129
Role of imaging in structural interventions M. Chrissoheris Attending, THV Department ROUND TABLE L Role of imaging in atrioventricular valves

Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

  • Upload
    others

  • View
    7

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Role of imaging in structural interventions

M. ChrissoherisAttending, THV Department

ROUND TABLE L Role of imaging in atrioventricular valves

Page 2: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Disclosures

- Proctoring activities for Abbott Vascular

I and the HYGEIA Hospital «Heart Team» have receivedresearch and/or travel grants and/or lecture fees from:- ABBOTT Vascular, Europe

HYGEIA Hospital Heart Team

Cardiologists: G. Kourkoveli, K Papadopoulos, A Halapas, M Chrissoheris, K Spargias

CT Surgeons: N Bouboulis, S Skardoutsos, A Tsolakis, S Pattakos

Anesthesiologist: C Nastoulis, I Nikolaou

Pediatric Cardiology: A Tzifa

Vascular Surgeons: I Belos, S Kaliafas

Radiologists: F Laspas, C Mourmouris

Electrophysiology: L Papavassiliou, S. Kourouklis, G Zervopoulos, T Apostolopoulos

Research Nurse: E. Dafnomyli

Page 3: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Daniel H. Steinberg et al. Eur Heart J Suppl 2010;12:E2-E9

Spectrum of Structural Heart Disease

Page 4: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Structural Heart Disease: Conditions for Successful Treatment

• Accurate assessment of nature of defect

• Patient selection

• Planning of procedure

• Precision and complication avoidance

– Guidance, monitoring, verification of success, early

identification of complications

• Patient follow up Imaging is Critical

Page 5: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

“…Are there not twelve hours in the day? If anyone walks in the day, he does not stumble, because he sees the light of this world. But if anyone walks in the night, he stumbles, because the light is not in him.” (Jn. 11:9)

Page 6: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Structural Heart Disease: Role for Echocardiography

• Diagnosis

• Guidance during therapeutic interventions (Interventional Echocardiography)

Page 7: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Echocardiographic Modalities

• Transthoracic

• Intracardiac

• Transesophageal

• Tomographic imaging

• 3-Dimensional

• Doppler

– Color

– Spectral

• Speckle tracking

• Contrast (agitated saline)

Page 8: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Transthoracic Echocardiography

• Portable

• Non-invasive

• No radiation or contrast

• Anatomic– Tomographic and 3-

Dimensional imaging

• Physiologic information– Doppler

– Contrast

• Common denominator for all patients referred for evaluation is a comprehensive examination (incl’d. ventricular size / function, valvularfunction, intracardiacshunt determination, hemodynamic assessment)

Page 9: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Intracardiac Echocardiography (ICE)

• US probe inserted via transvenous access and advanced in right atrium Near field structures

• Under conscious sedation

• Variability in image acquisition

• Not standardized image planes

• Single use Cost

Page 10: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Transesophageal Echocardiography

• Superior image resolution

• Standardized imaging planes

• No interference with sterility during procedures Ideal for guidance during interventions

• Ability for real time 3-Dimensional imaging

Page 11: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Real Time 3-Dimensional Echo

Focused wide sector (3D-Zoom) Narrow sector (Live 3D)

Page 12: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Fusion Imaging: Fluoroscopy and Echo

Page 13: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Structural Heart Disease and Catheter Based Intervention

• Closure of Intracardiac Shunts– Atrial Septal Defects

• Secundum

– Patent Foramen Ovale– Ventricular Septal Defects

• Congenital• Post Myocardial Infarction

– Ventricular pseudoaneurysms

• Valvular Heart Disease– Aortic Stenosis / Regurgitation– Mitral Regurgitation– Mitral Stenosis

• Prosthetic Heart Valve Dysfunction– Degeneration– Paravalvular Regurgitation

• Left atrial appendage closure• Pulmonary vein ablation• Other (patent ductus closure, coarctation of the aorta, congenital aortic /

pulmonic stenosis)

Page 14: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Atrial Septal Defects

• Type of defect

– Secundum, primum, venosum

– Coexistent pathology (e.g. anomalous pulmonary venous return)

• Hemodynamic impact

– Chamber size, function

– Shunt direction

– Qp/Qs calculation

– Pulmonary pressures

Page 15: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Atrial Septal Defects: Secundum

• Anatomy for Intervention

– Evaluation for rims

– Size, shape, complexity

• Transesophageal Echo

– Modality of choice

– Complemented by 3D

Arch Cardiol Mex 2012;82(1):37-47

Page 16: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Atrial Septal Defects: Secundum

