65
Sensor and Simulation Notes Note 257 October 1976 . A Study of Charge Probe Response ona Junction of Unequal Radii Thin Wires Santiago Singarayar B. M. Duff University of Mississippi CLEARED FORPil&LICRELEME p@39 5%+97 Abstract The experimental measurement of charge density induced on conducting surfaces often makes use of a short monopole as the charge probe. Measure- ments on structures containing junctions of unequal radii cylinders have been hampered by incomplete knowledge of the probe characteristics. An , investigation of charge probe response as related to the radius of the cylinder on which it is mounted is reported here. An experimental determin- ation of the ratio of probe response when mounted on two different radii cylinders has been conducted using a coaxial line geometry with a step change in radius of the inner conductor. In addition, a theoretical solution in numerical computations of the probe response in relation to the radius of the cylinder upon which it is mounted has been conducted. The results of these studies are presented in the following report.

Sensor and Simulation Notes Note 257ece-research.unm.edu/summa///notes/SSN/note257.pdfSensor and Simulation Notes Note 257 October 1976 . A Study of Charge Probe Response ona Junction

  • Upload
    others

  • View
    5

  • Download
    0

Embed Size (px)

Citation preview

  • Sensor and Simulation Notes

    Note 257

    October 1976 .

    A Study of Charge Probe Response ona Junctionof Unequal Radii Thin Wires

    Santiago SingarayarB. M. Duff

    University of Mississippi CLEAREDFORPil&LICRELEME

    p@39 5%+97

    Abstract

    The experimental measurement of charge density induced on conducting

    surfaces often makes use of a short monopole as the charge probe. Measure-

    ments on structures containing junctions of unequal radii cylinders have

    been hampered by incomplete knowledge of the probe characteristics. An,

    investigation of charge probe response as related to the radius of the

    cylinder on which it is mounted is reported here. An experimental determin-

    ation of the ratio of probe response when mounted on two different radii

    cylinders has been conducted using a coaxial line geometry with a step

    change in radius of the inner conductor. In addition, a theoretical solution

    in numerical computations of the probe response in relation to the radius of

    the cylinder upon which it is mounted has been conducted. The results of

    these studies are presented in the following report.

  • o

    TABLE OF CONTENTSN

    LIST OF FIGURES . . . . . . . . . . . . . . . . . . . . . . . . .“.3~

    CHAPTERw

    1. INTRODUCTION . . . . . . . . . . . . . . . . . . . . . ...5

    II. MEASUREMENT TECHNIQUE . . . . . . . . . . . . . . . . . . . @*

    lntroduc.tLon . . . . . . . . . . . . . . . . . . . . . . 8

    Analysis of Coaxial Transmission Line . . . . . . . . . . 12

    Description of the Measurement System . . . . . . . . . . 1.6

    Determination of Probe Calibration Factor . . . . . . . . 21

    Curve Fitting: Modified Prony’s Method . . . . . . . . . 2~”

    111. NUMERICAL ANALYSIS. . . . . . . . . . . . . . . . . . . . .28

    Charge Probe as a Scatterer: Theory and Solution . . . . 28 m

    Formulation of Straight Wire Integral Equation . . , . . 29

    Charge Probewith50flLoad . . . . . . . . . . . . . ..32

    Numerical Solukion Technique . . . . . . . . . . . . . . 33

    Expansion of Current and Formulation of Matrix . . . . . 35

    Determination of Probe Calibration Factor , . . . . . . . 39

    IV. PRESENTATION OF DATA AND CONCLUSION . . . . . . . . . . . .42

    Introduction . . . . . . . . . . . . . . . . .’. . . . .4z

    Measured Data for the Charge Distribution . . . . . . . .42

    Probe Calibration Factor . . . . . . . . . . . . . . . .~2

    Conclusions. . . . . ... . . . . . . . . . . . . . . . . .64

    2

    1

  • oLIST OF FIGURES

    FIGURE*

    2.1

    8 2,2

    2.3

    2.4

    2..5

    2.6

    3.1

    3.2

    3.3

    3.4

    3.5

    3.6

    4.1

    4.2

    4.3

    4.4

    4*5

    4.6

    Charge Probe . . . . . . . . . . . . . . . . . . . . . . . . .

    I/p Variation as a function of (p-a) . . . . . . . . . . . . .

    Coaxial System with Step Radius System . . . . . . . . . . . .

    (a) Zero Order . . . . . . . . . . . . . . . . . . . . . . .

    (b) Approximate actual distributions of Charge per unit

    length and Radial E field in Coaxial Line . . . . . . . .

    Coaxial Transmission Line . . . . . . . . . . . . . . . .

    Block Diagram of Measurement System . . . . . . . . . . .

    Charge Probe on a Charged Conductor . . . . * . . . . . .

    Charge ProbeasaScatterer . . . . . . . . . . . . . . .

    Lumped LoadZ1oadati= 0 . . . . . . . .. . . . . . . . .

    Triangles in Piecewise Linear Testing Scheme . . . . . . .

    Pulse Arrangement for Approximating current on Straight

    Wire. . . . . . . . . . . . (l. . . . . . . . . . . . . .

    Relative locations of-Pulses (Bases Set) and

    Triangles (Testing Set) on Straight Wire . . . . . . . . .

    Maximum Charge Condition - Probe Height 0.10”, al = 5/16”

    Medium Charge Condition - Probe Height 0.10”, al = 5/16” .

    Minimum Charge Condition - Probe Height 0.10”, al = 5/16”

    Max5.mumCharge Condition - Probe Height 0.08”, al = 5/16”

    Medium Charge Condition Probe Height 0.08”, al = 5/16” . .

    Minimum Ch&rge Condition - Probe Height 0.08”, al = 5/16”

    . .

    . .

    . .

    . .

    . .

    . .

    . .

    . .

    . .

    . .

    ● ☛

    ✎ ✎

    ✎ ✎

    ✎✌

    PAGE

    9

    11

    13

    14

    14

    15

    27

    30

    31

    31

    34

    38

    40

    43

    44

    45

    46

    47

    48

    3

  • -!- =

    .

    FIGURE

    4.7

    4,8

    4.9

    4.10

    4.11

    4.12

    4.13

    4.14

    4.15

    4.16

    4.17

    4.18

    4.19

    4.20

    Maximum Charge Condition - Probe Height 0.06”, al = 5/16” .

    Medium Charge Condition - Probe Height 0.06”, al = 5/16?’. .

    Minimum Charge Condition - Probe Height 0.06”, al = 5/16” .

    Maximum Charge Condition - Probe Height 0.10”, al = 3/161’ .

    Medium Charge Condition - Probe Height 0.10”, al = 3/16”. .

    Minimum Charge Condition - Probe Height O.10~’,al = 3/16” .

    Maximum Charge Condition - Probe Height 0.08”, al = 3/16” .

    Medium Charge Conditicm - Probe Height 0.08”, al = 3/16”. .

