Instruction Manual For use with methane gas only. Read and understand instructions before use. Patented SENSIT ® PMD PORTABLE METHANE DETECTOR MADE IN USA SENSIT Technologies is in compliance with ISO 9001:2008 Ex-Ox-Tox Gasdetectie Westerdreef 5V 2152 CS Nieuw-Vennep Telefoon: 0252 620885 E-mail: [email protected]l Website: www.exoxtox.nl

SENSIT PMD - Ex-Ox-Tox · PDF fileThe Sensit® PMD , referred to as the ... An internal pump provides rapid sampling into the ... • The PMD uses a simple “Sensit Style” user

Embed Size (px)

Citation preview

Instruction ManualFor use with methane gas only.

Read and understand instructions before use.

Patented

SENSIT® PMDPORTABLE METHANE DETECTOR

MADE IN USA

SENSIT Technologies is in compliance with ISO 9001:2008

Ex-Ox-Tox Gasdetectie Westerdreef 5V2152 CS Nieuw-Vennep Telefoon: 0252 620885 E-mail: [email protected] Website: www.exoxtox.nl

Warnings:

• Topreventignitionofflammableorcombustible atmospheresdisconnectpowerbeforeservicing.

• Removeandchargebatterypackinanareaknowtobe freeofcombustiblegases.

• UseonlySensitTechnologiesbatterypack.

• Servicemayonlybeperformedbyfactoryauthorized servicetechnicians

• Notforuseinenvironmentsgreaterthan21%oxygen.

Safety Precautions:

• Readandunderstandinstructionspriortouse.

• AlwaysstartthePMDinanareaknowntobegasfree.

• Tamperingwiththisproductmayvoidthewarranty.

• UseonlySensitTechnologiesapprovedparts and accessories.

• Neveruseaninstrumentknowntobedamaged, operatingunusually,oroutofcalibration.

For further information contact Sensit Technologies.

SensitTechnologies851TransportDriveValparaiso,Indiana46383USA

Tel:219/465-2700Fax:219/465-2701

Email:[email protected]:www.gasleaksensors.com

TABLE OF CONTENTS Page Section Title

4 GeneralDescription

5 PartsandAccessories

6 ProductSpecifications

6 Product Features

7 ElectronicFeatures

8 DisplayFeatures

9-10 HousingFeatures

11 Filters

12 FieldOperations

13 QuickMenu

14 UserMenu

15 Calibration

16-17 AdjustableFeatures

20 Warranty

4

TELESCOPICSURVEY BOOTand PROBE ROD

GRAPHIC DISPLAY INFRAREDCOMMUNICATIONS

ALARM LEDSOUNDER

FILTER

OPERATIONBUTTONS

BACKLIGHTSENSOR

General Description

TheSensit®PMD,referredtoasthe“PMD”,isdesignedtodetectmethanegasfrom1ppmupto100%volume.ThePMDmaybeusedforwalkingsurveysandvehiclesurveys(whenproperlyinstalled)forundergroundnaturalgaspipingsystems.Otherapplicationsincludelandfillmonitoringandcheckingpipingsystemswhereotherinterferinggasesmayprovideinaccuratereadings.

ThePMDsensesgasusingInfrared(IR)AbsorptionSpectroscopyincombinationwithanelectronicnarrowbandpassfilter.Thistechnologyutilizesaninfraredlightsourcewithanoutputthatischangedwhencertaingasesabsorbthelightoutput.Thefilteronlyallowsspecific lightwavelengthstobemonitoredandmeasured. Theconcentrationofgas isproportionaltotheamountofspecificIRlightabsorbedandisdisplayedinPPM,%LELand/or%Vol.

ThePMDhasalargedisplayindicatingconcentrationandotherinstrumentfunctionssuchasbatterychargeandperformance.Aninternalpumpprovidesrapidsamplingintothedetectionchamber.Audibleandvisualalarmsindicatewhenpresetalarmsareexceeded.Bluetoothdatatransmissionprovidescommunicationofrealtimeandstoreddata.OptionalGPSandDataLoggingallowforfurtherrecordingoftime,dateandlocationdata.

