Upload
luan-nguyen
View
682
Download
70
Tags:
Embed Size (px)
DESCRIPTION
ndt
Citation preview
1
TABLE OF CONTENT
INDEXPAGE
1.0 Scope
2.0 References
3.0 Definition
4.0 Qualification and Personnel
5.0 Radiographic Equipment
6.0 Preparation
7.0 Location & Datum
8.0 Extent of Inspection
9.0 Cassettes
10.0 Screens
11.0 Film
12.0 Geometric Relationship
13.0 Image Quality Indication ( IQI )
14.0 Sensitivity
15.0 Film Density
16.0 Scattered Radiation
17.0 Radiographic Technique
18.0 Film Processing
19.0 Film Identification
20.0 Viewing Condition
21.0 Interpretation of Radiograph
22.0 Radiographic Report
23.0 Safety
24.0 Acceptance Criteria
APPENDIX
Appendix 1 ( Exposure Technique )
Appendix 2 ( Instruction Technique Sheet )
1
1
1
1 & 2
2
2
2
3
3
3
3
4
4, 5, 6, 7, 8
8
8
8 & 9
9 & 10
10 & 11
12
12
12
1213
13
14
15
1.0 SCOPE
This procedure describes the radiographic techniques to be employed by TWI /CIBD-Johor for the radiography inspection of butt welds in accordance with ASME, Article 2,section V.
2.0 REFERENCES
ASME Sec V Article 2
Atomic Energy Licensing Act 1984
Ionizing Radiation Regulation (Licensing) 1986
Ionizing Radiation Regulation (Basic Safety Standard) 1988
Ionizing Radiation Regulation (Transportation) 1989
Qualification of Personnel- PCN/CSWIP/MLVK/ASNT/ACCP
3.0 DEFINITION
CIDB - Construction Industry Development Board
LPTA- Lembaga Perlesenan Tenaga Atom(Atomic Energy Licensing Board)
NDT- Non Destructive Testing
PCN- Personnel Certification in Non Destructive Testing
CSWIP-Certification Scheme in Welding Inspection Personnel
IQI- Image Quality Indicator
MLVK-Majlis Latihan Vokasional Kebangsaan
ASNT-American Society of Non-Destructive Testing
ACCP-ASNT Central Certification Program
1.0 QUALIFICATION OF PERSONNEL
Radiography inspection work shall be carried out by at least one Radiographer and one Assistant Radiographer.Each technician shall be required to hold the following:
a) Radiographer
A valid LPTA Radiographic Operator license and a current Certificate of Proficiency in Radiography Lev II to a nationally recognized scheme.(MLVK)
b) Assistant Radiographer
Registered with LPTA as a Radiation Worker and a current Certificate of Proficiency in Radiography Lev I or II to a nationally recognized scheme.
c) Radiographic Interpreter
A current Certificate of Proficiency in CSWIP/PCN Radiographic Interpreter Lev II or ACCP RT Lev II or Lev III. Alternative qualifications shall only be acceptable with the prior consent of TWI.
5.0RADIOGRAPHIC EQUIPMENTOnly X-ray units to be usedMuller 260kv Macrotank or equivalent.These may be directional or panaromic units,up to 300kv.
6.0 PREPARATION
Radiographic inspection shall only be carried out on weldment which has been visually inspected and any remedial work necessary has to be completed and accepted by TWI.In addition,Radiographer inspects the weld surface for any irregularities that may mask or confuse interpretation of the final radiograp
7.0LOCATION OF DATUM
All welds will be positively marked datum and each portion under examination shall be positively identified by markings along the length of the weld.
This may be achieved by taking measurements around the circumference from an agreed datum or by means of the clock wise method.For pipes,the datum line shall be positioned,as follows:
For horizontal and inclined pipes,the datum is the highest point in the joint when viewed from the lower numbered joint.
For vertical pipes,the datum is taken from north,when the joint is viewed vertically downwards.
