Upload
razvan-theodor-anca
View
223
Download
0
Embed Size (px)
Citation preview
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 1/183
ROBOTIC ASSISTED SUTURINGIN MINIMALLY INVASIVE SURGERY
By
Hyosig Kang
A Thesis Submitted to the Graduate
Faculty of Rensselaer Polytechnic Institute
in Partial Fulfillment of the
Requirements for the Degree of
DOCTOR OF PHILOSOPHY
Major Subject: Mechanical Engineering
Approved by theExamining Committee:
Dr. John T. Wen, Thesis Adviser
Dr. Stephen J. Derby, Co-Thesis Adviser
Dr. Daniel Walczyk, Member
Dr. Harry E. Stephanou, Member
Rensselaer Polytechnic InstituteTroy, New York
May 2002(For Graduation August 2002)
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 2/183
ROBOTIC ASSISTED SUTURINGIN MINIMALLY INVASIVE SURGERY
By
Hyosig Kang
An Abstract of a Thesis Submitted to the Graduate
Faculty of Rensselaer Polytechnic Institute
in Partial Fulfillment of the
Requirements for the Degree of
DOCTOR OF PHILOSOPHYMajor Subject: Mechanical Engineering
The original of the complete thesis is on filein the Rensselaer Polytechnic Institute Library
Examining Committee:
Dr. John T. Wen, Thesis Adviser
Dr. Stephen J. Derby, Co-Thesis Adviser
Dr. Daniel Walczyk, Member
Dr. Harry E. Stephanou, Member
Rensselaer Polytechnic InstituteTroy, New York
May 2002(For Graduation August 2002)
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 3/183
c Copyright 2002
by
Hyosig Kang
All Rights Reserved
ii
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 4/183
CONTENTS
LIST OF TABLES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
LIST OF FIGURES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . viii
ACKNOWLEDGMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xiv
ABSTRACT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xvi
1. INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1 Background and Motivation . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Problem Statement . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.3 Contributions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31.4 Outline . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2. LITERATURE REVIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.1 Robotic Devices in Surgical Applications . . . . . . . . . . . . . . . . 5
2.2 Augmentation of Human Capability in Surgical Tasks . . . . . . . . . 7
2.3 Motion and Force Control of Rigid Robot Systems . . . . . . . . . . . 8
2.3.1 Robot Motion Control . . . . . . . . . . . . . . . . . . . . . . 8
2.3.2 Robot Force Control . . . . . . . . . . . . . . . . . . . . . . . 9
3. MODELING AND IDENTIFICATION . . . . . . . . . . . . . . . . . . . . 11
3.1 EndoBot Design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.1.1 Manipulator Design Requirements . . . . . . . . . . . . . . . . 11
3.1.2 Current Prototype . . . . . . . . . . . . . . . . . . . . . . . . 12
3.2 Kinematic Modeling of the EndoBot . . . . . . . . . . . . . . . . . . 15
3.2.1 Forward Kinematics . . . . . . . . . . . . . . . . . . . . . . . 15
3.2.2 Inverse Kinematics . . . . . . . . . . . . . . . . . . . . . . . . 17
3.2.3 Jacobian Formulation . . . . . . . . . . . . . . . . . . . . . . . 18
3.2.3.1 Geometric Jacobian . . . . . . . . . . . . . . . . . . 18
3.2.3.2 Analytical Jacobian . . . . . . . . . . . . . . . . . . 19
3.2.4 Singularity Analysis . . . . . . . . . . . . . . . . . . . . . . . 19
3.3 Dynamics of the EndoBot . . . . . . . . . . . . . . . . . . . . . . . . 20
3.3.1 Lagrangian Formulation . . . . . . . . . . . . . . . . . . . . . 20
3.3.2 Properties of Dynamic Model . . . . . . . . . . . . . . . . . . 21
iii
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 5/183
3.3.3 Dynamic Model of the EndoBot . . . . . . . . . . . . . . . . . 22
3.4 Friction Modeling and Compensation . . . . . . . . . . . . . . . . . . 24
3.5 Identification of Dynamic Parameters . . . . . . . . . . . . . . . . . . 26
3.5.1 Input Signal Design . . . . . . . . . . . . . . . . . . . . . . . . 283.5.2 Dynamic Models for Parameter Identification . . . . . . . . . 34
3.5.2.1 Use of Differential Model . . . . . . . . . . . . . . . 35
3.5.2.2 Use of Energy Model . . . . . . . . . . . . . . . . . . 36
3.5.3 Experimental Identification of Dynamic Parameters . . . . . . 38
3.5.3.1 Identification of the Friction Coefficients in EnergyModel . . . . . . . . . . . . . . . . . . . . . . . . . . 39
3.5.3.2 Experimental Results . . . . . . . . . . . . . . . . . 41
3.5.4 Validation of Parameter Identification . . . . . . . . . . . . . . 43
4. SUTURING IN MINIMALLY INVASIVE SURGERY . . . . . . . . . . . . 45
4.1 Suturing Task Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . 45
4.2 Suture Length Tracking . . . . . . . . . . . . . . . . . . . . . . . . . 48
4.3 Motion Control in the Stitching Task . . . . . . . . . . . . . . . . . . 49
4.3.1 Dynamic modeling of Stitching . . . . . . . . . . . . . . . . . 49
4.3.1.1 Kinematics of the Suture Motion . . . . . . . . . . . 50
4.3.1.2 Static Model of the Suture Tension . . . . . . . . . . 50
4.3.2 Problem Formulation . . . . . . . . . . . . . . . . . . . . . . . 51
4.3.3 Region of the Feasible Motion . . . . . . . . . . . . . . . . . . 51
4.3.4 Problem Simplification . . . . . . . . . . . . . . . . . . . . . . 52
4.3.5 Problem Transformation . . . . . . . . . . . . . . . . . . . . . 52
4.4 Knot Tying . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
4.4.1 Ligation Algorithm 1 . . . . . . . . . . . . . . . . . . . . . . . 54
4.4.2 Ligation Algorithm 2 . . . . . . . . . . . . . . . . . . . . . . . 57
4.4.3 Ligation Algorithm 3 . . . . . . . . . . . . . . . . . . . . . . . 61
4.5 Placing the Knot . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
4.5.1 Problem Formulation . . . . . . . . . . . . . . . . . . . . . . . 624.5.2 Sliding Condition of Knot Placement . . . . . . . . . . . . . . 63
4.5.3 Trajectories of Suture Ends for Placing a Knot . . . . . . . . . 68
4.6 Controller Requirement and Architecture . . . . . . . . . . . . . . . . 70
4.6.1 Hybrid Dynamic System . . . . . . . . . . . . . . . . . . . . . 70
4.6.1.1 Hybrid State . . . . . . . . . . . . . . . . . . . . . . 70
iv
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 6/183
4.6.1.2 Hybrid Automaton . . . . . . . . . . . . . . . . . . . 71
4.6.2 Human Sharing Supervisory Controller . . . . . . . . . . . . . 72
4.6.3 Planning of Suturing Task . . . . . . . . . . . . . . . . . . . . 73
4.6.4 Development Environment . . . . . . . . . . . . . . . . . . . . 75
5. MOTION CONTROL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
5.1 Gravity Compensation in Manual Mode . . . . . . . . . . . . . . . . . 79
5.2 Motion Control in Autonomous Mode . . . . . . . . . . . . . . . . . . 79
5.2.1 Energy Based Output Feedback Control . . . . . . . . . . . . 80
5.2.2 Experimental Evaluation of Joint Space Control . . . . . . . . 82
5.2.3 Nonlinear Decoupled State Feedback Controller . . . . . . . . 86
5.2.3.1 Global Linearization of the Nonlinear Robotic System 86
5.2.3.2 Optimal State Feedback Control . . . . . . . . . . . 87
5.2.3.3 Velocity Estimation . . . . . . . . . . . . . . . . . . 91
5.2.3.4 Linear Quadratic Gaussian (LQG) Control . . . . . . 94
6. SHARED CONTROL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
6.1 Constraints Description . . . . . . . . . . . . . . . . . . . . . . . . . . 100
6.2 Control Objective . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
6.3 Task Space Control Design . . . . . . . . . . . . . . . . . . . . . . . . 102
6.3.1 Stability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
6.3.2 Controller Description . . . . . . . . . . . . . . . . . . . . . . 1046.3.3 Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
6.3.4 Experimental Evaluation of Task Space Shared Controller . . 105
6.4 Joint Space Shared Control Design . . . . . . . . . . . . . . . . . . . 109
6.4.1 Controller Description . . . . . . . . . . . . . . . . . . . . . . 112
6.4.2 Examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
6.4.3 Experimental Evaluation of Joint Space Shared Controller . . 114
7. TENSION CONTROL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
7.1 Securing a Knot . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1217.1.1 Principle of Direction of Securing a Square Knot . . . . . . . . 121
7.2 Tension Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
7.2.1 Basic Idea . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
7.2.2 Tension Estimation with a Base Force/Torque Sensor . . . . . 125
7.3 Force and Position Control . . . . . . . . . . . . . . . . . . . . . . . . 127
v
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 7/183
7.3.1 Problem Formulation . . . . . . . . . . . . . . . . . . . . . . . 127
7.3.2 Hybrid Force/Position Regulating Control . . . . . . . . . . . 128
7.3.3 Explicit Force Control . . . . . . . . . . . . . . . . . . . . . . 131
7.4 Stability Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132
7.4.1 Proportional plus Integral Control with Active Damping . . . 132
7.4.2 Stability of Time Delayed System . . . . . . . . . . . . . . . . 136
7.5 Experimental Study . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
7.5.1 Experimentation Environment . . . . . . . . . . . . . . . . . . 139
7.5.2 Experimental Evaluation of Tension Controller . . . . . . . . . 141
8. CONCLUSIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147
8.1 Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147
8.2 Future Work . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 148
LITERATURE CITED . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150
APPENDICES
A. INVERSE KINEMATICS USING SUBPROBLEMS . . . . . . . . . . . . . 158
B. LAGRANGIAN FOR A ROBOT MANIPULATOR . . . . . . . . . . . . . 159
B.1 Kinetic Energy of a Robot Manipulator . . . . . . . . . . . . . . . . . 159
B.2 Potential Energy of a Robot Manipulator . . . . . . . . . . . . . . . . 160
C. LEAST SQUARE METHOD . . . . . . . . . . . . . . . . . . . . . . . . . 161
D. PERSISTENT EXCITATION AND OBSERVABILITY GRAMIAN . . . . 163
E. DIFFEOMORPHISM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 165
vi
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 8/183
LIST OF TABLES
3.1 Parameters of the electrical system. . . . . . . . . . . . . . . . . . . . . 39
3.2 Coefficients of cross correlation. . . . . . . . . . . . . . . . . . . . . . . 40
3.3 Friction parameters of each link. . . . . . . . . . . . . . . . . . . . . . . 41
3.4 Estimates of the dynamic parameters of the EndoBots. . . . . . . . . . 42
5.1 Summary of friction compensation performance. . . . . . . . . . . . . . 84
5.2 Circle tracking error. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
6.1 Parameters for constrained line. . . . . . . . . . . . . . . . . . . . . . . 106
6.2 Desired and measured stiffness for shared control. . . . . . . . . . . . . 109
6.3 Effective constraint stiffness for linear trajectory . . . . . . . . . . . . . 117
6.4 Effective constraint stiffness for circular trajectory . . . . . . . . . . . . 118
7.1 Suture diameter-strength relationship. . . . . . . . . . . . . . . . . . . . 140
7.2 Suture pullout values. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
7.3 Comparison of force controller with active damping. . . . . . . . . . . . 142
7.4 Effect of the proportional gain. . . . . . . . . . . . . . . . . . . . . . . . 142
7.5 Performance of force controller with a modified force reference trajectory.143
vii
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 9/183
LIST OF FIGURES
3.1 Fulcrum effect in laparoscopic surgery. . . . . . . . . . . . . . . . . . . . 11
3.2 Mechanical overview of the EndoBot. . . . . . . . . . . . . . . . . . . . 13
3.3 Closeup on semicircular arches. . . . . . . . . . . . . . . . . . . . . . . . 13
3.4 Translational stage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.5 Picture of a pair of the EndoBots. . . . . . . . . . . . . . . . . . . . . . 14
3.6 Grasper and suturing tools. . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.7 Kinematic diagram of the EndoBot. . . . . . . . . . . . . . . . . . . . . 15
3.8 Simplified dynamic model of the EndoBot. . . . . . . . . . . . . . . . . 22
3.9 Model-based friction compensation. . . . . . . . . . . . . . . . . . . . . 24
3.10 Classical friction models. . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.11 Parametric system identification. . . . . . . . . . . . . . . . . . . . . . . 26
3.12 Generation of PRBS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
3.13 Schroeder-phased signal design. . . . . . . . . . . . . . . . . . . . . . . 32
3.14 Comparison of input signals in time domain. . . . . . . . . . . . . . . . 33
3.15 Comparison of power spectra of input signals. . . . . . . . . . . . . . . 34
3.16 Comparison of two identification methods. . . . . . . . . . . . . . . . . 35
3.17 Block diagram of velocity control loop. . . . . . . . . . . . . . . . . . . 40
3.18 Trajectories of the first joint. . . . . . . . . . . . . . . . . . . . . . . . . 41
3.19 Estimates of the dynamic parameters of the first joint. . . . . . . . . . . 42
3.20 Comparison on measured and simulated output trajectories . . . . . . . 43
4.1 Knot tying techniques. . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
4.2 Grasper with a needle in conventional suturing. . . . . . . . . . . . . . . 47
4.3 Shuttle needle device. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
4.4 Components of a suture. . . . . . . . . . . . . . . . . . . . . . . . . . . 49
viii
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 10/183
4.5 Autonomous simple knot tying algorithm 1. . . . . . . . . . . . . . . . . 56
4.6 Flexible hook for catching the suture. . . . . . . . . . . . . . . . . . . . 57
4.7 Autonomous simple knot tying algorithm 2. . . . . . . . . . . . . . . . . 59
4.8 Implementation of autonomous simple knot tying algorithm 2. . . . . . 60
4.9 Autonomous simple knot tying algorithm 3. . . . . . . . . . . . . . . . . 61
4.10 Forming a simple knot. . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
4.11 Two point contact model of the knotted suture. . . . . . . . . . . . . . 64
4.12 Free-body diagram for the knot sliding problem. . . . . . . . . . . . . . 64
4.13 Free-body diagram of one strand. . . . . . . . . . . . . . . . . . . . . . 65
4.14 Free-body diagram of one strand in symmetric tension. . . . . . . . . . 66
4.15 Geometric interpretation of the sliding condition. . . . . . . . . . . . . . 67
4.16 Sliding condition in a knot tying plane. . . . . . . . . . . . . . . . . . . 68
4.17 Evolving trajectory of the knot. . . . . . . . . . . . . . . . . . . . . . . 68
4.18 Trajectory of the loop end for placing the knot. . . . . . . . . . . . . . . 69
4.19 Trajectories of loop end. . . . . . . . . . . . . . . . . . . . . . . . . . . 70
4.20 State invariant space. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 724.21 Hybrid state transition diagram. . . . . . . . . . . . . . . . . . . . . . . 72
4.22 Human sharing supervisory controller. . . . . . . . . . . . . . . . . . . . 73
4.23 High level state transition diagram. . . . . . . . . . . . . . . . . . . . . 74
4.24 Manual state diagram. . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
4.25 Autonomous state evolution. . . . . . . . . . . . . . . . . . . . . . . . . 75
4.26 State transition diagram for stitching task. . . . . . . . . . . . . . . . . 76
4.27 State transition diagram for grasping suture tail. . . . . . . . . . . . . . 76
4.28 State transition diagram for creating the knot. . . . . . . . . . . . . . . 76
4.29 State transition diagram for securing the knot. . . . . . . . . . . . . . . 77
4.30 State evolution for creating the knot. . . . . . . . . . . . . . . . . . . . 77
ix
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 11/183
4.31 Calibration of two coordinates. . . . . . . . . . . . . . . . . . . . . . . . 77
4.32 Overview of the experimental environment. . . . . . . . . . . . . . . . . 78
5.1 Friciton approximation for compensation. . . . . . . . . . . . . . . . . . 83
5.2 Friction compensation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
5.3 Cartesian space experimental result. . . . . . . . . . . . . . . . . . . . . 84
5.4 Experimental results of circle tracking. . . . . . . . . . . . . . . . . . . 85
5.5 Bode plot for the numerical differentiating filter. . . . . . . . . . . . . . 92
5.6 Bode plot for the washout filter. . . . . . . . . . . . . . . . . . . . . . . 93
5.7 Observer based linear controller with feedback linearization. . . . . . . . 96
5.8 Experimental results of friction compensation in the LQG controller. . . 97
5.9 Tracking performance of the LQG controller. . . . . . . . . . . . . . . . 98
5.10 Comparison on error states with the different weighting. . . . . . . . . . 98
5.11 Comparison on control inputs with the different weighting. . . . . . . . 99
6.1 Block diagram of task space shared control. . . . . . . . . . . . . . . . . 104
6.2 Desired constrained path. . . . . . . . . . . . . . . . . . . . . . . . . . . 106
6.3 Measured task space positions. . . . . . . . . . . . . . . . . . . . . . . . 1076.4 Constraint forces. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
6.5 Constraint forces projected onto range space R(J T c ). . . . . . . . . . . . 108
6.6 Measured joint torques. . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
6.7 Experimental result of task space shared control with constrained line. . 110
6.8 Concept of coordinates mapping. . . . . . . . . . . . . . . . . . . . . . . 111
6.9 Block diagram of joint space shared controller. . . . . . . . . . . . . . . 112
6.10 Measured task space positions. . . . . . . . . . . . . . . . . . . . . . . . 115
6.11 Desired and measured joint position . . . . . . . . . . . . . . . . . . . . 116
6.12 Measured joint torque . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
6.13 Actual path with joint space shared controller . . . . . . . . . . . . . . 117
x
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 12/183
6.14 Desired circle trajectory . . . . . . . . . . . . . . . . . . . . . . . . . . . 118
6.15 Measured task space position . . . . . . . . . . . . . . . . . . . . . . . . 119
6.16 Desired and measured joint position . . . . . . . . . . . . . . . . . . . . 119
6.17 Measured joint torque . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
6.18 Actual trajectory of the end-effector in the task space for the circularconstraint . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
7.1 Suturing line. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
7.2 Direction of securing in the first throw. . . . . . . . . . . . . . . . . . . 122
7.3 Securing the knot. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
7.4 An unstable two half hitch knot with reverse-directed tension during
the second throw. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
7.5 Strain gauge transducer. . . . . . . . . . . . . . . . . . . . . . . . . . . 124
7.6 Vision sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
7.7 Base force/torque sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . 125
7.8 Transformation of wrenches. . . . . . . . . . . . . . . . . . . . . . . . . 126
7.9 Constrained motion. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129
7.10 Constrained motion for securing the knot. . . . . . . . . . . . . . . . . . 129
7.11 Simplified second order environment model. . . . . . . . . . . . . . . . . 132
7.12 Explicit PI force control. . . . . . . . . . . . . . . . . . . . . . . . . . . 133
7.13 Root locus under explicit PI force control with K p = 1. . . . . . . . . . 134
7.14 Root locus under explicit PI force control with K p = 10. . . . . . . . . . 135
7.15 Explicit PI force control with active damping. . . . . . . . . . . . . . . 135
7.16 Root locus under explicit PI force control with the active damping Kv =
1000 and K p = 1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 136
7.17 Explicit PI force control with time delay. . . . . . . . . . . . . . . . . . 136
7.18 Root locus of the time delayed system under explicit PI force controlwith K p = 10. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
7.19 Root locus of the time delayed system under explicit PI force controlwith active damping K v = 1000 and K p = 10. . . . . . . . . . . . . . . . 139
xi
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 13/183
7.20 Experimental testbed for tension control. . . . . . . . . . . . . . . . . . 140
7.21 Experimental data from integral control with K i = 0.5. . . . . . . . . . 143
7.22 Experimental data from integral control plus active damping with K i =
0.5 and K v = 1000. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144
7.23 Experimental data from integral control plus active damping with K i =0.5 and K v = 2000. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144
7.24 Experimental data from PI control plus active damping with K p = 0.3,K i = 0.5, and K v = 2000. . . . . . . . . . . . . . . . . . . . . . . . . . . 144
7.25 Experimental data from PI control plus active damping with K p = 0.5,K i = 0.5, and K v = 2000. . . . . . . . . . . . . . . . . . . . . . . . . . . 145
7.26 Experimental data from PI control plus active damping with K p = 0.2,
K i = 0.75, and K v = 2000. . . . . . . . . . . . . . . . . . . . . . . . . . 145
7.27 Experimental data from PI control plus active damping with K p = 0.2,K i = 0.75, and K v = 1000. . . . . . . . . . . . . . . . . . . . . . . . . . 145
7.28 Experimental data from PI control plus active damping with K p = 0.2,K i = 0.75, and K v = 500. . . . . . . . . . . . . . . . . . . . . . . . . . . 146
7.29 Experimental data from PI control plus active damping with K p = 0.2,K i = 0.3, K v = 2000, and the modified force reference trajectory withF des = −(f + 5)N. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146
B.1 Kinetic energy of a rigid body. . . . . . . . . . . . . . . . . . . . . . . . 159
C.1 Orthogonal decomposition. . . . . . . . . . . . . . . . . . . . . . . . . . 161
C.2 Interpretation on the least square solution. . . . . . . . . . . . . . . . . 162
E.1 One-to-one and not onto . . . . . . . . . . . . . . . . . . . . . . . . . . 165
E.2 Onto not one-to-one . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 165
E.3 One-to-one and onto . . . . . . . . . . . . . . . . . . . . . . . . . . . . 166
xii
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 14/183
To my wife Kyunga
and my daughters Dahyun and Hannah
with all my love...
xiii
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 15/183
ACKNOWLEDGMENT
With the completion of this thesis I am again at a crossroad in my life. The time Ihave spent here has been very fruitful and many people came into my life to inspire
and encourage this effort.
