Sample Analysis Design – Step 2 – Calibration/Standard ...asimonet/ENGV60500/Lecture_8_10_11_2011.pdf · Sample Analysis Design – Step 2 – Calibration/Standard Preparation

Embed Size (px)

Citation preview

  • Sample Analysis Design Step 2 Calibration/Standard Preparation

    Choice of calibration method dependent upon several factors:

    1. potential matrix effects

    2. number of samples

    3. consistency of matrix across samples

  • Sample Analysis Design Step 2 Calibration/Standard Preparation

    EXTERNAL CALIBRATION: Prepare a set of standard solutions to cover the

    expected range of analyte concentrations

    Fit a least squares regression line

    y = mx + b

    and calculate analyte concentration in unknowns

  • Sample Analysis Design Step 2 Calibration/Standard Preparation

    23Na calib curve (Medium resolution)

    y = 17557xR2 = 0.9992

    0

    100000

    200000

    300000

    400000

    500000

    600000

    700000

    800000

    900000

    1000000

    0.00 10.00 20.00 30.00 40.00 50.00 60.00

    conc ppb

    cps

  • Sample Analysis Design Step 2 Calibration/Standard Preparation

    44Ca calib curve (Medium resolution)

    y = 676.92xR2 = 0.9961

    0

    5000

    10000

    15000

    20000

    25000

    30000

    35000

    40000

    0.00 10.00 20.00 30.00 40.00 50.00 60.00

    conc ppb

    cps

  • Sample Analysis Design Step 2 Calibration/Standard Preparation

    Advantages of External Calibration

    Easy to prepare

    Quick

    Widely used technique

  • Sample Analysis Design Step 2 Calibration/Standard Preparation Disadvantages of External Calibration:

    Need to matrix match calibration solutions and samples

    If standards containing

  • Sample Analysis Design Step 2 Calibration/Standard Preparation Preparation of External Calibration Solutions:

    Need to evenly space calibration concentrations

    If the highest concentration is much higher than the rest, linear regression introduces bias favoring the high point

    X = independent variable = concentration

    Y = dependent variable = counts/second

  • Sample Analysis Design Standard Addition Method

    Aliquots of spike are added to unknown samples to increase the ion signal intensities for elements of interest

    Typically use at least three aliquots of sample spiked with evenly spaced amounts of analyte

    These spiked aliquots of sample are used to generate a calibration line and calculate the concentration in the sample

  • Sample Analysis Design Standard Addition Method

    S0 = unspiked sample

    S1 = sample spiked with analyte at concentration x

    S2 = sample spiked with analyte at concentration 2x

    S3 = sample spiked with analyte at concentration 3x

    S4 = so on and so on

  • Sample Analysis Design Standard Addition Method

    AMT

    y = 29387x + 279235R2 = 0.9992

    0

    50000

    100000

    150000

    200000

    250000

    300000

    350000

    400000

    450000

    500000

    0 1 2 3 4 5 6 7

    Concentration (ppb)

    Cps

  • Sample Analysis Design Standard Addition Method

    The concentration of the unknown solution is then determined by dividing the y-intercept value by the slope of the sample-spike mixing line.

    From example on previous slide,

    Conc. soln = 279235 / 29387 = 9.5 ppb

    If the original sample was a solution, then this is the concentration of the analyte in question in the solution

  • Sample Analysis Design Standard Addition Method

    If the original sample was in solid form that you digested and subsequently converted into a solution;

    then in order to determine the concentration of the analyte in question, you must factor in the amount of total analyte in the solution and the dry weight of the sample powder

  • Sample Analysis Design Standard Addition Method

    If we continue with the same example, the solution has a concentration of 9.5 ppb, and the original volume of the unknown solution was 10 ml (g) prior to aspirating some of it into the plasma for analysis, then the total amount analyte in the solution is:

    = 10 g x 9.5 ng/g (ppb) = 95 ng, or = 0.095 g

    If the amount of powder weighed out was 0.1 g, then the concentration of the element in question is:

    Conc. = 0.095 g/0.1 g = 0.95 g/g or ppm

  • Sample Analysis Design Standard Addition Method

    This method works best if the slope of the calibration line is not too shallow

    This will create more uncertainty in the location of the intersection between the cps of your unknown and the calibration line

  • Sample Analysis Design Standard Addition Method

    For maximum precision its necessary that the amount of sample be the same in each aliquot

    Also want the amount of spike added to be the same for each aliquot

    Amount of spike added should be as small as possible (usually 0.1 ml to 10 ml total volume)

  • Sample Analysis Design Standard Addition Method

    Ideally, the highest spike concentration should be approximately equal to the concentration of analyte in the unknown

    Need to have some idea of the concentration in the sample prior to analysis

  • Sample Analysis Design Standard Addition Method

    Advantages:

    Overcomes matrix differences More precise and accurate than external calibration

    Disadvantages:

    Requires at least three aliquots for each sample Run lengths become much longer and more

    preparation time is required

  • Sample Analysis Design Isotope Dilution

    Most accurate and precise calibration method available

    Requires analyte with two stable isotopes

    Monoisotopic elements cannot be determined via isotope dilution

    Spike natural sample with enriched isotope spike of analyte

  • Sample Analysis Design Isotope Dilution

    The amount of spike is selected so that the resulting ratio between spiked isotope and unspiked isotope is near unity maximizes precision

    Typically use the most abundant isotope as the reference -- maximizes sensitivity

  • Sample Analysis Design Isotope Dilution

    Check isotope ratio in unspiked sample to determine if the natural ratio in the sample matches with the predicted ratio

    If not -- interference in acting on one or both of the isotopes

    Always attempt to use interference free isotopes

  • Sample Analysis Design Isotope Dilution

    Prepare the spike to desired concentration

    Add spike as early as possible after equilibration of spike and sample you dont have to have complete sample recovery

    During any stage of the process complete equilibration is absolutely necessary

  • Sample Analysis Design Isotope Dilution

    Analyze the solution on the ICP using many repetitive scans (to maximize precision)

    Need to measure isotopic ratios on standards of a known ratio in order to correct for machine mass discrimination

    Use previous equation to calculate concentrations!

  • Sample Analysis Design Isotope Dilution

    Advantages:

    Most accurate and precise method for quantitative elemental concentrations

    Partial loss of analyte during preparation is compensated for since physical and chemical interferences are not an issue -- will cancel out as they will affect each isotope identically

    Ideal form of internal standardization since another isotope of the same element is used in this capacity

  • Sample Analysis Design Isotope Dilution

    Disadvantages:

    Generally only applicable to multiple-isotopic elements

    Need an enriched isotope spike for the analyte of interest - not always available or sometimes at very high cost

    Need two interference free isotopes VERY time consuming

  • Sample Analysis Design

    STEP 3 INTERNAL STANDARDIZATION & INSTRUMENT

    DRIFT CORRECTION

  • Sample Analysis Design Internal Standard

    Every sample should be analyzed with an internal standard (IS)

    What is an internal standard (IS)?

    element that is added to EVERY sample/ blank/calibration standard/QA sample/etc., that is not expected to be in the sample in appreciable quantities and is not an element of interest

    use IS to monitor machine drift (both short and long term) and matrix effects

  • Sample Analysis Design Internal Standard

    Choice of IS depends upon which elements you are quantifying

    The IS should have similar properties in the plasma as element(s) of interest

    ICP-MS: similar in mass/ionization potential

  • Sample Analysis Design Internal Standard

    Example:

    attempting to quantify U - use Th

    attempting to quantify most transition metals - use As

    attempting to quantify REEs - use Re

    115In and 103Rh are common IS for general use

    alternatively, you can add several IS to each sample

  • Sample Analysis Design Internal Standard

    From previous slide, we assume that samples have little or no Th, As, or Re

    Its important to have an idea of whats in your sample prior to quantitative analysis

    Solid samples can use a naturally occurring element as IS, provided that you know the concentration in each sample