In all cases the datum shall be indelibly marked on the welds. A meter band incorporating lead numerals along the weld shall be utilized to allow positive location of any discontinuity.
8.0 EXTENT OF INSPECTION
Coverage of inspection shall be 100% of weld,Heat Affected Zone & parent material adjacent to weld zone.
9.0 CASSETTES
Flexible envelope cassettes shall be used unless specifically stated otherwise.Ensure close Film-To-Screen and Cassette-To-Object contact.
10.0 SCREENS
Lead intensifying screens shall be used. The front and rear screen thickness shall be 0.125mm unless specified by TWI.
Screen shall be examined on a regular basis for dirt or blemishes which may transfer onto the film thus causing spurious indication which may mask out the actual discontinuity or defect.
11.0 FILM
11.1 Type of Flim
Radiographs shall be made using industrial radiographic film and shall be
selected from Table 1 below.
Class Brand Type
I Agfa
Kodak D4
MX
II Agfa
Kodak D7
CX
11.2 Overlap
Radiographs shall have sufficient overlap to ensure that no portion of weld remains unexamined.There shall be a specific lead marker visible on both radiographs in the area of overlap.
12.0 GEOMETRIC RELATIONSHIP
12.1 The geometric unsharpness is calculated as follows: -
Ug = F x OFD
FOD
F = Focal size
OFD = Object To Film Distance
FOD = Focal To Object Distance
12.2 Geometric Unsharpness of the radiograph shall not exceed the following: -
Material Thickness Ug Maximum(Inches)
Below 2 inches 0.020
2-3 inches 0.030
3-4 inches 0.040
Greater than 4 inches 0.070
13.0 IMAGE QUALITY INDICATOR
13.I ASME Type
The IQI provides a guide to the quality of radiograph produced and in strict compliance with ASME V, shall be appropriately choosen from Table 2 below.A DIN54 109 wire type IQI may be used as an alternative to ASME IQI.
a) Source Side Penetrameter(s)
The penetrameter shall be placed on the source side of the part being examined,except in the condition described in 12.1(b)
b) Film Side Penetrameter
Where inaccessibility prevents hand placing the penetrameters on the source side,it shall be placed on the film side in contact with the part being examined. A lead letter F shall be placed at least as high as the penetrameters indentification number in adjacent to or on the penetrameters but shall not mask the hole.
Radiography shall be performed with a technique that will produce a
sufficient sensitivity from the penetrameter image or specified hole, which are essential to assess the quality of a radiograph. Each radiograph shall display its own identification number.
Penetrameter
DesignationPenetrameter
Thickness,in 1T Hole
Diameter,in 2T Hole
Diameter,in 4T Hole
Diameter,in
5 0.005 0.010 0.020 0.040
7 0.0075 0.010 0.020 0.040
10 0.010 0.010 0.020 0.040
12 0.0125 0.0125 0.025 0.050
15
0.015 0.015 0.030 0.060
17 0.0175 0.0175 0.035 0.070
20 0.020 0.020 0.040 0.080
25 0.025 0.025 0.050 0.100
30 0.030 0.030 0.060 0.120
35 0.035 0.035 0.070 0.140
40 0.040 0.040 0.080 0.160
45 0.045 0.045 0.090 0.180
50 0.050 0.050 0.100 0.200
60 0.060 0.060 0.120 0.240
70 0.070 0.070 0.140 0.280
80 0.080 0.080 0.160 0.320
100 0.100 0.100 0.200 0.400
120 0.120 0.120 0.240 0.480
140 0.140 0.140 0.280 0.560
160 0.160 0.160 0.320 0.640
200 0.200 0.200 0.400 ..
240 0.240 0.240 0.480 ..
280 0.280 0.280 0.560 ..