I would first like to express my truly gratitude to my advisor, Dr. John T.
Wen. John is an exceptionally distinguished researcher with enthusiasm, but he
stepped down to my level in order to collaborate with me. He was always there to
answer my questions clearly and helped to make robotics and control fun for me.
Throughout the last five years, he provided not only the promising research subject
and motivation to challenging problems, but also the way of thinking and the way
of proceeding research. In every sense, non of this work would have been possible
without him. He has been a real partner and I will always keep unforgettable
memories of discussions and friendship we have had together.
I would like to acknowledge Dr. Harry E. Stephanou, director of the Center
for Automation Technologies, who provided me the opportunity of my journey at
Rensselaer Polytechnic Institute and wide latitude to explore interesting research
areas. I would also like to thank the other members of my doctoral committee,
Dr. Stephen J. Derby and Daniel Walczyk, both of whom provided many helpful
comments and suggestions.
My colleagues at the Center for Automation Technologies have created a very
pleasant atmosphere to work in. In particular, I want to thank Wooho Lee and
Jeongsik Sin for being my friend and for making me ponder many things about
my life. I firmly believe that we can work together again. I also want to thank
Ben Potsaid, for our fruitful conversations during many trips to Rio de Janeiro and
Alaska. I would like to thank my long time friend, YoungCheol Park, and my seniors
Youngkee Ryu, Hyunggu Park, Byunghee Won, and Hyungsoo Lee. Our occasional
telephone chats and emails have always provided a welcome relief from the lonely
journey.
I thank my parents for their love, support, and understanding through these
xiv
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 16/183
years. I also thank to my daughters, Dahyun and Hannah, who provided the joy of
life and the ability to make peace with chaos. They prayed for me and brought me
great comfort during times of extreme stress.
My final, and most heartfelt, acknowledgement must go to my wife, Kyunga,who supported my decision to embark in graduate studies despite the significant
changes in our life and provided stability to our family. Her support, encouragement,
and companionship has turned my journey through doctoral studies here into a
pleasure. She is my everlasting love.
Delight yourself in the LORD and he will give you the desires of your heart
(Psalm 37:4)
xv
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 17/183
ABSTRACT
Open surgery traditionally involves making a large incision to visualize the operativefield and to access human internal tissues. Minimally invasive surgery (MIS) is an
attractive alternative to the open surgery whereby essentially the same operations
are performed using the specialized instruments designed to fit into the body through
several pencil-sized holes instead of one large incision. It can minimize trauma and
pain, decrease the recovery time thereby reducing the hospital stay and cost, but
also offers more technical difficulties to surgeons.
Despite the lack of dexterity and perception, all surgeries are moving toward
MIS due to the benefits to patients. However, the demand on surgeons is much
higher during suturing, which is the primary tissue approximation method. Sutur-
ing is known as one of the most difficult tasks in MIS and consumes a significant
percentage of the operating time. Despite the important role and technical challenge
of suturing in MIS, there is little research on suturing.
Motivated by these observations, a new surgical robotic system, which we have
named EndoBot, is developed in this research. The focus of the research is the de-
velopment of the EndoBot controller that is capable of a range of MIS operations
including manual, shared, and supervised autonomous operations. Due to the chal-
lenge of the suturing task, the particular emphasis is placed on its automation. The
suturing operation is decomposed as knot forming, knot placement, and tension
control. New algorithms are developed and implemented for each subproblem and
integrated for the completely autonomous knot tying. To ensure safe operation,
a discrete event controller allows interruption by the surgeon at any moment and
continuation with autonomous operation after the interruption.
xvi
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 18/183
CHAPTER 1
INTRODUCTION
1.1 Background and Motivation
The use of robots in surgery is rapidly expanding in recent years. Robots
can improve precision, filter human motion tremor, extend human reach into the
body, and reduce the risk of infection. Surgical robots can be classified based on
the planning strategy: model-based robotic surgery and nonmodel-based robotic
surgery; or based on the level of surgical invasiveness: open surgery and minimally
invasive surgery.
Model-based surgery uses the geometric models obtained from scanned images
using the computer tomography (CT) or magnetic resonance imaging (MRI) during
the pre-operation stage. Typically, this requires a registration process in which pre-
operative model is matched with intra-operative model. If the registration process
can be accurately performed, the robotic surgery becomes as off-line robot motion
planning problem as in a CAD-CAM system. This approach is applicable to surg-
eries of hard tissues such as spine surgery, neurosurgery, hip replacement surgery,
total knee replacement surgery, plastic surgery, and eye surgery. Quality of the
surgery depends to a large extent on accuracy of the model.
Nonmodel-based robotic surgery typically deals soft tissues whose model is
either not available or not useful during operations due to the tendency of tissues
floating in the body or changing in shape. Most surgeries involving a body organ are
nonmodel-based. Surgeons either navigate a hand-held robot directly or manipulate
a input device in the case of teleoperation. The man/machine interface plays a
key role because surgeons are directly responsible for manipulating the surgical
robots. In this surgery, the surgeon’s expertise is the key factor on the quality of
surgery. Force feedback and cooperative control to guide surgeons can also improve
the surgical results in nonmodel-based robotic surgeries.
Surgery traditionally involves making a large incision to expose the operation
sites and to access human internal tissues. This is referred to as the open surgery.
1
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 19/183
2
The incision needed to allow surgeons to visualize the field tends to delay patient
recovery and causes most of the pain.
Minimally invasive surgery (MIS), also known as laparoscopic surgery in the
abdominal cavity, arthroscopic surgey in the joints, and thoracoscopic surgery inthe chest, has became increasingly an attractive alternative to open surgery. In
MIS, the same operations are performed with the specialized instruments which are
designed to fit into the body through the several pencil-sized holes instead of one
large incision. By eliminating the significant incision, MIS has many advantages
over conventional open surgery. It can minimize trauma and pain, and decrease the
recovery time thereby reducing the hospital stay and cost for patients.
However, compared to open surgery, MIS offers greater challenges to surgeons.
Instead of looking directly at the part of the body being treated, surgeons monitor
the procedures via a two-dimensional video monitor without the three-dimensional
depth information. Due to the inherent kinematic constraints at the incision points,
the motions of MIS instruments are restricted to four degrees of freedom. Despite of
lack of dexterity and perception, all surgeries are moving toward MIS to give more
benefits to patients at the expense of a more stressful environment to surgeons.
The demands to surgeons for dealing with these technical difficulties are higher
during the suturing task, which is the primary tissue approximation method and
has been known as one of the most difficult tasks in MIS operations and uses a
significant percentage of operating time.
1.2 Problem Statement
Several robotic systems have been developed for minimally invasive surgery
[10, 11, 12]. There are also a few commercial systems available on the market such
as AESOP and ZEUS (Computer Motion), daVinci (Intuitive Surgical), and Neu-
romate (ISSS/immi). It is important to note that in the various surgical robots
described above, the surgical procedures are still completely performed by the hu-
man surgeon High skill level procedures such as suturing and precise tissue dissection
continue to depend on the expertise of surgeons, and surgeon’s commands are mim-
icked by the robotic devices through computer control. Despite the important role
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 20/183
3
and technical challenge of suturing in MIS, there has been little research effort re-
ported an automated robotic suturing. Most results on robotic surgery focus on the
development of the robotic system and leave the suturing task to the surgeon.
This thesis presents the development of a new surgical robotic system, whichwe named the EndoBot, for MIS operations. The EndoBot is designed for the col-
laborative operations between surgeons and the robotic device. Surgeons can select
the device to operate completely in manual mode, collaboratively where motion of
robotic device in certain directions are under computer control and others under
surgeon’s manual control, or autonomously where the complete robot system is un-
der computer control and surgeon’s supervision. Furthermore, the robotic tools can
be quickly changed from a robotic docking station, allowing different robotic tools
to be used in operations.
For automated robotic suturing using the EndoBot, the suturing task needs
to be analyzed and the algorithms for knot tying task need to be developed to deal
with a flexible suture. In addition, knot placement and knot tension control should
also be considered.
Motivated by the above observations, the goal of this research is to develop a
robotic system that can collaboratively perform laparoscopic procedures with sur-
geons, conduct certain tasks autonomously to reduce the strain on surgeons, remove
the variability of surgeon’s training levels, and enhance the system efficiency by
decreasing the operation time.
1.3 Contributions
The major contributions of this research are summarized here.
• Robotic system for minimally invasive surgery. A new surgical robot system,
called the EndoBot, for MIS operation is developed and the dynamic param-eters are identified experimentally.
• Suturing task analysis and implementation. The suturing task is decomposed
into five subtasks and the technical difficulties on each task are analyzed. Knot
placement problem is formulated, and the sliding condition is proposed. The
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 21/183
4
knot tying algorithm is presented and implemented with the EndoBot.
• Supervisory controller. In order to guarantee the safe operation of sutur-
ing task in non-deterministic environments, a supervisory controller is imple-
mented based on the hybrid dynamic system framework.
• Shared control. Shared control using artificial constraints is proposed and
experimentally verified.
• Tension Control. A base sensor method is proposed to measure the tension in
suture. A hybrid force/position control is implemented to effectively regulate
the suture tension while achieving the desired direction.
1.4 Outline
This document is organized in the following manner. In Chapter 2, related
work is reviewed. Chapter 3 provides the kinematics and dynamics analysis of new
surgical robot and discusses the identification method for dynamic parameters. In
Chapter 4, a suturing task analysis is presented with the high level control architec-
ture. Chapter 5 discusses the motion controller design and the experimental results
and the shared control is presented in order to augment the human capability in
Chapter 6. Chapter 7 provides a hybrid force/position control strategy to effectively
regulate the tension with the experimental results. Finally, Chapter 8 summarizes
this document, outlines the future work.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 22/183
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 23/183
6
available on the market as the da Vinci Surgical System (Intuitive Surgical
Inc.). The six degrees of freedom slave manipulator was developed in [12],
and the parallel mechanism was chosen for gross motion to increase rigidity
and the tendon actuated millirobot was added. In [13], the workspace of laparoscopic extender with flexible stems was formulated. They performed
the optimization on design of flexible stems by defining the dexterity measure,
which is the ratio of areas of dexterous workspace and reachable workspace.
The redundant wrist with parallel structure, which has three degrees of free-
dom, was presented in [15], and they carried out an optimal design of the
mechanism.
Laparoscopic suturing Suturing is one of the most difficult tasks and takes asignificant percentage of operating time and involves in complex motion plan-
ning. The general description on manual laparoscopic suturing can be found
in [16, 17]. As a direct result of constraints in laparoscopic surgery, there is an
extended learning curve that surgeons must go through to gain the required
skill and dexterity. Furthermore, there is a great deal of variability even among
trained surgeons. As demonstrated in [21], time-motion studies of laparoscopic
surgery have indicated that for the operations such as suturing, knot tying,
suture cutting, and tissue dissection, the operation time variation between
surgeons can be as large as 50%. In suturing, it was noted that the major
difference between surgeons lies in the proficiency at grasping the needle and
moving it to a desired position and orientation, without slipping or dropping
it. Cao et al. The motion analysis on the suturing task using conventional
needle holders was performed in [21], and they broke down into five basic
motions such as reach and orientation, grasp and hold, push, pull, and release.
The teleoperator slave system with a dexterous wrist for minimally invasive
surgery was developed in [20], and they demonstrated the suturing task with
direct vision. More recently, the knot tying simulation for surgical simulation
with a spline of linear spring model was proposed in [85]. The future trends
in laparoscopic suturing can be found in [19]. However, these all publications
did not address the issues on the autonomous robotic suturing.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 24/183
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 25/183
8
During the operation, the human supervises each evolution of task with intervening
capability. This framework can be the basis of automating surgical tasks.
2.3 Motion and Force Control of Rigid Robot Systems2.3.1 Robot Motion Control
Robot manipulators have complex nonlinear dynamics and there are many con-
trol schemes in motion control. The selection of the motion controller may depend
on the required task, sensor availability, computational power, and the knowledge
of the values of the dynamic parameters. The basis in motion control of robot
manipulators is the PD (Proportional and Derivative) plus gravity compensation
scheme at the joint level. In [66], it was proved that the PD control with computed
feedforward terms was locally exponentially stable by using an energy-motivated
Lyapunov fuction. In [57], the stability of PD control with gravity compensation
by shaping the potential energy of closed loop system and injecting the required
damping was proved. This controller provides the globally asymptotic stability for
the regulation problem [58] and it is the most widely used in industrial robots. This
controller may be practically used in tracking control applications at the expense of
degrading the tracking performance. For the general trajectory tracking problem,
computed-torque control has been shown to have the globally asymptotic stability[59, 60, 47, 62]. The basic idea is to transform the nonlinear dynamic system into
a linear one by cancelling the nonlinearities of the robot dynamics with a nonlinear
inner feedback linearization loop [63]. The beauty of this approach is that we can
apply many linear control schemes in designing the outer feedback loop. Linear PD
or PID control is the common choice and linear quadratic optimal control is also
used for designing the outer feedback loop [64, 65]. For robot systems, which do not
admit exact feedback linearization, the outer loop controller must be robust in order
to handle the uncertainties. The robust control problem translated into the optimal
control problem in [67]. They designed an optimal LQ (Linear Quadratic) regulator
for the linearized system with uncertainties reflected in the performance index. A
survey on robust control of robots is given in [68, 69]. In many robot systems, only
a part of states can be measured, but the state feedback controller needs the knowl-
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 26/183
9
edge of positions and velocities. The immediate solution to get the velocity signals
is to numerically differentiate the position signals. Inherently numerical differenti-
ating is very sensitive to the noises because of the improper function characteristics.
When the system dynamics is given, the most effective method to estimate all statescan be to design a dynamic observer. The computed-torque controller with linear
observer was proposed in [70], and they proved the exponential stability for the
robot system, which has only position measurement. The performance comparison
on linear and nonlinear observers was given in [71] with the linear state feedback
algorithm for a two-link manipulator.
2.3.2 Robot Force Control
When robot manipulators interact with the environments, force control is es-
sential for successful execution of tasks. Force control has been a research subject
for many years, and various control schemes have been proposed. Basically, the
existing force control strategies can be classified into two categories. The first group
aims at performing indirect force control by controlling the relationship between the
manipulator position and the interaction force. Compliance control and impedance
control can be grouped into the indirect force control scheme. In [72], Hogan pro-
posed impedance control to achieve a desired dynamic behavior while compliance
control can achieve the static behavior in [73, 74]. These schemes are suitable for
tasks where accurate force regulation is not required and the force measurement is
not available. In order to regulate the exact force, the parallel position/force control
scheme was introduced in [79]. Parallel position/force control gives the priority to
force errors over position errors by closing an outer force control loop. The second
group aims at controlling the positions and the interacting forces simultaneously by
decomposing the task space into the directions of the admissible motions and the
interacting forces. Hybrid force/motion control, proposed in [75], belongs to thisscheme. This control with consideration on the dynamics of robot manipulators was
extended in [76, 77]. The state of the art of force control for robot manipulators
is surveyed in [78]. For explicit force control, integral force control with robustness
enhancements in order to reduce the initial impact force was proposed in [80]. In
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 27/183
10
[81], experimental results for explicit force control with an active damping term were
presented.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 28/183
CHAPTER 3
MODELING AND IDENTIFICATION
3.1 EndoBot Design
3.1.1 Manipulator Design Requirements
In laparoscopic surgery, a patient’s abdomen is inflated with carbon dioxide
(CO2) through a needle to lift the abdominal wall away from the organs so as to
expose and access the operating field, and then three pencil-sized small holes are
punctured on abdominal wall for fitting of laparoscopic instruments and camera as
shown in Figure 3.1. Due to the requirement of minimizing the tear of the incisionpoint, a manipulator for MIS has an inherent kinematic constraint (a spherical joint
at the incision point). This constraint is a primary design consideration. This
section describes the mechanical design issues of the EndoBot.
Figure 3.1: Fulcrum effect in laparoscopic surgery.
Fulcrum accommodability Due to the fulcrum constraint, it is required for theMIS surgical robot manipulator, which passes through a small hole, to have
an effective center of rotation at the incision points. The remote center of
rotation can be implemented with several mechanical design such as spherical
joint, spherical link mechanism, double-parallelogram mechanism [37, 20], and
11
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 29/183
12
also can be implemented with tool center position technique with conventional
industrial robot programming.
Backdrivability Backdrivability is needed in the cooperative control mode for the
surgeon to move the manipulator directly. It is also critical for removing the
robot from the patient in case of unanticipated power shutdown or emergency.
Rapid tool changability Surgical operations typically require many surgical tools.
Faster tool change can reduce the overall operation time.
3.1.2 Current Prototype
This section gives a brief overview on the EndoBot manipulator, which was
built by Bernard [18]. By comparing various mechanisms based on the above designconstraints, a simple spherical joint mechanism with semi-circular arches is found
to be a suitable choice because it gives a compact and light design and has mini-
mum number of joints and a mechanical fulcrum constraint, and can be operated in
multiple control mode such as manual mode, shared control mode, and autonomous
mode.
The EndoBot is capable of four degrees of freedom of motion and consists of
two parts as shown in Figure 3.2: rotational stage and translational stage. The
rotational stage creates the spherical motion based on a pair of motor-driven semi-
circular arches for yaw and pitch, and a sleeve that can generate rolling motion.
The translational stage carries the specific tool, which is actuated pneumatically
and is translated along the tool z-axis. All four actuators are DC servo motors and
linear motion for translational stage is performed by a lead screw. All axes are
back-drivable when the motors are not energized. The center of rotation of the arch
joints is at the incision point of the patient’s abdominal wall. Therefore, motion
of the EndoBot will not cause tearing of the incision point. Each tool can easilyslip through the locking hole in the translational stage and the sleeve and be locked
via a locking pin. Figures 3.3–3.4 show the closeups on the semicircular arches and
translational stage carrying grasper tool.
Figure 3.5 shows two manually operated EndoBots and Figure 3.6 shows the
closeup of two EndoBots with grasper and stitching tool.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 30/183
13
Figure 3.2: Mechanical overview of the EndoBot.
