Upload
vuongnhi
View
219
Download
0
Embed Size (px)
Citation preview
Scanning electron microscopy
Tabletop
Fei Quanta
Hitachi
Example: Tin soldier
Pb M
Sn L
Secondary electrons Backscatter electrons EDS analysis – Average composition
Learning goals: • Understanding the principle
– Of the instrument
– Of type of signals
– detectors
• Understand how the parameters that you can/must change influences the picture and chemical analysis.
• Lab: Demonstration and hands on experience with our three different SEMs– Table top SEM
– Environmental SEM
– High resolution SEM
Microscopes then and now
1930
1970
2018
Limits to resolution
Unaided eye ~ 0.1 mm
Light microscope ~ 0.2 mm
Scanning EM ~ 1.0 nm
Transmission EM ~ 0.1 nm
The higher the accelerating voltage, the smaller the wavelength of the electrons and the higher the possible achievable resolution.
In SEM, there are several electromagnetic lenses,including condenser lenses and one objective lens.Electromagnetic lenses are for electron probeformation, not for image formation directly, as inTEM. Two condenser lenses reduce the crossoverdiameter of the electron beam. The objective lensfurther reduces the cross-section of the electronbeam and focuses the electron beam as probe onthe specimen surface.
Instrument: SEM is it like a TEM ?
Objective lens
Objective lens
Cross sectionThe final probe-forming lens has to
operate with a relative long working
distance (WD= 10 mm), that is the
distance between the specimen and
lower pole-piece. This is necessary
so that the emitted radiation can be
collected and detected with desired
efficiency.
The long working distance increases
the spherical aberration of the probe-
forming lens, which increases the
size of the smallest attainable
electron-beam spot.
Final Lens
•Probe Forming
•Labeled as Focus
Objective lens designs
Electron beam
SED
Virtual lens
Specimen
Out lens In lens Semi-in-lens
☆Easy to observe magnetic sample
☆Possible to observe bigger sample
☆Topographical imaging☆Deep depth of Focus
☆Ultra high resolution☆High throughput observation
at sample exchange position☆Variety of signal detecting
system to optimize the contrast
Features Features Features
☆Ultra high resolution☆ Possible to observe bigger
sample at short WD☆Variety of signal detecting
system to optimize the contrast
SED
Electron beam
Virtual lens
Specimen
Specimen
Virtual lens
Electron beam
SED
SEDSED
Compare the three different designs of objective lens :
1:Conventional lens
Here the magnetic field of the lens is located inside the lens and the
specimen sits away from the field.
This leads to that the beam travels “unprotected” between lens and sample
and is more vulnerable to EM disturbance. On the other hand the distance
between lens and specimen only has limited effect on the signal strength.
2: Snorkel lens, semi-in-lens or immersion lens
Here the magnetic field is projected down below the lens to enclose the
specimen if it is located at short working distance. At the same time as the
beam is protected from EM disturbance the electrons are effectively captured
and led up through the lens to be detected by an in-lens detector. For longer
working distance or when a side illumination effect is wanted, a classical SE
detector is mounted in the chamber.
3: In-lens
Here the sample sits on a TEM-type holder and has a maximum size of ca.
4x4x9 mm.
Benefits here are excellent mechanical stability, high electron collection
efficiency and very high EDS count rates.
Relationship between resolution and focal length
Different e- sources and gun types
> 5 years18 months6 months1 monthLife time
0.2 eV0.5 eV1.5 eV2.0 eVEnergy spread, E
1000
2x109
500
5x108
10
1x107
1
1x106
Brightness
[A/cm2sr]
Room temp1500 C1500 C2300 CTemperature
3 - 5 nm10 – 25 nm1 – 2 mm1 – 2 mmSource size
Cold FESchottky FELaB6W
10-11 Torr10-9 Torr10-7 Torr10-5 TorrGun vacuum
20-30 nA>100 nA50 nA -1 mA50 nA -1 mAProbecurrent
2-3 %0.2 %0.2 %0.1 %Stability,%/h
Magnification of SEM is determined by the ratio of the linear size of the display screen
to the linear size of the specimen area being scanned. The linear magnification is given by
Electronbeam is scanned across the specimen and the procedure is known as Raster scanning. Raster scanning causes the beam tosequentially cover a rectangular area on the specimen. The signal electrons emitted from the specimen are collected by the detector,amplified and used to reconstruct the image according to one-to-one correlation between scanning points on the specimen and picturepoints on the screen of cathode ray tube (CRT). CRT converts the electronic signals to a visual display.
