lower the pressure level, the lower the bolus occlusion on releaseand the quicker the time to alarm. As soon as the line pressure exceeds theprescribed pressure level, the pump will go into occlusion alarm in the alarmwindow (16). An audible alarm will also be triggered. Once the occlusion has beenreleased, the visual occlusion symbol will flash, and both the general alarm !and audible alarm will continue until the START/STOP key (18) has been pressed.
Caution: Do not open the door if the system is under pressure!Page 11
6. Warranty
ARCOMED AG offers a 12-month warranty on every Volumed 5000 unit, effectivefrom the date of delivery.The warranty covers the repair and replacement of faulty parts due to manufacturingor material defects. This warranty is rendered null and void if the unit is tamperedwith by unauthorized persons and if the inspection/maintenance intervals are notobserved.
This warranty does not cover the elimination of problems caused by incorrectoperation, inappropriate handling or normal wear and tear.
The supplier only accepts responsibility for the safety, functional reliability andperformance of the unit providing that:- assembly, extension work, resetting, modification or repair work has been carriedout by personnel specifically authorized by him,- the electrical system at the operating site meets the IEC requirements, and the unitis operated in accordance with these instructions for use.
CAUTION:The Volumed® µVP5000 may only be used with accessories, expendable parts andconsumable items which have been authorized by ARCOMED AG as beingapproved as being safe.
The information provided in this manual applies to the currently prevailing situationand is given in good faith. We reserve the right to make modifications which may bein the interest of technical progress.
6.1. Design changes
ARCOMED endeavor to ensure that future improvements and modifications arecompatible with the earlier models.
NOTE: When ordering spares, always state the model, serial number and whereapplicable the color of the unit in question.
Page 12
7. Inspection and maintenance intervals Volumetric InfusionpumpVolumed® µVP5000 (according MDD)
Interval: After 24 m months or 10’000 h of use.The following checks must be done by an engineer with sufficient technical background to comply withthe safety regulations.
What to do How / Equipment Remarks Result
Visual Check
Housing Physical damageExternal empty container detector (ECD) (Easy Clip) Physical damageor Internal empty container detector (ECD) ( especially spring ) Physical damageDoor, door latch Clean, functionStop-flow lever Clean, functionCover for peristaltic blades Physical damageInscriptions, display Readable, damageDisplay - LED Function, display testMains plug, fuses Damaged, valuesAir in line Physical damage
Functional checks
o Spring plate manual check free motion
o Pressure checks: IV-set filled with water,manometer & syringe
o minimal mechanical pressure: preload system with watch manometer o alternative testsyringe to 0.7 bar PTD-50004 min. at rate 5 ml/h pressure always(press. limit 999 mbar) above 0.6 bar p min =
o maximal mechanical pressure: rate 400 ml/h pressure always(press. limit 999 mbar) below 1.5 bar p max =
o Pressure Sensor (option): rate 100 ml/h alarm reactionpressure limit 500 mbar withinmake occlusion on set 20 sec ±10 sec
o Rate check: Rate 100 ml/h ± 5 % accuracy o alternative testTotal of 100 ml Refer to trumpet VT-5000
curve (tech manual) % dev =while running on 100 ml/h
o simulate missing drops Take out drop chamber Visual and acousticof empty container detector alarm
o simulate Air in Line Take out IV - set Visual and acousticof Air Detector alarm
o check nurse call e. g. open door while signal at connectorpump is running
o (External pump stop only Option RS 232C)
Electrical safety according to IEC 601Safety Tester 601 IEC 601.1, section 19o Leakage current ≤ 75 µAo Resistance protective conductor ≤ 100 mOhm
The rates of the fuses must comply with the rates recommended by arcomed (producer):Conventional transformer 230V :100 mAT/250V, toroid transformer 230V: 250 mAT/250V,toroid transformer 115V: 500mA T/250V ( IEC127/III/SEV 1064).
Caution: After any work on the pump (e. g. adjustment of programming, change of parts, any opening ofthe pump) this inspection must be made and all checks must be documented with the serial number of thepump.Serial Number: Remarks: Date /Signature:
Page 13
8. Significance of trumpet curves for practical use
Trumpet curves indicate for 5 different observation windows the maximum and minimummean values of the flow rate in ratio to the preset flow rate.Known therefore is the discrepancy per time-window. For optimal use of the infusion pumpVolumed® µVP5000, the trumpet curve is an important factor in deciding whether the pumpcan be used with the prescribed drug.
Volatile drugs with short therapheutic half life demand high accuracy.
For a drug where the plasma-half life is e.g. 1 min. discrepancy of the flow rate of 15% perminute would mean the same discrepancy for the plasma level. Therefore, a predictableconstant impact of the drug would not be guaranteed.
Example:
Intraveinous infused Insulin has a therapheutic half life of 15 minutes. A flow deviation of ±15% within 40 minutes would have the same (and even greater) influence on deviation ofthe plasma level and therefore on its impact.
This is inacceptable to physicians and nursing personel.
It is important to know that the deviation in a short observation window depends strongly onthe preset rate. The Volumed® µVP5000 has at a rate of 5 ml/h a deviation if ±7% in awindow of 2 minutes and ± 3% in a window of 5 min.However, at a rate if 25ml/h, the deviation is smaller than 2% for a observation window of2 minutes (see also following table).