    Minimum Charge Condition - Probe Height 0.08”, al = 3/16” .

    Maximum Charge Condition - Probe Height 0.06”, al = 3/16” .

    Medium Charge Condition - Probe Height 0.06”, al = 3/16!’. ,

    Minimum Charge Condition - Probe Height 0,06”, al = 3/16” .

    Probe Calibration Factor for 5/16’~to 1/8” ratio of radii .

    Probe Calibration Factor for 3/16” to 1/8” ratio of radii .

    PAGE

    49

    50

    51

    52

    53

    54

    55

    56

    57

    58

    59

    60

    61

    63

    .

  • o

    0

    The

    antennas

    standing

    measurement of

    and scatterers

    Chaptez I

    INTRODUCTION

    the current and

    is important in

    charge distributions on

    order to gain an under-

    of their electromagnetic behavior and to aid in developing

    and checking of theoretical solutions, The basic techniques for\

    making these measurements have been established for some time and

    used by many investigators. However, during a recent experimental

    project, to be described below it became obvious that more accurate

    knowledge of the response characteristics of charge probes was

    necessary. The resulting investigat-ionof charge probe response

    forms the basis of this thesis,

    The experimental project which leadsto the investigat~on re-

    ported here, concerned the measurement of

    induced on structures containing junction

    radii. The purpose of that investigation

    charge distribution

    of thin wires of unequal

    was to experimentally

    determine the appropriate condition on charge which should

    applied in theoretical solution of this type of structure.

    A theoretical analysis and discussion of the junction

    be

    has been

    presented by T.T. Wu and R.W.P. King [3]. The probe used for the

    measurement of charge distribution consisted of.a very short mono-.

    pole mounted on the surface of.the cylindrical structure. For

    the experimental determinati.onof the charge-condition at the

    5

  • -junction,an accurate knowledge of the probe response as a function

    of the radius of the cylinder upon which it is mounted is necessary.

    A charge probe mounted on a cylindrical structure has been analysed

    theoretically by R.W.P. King [1]

    However, these investigations do

    solutions for the probe response

    and experimentally by Whiteside [2].

    not provide sufficiently accurate

    as a function of the radius of the

    cylindrical stru~tures. An independent experimental and theo~’etical

    investigation of the charge probe response has been conducted, the

    results of which are presented in this thesis.

    Chapter II deals with the measurement technique by which the

    probe calibration factor can be determined. This probe calibration

    factor has been determined for 3 different probe heights for 2

    junctions of thin wires of 5/16” to 1/8” radii and 3/16” to 1/8”

    radii. To make sure that the probe calibration factor is the same

    for all charge conditions at the junction, the probe calibration

    factor has been determined for 3 different charge conditions namely

    ,

    8

    maximum, minimum, and a condition in between maximum

    The experimental data was analysed by using modified.

    A numerical solution for the current induced in

    and minimum.

    Prony’s method.

    the probe

    has been carried out using moment method techniques [4,5]. The

    analysis models the probe as a monopole receiving antenna perpen-

    dicular to an infinite plane condcctor, but includes the incident

    ~ field variation which is to be expected when the probe is mounted

    on a cylinder. Chapter 111 deals with the methods used for this

    numerical analysis.

    6

  • Finally, in Chapter IV, the experimental data, the results and

    the conclusion are presented.

    w

    .

  • CHAPTER 11

    MEASUREMENT TECHNIQUE

    (i) Introduction

    The measurement of the surface charge density on a conductor may

    be related to ameasurementof ~ field by using the boundary condition,

    fi.cE = rjo

    (2.1)

    where &o

    is the relative permittivity, 6 is the unit normal vector to.

    tfiesurface of

    surface charge

    charge density

    the test structure under consideration and q is the

    density. It is therefore possible to obtain the surface

    by using a sensor whose output is proportional to the

    normal component of the E field at the surface. The most common method

    uses a short monopole receiving antenna, as shown in Fig. (2.l),which

    is perpendicular to the surface of the test structure. The signal

    induced in the load is however related to the ~ field along the entire

    length of this monopole rather than simply the field at the surface.

    If the normal component of the ~ field is constant over the length of

    the probe there is no problem. But generally this is not the case.

    In order to relate the monopole probe response to the charge density

    on the test structure it is therefore necessary to have some knowledge

    of the manner in which ~ varies over a short distance away from the

    surface.

    .

    8

  • w

    ‘8 f

    ;.

    Coaxial cable

    -.

    f 1

    --L_+--—.—. .—.! .—. —. —.—. -(!9* .—.—I*(a) Probe carriage for smaller radius tube.

    a *

    .—— —7T’—

    . — . —. —.— . —.t“ ‘—1t1

    .

    (b) Probe carriage for

    —.—. e ,’.—— .1 .—1!Ilarger radius tube.

    Fig. (2.1) CHARGE PROBE.

    I

  • The present work considers only test structures composed of

    electrically thin conducting cylinders. The field produced by a

    1uniformly charged infinite cylinder is known to vary as —. If the

    P

    test structure (a cylindrical antenna) is of electrically thin radius

    and the probe is electrically short, then ~ over the length of the

    probe is controlled primarily by the local charge distribution. The

    field then becomes.essentially the same as that of an infinite cylinder

    and ~ is very nearly

    near discontinuicies

    of this variation on

    1proportional to –. This assumption is not valid

    P

    of the conducting surface, however. The effect

    the response of the probe depends on the radius

    of the test structure. To illustrate this effect, Fig. (2.2) presents

    ~plotted as a function of (p-a) for three different cylinder radii,

    For a given probe height, the variation of EP

    over the length of the

    probe is less when the probe is mounted on a large radius cylinder.

    The effect of the: variation ntaybe reduced by reducing the

    probe height. However, the signal induced across the load is also

    reduced and so the signal to noise ratio. A second effect of changing

    the radius of the test structure is that the image plane for the

    monopole probe is of different radius of curvature. The fact that the

    charge probe is of finite size and it responds to fields away from the

    suiface of the test structure makes it necessary to determine a probe

    calibration factor as a function of cylinder radius in an exact manner.

    To determine the probe calibration factor experimentally, a system

    containing cylinders of different radii and for which the variation of

    the charge density as a function of cylinder radius can be predicted

    o

  • t . *

    e

    ,. wP (b

    0.04 0.06 0.08 0.10 0.120.0 0.02

    Fig. (2.2) ; VARIATION AS A

    (p-a) in inches.

    FUNCTION OF (p-a).

    1’

    I

  • is needed. The system chosen was a coaxial line with step change in

    the radius of the inner conductor. This system is shown in Fig. (2.3)

    and has been analysed by Marcuvitz[6]. The TEM mode ~ field dis-

    tribution of the above system is shown in Fig. (2.4). The ~ field is*

    uniform and proportional to ~ inside the coaxial line except near theP

    junction. The local effect near the junction can be replaced by a

    shunt capacitance.so that the Kirchoff~s current law is satisf~ed.

