Spark probability measurement Spark probability measurement for GEM for CBM for GEM for CBM (Summary of the beam test at CERN SPS, (Summary of the beam test at CERN SPS, October 2011) October 2011) Saikat Biswas, A. Abuhoza, U. Frankenfeld, C. Garabatos, J. Hehner, T. Morhardt, C.J. Schmidt, H.R. Schmidt, J. Wiechula GSI Detector Laboratory RD51 Mini week, 13-15 June 2012, CERN 1

Saikat Biswas , A . Abuhoza , U . Frankenfeld , C. Garabatos ,

Embed Size (px)

DESCRIPTION

Spark probability measurement for GEM for CBM (Summary of the beam test at CERN SPS, October 2011). Saikat Biswas , A . Abuhoza , U . Frankenfeld , C. Garabatos , J. Hehner , T. Morhardt , C.J. Schmidt , H.R. Schmidt, J. Wiechula GSI Detector Laboratory. - PowerPoint PPT Presentation

Citation preview

Page 1: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

Spark probability measurement for Spark probability measurement for GEM for CBM GEM for CBM (Summary of the beam test at CERN SPS, (Summary of the beam test at CERN SPS, October 2011)October 2011)Saikat Biswas, A. Abuhoza, U. Frankenfeld, C. Garabatos, J. Hehner, T. Morhardt, C.J. Schmidt, H.R. Schmidt, J. Wiechula

GSI Detector Laboratory

RD51 Mini week, 13-15 June 2012, CERN

1

Page 2: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

Outline of the talkOutline of the talkMotivationTest set-upAnalysis and ResultsSummary and future plan

2

Page 3: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

GEM for CBM GEM for CBM Triple GEM as a precise tracking

detector in the Muon Chamber (MUCH) under the extreme conditions of the CBM experiment

3

Page 4: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

ObjectiveObjectiveTo measure the properties of

GEM with shower and in particular Spark probabilities of Double mask and Single mask GEM

4

Page 5: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

Summary of beam testSummary of beam testDetectors

◦ 2 Double mask GEM

◦ 1 Single mask GEM

Measurement with◦ Pion beam◦ Pion beam with

absorber: Shower

Measured parameters◦ Current◦ Voltage◦ Trigger and GEM

Counts◦ GEM signal

5

Page 6: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

Voltage distribution in Voltage distribution in GEMGEM

6

Page 7: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

Details of the set upDetails of the set upGas mixture: Ar/CO2: 70/307 channel HVG210 power supply2 sum-up boards are used for

signal (2×128 6×6 mm2 pads) for DM GEM

4 sum-up boards are used for signal (4×128 4×4 mm2 pads) for SM GEM

PXI LabView based DAQ is used 7

Page 8: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

8

Set-up for Pion beamSet-up for Pion beam

Page 9: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

9

Set-up for showerSet-up for shower

Page 10: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

Ref. A. Senger 10

Particle production during Particle production during shower from FLUKA simulationshower from FLUKA simulation

Page 11: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

11

Comparison of shower number Comparison of shower number from measurement and from measurement and simulationsimulation

Page 12: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

CurrentCurrent

12

Page 13: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

13

Current and GEM counting rate:Current and GEM counting rate:Pion beam 300 kHzPion beam 300 kHz

Page 14: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

14

Current and GEM counting rate Current and GEM counting rate during Shower: Beam rate300 during Shower: Beam rate300 kHzkHz

Page 15: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

Pion beam Pion beam with absorber

15

Current as a function of Current as a function of rate for DM GEMrate for DM GEM

Page 16: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

Charge Vs. current for DM Charge Vs. current for DM GEMGEM

16

Pion beamPion beam with absorber

Slope: -1.38×10-12

Slope: -2.04×10-12

Page 17: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

Pion beam Pion beam with absorber

17

Current as a function of rate for Current as a function of rate for SM GEMSM GEM

Page 18: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

Charge Vs. current for SM Charge Vs. current for SM GEMGEM

18

Pion beamPion beam with absorber

Slope: -1.35×10-12

Slope: -1.52×10-12

Page 19: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

19

EfficiencyEfficiency

Page 20: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

20

Efficiency during showerEfficiency during shower

Page 21: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

21

Efficiency as a function of rate Efficiency as a function of rate during showerduring shower

Page 22: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

Efficiency for pion beamEfficiency for pion beam

22

Page 23: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

23

Efficiency vs. rate for pion Efficiency vs. rate for pion beambeam

Page 24: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

24

Spark probability Spark probability measurementmeasurement

Page 25: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

Methods of Spark detectionAbsence of signal

◦Drop in the counting rate of GEM signals

◦Data from sampling ADCDetection of high current

◦Sudden increase in the Current (Slow)

◦Built in Trip checker in HVG210 Power supply (Fast)