• Anatomy for Intervention

– Evaluation for rims

– Size, shape, complexity

• Transesophageal Echo

– Modality of choice

– Complemented by 3D

Page 17: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Secundum ASD: TEE Evaluation for Closure Device

Arch Cardiol Mex 2012;82(1):37-47

Page 18: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Secundum ASD: Device Closure under Real Time 3D Echocardiography

Intraoperative

guidance

– Follow catheters

– Monitoring while

device is

positioned

– Accurate

positioning

– Residual shunt

Page 19: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Patent Foramen Ovale

• Common congenital abnormality (up to 35%)

• Associations with:

– Cryptogenic stroke, paradoxical thromboembolism, migraines, decompression illness, orthodeoxia)

• Agitated saline infusion for diagnosis

Otto C. The Practice of Clinical Echocardiography, 4th Edition

Page 20: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Patent Foramen Ovale

• Common congenital abnormality (up to 35%)

• Associations with:

– Cryptogenic stroke, paradoxical thromboembolism, migraines, decompression illness)

• Agitated saline infusion for diagnosis

J A C C : C A R D I O V A S C U L A R IMA G I N G V O L . 7 , N O . 3 , 2 0 1 4

Page 21: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

PFO: Device Closure under Real Time 2D Echocardiography

Page 22: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Ventricular Septal Defects

• Type of defect

– Congenital (perimembranous, muscular, supracrystal, inflow)

– Post myocardial infarction

• Hemodynamic impact

– Chamber size

– Shunt direction

– Pulmonary pressures

– Qp/Qs calculation

Post MI VSD

Dan G. Halpern et al. Eur J Echocardiogr 2009;10:569-571

Page 23: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Ventricular Septal Defects

• Anatomy for intervention– Location

– Adjacent structures (aortic valve cusps, tricuspid valve)

– Sizing with multiplanar reconstruction

• Intraoperative guidance– Follow catheters

– Monitoring while device is positioned

– Accurate positioning

– Residual shunt

J A C C : C A R D I O V A S C U L A R I M A G I N G V O L . 6 , N O . 1 , 2 0 1 3

Page 24: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Ventricular Septal Defects

• Anatomy for intervention– Location

– Adjacent structures (aortic valve cusps, tricuspid valve)

– Sizing with multiplanar reconstruction

• Intraoperative guidance– Follow catheters

– Monitoring while device is positioned

– Accurate positioning

– Residual shunt

Dan G. Halpern et al. Eur J Echocardiogr 2009;10:569-571

Page 25: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Post MI Ventricular Septal Rupture

Circ Cardiovasc Interv. 2013;6:59-67.

Page 26: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Ventricular Pseudoaneurysms

• Complication of

myocardial infarction or

cardiothoracic surgery

• Contained rupture

• High risk for reoperation

• Feasibility of

transcatheter device

closure under 3D-TEE Circ Cardiovasc Interv. 2011;4:322-326.

Page 27: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Dysfunction of Prosthetic Valves

• Paravalvular Leak

• Degeneration of (bio) Prosthesis

Page 28: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Baseline Mitral Valve and 3D Color

Medial

AO

LAA

Posterior AO

PosteriorMedial

Medially located PVL with severe MR

Page 29: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Medial PVL Sizing using Vena Contracta Area

• Crescent shape• Vena Contracta 8x2mm

Page 30: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Access via PVL space

Diagnostic Multipurpose Catheter 6F, Terumo hydrophylic wire

Page 31: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Amplatz Duct Occluder 12x10mm

7F Delivery Catheter via PVL

Page 32: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Final Result

Page 33: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Dysfunction of Prosthetic Valves

• Paravalvular Leak

• Degeneration of (bio) Prosthesis

Page 34: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Baseline TEE

MG 8mmHg MVA 0.8cm2

Page 35: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Heart Team Assessment• Surgical re-do MVR and Tricuspid valve annuloplasty or replacement,

but patient was considered at high surgical risk / inoperable due to coexistent right heart failure, severe pulmonary hypertension and overall frailty

• Alternative: Implantation of a Sapien XT bioprosthesis within the degenerated Perimount in the mitral position

Page 36: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Thoracotomy Exposure of

Cardiac Apex

Valve in Valve with Sapien XT-29mm

Page 37: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Final Result

Mean Gradient 3mmHg

Page 38: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

3D-TEE

MVA 2.57cm2

Page 39: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Diastole: LVOT area 1.92cm2