    At the junctton the voltage is continuous making it easy to analyse

    the charge condition at the step in the following way.

    (ii) Analysis of Coaxial Transmission Line

    Fig. (2.5) illustrates a coaxial transmission line with a TEM

    mode propagation. In cylindrical coordinates the two dimensional

    Laplace equation is,

    1 32@~*(p?&=-= oP a~2

    For a potential function independent of angular coordinate+,

    eq. (2.2) reduces to

    The solution of equation (2.3) may be expressed as,

    (2.2)

    (2.3)

    -(2.4)

    Imposing the bo!.mdaryconditions @ = VO at p = a and @ = O at p = b,

    eq. (2.3) yields the final solution for the potential as,*

    12

  • ● 1

    e

    1,,

    I

    PLJ

    .

    4I P’

    Charge Probe I $dvable Short “1a ( .--, --- ---I

    II

    a

    1---- ---- ---

    I 4/

    iJ _ .If.—.—.—.—-—.—“z

    I----- ---- -- ---,I ------ ---

    / 4

    Step radius system/ I Coaxial system ~

    Ito#

    t1

    I,,

    ;1 ,1

    1’

    Fig. (2.3) COAXIAL SYSTEM WITH STEP RADIUS SYSTEM.

  • F.b

    m“

    ------- .

    ---- ---

    .— --- -

    -------- --

    ------ --

    -“

    ------ -

    .z----.----

    .

    r

    .

    ---’--------—

    -----------

    ---------- --

    ----- ----”

    ---- ---- -

    +n>

    4.

    t

    +

    +

    +4.

    +

    +

    +

    +

    +—

    ,------ ---

    ----- --

    -.-...+_ --

    ----% ---

    -------

    ..-.-------m-------

    -. . .------ -

    --‘. ------ -

    -- --~------

    -------- ----

    ----- ----

    [

    r

    ------ ---

    ----- --- -.----- -------- ___

    ---—- ------- ---------- -_-,----- -------- -------- --—_ .

    1

    Iv!-cr~”

    I

    _____ ____

    —--- --- ---

    ------- ----

    --- ------ _.

    ---- - ----

    1I

    i1

    1IIIII1

    .a>.!

    I

    I1III

    III

    ;T

    *NN

    0w ;

    P.-—-- ---

    ----- -------- --

    ----- --~

    --- -— --”

    ---- —-

    ---- ----

    -- -----

    ----- ---

    ---- --- -.- —- --- ----

    ----- ----

    ---- -----

    ----- ---- ---

    ---- ------ -

    N&nN g

    ri

    -,,, ,

    i

  • ----

    ---

    ,)-

    k field

    ~ field_—

    Fig. (2.5) Coaxial transmission line.

  • an associated charge density which differs from that of a TEM mode as

    indicated in Fig. (2.4)(b).

    The voltage

    different radii,

    at the junction.

    the junction are

    is continuous at the junction but because of the

    a difference in the magnitude of charge density exists

    The ratios of linear and surface charge densities at

    given by L

    L .LL1 L,-=— .

    (iii) Description of the Measurement System

    A block diagram of the measurement system

    (2.9)(a)

    (2.9)(b)

    .

    .

    is given in Fig. (2.6).”

    The commercial instruments used in the system are given in the block

    diagram itself.

    The power at 300 MHz is fed to the amplifier. The amplifier sig-

    nal is fed to the coaxial system through a balanced feed to avoid

    higher order mode propagation. Matching of the amplifier output to

    the coaxial system is done through a single stub tuner. The reference

    signal to the vector voltmeter reference channel is taken from the

    amplifier outpu~ with proper attenuation. The signal to the frequency

    meter is also is taken from the amplifier output with proper attenua-

    tion. The signal from the charge probe is fed to the test channel

    16

  • o

    ~ field of TEM mode in +Z direction,

    where k is the propagation constant.

    Using eq. (2.5) ~ may be expressed as,

    The surface density of

    at z=O may be obtained

    eq. (2.1) as,

    ‘O 2Pe-jkz——In(&) p

    al

    chafge ?IIon an inner conductor

    from eq. (2.7) and the boundary

    (2.5)

    (2.6)

    (2.7)

    of radius a,,1.

    condition

    (2.8)

    -.Equation (2.8) can be written as,

    q~ = c1‘o

    where Cl is the capacitance per unit length and ql = 2na1Tllis the

    linear charge density. The same.relationship applies to the inner

    conductor of radius a2“

    It must be remembered that the above analysis assumes that only

    TEM mode fields exi’stand that the junction is represented by a lumped

    shunt.capacitance. The effect of the step change of radius is actually

    distributed.over some short distance on either side of–the junction with.

    #

    17

  • of the vector voltmeter. The power level is monitored by the broad-

    band sampling meter to compensate for the drift in power level.

    The magnitude or the phase of the signal is switched to the.

    digital voltmeter by means of the relay actuator. The calculator

    samples this output 5 to 10 times and takes the average to minimize .

    the effect of the minute-fluctuations in the probe signal. The magni-

    tude and phase quantities are stored in the calculator for lat~r

    plotting and for further use.

    The stepped radius coaxial line was

    brass tubing. The three different inner

    constructed from cylindrical

    conductor radii 0.3125’!

    (0.79375 cm), 0.1875’’(0.47625cm), and 0.125’’(0.3175cm) were selected

    as being available readily. Two inner conductor systems were con-

    structed one composed of 0.3125T’and 0.1251’radius tubing, the other

    of 0.1875” and 0.125’!radius tubing.

    was used as the outer conductor. The

    conductor was filled with styrofoam.

    A tube of 1.0” inside diameter

    space between inner and outer

    A slot of 1/19’’(0.13368cm) was

    cut axially in the inner conductor tubing to allow movement of the

    charge probe along the structure. The probe carriage was designed

    to slide inside of the inner conductor tube and extend through the

    slot to the surface of the tube so that the structure is electrically

    continuous. The charge probe was constructed on a probe carriage

    using a semi rigid coaxial cable. The charge measurement was con-

    ducted on two different radii tubes. If two different probes had been

    used it would have been necessary that they be constructed identically.

    .

    0

    .-

  • @Othe~ise, it would have become necessary to calibrate them indepen-

    dently. To avoid this problem, a single charge probe with two different

    probe carriages was used. The wall thickness of the tubes had to be

    taken into consideration while making the probe carriage. For example.

    for the 0.3125” radius tube wall thickness was 0.06’’(0.1524cm) and the

    0.125” radius tube was 0.04”. The probe carriage fit exactly inside.

    of the tube and through the slot to the surface so that the charge

    probe was effectively mounted on the surface of the inner conductor

    tube. The cable used for making .the probe was UT 20 manufactured by

    Uniform Tubes, Inc., with the following specifications.