“D”RING

5

Accessories

Standard Accessories SurveyProbewithrubberboot ShoulderStrap LithiumIonRechargeableBatteryPack BatteryCharger BatteryPackRemovalTool InstructionManual PMDHydrophobicFilterDisc(10pack) SurveyProbeFilter(blue)

Optional Accessories FiberglassBarHoleProbe BrassBarHoleProbe HDFiberglassBarHoleProbe CalibrationKit BumpGasCylinder(50ppmCH4) HDCarryingCase(PelicanStyle) VehiclePowerAdapter SDdataloggingcard

Replacement Parts ShoulderStrap“D”Rings FilterCap FilterCap“O”Rings(10) PMDHydrophobicFilterDisc(10pack) SurveyProbeFilter(blue)

6

Product Specifications

Operational Specifications

Temperature: -4to122FDutyCycle: ~7hoursResponsetime: T50<3secondsAlarms: Audible-95db@15inches Visual–Redbacklightondisplay Visual–Displayinfo Visual–Failureindicators Settings–PPMandLELPower: LithiumIonRechargeablebatterypack OptionalVehiclepoweradapterPumpFlow: 1.5lpmDetectionRange: 0-5000ppm 10–100%LEL 5-100%V/V

Physical Specifications

Size:Weight: 6.1LbsConstruction: CastAluminum/ABSPlastic

Sensor Specifications

IRSensor: Wavelength–3.4m DetectionRange1ppm–100%v/v Warmup<10min Calibrationrequired-6months

Product Features• ThePMDisconstructedofdurableABSplastic• ThehousingisdesignedtomeetIP54protection• ThePMDisdesignedtomeetCat3IntrinsicSafetyrequirements• Batterypackisdesignedforeasyfieldreplacement• FieldorSmartCalcalibrationiseasilyperformed• Externalfiltersareinexpensiveeasytochange• TheLCDdisplayiseasytoread• ThePMDusesasimple“SensitStyle”userinterface• CommunicationwithotherdevicesiseasywithBluetoothinterface• RealtimeGPSisoptional• RealtimeDataloggingwithremovablestorageoption

7

Electronic Features

User Interface

Theuserinterfaceincludesthefollowingfeatures:• Threebuttonoperation• LEDStatusindicators• Sounder• Photocell(automaticbacklighting)• IRcommunicationport• LCD

Button Operation

TheSensitPMDisoperatedwiththeuseofthreebuttonsbelowthedisplay.

TheyarelabeledA,B,C.

The“A”button:• activates/deactivatestheinstrument• mutestheaudiblealarms

The“B”button:• Accessesthequickmenuandusermenu• DigitallyadjuststheTICKsoundforleaklocating

The“C”button:• Savesdisplayedreadingstotheinternalmemory• Performsamanualzero.

8

Display Features

Thedisplayprovidesallinformationincluding:• GasconcentrationinPPM,LELAND%Vol• Batteryvoltagestatus• PumpStatus• GPSactivationandcoordinates• Bluetoothoperation• Dataloggingoperation• Warningindication

Iconsonthedisplayindicatethestatusofvariousfunctions.

-FlowOK -Flowblocked

-Alarmnotmuted -PPMalarmmuted

-%LELalarmmuted -%GASalarmmuted

-Warning/trouble/failuresymbol

-BluetoothOnandConnected -Bluetoothonbutnotconnected

-TickON

-Dataloggingon–flashoncefordatasave

-DataSaved–flashonceformanualdatasave

-GPSConnected -GPSenabled-notconnected

Alarm settings

Thereare3adjustablealarmsettings.

1.PPM a. Default:10ppm b. LowLimit:0 c. Highlimit=500 d.DisablePPMalarm:SetthePPMalarmto0 e. Alarm:Audible,LED,Redbacklightifunitindark

2.%LEL a. Default:50%LEL b. LowLimit:10%LEL c. Highlimit=90%LEL(Ifnoautorangeto%gas) d.DisableLELalarm:Cannotdisablethisalarm(muteonly) e. Alarm:Audible,LED,Redbacklightifunitindark

3.%GAS a. Highlimit=20%Vol(Default17%) b.Disable%GASalarm:Cannotdisablethisalarm c. Alarm:LED,Redbacklightifunitindark

AdjustmentoftheppmalarmiscoveredintheUserMenuportionofthismanual.