Table 2: ASME Hole Type IQI Designation
13.2 DIN 54109 Wire Type IQI
The IQI provides a guide to the quality of radiographs produced and shall be appropriately chosen from one of the types recommended in DIN 54109(Table 3) as an alternative to ASME penetrameter.
a) Source Side Penetrameter
The penetrameter shall be placed on the source side of the part being
examined,except in the condition described in 13.2 (b).
b) Flim Side Penetrameter
Where inaccessibility prevents hand placing the penetrameter on the source side it shall be placed on the film side in contact with the part being examined.
A lead letter F at least as high as the penetrameter indentification numbers shall be placed adjacent to the penetrameter.
Radiography shall be performed with a technique that will produce a
sufficient sensitivity from the penetrameter image or specified hole, which are essential to assess the quality of a radiograph. Each radiograph shall display its own identification number.
Table 3: DIN54109 Wire Type IQI Designation
Penetrameter Designation Wire TypeWire Diameter,mm
1-7
1
2
3
4
5
6
7 3.200
2.500
2.000
1.600
1.250
1.000
0.800
6-12
6
7
8
9
10
11
12 1.000
0.800
0.630
0.500
0.400
0.320
0.250
10-16
10
11
12
13
14
15
16 0.400
0.320
0.250
0.200
0.160
0.125
0.100
13.2 Number of Penetrameters
For component where one or more film holders are used for an exposure,at least one penetrameter image shall appear on each radiograph except the following:-
a) For piping or tubular component ,where the source is located at the center of the component and one or more film holders are used for a single exposure of complete circumference(Panaromic Technique),at least three penetrameters shall be spaced approximately 120 apart.
b) For piping or tubular component,where the source is placed on the axis of the object(Double- Wall Double- Image Viewing) and four or more film holders are used for a single section of the circumference,at least three penetrameters shall be used.One penetrameter shall be in the approximate center of the section exposed and one at each end.
c) For piping or tubular component where the source is located at the center of the component (Single-Wall Single-Viewing Technique) and four or more film holders are used for an exposure of a circumference weld,at least three penetrameters shall be used.One penetrameter shall be in the approximate center of the section exposed and one at each end.
14.0SENSITIVITY
14.1 The required sensitivity for each radiograph using ASME Hole Type IQI
is 2-2T or better.ASME Hole Type IQI is calculated using the following
formula:-
IQI Sensitivity = (HT/2) x 100%
Thickness of Penetrated Material In Same Region
H = Diameter of The Smallest Hole Visible on Radiograph
T = Thickness of Penetrameter Used
14.2 The required sensitivity for each radiograph using DIN 54109 Type IQI
is 2% or better.DIN54109 Wire Type IQI is calculated using the
following formula:-
IQI Sensitivity = Diameter of Thinnest Visible Wire On Radiograph x 100%
Thickness of Penetrated Material In Same Region
15.0 FILM DENSITY
For radiographs made with an X-Ray tube,film density will be measured in the area of interest and shall be not less than 1.8 and not greater than 3.0.
For radiographs made with Gamma source,film density will be measured in the area of interest and shall be not less than 2.0 and not greater than 3.0.
16.0 SCATTERED RADIATION
Wherever necessary 6mm thick lead sheet shall be placed behind the film cassette to prevent scattered radiation from fogging the film.
The cassette shall have a lead letter Ba minimum of 12.5mm high, fastened to the back and if a light image of the letter B appears on the film, additional
backing sheets are required. A light image of the letter B shall be called rejection of any radiographic film.
17.0 RADIOGRAPHIC TECHNIQUE
Radiographic Inspection techniques shall be carried out in accordance with the current edition of ASME Section V Article 2.
The specific radiographic parameters to be used shall be detailed on the technique sheets attached to this procedure in Appendix 2.
A single wall exposure technique shall be used for radiography whenever practical. When it is not practical to use a single-wall technique, a double wall technique shall be used.
17.1 Single-Wall Technique
In a single-wall technique, the radiation passes through only one wall of the weld(material),which is viewed for acceptance on the radiograph. An adequate number of exposures shall be made to demonstrate that the required coverage has been obtained.