    Impedance: 50 Ohms

    o

    Outside diameter: 0.023’’(0.05842 cm)

    Center conductor diameter: 0.0045’’(0.011.43cm)

    The outer–conductor of–the cable was soldered to

    cut at the surface. The dielectric insulator on

    was removed so that the inner conductor extended

    the probe carriage and

    the inner conductor

    beyond the surface to

    form the monopole antenna. It was necessary that this monopole be

    straight and perpendicular to the test structure. The signal from the

    probe was carried by a coaxial cable contained inside of the inner

    conductor tube. Care was taken to see that stray pickup did not affect

    the signal by shielding the signal cables. A 1/8” diameter tube

    attached to the probe carriage was used both to protect the signal

    cable and to provide a means of positioning the probe.

  • aThe

    actual E

    measured

    charge density measurement was made at intervals of 1 cm. The

    field distribution near the step due to local effect made the

    result unreliable in this region. Moreover the charge probe

    was mounted at the center of the probe carriage which made it impos-

    sible to reach the step in the large radius tube. For analysis of the

    data, afew data points were eliminated near the junction to eliminate

    the local effect and the data was analysed by using a modofied Pronyfs

    method to fit curves to the TEM mode distribution. This procedure

    will be discussed separately.

    Measurements were made for probe heights of 0.10’’(0.2540cm) to

    0.06’’(0.1524cm) at an interval of 0.02’’(0.0508cm). The charge

    condition at the junction was varied by using the movable short and

    the experiment was carried out for three charge conditions at the

    junction, namely charge maximum, charge minimum and a case in between

    maximum and minimum. The experiment was repeated for consistency.

    Even though frequency drift was observed, the change in electric length

    of the test structure was of only of the order of ~ 0.00IA which was

    considered negligible. Any change in the power output

    later was

    broadband

    Care

    compensated for

    sampling meter.

    w’astaken while

    probe carriage fits flush

    by monitoring the oscillator

    making the probe to see that

    with the test structure. To

    of the oscil-

    output using a

    the top of the

    achieve good

    elect~ical contact between the probe carriage and the test structure,

    .

    0

    silver conducting grease (Eccoshield, manufactured by Emerson & Cuming)

    20

  • o was applied along the sides of the probe carriage and the sidewallsof the probe. The signal carrying probe tube was coated with an

    absorber (Eccosorb by Emerson & Cuming) to avoid the possibility that

    . the probe tube would become the center conductor of a TEM structure

    inside of the inner conductor of the test structure and give rise to

    .propagation and unwanted resonances.

    After obtaining the extrapolated value of the probe reading at the

    junction the probe calibration factor was obtained as described in the

    following section.

    (iv) Determination of Probe Calibration Factor

    Let PI and P2 be the probe reading obtained by extrapolating the

    measured data to the junction of–the large and small tubes of radii a1

    0

    -.

    *

    and a .2

    Let N1 and N2 be the normalizing factors.

    ‘1 = Nlrj

    ‘2= N2?12

    therefore

    ~1 = ‘1 ‘2.—~ ‘2 ‘1

    nlfrom eq. (2,9)(b) the ratio of— is given by,

    ~2

    a2—

    al

    (2.10)

    (2.11)

    21

  • N,‘HIe

    (v)

    probe calibration factor, $ is then given by,QL

    Curve Fitting: Modified Prony~s Method

    ,

    (2.12)

    The experimental data obtained has magnitude and phase so that it

    is a complex quantity. Curve fitting to the experimental data is done

    by using a modified Prony’s method. Prony’s method [7] fits a set of

    data to a series of exponential. By assuming a function of the form

    f(z) = Aleyz + A2e-yz

    because of the TEM mode transmission,

    the exponential series of the general

    Y= (a +

    only two terms

    Prony’s method

    j $)

    (2.13)

    are selected from

    with

    (2.14)

    .

    .

    .

    where a is the attenuation constant, B is the propagation constant.

    = A e(a+j6)z+A2e-(a+j6)’f(z) ~ (2.15)

    where Al and A are complex constants, A2

    ~, A2, a, 6 are to be determined.

    Let Az be the spacing between the data points and

    ~yAz‘1 =

    (~.16)

    -yA!z .

    ‘2=e (2.17)

    220

  • then

    Assume U1 and U2 are

    f ~+l(!dz) = A1u1k+A2u2k

    two roots of

    U2+(%U+UO=01

    where aoand a

    1are unknown.

    Equation (2.18) can be expanded to

    ‘1=A1+A2

    ‘2 = ‘cl 4“A2U2

    2 2

    ‘3 = ‘I”l ‘-‘2U2

    ‘k= A1ulk--l+A2U;-1

    (2.18)

    (2.19)

    (2.18)(a)

    (2.18)(b)

    (2.18)(c)

    (2.18)(d)

    Since the above set of equations satisfy equation (2.19)

    ‘3 + ‘2al ‘E‘Iao =o

    ‘4 + ‘3al ‘p‘2ao = 0

    The unknowns an, a, can be determined from the above equations. Onceu L

    a.‘ al are‘nom’the roots Ul, U2 can be determined from equations(2.19). Once U1 and U2 are known, y may be obtained from equations

    (2.16) or (2.17) or,

    Y = ln(ul)/Az

    -y = kn(u2)/Az

    23

    (2.20)

    (2.21)

  • With y known Al and A2

    Since the roots U1 and

    stituting in (2.19) we

    are then determined from (2.18}(a) and (2.18)(b).

    1

    %!are reciprocal by definition, U2 = ~ . Sub-1get

    2‘1 ‘W+ao= 0

    ~2+ct ~+ao=o1U2

    ‘2

    J.

    (2.22)

    (2.23)21 + alu2 ‘%U2 = 0

    Comparing (2.19j and (2.23) we obtain,

    ao=l

    U*+alu+l= o (2.24)

    If we determine al, then the roots can be found easily in the following

    way,

    fl+f2~1+f3=o ; f2a1 = -(fl + f3)

    f2+f3c11+f4=o ; ‘3% =-(f2 + f4j

    ‘3+f4%+f5=0; f4cx1= -(f3 + f5)

    With three datapoints one can easily determine al from which U1 and U2

    can be determined.

    The method descrj>ed so far is not satisfactory for the set of

    data’where experimental noise is present. A larger number of datapoints

    24 ●

  • used and a least square curve fitting technique isis employed.

    3M=N, where NLet MAz be the spacing between datapoints. When

    is —,.th_etotal_nurnber_ofdata_po_ints,

    6f ti+f = om+l 2m+l

    fl +

    ‘2 +

    f3 -1-

    fm +

    ; fm+la=-(fl+ f2m+1)

    .