Housing Features

Battery Pack

Thebatterypackislocatedonthebacksideoftheinstrumenthousing.ThebatteriesareLithiumIonrechargeable.ThebatteriesareonlyavailablefromSensitTechnologies.

Toremovethebatterypackturntheretainingscrews¼turn.Pullthebottomofthepackawayfromthemainhousing.Rechargingofthebatterypackrequiresremovalfromthemainhousing.

Toreplaceslidethetoptabsofthebatterypackintotheretainersatthetopofthemainhousing.Pushthebottomintoplaceandturnandlocktheretainingscrews.

9

10

ChargeComplete

Charging Status ChargerConnection

ThebatterypackhasLED’stoindicatechargestatus.Chargingshouldonlybeperformedinanareaknowtobefreeofcombustiblegases.

Gas Inlet and Outlet

Locatedonthefrontofthemainhousingisthegasinletandoutlet.Botharefittedwithluer-styleconnectionsforeasyconnectiontoprobeaccessories.

Thecenterconnectionisthegasinlet.Totherightistheoutlet.Donotblockgasoutlet.

Shoulder Strap attachment

“D”ringsarelocatedoneachsideofthemainhousingforshoulderstrapattachment.Clipsonthestrapwillattachsecurely.

11

Hydrophobic Filter Assembly

WARNING:Donotoperatewithoutproperfilter.Damagemayoccurtothepumpandotherinternalparts.Onlychangefilterwithpumporinstrumentoff.

Thegasinletisprotectedbyahydrophobicfilter.

Itisaccessedby:1.Detachanyprobeassembly2.Twistandremovefiltercap3.Removeandreplacefilterdisc4.Check“O”ringinsidecap5.Replacefiltercap6.Performedflowblocktest

Additionalfiltersarelocatedineachoftheaccessoryprobeassemblies.

12

Operation and Field Use

ToactivateinstrumentpushPOWERbutton.Thefollowingwarm-upsequencewilloccur:

1. Warm-upsequence(a-i)withinfobelow.

a. Logo b.SerialNumber c.Revision d. SystemCheck i. LEDcheck ii. Backlightcheck iii.Date/Time e. Caldue(upcoming)orCalPastDue f. ActivationAcknowledgments i. Pump ii. Dataglog iii.GPS g. Sensorwarm-up(8-20Min)and“PleaseWait” h. Autozero i. Workingdisplay 2. Attachtheappropriateprobeassemblyandperformaflowblockcheck.An alarmsound,displayindicationandflowiconwillindicateafailedtest. 3. Followcompanyproceduresforperformingleaksurvey/investigation

4. Ifthealarmsettinghasbeenexceededpressandreleasethe“A”buttonto disablethealarm.Thealarmsoundwillautomaticallyresetwhenthe concentrationhasgonebelowthethresholdvalue.Secondaryalarmsathigher concentrationwillactivatewhenvaluesareexceeded.Pressandreleasethe“A” buttonagaintomutesecondaryalarms.Duringanyalarm,thedisplaybacklight willbeREDincolor.

5. Checkfiltersdailyforbestresults.

6. Batterypowerisdisplayedusingabatteryvoltageicon.Fullbatteryis12.0v. IconwillflashandBATLOWwillintermittentlyflashonthedisplay~8v. BATFAILwillbedisplayedwheninsufficientbatterypowerisavailable.

WhenBATLOWisindicated,pressanybuttontoacknowledgethewarning.Failure toacknowledgewithin30secondswillresultinautomaticpowerofftoprotect batteriesfromexcessivedischarge.

7. Toturnofforputintostandbymodepushandholdthe“A”button.Itwill display“SHUTDOWN”onleftsideand“STANDBY”onrightsideofthedisplay. Pressthe‘A’buttontoturntheinstrumentoff.Press‘C’toputunitinstandby modewhichwillreducepowerconsumptionandwillallowyoutoreducethe warmuptimetoapproximatelytwominutes.Werecommendusingstandby modeforlessthan1hour.