Condenser Lens Current and Resolution.The current in the condenser lens changes the spot size or
diameter of the beam of electrons that scans the sample.
More detailed information will be collected when the
electron beam scans the same area with a smaller spot size.
An increased current or a higher number for the condenser
lens (CL) setting, will produce a smaller spot size and in
general will result in a better resolution (A).
DiatomeCourtesy of GokhanTaken by Quanta SEM microscopeMagnification: 30000xSample: DiatomeDetector: ETDVoltage: 4.0kVVacuum: 4.01e-4PaHorizontal Field Width: 9.93μmWorking Distance: 7.7Spot: 2.0https://www.fei.com/image-gallery/
Magnification ?Same picture - different size
Beam samples interaction:
Scattering“Inelastic” scattering refers to a variety of physical processes that act to progressively reduce the energy of the beam electron by transferring that energy to the specimen atoms through interactions with tightly bound inner-shell atomic electrons and loosely bound valence electrons.
Although the various inelastic scattering energy loss processes are discrete and independent, Bethe (1930) was able to summarize their collective effects into a “continuous energy
where E is the beam energy (keV), Z is the atomic number, ρ is the density (g/cm3), A is the atomic weight (g/mol), and J is the “mean ionization potential” (keV) given by
Beam interaction volumes
Images are formed because of beam interaction with the sample
This happens in a volume, not in a point
The size of this volume varies with beam energy...
1μm
Vacc : 10kV Vacc : 1kV
Sample = Si
Beam Excitation Volumes
Elastic scattering
The elastic scattering crossection, can be used to estimate how far the beam electron must travel on average to experience an elastic scattering event, a distance called the “mean free path,” λ
Simultaneously with inelastic scattering, “elastic scattering” events occur when the beam electron is deflected by the electrical field of an atom (the positive nuclear charge as partially shielded by the negative charge of the atom’s orbital electrons), causing the beam electron to deviate from its previous path onto a new trajectory, as illustrated schematically in the figure.
The mean free path is of the order of nm. Elastic scattering is thus likely to occur hundreds to thousands of times along a Bethe range of several hundred to several thousand nanometers.
Accelerating voltage
一
一
一
一
一SE
Surface informationSE I
SEInner information
Electron beam
Z
SE escape depth
Low energy BSE (II)
sample
SE II
一
一一
一一
一
一
一一
一
SE
一一
一
High energy BSE (I)
Z
BSE escape depth
Signals in the SEM
Emitted electrons
1 10 100 1000 10000
Energy of signal electron (eV)
Ele
ctr
on
am
ou
nt
Secondary electron Back scattered electron
(SE) (BSE)
50eV
When a sample is hit by an electron beam a variety of electron emissions are available.(Notice that the scale is logarithmic)
“Huskelapp”
The group of secondary electrons (SE) can be divided into 4 groups.
There are SEI, SEII, SEIII, and SEIV.
These electrons are classed according to how they are generated.
A SEI is an electron is an electron that is generated at the point of primary beam impingement in surface of the
specimen. Thus, it carries the highest resolution information.
A SEII is an electron that is generated when a backscattered electron leaves the surface of the specimen. Due to
the energy of backscattered electrons, this SEII could leave the surface of the specimen microns away from the
primary beam impingement site. SEIIs hurt the resolution of the image, but add greatly to overall image brightness.
A SEIII is an electron released when an energetic backscattered electron strikes the interior of the specimen
chamber, causing a SE to be released.