    ; f~+2~ =-(f2 + f )2m+2f~,a + f

    2m+2 =o

    fm+3‘+f2m+3=0 ; ‘m+3

    ~ =-(f3 + f )2m+3

    f2ma -1-f3m = o .$ f2mcl=-(fm + f3m).,

    In general

    [F]cx= -(G)

    .[FTF]cx= -[FT] [G] whereo

    [F’] is the transpose matrix of-[F]

    Therefore u = - [FTF]-l [FT] [G] (2.25)

    The residues Al and A2 can be determined in the following way

    ‘1 = Al -1-A2

    A ;yAz ●1

    A e-yAz2‘2 =

    i

    A1(eyAz)N-l +A2(e-yAz)N-lfN =

    [F] = [G] [A]

    [A] = [GTG]-l [F][GT] (2.26)

    25

    0

    .-

  • with Al, A2, Y1 and Y2 thus determined, a continuous curve may then

    be computed from equation (2.15) and extrapolated to the junction. The

    value of P and P in equation (2.12) were obtained by this procedure.1 2 .

    a

    .

    .

    26

  • :

    . ‘1

    ‘1

    I‘1,!II,:

    1’,,

    II Nd

    ,’,,

    II ,/

    nFREQUENCY r SINGLEMETER STUBHP 5345A TUNERBROADBANDSAMPLINGVOLTMETER ~HP 3406A1

    50L?TEEf

    1 5ofJ

    r--

    HP 778A DUALDIRECTIONAL COUPLE

    OSCILLATOR AMPLIFIER

    GR 1362 BOONTON 230A ---- -. ---- -.

    i 4 50QTEE

    =4MAG.

    VECTORVOLTMETERHP 8405A

    H&E

    I I & 50Q I4 1

    11% I FEEDS CHARGE PROBE

    1AII f

    REI?.

    PEST.1

    poWER

    DIVIDERa- ! COAXIAL MOVABLE SHORr7

    ~1’STVM

    ‘3A -E!ZI”FFig. (2.6) BLOCK DIAGW OF MEASUREMENT SYSTEM.

    “uPLOTTERHP 9862A

  • #

    CHAPTER 111

    NUMERICAL ANALYSIS

    .

    (i) Charge Probe as a Scatterer: Theory and Solution

    The preceding chapter discussed the experimental determination of

    the response of a charge probe. A theoretical analysis and numerical

    solution for the probe response and calibration factor are presented

    in the present chapter. For the purpose of this analysis, the probe

    is modeled as a monopole scatterer perpendicular to an infinite plane

    conductor and planar

    Chapter II the major

    the probe is mounted

    image theory is employed.

    effect of the cylindrical

    1is the ~ variation of the

    As discussed in

    conductor upon which

    ~ field produced byIJ

    the charge on the test structure. In the treatment of the probe as a

    scatterer the correct ~ dependence of the incident field is imposed.

    It is recognized, however, that some error is introduced by using

    planar image theory rather than the Green’s function for a source ex-

    terior to a cylindrical conductor.

    From the specification given in Chapter 11, for the cable from

    which the probe is constructed, the electrical radius, ka, is found to

    be .00036 for an operating frequency of 300 MHZ. It is therefore

    appropriate to make the usual thin wire approximations.

    .

    a

    .,

    28

  • (ii) Formulation of Straight Wire Integral Equation

    The charge probe of length ~ slides in a charged conductor of

    radius a~ in a homogeneous medium characterised by (u, e, u = O) is-.

    shown in Fig. (3.1). The z axis of a cylindrical coordinate system

    lies at the center of the charged conductor as shown in Fig. (3.1).

    The probe is perpendicular to the charged conductor and extends frcm

    PH=

    al ‘0 p.= al+z”For convenience a coordinate

    used with a new variable < = (p-al) measured along

    shown in

    The

    ES(C)on

    where Ac

    —J-

    Fig. (3.2).

    current and charge on the probe produce an:,

    the probe surface which has a L component,

    — ~;(~) ..5U!! (t

    k2 3

  • 9

    0

    4PI,

    slot.- . ..- ------ ---

    Coaxial Line 1

    1-. —.- ----- . . . —--+

    Charged Conductorz

    Fig. (3.1) CHARGE PROBE ON A CHARGED CONDUCTOR.

    *.I t

    .

    e

  • ,’

    I

    a

    L.-lF-J

    +H/2

    o

    III

    -H/2 i

    II‘1

    ‘:

    1III

    I

    Fig. (3.2) CHARGE PROBE AS

    +H/2

    0

    -H/2

    I

    .

    50S250Q

    :

    .> A

    7

    +H/2 1

    Ak

    -H/2

    Fig. (3.,3) LUMPED LOAD ‘LOAD AT C=O.

    *

    ‘,

    ‘,

    t “1’

    ‘x”

    I

  • where a = radius of probe, and @ = angle variable’of a cylindrical

    coordinate system with C as the axis.

    Applying the boundary condition that the tangential electrical

    field on the probe surface be zero, we obtain,

    +0 +E;(G) = Oand therefore

    @ = -E;(C)

    Thus the integro-differential equation becomes,

    (,iii) Charge Probe with 50Q Load

    (3.4)

    The charge probe sees the 50 0 input impedance of the coaxial

    cable, so that loading on the probe should be considered. For a

    lumped load of impedance Zload inserted at L = O as illustrated in

    Fig. (2.3), the current IC in ZIoad causes a scalar potential drop

    A@g across the load of

    The approximate electric field on the surface of

    which must be added to the Es is,

    the load element ,

    To achieve more useful representation for the electric field due

    to the IC in Zload

    , we look upon the effect of the load at K = O as a

    32

  • local behavior at C = O and write,

    .

    In the case of a.scatt$rerz

    As explained in the

    tional to ~ where p

    (3.5)

    beginning E;(3) has been selected to be propor-

    = (~+al). Equation (3.5) with (3.2) is the final

    form of the integral equation for the probe current. Equation (3.5)

    has been solved numerically by the method described in the next section.

    (iv) Numerical Solution Technique

    In this section a technique for solving the integro-differential

    equation is presented. This technique uses piecewise linear testing

    and pulse expansion. One performs the testing of the equation to be

    solved with elements t

  • ,

    ww-

    C-m c-m+l c-m-l-2

    -H/2

    Fig. (3.4) TRIANGLES IN PIECEWISE LINEAR TESTING SCHEME.

    e,..,,....._,-----------.....-.———...II II

    o L—___ . . . . . . -t .,f

    0. . . ... ,— I

  • @

    (B

    ‘(3.6)(a,))AC+Z ~oadIc(o)8(c),t><

    E; ,m tm >

    .where the scalar product is

    -- ..:-+, $=J+H:2,*(3.6)(b)

    type considered here where

    .