Figure 3.3: Closeup on semicircular arches.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 31/183
14
Figure 3.4: Translational stage.
Figure 3.5: Picture of a pair of the EndoBots.
Figure 3.6: Grasper and suturing tools.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 32/183
15
3.2 Kinematic Modeling of the EndoBot
3.2.1 Forward Kinematics
The product of exponentials formula [38] is used to derive the kinematics equa-
tion. Consider the following manipulator with spherical joint shown in Figure 3.7.It consists of four joints - three revolute joints and one prismatic joint, and the base
and tool frame are shown in Figure 3.7. Let hi ∈ 3 be a unit vector, which spec-
ifies the direction of rotation or translation and q i ∈ be the angle of rotation or
linear displacement and pi,i+1 ∈ 3 be the position vector between ith and (i + 1)th
link frame. The transformation between base frame, E 0, and tool frame, E T , at
Figure 3.7: Kinematic diagram of the EndoBot.
q i = 0, i = 1,..., 4 is given by
p0T (0) =
0
I 3×3 0
0
0 0 0 1
, (3.1)
where
I 3×3 =
1 0 0
0 1 0
0 0 1
. (3.2)
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 33/183
16
Then the hi and pi,i+1 can be expressed in base frame as follows:
h1 =
1
0
0
h2 =
0
1
0
h3 =
0
0
1
h4 =
0
0
1
(3.3)
p01 = p12 = p23 =
0
0
0
p34 =
0
0
q 4
p4E =
0
0
0
. (3.4)
The product of exponentials formula gives the forward kinematics map:
R0(q ) = R04 = R12R01R23R34 = eh2q2
eh1q1
eh3q3
eh4q4
(3.5)
p0T (q ) = p04 = p01 + R01 p12 + R12R01 p23 + R12R01R23 p34 = eh2q2eh1q1eh3q3 p34. (3.6)
The individual exponentials are given by
eh1q1 =
1 0 0
0 c1 −
s1
0 s1 c1
(3.7)
eh2q2 =
c2 0 s2
0 1 0
−s2 0 c2
(3.8)
eh3q3 =
c3 −s3 0
s3 c3 0
0 0 1
(3.9)
eh4q4 = I 3×3. (3.10)
where, ci and si are abbreviation of cos(q i) and sin(q i) respectively. Expanding the
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 34/183
17
terms in the product of exponentials formula results in
R0T (q ) =
c2c3 −c2s3 s2
s1s2c3 + c1s3 −
s1s2s3 + c1c3 −
s1c2
−c1s2c3 + s1s3 c1s2s3 + s1c3 c1c2
(3.11)
p0T (q ) =
s2q 4
−s1c2q 4
c1c2q 4
. (3.12)
We parameterize the joint vector q as yaw angle (rotation about the inertial x axis),
pitch angle (rotation about the inertial y axis), roll angle (rotation about the body
z axis), and translation (extension of the tool along the body z axis). Since each
EndoBot only has four degrees of freedom, we can define the configuration of tool
frame as xT = (rT (q ), φ(q 3)) ∈ 4, where rT (q ) = p0T (q ) is the Cartesian position
of the tool frame and φ(q 3) = q 3 is the roll angle. Then the forward kinematics can
be viewed as the mapping xT = f (q ):
xT =
rT
φ
=
x
y
z
φ
=
s2q 4
−s1c2q 4
c1c2q 4
q 3
.
(3.13)
3.2.2 Inverse Kinematics
Due to the spherical joint at the incision point, the inverse kinematics can be
easily solved with the following algebraic identity.
Translational distance, q 4 The translational distance q 4 can be found through
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 35/183
18
the magnitude of rT (q ) =
x
y
z
:
q 4 = rT =
x
y
z
. (3.14)
Second joint angle, q 2
q 2 = tan−1(xc1
z ). (3.15)
First joint angle, q 1q 1 = tan−1(
−y
z ). (3.16)
Roll angle, q 3
q 3 = φ. (3.17)
3.2.3 Jacobian Formulation
3.2.3.1 Geometric Jacobian
Let ω and ν be the angular velocity vector and linear velocity vector of theend-effector. The vector Jacobian of the above manipulator in inertia frame can be
expressed in as follows:
ω
v
=
h1 R01h2 R02h3 0
h1( p1T )0 R01h2( p2T )0 R02h3( p3T )0 R03h4
(3.18)
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 36/183
19
where,( piT )0 = R0i pi,i+1+R0i+1 pi+1,i+2+R04 p4T . Calculating the associated elements
yields the geometric Jacobian:
J =
0 s21c2q 4 + c21c2q 4 0 s2
−c1c2q 4 s1s2q 4 0 −s1c2
−s1c2q 4 −c1s2q 4 0 c1c2
1 0 s2 0
0 c1 −s1c2 0
0 s1 c1c2 0
. (3.19)
3.2.3.2 Analytical Jacobian
The analytical Jacobian can be computed by direct differentiation of the for-
ward kinematics equation. Let f : n → p represent a mapping from joint space
to task space, then the forward kinematics is represented by xT = f (q ) and by the
chain rule:
xT = ∂f
∂q q = J a(q )q, (3.20)
where the matrix J a(q ) = ∂f ∂q
is termed analytical Jacobian.
In general, the analytical Jacobian, J a(q ), is different from the geometric Jacobian,J (q ), for orientation part. In case of the EndoBot, since the roll angle can be avail-
able in direct form, the analytical Jacobian and geometric Jacobian are identically
same as follows
J =
0 s21c2q 4 + c21c2q 4 0 s2
−c1c2q 4 s1s2q 4 0 −s1c2
−s1c2q 4 −c1s2q 4 0 c1c2
0 0 1 0
. (3.21)
3.2.4 Singularity Analysis
Since the Jacobian is a function of the configuration q , the singularity can be
occurred at some configurations with which the rank of J is decreased, i.e., two or
more of the columns of J becomes linearly dependent. When det(J ) = 0, the robot
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 37/183
20
is in a singular configuration, implying that there is certain task space motion that
cannot be achieved (and, conversely, small task space motion would result in large
joint space motion). In order to get a reliable operation, it is necessary to identify
singular configurations of a manipulator, if they exist. To analyze the rank of theJacobian matrix, consider the determinant given by
det(J ) = c2q 24. (3.22)
From (3.22), it is easy to find that the determinant of the Jacobian vanishes when
q 2 = ±π
2, q 4 = 0. (3.23)
This means singular configuration occurs when q 2 = ± π2 or q 4 = 0. In the first
case, the EndoBot would have to lie horizontal, and in the second case, the tip of
the EndoBot would have to be in the center of the spherical joint. Both cases are
outside of the normal workspace and therefore do not have to be considered.
3.3 Dynamics of the EndoBot
The closed-form dynamic equation of robot manipulator can be obtained
through the Lagrangian-Euler formulation based on either D’Alembert’s principle
or Hamilton’s principle. The Lagrange-Euler equation can be written in the form:
d
dt(
∂L
∂ q ) − ∂ L
∂q = Q (3.24)
where L(q, q ) = T (q, q )−P (q ) represents the Lagrangian function, T (q, q ) the kinetic
energy, P (q ) potential energy, q ∈ n the vector of generalized coordinates, q ∈ n
the vector of generalized velocities, Q
∈ n the vector of generalized forces.
3.3.1 Lagrangian Formulation
The lagrangian function of a robot manipulator with n joints is thus given by
L(q, q ) = 1
2 q T M (q )q − P (q ). (3.25)
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 38/183
21
Thus, we have
∂L
∂ q = M (q )q (3.26)
ddt( ∂L∂ q ) = M (q )q + M (q, q ) (3.27)
∂L
∂q =
∂
∂q (
1
2 q T M (q )q ) − ∂ P (q )
∂q . (3.28)
The general equations of motion for a rigid robot manipulator can be written as
M (q )q + M (q, q )q − ∂
∂q (
1
2 q T M (q )q )
C (q,q)q
+ ∂P (q )
∂q G(q)
= τ, (3.29)
where τ is the vector of generalized actuator force. Define C (q, q )q to be the vector
of Coriolis and centrifugal forces and G(q ) the gravity force:
C (q, q )q = M (q, q )q − ∂
∂q (
1
2 q T M (q )q ) (3.30)
G(q ) = ∂P (q )
∂q . (3.31)
Thus, (3.29) becomes
M (q )q + C (q, q )q + G(q ) = τ. (3.32)
3.3.2 Properties of Dynamic Model
The equation of motion has several properties, which are useful for stability
analysis on deriving an energy based control algorithm and dynamic parameter
identification. These properties follows below.
Symmetric and positive definite matrix, M (q ) - The mass-inertia matrix M (q )
is symmetric and positive definite.
Skew-symmetric matrix, M (q ) − 2C (q, q ) - This property is often referred to as
the passivity property and it is useful for deriving control law. Skew-symmetric
means:
xT ( M (q ) − 2C (q, q ))x = 0 for all x ∈ Rn. (3.33)
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 39/183
22
Linearity in the dynamic parameters - Through the reparametrization, we can
find a parameter vector X that is linear in system dynamics. This property
plays an important role on experimental identification of dynamic parameters.
In view of this linearity property, we can write the equation of motion in (3.32)more compactly
τ = ΦT (q, q, q )X, (3.34)
where ΦT (q, q, q ) is the regressor, which is a function of the positions, velocities
and accelerations and X is vector of dynamic parameters.
3.3.3 Dynamic Model of the EndoBot
Due to the spherical joint motion of first three joints of the EndoBot, it would
be convenient to be considered as a system with two links composed of rotational
body and translational body for the dynamic analysis purpose as shown in Fig-
ure 3.8. Define the inertial frame, B, and attach the body frame of first rotational
Figure 3.8: Simplified dynamic model of the EndoBot.
link, O1, to the center of spherical joint and the body frame of second translational
link, O2, to the center of mass of 2nd link. Let lci be the distance between B and
the center of mass of the ith link and mi be the mass of each link. For simplicity,
the inertia tensor relative to a frame attached at the center of mass of each link is
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 40/183
23
assumed to be diagonal:
I ci =
I ix 0 0
0 I iy 0
0 0 I iz
. (3.35)
Then, the total kinetic and potential energies of the EndoBot are given by
T = 1
2 q T J T 1 (q )M 1J 1(q )q +
1
2 q T J T 2 (q )M 2J 2(q )q (3.36)
P = m1gT ( pc1)0 + m2gT ( p02 + pc2)0. (3.37)
Finally, after substituting the Lagrangian L = T − P into Lagrange’s equation, we
have the mass inertia matrix, M (q ) ∈ 4×4, for the EndoBot:
M (q ) =
M 11 M 12 M 13 0
M 21 M 22 M 23 0
M 31 M 32 M 33 0
0 0 0 M 44
, (3.38)
where
M 11
= (I 1x
+ I 2x
)c22 + (I
1z + I
2z)s2
2 + m
2(lc2 + q
4)2c2
2
M 12 = M 21 = −(I 1z + I 2z)s1s2
M 13 = M 31 = (I 1z + I 2z)s2
M 22 = (I 1x + I 2x)c21 + (I 1z + I 2z)s21 + m2(lc2 + q 4)2c21
M 23 = M 32 = −(I 1z + I 2z)s1
M 33 = I 1z + I 2z
M 44 = m2. (3.39)
The Coriolis and centrifugal force terms are given by
C 1(q, q )q = 2[I 1z + I 2z − I 1x − I 2x − m2(lc2 + q 4)2]c2s2 q 1 q 2 + 2m2(lc2 + q 4)c22 q 1 q 4
+ [I 1x + I 2x − I 1z − I 2z + m2(lc2 + q 4)2]c1s1 q 2
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 41/183
24
+ (I 1z + I 2z)(c1 q 2 q 3 + c2q 2q 3 − s1c2 q 22)
C 2(q, q )q = 2[I 1z + I 2z − I 1x − I 2x − m2(lc2 + q 4)2]c1s1 q 1 q 2 + 2m2(lc2 + q 4)c21 q 1 q 4
− (I 1z + I 2z)(c1s2 q 21 + c1 q 1 q 3 + c2 q 1 q 3 + c2s2 q 21)
+ [I 1x + I 2x + m2(lc2 + q 4)2]c2s2 q 21
C 3(q, q )q = (I 1z + I 2z)(c2 − c1)q 1 q 2
C 4(q, q )q = −m2(lc2 q 21 + c21 q 22). (3.40)
The gravity terms are
G1(q ) = −[m1lc1 − m2(lc2 + q 4)]gs1c2
G2(q ) = −[m1l
c
1 − m2(l
c
2 + q 4)]gc1s2
G3(q ) = 0
G4(q ) = m2gc1c2. (3.41)
3.4 Friction Modeling and Compensation
For many servo applications, the joint friction is the main limitation to preci-
sion and performance. It could lead to stick-slip motions, static positioning errors,
or limit cycle oscillations. Systematic lubrication should be implemented from thedesign stage to reduce frictional disturbance. Stiff (high gain) position control can
reduce the frictional positioning error at the expense of possibly destabilizing ef-
fect. Integral action is also a common alternative to reduce the steady-state error
in constant-velocity application. When the friction behavior can be predicted , it
may be compensated by feedforward compensation as in Figure 3.9.
Figure 3.9: Model-based friction compensation.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 42/183
25
The friction compensation requires the knowledge of the friction model and
corresponding parameters. Friction is usually modelled as a map between an instan-
taneous friction force and velocity. Typical friction models are shown in Figure 3.10:
Figure 3.10: Classical friction models.
a) τ f (q ) = F csgn(q ) (3.42)
b) τ f (q ) = F csgn(q ) + F v q (3.43)
c) τ f (q ) =
F ssgn(q ) if q = 0
F csgn(q ) + F v q otherwise(3.44)
d) τ f (q ) = F ssgn(q ) if q = 0
F csgn(q ) + (F s − F c)e−|q/qs|δs + F v q otherwise. (3.45)
The friction parameters could be estimated by observing the static friction-
velocity curve. In [40, 39], a method of finding the friction parameters, F c and F v,
experimentally was proposed. A series of constant input torques is applied to each
joint and the correlating steady state joint velocities are recorded. Fitting the steady
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 43/183
26
state velocities versus input torques would then give the friction parameters.
3.5 Identification of Dynamic Parameters
The advanced model-based control methods to improve the performance re-quire the knowledge of the dynamic parameters. In most cases, computing the
numerical value of the dynamic parameters directly from the design data of the
mechanical structure is not feasible due to the complexity. One way to remedy
this problem is to identify the dynamic parameters experimentally. Identification of
robot dynamic parameters is basically grey-box approach requiring the measured in-
put and output signals with a given system structure resulting from analytic (white-
box) modeling. Due to the fact of linearity in the dynamic parameters the equation
of motion can be given with measurements of φ(t) and τ (t):
τ (t) = φT (t)θ + ε(t), (3.46)
where θ ∈ n is the vector of unknown parameters and ε(t) is the unmeasured noise
signal. More compactly with m measurements at given time instants t1,...,tm,
Figure 3.11: Parametric system identification.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 44/183
27
τ = φθ + ε, (3.47)
τ =
τ (t1)
τ (t2)
.
.
.
τ (tm)
θ =
θ1
θ2
.
.
.
θn
ε =
ε(t1)
ε(t2)
.
.
.
ε(tm)
φ =
φ1(t1) φ2(t1)
· · · φn(t1)
φ1(t2) φ2(t2) · · · φn(t2)
· · · · · ·· · · · · ·· · · · · ·
φ1(tm) φ2(tm) · · · φn(tm)
,
where φ ∈ m×n is the matrix in which each row represents regressor at each time,
τ
∈ m is a vector with τ (tk), and ε
∈ m is a vector with ε(tk). Since ε is unknown,
the exact solution to θ cannot be obtained. One attractive solution may be to seek
to find θ so as to minimize the norm of ε. Selecting Euclidean norm results in the
standard least squares problem:
minθ
φθ − τ . (3.48)
Therefore, the identification problem turns out to find unknown parameters θ from
two sets of measurement data, data about motions and data about torques. The
simplest linear estimation method, the least square method, can be applied with
the overdetermined regressor matrix as in Figure 3.11. Appendix C gives a brief
summary of the least square method. This can also be viewed as an unconstrained
optimization problem with the cost function given by
minθ
J = minθ
1
2εT ε = min
θ
1
2(φθ − τ )T (φθ − τ ). (3.49)
The necessary and sufficient conditions for a minimum are
∂J
∂θ = φT φθ − φT τ = 0 (3.50)
∂ 2J
∂θ2 = φT φ > 0. (3.51)
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 45/183
28
Due to convexity of the cost function, the solution is a global optimum and it can
be obtained from the necessary condition with the gradient of J :
θ∗LS = (φT φ)−1φT τ . (3.52)
It is important to note that the matrix (φT φ) is identical to the Hessian of the cost
function. In order to get the minimum, the Hessian must be positive definite and
this is an equivalent condition of the persistent excitation on input signals, which
will be shown in the next section.
3.5.1 Input Signal Design
When the parameter identification comes to real work, the practical questionmay be how to excite the systems. It is desirable to have systematic guidelines for
designing the input signals. The following factors are considered in this research.
1. Persistent excitation
Let φ be the regression matrix and θ unknown parameter vector. The degree
of persistence of excitation should be high enough with respect to the number
of unknown parameters in θ such that the contribution of each element in θ
can show up separately and consequently θ can be identifiable with invertible
φT φ. Matrix φT φ is often called the persistent matrix or the input correlation
matrix. Therefore, persistence of excitation can be checked with either the
input trajectory or the persistent matrix.
A regression matrix φ is called persistent exciting if there exist positive con-
stants α1, α2 and δ such that
αaI ≤ t+δt
φT φdτ ≤ α2I. (3.53)
In frequency domain, input signal u(t) is said to be persistently exciting of
order p, if the condition on the power spectrum ,Ψu(w) = 0, is satisfied at at
least p distinct points in the interval −π < ω ≤ π. The strict requirement
on the degree of persistent excitation under the noise signal is not yet fully
investigated. However, since input signal satisfying the persistent exciting
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 46/183
29
condition can make the regressor uncorrelated, persistent exciting also can be
obtained with the nonsingular persistent matrix. Clearly, this invertibility of
φT φ is the requirement of existence of least square solution. The main objective
of the input design is to make regressor vectors as uncorrelated among themas possible.
2. Condition number of regressor
This property is a quality tag to the input signals to represent the conditioning
of the persistent excitation. The condition number c(A) of the nonsingular
matrix A is defined
c(A) = σ1
σn, (3.54)
where σ1 ≥ ...σn > 0 are the singular values of matrix A. This number also
represents the sensitivity of estimation to noise and unmodeled dynamics.
The smaller condition number of regression matrix c(φT φ) results in better
identification.
3. Constraints on input amplitudes
Large values of input signals in amplitude are desirable to provide better
signal-to-noise ratio (SNR) in the identification process. However, there exist
the constraints on the input signals in most applications. The input signals
u(k) should be bounded so as to avoid the possibility introducing the nonlin-
earity from input saturation:
umin ≤ u(k) ≤ umax, (3.55)
where umin and umax represent the minimum and maximum values of input
signals respectively. It is desirable to have a way to impose this constraint
into input design process.
4. Input spectrum
The spectrum of the input signal determines the frequencies where the exciting
energy is allocated. The input signal with the well distributed spectrum over
the frequency bands where the system operates mostly results in high quality
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 47/183
30
of parameters. In this respect a white input signal may be the ideal input
signal since it excites all frequencies with same weights. However, most of real
systems have a limited bandwidth as in Figure 3.11. It makes no sense to put
a lot of energy beyond this bandwidth since the systems act as low-pass filters.In this respect, the input signal swith a user specified design parameters to
allocate the input powers over desired frequency bands will be advisable.
There also exist many efforts seeking to find an optimal input trajectory for iden-
tifying robot dynamic parameters. One of the common criteria is the minimization
of the condition number of the regression matrix. This minimization problem is dif-
ficult to solve and usually is a very time consuming procedure. For this reason, theexisting off-the-shelf input signals are investigated for experimental identification.