SEIVs are formed when the primary beam strikes an aperature within the electron column.
SEIIIs and SEIVs contribute noise to the image. By understanding signal formation, the specimen can be properly
prepared for analysis.
SE yield variation
The rapid change in the incident electron beam range causes a large, characteristic variation in the SE yield
Typically the yield rises from ~0.1 at 30keV to in excess of 1 at around 1keV, and as high as 100 for some materials
Experimental SE yield data for Ag
Why the SE yield changes
SE escape depth is ~ 3-5nm
At high energies most SE are produced too deep to escape so the SE yield d is low
But at lower energies the incident range is so small that most of the SE generated can escape so the SE yield rises rapidly
At very low energies fewer SE are produced because less energy is available so the SE yield falls again
interaction volumes
low
voltagehigh
voltage
Do high and low kV SE images look the same?
compared to the high energy
•The image looks less 3-D
•Highlighting is absent
•Surface junk is more visible
28
Emitted electrons – SE:s
Occurrence of edge effect in fine structured surfaces.
Change in SE yield at different incident e-beam angles.
Resolution
At high energy the SE1 signal typically comes from a volume 3-5nm in diameter, but the SE2 signal from a volume of 1-3µm in diameter.
But at low energythe SE1 and SE2 electrons emerge from the same volume because of the reduction in the size of the interaction volume
Resolution at ultra-low energies
• Because Cs and Cc decrease with the landing energy the imaging resolution is only limited by diffraction
500eV30eV
100nm
BSE – Backscatter electrons
一
一
一
一
一SE
Surface informationSE I
SEInner information
Electron beam
Z
SE escape depth
Low energy BSE (II)
sample
SE II
一
一一
一一
一
一
一一
一
SE
一一
一
High energy BSE (I)
Z
BSE escape depth
BSE vs. SE detection
Sample : Polyvinyl AlcoholAccelerating Voltage : 3kV, Vacuum: 60Pa, Mag. : 1,000x
Sample : Polyvinyl AlcoholAccelerating Voltage : 3kV, Vacuum: 60Pa, Mag. : 1,000x
BSE detector SE
33
Emitted electrons – BSE:s
The count ratio of BSE:s depends on the mean atomic number of the specimen.
Detectors on the SEM•Electron Detectors:
•° Everhart-Thornley (E-T) detector
•Located on one side and has a small solid angle of detection.
Detectors on the SEM
•Electron Detectors:
•° Solid-State Diode detector•Located on pole piece and has a large solid angle of detection
•Electron-hole pairs are produced by the action of high energy
•backscattered electrons. Annular detector split into two
•semi-circles, A and B.
•A+B: compositional mode
•A-B: topographic mode
SU8000Series standard optics
TopUpper
Lower
STEM Detector
Signal type
Signal name
Detector information
BSEHA-BSE
Top Composition, crystal
BSELA-BSE
UpperComposition + Topo(Charge suppression)
SE SE UpperSurface information
(Including voltage contrast)
SE Lower Lower Topo
STEMBF-
STEM※1
STEMSample internal information +
Crystal
STEMDF-
STEM※1
LowerSample internal information +
Composition
SE
BSE
STEM
Electrode
Variety of signal detection system in SU8200-series
37
Upper
Electrode
Sample
EXB
SE BSE
Top
Topographical image with shadowLower
Sample : Photocatalyst
Vacc : 3.0V
Mag. : x 50k
courtesy of :
Nagaoka University of Technology, Faculty of Engineering, Dr. Kazunori Sato
Model :SU8020
Lower detector
LOWER(SE-L)
Upper
EXB
Lower
Top
SE BSE
UPPER(SE)