13

QUICK MENU

Pressandrelease‘B’button.Thiswillprovideaccesstothefollowingfunctions.Pressthe“C”buttontoscrollthroughtheoptions.Use‘B’buttontoselect.Use‘B’buttonagaintoturnitOFFor‘C’buttontoturnON.Press‘A’buttontoselectyourchoiceandexitmenu.

1.TICK–Audibleleakdetectionmode.NOTE:Activatingthetickwilldisable soundalarm.

a.Whenactivatedatickwillbeheardeverysecond.Frequencyoftickwill increaseasgasconcentrationincrease. b. Pressandrelease‘B’buttonwillresetthetickfrequencyto1tickper second.Thetickfrequencywillincreaseifgasconcentrationincreasefrom thatconcentration. c. Toturnthetickoff,pressandrelease‘A’button.Pressandrelease‘A’button againtoactivatethealarmsound.

WhenviewingthefollowingselectionsUse‘B’buttontoselect.Use‘B’button againtoturnitOFFor‘C’buttontoturnON.Press‘A’buttontoselectyourchoice andexitmenu. 2.BHTest–Barholetestfunction

3.PUMP-Use‘B’buttontoselect.Use‘B’buttonagaintoturnitOFFor‘C’ buttontoturnON.Press‘A’buttontoselectyourchoiceandexitmenu.

4.GPS–RealtimeGPSX/Ycoordinates

5.Warning-Usethisfeaturetolistcurrentfailuremode.

Bluetooth Activation

WhentheBluetoothisactivatedthedisplayreadingscanbetransmittedtoanydevicecapableofreceivingBluetoothdata.Thedatamayalsobeviewedbyusing“HyperTerminal”.Thewebsiteiswww.lgraeve.com

BluetoothPasskey:5DigitSerialNumber

SerialCommunicationsSettings:BitsPerSecond:57600Databits:8Parity:NoneStopbits:1Flowcontrol:None

DataStreamcanbeadjustedtosenddataevery1-10sec.youcanfindsettingstochangethisinExpertMenu->DataStream.YouneedtoturnDataStreamONinUserMenu.

OnceconnectedtotheBluetoothdevice,PMDsendsthe7digitPPMreading.

14

USER MENU

Toaccesstheusermenupushandholdthe“B”buttonuntilUSERMENUisdisplayed.Fromthismenuthefollowingfeaturescanbeviewedbypressingthe“C”buttontoscrollthroughtheoptions.Entertheselectionbypressingthe“B”buttonandadjustingbyusingthe“C”buttonortheonscreenprompts.Pressingthe“A”buttonrepeatedwillreturnyoutotheworkingdisplay.

Date and Time Viewthedateandtime. Timeisin24hourformat.Dateismm/dd/yyformat.

Print Menu Allowsaccesstoprintmenuforcalibrationandoperationalsessions.

Bump Test Performtimedtestforgasresponseinppmrange.Pumpmaybe selectedONorOFFforthisfunction.

Calib Menu Calibrationto1000ppm,2.5%and100%methanegasconcentrations. ForinstructionrefertoCALIBRATIONsectionofthismanual. •ZeroAir-ToSetZeroPoint PumpOn/Off OnscreenInstruction •1000PPM-Tocalibrate1000ppmlevelonly (ZeroAircalibrationrequiredfirst-automatic) •2.5%V/V-toCal2.5%(50LEL)CH4 Only •100%V/V-toCal100%(100LEL)CH4 Only •FullCalib-Automaticallysequencezero,1000,2.5%V/V,100%V/V (ThisupdatesCalDue)

Datalog Activatesthedataloggingsystem.(IfActivatedw/memorycard)

Data Stream Activatesrealtimedatastreamingofdisplayreadingseverysecondto anotherrecordingdeviceusingBluetooth.ContactSENSITforsettings.

Alarm PPM AdjustmentforPPMalarmvalue. Settingat0disablesalarm. Rangeis0-500ppm

Set Clock Adjustdateandtime.

View Sess Log Viewmanuallysaveddata.

View BHLOG Viewthebarholetestlogs(IfActivated).

View Auto ViewAutologdataofoperationalpeakvalues.

View Cal Log Viewcalibrationlog.