Summed multisinusoidal signal
A sum of different sinusoidal signals is effective with relatively simple dynamic
model requiring small number of parameters. It has a discrete set of point
spectrum and can provide excitation at certain frequencies. Conceptually it
is simple to generate the signals and has the advantage of long duration of
excitation signal at each frequency. However, it demands many trial and erroriterations to select the appropriate discrete frequencies and relative amplitudes
of each frequency. The quality of identification can vary with the selected
frequencies and amplitudes.
Chirp signal
A chirp signal is a single sinusoid with a time varying frequency:
u(t) = sin(2πf (t)t), (3.56)
where f (t) is the time varying frequency. The common choice of f (t) may be
a linear or logarithmic function of time. The advantage of chirp signal is that
the system can be excited over all specified frequency bands. One shortcoming
may be the trade-off between the duration of excitation at each frequency and
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 48/183
31
the overall identification time. The overall identification time can be reduced
with the fast sweep signal from the lowest to the highest frequency, which
results in relatively short duration of the excitation at each frequency.
PRBS(Pseudo random binary sequence)
A pseudo random binary sequence is a signal that shifts between binary level
in a certain sequence. The difference from RBS (random binary sequence) is
a periodic and deterministic characteristics and the exact sequence can be re-
generated with same starting point. The PRBS can be generated by means of
shift registers and Boolean algebra as in Figure 3.12. The PRBS is character-
Figure 3.12: Generation of PRBS.
ized by two parameters: the switching time (T sw) and the number of registers(nr). The switching time is the minimum time between changes in the binary
level and the maximum length of a sequence is 2nr − 1.
The main usefulness of the PRBS signal lies in the fact that it resembles a
white noise in discrete time and thus excites all frequencies equally well. The
PRBS also uses the maximum power of input signals and consequently has
a high signal-to-noise ratio. The shortcoming with PRBS may be that it is
not possible to design an arbitrary spectrum profile of input signals such as
multiple bands. For the mechanical systems, the exciting with PRBS may
result in tendency of noisy signals in velocity and acceleration signals that can
be obtained from numerically differentiation of measured position signals.
Schroeder-phased signal
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 49/183
32
Consider the periodic and deterministic multisinusoidal signal:
us(k) = λnsi=1
√ 2αi cos(ωikT + φi), (3.57)
where λ is the scaling factor to insure the bounded amplitudes in time-domain
with ±usat, T is the sampling time, ns is the number of sinusoids in one signal
period, αi is the relative power in each component, ωi = 2πiN sT
, N s is the number
of samples in one signal period (ns ≤ N s2 ), and φi is the phase shifting of each
harmonic signal. In order to minimize peaking of the excitation in time domain
while maintaining the same frequency contents, Schroeder [48] proposed that
the phase of each sinusoid has a form:
φi = 2πi
j=1
jαi. (3.58)
The total power is normalized as
nsi=1
αi = 1, (3.59)
where the relative power {αi > 0, i = 1,...,ns} is specified. ns directly de-
termines the order of the persistent excitation and N s determines the low
frequency end of the Schroeder-phased spectrum. The Schroeder-phased sig-
Figure 3.13: Schroeder-phased signal design.
nal can produce almost flat spectral characteristics. The most distinguished
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 50/183
33
beauty against the other signals may be the capability of designing the arbi-
trary desired power spectrum profiles.
0 0.1 0.2 0.3 0.4−1.5
−1
−0.5
0
0.5
1
1.5
time(s)
u ( t )
(a) multisinusoid
0 0.1 0.2 0.3 0.4−1.5
−1
−0.5
0
0.5
1
1.5
time(s)
u ( t )
(b) chirp
0 0.1 0.2 0.3 0.4−1.5
−1
−0.5
0
0.5
1
1.5
time(s)
u ( t )
(c) PRBS
0 0.1 0.2 0.3 0.4−1.5
−1
−0.5
0
0.5
1
1.5
time(s)
u ( t )
(d) Schroeder−phased
Figure 3.14: Comparison of input signals in time domain.
Figure 3.14 compares the input signals in time domain and Figure 3.15 shows
the power spectra of each signal. As shown in Figure 3.15, the flat spectrum profile
can be achieved with the Shroeder-phased signal.
Input signals play an important role and have a major effect on parameter identifi-
cation. Each signal described previously has the tuning parameters that need to be
tweaked in order to excite the system with rich frequency contents. The Schroeder-
phased signal provides several attractive properties compared to other signals, and
therefore, the identification in this thesis was performed with this signal.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 51/183
34
0 1000 2000 3000 40000
1000
2000
3000
4000
w(rad/sec)
p
(a) multisinusoid
0 1000 2000 3000 40000
100
200
300
400
500
w(rad/sec)
p
(b) chirp
0 500 1000 1500 2000 25000
200
400
600
800
1000
1200
1400
w(rad/sec)
p
(c) PRBS
0 500 1000 1500 20000
20
40
60
80
100
120
140
w(rad/sec)
p
(d) Schroeder−phased
Figure 3.15: Comparison of power spectra of input signals.
3.5.2 Dynamic Models for Parameter Identification
Two identification models have been developed for general robot dynamics.
The first model is the differential model, which required the measurements of po-
sition, velocity, acceleration, and joint torque. In order to carry out a practical
identification using this model, accurate acceleration measurement has been key to
the successful identification, but it is a substantial challenge. Most of robots are
seldom equipped with accelerometers from which one can measure the acceleration
directly due to the cost and installation problem. The alternative is a numerical
reconstruction of acceleration by taking a second derivative of the joint positions at
the expense of amplifying the noise.
The second model is integral model, often referred to as the energy model because of
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 52/183
35
coming from the energy theorem, which only requires the measurement of positions,
velocities, and joint torques. It means that this model does not depend explicitly on
the accelerations. Figure 3.16 shows the principles of two identification methods.
Figure 3.16: Comparison of two identification methods.
3.5.2.1 Use of Differential Model
The dynamic equation with friction of the robot’s rigid body model can be
written as
M (q )q + C (q, q )q + G(q ) + N (q, q ) = τ , (3.60)
where N (q, q ) is friction.
More compactly, it can be written as
τ = D(q, q, q )X. (3.61)
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 53/183
36
X is the unknown dynamic parameter vector and is composed of the inertial param-
eters and the friction parameters. D(q, q, q ) is the kinematic information matrix,
which describes the motions of the robot. It has been referred to as the regression
matrix or the observation matrix. If measurements are made at given time instantst1,...,tm along a given trajectory, we may write
τ =
τ (t1)...
τ (tm)
(3.62)
¯D =
D(t1)
..
.D(tm)
. (3.63)
Once we got enough measurements so as to obtain overdetermined Matrix D(q, q, q ),
the unknown dynamic parameter vector X can be determined by the least-square
estimation method:
X = ( DT D)−1 DT τ , (3.64)
where ( DT D)−1 DT is the left pseudo-inverse matrix of D. Note that it is important
to use the minimum set of (nonredundant) dynamic parameters referred as the base
dynamic parameters by grouping parameters together such that the matrix D is
always full-rank [47].
3.5.2.2 Use of Energy Model
In order to eliminate any numerical derivation of velocity in the identification
process to get the joint accelerations, the model based on the energy theorem has
been used [49, 50, 55]. From the energy theorem, which states that the total of
mechanical energy applied to the system is equal to the change of the total energy
of the system, it is tbta
τ T e qdt = H (tb) − H (ta), (3.65)
where E (ti) is the Hamiltonian of the system at time instant ti and τ e does not
include friction torques. Since the total energy(kinetic and potential energy) is
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 54/183
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 55/183
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 56/183
39
By moving only each joint, we can get the following inertial parameters and the
friction coefficients for each joint expressed by 4 × 1 vector:
X T = [X 1 X 2 X 3 X 4] = [I ml F c F v]. (3.75)
With the assumption of neglecting the electrical dynamics for the permanent magnet
brush DC motors, we can get the joint torques from the input control voltages in
current drive mode:
τ = K a · K t · K g · vc, (3.76)
where K a is the amplifier gain, K t is the torque constant, which characterizes the
electromechanical conversion of armature currents to the torques, K g is the gear
ratio, and vc is the input control voltages. Table 3.1 shows these parameters used
in experimental identification.
Table 3.1: Parameters of the electrical system.
Parameters Joint 1 Joint 2 Joint 3 Joint 4K a [amp/volt] 0.375 0.375 0.375 0.375
K t [Nm /amp ] 0.01342 0.01342 0.00652 0.00996K g 134 134 72.88 66
3.5.3.1 Identification of the Friction Coefficients in Energy Model
The vectors in the regression matrix associated the Coulomb and viscous co-
efficients are prone to be linearly dependent in the energy model. The functions
|q | and q 2 are correlated and thus, a strong linear dependency between the vectors
exists. This fact gives rise to the difficulty of distinguishing each contribution to
regressor. Consequently, it is very difficult to identify the each coefficient separatelyin energy model. As an experimental verification, the Schroeder-phased input sig-
nals were applied to the first joint of the EndoBot and the regression matrix was
obtained. The comparison on the coefficient of cross correlation between column
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 57/183
40
Table 3.2: Coefficients of cross correlation.
column 1 2 3 41 - -0.8219 0.0004 0.0
2 - - 0.1651 0.21133 - - - 0.94424 - - - -
vectors in regression matrix gives an indication of correlation:
rXY = cov(X, Y )
σX σY , (3.77)
where rXY is the coefficient of cross correlation between X and Y vectors, cov(X, Y ) =E [(X − E (X ))(Y − E (Y ))] is the covariance , and σX and σY are the values of stan-
dard deviation. The Table 3.2 shows the comparison on the coefficient of cross cor-
relation between vectors. Clearly, the correlation between third and fourth column
is greater than those between others. This problem can be solved by identifying the
friction parameters first and then compensating the corresponding friction torques
by subtracting them form the generalized torques. The identification of friction pa-
rameters is performed by implementing the velocity control loop as in Figure 3.17.
The Table 3.3 shows the experimental results with friction parameters. E ij stands
Figure 3.17: Block diagram of velocity control loop.
for jth link of ith EndoBot.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 58/183
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 59/183
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 60/183
43
where p determines where the roll-off occurs. The discrete washout filter was im-
plemented by the backward approximation with the location of pole, p = 100:
f washout(z ) = p(z − 1)
( pT s + 1)z + 1, (3.79)
where T s = 0.001s is the sampling time.
3.5.4 Validation of Parameter Identification
The validation of the identified parameters is a crucial step before proceeding
the controller design. In this research, the parameters are validated by comparing the
simulation with the obtained parameters and real output with a given same torque
trajectory. The difference between the measured outputs y(t) and the estimated
0 2 4 6 8 10 12 14 16 18 20−1
−0.5
0
0.5
1
time
i n p u t t o r q u e [ N m ]
0 2 4 6 8 10 12 14 16 18 20−1
−0.5
0
0.5
1
time
p o s i t i o n ( r a d )
simulated trajectory
measured trajectory
Figure 3.20: Comparison on measured and simulated output trajectories
outputs φT θ is called the residual. The residual error analysis with the inputs was
performed to see the independence of the input:
ε(t) = y(t) − y(t|θ) = y(t) − φT (t)θ. (3.80)
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 61/183
44
The coefficient of cross correlation between the residual and input is very small (rεu
= 0.0818). The result indicates that there is very small correlation between the
residuals and the inputs and therefore the obtained model can be used. Figure 3.20
shows the comparison on simulated and measured output trajectories. There existsthe small difference between these two trajectories resulted from wishful assumptions
on model such as the ignored electric dynamics to simplify the analysis, the position
dependent friction coefficients, which are assumed to have a position independent
characteristics, and unmodeled nonlinearities. Therefore, it is desirable to design
the robust controller that can withstand these variations.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 62/183
CHAPTER 4
SUTURING IN MINIMALLY INVASIVE SURGERY
4.1 Suturing Task Analysis
From closing the wound with an ant’s head in ancient time to today’s laparo-
scopic stitching with absorbable materials, suturing has a long history and is one
of the most difficult tasks and uses a significant percentage of operating time in
MIS. Although there is ongoing research on alternatives to suturing technique such
as tissue gluing, stapling, and thermal tissue bonding, suturing is still the primary
tissue approximation method. However, the laparoscopic suturing task demands
high level of skills and endurance on surgeons. Despite the needs to be performed
autonomously, no research efforts have been published. It is worth while to decom-
pose the suturing task into the subtasks and figure out what technical difficulties is
encountered with each subtask. There are several knot tying techniques in surgery
[17] and two of them are shown in Figure 4.1.
Square Knot - This is the easiest and reliable method for all suture materials.
Basically it consists of two half knots in opposing directions.
Surgeon’s Knot - This knot requires an additional loop in the first throw and thus
the surface contact of suture is increase. It results in doubling the friction for
initial half knot and therefore more stable than the square knot. This knot
generally is used in open surgery and sometimes referred to as a friction knot.
Figure 4.1: Knot tying techniques.
45
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 63/183
46
The most common knot technique in surgery is the square knot. It is an ideal knot
for laparoscopic surgery because it requires fewer steps than the surgeon’s knot [17].
For this reason, the only square knot technique has been considered for analysis
in this thesis. Based on the observations of the manual suturing operations, thesuturing task can be broken down into the following five subtasks:
1. stitch,
2. create a suture loop,
3. develop a knot,
4. place a knot, and
5. secure a knot.
The following are the technical difficulties related to each subtask.
Stitch The stitching subtask involves in entrance and exit bites, extracting the
needle, and pulling out the suture. The task analysis shows that the greatest
difficulty for this subtask is to manipulate the curved needle between two
graspers without slipping or dropping the needle.
Create a suture loop In the subtask of loop creating, the wrapping motion is
required to create a suture loop with the loop strand. It is the most difficult
motion to be autonomously implemented because the trajectories of suture
are not predictable due to the flexibility. The surgeon creates the loop based
on visual feedback information and his personal experience.
Develop a knot Once created the loop, we need to pass either needle or short tail
through the loop to develop a knot. The difficulty will be grasping the short
tail with one of the grasper and pull the suture through the loop.
Place a knot Once developed the knot, the knot needs to be placed into the niche.
Placing of the knot requires sliding of the knot. The difficulties will be to de-
termine the sliding condition and obtain the optimal trajectory that minimizes
the tearing trauma on the tissue.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 64/183
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 65/183
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 66/183
49
Figure 4.4: Components of a suture.
on the suture in order for the length to increase. Let P o be the position of the exit
point and P be the current tip position of needle holder. Define ξ = (P − P o). P is
the velocity vector of the needle and L (a scalar) is the rate of change of the suture
length. To keep track of the suture length, we need to find a relationship between L
and P . If P and ξ form an obtuse angle ( P ·ξ < 0), then the suture is not in tension,
therefore L = 0. If P and ξ form an acute angle ( P · ξ > 0), we need to consider
two possibilities. If ξ ≥ L, then the suture would be in tension. Otherwise, the
suture would not be in tension and L = 0. The rate of change for the suture length
is summarized below:
L(t) =
0 if ξ (t) · P (t) ≤ 0 or ξ (t) < L(t)
ξ (t) · P (t) if ξ (t) · P (t) > 0.(4.1)
4.3 Motion Control in the Stitching Task
4.3.1 Dynamic modeling of Stitching
Whenever the robot is in the task of stitching, additional forces are are required
to pull out the suture from the tissue. Let f denote the force applied to the end of suture. Then, the dynamic equation of the EndoBot during stitching can be written
as
M (q )q + C (q, q )q + G(q ) = τ − J T f, (4.2)
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 67/183
50
where q is the joint displacement vector, M (q ) is the inertia matrix, C (q, q )q is the
vector function characterizing Coriolis and Centrifugal forces, G(q ) is the gravita-
tional force, J (q ) is the Jacobian matrix, τ is the applied joint torque, and f is the
suture tension force exerted by the robot at the end-effector.
4.3.1.1 Kinematics of the Suture Motion
The suture extension is the displacement of the end of the suture from the
stitch position on tissue and certainly it is the function of the joint angle. The
peculiarity is that the suture extension has a directionality and the rate must be
positive. Let the stitch position be in the origin and x(t) represent the current
position in Cartesian coordinate from forward kinematic mapping, then,
(t) =
xT (t)x(t) if in tension
pmax otherwise,(4.3)
where pmax is the previous maximum extension.
Then, the suture extension rate can be obtained
(t) =
12
2xT (t)x(t)√ xT (t)x(t)
= xT (t)x(t)x(t) = K (q(t))T J q(t)
K (q(t)) if in tension
0 otherwise.
(4.4)
4.3.1.2 Static Model of the Suture Tension
For simplicity, we assume that the suture is massless and hence has no dynam-
ics. In practice, this is a good approximation since the suture has negligible mass.
We further assume that the suture has inelasticity. During pulling out the suture,
the force applied to the suture in tension is given by
f = x(t)
x(t)(F
ssgn(l) + F
vl), (4.5)
where, l is the suture extension rate as pulled out by the robot and F s and F v re-
spectively denote the static and viscous friction coefficients between the suture and
the tissue. The important point here is that the suture tension force f must be pos-
itive because suture cannot generate negative tension force due to the positiveness
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 68/183
51
of (t).
Then, the dynamics become
M (q )q + C (q, q )q + G(q ) = τ − J T x(t)
x(t)(F ssgn(J q ) + F v J q ). (4.6)
4.3.2 Problem Formulation
During the autonomous stitching, the tension applied to the end of suture
must be overbounded with f max and the joint variables should be regulated about
the desired position independently of the initial conditions. The control objective
here is to choose the controller to solve the following regulation problem.
Given dynamic equation (4.6) and q d, d, and f max with initial conditions of
q (0), ˙q (0) = 0, and (0), determine a feedback control law, τ , so that the closed-loop
system satisfies
q (t) → q d as t → ∞
q (t) → 0 as t → ∞
(t) → d as t → ∞
f (t) ≤ f max ∀ t,
where q d is the desired position, d is the desired suture length after stitching, and
f max is the force limit above which the tissue will be torn off.
4.3.3 Region of the Feasible Motion
Once a needle has been passed through the tissue for new stitching, the geo-
metric constraint from the finite suture length is created. The norm of tip position
of the EndoBot must be radially overbounded with d:
K (q (t)) − K (q (0)) ≤ d. (4.7)
It means that the all motions as well as the desired position must be in the domain
of the feasible stitching motion.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 69/183
52
4.3.4 Problem Simplification
From the requirement of desired suture extension, the control objective can be
broken down into the following two sequential steps.
First, the needle must hit on the surface of region of feasible stitching motion whilesatisfying the tension inequality condition:
q (t) → q d as t → ∞, f (t) ≤ f max ∀ t, (4.8)
where q d is the desired joint position with which the needle lies on the region and
it is equivalent to (t) → d.
Then, the robot can move to the desired position:
q (t) → q d as t → ∞, f (t) = 0 ∀ t. (4.9)
Obviously, the second problem is the purely position control due to the directional
characteristics of the suture tension and the stability is already well proved.
The simplified control problem is given by
Given d and f max, design a feedback control raw, τ , such that (t)
→d
as t → ∞ and f (t) ≤ f max ∀t.
4.3.5 Problem Transformation
• Case I.
It is worth while to take a look at the force applied to the suture in tension in
(4.5). The tension force has a positive value only if the suture extension rate is
positive and certainly the suture extension rate, (t), comes from the motion
of the robot. From this point of view, we can transform the force inequalityconstraint into a motion constraint. Let f max be the maximum tension force
under which the suture is pulled out:
f max = F ssgn(max) + F v max (4.10)
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 70/183
53
max = f max − F s
F v. (4.11)
Then, the simplified control problem described above might be written as fol-
lows.
Given d and max, design a feedback control raw, τ , such that
(t) → d as t → ∞ and (t) ≤ max(t) ∀t.
Clearly, the motion control problem with overbounded tension force can be
transformed into the pure motion control problem with a constraint of upper-
bounded velocity. What is the benefit from this transformation?
In case of that the force information is not available, we still can bound the
tension force by f max in order to do a stitch without the problem of tearingoff the tissue.
• Case II.
Suppose we know the maximum tension force, f max, and let f d be less than
f max. Now, the control problem is a regulation problem with a desire position
and a desired force.