Note: Appendix 1 Figure 1 & 2 illustrate the single-wall exposure setup that may be used.
17.2 Double-Wall Technique
When it is not practical to use a single wall technique, one of the following double wall technique shall be used.
17.2.1 Single-Wall Viewing
For materials and for weld components, a technique will be used in which the radiation passes through the two walls and only the weld (material) on the film sidewall is viewed for acceptance. An adequate number of exposures shall be made to demonstrate that the required coverage has been obtained.
Wherecomplete coverage is required for circumferential welds (material), a minimum of three exposures taken 120 to each other shall be made.
17.2.2 Double-Wall Viewing
For materials and for welds in components 3-1/2 inch or less in nominal outside diameter, a DWDI technique will be used in which the radiation passes through two walls and the weld (material) in both walls is viewed for acceptance on the same radiograph. For double wall viewing only a source side penetrameter shall be used. Care should be exercised to ensure that the required geometric unsharpness requirement is not exceeded,if so then a single wall viewing shall be used.
a) For welds, the radiation beam will be offset from the plane of the weld at an angle sufficient to separate the images of the source side and film side portions of the weld so that there is no overlap of the areas to be interpreted. When complete coverage is required, a minimum of two exposures taken 90 to each other shall be made for each joint.
b) As an alternative, the weld may be required to be radiographed
with the radiation beam positioned so that the images of both
wall are superimposed. When complete coverage is required a
minimum of three exposures taken at 60( or 120( to each other
shall be made for each joint.
c) Additional exposed shall be made if the required radiographic
coverage cannot be obtained using the minimum number of exposures indicated in a) or b) above.
Note: Appendix 1 Figures 3,4 and 5 illustrate the double-wall exposure setups that
that may be used.
18.0 FILM PROCESSING Processing should be carried out using the standard developer and fixer solution, and shall be in accordance with manufacturers recommendations. The performance of the processing will be checked daily using control strips of X-ray film exposed to a known density and processed. From the developed film, a check is taken and any density change is marked on a process control chart. Upper and lower limits are marked on the chart to provide controls against which the plot can be monitored. The results are interpreted and corrective action taken if necessary. The radiographs shall be free from imperfections due to processing or any other defects which would interfere with interpretation.
18.1.2 Developing
Films will be placed in hangers and immersed in developer solution with constant agitation for even development. Developing time shall be in accordance to that recommended by film manufacturer. It shall normally be 5 min. At 20( C or (68(F). If temperature of the developer varies, development time should be adjusted accordingly.
18.1.2 Stop Bath or Rinse
After development is completed, the film shall be immersed in stop bath or rinsed in clean running water to stop the activity of the developer.
18.1.3 Fixing
Fixing shall be done twice the cleaning time but not more than 15 min. in fresh fixer.
18.1.4 Washing
Washing shall be done with constant flow of clean fresh water. Rate of water flow shall be approximately 5 to 6 times the volume of water tank per hour. After washing, the film shall be rinse in bath of Photo flo (wetting agent) for 1 min to minimize the clinging of water droplets on the films.
18.1.5 Drying
The film shall either be left to dry on the drying racks or dried by the drying machine (Agfa Gavaert Structurix dryer).
18.1.6 Chemical Used
Developer Agfa Structurix G 128
Fixer Agfa Structurix G 328
Film and processing chemical shall be used from the same compatible brand.
19.0 FILM IDENTIFICATIONEach film will be identified by the used of lead symbols to indicate the following minimum information: -
a) The job or workpiece
b) The weld number
c) The section of the weld
d) The date of the test
e) The welders identification
20.0 VIEWING CONDITIONS
The radiographs shall be examined in a darkened room. The radiograph viewer shall provide a light source sufficient for the essential IQI wire/hole to be visible for the specific density range. The light from around the outer edges of the radiographs will be masked.