QCal Log Viewrecordingofindividualcalibration(NotFullCalib)

15

Calibration Check

Toverifytheproperoperationofthisinstrumentapplyaknownconcentrationof50-1000ppmmethanetothesamplesystempriortouse.

Thisisnotarequiredprocessbutisrecommendediftheproducthasnotbeenusedformorethan30days.

Calibration

ThePMDrequirescalibrationevery3months.Anautomaticreminderwillnotifytheuserthatcalibrationisdueatstartup.

Calibrationcanonlybeperformediftheinstrumenthasbeenoperationalfor30minutes.

Tocalibratetheareamustbeknowntobegasfree.Ifthereismethaneintheareaabottleofzeroairwillbenecessary.Othergasesrequiredinclude1000ppm,2.5%and100%methane(pipelinegasisalsoacceptable).Somegasescanbecalibratedindividually.ContactSENSITTechnologiesServiceDepartmentformoreinformation.

Calibrationisperformedwiththepumpoffifthereisnoaccesstodemandflowregulatorsforthecalibrationbottles.Turn-offisselectedpriortocalibration.

CalibrationisperformedthroughtheUserMenu.Usingthe“B”buttonaccesstheUserMenu.Usethe“C”buttontoscrollto“CalibMenu”.

Tobeginpressthe“B”button.Iftheinstrumenthasnotoperatedfor30minutesacountdowntimerinsecondswillbedisplayed.Whentheproperamountoftimehasbeenmetthedisplaywillshow“Proceed”.Pressanybutton.Abeepwillbeheard.

SelectCalTypeandfollowon-screenprompts.

Applyzeroppmmethaneorcleanair.Press“C”buttontobeginzeroaircalibration.Displaywillshow“waitingforstableread”followedby“success”whencompleted.Pressanybuttontocontinue.

Displaywillread“pleaseapply1000ppmmethane”.Attachandapply1000ppmmethane.Theprocessrequires1-5minutes.Displaywillshow“waitingforstableread”followedby“success”whencompleted.

Displaywillread“pleaseapply2.5%methane”.Attachandapply2.5%methane.Theprocessrequires1-5minutes.Displaywillshow“waitingforstableread”followedby“success”whencompleted.

Displaywillread“pleaseapply100%methane”.Attachandapply100%methane.Theprocessrequires1-5minutes.Displaywillshow“waitingforstableread”followedby“success”whencompleted.

Uponcompletionpressanykeytoexittoworkingdisplay.Activatethepumpandallowtoclearpriortoturningoff.Thecalibrationreminderwillautomaticallyupdateaftersuccessfulcalibration.

16

Adjustable Features (passwordprotected)ContactSensitTechnologiestogainaccesstothefollowingadjustablefeaturesnotintheusermenu.

Service Contrast: Adjustthedisplaycontrastasneeded.

Quick Scan: OptimumTempAdjust(5-30RQd) Orientation Cal: Cal Flow: Settingflowalarm

Tick SENS Lo,Med,orHigh

Alarm %LEL AdjusttheLELalarmvalue.Defaultis50%.

Alarm %V/V Adjustthepointwhenthealarmdeactivates.Defaultis17%.

Show SESS Allowdisplayofmanuallysaveddata.Defaultison.

Show BHLOG AllowdisplayofBHtests.Defaultison

Show AUTO AllowdisplayofAutologdata.Displayisoff.

BH Time Setthetimeinsecondsfordurationofthebarholetest. Defaultis15seconds.

Purge Time Setpurgetimeforinstrumentpriortoshutdown. Defaultis10seconds.

Standby Timeofinactivityorlackofmotioninminutesthatwillcausethe instrumenttoautomaticallygointolowpowermode.Defaultisoff.

Shutdown Timeinminutesthatshutdownwilloccurwheninstandbymode. Defaultis60.

Bump Limit Concentrationofgastobeacceptablewhenperformingabumptest. Rangeis40–1000ppm.Defaultis1000.SetLowerandUpperLimit.

Bump Time Timerequiredinsecondsforunittopassabumptest. Rangeis10-60.Defaultis45.

Log Period TimebetweenDatalogpointsinseconds.Range1-60.Defaultis2.

Data Stream Timeinsecondsfordatatobestreamedthrough Bluetoothcommunication.