End-Systole:LVOT area 0.71cm2

Mitral VIV Complex: LVOT Interaction X-Plane at LVOT level shows interaction of VIV complex with LVOT during cardiac cycle

Page 40: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Recurrent mitral regurgitation after surgical annuloplasty

Page 41: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

The Patient

• 80 year old female, NYHA IV

• CABG (3-vessel), MV repair (ring 32mm), TV repair (34mm ring) in 2015

• Recurrent severe MR (ERO 40mm2, RVOL 64ml)

• LVEF 35%, moderate TR, pulmonary hypertension sPAP 65mmHg

• Comorbidities: Chronic atrial fibrillation, diabetes mellitus II, transient

ischemic attacks

• Log Euroscore 43%, STS mortality 13.4%

• Decision for valve in ring via the transapical approach

Page 42: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Echocardiography:

Severe Mitral Regurgitation

Page 43: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Multidetector Computed Tomography

Note: Non-planar annuloplasty ring with highest points in the anteroposteriordirection and lowest points in the commissures

Page 44: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Valve in RingBalloon Expandable 29mm

Page 45: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Significant MR: What is the mechanism?

• Significant transvalvular regurgitation due to uncovered cell below the annuloplastyring near the lateral commissure

• Annuloplasty ring sits lower near the commissures and higher in the anteroposteriordirection (physio ring) Possibly contributes to MR from the lateral commissure

Page 46: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Post-dilatation 29mm balloon

Unsuccessful in decreasing MR grade

Page 47: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Valve in Valve (in Ring)29mm balloon expandable valve

Resolution of mitral regurgitation

Page 48: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Mitral Valve Dysfunction

• Mitral Regurgitation

• Mitral Stenosis

Page 49: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Functional Mitral RegurgitationLive 3D with Color Flow Doppler

Page 50: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Simultaneous X-Plane with Color

Page 51: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Degenerative MR : P2 Prolapse with ruptured chords

Page 52: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Simultaneous (X-Plane) Views Commissural and Long Axis with Color

Page 53: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Transeptal Puncture

• Critical first step to entering the left atrium for left sided interventions

• Position of puncture depends on procedure

• e.g. for MitraClip, puncture at the high and posterior aspect of fossa ovalis

Page 54: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Transeptal Puncture under 2D TEE

Page 55: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Clip Delivery System in LA

Page 56: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Positioning for Leaflet Capture

Page 57: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Grasping of Leaflets

Page 58: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Final Double Orifice Mitral Valve

Page 59: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar
Page 60: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar
Page 61: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar
Page 62: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Case 1: PASCAL in patient with FMR

• 75 year old male, NYHA IV

• Coronary bypass surgery (1993, LIMA LAD, SVG Diagonal and OM)

• Echo: LV-EF 30%, severe FMR (ERO 1cm2) sPAP 70mmHg

• Stent Graft Abdominal Aorta (2011)

• Carotid artery disease, atrial fibrillation, pacemaker, mild kidney

disease

Page 63: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Baseline 3D Color: Severe MR

Page 64: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Baseline MR

Page 65: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Baseline MVQ analysis and MV area

Page 66: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Baseline MVQ analysis and MV area

Page 67: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar
Page 68: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

PASCAL: Transition to Final Configuration

Page 69: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar
Page 70: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Final 3D Zoom: Double Orifice Mitral Valve

Page 71: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Final Result

3D Narrow Sector Full Volume Fluoroscopy

Page 72: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Final MV Orifice Evaluation

Medial 1.07cm2 Lateral 1.09cm2

Page 73: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Final 3D Color: Mild MR

Extubation in OR, fluoro time 22min, discharge home on day #4, clinically much improved at the 3 month follow up

Page 74: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Example of other evolving Transcatheter Mitral Valve Repair Technologies

Page 75: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Transcatheter Direct Annuloplasty: Cardioband

Page 76: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar
Page 77: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Transcatheter Chordal Implants

• Synthetic chords attached to LV myocardium and the leaflets for degenerative MR

J Am Coll Cardiol Intv 2011;4:1–13)

Page 78: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Neochord Implantation

Page 79: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Transcatheter Mitral Valve Implantation: Intrepid TWELVEFor Degenerative Mitral Regurgitation