Given d and f d, design a feedback control raw, τ , such that (t)
→d
as t → ∞ and f (t) = f d < f max ∀t.
4.4 Knot Tying
In this section, we present results on autonomous suturing using a pair of the
EndoBots. One of the EndoBot has a grasping tool and the other a stitching tool.
Both tools are built from disposable tools made by US Surgical Corp. (USSC).
The mechanical handles were cut and mated with pneumatic drives. The grasping
tool can be commanded open or close. The stitching tool contains two jaws, eachcan lock in the needle. The tool can also pass the needle between the jaws. The
challenges of suturing operation include,
• Only the needle position is known. The suture position is not directly mea-
sured.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 71/183
54
• The suture and the tissues are both flexible.
• The workspace is limited.
We will consider three algorithms of automatic ligation in this section. The
first one uses a standard manual stitching tool instrumented for robotic operation.
The second one modifies the grasping tool with a flexible hook to facilitate the knot
tying process. The last one modifies the grasping tool to have an articulated finger.
At the present, the first and second methods have been implemented and the last
method is currently under development.
4.4.1 Ligation Algorithm 1
The key observation for tying a simple knot is that if the suture can be placedover the jaw carrying the needle, then a loop can be formed by passing the needle
to the other jaw. For a human surgeon, this step is performed by putting the jaws
over the thread and then pass the needle. This is not possible for the EndoBots
since the thread is flexible and the position is not directly measured. Instead, we
use the rigidity of the grasper to guarantee that the suture is placing over the jaw.
Automatic tying a simple knot can then be accomplished through the following steps
(shown schematically in Figure 4.5):
1. Make a single stitch near the wound and pull out the suture so that leave
a small suture tail. This may be done manually by the surgeon using the
manual mode or semi-autonomously. In the latter case, the surgeon manually
grasps both sides of the wound with the grasper and uses the foot pedal to
command the stitcher to make a stitch. The stitcher then retracts until a
specified amount of suture has been pulled through the suturing point.
2. Grab the suture tail with the grasper tip. This may be done manually by
the surgeon using the manual mode or semi-autonomously. In the latter case,
the grasper predicts the location of the suture tail based on the location and
direction of the suture performed in the previous step.
3. Move the stitcher so that the open jaw (the jaw without the needle) touches
the front of the grasper stem. This location should be as far up the grasper
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 72/183
55
stem as the workspace would allow. Denote the point of contact between the
stitcher jaw and grasper stem P .
4. Rotate the stitcher 180◦ about the axis OP where O is the center of the
spherical joint. The angle form between the grasper and the stitcher would
have to be sufficiently large to guarantee that the tip of the needle does not hit
the stem of the grasper. At the completion of this step, the thread has to lay
over the open jaw. Otherwise the surgeon would have to intervene manually.
5. Move the stitcher towards the grasper until the grasper stem is within the
open jaw of the stitcher. The grasper stem could fit through the jaw opening,
but it is too thick for the stitcher jaws to close. The stitcher therefore needs
to move along the grasper stem until it reaches the grasper tool tip.
6. Rotate the grasper so that the narrow side faces the jaw opening. The stitcher
can now close and pass the needle to the other jaw.
7. Retract the stitcher to tighten the knot.
The above procedure would create a simple knot. Two simple knots may be com-
bined to form a square knot. However, the second simple knot must be the mirror
image of the first. Otherwise, the knot is known as the granny knot, which is not
secure. The only modification is that at Step 3, the stitcher should touch the back
of the step and in Step 4, the rotation should be −180◦. The square knot procedure
may then be repeated to form a surgeon’s knot.
The above procedure works well most of the time in the laboratory, but has
the following drawbacks:
• In the Step 4 above, a large angle is required between the two instruments.
For knots near the center of the workspace, this may not be feasible.
• Because of errors in positioning, the thin suture thread could fall between the
open jaw and the grasper stem.
• During retraction stage, the thread could get tangled.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 73/183
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 74/183
57
4.4.2 Ligation Algorithm 2
The first algorithm is a step toward automatic laparoscopic suturing, but suf-
fers several drawbacks as to render the procedure less than robust. To address these
issues, we made a small modification of the conventional grasping instrument. Thekey observation is that if we can hold on to any part of the suture at a known po-
sition, the rotation of the stitcher would be unnecessary. To achieve this, we added
a reciprocating actuator connected to a flexible hook over the grasper hinge so that
it can be extended or extracted as needed (see Figure 4.6).
Figure 4.6: Flexible hook for catching the suture.
The simple knot algorithm is now modified as described below (shown schemat-
ically in Figure 4.7 and pictures from the experiment in Figure 4.8):
1. Perform Steps 1–2 in Algorithm 1.
2. Extend the flexible hook. Move the stitcher over the hook from the front to
back so that the suture hangs over the hook.
3. Perform Steps 5–6 in Algorithm 1.
4. Retract the flexible hook.
5. Retract the stitcher to tighten the knot.
Again, a mirror image of the first simple knot needs to be performed to ensure a
secured square knot. It is done by replacing the motion over the hook to back to
front (instead of front to back). This algorithm does not have the angle limitation
as in Algorithm 1. The motion over the hook can be designed so the thread is
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 75/183
58
always snared by the hook, so the positioning requirement is not as tight as before.
Finally, since the hook is close to the grasper tip, the grasper could retract by a
small amount as to avoid thread tangling.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 76/183
59
Figure 4.7: Autonomous simple knot tying algorithm 2.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 77/183
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 78/183
61
4.4.3 Ligation Algorithm 3
The two ligation algorithms described above require the use of the special
stitching tool designed for endoscopic surgeries (specially, the EndoStitch by USSC).
For many surgeries, for example, anastomosis (connection between blood vessels), asemi-circular type of needle is used with two graspers. In such cases, the algorithms
presented so far are not applicable. In this section, we consider another type of
grasper (such as in [11]), which contains an additional universal joint. Then forming
a loop can be accomplished by just wrapping the thread around the bent arm.
The detailed sequence for tying a simple knot is as described below (see Fig-
ure 4.9):
1. Hold the needle between needle holder and make a single stitch near the wound.
Pull out the suture so that leave a small suture tail.
2. Move the needle holder around the bent grasper tip to create a loop.
3. Move the two instruments together so that the bent grasper grabs the tail of
the suture while maintaining the loop wrapping around the bent stem.
4. Retract the grasper to tighten the simple knot.
To form a square knot, the second simple knot needs to be done with the loopformed in the opposite direction. Note that the only reason that a bent tip grasper
is needed is to ensure that the loop does not slip off the grasper.
Figure 4.9: Autonomous simple knot tying algorithm 3.
This algorithm has several advantages over the previous ones. There is no
limitation on the relative positions between the two endoscopic instruments, and
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 79/183
62
it does not require the use of a special stitching tool (and the associated needles).
This algorithm can also be extended to more general knots. For example, looping
twice around the bent arm would make a friction knot. We are also currently
evaluating the possibility of using the grasper with a hook described in Algorithm 2to implement this algorithm.
It should be noted that each step in above knot tying algorithms is performed
autonomously under the surgeon’s supervision on suture entanglement and failure
of suture tail grasping to confirm the advance to the next step. In the future, it
might be anticipated to perform suturing task fully autonomously with additional
sensing capability such as vision system with less intervention or self confirmation.
4.5 Placing the Knot
4.5.1 Problem Formulation
Once developed the knot, the next step would be to place the knot on the
tissue. Seating the knot on the tissue involves the magnitude of applying tension
as well as the direction of tension. To investigate the motion of seating the knot,
we can consider the simple knot tying in Figure 4.10. The knot forming procedure
is illustrated in the left figure. Let a post strand represent the suture that the
knot will be tied around and a loop strand represent the suture, which is loopedaround the post strand as shown in Figure 4.10. A simple knot can be developed by
wrapping around the post strand with the loop strand. After developing the knot,
applying the tension at both post and loop end results in a simple knot as shown in
Figure 4.10. Define pa represent the position where the needle enters to the tissue
and pb the position where the needle exits from the tissue. Let pn be the position
of knot and p0 the position at which the knot will be seated.
It was reported that applying tension in nearly horizontal would be desirable in
order not to get stuck in proceeding the placement of the knot. No report, however,
was published about the sliding condition, which guarantees the proceeding the
placement of the knot. The following section will consider the motion of a simple
knot and derive the sliding condition. The problem of the placement of a knot can
be formulated as follows.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 80/183
63
Figure 4.10: Forming a simple knot.
• Given geometrical information of knot ( pa, pb, pn, p0) and friction coefficient of
suture ( µ), determine the sliding condition in which the placement of a knot
can be proceeded.
• Given a desired knot placing trajectory, pn(t), in terms of placing rate ( s),
determine the trajectories of suture end, p1e(t) and p2e(t), that bring the knot
on trajectory of pn(t).
4.5.2 Sliding Condition of Knot Placement
Placing the knot down on the tissue results from sliding of two strands rela-
tively. To investigate the sliding condition, we need to have a model of the knot,
which basically is consisted of two sutures wrapped around each one. The modeling
of a knotted suture is quite difficult and very complicated. Due to the fact that
the proceeding the placement of a knot is highly dominated by the friction between
knotted sutures, the suture local deformation is not considered in this research. A
simplified model of knotted sutures can be obtained from the two point contact
model with assumption of maintaining the half knotted configuration where the
equivalent normal forces and the friction forces pass through each point as shown
in Figure 4.11. Then, the free-body diagram can be drawn in Figure 4.12.
The problem can be stated as follows: Given F 1a, F 2a, F 1b F 1b
, and F 2b F 2b
, de-
termine the sliding condition. Define e1a = F 1a F 1a
, e2a = F 2a F 2a
, e1b = F 1b F 1b
, and
e2b = F 2b F 2b
and let en = e1a+e1ae1a+e1a
be the unit vector along which the normal forces
apply and et be the unit vector along which the friction forces apply. When the
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 81/183
64
Figure 4.11: Two point contact model of the knotted suture.
Figure 4.12: Free-body diagram for the knot sliding problem.
knot is in equilibrium, F = 0, so
F 1ae1a + F 2ae2a + F 1be1b + F 2be2b = 0. (4.12)
It has two unknown variables, F 1b and F 2b, and two component equations as follows:
eu · (F 1ae1a + F 2ae2a + F 1be1b + F 2be2b) = 0 (4.13)
ev · (F 1ae1a + F 2ae2a + F 1be1b + F 2be2b) = 0. (4.14)
It becomes
F 1a cos β 1 − F 2a cos β 2 − F 1b cos γ 1 + F 2b cos γ 2 = 0 (4.15)
F 1a sin β 1 + F 2a sin β 2 − F 1b sin γ 1 − F 2b sin γ 2 = 0. (4.16)
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 82/183
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 83/183
66
sliding trajectory, which lies in the normal direction to the tissue, henceforth the
symmetric case will be considered as shown in Figure 4.14:
β = β 1 = β 2 (4.21)
γ = γ 1 = γ 2 (4.22)
F 1a = F 2a. (4.23)
Then,
Figure 4.14: Free-body diagram of one strand in symmetric tension.
F 1a cos β − F 1a cos β − F 1b cos γ + F 2b cos γ = 0 (4.24)
F 1a sin β + F 1a sin β − F 1b sin γ − F 2b sin γ = 0. (4.25)
From (4.24) and (4.25),
F 1b = F 2b = sin β
sin γ F 1a. (4.26)
It means that the tension in the bridge strand can be determined from only the
applying force, F a, and the geometric conditions, β and γ .
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 84/183
67
Moment equations about p and p give
f 1an = F 1a sin β (4.27)
f 1bn = F 1b sin γ. (4.28)
The coulomb’s law states the sliding occurs when
F 1a cos β − F 1b cos γ > µ(f 1an + f 1bn). (4.29)
It becomes
1
tan β − 1
tan γ
> 2µ. (4.30)
It gives the following sliding condition:
0 < β < tan−1( tan γ
2µ tan γ + 1). (4.31)
It means that the sliding begins when the β is sufficiently small to satisfy the in-
equality condition dependent on the coefficient of friction between the strands and
γ . The sliding condition of (4.31) can be represented geometrically as shown in
Figure 4.15. It is a three dimensional shape, which can be obtained by extracting
a cone from a cylinder shape. It means that the force lying in this shape can slide
down the knot. Due to the constraint, which requires the applied tension to lie on
Figure 4.15: Geometric interpretation of the sliding condition.
the plane consisting of pa, pb and pn, the sliding condition can be represented with
two right-angled triangles in a plane as shown in Figure 4.16.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 85/183
68
Figure 4.16: Sliding condition in a knot tying plane.
4.5.3 Trajectories of Suture Ends for Placing a Knot
The knot position can be determined from the end motion of the sutures due
to the length invariant constraints. Let q be the current knot position and p1e and
p2e be the current position of each end of strands. During motion, we assume that
Figure 4.17: Evolving trajectory of the knot.
the length of each strand is constant:
l1 + l1e = l1 + l1e = c1 (4.32)
l2 + l2e = l2 + l2e = c2. (4.33)
When the suture ends move to new positions, the knot position q
can be obtainedfrom the following length invariant constraint equations (which is imposed so that
the strands remain in tension):
p1e − q + q − p10 = c1 (4.34)
p2e − q + q − p20 = c2. (4.35)
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 86/183
69
By choosing the suture end position trajectories, we can shape the knot position
trajectory. We focus on the case that the desired knot trajectory is a straight line
normal to the tissue. This case can be achieved with only symmetric pulling. The
problem becomes finding p1e(t) to move the knot along pn(t) as shown in Figure 4.18.Let ξ denote the unit vector along the desired direction of knot motion, and s be
the desired sliding rate. The angle γ (t) and the distance between pb and pn evolve
Figure 4.18: Trajectory of the loop end for placing the knot.
according to
pn(t) = pn0 + (st)ξ (4.36)
δ (t) = pn0 − pb − pn(t) − pb (4.37)
γ (t) = tan
−1
( pn0
−st
pb ). (4.38)
Due to the sliding condition, (β min ≤ β (t) ≤ β max(t) = f (γ (t), µ)), the trajectory of
the suture end to just start knot motion is
p1e(t) = p1e0 + δ (t)(cos(β (t))u + sin(β (t))v). (4.39)
Excessive tension at the tissue at pa and pb could damage the tissue. The optimal
trajectory in the sense of minimizing the tension force in the bridge strand can be
obtained by using the minimum β (t) = β min:
p1e(t) = p1e0 + δ (t)u + (st)ξ. (4.40)
Figure 4.19 compares the trajectories with β (t) = β max(t) and β (t) = 0.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 87/183
70
−1 −0.5 0 0.5 1 1.5 2 2.5 3
0
0.5
1
1.5
2
2.5
3
3.5
4trajectories of loop end
u [cm]
v [ c m ]
PnBeta*
Beta=0
loop end
Figure 4.19: Trajectories of loop end.
4.6 Controller Requirement and Architecture
In surgical robotic applications, it is not possible to get the models, which
completely predict all dynamics of the system over the entire range of operating
due to the non-deterministic nature of behavior of environments such as flexible
sutures and soft tissues. In order to guarantee the safe operation in surgical tasks,
the controller architecture reflects that the proceeding of task must evolve under
surgeon’s supervision. This supervisory controller gives the capability of dealing
with uncertainty and interactive commands from surgeons according to the real
situation without increasing the complexity. In this section, the architecture of the
EndoBot’s controller will be described.
4.6.1 Hybrid Dynamic System4.6.1.1 Hybrid State
From the planning perspective, suturing can be seen as sequencing states of the
dynamic system whose states evolve continuously with time or discretely according
to asynchronous events. Inherently it is a hybrid dynamic system in nature. In
order to model the continuous time systems as well as the discrete event systems,
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 88/183
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 89/183
72
Figure 4.20: State invariant space.
to next state. In robotic applications, the invariant space for continuous state x(t)
can be defined from trajectory planner as shown in Figure 4.21. In Figure 4.21, ei
Figure 4.21: Hybrid state transition diagram.
represents the planned event, which occurs when the evolution of continuous state in
si−1 reached the previously defined state and ei represents the unpredictable event
due to uncertainty of the model.
4.6.2 Human Sharing Supervisory Controller
A supervisory controller can be thought of as a discrete event system, which
issues the transition of states and receives events, which indicate either the success-
ful evolution of current state or the occurrence of an error condition. Due to the
fact that all possible events cannot be predicted during the planning phase, the con-
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 90/183
73
troller architecture must reflect that the proceeding of task evolve under surgeon’s
supervision. Figure 4.22 shows the controller architecture implemented in this re-
search. In this research, the human sharing supervisory control is used to supervise
Figure 4.22: Human sharing supervisory controller.
the surgical operations. It can give the systems decision making and fault recovery
capabilities through the surgeon. The Autonomous Discrete Event System (ADES)
takes the events and causes the state transition in autonomous mode. The Human
Discrete Event System (HDES) takes the events which either the surgeon or the
event generator issues in order to correspond to the error occurrence.
4.6.3 Planning of Suturing Task
Each Endobot can have three states in high level as shown in Figure 4.23:
Lock, manual, and autonomous states. The transition to the lock state can be
started from issuing the lock event by either the surgeon or the ADES. In lock
state, the robot manipulator is under a control law, which regulates the current
position, and only two events for transition to the manual or autonomous state can
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 91/183
74
Figure 4.23: High level state transition diagram.
be received. The Inv can be defined as
Inv(S L) = {x(t) ∈ Inv(S L) : x1(t) = xcp, x2(t) = 0}, (4.44)
where x(t) = (x1(t), x2(t)), x1(t) represents the position state vector, x2(t) represents
the velocity state vector, and xcp denotes the current position vector. In manual
state, the surgeon can take all control commands and HDES (Human Discrete Event
System) is active to correspond to the event issued by the surgeon. The manual
state, S M , has substates, S Mi, according to values of discrete states as shown in
Figure 4.24. In contrary to the lock state in (4.44), the Inv of the manual substate
can be viewed as a whole workspace:
Inv(S Mi) = {x(t) ∈ Inv(S M i) : xmin ≤ x(t) ≤ xmax}, (4.45)
where xmin and xmax define the workspace. The autonomous sequencing of suturing
algorithm can be evolving in the autonomous state under the surgeon’s supervision
and the surgeon can issue the event, which causes the transition to the lock state in
order to gain the control power for dealing with uncertainties.
In Figure 4.25, ei represents the planned event issued by the ADES when the
evolution of the current state ends and ei represents the unpredictable event duringthe evolution due to the real situation, which cannot be predicted. The ei makes
the transition to the NOT BE state, si. In this case, it can transit to the previous
state and try it again by generating the event ei−1. The surgeon also can take the
control in this case and he or she can recover the error in manual state and continue
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 92/183
75
Figure 4.24: Manual state diagram.
Figure 4.25: Autonomous state evolution.
to evolve the algorithm by transiting into the autonomous state. Figure 4.26-4.29
show the state transition diagram for suturing task and Figure 4.30 shows the state
evolution for creating a knot. Two local coordinates of the EndoBots, E 1 and E 2,
are calibrated with 3D Digitizer with respect to the predefined base coordinate, B ,
such that the transformation between these frames can be achieved as shown in
Figure 4.31.
4.6.4 Development Environment
In this research, all controllers, the high level discrete event system controller
and the low level motion controller, are implemented in Simulink with ARCSlink
library (ARCS, Inc.), which contains the Simulink device drivers of the DSP Lighting
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 93/183
76
Figure 4.26: State transition diagram for stitching task.
Figure 4.27: State transition diagram for grasping suture tail.
Figure 4.28: State transition diagram for creating the knot.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 94/183
77
Figure 4.29: State transition diagram for securing the knot.
Figure 4.30: State evolution for creating the knot.