17

Adjustable Features (Continued)

Rest UCal Resetcaldatatofactorysettings(PasswordProtected)

Set Cal Due TimeInterval(90Default)

Cal Due Ack RequiresbuttonpushafterCalDuereminder

DISP Callog CalDetail

Display Options SettheconcentrationdisplayPPM,LEL,%Vol. Configurations include: PPM0-10,000.Defaultis5000

PPM Option: PPMAutorangetoLEL.Defaultison.

PPM Handoff: HandofffromPPMatPPMsetting.

% LEL Option: LELfrom0.1-100%.DefaultisOn 100% LEL Equiv: SetLELlimitsfrom4.0–5.0%gasvolume.Defaultis5%. PPM+%LEL: DisplaybothPPMand%LEL.DefaultisOff NG Factor: Automaticcalculationforlowmethanepipelinegas.

UCal Date Dateoflastfullusercalibration Erase Auto EraseAutolog

Erase Sess ErasemanuallysaveddataLog Erase BHLog Erasebarholetestlog

SCal Pump On/OffDuringCal(foreachgastype)

Rezero REQD DowndriftAlert,On/Off,Value5Default

Language Selectlanguagefordisplay

Auto Bump Requirebumptestbeforeuse

Erase Bump EraseAutoBumpLog

18

Saving SD Card Data Log Information

RemoveSDMemoryCardfrominsidethebatterycompartment.InstallSDMemoryCardincomputer.SaveFileindesiredformatandlocation.MicrosoftExcel.CSVisthepreferredformat.

GPS Data Conversion & Coordinate Mapping

GPSDataconversionandmappingcanbeaccomplishedusingSmart-LinkDataViewerSoftwareandsoftwarefromgpsvisualizer.com

1.UsingSmartLinkselectthedesiredreports.(See“Howtouse”and“Reports”sectionofSmartLinkHelpfileformoreinformation.)

2.Click“SaveReport”button.Savereportas.CSV(CommaDelimited)fileformat.This formatwillonlysaveloginformationforwhichvalidGPScoordinateinformationisavailable.

3.Goto:http://www.gpsvisualizer.com/Select“Google Maps”.

Import your .CVS file:

1.Usingbrowsebutton,selectthefileyouwishtoview.

2.Selectthemaptype.

Way-points:Todisplaycoordinates,gasconcentrationdetectedandtimeatspecificlocations.ItisbesttousewithSession-log,Bar-holelogandAuto-log.

Routepoints:The“route”wouldrepresenttheroad,trail,path,etc.thatyouhavetaken.Italsodisplaygasconcentrationdetectedandtimeatspecificlocations.ItisbesttousewithData-logtoviewareasurveyed.

3.Click“Drawthemap”button.

19

GPS Data Conversion & Coordinate MappingSampleofthedatalogreportwithmaptypeas“routepoints”isdisplayedbelow.

Warranty

YourSensitPMDiswarrantedtobefreefromdefectsinmaterialsandworkmanshipforaperiodoftwoyearsafterpurchase(excludingcalibration).Ifwithinthewarrantyperiodtheinstrumentshouldbecomeinoperativefromsuchdefectstheinstrumentwillberepairedorreplacedatouroption.Thiswarrantycoversnormaluseanddoesnotcoverdamagewhichoccursinshipmentorfailurewhichresultsfromalteration,tampering,accident,misuse,abuse,neglectorimpropermaintenance.Proofofpurchasemayberequiredsbeforewarrantyisrendered.Unitsoutofwarrantywillberepairedforaservicecharge.InternalrepairormaintenancemustbeperformedbyaSensitTechnologiesauthorizedtechnician.Violationwillvoidthewarranty.Unitsmustbereturnedpostpaid,insuredandtotheattentionoftheservicedepartmentforwarrantyorrepair.

Thiswarrantygivesyouspecificlegalrightsandyoumayhaveotherrightswhichvaryfromstatetostate.

SensitTechnologies851TransportDriveValparaiso,Indiana46383USA

Tel:219/465-2700Fax:219/465-2701

Email:[email protected]:www.gasleaksensors.com

SENSITPMDINSTRUCTIONMANUAL V2.15-6-2013