Page 80: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

• Conformable Outer Stent engages the annulus and leaflets providing

fixation & sealing while isolating the inner stent from the dynamic

anatomy

• Circular Inner Stent houses a 27 mm tricuspid bovine pericardium valve

• Conformable Brim aids imaging during delivery & subsequent platform

for healing

80

MEDTRONIC INTREPIDTM TMVIDUAL STENT DESIGN

CAUTION: INVESTIGATIONAL DEVICE. LIMITED BY FEDERAL LAW (USA) TO INVESTIGATIONAL USE. These tests may

not be indicative of clinical performance. These statements have not been evaluated by the FDA and are not

intended to represent claims of human clinical performance or serve as a substitute for medical judgment

Page 81: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

• Variable stiffness along the height of the Outer Stent helps wedge the

implant – similar to a champagne-cork

• Outer stent engaging with, and conforming dynamically to, the annulus

• Circular inner stent isolated from the fixation and sealing

81

MEDTRONIC INTREPIDTM TMVIFIXATION AND SEALING

CT images from a human implantation Echo image from a human implantation – atrial view

CAUTION: INVESTIGATIONAL DEVICE. LIMITED BY FEDERAL LAW (USA) TO INVESTIGATIONAL USE. These tests may

not be indicative of clinical performance. These statements have not been evaluated by the FDA and are not

intended to represent claims of human clinical performance or serve as a substitute for medical judgment

Page 82: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

• Radial interference, small cleats, frictional elements & tissue ingrowth

• Excellent healing response in longer term porcine and ovine studies

• Gross and histological evaluation of pre-clinical implants reveals the

implants are well integrated with the native tissue, with neo-endocardial

coverage

82

MEDTRONIC INTREPIDTM TMVIFIXATION AND SEALING

Chronic Porcine Pre-clinical Implants at 90 Days

Atrial View

Ventricular View

CAUTION: INVESTIGATIONAL DEVICE. LIMITED BY FEDERAL LAW (USA) TO INVESTIGATIONAL USE. These tests may

not be indicative of clinical performance. These statements have not been evaluated by the FDA and are not

intended to represent claims of human clinical performance or serve as a substitute for medical judgment

Page 83: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

12-Month Follow-up: Stable Position & In-growth

MEDTRONIC INTREPIDTM TMVIRESULTS

83CAUTION: INVESTIGATIONAL DEVICE. LIMITED BY FEDERAL LAW (USA) TO INVESTIGATIONAL USE. These tests may

not be indicative of clinical performance. These statements have not been evaluated by the FDA and are not

intended to represent claims of human clinical performance or serve as a substitute for medical judgment

Page 84: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

MEDTRONIC INTREPIDTM TMVIRESULTS

84

12-Month Follow-up: Stable Position & In-growth

CAUTION: INVESTIGATIONAL DEVICE. LIMITED BY FEDERAL LAW (USA) TO INVESTIGATIONAL USE. These tests may

not be indicative of clinical performance. These statements have not been evaluated by the FDA and are not

intended to represent claims of human clinical performance or serve as a substitute for medical judgment

Page 85: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

The Patient

• 78 year old female, NYHA IV

• Multiple hospitalizations for CHF decompensation

• Severe mitral valve regurgitation (4+) : Degenerative etiology

• Coronaries without stenosis

• Chronic lung disease, FEV1 1.03 l/min, bronchodilator use

• Prior cardiac surgery for removal of a left atrial myxoma (2011)

• Chronic atrial fibrillation, on anticoagulation

• Pulmonary hypertension • sPAP 60mmHg, dPAP 30mmHg, mean 42mmHg

• LE 42%, STS mortality 8.9%, STS morb/mort 46.3%

Page 86: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Baseline Color in X-plane View

Page 87: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Delivery catheter advancement

Page 88: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Valve Positioning

Page 89: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Opening of the device in the left atrium, viewed in 3D-zoom mode

Page 90: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Device positioning under rapid ventricular pacing

Page 91: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar
Page 92: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Final Result

Page 93: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar
Page 94: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Final 3D color without MR

Page 95: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Mitral Valve Dysfunction

• Mitral Regurgitation

• Mitral Stenosis

Page 96: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Mitral Stenosis

Diagnosis

• Multiplanar reconstruction for stenosis assessment

• Exclude left atrial thrombus prior to intervention

• Wilkins score for valvuloplasty

Page 97: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Mitral Stenosis

Diagnosis

• Multiplanar reconstruction for stenosis assessment

• Exclude left atrial thrombus prior to intervention

• Wilkins score for valvuloplasty

Page 98: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Novel RT-3D Echocardiographic Score for Mitral Balloon Valvuloplasty