Table 1: Flow Accuracy of the Volumed® µVP5000 (typical values)
Observation window
2 min 5 min Rate (ml/h) Max Min Max Min
5.0 +5.25% -6.17% +2.60% -2.33%25.0 +1.85% -1.82% +0.75% -0.77%50.0 +1.38% -1.17% +0.38% -0.50%
Rate (ml/h) Measured Flow Rate Standart Test SampleFlow Rate Error Deviation Duration Duration
(ml/h) (%) (ml/h) (h.min) (h.min)
5.0 5.083 1.652 0.497 2.00 1.0025.0 25.364 1.458 0.642 2.00 1.0050.0 50.235 0.470 0.868 1.30 1.00
Page 14
Flow [ml/h]:
0.0
12.5
25.0
37.5
50.0
0:00 0:10 0:20 0:30 0:40 0:50 1:00 1:10 1:20 1:30 1:40 1:50 2:00
Flow: ø Flow:Weight [mg]: Time:ø Error[%]:
File Name:
µVP5000 25ml/h# Scans:
240Interval (s):
30Evaporation:
0.19Rate (ml/h):
25.0Rec. date: Rec. time:
6.2.2001 8:55:37 Uhr
Trumpet Curve:
-10.0
-5.0
0.0
5.0
10.0
0:00 0:10 0:20 0:30
2 Min (max):
1.44
5 Min (max):
0.36
2 Min (min):
-1.63
5 Min (min):
-0.76
25.87 25.316 1.267 50243.7 120.30
6.2.2001 8:55:37 Uhr
Flow [ml/h]:
0.0
2.5
5.0
7.5
10.0
0:00 0:10 0:20 0:30 0:40 0:50 1:00 1:10 1:20 1:30 1:40 1:50 2:00
Flow: ø Flow:Weight [mg]: Time:ø Error[%]:
File Name:
µVP5000 5ml/h# Scans:
240Interval (s):
30Evaporation:
0.19Rate (ml/h):
5.0Rec. date: Rec. time:
5.2.2001 8:29:10 Uhr
Trumpet Curve:
-10.0
-5.0
0.0
5.0
10.0
0:00 0:10 0:20 0:30
2 Min (max):
3.93
5 Min (max):
3.09
2 Min (min):
-3.56
5 Min (min):
-2.25
4.64 4.971 -0.569 9561.1 120.30
5.2.2001 8:29:10 Uhr
9. Technical description9.1 Introduction
The microprocessor controlled peristaltic pump Volumed® µVP5000 is suppliedwith a step motor drive and comprehensive software monitoring. The operatingrange of the pump allows rates from 0.1-999 ml/h to be selected. An integrated,rechargeable battery permits mains-independent operation for emergencies or forambulatory use.A mechanical stop flow is installed behind the door to prevent the infusion solutionfrom flowing freely in the event of the pump door being opened inadvertently. Theperistaltic system is driven by a step motor via a positive-engagement toothed belt,with the individual blade movements being controlled by an eccentric camshaft.These blades are fully protected by a rubber cover.All important operating parameters are clearly reproduced on an LED display. Thedesired values are entered via a keypad. The unit incorporates state-of-the-art SMDcircuit technology.
9.2 Circuit description
9.2.1 Microcontroller
Microcontroller D1 monitors the following signals :
Drip detectorThe drip chamber is monitored by a light barrier.Each falling drop produces a signal which is amplified in the OP N2, lengthened inmonoflop D7 and processed by the microcontroller.
Motor controllerMotor controller D10 controls the step motor driver D11 which activates the stepmotor in chopper mode. Simultaneously, after eight step motor pulses it supplies anacknowledge signal, the arrival of which in the motor controller is monitored by thesoftware.
Pump camshaft rotation pulsesMounted adjacent to the pump camshaft is a hall sensor N12 which produces apulse for each revolution of the pump camshaft. The chronological frequency of thisis monitored in the microcontroller by the software.
Air detectorIf air is present, the air bubble detector produces a signal which is monitored in themicrocontroller in combination with the delivery rate.The air bubble detector is periodically checked for correct functioning(AirDetectorTest). This high frequency detector is capable of monitoring both clearand opaque fluids.
Pressure transducer (optional)The pressure from the tubing is measured with the temperature compensatedpressure transducer (Whetstone bridge) which transmits an analog signalcorresponding to the pressure in the line. The signal is converted with an A/Dconverter in the microprocessor conditioned and monitored in the microcontroller.The pressure can be displayed and the pressure limit can be set by the user. If thepressure exceeds the set pressure limit, an occlusion alarm stops the pump. Alsoopening the door is detected and the pump can not be restarted until the door isclosed.
Page 17
WatchdogThe watchdog circuit D6 monitors the program run time. If the time limits areexceeded or fall, semiconductor relay V1 interrupts the supply to the step motor. Afault is signaled to the microcontroller via inverter D5. The relay V1 is periodicallytested via the SM-Test connection. A second watchdog is located in themicrocontroller which also monitors the program run time.
Operating voltagesThe supply voltage, the 5V - and reference voltage are measured and monitored viadisc resistors by the analog/digital converter in the microcontroller.
Mains/battery operationThe level of the operating voltage determines whether operation is from mains orbattery.
Battery voltageWith battery operation the battery alarm is activated if the minimum operatingvoltage falls below limits and in the case of exhaustive discharge voltage, the unit isswitched off.
Microcontroller D1 activates the following signals :
Step motor pulsesThe step motor pulses are delivered by the microcontroller from the quartzfrequency corresponding to the desired delivery rate.
Step motor Start/StopThe microcontroller produces the Start/Stop signal for the step motor on the basis ofthe operating status. This is supplied to the motor controller D10 via the inverter D4.
WatchdogWatchdog D6 is periodically triggered after each program run.
Buzzer (audible alarm)Dependent on the operating status the microcontroller activates the audible alarmtransmitter H1 via driver V8.
Alarm relayDepending on the operating status the microcontroller activates the alarm relay K1(nurse call) via inverter D5 and driver V9.
On/Off functionAfter switching on, the microcontroller activates the holding of semiconductor relayV18. This maintains the supply voltage on. On actuating the key "I/0" or in the caseof a discharged battery, the microprocessor cuts off the supply via semiconductorrelay V18.
A/D converterThe microcontroller controls the internal A/D converter and evaluates the results.
EEPROM memoryImportant data and preset values are stored in the non-volatile memory D4. Also incase of a completely discharged battery the data will be maintained and read backafter the next power on.
Page 18
9.2.2 Display
The status and the important data are displayed by pictograms and 7 segment LEDdisplays. The display drivers M2, M3 and M4 are controlled by the circuits D3 andM1 and multiplexed by the microcontroller.
Rate 3-digit numerical displayTotal 4-digit numerical displayMl infused 4-digit numerical displayAlarm pictograms red LEDsStop red LEDInfusion complete yellow LEDBattery supply green LEDDrop falls green LED
The green mains-LED is supplied directly from the transformer.