Sample : Photocatalyst
Vacc : 3.0V
Mag. : x 50k
courtesy of :
Nagaoka University of Technology, Faculty of Engineering, Dr. Kazunori Sato
Model :SU8020Surface information(incl. Voltage contrast)
Control electrode
Electrode
Upper detector: Pure SE
Sample
Upper
EXB
Lower
Top
SE BSE
Sample : Photocatalyst
Vacc : 3.0V
Mag. : x 50k
courtesy of :
Nagaoka University of Technology, Faculty of Engineering, Dr. Kazunori Sato
Model :SU8020Topographical + Compositional information
UPPER(LA-BSE)Electrode
Upper detector: SE Filtering, LA-BSE signal
Control electrode
Sample
Upper
EXB
Top
Lower SE BSE
Sample : Photocatalyst
Vacc : 3.0V
Mag. : x 50k
courtesy of :
Nagaoka University of Technology, Faculty of Engineering, Dr. Kazunori Sato
Model :SU8020Compositional + Crystal information(Less topographical information)
TOP(HA-BSE)Electrode
Top detector: High-angle BSE
Control electrode
Sample
LOWER (SE-L) UPPER (SE)
UPPER (LA-BSE)TOP (HA-BSE)
Deceleration (cathode lens)
-VR
Vi = Vo - VR
Objective lens
Specimen
Primary beam
VR = 0 V
Vi = 0.5kV
VR = 1.5kV
Vi = 0.5kV
In deceleration, a negative voltage is applied to the specimen to decelerate primary electrons before arriving at specimen surface.
Deceleration
Landing voltage: Vi
-VR: Deceleration voltage
Expands
Optimum αiImproved resolution
Normal mode
TopUpper
Accelerated BSE
SEAccelerated SE
Vd
1 Both of the SE and BSE are accelerated bythe deceleration voltage(negative bias)
2 Makes it more difficult to only select the low energy electrons
3 Sample bias and the accelerated BSE:s increase the signal and contrast
Top : Low energy signal
Upper : High energy signal
But if we accelerate the SE:s ?
Lecture 2
Scanning Electron Microscopy
Lecture II – EDS, EDX
BackscatteredElectrons (BSE)
Cathodoluminescence
AugerElectrons
Characteristic X-rays
SecondaryElectrons (SE)
Transmitted Electrons
Absorbed Electrons
Electron Beam
Heat
Parameters , imaging , conclusion?We can change1. Accelerating voltage2. Magnification3. Spot size4. Working distance5. Beam current6. Detectors
We want:High resolution
We have choosen low Accelerating voltage
Ask questions in lab:
We will repeat inlens detectorsNext week
Plan:
• X-rays
• Detectors
• Qualitative analyses
• Quantitative analyses
• Data from lab(?)
• Rapport
SE and BSE
Average
PbM
Sn L
Atomic energy levels and line transition
Transitions have different probabilities
Lines have different intensities
EDS spectra: Origin of Bremsstrahlung and characteristic peaks
1 keV = 1.602677·10-16 J
continuum or Bremsstrahlung (breaking radiation)
• results from deceleration of beam electrons in the electromagnetic field of the atom core
• combined with energy loss and creation of an X-ray with the same energy
EDS spectra: Origin of Bremsstrahlung and characteristic peaks
Hmmm …….
Observerte data fra SEM + EDS
- Characteristic X-rays are formed by excitation of inner shell electrons
- Inner shell electron is ejected and an outer shell electron replaces it
- Energy difference is released as an X-ray
2 4 6 8 10 12 14
keV
0
1
2
3
4
5
cps/eV
C Si Cr Cr Mn Mn
Fe
Fe
Ni Ni
EDS spectra: Origin of Bremsstrahlung and characteristic peaks
If beam energy E > EK then a K-electron may be excited
Energy of emitted photon can be calculated:
EPhot = E1 – E2
e.g.: Fe L → K
E1 = EK = 7.11 keV
E2 = EL = 0.71 keV
EKa = 6.40 keV
X-ray energy is the difference between two energy levels !