• Mild MV involvement <8 points• Moderate MV involvement 8-13• Severe MV involvement ≥ 14

J Am Soc Echocardiogr 2010;23:13–22

Page 99: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Mitral Stenosis: Percutaneous Balloon Valvuloplasty

Intervention

• Follow catheters / balloon valvuloplasty / assess results / complications

Page 100: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Mitral Stenosis: Percutaneous Balloon Valvuloplasty

• Complicated mitral valvuloplasty with tear of the anterior leaflet and severe MR

J A C C : C A R D I O V A S C U L A R IMA G I N G V O L . 6 , N O . 1 1 , 2 0 1 3

Page 101: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Severe Mitral Valve Calcification

• 80 year old male, NYHA IV

• Comorbidities: Diabetes II, coronary artery disease PCI LCX (at age 64), pacemaker, lung Ca operated, prostate Ca, stroke, COPD, porcelain aorta

• Transcatheter aortic valve replacement (at age 77) for severe aortic stenosis Sapien XT 23mm

• Multiple hospitalizations for recurrent acute pulmonary edema

• Coronary artery disease (PCI LAD in 12/2015 and 9/2016)

• Moderately severe mitral stenosis and regurgitation, degenerative etiology

Page 102: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Severely degenerated mitral valve in 3D-Zoom

Page 103: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Mixed Mitral Valve Disease

Page 104: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Mixed Mitral Valve Disease

Page 105: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar
Page 106: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Mitral regurgitation and stenosis

Page 107: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Mitral Annulus Measurements

Page 108: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

3D-Printing: Implantation of Sapien XT in model

MV

TV

Page 109: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Transapical Sapien 3-29mm Implantation in calcified mitral annulus

Page 110: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Final Result

Page 111: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Aortic Valve Pathology

• Stenosis

• Transcatheter aortic valve replacement

Page 112: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

3D-Planimetry of the Aortic Valve

• Multiplanar reconstruction in mid systole

• Evaluate AV orifice at a plane perpendicular to the tip of the three cusps

AVA 1.27cm2

Page 113: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

3D-Planimetry: Bicuspid Aortic Valve

AVA 1.7cm2

Page 114: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

3D-Evaluation of Aortic Annulus

Multiplanar reconstruction at a level corresponding to the leaflet insertion points Virtual annulus of aortic valve

Page 115: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Transcatheter Aortic Valve Replacement

• Echocardiographic Image Guidance not routinely used – fluoroscopy and angiography suffice in majority of patients, except for

– Valve in valve for stentless bioprosthetic valves

– Native valve stenosis without significant calcification

• Echo for pre- and post procedure evaluation mandatory for assessment of final result and early detection of complications

Page 116: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

J A C C : CARDI O V A SC U LAR IMAGI NG VOL. 8, NO. 3, 2015

Page 117: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Echocardiography During TAVR

Page 118: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Baseline X-Plane Imaging

Page 119: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Valve in Valve: Annulus SizingComparison of 3D-TOE and MDCT

3D TEE : Aortic Root MultiplanarReconstruction : Area of Annulus 530mm2

MDCT : Aortic Root MultiplanarReconstruction : Area of Annulus 492mm2

Selection of CoreValve 29mm prosthesis

Page 120: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

TAVR - Valve in Valve: CoreValve 29mm positioning and release

Page 121: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

TAVR - Valve in Valve: Final Result

Page 122: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Echocardiography post TAVR: Evaluation of Degree and Etiology of

Paravalvular Regurgitation

Page 123: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Severe Paravalvular Regurgitation: Incomplete expansion of the prosthesis

Valve not well apposed to the annulus

Calcium deposits not allowing stent frame to fully expand

Page 124: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Severe Paravalvular Regurgitation:Very low position

Page 125: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Tricuspid Valve Regurgitation

Page 126: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

Severe Tricuspid Valve Regurgitation

Page 127: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar
Page 128: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar
Page 129: Role of imaging in structural interventions · pulmonic stenosis) Atrial Septal Defects •Type of defect –Secundum, primum, venosum ... tricuspid valve) –Sizing with multiplanar

In Summary

• Echocardiography is the “first stop” imaging modality for accurate evaluation of structural heart disease

• Even more, it is an indispensable tool during transcatheter interventions for guidance and evaluation of the results

• Future developments will aim at fusion imaging to integrate echocardiography with other available imaging modalities