9.2.3 Setting of the desired data
All the inputs are actioned by the keys on the front panel and are periodicallychecked by the microcontroller. Only the „ON/OFF“-key is separately connected andcontrols the power on circuit. Power off is however controlled via the microcontroller.The keys have the following meaning:
function key
On/Off power ON/OFF (20)Rate setting Rate 100 ml/h up/down (11)
Rate 10 ml/h up/downRate 1 ml/h up/down
Total volume setting Total 1000 ml/h up/down (13)Total 100 ml/h up/downTotal 10 ml/h up/downTotal 1 ml/h up/down
Start/Stop of the pump Start/Stop (21)Alarm tone-mute (for 2 minutes) Alarm tone mute (22)
9.2.4 Power supply
The power supply from the mains is effected via the mains transformer and rectifierV14. In the case of a power supply failure, power is supplied without interruptionfrom the integrated battery unit. Charging circuit V16 provides for constant batterycharging while the unit is connected to the mains (also when switched off).
The electronic relay V18 switches the supply on or off. This function is activatedfirstly by the ON/OFF key and secondly by a hold command from the microprocessorvia inverter D5.Additionally the microprocessor can disconnect the supply in the same manner.Switching controller N3 produces the +5V logic voltage.
Page 19
9.2.5 Monitoring
Air bubbles: 50 µl detection limit, integration over 15 min,individual bubbles > 50 - 250µl adjustable
Air detector test: every 100 ms.Pressure (optional): Contious monitoring of pressure in the line in mbar or mmHg.
Pressure limit 1…999 mbar/mmHg settable.
Pump camshaft: One revolution of the pump camshaft requires 1164step motor pulses. For each complete revolution of the pumpcamshaft, the hall pulse initiates a comparison with thenumber of step motor pulses supplied. Tolerance perrevolution +/- 20% pulses.
Semiconductor relay: Functional check every 50 msec.Keypad: Key actuated in excess of 30 seconds triggers a fault signal.Microcontroller: When the unit is switched on, the registers are
checked for their initial status and a continuous ROM, RAMand CPU test is conducted.
Program run time: External watchdog monitors program run time; tolerance is approx. +/-15%.
Operating voltages: The microprocessor monitors the following voltages:- mains/battery changeover threshold: 15.0 V- battery charged: 12.9 V- battery discharged: 10.5 V- 5 V monitoring: <4.75 V: fault signal
>5.46 V: fault signal- A/D converter: fault > 2.5%: fault signal
Other functions: - audible alarm in standby mode after 4 minutes- Suppression of audible alarm: 2 min.- audible alarm repetition: On: 0.6 sec.
Off: 3.0 sec.- Nurse call pulse: 0.2 sec to 12 sec or static.- Interface (option): 20 mA passive
Page 20
10. Trouble shooting
In the case of a "DEFECT" signal a fault code is displayed automatically. A two-digitnumber appears in the total display (F-xx). The associated causes are listed in thefollowing table:
Code Meaning:(Total Ml)
0 Program sequence1 Air detector test2 Step-motor too fast3 Step-motor too slow4 AD converter5 Reference voltage6 Supply voltage7 Key locked8 Relais test9 CRC-test EPROM
10 Program timing11 CRC-test EEPROM12 Watchdog13 Initial CPU test14 Register test15 RAM test16 Bit-walk test RAM17 CPU instruction set18 Sync. watchdog19 Home counter high
Page 21
11. Replacements of parts
Caution: The unit must be switched off and disconnected from the mains supplybefore any maintenance work is carried out.
The Volumed µVP5000 is dismantled as follows:
a) Remove the 4 screws in the rear of the pump and separate the two housings.Disconnect the plugs only if needed, otherwise the correct connection has to beverified before reassembly (see topography main PCB).
b) removal of the main PCB:Disconnect the 230V/240V cable from the mains socket, the connector to thebattery and the connector to the drip detector. Remove the two bolts in the blueclip and take out the the rear chassis. Disconnect all plugs from the PCB andremove the two screws holding the main PCB. Carefully remove the main PCB.Caution: Improper handling of the main PCB can damage the sensitiveelectronic components as a result of electrostatic charging.
c) removal of the display PCB:Unscrew the 5 nuts inside the housing and carefully remove the display PCB. Thefront panel can be removed from the front side.After reassembly use the spacers between the display PCB (2.5mm) and thehousing , then adjust with the 5 screws the distance between display and frontpanel to ensure proper function of all keys (audible snapping while pushing thekeys).
d) removal of the front door:Open the door approx. 45 ° and pull out. Preserve white door stop on axle.
e) removal of the pump unitUnscrew the 4 spacers, remove the two beams and air bubble detector andunscrew the two lower spacers and screws of the pump unit. The entire pump unitcan than be taken out.
Caution:When replacing the pump block, a space of 79.5 mm is to be set between the centerof the door hinge axis (93152) and the center of the door catch axis (93340) beforethe fastening screws are finally secured (to ensure that the door latch functionsproperly when the door is closed).
Assembly takes place in reverse order. Check proper fit of the front chassis intocentering recesses. Check pump housing gasket‘s sealing for proper joint. Replaceif damaged. Replace and align white door stop on axle before refitting pump door.
Page 22
11.1 Inspection and maintenance intervals
Cleaning / Desinfection
CAUTION: The pump must be switched off and disconnected from the mains powersupply before cleaning and disinfecting.
The pump must be kept clean and dry. Remove any spillage immediately. Thepump must not be placed in an autoclave.The unit is disinfected by wiping over with a cloth which has been damped slightlywith an alcohol-based disinfectant. Take care when cleaning that no liquid entersthe inside of the pump case. Wait at least 30 seconds after disinfecting beforeswitching the pump on.
Maintenance intervals with safety testing
With the exception of the nickel-cadmium (or ‘green‘ nickel hydrid) battery, the unithas no wearing parts which are subject to regular replacement. As a consequence,the condition of the battery and a visual inspection of the inside of the unit forcontamination or loose or damaged parts is to be performed. If necessary, theinterior of the unit is to be cleaned.The battery is checked by switching off the unit and leaving it connected to themains for 15 hours so that the battery can be fully charged. This is followed byswitching the unit on and allowing it to run from the battery, with the time beingmeasured until the battery alarm is triggered. It should amount to a minimum of 3hours (1.2Ah battery). A battery can be regenerated in some cases by repeatedcharging and discharging (memory effect).
