SD BIOLINE
HBsAg WB
Quick Reference GuideTest pouch
(Test Device)
Preparation
Check Point !
QU-01FK10WEN Rev. 1402-00
Tel : +82-31-899-2800 Fax : +82-31-899-2840 E-mail : [email protected]
Test Procedure
Check the expiry date.If expiry date has passed, use another kit.Check the desiccant.If the color is green, use another kit.
1
Add drawn sample into the sample well (“S” marked).
6
Prepare the serum/plasma/whole blood sample.
4
Wait 20 minutes after adding sample.Don’t read test results after 20 minutes.Reading too late can give false results.
7
Put on the gloves.
2
Open the test pouch and write the patient’s name on the test device.
3
Use the micropipette to collect sample (100µl).
5
YYYY/MM/DD
Caution: Each test can be used ONLY ONE TIME.
Do not use the test more than once.
www.standardia.com
Yellow Color Green Color
Overview of the Device
Desiccant
Timer(Not Provided)
Disposable Gloves(Not Provided)
Micropipette & Disposable Tip(Not Provided)
Interpretation
No “C” line in the result window.The specimen should be retested using a new test kit.
POSITIVE RESULT NEGATIVE RESULT INVALID RESULT
HBsAg WB HBsAg WB
HBsAg WB HBsAg WB
HBsAg WB HBsAg WB
HBsAg WB HBsAg WB
HBsAg WB HBsAg WB
The presence of two color bands (“T” band and “C” band) within the results window no matter which band appears first indicates a positive result.
The presence of only one band (“C” band) within the results window indicates a negative result.
HBsAg WB
HBsAg WB
Control Line
Test Line
Result Window
SampleWell
100µl