Figure 4.31: Calibration of two coordinates.
motion controller (ARCS, Inc.). Once created the Simulink block, C code can be
automatically generated using Real-Time Workshop (MathWorks, Inc.) and can be
compiled and linked with Texas Instruments C compiler so that it can be downloaded
to the DSP motion controller through AIDE (ARCS, Inc.), which can provide the
run time environment through ISA bus. Once downloaded the execution file, AIDE
can communicate with the target system. Figure 4.32 shows the overview of these
development environments.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 95/183
78
Figure 4.32: Overview of the experimental environment.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 96/183
CHAPTER 5
MOTION CONTROL
5.1 Gravity Compensation in Manual Mode
In the manual mode, the motors are back-drivable allowing the surgeon to
manual move the tip of the tool in roll-pitch-yaw and translation directions. As
early mentioned, most procedures in MIS are time consuming and in manual mode
the surgeon might be tired to deal with operating two EndoBots. Since the manual
operation is performed with low speed and the gravity term will tend to be the key
external static force, the gravity compensation is applied to minimize the surgeon’s
command input. In the manual mode, the joint torque is applied to compensate
gravitational force and moment:
τ = τ g, (5.1)
where the gravitational vector τ g is given by
τ g =
−[m1lc1 − m2(lc2 + q 4)]gs1c2
−[m1lc1
−m2(lc2 + q 4)]gc1s2
0
m2gc1c2
. (5.2)
5.2 Motion Control in Autonomous Mode
As the surgeon gains greater confidence in using the EndoBots, certain pro-
cedures can be performed autonomously under the surgeon’s supervision. In this
section, two motion control schemes implemented in the EndoBots system will be
discussed. First, the output feedback controller is motivated to consider from the
fact that only position state is possible to measure in the EndoBots system. The
second control scheme is the state feedback controller with the velocity observer
based on the global linearization of the nonlinear robotic systems.
79
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 97/183
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 98/183
81
V becomes a negative semi-definite:
V = −eT K de. (5.11)
Lyapunov’s direct method leads to only stability of the equilibrium point. By in-
voking the Lasalle’s invariance principle, it can be shown that the largest invariant
set S is given by
S = {e, e : e = 0, e = 0}. (5.12)
Consequently e = q − q d, tends to zero asymptotically. The beauty of this approach
is that we can easily show the globally asymptotical stability of the the manipulator
under PD control with gravitational and frictional compensation.
• Control Gain Selection
The error dynamics are given by
M (q )e + C (q, q )e + K de + K pe = 0. (5.13)
It can be simplified with assumption of neglecting the Coriolis and centrifugal
terms and an approximation of inertia matrix, M (q ), by a constant diagonal
inertia matrix M due to the high gear ratio:
M e + K de + K pe = 0. (5.14)
Since the simplified error dynamics are linear differential equation, it is possible
to tune the PD controller gains to achieve a desired performance specification.
The gain matrix K p and K d are chosen such that the error dynamics becomes
critically damped. Let ωe be the desired natural frequency of error dynamics,
then for being critically damped error dynamics, this gives
K p = w2eM (5.15)
K d = 2
K pM. (5.16)
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 99/183
82
5.2.2 Experimental Evaluation of Joint Space Control
For experimental evaluation, we choose the closed loop natural frequency to
be 7 Hz and approximate the inertia term at the center of the workspace. For being
critically damped error dynamics, this gives
K p =
97 0 0 0
0 97 0 0
0 0 1.93 0
0 0 0 193
(5.17)
K d =
4.4 0 0 0
0 4.4 0 00 0 0.088 0
0 0 0 8.8
. (5.18)
• Friction Compensation
To compensate the friction effect, the estimated amount of Coulomb friction
is added into the input torque as follows
τ f = F csgn(q ), (5.19)
where F c is estimate for F c.
Since the joint velocities are not too high, the amount of compensated viscous
friction is small enough to be neglected. In Figure 5.1, the solid line shows the
friction approximation for compensation by only the Coulomb friction while
dotted line the real friction. Figure 5.2 shows the experimental results on fric-
tion compensation in the first two joints. Figure 5.2 (a) and (b) show that the
friction leads to almost no motion in moments of changing of velocity direction
without friction compensation. As seen in Figure 5.2 (a) and (b), after some
time, the control output from the controller is big enough to overcome the
friction and starts to catch up the desired motion. Figure 5.2 (c) shows the
improved performance with the same reference trajectory with friction com-
pensation as proposed above with first joint. Figure 5.2 (d) shows the same
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 100/183
83
Figure 5.1: Friciton approximation for compensation.
0 2 4 6 8−15
−10
−5
0
5
10
15(a)
time
p o s i t i o n [ d e g ]
0 2 4 6 8−15
−10
−5
0
5
10
15(b)
time
p o s i t i o n [ d e g ]
0 2 4 6 8−15
−10
−5
0
5
10
15(c)
time
p o s i t i o n [ d e g ]
0 2 4 6 8−15
−10
−5
0
5
10
15(d)
time
p o s i t i o n [ d e g ]
Figure 5.2: Friction compensation.(a) First joint position without friction compensation(b) Second joint position without friction compensation(c) First joint position with friction compensation(d) Second joint position with friction compensation
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 101/183
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 102/183
85
performed with three difference angular velocity of ω1 = 0.1Hz(0.62radsec ), ω2 =
0.25Hz(1.57radsec ), and ω3 = 0.5Hz(3.14 rad
sec ) with radius r1 = 2.5mm, r2 =
5mm, and r3 = 10mm. Gravity and Friction compensation were included in
all experiments.
−15 −10 −5 0 5 10 15−15
−10
−5
0
5
10
15
x [mm]
y [ m m ]
Circle Tracking
Figure 5.4: Experimental results of circle tracking.
Table 5.2 quantitatively summarizes the tracking performance of joint spacecontroller. As shown in the table, higher angular velocity results in large
tracking error.
Table 5.2: Circle tracking error.
Angular r1 r2 r3Velocity (mm) (mm) (mm)
ω1 0.409 0.463 0.487ω2 0.419 0.551 0.588
ω3 0.445 0.544 0.757
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 103/183
86
5.2.3 Nonlinear Decoupled State Feedback Controller
The motivation of this approach is to globally linearize the nonlinear robot
dynamics with suitable choice of the input. This approach is attractive in the sense
of availability of rich linear controller design schemes with linear system such as theoptimal or robust controller.
5.2.3.1 Global Linearization of the Nonlinear Robotic System
Let the rigid robot dynamics be given in general compact matrix form by
M (q )q + C (q, q )q + G(q ) + N (q ) = τ . (5.20)
The idea is to cancel the nonlinearities by choosing control input as
τ = M (q )u + C (q, q )q + G(q ) + N (q ), (5.21)
where M (q ), C (q, q ), G(q ), and N (q ) denote the approximated matrix of real robot
dynamics. Substituting (5.21) into (5.20) gives
q = u + (M (q )−1 M (q ) − I )
η1(q)
u + (M (q )−1∆H (q, q )
η2(q,q)
, (5.22)
where ∆H (q, q ) = ( C (q, q ) − C (q, q )) + ( G(q ) − G(q )) + ( N (q ) − N (q )).
Then it becomes
q = u + η1(q )u + η2(q, q ), (5.23)
where η(u,q, q ) = η1(q )u + η2(q, q ) represents the disturbance of the linearized robot
dynamics due to the lumped model uncertainties. The state equation can be given
by defining a new set of the state variables to be
x =
q
q
. (5.24)
Then
x = Ax + B(u + η1(x)u) + Bη2(x), (5.25)
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 104/183
87
where
A =
0 I
0 0
B =
0
I
. (5.26)
With assumption of neglecting the uncertainty term η(u,q, q ), we have
x = Ax + Bu. (5.27)
Clearly, this is globally linearized system. Hence, it can be concluded that the
manipulator can be regarded as a set of fully decoupled second-order time-invariant
linear systems. In practice, it is not possible to have the exact matrices of robot
dynamics, and therefore the linear controller based on this approximated linear
system, η(u,q, q ) ≈ 0, needs to have a robustness against the inevitable uncertaintiesin terms of modeling and parameters.
5.2.3.2 Optimal State Feedback Control
The objective of optimal control is to design the control law that minimizes
a cost function while satisfying the equality constraints imposed by the differential
equation of system dynamics:
minu(t) J (x(t), u(t)) = φ(x(tf )) + tf t0
L(x,u,t)dt subject to x(t) = f (x,u,t).
(5.28)
It is, in general, impossible to solve analytically so that we can use the explicit
control law. In the case of the linear time-invariant system with the quadratic cost
functions, the explicit control equation, which is basically the full state feedback
controller, can be obtained. Given the system dynamics:
x(t) = Ax(t) + Bu(t) x(t0) = x0, (5.29)
design a control law such that it minimizes the cost function:
J = 1
2x(tf )
T Sx(tf ) + 1
2
tf t0
(xT (t)Qx(t) + u(t)T Ru(t))dt, (5.30)
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 105/183
88
where S and Q are real symmetric positive semidefinite matrices, and R is a real
symmetric positive definite matrix. This is a constrained optimization problem and
it can be converted into an unconstrained optimization problem using Lagrange
multipliers. Then, the augmented cost function is given by
J a = 1
2x(tf )
T Sx(tf ) +
1
2
tf t0
(xT (t)Qx(t) + u(t)T Ru(t) + λ(t)T (Ax(t) + Bu(t) − x(t)))dt
= 1
2x(tf )
T Sx(tf ) + 1
2
tf t0
(H (x,u,t) = λ(t)T x(t))dt, (5.31)
where H (x,u,t) denotes the Hamiltonian function and is defined
H (x,u,t) = L + λT f = 12
(xT (t)Qx(t) + uT (t)Ru(t) + λT (t)(Ax(t) + Bu(t))). (5.32)
Using the theory of the calculus of variations, we have the Euler-Lagrange equations
for the states, the costates, and the optimal control inputs as
state equation : x(t) = ∂H
∂λ = Ax(t) + Bu(t) (5.33)
costate equation : λ(t) = −∂H
∂x = −Qx(t) − AT λ(t) (5.34)
control equation : ∂H ∂u
= 0 ⇐⇒ u(t) = −R−1BT λ(t) (5.35)
with the following the boundary conditions at t = t0 and t = tf :
t = t0 : x(t0) = x0 (5.36)
t = tf : λ(tf ) = ∂φ
∂x(tf ) = Sx(tf ). (5.37)
Optimal control problem now becomes a TPBVP (Two Point Boundary Value Prob-
lem), which has n boundary conditions at the initial time t0 and n conditions at
the final time tf . The state and costate equations can be given by the Hamiltonian
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 106/183
89
matrix H :
x(t)
λ(t)
=
A −BR−1BT
−Q
−AT
x(t)
λ(t)
≡ H
x(t)
λ(t)
. (5.38)
It is a homogeneous linear differential equation and the solution can be written by
x(t)
λ(t)
= eHt
x(t0)
λ(t0)
, (5.39)
where λ(t0) is not given. It can be solved, in general, numerically by iterating the
boundary condition, called a shooting method . It basically sequences the state and
costate equations in order to get ˆλ(tf ) with the guessed initial values of λ(t0) and
perturbs λ(t0) until the boundary conditions at the final time are satisfied to the
desired accuracy, λ(tf ) = λ(tf ). In case of LTI system with the quadratic cost
function, the problem can be solved analytically using the sweep method proposed
by Bryson and Ho [61]. The idea of the sweep method is the assumption of a linear
relationship between the costate and the state, which can be given by
λ(t) = P (t)x(t). (5.40)
This transforms the TPBVP with respect to x(t) and λ(t) into a single point bound-
ary problem in P (t) as
− P (t) = AT P (t) + P (t)A + Q − P (t)BR−1BT P (t) (5.41)
with the boundary condition P (tf ) = S . This is called the time varying Riccati
equation. It means that the solution to the Riccati equation gives rise to the optimal
full state feedback control law:
u(t) = −K (t)x(t) = −R−1BT P (t)x(t), (5.42)
which minimizes the cost function while satisfying the equality constraints. This
optimal controller is called a linear quadratic regulator (LQR). Note that even for
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 107/183
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 108/183
91
5.2.3.3 Velocity Estimation
Friction and velocity estimation are two main factors, which can affect the
performance in motion control applications. The friction can be identified and
compensated as mentioned before. In motion control applications, velocity can bemeasured directly with velocity sensors at the expense of high cost. In general, only
position signals are available and it is true in our research. In order to implement
the optimal state feedback controller, the state corresponding to the velocity should
be estimated.
Numerical Differentiating The easiest way to get the velocity information is to
numerically differentiate the position signals. Inherently numerical differenti-
ating is very sensitive to the noise. From the filter perspective, the numerical
differentiating has the following filter characteristics:
F nd(s) = 1 − e−sT S
T s, (5.48)
where T s represents the sampling time. It is an irrational transfer function and
can be approximated by creating a series expansion. With the second order
expansion, it becomes
F nd(s) ≈ 1 − (1 − sT s + 12
s2T 2s ) = sT s(1 − 12
sT s). (5.49)
Clearly, it is an improper function and the Bode plot corresponding to the
second order approximated filter with T s = 0.001 is shown in Figure 5.5. As
can be seed in Figure 5.5, it amplifies the signal in high frequency band where
noise signal mostly lies in. In order to get around this problem, the low-pass
filter can be used with the numerically differentiated signals.
Washout Filter The alternative is to combine the differentiating and smoothing
into one filter, called a washout filter . The washout filter can be represented
as
F wo(s) = ss p
+ 1, (5.50)
where p determines the location of pole and in general can be located 2 ∼ 3
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 109/183
92
Bode Diagram
Frequency (rad/sec)
P h a s e ( d e g )
M a g n i t u d e ( d B )
−20
−10
0
10
20
30
40
102
103
104
0
45
90
Figure 5.5: Bode plot for the numerical differentiating filter.
times faster than the fastest system pole. It is a proper function and the Bode
plot corresponding to the washout filter is shown in Figure 5.6. It is basically
a high-pass filter with a slope of 20 dB per decade to corner frequency and a
unit amplifying gain over the pass band.
Observer Design Once the system dynamics is given, the most effective method
to estimate the states can be the dynamic state observer. When noise is
present, the system can be represented as
x(t) = Ax(t) + Bu(t) + w(t) (5.51)
y(t) = Cx(t) + v(t), (5.52)
where w(t) represents the process noise and v(t) represents the sensor(measurement)
noise. The observer is constructed as follows
˙x(t) = Ax(t) + Bu(t) + L(y(t) − C x(t)), (5.53)
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 110/183
93
Bode Diagram
Frequency (rad/sec)
P h a s e ( d e g )
M a g n i t u d e ( d B )
36
36.5
37
37.5
38
38.5
39
39.5
40
102
103
104
0
30
60
Figure 5.6: Bode plot for the washout filter.
where x(t) is the estimate of x(t). The observer poles can be chosen arbitrarily
faster enough such that the controller poles can dominate the system dynamics.
Too fast poles can amplify the measurement noise signals. As a rule of thumb,
2 ∼ 3 times faster poles than the closed system poles can be used. In an
optimal manner, the observer can be designed so that it has an optimal balance
of dependency on the model or the measurements.
The Kalman Filter Design The Kalman filter gain L can be obtained such that
it minimize the mean square error:
J = E [(x(t) − x(t))(x(t) − x(t))T ]. (5.54)
Let Qo and Ro be the covariance matrices of the process and measurement
noise:
Qo = E [w(t)w(t)T ],
Ro = E [v(t)v(t)T ].
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 111/183
94
Then, the steady-state optimal gain is computed from
L = ΣC T R−1o , (5.55)
where Σ represents the intensity of the estimation error and it is the solution
of the following the ARE:
AΣ + ΣAT + Qo − ΣC T R−1o C Σ = 0. (5.56)
This result is the dual of that of the LQR design.
5.2.3.4 Linear Quadratic Gaussian (LQG) Control
The difference between LQR and LQG can be represented as
LQR : u(t) = K LQRx(t)
LQG : u(t) = K LQGx(t) = K LQG(f (y(t))).
The LQR has guaranteed stability margins, but a major limitation of the LQR is
to require the measurement of all states. In contrary to the LQR, the LQG control
provides the optimal control with partially available states. From the separation
principle, the LQG optimal controller can be obtained:
T he Robot Dynamics : τ = M (q )q + C (q, q )q + G(q ) + N (q )
T he F eedback Linearization : τ = M (q )u + C (q, q )q + G(q ) + N (q )
T he Linearized Plant :
x(t) = Ax(t) + Bu(t) + w(t)
y(t) = Cx(t) + v(t)
T he Optimal Controller : u(t) = K x(t)
K = −R−1BT P
0 = AT P + P A + Q − P BR−1BT P
T he Optimal Observer :
˙x(t) = Ax(t) + Bu(t) + L(y(t) − C x(t))
L = ΣC T R−1o
0 = AΣ + ΣAT + Qo − ΣC T R−1o C Σ.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 112/183
95
The LQG controller is basically a regulator, which has a zero command input. In
order to track a reference input, the state error feedback controller is considered via
coordinate translation in this work. Define the error vector as
x(t) = x(t) − xd, (5.57)
where xd is the desired state and xd = xd = 0. Then,
˙x(t) = Ax(t) + Bu(t) (5.58)
y(t) = C x(t). (5.59)
State feedback controller in the second-order system can be viewed as a generaliza-
tion of proportional-derivative control. To eliminate the steady-state error, we can
augment the system by taking the integral of the error as an additional state:
xi(t) =
(y(t) − r)dt =
(Cx(t) − r)dt =
C x(t)dt. (5.60)
The augmented state equation is then,
˙x(t)
xi(t) = A 0
C 0 x(t)
xi(t) + B
0 u(t).
It can be written with augmented error state denoted by adding a bar:
˙x(t) = Ax(t) + Bu(t) (5.61)
y(t) = C x(t). (5.62)
Figure 5.7 shows the block diagram of the LQG controller with augmented error
state. Figures 5.8–5.11 show the experimental results of the LQG controller on the
first joint. Figure 5.8 compares the tracking performance of LQG controller when
the friction compensation is active and not active. As seen in the figure, friction
leads to no motion in moments of changing of velocity direction without the friction
compensation and the friction compensating improves performance. Figure 5.9 (b)
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 113/183
96
Figure 5.7: Observer based linear controller with feedback linearization.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 114/183
97
compares the measured position error state and the estimated position error state
and (c) shows the estimated velocity error state and the corresponding control input
signal is shown in (d). Figure 5.10 compares the measured position error states
according to different weighting factors on the position error state. Figure 5.11shows the corresponding control input. As seen in Figures 5.10–5.11, the error state
is decreased and the control input is increased by increasing the weighting factor on
q 1 as expected. .
0 5 10 15−40
−30
−20
−10
0
10
20
30
40
time
p o s i t i o n [ d e g ]
desiredwithout friction comp.with friction comp.
Figure 5.8: Experimental results of friction compensation in the LQGcontroller.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 115/183
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 116/183
99
0 5 10 15−30
−20
−10
0
10
20
30
time
c o n t r o l i n p u t
q1=1q1=5q1=10
Figure 5.11: Comparison on control inputs with the different weighting.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 117/183
CHAPTER 6
SHARED CONTROL
Shared control, as the name implies, enables the human operator and the computer
control the manipulator in parallel. It means that the human operator controls some
axes while the computer concurrently controls other axes. This would be useful in
the following scenarios:
• Surgeons want to move the tool along the tool axis for drilling. The roll-pitch-
yaw rotation of the docking station would be computer controlled while the
surgeon manually controls the tool translational motion and operates the tool(for example, for drilling or cutting).
• The surgeon wants to control the tip of the tool along a straight line. For
example, the surgeon may want to perform precision cutting and stitching.
The computer would actively control the tool to stay in a “valley” along which
the surgeon is free to move the docking station and operate the tool manually.
The surgeon may specify the direction of manual control through a 3D input device.
Our current implementation allows the surgeon to manually operate one of the
EndoBot as the pointing device and use a foot pedal to register the selected points.
In the shared control case, the computer control algorithm needs to be modified to
ensure only the deviation from the specified path is corrected, but not the motion
along the path.