The safety testing can only be done by qualified personnel and has to be inaccordance with the safety requirements.
The inspection and safety testing of the unit is to comprise cleaning as well asperforming visual checks for damage, checking alarm operation and verification ofthe accuracy of delivery. No intervention inside the unit is necessary.The alarm functions are checked as follows :
- Install a water filled tubing, but remove the drip chamber from the emptycontainer detector.
- The display test is actuated by pressing the "START/STOP" key while the unit isswitched on. Check that all indicator lamps and the audible alarm functionproperly.
- Set rate 20 ml/h and start. The bottle/container empty alarm must be activatedafter 60 sec.
- Set rate 500 ml/h and start. The drop alarm must be activated after approx. 2 sec.- Simulate drops by placing a finger through the drip detector light barrier. After 22
drops the alarm must be activated.- Remove the tubing from the air detector, the air alarm will be activated.- Optional pressure transducer: Test of the response to a complete occlusion.
Tubing must be filled completely with water. Rate 100 ml/h. Pressure limit set to500 mbar (360 mmHg) above actual pressure reading. Response time approx.15 sec. (± 10 sec). Long term test: Let the pump run for 24h at different rates (e.g.
Page 23
25 ml/h, 100 ml/h, 999 ml/h). No unexpected alarms.- Switch off the unit.
The delivery accuracy is checked for instance by entering a "TOTAL" of 25 ml witha rate of 250 ml/h. Over a period of six minutes, 25 cm3 or 25 g +/- 5% of watershould be delivered into a measuring beaker.
Electrical safety testing is to be carried out according to IEC regulations and is tocomprise measuring the protective conductor as well as checking the leakagecurrent.The protective conductor connection in the unit is to be tested according to IEC 601part 1, section 18 and is not to exceed 0.1 ohm.
The leakage current is to be tested according to IEC 601 part 1, section 19. In noevent is it to exceed 0.1 mA.
Page 24
12. List of spares 5000
Please mention serial number of pump with your order. For spare parts of older versions(befor AC 040) of Volumed 5000 and older arcomed pumps see their manuals.Following to this spare parts list you find a table of revisions with orderingnumbers for main pcb, EPROM, fuse holder, and battery respectively doorand rubber cover.
Spare part number Description
0 0 1 1 0 Mains supply cable Switzerland 5000 final0 0 1 2 0 Mains supply cable Germany 5000 final0 0 1 4 0 Mains supply cable UK 5000 final0 0 1 6 0 Mains supply cable Italy 5000 final1 0 0 3 0 Equipotential plug connector f inal1 0 7 1 0 Molex 4-pole plug housing f inal1 0 7 8 0 Molex 8-pole plug housing f inal8 0 0 5 0 Upgrade plate door 5000 AC105 upgrade / Final8 0 1 0 2 Instructions for use 5000-E document8 0 1 0 4 Instructions for use 5000-esp document8 0 1 0 5 Instructions for use 5000-it document9 0 1 1 0 Knurled screw for stand f inal9 0 4 4 0 Pulley 32 XL 025 compete pump bloc9 0 4 5 0 Pulley 11 XL 025 with flange ring pump bloc9 0 4 6 0 Toothed belt 90 XL 025 pump bloc9 1 0 6 0 IV pole 450 mm final9 1 2 9 0 Spring plate door9 1 2 9 1 Spring plate AC050 door9 1 3 3 0 Compression spring 0 door9 1 3 3 1 Compression spring 1 door9 2 2 0 1 Pump housing (blade housing) AC040 pump bloc9 2 2 3 1 Permanent magnet dia. 6x3 pump bloc9 2 4 5 0 Camshaft complete pump bloc9 2 4 6 1 Blade AC040 pump bloc9 3 0 0 1 Air in line detector (blue) submodul / final9 3 0 0 3 Air in line dummy (blue) submodul / final9 3 1 2 1 Hall pcb complete AC040 pump bloc9 3 1 4 0 Door (blue) door9 3 1 4 1 Door (brown) door9 3 1 4 2 Door (blue) AC050 door9 3 1 5 2 Door latch (blue) AC050 door9 3 2 1 0 Silent bloc 10/10 pump bloc9 3 2 7 1 Stop flow lever AC040 pump bloc9 3 2 8 1 Red cover for stop flow lever AC040 pump bloc9 3 2 9 1 Spring stop flow lever AC040 pump bloc9 3 3 0 1 Locking pawl AC040 pump bloc9 3 3 1 1 Spring locking pawl AC040 pump bloc9 3 3 4 0 Spindle for door latch pump bloc9 3 3 5 1 Spindle for door AC040 pump bloc9 3 3 8 0 Battery 12V / 1.4 Ah f inal9 3 6 2 0 Step motor pump bloc9 3 9 1 1 Counter bearing AC040 pump bloc
Page 25
9 3 9 2 0 Motor flange pump bloc9 3 9 3 0 Clutch complete pump bloc9 3 9 8 0 Ball bearing dia. 6/19x6 pump bloc9 3 9 9 1 Spacer plate AC040 pump bloc9 4 0 5 1 Side wall. right AC040 pump bloc9 4 0 6 1 Side wall. left (small) AC040 pump bloc9 4 0 7 0 Cable for nurse alarm 3 m final9 4 1 1 1 Pump block V2.00 complete. less door AC040 final9 4 3 2 1 Door complete Silicone (blue) submodul / final9 4 3 2 7 Door complete Silicone (blue) BC110 AC050 submodul / final9 8 0 0 1 Housing front (blue) f inal9 8 0 1 1 Housing rear (blue) f inal9 8 0 1 3 Housing rear ec(blue) f inal9 8 0 1 6 Housing rear ec complete (blue) f inal9 8 0 2 0 Chassis front 1.00 pump bloc9 8 0 2 1 Chassis front AC040 pump bloc9 8 0 3 0 Chassis back f inal9 8 0 4 0 Chassis beam1 final9 8 0 5 0 Chassis beam2 final9 8 0 6 1 Clip (blue) f inal9 8 0 7 0 Threaded spindle M8x50 final9 8 0 9 1 Adapter Diabolo (blue) f inal9 8 1 0 0 Spacer M5/60mm thread inside f inal9 8 1 1 0 Spacer M5/60mm final9 8 1 2 