21.0 INTERPRETATION OF RADIOGRAPHS
A calibrated densitometer or calibrated density strip will be kept in close proximity to the viewer and a spot high-density illuminator will be available. All comment with regard to film quality and interpretation will be recorded on the film folder and on the radiographic report form.
The density of density strips and densitometer calibration shall be verified by comparison with a calibrated step wedge film traceable to a national standard. The densitometer shall be calibrated in accordance with paragraph 5 of SE-1079, Calibration of Transmission Densitometer.
22.0 RADIOGRAPHIC REPORT
Written report shall be prepare giving the result of all the tests performed using the TWI/CIDB Radiographic Report Format containing all relevant information in regards to Radiography Inspection.
23.0 SAFETY
All radiographic work will be carried out in accordance with Statutory Safety Requirement and with TWI/ CIDB (Johor) Sdn. Bhd. Radiation Safety Standing Instructions.
All radiography Inspection work shall be in accordance with LPTA ruling and safety requirement.
24.0 ACCEPTANCE CRITERIA
The ruling code for defects evaluation on plate and piping welds shall be in accordance with ASME IX 1999 edition (Article 1 section QW- 191.2.2)
APPENDIX 1
( EXPOSURES TECHNIQUE )
APPENDIX 2
( INSTRUCTION TECHNIQUE SHEETS)
TECHNIQUE SHEET PAGE 1 of 3
RADIOGRAPHIC EXAMINATION OF WELDS
(PLATE) RT-01
1.0 Material: Carbon Steel or Stainless Steel
2.0 Scope: Applicable to butt weld on plate 6 mm thick and above
3.0 Equipment type: Muller 260 kV Macro tank
4.0 Radiation Source: 2.3 mm x 2.3 mm, 260 kV X-ray machine
5.0 Technicians Qualification:
a) Operator hold valid LPTA Operator License PCN/ CSWIP/ACCP RT Lev.
b) Assistance Operator registered with LPTA as with LPTA as Radiation
worker and hold PCN/ CSWIP/ ACCP RT level
c) Interpreter Holds CSWIP / PCN RI Lev. or ACCP RT Lev. /III
6.0 Technique: Single wall single image (SWSI)
7.0 Geometric relationship: Radiation beam at 90 to weld and film
8.0 Focal to Film Distance: 16 inch or greater
9.0 Film type: Agfa D7 or equivalent grain size
10.0 Film coverage: Film will be equidistance spaced plus 50mm overlap between
Adjacent exposure
11.0 Quantity of film: ) 4 x 10 Length of interest 8
) 4 x 15 Length of interest 13
12.0 Intensifying screen: Lead (Pb) 0.125mm front and rear 0.125mm
TECHNIQUE SHEET PAGE 2 of 3
RADIOGRAPHIC EXAMINATION OF WELDS
(PLATE) RT-01
13.0 Exposure: mAmin x FFD2 FFD1 where mAmin is as follows
Thickness mAmin Kv Ma used
6 mm 21 mAmin 140 Kv 4 mA
9 mm 35 mAmin 140 Kv 4 mA
14.0 Film factor: Change of film from Cx to D7
Film Type Cx D7
Film Factor 2.5 3.5
Film Factor D7 X Original exposure Film Factor CX
Original exposure (plate 6mm) = 1.03 min
New exposure = 3.5 X 1.03
2.5
New exposure time = 1min 26 sec
15.0 Image Quality indicator (IQI): Din 62 Fe wire type, shall be placed on the source