25

Page 26: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

Double Mask GEM with Fe Absorber

Gas: Ar/CO2 : 70/30, Gas flow rate: 5 lt/hr, Particle rate: ~300 kHz, Pion beam

415_410_405

26

No spark during a spillNo spark during a spill

Page 27: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

Double Mask GEM with Fe Absorber

Gas: Ar/CO2 : 70/30, Gas flow rate: 5 lt/hr, Particle rate: ~300 kHz, Pion beam

415_410_405

27

Drop in GEM counting rateDrop in GEM counting rate

Page 28: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

Double Mask GEM with Fe Absorber

Gas: Ar/CO2 : 70/30, Gas flow rate: 5 lt/hr, Particle rate: ~300 kHz, Pion beam

415_410_405

28

Sudden increase in Sudden increase in currentcurrent

Page 29: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

Double Mask GEM with Fe Absorber

Gas: Ar/CO2 : 70/30, Gas flow rate: 5 lt/hr, Particle rate: ~300 kHz, Pion beam

412 - 407 - 402

415_410_405

29

Two sparks during a spillTwo sparks during a spill

Page 30: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

Spark probability vs. global Spark probability vs. global voltagevoltagefor showerfor shower

Discharge probability:

No. of Discharge/ No. of incident particle

30

Page 31: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

31

Spark probability vs. global Spark probability vs. global voltagevoltageshower and pion beamshower and pion beam

Page 32: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

32

Spark probability vs. gainSpark probability vs. gainshower and pion beamshower and pion beam

Page 33: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

33

Spark probability vs. global Spark probability vs. global voltagevoltageSM and DMSM and DM

Page 34: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

34

Off spill spark rate as a function Off spill spark rate as a function of global voltageof global voltage

Page 35: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

SummarySummary SPS test line has good conditions for our purpose 2 mm drift gap sub-optimal (3 mm standard!) Efficiency

◦ Rate dependency of efficiency observed Pion (signal close to threshold!) Shower (signal below threshold! Pick-up noise)

Spark probability◦ Spark measurement reliable also with noise (high

thresholds)◦ Comparable spark probability for pion beam and

shower (high rate) !◦ Higher spark probabilities for lower intensities

(shower) SM GEM

◦ Was in conditioning phase.◦ No indication for different performance

35

Page 36: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

Future plan: test beam Future plan: test beam Optimized drift gap (3 mm)Conditioned counters (SM and

DM)Pixel readout ?

36

Page 37: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

AcknowledgementAcknowledgement

Thanks to the RD51 collaboration for their support in the beam test….

Thank you for your kind attention !37

Page 38: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

Back up slidesBack up slides

38

Page 39: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

ConclusionConclusionThe spark probability for pion

beam is high.May be the gain is not measured

correctly!!Effect of space charge !!Investigated in different

conditions. to be understood the different

spark probabilities.

39

Page 40: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

40

Pulse height distributionPulse height distribution

Page 41: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

100 sample is taken

Difference of the maximum and minimum value of the channel is taken as pulse height

41

MethodMethod

Page 42: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

Resolution ~17.6%

42

FeFe5555 spectrum @ 400-395- spectrum @ 400-395-390 V390 V

Page 43: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

43

For DM GEM at 400-395-390 For DM GEM at 400-395-390 with pion beam: Rate 300 kHzwith pion beam: Rate 300 kHz

Page 44: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

44

For DM GEM at 415-410-405 For DM GEM at 415-410-405 during shower: Beam rate: 300 during shower: Beam rate: 300 kHzkHz

Page 45: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

45

For SM GEM at 400-395-390 For SM GEM at 400-395-390 with pion beam: Rate 300 kHzwith pion beam: Rate 300 kHz

Page 46: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

46

For SM GEM at 405-400-395 For SM GEM at 405-400-395 during shower: Beam rate: 300 during shower: Beam rate: 300 kHzkHz

Page 47: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

Geometry of the experimental set-Geometry of the experimental set-upup

47

Page 48: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

For SM GEM at 400-395-390 with For SM GEM at 400-395-390 with pion beam: Rate 300 kHzpion beam: Rate 300 kHz

48

Page 49: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

Spill:Rbeam

<R*beam> > 0.5

Spark:

CGEM

Cbeam

<RGEM>

<Rbeam>

< 0.2

Spill

49

DefinitionsDefinitions

Page 50: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

Gain as a function of global voltage Gain as a function of global voltage for SM GEMfor SM GEM

50

Page 51: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

Corrected voltage for Corrected voltage for GEM3GEM3

51

Page 52: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

Discharge probability:

No. of Discharge/ No. of incident particle

Ref: S. Bachmann et al., NIM A 470 (2001) 548–561

52

Discharge probability as a Discharge probability as a function of gainfunction of gain

Page 53: Saikat Biswas , A .  Abuhoza , U .  Frankenfeld , C.  Garabatos ,

Spark rate as a function of Spark rate as a function of global voltageglobal voltage