0 Spacer M5/50mm final9 8 1 3 0 Spacer M5/35mm thread inside.Polyamid f inal9 8 1 4 0 Screw M5/40mm final9 8 1 5 0 Screw M5/16mm final9 8 1 6 0 Screw M3/10mm drop9 8 1 7 0 Screw M3/10mm final9 8 2 0 0 Nut M6 final9 8 2 1 0 Nut M3 final9 8 2 5 0 Washer M6/6.4 f inal9 8 2 6 0 Washer M5/2mm final9 8 2 7 0 Washer M4/4.15mm final9 8 2 8 0 Washer M3/3.2 f inal9 8 2 9 0 Screw M5/12 f inal9 8 3 0 0 Security clamp for rubber feet 8mm final9 8 3 2 0 Washer 6/3 AC050 door9 8 4 0 4 Front plate (UK) f inal9 8 4 2 1 Label set 5000 (blue) f inal9 8 4 4 4 Condensed Instructions 5000 UK final9 8 4 7 4 Alarm sticker UK final9 8 5 0 1 IR-light barrier (drip detector) (blue) submodul / final9 8 5 0 2 Easy Clip submodul / final9 8 5 1 1 Clamping spring for light barrier drop detector9 8 6 0 1 Pressure sensor Silicone submodul / final9 8 6 0 3 Pressure sensor Dummy submodul / final9 8 6 4 1 Adapter P-Sensor (gray)9 8 6 9 0 Ball ø 2mm p sensor9 8 7 0 0 P-Sensor A1 p sensor9 8 7 8 0 Magnet EC AC045 Easy Clip9 8 8 1 0 Anti Slide 25/140mm final9 8 8 7 0 Screw M4/32 drop detector9 8 9 0 1 Rubber foot front (blue) f inal
Page 26
9 8 9 1 1 Rubber foot rear (blue) f inal9 8 9 2 0 Gasket (Seal ) HF final9 8 9 3 0 Rubber cover for peristaltic blades old pump bloc9 8 9 3 1 Rubber cover for peristaltic blades (4 h AC040 pump bloc9 8 9 3 2 Rubber cover for peristaltic blades (pre AC040 pump bloc9 8 9 7 0 Seal for air detector f inal9 9 0 0 1 Main pcb board V2.02 AC100 final9 9 0 0 2 Main pcb board V2.03 f inal9 9 0 0 3 Main pcb board V2.04 f inal9 9 0 1 0 Connector pcb board submodul / final9 9 0 2 0 Display pcb board submodul / final9 9 0 3 0 HW-Key (red) submodul / final9 9 0 4 0 Tec-Key (yellow) submodul / final9 9 0 5 0 PCB P-Sensor p sensor9 9 1 0 0 Ribbon cable 26-pole f inal9 9 1 1 0 Ribbon cable 14-pole f inal9 9 1 2 0 Earth cable f inal9 9 1 2 1 Earth cable long (UK) f inal9 9 1 3 0 Buzzer SMA24L final9 9 2 1 1 Transformer 7.5VA final9 9 2 1 2 Toric transformer 15VA 230V AC030 final9 9 2 1 3 Toric transformer 15VA 110V final9 9 2 2 0 Micro fuse 1AT batttery f inal9 9 2 2 1 Mains plug (screw) AC055 final9 9 2 3 2 Fuse-holder 230/240/115V final9 9 2 4 2 Fuse 250mAT/250V AC030 final9 9 3 0 1 EPROM 5000 12V battery f inal9 9 3 1 0 Opto coupler 4100 interface9 9 3 2 0 Opto coupler 4200 interface9 9 5 3 0 Adapter Magnet EC (blue) with magnet AC045 upgrade / Easy Clip9 9 5 5 0 Screw M5/30 f inal9 9 5 6 0 Screw M5/20 f inal9 9 5 7 0 Nut M4 final9 9 5 8 0 Screw M3/6 f inal9 9 5 9 0 Washer M4/4.1 f inal9 9 6 0 0 Nut M3 AC055 final9 9 6 1 0 Screw M3/10 AC055 final9 9 6 2 0 Protection main PCB final9 9 6 9 0 Bolt locking pawl AC040 pump bloc9 9 7 0 0 Bearing locking pawl AC040 pump bloc9 9 8 3 0 Washer 4/9.3/0.7 pump bloc9 9 8 7 0 Screw M3/3 pump bloc9 9 8 8 0 Screw M3/4 pump bloc9 9 9 1 0 Precision washer 4/8/0.5 pump bloc9 9 9 3 0 Screw M3/5 pump bloc9 9 9 7 0 Washer M3/3.2/7/0.5 AC055 final
Page 27
Volumed µVP5000 (engl) Ordering numbers main pcb - EPROM - battery - fuse holder
Version BC005 (>030305) BC030 (>350405) BC100 (>438603) BC115 (>100610)12V battery 12V battery 9.6V battery 12 V battery
charge transistor 12V charge transistor 12V charge circuit 9.6V charge circuit 12Vtransformator 7.5VA toroid trafo 15VA toroid trafo 15VA toroid trafo 15VA
main pcb 990021) 990022) 990023) 99002(upgrade: ‘12/95’) (upgrade: ‘12/95’) (upgrade:‘12/95’) ‘12/95’
EPROM (Rev-Nr 99301 99301 99304 99301to be mentioned ) 2.xx 2.xx 2.xx. 2.xx
battery 93380 93380 93381 9338012 V battery 12 V battery 9.6V battery 12V battery
10 cells 10 cells 8 cells 10 cells
fuse holder 99231 992324) 99231 99232230/200/115V (upgrade: 230/240/115V) 230/200/115V 230/240/115V
(only with 99002) was: new: was: new: was: new: was: new:
Ordering numbers door - rubber cover - pump block
Version BC005 (>030305) BC040 (>217503) BC050 (>100502) BC110 (>400606)door old design door old design door new design door new design
metal spring plate metall spring plate plastic spring plate ribbed spring platepump block long pump block short pump block short pump block short
door complete 94321 94327 94327 94327
rubber cover 98930 989315) 98931 5) 98932v2.00 v3.00 v3.00 v3.14
6 holes 4 holes 4 holes Preshaped1 For pumpsbefore BC30 new main pcb with charging circuit 12V (99002) will be delivered. For complete upgrade to 15VA order toroid trafo (99212) and new fuseholder(99232) in addition and change these parts too. 2 For pumps from BC030 to BC100 new main pcb with charging circuit 12V (99002) and new fuse holder (99232) should be ordered. 3 For pumps BC100 to BC 110 new main pcb with charging circuit 12V (99002) is delivered (battery stays 9.6V). If you wish to upgrade your pump to 12 V you have to order and change battery 12V (99301), EPROM (99301) and fuseholder (99232) . caution: main pcb with charging circuit 9.6V (99001) can not be used with 12V battery!4 caution: only change fuseholder (99232) if new main pcb 12V (99002) is used.5 New rubber cover preshaped (98932) to be used with ripped spring plate in door (transformation kit 80050) only . 18.6.2001/boe/bestellNr. bis BC115
wa
sw
as
new
: (s
par
es)
new
: (s
par
es)
Drawings
ALARM
RATE
KVO
DEFECT!