6.1 Constraints Description
In shared control mode, we add artificial constraints to relieve the operatorof some sub task that would be tiring for an operator to concentrate on such as
tool alignments. These artificial constraints are assumed to be holonomic. In or-
der to control robot systems with holonomic constraints, we have to represent the
constraints mathematically. Let the forward kinematics (mapping from the joint
coordinate, q = (q 1, q 2, q 3, q 4) to the end effector coordinate x = (rT , φ)) be denoted
100
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 118/183
101
by x = k(q ). Specifically, suppose that the surgeon specifies that the end effector
position rT and roll angle φ to be constrained as C (rT , φ) = 0. The constraints that
the controller needs to enforce are C (k(q )) = 0. Writing them more compactly, the
holonomic constraints can be represented on the task space variables x(t) ∈ n
asfollows
C (x) = 0. (6.1)
Let the gradient of C (x) be J c(x):
J c(x) = ∂C (x)
∂x . (6.2)
6.2 Control Objective
In shared control, the control task is to restrict the motion of manipulator
within the constraint space, which can be represented as mapping C (x) = 0. The
control objective here is to choose the controller to solve the following regulation
problem with constraints. Given dynamic equation and constraints, determine a
feedback control law, τ , so that the closed-loop system satisfies
C (x(t)) → 0 as t → ∞. (6.3)
It means that the motion will be lie in the following Constraint manifold:
S = {x, x : C (x) = 0, J c(x)x = 0}. (6.4)
A suitable stabilizing controller would be
V = −J T c K C (x), (6.5)
where K is a PID or lead/lag type of controller.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 119/183
102
6.3 Task Space Control Design
The dynamic model in task space becomes
M x(q )x + C x(q, q )x + Gx(q ) = J
−T
τ. (6.6)
Let xcd be the desired position of constrained variables in task space and xc be the
current position of constrained variables. We apply the following a PD feedback
control law with gravity compensation so that the constrained space error xc − xcd
tends asymptotically to zero:
τ = −J T J T c ( K p∆xc + K vJ c x ) + G(q ), (6.7)
where ∆xc = xc − xcd.
6.3.1 Stability
Choose the following positive definite quadratic form as a Lyapunov function
candidate:
V = 1
2 q T M (q )q +
1
2∆xc
T K p∆xc. (6.8)
Differentiating (6.8) with respect to time,
V = q T M (q )q + 1
2 q T M (q )q + ∆xT
c K p∆xc. (6.9)
Substituting M (q )q gives
V = q T (τ − C (q, q )q − G(q ) + 1
2M (q )q ) + ∆xT
c K p∆xc (6.10)
= 1
2 q T ( M (q ) − 2C (q, q ))q + q T (τ − G(q )) + ∆xT
c K p∆xc (6.11)
= q T
(τ − G(q )) + ∆xT c K p∆xc. (6.12)
Since
τ = J T F = J T J T c F c (6.13)
∆xc = xc − xcd (6.14)
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 120/183
103
= C (x) − xcd (6.15)
∆xc = J c x, (6.16)
then,
V = q T J T J T c F c + (J c x)T K p∆xc − q T G(q ) (6.17)
= (J q )T J T c F c + xT J T c Kp∆xc − xT JG(q ) (6.18)
= xT (J T c (F c + Kp∆xc) − JG(q )). (6.19)
This equation suggests the structure of the controller. By choosing the control law
for constraint force:
F c = −K p∆xc − K vJ c x + J −T c JG(q ), (6.20)
where K p and K v are symmetric positive definite matrix.
V becomes
V = −xT J T c K vJ c x. (6.21)
The function V is only negative semi-definite, since
V = 0 for x = 0, ∀∆xc. (6.22)
Hence Lyapunov’s direct method leads to only stability of the equilibrium point. By
invoking the Lasalle’s invariance principle, it can be shown that the largest invariant
set S is given by
S = {x, x : x = 0, ∆xc = 0}. (6.23)
Consequently the constrained space error, ∆xc = xc − xcd, tends to zero asymptot-
ically. The feedback control law becomes
τ = J T J T c F c
= −J T J T c (K p∆xc + K vJ c x) + G(q ). (6.24)
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 121/183
104
6.3.2 Controller Description
The block diagram of the shared control algorithm is shown in Figure 6.1.
Basically, the constraint Jacobian, Jc, decomposes the Cartesian forces into con-
Figure 6.1: Block diagram of task space shared control.
strained space and unconstrained space. It is worth while pointing out that applying
the constrained torques from (6.24) is equivalent to adding virtual mechanisms with
passive springs and dampers normal to constrained space. Suppose that we have a
constrained line parallel to z axis, and then applying shared control has same effecton adding virtual springs and dampers lying on the x-y plane. From this point of
view, intuitively the closed-loop system with control law of (6.24) is always stable
since the robotic system is passive.
These virtual springs and dampers would be very useful for actively guiding
the surgeon. When the surgeon is trying to pull out the suture, the constrained path
will be helpful to guide the surgeon to indicate preferred direction with adjustable
impedance.
6.3.3 Example
To illustrate the procedures of shared control, we will consider two simple
examples of constrained equations and show how to get the constraint Jacobian,
J c(x).
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 122/183
105
• Move along the Straight Line
Supposed that the end-effector is required to stay along a specific line. In this
case, the surgeon can control the roll as well as translation. Let p(x) ∈ 4
represent the tip of the tool and x axis be the preferred line. Let n be thenumber of task space variables and m be number of constraints. Then, the
constraint space can be described as the set of all points satisfying the following
constraint equations:
x = 0, y = 0. (6.25)
These constraints become
C (x) =
x
y
= 0. (6.26)
The constraint Jacobian becomes m × n matrix as follows
J c(x) = ∂C (x)
∂x =
1 0 0 0
0 1 0 0
. (6.27)
6.3.4 Experimental Evaluation of Task Space Shared Controller
This section presents the experimental results that illustrate the performancecharacteristics of shared controller described in the previous section. For the purpose
of evaluating the shared control algorithm experimentally, a straight line through
the center of the spherical joint was selected as a constrained path. The straight
line can be parameterized asx
a1=
y
a2=
z
a3. (6.28)
The constraints can be represented by
C 1(xT ) = x
a1− y
a2= 0 (6.29)
C 2(xT ) = x
a1
− z
a3= 0. (6.30)
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 123/183
106
By combining, we obtain x
a1− y
a2
xa1
− za3
= 0. (6.31)
The constraint Jacobian becomes 2 × 4 matrix:
J C (xT ) =
1
a1− 1
a20 0
1a1
0 − 1a3
0
. (6.32)
Table 6.1 shows the parameters used to specify the constraint line.
Table 6.1: Parameters for constrained line.
a1 a2 a3
1 1 -2
00.05
0.10.15
0.2
0
0.05
0.1
0.15
0.2
−0.2
−0.18
−0.16
−0.14
−0.12
−0.1
−0.08
−0.06
−0.04
−0.02
0
x [m]
y [m]
z [ m ]
Desired trajectory
Figure 6.2: Desired constrained path.
Figure 6.2 shows the desired path, which can be obtained with the above
constraint Jacobian. The surgeon can slide up and down the tip of tool along the
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 124/183
107
desired trajectory and roll the tool freely. In shared control, the operator can be
interpreted as an actuator in the unconstrained space, but also can be a disturbance
source in the constrained space. It should be noted that the performance of shared
controller depends on the human operator. For this reason, the operator was tryingto move the tool in all direction with same impedance for performance evaluation.
Figure 6.3 shows the measured task space position during the motion with shared
0 2 4 6 80
0.02
0.04
0.06
0.08
0.1
time
x [ m ]
0 2 4 6 80
0.02
0.04
0.06
0.08
0.1
time
y [ m ]
0 2 4 6 8−0.2
−0.15
−0.1
−0.05
0
time
z [ m ]
0 2 4 6 8−1.5
−1
−0.5
0
0.5
1
1.5
time
r [ r a d ]
Figure 6.3: Measured task space positions.
controller. As can be seen, the resulting motion was completely satisfactory. For
sliding up and down and rolling, the operator can easily move the robot.
Figure 6.4 shows the constraint force from the shared controller and Figure 6.5
shows the constraint forces mapped into range space R(J T c ). The range space R(J T c )
represents the set of all possible task space forces that make the constraints. Onthe other hand, it represents the force exerted from virtual springs and dampers.
Figure 6.6 shows the corresponding joint torques. The torques in first, second and
fourth joint keep the tip of tool along the constrained path.
Figure 6.7 demonstrates the performance of shared controller in Cartesian
space. The dotted line shows the desired constrained path and the solid line rep-
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 125/183
108
0 1 2 3 4 5 6 7 8 9−10
−5
0
5
10
time
F
c 1 [ N ]
0 1 2 3 4 5 6 7 8 9−10
−5
0
5
10
time
F c 2 [ N ]
Figure 6.4: Constraint forces.
0 2 4 6 8−10
−5
0
5
10
time
J c ‘ F c 1 [
N ]
0 2 4 6 8−10
−5
0
5
10
time
J c ‘ F c 2 [
N ]
0 2 4 6 8−10
−5
0
5
10
time
J c ‘ F c 3 [ N ]
0 2 4 6 8−10
−5
0
5
10
time
J c ‘ F c 4 [ N ]
Figure 6.5: Constraint forces projected onto range space R(J T c ).
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 126/183
109
0 2 4 6 8−2
−1
0
1
2
time
t a u 1 [ N m ]
0 2 4 6 8−2
−1
0
1
2
time
t a u 2 [ N m ]
0 2 4 6 8−2
−1
0
1
2
time
t a u 3 [ N m ]
0 2 4 6 8−2
−1
0
1
2
time
t a u 4 [ N m ]
Figure 6.6: Measured joint torques.
resents the measured task space positions from the shared controller. The shared
control algorithm brings the tip of tool onto the desired trajectory.
Table 6.2 summarizes the performance of shared control with respect to de-
sired stiffness and actual stiffness getting from maximum deviation and maximumconstraint force in constrained space without considering the dynamic effects.
Table 6.2: Desired and measured stiffness for shared control.
Max(|e|) Max(|F c|) Measured Stiffness, K a Desired Stiffness, K d[m] [N] [N/m] [N/m]
ec1 0.0047 4.680 1003 1000ec2 0.0041 4.097 1002 1000
6.4 Joint Space Shared Control Design
Let the motion constraints in the task coordinate xT be specified as
xc = c(xT ) = 0, (6.33)
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 127/183
110
00.05
0.1
0.150.2
0
0.05
0.1
0.15
0.2
−0.2
−0.18
−0.16
−0.14
−0.12
−0.1
−0.08
−0.06
−0.04
−0.02
0
x [m]y [m]
z [ m ]
Actual trajectory
Desired trajectory
Figure 6.7: Experimental result of task space shared control with con-strained line.
where c is a constraint function.
Let f (·) be the complement of c so that
d(xT ) = c(xT )
f (xT )
(6.34)
is a diffeomorphism (i.e., d is one-to-one and onto, and d and d−1 are both con-
tinuously differentiable). The basic concept of shared control implemented in this
section is that it is suitable to choose the generalized coordinate vector in order
to describe the constrained motions. Due to the presence of k constraints and the
resulting lack of k degree of freedom, the description of constrained motions in
Cartesian coordinates is not efficient. Clearly, the mapping d(xT ) transforms the
position vector expressed in Cartesian coordinates into generalized coordinate space,
which marches on the constraint surface. With this generalized space, the motions
can be easily decomposed into one in constrained space and the other in free space.
The corresponding vector in Cartesian coordinates can be obtained with the inverse
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 128/183
111
mapping as shown in Figure 6.8 . The assumption of diffeomorphism ensures the
bijection mapping of both position and velocity vectors. If we use the same joint
Figure 6.8: Concept of coordinates mapping.
space controller as described earlier, shared control can be achieved by removing
the free motion component in the desired task space trajectories. Specifically, let
q be the measured joint positions. The desired free motions should be the same as
the actual free motions, therefore, f (xT d) = f (k(q )). The constraint motions are
required to be zero, therefore, c(xT d) = 0. The desired task space set points canthen be obtained from
xT d = d−1
0
f (k(q ))
, (6.35)
where d is from (6.34).
The desired task space velocities can be found similarly:
˙xT d =
J c
J f
−1
0
J f J q
, (6.36)
where J c = ∂c∂x
and J f = ∂f ∂x
.
Once xT d is found, the same joint space controller can be used:
τ = −K p(q − k−1(xT d) − K d(q − J −1(q ) xT d) + G(q ) + N (q ). (6.37)
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 129/183
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 130/183
113
6.4.2 Examples
• Move along the Straight Line
Let the straight line be parameterized as
x
a1=
y
a2=
z
a3. (6.38)
Then the constraint function is
c(xT ) =
1
a1x − 1
a2y = 0
1a1
x − 1a3
z = 0
=
1
a1− 1
a20 0
1a1
0 − 1a3
0
xT . (6.39)
The free motion function, f , can be chosen to be orthogonal to c. In this case,
we may choose f to be
f (xT ) =
a1 a2 a3 0
0 0 0 1
xT . (6.40)
The Jacobian for constrained motion and free motion are
J c =
1a1
− 1a2
0 0
1
a10
−1
a30
(6.41)
J f =
a1 a2 a3 0
0 0 0 1
. (6.42)
• Move along the Circle
Suppose that the tip of tool is needed to stay along a circle, which lies on x-y
plan. Let the radius of this circle be r, and the center of the circle at the point
(x0, y0, z 0). The constraint equations can be represented
c1(xT ) = (x − x0)2 + (y − y0)2 − r2 (6.43)
c2(xT ) = z − z 0. (6.44)
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 131/183
114
Then, the constraint function is
c(xT ) =
(x − x0)2 + (y − y0)2 − r2
z
−z 0
= 0. (6.45)
The constraint Jacobian is now
J c =
2(x − x0) 2(y − y0) 0 0
0 0 1 0
. (6.46)
Let θ denote the angle between the x axis and the vector r:
x = r cos(θ) (6.47)
y = r sin(θ). (6.48)
The free motion can be described by
f 1(xT ) = θ = tan−1(y
x). (6.49)
Let the tip of tool be free to rotate about an axis parallel to the tool axis, then
f 2(xT ) = φ. (6.50)
J f becomes
J f ==
−y
rxr 0 0
0 0 0 1
. (6.51)
6.4.3 Experimental Evaluation of Joint Space Shared Controller
• Linear Trajectory For the purpose of evaluating the joint space shared con-
troller experimentally and comparing with the task space controller, same
straight line used in evaluating the task space shared control was selected.
Figure 6.10 shows the measured task space positions. The tip of tool is sliding
down and up and the position on x and y axis is increased and decreased cor-
responding to position of z axis. Figure 6.11 shows the desired and measured
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 132/183
115
0 2 4 6 80
0.02
0.04
0.06
0.08
0.1
time
x [ m ]
0 2 4 6 80
0.02
0.04
0.06
0.08
0.1
time
y [ m ]
0 2 4 6 8−0.2
−0.15
−0.1
−0.05
0
time
z [ m ]
0 2 4 6 80
0.2
0.4
0.6
0.8
1
time
r [ r a d ]
Figure 6.10: Measured task space positions.
positions in joint space and Figure 6.12 shows the corresponding controller
outputs.
Figure 6.13 demonstrates the performance of shared controller. The dotted line
shows the desired constrained path and the solid line represents the measured
task space position from the shared controller. The constraint force in the
direction normal to the free motion can be obtained as follows:
τ = J T J T c F c, (6.52)
where F c is the constraint force in task space:
F c = (J T c )+J −T τ. (6.53)
Then, the effective stiffness in constrained space can result in
K c = max |F c| /max |xc| . (6.54)
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 133/183
116
0 2 4 6 8−0.5
0
0.5
time
j o i n t 1 [ r a d ]
qdqqe
0 2 4 6 8−0.5
0
0.5
time
j o i n t 2 [ r a d ]
qdqqe
0 2 4 6 8−1
−0.5
0
0.5
1
time
j o i n t 3 [ r a d ]
qdqqe
0 2 4 6 8−0.5
0
0.5
time
j o i n t 4 [ r a d ]
qdqqe
Figure 6.11: Desired and measured joint position
0 2 4 6 8−2
−1
0
1
2
time
t a u 1 [ N m ]
0 2 4 6 8−2
−1
0
1
2
time
t a u 2 [ N m ]
0 2 4 6 8−2
−1
0
1
2
time
t
a u 3 [ N m ]
0 2 4 6 8−2
−1
0
1
2
time
t
a u 4 [ N m ]
Figure 6.12: Measured joint torque
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 134/183
117
00.05
0.1
0.150.2
0
0.05
0.1
0.15
0.2
−0.2
−0.18
−0.16
−0.14
−0.12
−0.1
−0.08
−0.06
−0.04
−0.02
0
x [m]y [m]
z [ m ]
Desired trajectory
Actual trajectory
Figure 6.13: Actual path with joint space shared controller
Table 6.3 summarizes the performance of joint space shared controller with
the linear constrained motion.
Table 6.3: Effective constraint stiffness for linear trajectory
Max(|e|) Max(|F c|) Effective Stiffness, K e[m] [N] [N/m]
ec1 0.0013 8.1838 6268ec2 0.00062 9.763 15570
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 135/183
118
• Circular Trajectory Figure 6.14 shows the desired circle trajectory, which is
centered at the (0, 0, -0.15 m) with radius of 0.05m. The experimental data
of the joint shared controller with the circular trajectory are illustrated in
Figures 6.15–6.18. Table 6.4 summarize the performance of joint space sharedcontroller with the circular motion.
00.05
0.10.15
0.2
0
0.05
0.1
0.15
0.2
−0.2
−0.18
−0.16
−0.14
−0.12
−0.1
−0.08
−0.06
−0.04
−0.02
0
y [m]x [m]
z [ m ]
Desired trajectory
Figure 6.14: Desired circle trajectory
Table 6.4 summarizes the performance of joint space shared control with linear
motion.
Table 6.4: Effective constraint stiffness for circular trajectory
Max(|e|) Max(|F c|) Effective Stiffness, K e
[m] [N] [N/m]ec1 0.0032 102.02 31400ec2 0.002 3.49 1770
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 136/183
119
0 2 4 6 8−0.1
−0.05
0
0.05
0.1
time
x [ m ]
0 2 4 6 8−0.1
−0.05
0
0.05
0.1
time
y [ m ]
0 2 4 6 8−0.2
−0.15
−0.1
−0.05
0
time
z [ m ]
0 2 4 6 80
0.2
0.4
0.6
0.8
1
time
r [ r a d ]
Figure 6.15: Measured task space position
0 2 4 6 8−0.5
0
0.5
time
j o i n t 1 [ r a d ]
qdqqe
0 2 4 6 8−0.5
0
0.5
time
j o i n t 2 [ r a d ]
qdqqe
0 2 4 6 8−1
−0.5
0
0.5
1
time
j o
i n t 3 [ r a d ]
qdqqe
0 2 4 6 8−0.5
0
0.5
time
j
o i n t 4 [ m ]
qdqqe
Figure 6.16: Desired and measured joint position
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 137/183
120
0 2 4 6 8−2
−1
0
1
2
time
t a u 1 [ N m ]
0 2 4 6 8−2
−1
0
1
2
time
t a u 2 [ N m ]
0 2 4 6 8−2
−1
0
1
2
time
t a u 3 [ N m ]
0 2 4 6 8−2
−1
0
1
2
time
t a u 4 [ N m ]
Figure 6.17: Measured joint torque
00.050.1
0.150.2
0
0.05
0.1
0.15
0.2
−0.2
−0.18
−0.16
−0.14
−0.12
−0.1
−0.08
−0.06
−0.04
−0.02
0
x [m]y [m]
z [ m ]
Desired trajectoryActual trajectory
Figure 6.18: Actual trajectory of the end-effector in the task space forthe circular constraint
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 138/183
CHAPTER 7
TENSION CONTROL
7.1 Securing a Knot
After the knot is formed, securing the knot can be performed by applying the
tension. For the square knot, two throws are required and each throw must be done
in opposite direction and with a different tension. For the first throw, the proper
tension should be applied to place a simple knot on the wound surface such that it
will not contribute on tissue strangulation and also it would have enough traction
force to remain the knot as it is laid down. Applying too much tension might cause
breaking out the tissue. After the second throw, applying the precise tension would
be necessary so as not to break the knot or loose the knot.