side on each center of the diagnostic length of
radiograph
t < 20mm 10-16
20 < t > 50 mm 6-12
t > 50 mm 1-7
16.0 Geometric Unsharpness: Less than 0.5mm
TECHNIQUE SHEET PAGE 3 of 3
RADIOGRAPHIC EXAMINATION OF WELDS
(PLATE) RT-01
17.0 Processing (manual): Developer - Agfa Structurix G 128
Fixer - Agfa Structurix G 328
Stop bath - Agfa
Photo Flo - Agfa
18.0 Sensitivity: 2 % or better
19.0 Density: Not less than 2.0 and not greater than 3.0 through weld metal
20.0 Acceptance Criteria: ASME 1X 1999 edition (Article 1 section Q rev- 1.91.2.2)
FORMULATED BY: ____________________________ DATE: ______________
Shajahaan Prem
ASNT / ACCP Lev. 115270
APPROVED BY: _____________________________ DATE: ______________
QUALITY ASSURANCE MANAGER
TECHNIQUE SHEET PAGE 1 of 4
RADIOGRAPHIC EXAMINATION OF WELDS
(PIPE above 2- DWSI) RT-02
1.0 Material: Carbon Steel or Stainless Steel
2.0 Scope: Applicable to butt weld on pipe OD above 2, 7 mm thick and above
3.0 Equipment type: Muller 260 KV Macro tank
4.0 Radiation Source: 2.3 mm x 2.3 mm 260 kV X-ray machine
5.0 Technicians Qualification:
b) Operator hold valid LPTA Operator License PCN/ CSWIP/ACCP RT Lev.
b) Assistance Operator registered with LPTA as with LPTA as Radiation
worker and hold PCN/ CSWIP/ ACCP RT level
c) Interpreter Holds CSWIP / PCN RI Lev. or ASNT RT Lev.
6.0 Technique: Double wall single image (DWSI)
7.0 Geometric relationship: Radiations beam offset just outsides weld toe
8.0 Focal to Film Distance: Outside diameter (OD) + distance from focal spot (x)
9.0 Film type: Agfa D7 or equivalent grain size
10.0 Number of exposure: 3 taken of 120 to each other
11.0 Film coverage: Film will be equidistance spaced plus 50mm overlap between
adjacent exposure
12.0 Quantity of film: ) 4 x 10 OD x 3.14 for each exposure
24
) 4 x 15 OD x 3.14 for each exposure
39
TECHNIQUE SHEET PAGE 2 of 4
RADIOGRAPHIC EXAMINATION OF WELDS
(PIPE above 2- DWSI) RT-02
13.0 Intensifying screen: Lead (Pb) 0.125mm front and rear 0.125mm
14.0 Exposure: mAminX FFD2 FFD1 where Ma-min is as follows
Thickness Ma - min Kv Ma used
7.1 mm 35 mAmin 160 Kv 4 mA
10.97 mm 40 mAmin 180 Kv 4 mA
15.0 Film factor: Change of film from Cx to D7
Film Type Cx D7
Film Factor 2.5 3.5
Film Factor D7 X Original exposure Film Factor CX
Original exposure (7.1 mm) = 1.72 min
New exposure = 3.5 x 1.72
2.5
New exposure time = 2 min 24 sec
Original exposure (10.97 mm) = 1.97 min
New exposure = 3.5 x 1.97
2.5
New exposure time = 2 min 45 sec
TECHNIQUE SHEET PAGE 3 of 4
RADIOGRAPHIC EXAMINATION OF WELDS
(PIPE above 2- DWSI) RT-02
16.0 Image Quality indicator (IQI)
One to three films in single exposure
Din 62 Fe wire type, one on each center of the diagnostic length of radiograph
Four or more film in single exposure
Din 62 Fe wire type, one at center of each section exposed and one at each end.