EDS spectra: Origin of Bremsstrahlung and characteristic peaks
0.71 keV
7.11 keV
6.40 keV
X-ray and AUGER generation process
Emission of Auger electronEmission of X-ray
Auger and X-ray yield are competing processes
C
Ge
Fluorescence yield (ω)
• ω= fraction of ionisation events producing characteristic X-rays (rest produce Auger electrons)
+ A = 1- ω increases with Z- ω for each shell: ωK ωL ωM- Auger process is favoured for low Z, - fluorescence dominates for high Z
0.005 for C K
0.5 for Ge K
L-familie
Kα
Kβ
Fe K-familie
• Energy of characteristic peaks is defined by element• The higher the atomic number Z the higher the peak energy
Characteristic peaks: K, L, M series
S (16)
K1,2 2308 eV
K 2464 eV
(K - K1,2) 156 eV
S (16) Ca (20)
K1,2 2308 eV 3692 eV
K 2464 eV 4013 eV
(K - K1,2) 156 eV 319 eV
S (16) Ca (20) Mn (26)
K1,2 2308 eV 3692 eV 5900 eV
K 2464 eV 4013 eV 6492 eV
(K - K1,2) 156 eV 319 eV 592 eV
The K-family of lines
Mo (42) Ce (58)
L1 2292 eV 4839 eV
L 2394 eV 5262 eV
Ll 2014 eV 4287 eV
Mo (42)
L1 2292 eV
L 2394 eV
Ll 2014 eV
The L-family of lines
Line intensity relations (2)
Lα1
Lβ1
Lβ2
Lγ1LlLγ2/3
Lα2
Spectrum Barium L series, 15 kV
α1 : β1 : γ1 = 100 : 52 : 10
Line intensity relations (3)
Spectrum Barium L-series, 15 kV
α1 : β1 : γ1 = 100 : 52 : 10
Ti-Kα1
Ti-Kβ1
BaTiO3
Barium
Spectrum BaTiO3, 15 kV
Overlapped Ba L-series and Ti K-series
Lα1 (100)
Lβ1 (31)
Lβ2
Lγ1 (5)Ll
Lγ2/3
Lα2
Lα1
Lβ1
Lβ2
Lγ1LlLγ2/3
Lα2
Intensity and energy of characteristic lines
- Energy of line is defined by
- Element
- Type of transition
- Intensity of line is defined by
- probability of producing a hole (vacancy)
- probability of electron transition
- probability of x-ray emission
- concentration
Probability of producing a hole: Ionization cross-section for electrons
• Ionization cross section: probability of excitation
• maximum ionization cross section: 2,5 x Ebind
Ionisation cross section for electrons
Important for the selection
of accelerating voltage.
Ionization cross section for electrons
Fe
Cr
Ni
Cr: 33%
Fe: 33%
Ni: 33%
U = 10 keV
:bind
exc
E
E Cr Fe Ni
1,847 1,561 1,337
A sample with equal amount of Cr,Fe and Ni.
30 kV and 10 kV
Fe/Cr=0,8Ni/Cr=0,7
Fe/Cr=0,5Ni/Cr=0,15
• Isolated Atoms
• When isolated atoms are considered, the probability of an
• energetic electron with energy E (keV) ionizing an atom by
• ejecting an atomic electron bound with ionization energy Ec
• (keV) can be expressed as a cross section, QI:
where ns is the number of electrons in the shell or subshell
(e.g., nK = 2), and bs and cs are constants for a given shell
(e.g.,bK = 0.35 and cK = 1)E= beam energy Ec= ionization energy
http://www.med.harvard.edu/jpnm/physics/refs/xrayemis.html
5.) X-ray range
Different excitation ranges for: - characteristic x-ray radiation and Bremsstrahlung,
- secondary electrons (SE)- back-scattered electrons (BSE)
sample surface
electron beam (E0)
secondary electrons
ca. 0.5 ... 5 µm
ca. 10 µm3
back-scattered electrons
bremsstrahlung
X-rays
MEF4010 Scanning electron microscopy
Detectors on the SEM
•X-ray Detectors
•° EDS Spectrometer
•° WDS Spectrometer
X-RAY-detectors
WDX wavelength dispersive EDX Energy dispersive
Energy resolution is 100 times better with the WDX.