    -{Jreal testing set t ofmFor atin(c)

    the subdomain

    is zero outside

    . .Clz

    Equation (3.6) becomes

    “(cm-l,Cm+l)

    !-

    <

    Jm+l

    .=

    E;(c) tin(c)dc

    c =Cm-l

    A A-1 C- gm ISetting tm =Am =A

    in the above, the equation becomes

    (3.7)

    (3.8)

    (Lm)

    ‘r

    +

    %1

    +-“JIG(0)

    35

  • with xn=O~l,~2,~3,~ 4,.., ~(M-l) - (3.10)(a}

    and

    For a function f sufficiently

    one employs the approximation

    %-f-l

    (3.10)(b).

    smooth over the interval (Gm-l,~til)

    The final equation becomes

    +-AZ I (0)60m= AE; (Qm)load G

    where 6 is the kronecker delta

    &=~{

    1, m=nO, m#n

    m= 0, ~1, -& 2,~ 3, +4,...,& (M-1)—

    (3.11)

    (3.12) ● -

    (v) Expansion of Current and Formulation of Matrix

    The current Icis represented by expanding in known expansion

    functions each weighted with an unknown coefficient.

    coefficients ultimately determine I . In terms of aL

    uPn , as the known functions and the complex unknown

    These unknown

    pulse basis set,

    coefficients In?

    the current becomes

    .

    36

    J

    ,

    0= —

  • (0

    IC(i) = ~ InPn(G)n=-N

    (3.13)

    The pulse arrangement for approximating the current on straight wire

    is shown in Fig. (3.5)..

    The vector potential can be written as

    o

    where

    ‘c!:)= f InALn(L)n=-N.

    J

    .H2

    ‘ALn(c) =% pn(C’)K(C-C’)dC’

    G’=-H/2

    (3.1.4)

    (3.15)

    @ a partial vector potential contributed by the pulse basis element

    Pn(c)* In convenience, the following quantities are defined:

    z =mn

    +

    Vm =

    In terms of.the above definition equation

    N~ ZmnIn = V

    mn=-N m=O, +1, +2,. ... ~(1)l)——

    37

    (3.16)(a)

    (3.16)(b)

    (3.17)

  • LJm

    I I 1,1,

    IL(O

    ‘“w

    G-N C-N+l ‘-N+2

    -H/2

    Lb

    -.””.-

    A

    ‘\ $.J+\

    ‘N-2 ‘N-1 ‘N

    Fig. (3.5) PULSE ARRAIJGEKENTFOR APPROXIMATING CURRENT ON STRAIGHT WIRE.

    ,, ,, 1,,

    +H/2

    *

  • @

    ends

    The boundary condition that the current goes to zero at the

    is incorporated by directly setting p_N = pN = O. Equation

    (3.17) can

    Therefore

    be conveniently written in matrix form as,

    ‘“[Z][1] = [V] (3.18)

    [1] = [Z]-l[V]

    The total scheme of testing antiexpansion is shown in Fig. (3.6).

    (vi) Deterrni.nationof Probe Calibration Factor

    Once the current induced in the load can be found as a function

    of a the probe calibration factor can be determined as follows.1’

    Let 11 be the current induced in the load when the probe is mounted

    on the larger radius tube and I~ be the current induced in the load

    when the probe is mounted on the smaller radius tube, then relating

    these to surface charge density

    11 = ‘1’h

    12‘:‘2%

    If the linear density of charge is the same for both tubes then

    q~ =q2

    .

    nl a2-=’ (—)‘2 al

    39 L

    (3.19)

    (3.20)

  • .

    4(9TestingTriangl I_A

    C-5 C-4

    -H/2

    ‘Opo

    *-2

    &‘-3

    ‘-3 c-z

    k!!L

    Note: m=~=o, ~l,~z,...

    L

    L-J A

    13 “’

    a“‘4 15=~ (Ha~f_pulse)‘3 ‘4

  • @

    From eq. (3.19) and (3.20~, ‘1the probe calibration factor ~ isN 2

    obtained as,

    ‘1 11 al—=— ._ (3.22)‘2 12 a2

    The computed value of this ratio for different probe heights for two

    different radii of cylinders are shown in Fig. (4.19) and (4.20).

    41

  • CHAPTER IV

    PRESENTATION OF DATA AND CONCLUSION

    (i) Introduction

    Chapter 11 and 111 have discussed the measurement technique and

    the numerical analysis respectively. It is the purpose of the

    present chapter to present the data resulting from both approaches

    and discuss these results.

    (ii) Measured Data for the Charge Distribution

    As explained in Chapter 11 on measurement the charge density

    measurement was made for 3 charge conditions at the junction near

    maximum near minimum and a condition in between minimum and maximum.

    The figures (4.1) to (4.9) present the charge distribution of

    the step radius system with al= 5/16” and a~= 1/8” for 3 different. .

    probe heights 0.10”, 0.08”, and 0.06”. For each probe height,

    measured data for 3 different charge conditions at the -junctionare

    presented along with the curve determined from the modified Prony’s

    method for the magnitude of the measured data. Similarly the

    Figures (4.10) to (4.18) present the data for the step radius system

    with a~= 3/16” and a2= 1/8”.

    (iii) Probe Calibration Factor

    The probe calibration factor (N1/N2) as a function of probe

    height as computed from the measured data is shown in Fig. (4.19) for

    al= 5/16” and a2= 1/8”. The theoretical probe calibration factor

    42

  • :

    4-u

    ,,

    CHFIRGE PROBE RESPONSE: PROBE HEIGHT 0.10”c1 PHFtSE 91 = 5/16” 92 = 1/8”x MRGNITUDE

    o ,—.+ CURVE FIT

    0.00 -50.00 -io.oo .10.00 30.00 :X10”

    Z(MMI

    Fig. (4.1) MAXIMUM CHARGE CONDITION - PROBE HEIGHT = O.1O”, al= 5/16”

    o(3

    G’N,

    o0

    G-+Hod

    x’

    fg-(3.

    gel .LLJC3

    o-‘?LIJo~~a

    sL’

    o0

    &N

    !00

    . ..- ,-.

    11,-

    1

    — —

  • tIII

    Fig. (4.2)

    CHRRGE PROBE RESPONSE:cl PHRSEx MFIGNITUIIE

    o CURVE FIT

    (9

    PROBE HEIGHT 0.10”91 = 5/16n 82 s 1/8”

    0.00

    MEDIUM CHARGE

    -30”00 -’20.00 . lb.oo 3i?.oo)kl~i

    Z[MMjCONDITION - PROBE HEIGHT = 0.10”, al= 5/16”

    5

    .—..._ %. . ~

    I

  • ,.) ,

    :. ,1

    I

    ,1’

    ,’

    1,

    CHflRGE PROBE RESPONSE:

    % MRGNITUDE

    ‘z CURVE FIT

    o0

    G’

    Q

    PROBE HEIGHT 0.10”fll = 5/16” W = 1/8”

    .,

    0.00 -30.00 -io.oo - lb.oo 3tl.oo 5

    o’0 1,

    >4 -iN

    ,1

    00.

    .s:00

    )KltllZ(rfrl)

    Fig. (4.3) MINIMUM CHARGE CONDITION - PROBE HEIGHT = 0.10”, a,= 5/16”.L .—.