2 MINON
OFF
START
STOP
ML/H TOTAL ML
ML INFUSED
INFCOMPL
98001
98404
9860198603
9890198300
94327
9842193001
Housing front 5000
15.3.2001/mvo
981209813099560
98050
981009811099550
94111
98021
9815098260
98040
99020
Housing front 5000 - inside view
982109827098280 98920
98970
15.3.2001/mvo
98421
98501 98511
98290
90110
98061
992211003099120
9891198300
98140
98421
98070
98421
9801198013 (EC o. Magnet)
98016 (EC kompl)
Housing rear 5000
98502
15.3.2001/mvo
99232 (AC115)99241 (100mAT)99242 (250mAT)
71271
99212 (230V)99213 (110V)
98030
93380 (10 Zellen)93381 (8 Zellen)
9901099110
99301 (10 Zellen)99304 (8 Zellen)Version angeben !
99003 (alleVersionen, sieheVersionsmatrix)99100
15.3.2001/mvo230 V 110 V
(set fuse holder to 115V)
black
orangegrey
black
orangegrey
1
3
2
Housing rear 5000 - inside view
Door 5000
door complete without door cover 94327upgrade to ribbed plate 80050(please mention pump typ and iv set; for doors later than BC 050 the ribbed plateonly is available)
93152
93142
VO LUMED
98421
µVP5000
Springs4 x Nr.2 91332(Standard for 5000)
4 x Nr.1 91331
91291ribbed
(BC050 / 110)
98641
Pump block(AC040 / 110)
98021
93351
92450
93980
92201
93620
93930
93920
93210
94051
90460
90450
90440
92231
93121
9399193980
92461
94061
93281
93301
93271
Spring 93291
Spring 93311
10780
10710
93991
93340
93911
9893198932 preshaped
6x12x0,1
31.05.96/mvo
6x12x0,1
6x12x0,5
6x12x0,5
99690
99700
Schematics
DisplayTotal ml
Current
D11Motor Driver
Motor
Hall
AirDetector
ACMain
V12 V13 V14
OnOff
SR
+5V
Vcc
D4 Inverter
Power Hold
Battery
N1Reset
SerialInterface
K1Relay
RelayDriver
D4 Inverter
Air. Det. Test
Air Bubble
BuzzerDriver
Ak. Alarm
V1D6
Watchdog
D4 Inverter
D4 InverterD10Motor
Controller
Start / Stop
Clock
Home
Fail
D4 Inverter
WDT
SM-Test
Hall
=
VoltageControl
Ref. 2.5V
Ref. Ctrl.
18V Ctrl.
5V Ctrl.
P1.5
P4.7
P3.5
INT 1
P1.4
P4.6
P1.1
P4.5
P1.2
TXD
RXD
PWM0
P3.2
AVref+
ADC0
ACD1
ACD2
RST
D1µP
P1.0
D3Segment
Driver
D2Eprom
Indic.LED
Drop-DetectorCircuit
Drop
ON/OFF-Key
EEPROM
Digit Select + Driver
Display-Rate ml/h
DisplayTotal ml
Displayml infused
InternalDrop Detector
ExternalDrop Detector
P4.0-4.3
WR
DataBus
AddressBus
P4.4
P1.3
Block schematics µVP5000
SCL
SDAPressureDetector
ACD3Pressure
+5V
PWM1
P3.7
Key
Display Print
AlarmLED
mvo 15.6.2001
Reset
Extern12V
10
111
1918171615121314
20 abcdefgh
30
7
8
9
10
11
3
6
49
13
12
67
68
1
59
23
22
25
24
62
6564
66
14 21
2015
8 19
187
176
5 16
14
1312
11
31
29
28
27
26
5
4
2
15
34
35
+5V
D4 9D3 8D6 7D5 6D0 5
D2 3D7 4
D1 2
18 319
16
17 5
4
615
54
12
13 8
7
55
56 911
22
48
57
47
10
1
202726 223
2421
25
4645444342414039A8
A9
+5V
51
52
53
50
+5V
1K0
R2
K9
+5V
0K
+5V
0K
+5V
C7
22pF
C8
22pF
C20,1µF
Reset
XTAL1
XTAL2
RST
VDD AVDD
61
C10,1µF
+5V
Q112MHz
R115x10K
Bl.3
Air Det. Test Bl.4
Air bubble Bl.4
Fail Bl.2
SM Home Bl. 9
Ak.Alarm Bl.4
SM Test Bl.2
SM-Clock Bl. 9
R115x10K
Power-hold Bl.7Hall Bl.4
Nurse call Bl.4
WDT Bl.2
SM Start/Stop Bl.9
I/O Key Bl. 7
Drop Bl.5
RXD Bl.4
TXD Bl.4
2.5 V Ref. Bl.3Ref.Ctrl Bl.3
18V Ctrl Bl.35V Ctrl Bl.3
PWM0
PWM1
P3.2
INT1
P3.4
P3.5P3.7
P4.4
P4.5
CMT1
P1.0
P1.1
P1.2
P1.3
P1.4
P1.5
P5.4P5.5
P5.7
RXD
TXD
SCL
SDA
A Vref+ADC0
ADC1
P4.6
D1P80C552EBA
ADC2
AVss Vss A Vref-
60 36,37 58
P4.0P4.1
P4.2
P4.3
STADC
EW
EA
WR
C61nF
R3
Key
Display-Print
21
6
5
4
3
10987
131211
14
X4
C50,1µF
Segment
D6D7D4D5D2D3D0D1
OECP
D3
74AC574
Gnd
Q6Q7Q4Q5Q2Q3Q0Q1
A12A13A14A15
A11A10
ALE
PSEN
A8A9
A12
CE
A11A10
OE
ALE
A13A14
A0O0A1O1A2O2A3O3A4O4A5O5A6O6A7O7