  • Sample Analysis Design Internal Standard

    Procedure for IS use:

    Calculate the concentration of the IS in each centrifuge tube the latter will contain an aliquot of your sample and an aliquot of the IS

    Divide the measured ion signal (CPS) by the concentration of your IS to derive the factor = CPS/ppb

    Divide CPS/ppb of each tube by the CPS/ppb for those measured for the blanks since these are not influenced by possible effects due to sample matrices

    The latter yields a dimensionless correction factor (I refer to it as a normalization factor)

    Use correction factor to adjust analyte counts for drift or matrix effects

  • Sample Analysis Design Internal Standard

    Advantages:

    Fluctuations are monitored in each sample/ calibration / blank

    Disadvantages:

    Assume that behavior of IS is the same as the analyte

  • Sample Analysis Design Instrumental Drift

    Correct for instrument drift with:

    Internal standardization is a common procedure

    Use of drift corrector solutions (DCS)

  • Sample Analysis Design Instrumental Drift

    Drift Corrector Solutions (DCS):

    Measure the same solution intermittently throughout the course of the analytical session

    Change in ion signal is assumed to be linear between each DCS measurement

  • Sample Analysis Design Instrumental Drift

    The DCS should contain all elements of interest and can be matrix matched to samples

    Example: use standard reference materials (SRMs) for DCS

  • Sample Analysis Design Instrumental Drift

    Apply a linear correction to samples between DCS solutions

    DCS1 + ((DCS2 - DCS1)*F)

    F = position dependent fraction

  • Sample Analysis Design Instrumental Drift

    Advantages of DCS correction:

    all analytes are monitored for drift

    nothing added to sample solutions

    Disadvantages of DCS correction:

    assume change is linear

    cannot easily monitor matrix effects

  • Sample Analysis Design Background & blanks

    Standard blank - blank used to monitor polyatomic ion interferences, gas peaks, and contamination from reagents; used for background subtraction

    Procedural blank - blank used to monitor contamination acquired during all stages of sample preparation; grinding, digestion, acidification, powdering, etc

  • Sample Analysis Design Background & blanks

    Use of blanks during an analytical session:

    ALWAYS begin an analytical session with at least one standard blank

    Analyze standard blanks periodically throughout the course of the session in particular to monitor memory effects

    Process and analyze at least one procedural blank at some point during your research study; for its analysis, its preferable to measure it early in order to avoid any potential memory effects

  • Sample Analysis Design Background & blanks

    The more standard blanks that are run during an analytical session, the more information you will have with regards to monitoring change(s) in background levels throughout the entire session

  • Sample Analysis Design Background & blanks

    How to determine the background:

    1. just use the first standard blank

    2. average all standard blanks

    3. take median of all standard blanks

    4. apply statistical analysis to standard blanks and select some of them

  • Sample Analysis Design Background & blanks

    Outlier tests:

    1. I know the truth

    2. Looks different

    3. Statistical proof

  • Sample Analysis Design Background & blanks

    Option 1 should be avoided - unscientific and invalid

    Option 2 is better but only if the measurement is repeated

    Option 3 is the best approach, but needs to be carried out carefully in order to avoid false negatives and positives

  • Sample Analysis Design Background & blanks

    Huber Outlier Test

    take median of all values

    calculate absolute deviation |xi - xm|

    take mean of absolute deviations (MAD)

    multiply MAD by coefficient (k = 3-5)

    anything higher than k*MAD is rejected as outlier

  • Sample Analysis Design Background & blanks

    Calculation of Limit of Detection (DL) and Limit of Quantification (QL)

    Easy way: LOD = 3*STDEVblank; LOQ = 10*STDEVblank

  • Sample Analysis Design SUMMARY

    A good analytical method will:

    1. provide the means to calculate an accurate background level

    2. allow for correction of instrument drift

    3. use Internal standardization to monitor matrix effects

    4. provide some method for monitoring/ correcting interferences

    5. Use a proper calibration strategy

  • Example Calculation Determination of Ca and Na in

    beetle blood

    Using External Calibration Method