125-150 eV EDX
10 eV WDX
Kvalitativ analyse
Hva har vi tenkt på
• Valg av akselerasjonsspenning
• Utsnitt på prøven
• Vinkel – tilting av prøven
• Arbeidsavstand (WD)
• Telletid
• Oppladning (?)
• Mål:
• Gode data med god tellestatistikk og god oppløsning
Vi kan finne ut mer
Quantitative
5.) X-ray range
• Monte Carlo electron-trajectory simulations of interaction volume in iron as function of primary beam energy
With higher primary electron energy penetration depth is increasing
Rd ≈ 2,5 µmRd ≈ 1,3 µmRd ≈ 0,4 µm
EHT = 10 kV EHT = 20 kV EHT = 30 kV
http://www.gel.usherbrooke.ca/casino/What.html
5.) X-ray range
• Monte Carlo electron-trajectory simulations of interaction volume as function of atomic number (EHT = 15 kV)
With higher density penetration depth is decreasing
Carbon Rd ≈ 2 µm Iron Rd ≈ 0,6 µm Gold Rd ≈ 0,2 µm
The kV compromise
Ichar increases with increasing E0/Ec
X-ray signal improves
Rx increases with increasing E0/Ec
X-ray spatial resolution degrades
00
C
EU 2 ... 2,5
E
Optimum overvoltage
Without sufficient overvoltage, x-ray production is dramatically lowered.
QE (cm2 keV2)
U
U = Eo/Ec
Optimum overvoltage U is around 2-2.5
The ionization cross-section describes the probability that a particular event will take place:
For X-rays: Q = 6.5x10-20nsbs
UEc
ln(csU)
ns = number of electrons in a shell
Ec = critical excitation voltage
bs & cs = constants related to the electron shell
U = overvoltage (Eo/Ec)
ZAF
Can't calculate ZAFs unless concentration is known. But we don’t know the concentration??
Use k values (I/I° = k) to estimate compositions of each element. Then calculate ZAFs, and refine by iteration
Z – at. no. correctionFunction of electron backscattering factor &electron stopping power - depend upon the average atomic number of unknown and standardDependent on ionization cross section
Varies with composition and accelerating voltage
A – absorption correctionVaries with m, takeoff angle,accelerating voltage
F – fluorescence correctionprimary fluorescent x-rays ––>secondary fluorescent x-rays
Varies with composition and accelerating voltage
CalcZAF
http://www.probesoftware.com/Technical.html
http://www.probesoftware.com/download/PROBEWIN.pdf
kV=15 vinkel = 40
Mass absorption coefficientsMass absorption coefficients are stored as a matrix of numbers of absorption of a particular X-ray line (the emitter) by an absorber: For example, a portion of the MAC matrix for Kα X-rays for Z = 23 to 29 is shown below.
http://www.ammrf.org.au/myscope/analysis/eds/quantitative/
Mass absorption coefficients
Emitter V 4952 eV
Cr 5415 eV
Mn 5899
eV
Fe 6403 eV
Co 6930
eV
Ni 7478 eV
Cu 8048
eV
V 94.6 73.8 498.4 403.4 328.2 268.5 220.8
Cr 111.1 86.7 68.4 454.8 370.8 303.9 250.4
Mn 125 97.6 76.9 61.3 401.9 330.1 272.4
Fe 145 113.2 89.3 71.1 57.1 370.2 306
Co 160.9 125.7 99.1 79 63.5 51.4 329.4
Ni 187.9 146.8 115.8 92.3 74.1 60 49
Cu 200.7 156.8 123.8 98.7 79.3 64.2 52.4
Note that the MAC for absorption of Fe Kα by Co (79.0) is different from the MAC for absorption of Co Kα by Fe (57.1).
Sample surface and absorption
rough surface
polished surface
tilted polished surface
X-rays
electron beam
take-off angle
negative tilt angle
-TA
SDD
d1
d2
d1: distance to (imaginary) polished surface
d2: actual distance to specimen surface