D2EPROM87C257
Gnd
Vcc
28
C4
0,1µF
14
Main PCB Microcontrollersheet 1
SCL
SDA
6
5D4
EEPROMST24C16M1
A1A2Vss
A0
7
1
2
3
4
VDD
+5V
8
4
3
9
11
24
23
22
2
1+5V
Pressure Bl.4
2
Current Bl.9
ADC3
P1.6
P1.7
63 P5.6
C30.1µF
mvo 15.6.2001
Vcc
nc
21
Display Bl.1
Display Bl.1 20
38
33
32
181615
19
Ext.1 Bl.6
Hall2 Bl.4
Press Test Bl.4
R1
R2
Vcc
RC2
Q2
A2
T2GndT1A1
B1
B2
Q1
RC1
D6HC4538
10K20
K
+5V
C933nF/1%
14,0 ms
17,5 ms
WDT Bl. 1
TP5
WDT
6
11
5
4 1 8
15
12
10
14
16
13
3
R4
SM TestBl. 1
R5
33nF/1%
C110,1µF
C10
18,5 ms
17,5 ms
10K
R12V2
BC846 B
R8
10K
V6BC856 B
100K
10K
R6
R9
2
1
V1IRFD 9120
10K
R13V3
BC846 B
K0 R14
3,4
2K7
1KO
R10
R10b
7K
+ 5V
V5SB 140
V4BC846 B
R11
9 8
D51/6 HC04
Bl. 1
Failure
R7
1R0
SM + 18 VBl. 9
Main PCB WatchdogSheet 2
+18V*Bl. 7
2
mvo 15.6.2001
Bl. 3Reset
0.5W
9
7
65D5
7
100K
0K
+18V*Bl. 7
R16
R17
1,64V18V* Ctrl
Bl.1
C120,1µF
0K R21
+5V +5V0K R22
+
7 8C17
0,1µF
TP4
ResetBl.1
4
Gnd
C16
0,1µF
C152,2µF
TP3
61
3
2
ResinSens. Vcc
Reset
N1
TL7705AD
CT
Vref
+5V
R23
R24C18
0.1µF
1.79V 5V CtrlBl.1
470R R18
1
2,3,6,7
8
V7TL431 C
+2.5V Ref.
+
C132,2µF
C140.1µF
R207K7K
+1,25V Ref.Ctr.Bl.1
Drop det.Bl.5
2.5V Ref.Bl.1
TP6
Main PCB Peripheral 1Sheet 3
R19
mvo 15.6.2001
ResetBl.2
5
Reset
0K R15
+5V
7K5K
+5V
1
2
3
4
Airdetector
80msec
Air Det. TestBl. 1
Air bubbleBl. 1
+5V
X6
R26
+22V*
Nurse call
X12
13
9
11
+5V
121310K
1/6 HC0414
D5
7V9
BC846B
Nurse callBl.1
+5V
V28N4148
3
8 K1V23026
10
1
5
TXD Bl.16
5
Vin
Gnd
+5V
8
I+
I-4
D8
HCPL-4200
X12
36
8
10
12
+5V
8
5
6
7
Vo
VE
GndI-
I+D9
HCPL-4100 Transmitter
20 mA
Current Loop
Interface
Receiver
RXD Bl. 1
Main PCB Peripheral 2 (2.06)Sheet 4
1
2
3
4
PressureSensor
+18V*Bl. 7
X11
Pressure Bl.1
1
2
3
4
Hall Sensor
X1
HallBl.1
+5V
mvo 6.9.2001
1210K
H1KPE208H
V8
BC846B
C190,1µF
R25Ak.AlarmBl.1
1
2
X15
Vcc
Vcc
2
1
C470.1µF
+5V
C480.1µF
7 1 2
5
7
14
4 3
R100
560
Ohm
+16V* / 12 mA
7
6
0K
K7
00
7K
+
3M3
10K
100K
0K
K7
0K
1M0
M0
Ref. + 2,5VBl.3
+5V
R28
R27Built-indrop-detector X8
1
4
2
3
+0,16V
V12BC846B
V11BC846B
X12
externaldrop-detector
R34 R35
+5V
R36
+2,6V
1
4
3
2
+1,85V
C2022µF
R30
C21
C22
0,1µF
R33
R32
R31
+1,7V
+0,15V
R37
0,1µF
N2TLC271
+ 5V
C230,1µF
R382
3
4
11
4
TP7
3 5 13 16
+ 5V
121 8 15
A1
D7
HC4538
T1 Gnd A2 T2
Q1
RC1
7
2
CI1 B1 CI2 VccB2T = 134 ms
C240,1µF
C250,1µF
DropsDrop Bl.1
R29
Main PCB Drop-detectorSheet 5
V13BC846B
+
-
mvo 15.6.2001
8
1K0
R30b
C21b0,1µF
1
5
6
10
9
14
N
VAC
A
Ph
X1
F2
1
X2
23
T1 10
15V
11
1
2F1
+-
~
~
V14B80C1500
0,8Ap
R43
C261000µF
(35V)
5K6 +
C270,1µF
V15LL4002
V17SB140
+18VBl. 7
R39 R40
180180
180 180 26
25
X4
Display-print
R41 R42
+
-
Ni-Cd /Ni Hydr.battery
1 2 V1.2 …1,7Ah
X31
4
Main PCB Transformer (2.03/AC115)Sheet 6
mvo 15.6.2001
X131
3
V17b
SB140
12V external
F31AT/250V
L22*27µH
F41AT/250V
22
3
LED
GND
VS
Test
Vc
shunt
sense
osc
1
2
3
4
8
7
6
5
R4515 Ω
30K
C502.2nF
10Ω
C51100µF
V29HLMP1503
R44
R68 D1
2U
2403
BF
P
14h
Charge current:100 + 45 mA
4 Ext.1 Bl1
V30BZW04-5V58
trickel current:45 mA
R70220Ω
V31HLMP1400
2Ω R71
(45mA)
22…30V
for batt. charge:22…30V
black
grey
orange
230V
240…260V
+18 VBl. 6
DisplayprintI/O Key
0 K
R46 5K6
R47 0K R48
C28
22 µF
C2922 µF
+
+
2 3,4
1
17 100
X4R50
V21
TL431C
81
2,3,6,7
V19
LL4148
V20LL4148
+5V
R53R54
1KO
10K
I/O Key
Bl. 1
10K R49
10K
R51
R52
8K22 1
D5
1/6 HC04
Power-HoldBl. 1
+18V*Bl. 2Bl. 3Bl. 4Bl. 8
Main PCB ON/OFFSheet 7
V18
IRFD9120
mvo 15.6.2001
+18V*Bl.7 1R0
R55
0,5W
C300,1µF
C31100µF
C320,1µF
C330,1µF
15KR56
R57 K7
C353,3 nF
f ca. 100 kHz
1
3
5 4 6
7In
Comp.