t < 20mm 10-16
20 < t > 50 mm 6-12
t > 50 mm 1-7
17.0 Geometric Unsharpness: Less than 0.5mm
18.0 Processing (manual): Developer - Agfa Structurix G 128
Fixer - Agfa Structurix G 328
Stop bath - Agfa
Photo Flo - Agfa
19.0 Sensitivity: 2 % or better
20.0 Density: Not less than 2.0 and not greater than 3.0 through weld metal
21.0 Acceptance Criteria: ASME 1X 199 edition (Article 1 section Q rev- 1.91.2.2) TECHNIQUE SHEET PAGE 4.of 4
RADIOGRAPHIC EXAMINATION OF WELDS
(PIPE above 2- DWSI) RT-02
FORMULATED BY: ____________________________ DATE: ______________
Shajahaan Prem
ASNT / ACCP Lev. 115270
APPROVED BY: _____________________________ DATE: ______________
QUALITY ASSURANCE MANAGER
TECHNIQUE SHEET PAGE 1 of 4
RADIOGRAPHIC EXAMINATION OF WELDS
(PIPE 2 & below- DWDI) RT-03
1.0 Material: Carbon Steel or Stainless Steel
2.0 Scope: Applicable to butt weld on pipe OD below 2, 5.5 mm thick and above
3.0 Equipment type: Muller 260 KV Macro tank
4.0 Radiation Source: 2.3 mm x 2.3 mm 260 kV X-ray machine
5.0 Technicians Qualification:
c) Operator hold valid LPTA Operator License PCN/ CSWIP/ACCP RT Lev.
b) Assistance Operator registered with LPTA as with LPTA as Radiation
worker and hold PCN/ CSWIP/ ACCP RT level
c) Interpreter Holds CSWIP / PCN RI Lev. or ASNT RT Lev.
6.0 Technique: Double wall double image (DWDI)
7.0 Geometric relationship: Radiations beam offset, of local to film distance
approximately 4.
8.0 Focal to Film Distance: 16 or greater
9.0 Film type: Agfa D7 or equivalent grain size
10.0 Number of exposure: 2 taken at 90 to each other
11.0 Quantity of film: One (1) for each exposure
12.0 Intensifying screen: Lead (Pb) 0.125mm front and rear 0.125mm
TECHNIQUE SHEET PAGE 2 of 4
RADIOGRAPHIC EXAMINATION OF WELDS
(PIPE 2 & below- DWDI) RT-03
13.0 Exposure: mAminX FFD2 FFD1 where Ma-min is as follows
Thickness Ma - min Kv Ma used
5.5 mm 60 mAmin 140 Kv 4 mA
14.0 Film factor: Change of film from Cx to D7
Film Type Cx D7
Film Factor 2.5 3.5
Film Factor D7 X Original exposure Film Factor CX
Original exposure (5.5 mm) = 2.96 min
New exposure = 3.5 x 2.96
2.5
New exposure time = 4.14 min
TECHNIQUE SHEET PAGE 3 of 4
RADIOGRAPHIC EXAMINATION OF WELDS
(PIPE 2 & below- DWDI) RT-03
15.0 Image Quality indicator (IQI):
One to three films in single exposure
Din 62 Fe wire type, one on each center of the diagnostic length of radiograph
Four or more film in single exposure
Din 62 Fe wire type, one at center of each section exposed and one at each end.
t < 20mm 10-16
20 < t > 50 mm 6-12
t > 50 mm 1-7
16.0 Geometric Unsharpness: Less than 0.5mm
17.0 Processing (manual): Developer - Agfa Structurix G 128
Fixer - Agfa Structurix G 328
Stop bath - Agfa
Photo Flo - Agfa
18.0 Sensitivity: 2 % or better
19.0 Density: Not less than 2.0 and not greater than 3.0 through weld metal
20.0 Acceptance Criteria: ASME 1X 199 edition (Article 1 section Q rev- 1.91.2.2)
TECHNIQUE SHEET PAGE 4 of 4
RADIOGRAPHIC EXAMINATION OF WELDS
(PIPE 2 & below- DWDI) RT-03
FORMULATED BY: ____________________________ DATE: ______________
Shajahaan Prem
ASNT / ACCP Lev. 115270
APPROVED BY: _____________________________ DATE: ______________
QUALITY ASSURANCE MANAGER
PAGE 1
_1123677998.psd