    1,

  • ‘k-0

    1’

    0

    CHftRGEax

    9 .—.*

    RESPONSES PROBE HEIGHT 0.08”PHflSE RI = 5/16” f%? s 1/8”MRGNITUIIECURVE FIT

    1

    0=00 -30 ●OO -io.oo lt).oo 30.00 5M@

    Z(Mtl)

    Fig. (4.4) ?4AXl141JlfCHARGE CONDITION - PROBE HEIGHT = 0.08”, al==5/16”

    0

    c)SJ

    ;0LLJo

  • :

    CHFIRGE PROBE RESPONSE: PROBE HEIGHT 0.08” I

    c) PHFISE ,Rl = 5/16” f12 = 1/8”x tlRGNITUllE I

    o* CURVE FIT ~

    I Io’0

    .

    I

    m.l--m -d)

    ,!

    -1-*-~w

    ,,I!’

    >’w

    -1

    -1’0-INx “.-f=

    0

    I

    t

    (-J-

    0.GoLIJ0

    ,,

    (3

    -c---l h.1

    Io-:WIom ,i I

    -—,

    oc)

    t4- 1.

    1I

    1 1 I

    0.00 -30.00 -Lo’-ou-- .10.00 30.00 50’!003KmL

    i!trml

    Fig. (4.5) MEDIUM CHARGE CONDITION - PROBE HEIGHT = 0.08”, al= 5/16”

  • I

    (4

    mal-(-o

    LLJ

    00

    2’.

    PROBE RESPONSE: PROBE HEIGHT 0.08”PHI?SE 91 = 5/16” 82 = 1/$”MflGNITUOECURVE FIT

    Fig. (4.6) MT.NIMUM

    a

    0.00 -30.00 -10.00 10.00 30.00

    1 1,

    IEIQ1Z(MMI

    CONDITION - PROBE HEIGHT = 0.08”, al= 5/16”

    m a

    q ‘,!1

  • :

    .c-W

    b .

    :

    .

    ,1

    CHflRGE PROBE RESPONSE: PROBE HEIGHT 0.06”a PtifW3E 91 = 5/16” fl12s 1/$”% I’IRGNITUDE

    ~,~

    -1(!)

    I I

    m

    .,

    ,,

    In

    on

    09

    ~c5I.LJo

    0-

    31fll)1Z?(MM]

    ,,

    I

    Fig. (4.7) MAXIMUM CHARGE CONDITION - PROBE HEIGHT = 0.06”, al= 5/16”

    I

  • u

    CHRRGE PROBE RESPONSE: PROBE HEIGHT 0.06”Q PHRSE 81 = 5/16” W? = 1/8”x MRGNITUDE

    cl_v

    CURVE FIT

    c?)

    0.00 -!30 .00 -ioaoo . lb.oo 30.00 53E101

    Z(Mtll

    J?ig. (4,8) MEDIUM CHARGE CONDITION - PROBE HEIGHT = 0.06”, al” 5/16”

    O . ,

    ao

    &N

    o0

    r-l

    x

    Q

    o-%W

    1!00

    *

    ..

  • i .:

    CHRRGE PROBE RESPONSE: PROBE HEIGHT 0.06”a PHfiSE fll = 5/16” ff2 s 1/8-# MFIGNITUDiE

    o* CURVE FI’T 1:

    ‘o

    -OJE“-0

    IJJo

    ~o

    (2)

    o

    0

    IS”

    (3

    (3

    c1

    I

    v K0.00 -30.00 -10.00 10*OO 30.00 &c

    XIO1

    Z[MM)

    Fig. (4.9) MINIMUM CHARGE CONDITION - PROBE HEIGHT = 0.06”, al= 5/16”

    I

  • o

    CH9RGE PROBE RESPONSE:(9 PHRSEM tlflGNITUOE

    o

    N— CURVE FIT

    Fig. (4.10)

    ,

    V@

    PROBE HEIGHT 0.10”fal = 3/16” R2 = 1/8”

    Q

    II 1

    0.00 -50 ●OO -io.oo .10.00 30=00 53E101

    Z(MMI

    MAXIMUM CXMRGE CONDITION - PROBE HEIGHT = 0.10”, al= 3/16”

    @!

    II II . 1,

  • ml .. : *

    .,

    CHRRGE PROBE RESPONSE: PROBE HEIGHT

    c1 PHflSE “ ,91 = 3/16”

    % Mf!GNITUOE

    o*lo- 1

    92”= 1/8”

    E CURVE FIT I

    0.00 -30.00 -io=ooI I

    10*OO 30.00 5IE1(I1

    Z(MMI

    I?ig, (4.11) MEDIUM C~GE CONDITION - PROBE HEIGHT = 0.10”, a~= 3/16”

    ,’!

    0’<0:

    .&do4

    N

    0

    O(J-J

    Q

    o

    0.

    ,R1!00

    I

  • —-- n..

    WI.P-

    Ct!FIRGE PROBE RESPONSE: PROBE HEIGHT 0.10”

    a PHRSE 91 = 3/16” F12 = 1/8”

    x MRGNITUDEo@J — CURVE FIT

    t?)

    @)

    ;0.00 -30.00 -io.oo lb=oo 30.00 Ixmr

    Z[MM)

    Fig. (4.12) MINIMUM CHARGE CONDITION - PROBE HEIGHT = 0.10”, al= 3/16”

    ,=*

    , ,

    I ,,I

  • m’.+ ●

    :

    1!

    CHflRGE PROBE RESPONSE: PROBE HEIGHT 0.08”c) PHflSE .Ifll = 3/16n 92 = Ii/8”% MRGNITUDE

    oN CURVE FIT

    -1

    wo

    o

    0.

    0

    c)

    0.00 -30.00 -io.oo lU.00 30.00 5MIO1

    ZIMMI

    1’,

    ‘1

    o~iIOi!

    0-

    Dc)

    .

    DN

    !00

    Fig. (4.13) MAXIMUM CHARGE CONDITION - PROBE HEIGHT = 0.08”, al= 3/16”

  • ,

    CH13RGE PROBE RESPONSESc1 PHWSEx flRGNITUilE

    aml CURVE FIT

    PROBE HEIGHT 0.08”RI = 3/16” R2 = lf8”

    c)

    c1

    Q

    C9

    (3

    0.00 -30.00 -iOaOO .10.00 30.00*IQ’

    2(flM)

    Fig. (4.14) MEDIUM CHARGE CONDITION - PROBE HEIGHT = 0.08”, al= 3/16”

    * , 1.

  • . w

    m-J

    :

    CHf3RGE PROBE RESPC?NSEX PROBE HEIGHT 0.08”c) PHf7SE 91 = 3/16” 92 = 1/8”x tlfiGNITUDE

    ~R— CURVE FIT .

    .CJ

    -

    Coml-.-l”C!3°

    ‘>n-1

    wo3

    I~l.g

    b-i.