N3
L4960
RC Gnd Start
Sense
Out
C3422 nF
C362,2 µF
L125 µH
1
2
V22SB140
2
+
+
C370,1µF
C381000µF
+V23
BZW04-5V5
+5V / 0,35A
Main PCB DC-ConverterSheet 8
J1
mvo 15.6.2001
V23b
BZW04-5V5
0K
27K
+5V / 72 mA
14,5kHzR58
10K
K0
C393n3 16 12
Osc Vs11
19
17
Control
Hall/Full
CW / CCW
18 CISM ClockBl. 1
SMStart / Stop
D51/6 HC04
11 10
20En
Reset+5V
Bl. 1
10
SM Home4 3
R59
3Home
D51/6 HC04
Bl. 1
15R62
C440,1µF R63
+5V VrefGnd
2
S1
+
4A
B6
C 7
D9
Inh15
Inh28
S213
14
1K0
R2
R2
1K0
R6417
14
C461nF
C45 R67
R667
4
1nF
R65
C40100µF
C410,1µF
20
Vss
10
Vs
+C420,1µF
C43100µF
SM +18V / 100 mABl. 2
2
9
12
1
11
InA
InB
InC
InD
Inh1
Inh2
19
SD
SC
SA
SB
Gnd
5,6,15,16
Out13
Out2
Out3
Out4
8
13
18
M 1
V24EDF1
4
2 1
3
1
2
3
4
X10
sz/w
o
sz
EDF1V25
12
3
4
6
5
7
8
o/w
gb/w
r/w
r
gb
Main PCB step motorSheet 9
D10L297
D11L293E
10K
R6110K
R60
+5V
V26BC856BCurrent
Bl. 1
mvo 15.6.2001
C492,2nF
1
Nurse call
X5
1
2
4
X7
1
2
3
4
5
6
7
8
Transmitter
20 mA
Current Loop
Interface
Receiver
3
X9
Externaldropdetector
1
2
4
X14
3
1
2
4
6
8
10
12
14
13
11
3
Connector PCBSheet 10 mvo 15.6.2001
9
1
2
3
4
X1
R1
+5V
Hall
N12
UGN3020T
0K
Hall sensorSheet 11
mvo 15.6.2001
X5
2
1
4
+5V
3
8
K1
10
1
5
Gezeichnete Stellung: Pumpe im Alarm
Nurse callSheet 12 mvo 15.6.2001
S1
74HC154
D1
Aa
b
c
d
e
f
g
dp
V1
S2
D2
Aa
b
c
d
e
f
g
dp
V2
S3
D3
Aa
b
c
d
e
f
g
dp
V3
S4
D4
Aa
b
c
d
e
f
g
dp
V4
S8
D8
Aa
b
c
d
e
f
g
dp
V8
S9
D9
Aa
b
c
d
e
f
g
dp
V9
S10
D10
Aa
b
c
d
e
f
g
dp
V10
S11
D11
Aa
b
c
d
e
f
g
dp
V11
S5
D5
Aa
b
c
d
e
f
g
dp
V5
S6
D6
Aa
b
c
d
e
f
g
dp
V6
S7
D7
Aa
b
c
d
e
f
g
dp
V7
S12
V12
D12A
C
S13
V13
S14
V14
S15
V15
S16
V16
UDN2580 UDN2580
S17
abcd 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
I0 I1 I2 I3 I4 I5 I6 I7
O0 O1 O2 O3 O4 O5 O6 O7
D26
AC
3456
7
8
9
10
11
12
13
14
25
26
1,2
22,23,24
+
C1 C2
C3
C4C5
R1 R410K
10K R5
R2R3
DisplaySheet 13
mvo 15.6.2001
21
M1
M2 M3O0 O1 O2 O3 O4 O5 O6 O7
I0 I1 I2 I3 I4 I5 I6 I7
17
G1 G2
+
C6
+5V
+5V
+5V
+5V
UD
N2
59
5A
GNDVs
C7
I0
I1
I2
I3
I4
I5
I6
I7
O0
O1
O2
O3
O4
O5
O6
O7
M4
D13A
C
D14A
C
D15A
C
D16A
C
D17A
C
D18A
C
D19A
C
D20A
C
D21A
C
D22A
C
D23A
C
Component layout
Batterie
X4
Topography main PCB
X12 X8
X11X3X2X13
EpromµP
InterfaceRelay Buzzer
Drop Det.
Motor Cont.
X10
Seg. Driver
Reset
EEprom
WatchDog
AC/DC
X6
X1
Stepping motor
Hall sensor
TransformerBattery
Air bubble detector
Pressure sensor
Display PCBDrop detectorConnector PCB
Option12 V
Component layout Main PCB µVP5000
Version 2.05 / 7.11.00
Component layout Plug PCB µVP5000
Version 1.11 / 15.11.95
Component layout Display PCB µVP5000
Version 1.11 / 27.6.96