    ~ 0“(3

    o0

    &’

    ‘o

    I1’

    (

    )c

    I I I I

    0.00 -30.00 -10.00 10.00 30.00 5*101

    Z(MMJ

    Fig. (4.15) MINIMUM CHARGE CONDITION - PROBE HEIGHT = 0.08”, al= 3/16”

    ~

    ,,‘1

    uw

    aa

    .

    *

    -.

    I

    ,1.,

  • CHRRGE PRi3BE RESPONSE: PROBE HEIGHT 0.06’a PHFISE 91 = 3/16” f12 = 1/8”x MRGNITUDE

    ww CURVE FIT

    4

    m

    . . I

    ?%lO1Z(MM1

    - PROBE HEIGHT = 0,06”, al= 3/16”Fig. (4.16) MAXIMUM CHARGE CONDITION I I

    ●b +

    .—,“

    ,

    * ..

  • I

    II I

    1,,, I

    ,!z

    :

    CHflRGE PROBE RESPONSE: PROBE HEIGHTa PHfWEx MRGNITUDE

    oN CURVE FIT

    III

    I

    ,,

    .0

    n

    mml--d

    .go

    >-

    -1.-l1--1:z“-0

    I.lJn

    ~o

    C5

    o0

    .0

    91 = ,3/16=0.06”

    R2 = 1/0”

    c1

    I v1 I 1 I

    0.00 -30.00 -10.00 .10.00 30900 53K10X

    Z(MMJ

    Fig. (4.17) MEDIUM CHARGE CONDITION - PROBE HEIGHT = 0.06”, al= 3/16”

    .0N

    o0

    .

    ,=

    1!00

    ,,

    ‘1,

    I ,)

  • Fig.

    .

    1.’1’. .1, .

    (4.18)

    <

    PROBE RESPONSE: PROBE HEIGHT 0.06”PHRSE R1 = 3/16” F12 = 1/8”

    tli3GNITUOEC3N CURVE FIT

    Q

    (!)

    MH!

    i I I i0.00 -30.00 -10”00 .10.00 30.00

    MINIMUM CHARGE

    MI(Y2(MMJ

    CONDITION - PROBE HEIGHT = 0.06”, al= 3/16”

    o“

    C20

    CJ(N

    .

    c)-?Ll

    r2-

    08

  • :

    1.4

    1.32

    1.24

    N1/N2

    1.08

    1.0

    I

    ,! I, I Charge Condition at the’1 :,

    I Junction ~1,

    1,

    4~I I I

    JI l,’

    PROBE HEIGHT IN INCHES

    MinimumMedium

    Maximum

    Theoretical

    ,,,,II

    ,,

    I

    .. ,

    0.04 0.06 0.08 0.10 0.12

    Fig (4.19) PROBE CALIBRATION FACTOR FOR 5/16” to 1/8” RATIO OF RADII

  • obtained by the technique explained in Chapter III is also shown in

    Fig. (4.19). Similarly Fig. (4.20) presents the probe calibration

    factor for al = 3/16” and a2 = 1/8”.

    Examination of Figs. (4.19) and (4.20) reveals differences of about

    8% in the experimental values of N1/N2 for different charge conditions

    are the junction. With the assumption that the probe responds to the

    local charge density at its location on the cylinder N1/N2 should be a

    function of probe height and cylinder radii only. The observed effect

    of the position of the junction with respect to the standing wave pattern

    is believed to result from inaccuracies in the extrapolation of the data

    to the,junction. The extrapolation,

    is accomplished by fitting the curve

    data 5s available over a distance of

    the accuracy in determination of the

    limited. For these reasons the data

    and hence determination of N /N12

    to the measured data. Since the

    only A/2 each side of the junction

    parameters for this curve fs

    obtained for the case of maximum

    charge at the junction is the most reliable since the slope of the curve

    in the extrapolation region is near zero. In future use of these experi-

    mentally determined probe calibration factors greatest weight will, there-

    fore, be given to the values determined for the maximum charge condition.

    Examination of Figs. (4.19) and (4.20) also reveals a consistent

    difference between the measured and theoretical values of N1/N2. The

    reasons for this difference are not fully understood; however, it must

    be remembered that the theoretical solution is based on a planar image

    theory. The effects of the cylindrical geometry are the most likely

    62

    ..= .-

  • m w : ●✌‘e

    1.25

    I

    I 1.20

    1.15

    N1/N2

    1.10

    1.05

    I

    ),,

    !

    Charge Condition at;theJunction

    I

    I

    PROBE HEIGHT IN INCHES

    1 * #

    Medium

    Minimum

    Maximum

    Theoretical

    0.04 0.06 0.08 0.10 0.12

    Fig. (4.20) PROBE CALIBRATION FACTOR FOR 3/16” tO 1/8” RATIO OF RADII

  • .

    ●source of the observed differences between experimental and theoretical

    values. Until more accurate data and theoretical solutions are available,

    the experimental value of N1/N2 will be used as the

    (iv) Conclusions

    The results of this investigation have clearly

    effect of cylinder radius on the response of charge

    most reliable.

    demonstrated the

    probe and have taken

    a considerable step toward better understanding of charge measurements.

    The probe calibration factors obtained can now be used to correct charge—

    . distribution measurements on structures containing junctions of unequal

    radii cylinders.

    Further work, however, is needed for complete understanding of the

    response of monopole probes on nonplanar surfaces. In”particular, an.—

    effort should be made to include the cylindrical image plane geometry o

    into the theoretical treatment presented here. In addition, other probe

    configurations should be examined and evaluated with respect to their

    response on nonplanar surfaces.

    64

  • [El

    [2]

    [3]

    [4]

    [5]

    [6]

    [7]

    REFERENCES

    R.W.P. King, “Quasi-Stationary and Nonstationary Currents inElectric Circuits,” pp. 267-270, Encyclop-ediaof Physics,Edited by S. Flfigge,Vol. XVI: Electric Fields and Waves,Springer-Verlag, 1958.

    H. Whiteside, “Electromagnetic Field Probes,”Technical Report SO. 377, Harvard University,setts, 1962.

    T.T. Wu and R.W.P. King, “The Tapered Antenna

    Cruft LaboratoryCambridge, Massachu-

    and its Applicationto the Junction Problem for Thin Wires,” IEEE Transactions onAntenna and Propagation, Vol. AT-24, pp. 42-45, January 1976, andalso Interact~on Note 69, January 1976.

    C.M. Butler, K.R. Umashankar and C.E. Smith, “Theoretical Analysisof Biconical Antenna,” report submitted to the U.S. Army,STFL4TCOM,1975.

    R.F. Barrington, Field Computation by Moment Methods,Company, New York, 1968,

    Macmillan

    N, Marcuvitz, Waveguide Handbook, McGraw-Hill Book Company,New York, 1957.

    Hildebrand. Introduction to Numerical Analvsis. McGraw-HillBook Company, New York, 1956.

    65

    !-