7.1.1 Principle of Direction of Securing a Square Knot
Consider the suturing line locating on the tissue plane in Figure 7.1. Let
the point pa and pb be the entry point and exit point of suture, respectively. For
simplicity without the loss of generality, the base coordinate frame B can be attached
to the point where the suturing line and the line connecting two suturing point pair.Define ha and hb to be the unit vectors heading point pa and pb in the frame B.
After the needle is drawn through the tissue, the loop end, which holds the needle,
is at the exit point pb and the post end is at pa. Define ple and p pe to represent the
position of the loop end and post end and pn to represent the knot position. Let
el = ple− pn ple− pn
and e p = ppe− pn ppe− pn
be the unit vectors of the directions along which the
tensions apply. In the current configuration of the Endobots, ple represents the tip
position of stitcher and p pe represents the tip position of grasper.
• Direction of the first throw of a square knot
For the first throw of the square knot, the stitcher holding the needle is forced to
move along the direction ha and the grasper holding the suture tail is required to
121
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 139/183
122
Figure 7.1: Suturing line.
move along the hb as shown in Figure 7.2:
en · ha = 1
et · hb = 1.
(7.1)
Figure 7.2: Direction of securing in the first throw.
• Direction of the second throw of a square knot
After the second throw, the direction of securing the knot should be opposite to the
first throw. The stitcher is forced to move along the direction hb and the motion of
grasper is required to be done in the direction ha as shown in Figure 7.3:
en · hb = 1
et·
ha = 1.(7.2)
These requirements can be written by
en(iT ) · hb = (−1)iT
et(iT ) · ha = (−1)iT ,(7.3)
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 140/183
123
Figure 7.3: Securing the knot.
where iT represents each throw, iT = 1, 2.
If these requirements are not satisfied, the desired square knot is not achieved. For
example, if the tension is applying in the opposite direction during the second throw
as
en(2T ) · hb = −1
et(2T ) · ha = −1,(7.4)
it yields the following unstable a two half hitch knot as shown in Figure 7.4.
Figure 7.4: An unstable two half hitch knot with reverse-directed tensionduring the second throw.
7.2 Tension Measurement
7.2.1 Basic Idea
Measuring a tension in the suture in minimally invasive surgery is very difficult
because of the geometric constraints on mounting sensor and the sterilizing problem.
Strain gauge transducers are the most widely used sensors in the applications of
force measurements. They can be mounted on the laparoscopic instruments and
measure the strains of the instruments according to the external forces as shown
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 141/183
124
in Figure 7.5. These sensors have high sensitivity, accuracy, and bandwidth with
simple electronic circuit. With these sensors, the tension can be measured directly.
The main drawback of using the strain gauge sensors in minimally invasive surgery is
that they are very sensitive to temperature variation. The contact with the humansoft tissues during the surgery may cause drift on the outputs. Another problem is
that the surgical instruments go through in high temperature process in order to be
sterilized for the repeated uses. It forces to have a disposal type strain gauge sensor.
Calibrating with the each tool should be considered.
Figure 7.5: Strain gauge transducer.
Vision sensors can be used with the pre-modeled and pre-marked sutures. If
the sutures have the marks at predetermined spacing, the tension can be estimated
by measuring the elongation between two marks as shown in Figure 7.6. The diffi-
culty with this method is that the mark may not be visible due to the contamination
of blood or the occlusion of the sutures for the instruments. It is also difficult to
have a high bandwidth and resolution.
The idea to measure the tension in this research is to put a force/torque sensor
in the base of the manipulator, called a base sensor, as shown in Figure 7.7. It has
two main benefits such as
• The sensor does not need to be sterilized,
• Force measurement (and hence control) is independent of the tools.
Beside with these benefits, we can use the sensor information to compensate the
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 142/183
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 143/183
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 144/183
127
Transformation of the wrench between two coordinate frames can be achieved with
the adjoint transformation, Φ : 6 → 6,
Φ = R 0
ˆ pR R . (7.8)
Consider B is an inertial frame and E is a frame attached to the rigid body. Let
F e represent an applied wrench at the origin of E with respect to the E coordinate
frame. Then, the wrench in B’s coordinate frame is given by
F b = ΦT ebF e.
Thus, the wrench in the end-effector frame, F e, is calculated from the measured base
wrench:
F e = ΦT beF b.
It can be expanded as
F e =
Reb −Rebˆ pbe
0 Reb
F b. (7.9)
As a result, the corresponding joint torque can be obtained as
τ = J T F e = J T ΦT beF b = (J b)
T F b, (7.10)
where J b = ΦbeJ is the base Jacobian matrix.
7.3 Force and Position Control
7.3.1 Problem Formulation
As mentioned before, the tension applied to the end of suture should be reg-
ulated about the desired value and the end-effector of robot is forced to lie on the
straight line. In order to maintain the direction we can impose artificial constraints
so that the motion can be only occurred along the constraint line. The control ob-
jective here is to design a controller to solve the following problem.
Given dynamic model for robot manipulator, environment model and
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 145/183
128
f des, C (x) = 0 with initial condition of f (0) = 0, q (0), q (0) = 0, determine
a feedback control law, τ , so that the closed-loop system satisfies
f (t)→
f des as t→ ∞
C (x(t)) → 0 ∀ t
where f des is the desired tension, C (x) = 0 represents the holonomic constraint on
the task space variables and defines the desired direction along which the tension
would be applied. Clearly, the tension control problem during securing the knot can
be turned into force and motion regulation problem, which will be reviewed in the
following section.
7.3.2 Hybrid Force/Position Regulating Control
Since artificial constraints are imposed, securing the knot can be considered as
a geometrically constrained problem. When the whole information on environment
geometry is available, an effective strategy to embed the capability of position and
force control would be a hybrid force/motion control scheme, also referred to as a
hybrid control. The basic concept is the partition of the task space into direction of
motion and force based on orthogonality. To this goal, the overall control torques
can be split into two components:
τ = τ m + τ f (7.11)
where τ m is the torque vector applied to motion control and τ f is the torque vector
for force control. The hybrid controller, which will be addressed here, is similar in
many respect to the dynamic hybrid control approach. Consider a simple problem
of the three-link planer manipulator in Figure 7.9. The constraint can be expressedby C (x) = 0, where x ∈ 2 is the generalized position vector in Cartesian space. To
express the end-effector position on the constraint line, a free motion function, f (x),
can be chosen such that c(x) and f (x) can be mutually independent. Then d(x) can
be defined by d(x) = [c(x)T f (x)T ]T and it represents the end-effector position on the
constraint hypersurfaces. Let J c(x) = ∂c(x)/∂x be the Jacobian of the constraint
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 146/183
129
Figure 7.9: Constrained motion.
equation, and J f (x) = ∂f (x)/∂x be the Jacobian of the free motion equation. In
classical dynamic hybrid control literatures, the objective is to design a feedbackcontroller such that it can move the end-effector along the line, J f , while applying
a constraint force in the normal direction to the constraint, J c. In other words, the
Jacobian J c and J f represent the force and position control directions, respectively.
In knot securing case, the objective is to find a feedback control law so that
it can regulate the force along the direction of J f while constraining the motion
in the direction of J c. It can be considered as a problem with planar manipulator
moving in a slot with attached spring. The objective is to regulate the spring force
while applying no force against the side of the slot. The constraint equation can be
Figure 7.10: Constrained motion for securing the knot.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 147/183
130
written in joint space as
Φ(q ) = C (k(q )) = 0 Φ(q ) = [Φ1(q ),..., Φm(q )]T , (7.12)
where k(q ) represents the forward kinematics. Then,
J Φ(q ) = ∂ Φ(q )
∂q =
∂c(x)
∂x
∂k(q )
∂q = JcJ J Φ(q ) ∈ m×n. (7.13)
Since there is no applied force against the side of the virtual slot, the required force
in the Cartesian space to generate the desired tension can be written as
F = J T f f f F
∈ n, (7.14)
where f f ∈ n−m is the force generating tension in the force controlled subspace.
Then, the joint torque τ f generating the force, f f , along the J f direction is given by
τ f = J T F = J T J T f f f = J T Φf f . (7.15)
The force f f can be obtained from the measured generalized force F e:
f f = [0 I n−m] J c
J f
−1
F e. (7.16)
Then, the overall feedback control law is given by
τ = τ m + τ f (7.17)
τ m = M (q )um + G(q ) (7.18)
um = −K p(q − k−1(d−1( 0
f (k(q ))
))) − K d(q − J −1 J c
J f
−1 0
J f
J q ) (7.19)
τ f = J T J T c
0
uf
. (7.20)
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 148/183
131
The position regulation to satisfy the constraint is already verified in shared control
section, therefore hereafter only force regulation problem designing the force, uf , will
be considered. The existing force control scheme can be grouped into two categories.
The first category is implicit force control, which achieves desired compliant behaviorthrough position error based control and is suitable to avoid excessive force buildup.
Compliance control and impedance control are belonged to this category. The second
category is explicit force control control. As the name implies, explicit force control
has an explicit closure of a force feedback loop such that the output force can be
regulated. Since the tension should be regulated about the desired value and the
measurement of tension is available, the explicit force control is suitable.
7.3.3 Explicit Force Control
The general explicit force control scheme used here is
uf = K f f r + K p(f r − f m) + K d( f r − f m) + K i
(f r − f m)dt − K v xm, (7.21)
where f r is the reference force, f m is the measured force, K f is the feedforward force
gain, K p is the proportional force gain, K d is the derivative force gain, K i is the
integral force gain, K v is the velocity gain for the active damping, and xm is the
measured velocity.
In order to achieve the compliant transition during a contact period, damping
needs to be added to the systems. With damping, systems will be more stable and
can reduce the spike of force response. The derivative force gain can be used to
obtain damping with incorporating of numerical differentiating the force signals.
However, due to presence of noise in the force signals, implementing of derivative
force control essentially is not feasible. Filtered derivative control may be used with
incorporating low-pass filter, which can reduce the noise at the expense of phase
lag. An alternative is to actively add the damping to the systems. Active damping
method is widely used to deal with impact problem and it is much considerably
effective for soft environment same as a suture model. For soft environment, the
velocity trajectory is considerably smoother than for stiff one. For this reason, active
damping component is added. In addition, to ensure steady state performance and
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 149/183
132
robustness, it is advisable to endow the controller with an integral action. The
feedforward component usually needs to be added for the case of time-varying desired
force tracking problem, but in this study it is not considered. Finally, the explicit
force control law is given by
uf = K p(f r − f m) proportional
force
feedback
+ K i
(f r − f m)dt integral
force
feedback
− K v xm. active
damping
forward
(7.22)
7.4 Stability Analysis
7.4.1 Proportional plus Integral Control with Active Damping
Figure 7.11: Simplified second order environment model.
In this research, the suture is modeled as a second order system as shown in
Figure 7.11. F is the control input and m, b, and k represent the modeled environ-
mental mass, damping, and stiffness. The interacting force f m can be assumed to
be measured as a position dependent linear relation:
f m = kx. (7.23)
The environmental dynamics can be described as
mx(t) + bx(t) + kx(t) = F (t). (7.24)
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 150/183
133
With PI control, it becomes
mx(t) + bx(t) + kx(t) = −K p(f m(t) − f r(t)) − K i
(f m(τ ) − f r(τ ))dτ, (7.25)
where f r is the desired reference force. After substituting with f m = kx, taking the
Laplace transformation becomes
{ms2 + bs + k(1 + K p) + kK i
s }X (s) = (K p +
K is
)F r(s). (7.26)
The transfer function of this system is
Figure 7.12: Explicit PI force control.
X (s)
F r(s) = K ps + K i
ms3 + bs2 + k(1 + K p)s + kK i . (7.27)
The characteristic polynomial of system is
Π(s) = ms3 + bs2 + k(1 + K p)s + kK i. (7.28)
The Routh criterion yields the stability bound on the values of K i and K p:
0 < K i <
b
m(1 + K p). (7.29)
Note that the stability condition does not contain the environment stiffness, k.
The stability can be achieved as long as the gains are bounded. From this stability
condition, since the impedance parameters, m, b, are fixed, the way to increase the
range of the integral gain K i is to either increase the proportional gain K p or add
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 151/183
134
Root Locus
Real Axis
I m a g A x i s
−5 −4 −3 −2 −1 0 1 2 3 4 5−10
−8
−6
−4
−2
0
2
4
6
8
10
Figure 7.13: Root locus under explicit PI force control with K p = 1.
the active damping to the system.
The root locus of this system is shown in Figure 7.13 with parameters selected for
simulation purpose(m = 1, b = 1, k = 10), the proportional gain K p = 1 and the
range of integral gain, 0 ≤ K i ≤ 10. As shown in the root locus plot, the increasing
the K i gain causes the poles moving to the left and eventually makes the system
unstable with K i > 2 as predicted. Figure 7.14 shows the root locus for the system
with K p = 10. As can be seen in the plot and expected, the stability conditions are
always satisfied with the range of 0 ≤ K i ≤ 10 and the poles never across over the
imaginary axis. When the active damping is added, the equation of motion is
mx + bx + kx = −K p(f m − f r) − K i
(f m − f r)dt − K v x. (7.30)
The transfer function of this system is given by
X (s)
F r(s) =
K ps + K ims3 + (b + K v)s2 + k(1 + K p)s + kK i
. (7.31)
The characteristic polynomial of system is
Π(s) = ms3 + (b + K v)s2 + k(1 + K p)s + kK i. (7.32)
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 152/183
135
Root Locus
Real Axis
I m a g A x i s
−1 −0.8 −0.6 −0.4 −0.2 0 0.2−20
−15
−10
−5
0
5
10
15
20
Figure 7.14: Root locus under explicit PI force control with K p = 10.
Figure 7.15: Explicit PI force control with active damping.
It is Hurwitz if and only if the gains are chosen such that
0 < K i < (b + K v)
m (1 + K p). (7.33)
Clearly, the upper bound of K i increases with active damping as expected and it
can be seen in Figure 7.16. Figure 7.16 shows the root locus of the system with the
active damping under same parameters of Figure 7.13 where the crossover of j -axis
can be observed as K i is increased without the active damping term. However,
the active damping provides the large upper bound of K i and thus the stability
condition is satisfied. Note that another stationary pole is not shown in the plot for
enlargement of root locus.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 153/183
136
Root Locus
Real Axis
I m a g A x i s
−0.2 −0.15 −0.1 −0.05 0 0.05 0.1 0.15 0.2−0.4
−0.3
−0.2
−0.1
0
0.1
0.2
0.3
0.4
Figure 7.16: Root locus under explicit PI force control with the activedamping Kv = 1000 and K p = 1.
7.4.2 Stability of Time Delayed System
The time delay causes phase lags in control systems and consequently affects on
the stability of the system. Due to the signal processing or transportation lag, there
exists a time delay on force measurement. To investigate the influence of delayed
force measurement, the equation was modified with a time delay in measured force
term. The transfer function of time delay is given by the transform of the delayed
impulse response with T sec:
Figure 7.17: Explicit PI force control with time delay.
Lh(t) = Lδ (t − T ) = e−sT . (7.34)
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 154/183
137
Clearly, the transfer function of a time delay is not a rational function. With a
delayed force measurement, the closed loop system is
mx(t) + bx(t) + kx(t) =−
K p(f m(t−
T )−
f r(t))−
K i (f m(τ −
T )−
f r(τ ))dτ.
If we take the Laplace transform, we obtain
(ms2 + bs + k + kK pe−sT + kK i
s e−sT )X (s) = (K P +
K is
)F d(s). (7.35)
The irrational transfer function of a time delay can be approximated with Taylor
series expansion
e−sT = 1
−sT +
s2T 2
2 − s3T 3
6
+
· · ·. (7.36)
Taking a first order approximation gives a rise to
[ms2 + (b − kK pT )s + k(1 + K p − K iT ) + kK i
s ]X (s) = (K P +
K is
)F d(s). (7.37)
The transfer function of this system is
X (s)
F r(s) =
K ps + K ims3 + (b
−kK pT )s2 + k(1 + K p
−K iT )s + kK i
. (7.38)
The characteristic polynomial of system is
Π(s) = ms3 + (b − kK pT )s2 + k(1 + K p − K iT )s + kK i. (7.39)
The Routh criterion yields the following stability conditions:
(i) (b − kK pT ) > 0 −→ K P < b
kT
(ii) kK i > 0 −→ K i > 0 (7.40)(iii) (b − kK pT )k(1 + K p − K iT ) − mkK i > 0 −→ K i <
b − kK pT
m + (b − kK pT )T λ
(1 + K p).
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 155/183
138
Thus, we conclude that the stability bounds of a time delayed system can be obtained
as
K p < b
kT , 0 < K i < λ(1 + K p). (7.41)
The effect on stability condition of a time delay can be investigated with the gain λ.
The upper bound gain of K i, λ, depends on the the impedance parameters m, b, k
as well as the time delay T. Let λ be the reciprocal of λ, then
λ
= 1
λ =
m
b − kK pT >0
+ T, (7.42)
where b − kK pT is always positive due to the first stability condition of (7.40).
Clearly λ
is increased as T increased, and consequently the upper bound of K i,
λ, is decreased. Thus the effect of a time delay is decrease the stability range of
the proportional gain K p as well as K i. To show the effect of the time delay on
stability of the system, the root locus with K p = 10 is shown under small time
delay, T = 0.002s, which is selected to satisfy the first stability condition in (7.40).
As mentioned before, the system is always stable with K p = 10 and 0 ≤ K i ≤ 10
in case of without a time delay. Introducing the small time delay decreases the
upper bound of K i to 8.82, and gives rise to violation of stability condition with K i
beyond that value. The stability bounds of the time delayed system with the active
damping K v can be obtained by substituting b with b + K v:
K p < b + K v
kT , 0 < K i < κ(1 + K p), (7.43)
where the reciprocal of κ can be written as
κ
= 1
κ
= m
(b + K v) − kK pT >0
+ T. (7.44)
It is seen that the effect of adding the active damping for the time delayed system
is to increase the range of K i by providing the wide range of K p as expected. The
resulting root locus is shown in Figure 7.19. As can be seen in the root locus plot,
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 156/183
139
Root Locus
Real Axis
I m a g A x i s
−1 −0.8 −0.6 −0.4 −0.2 0 0.2−20
−15
−10
−5
0
5
10
15
20
Figure 7.18: Root locus of the time delayed system under explicit PI forcecontrol with K
p = 10.
the active damping make the system more stable by setting the upper bound of K i
further high.
Root Locus
Real Axis
I m a g A x i s
−0.2 −0.18 −0.16 −0.14 −0.12 −0.1 −0.08 −0.06 −0.04 −0.02 0 0.02−0.4
−0.3
−0.2
−0.1
0
0.1
0.2
0.3
0.4
Figure 7.19: Root locus of the time delayed system under explicit PI forcecontrol with active damping K v = 1000 and K p = 10.
7.5 Experimental Study
7.5.1 Experimentation Environment
A six axis force/toqrue sensor (ATI Model 15/50) with force range of 65N
and torque range of 5Nm is mounted between the base plate and the mounting
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 157/183
140
block as in Figure 7.20. The sensor outputs are interfaced to the PC through
Figure 7.20: Experimental testbed for tension control.
the serial communication and then, through the Ethernet, to a DSP-based motion
controller from ARCS Inc. The suture, which is used in this experimental study,
is USP (the United States Pharmacopeia) 2-0 suture (coated, braided silk suture).
Table 7.1 summaries the suture diameters according to suture sizes with knot-pull
tensile strength [82]. The knot-pull tensile strength is the limited average tension
Table 7.1: Suture diameter-strength relationship.
USP Metrix Diameter [mm] Knot-Pull Tensilesize size min max Strength(Kg)
0 3.5 0.35 0.399 2.162-0 3 0.30 0.339 1.443-0 2 0.20 0.249 0.964-0 1.5 0.15 0.199 0.605-0 1.0 0.10 0.149 0.406-0 0.7 0.07 0.099 0.20
strength by USP at which breaking failure occurs in knotted suture. Seeking to
find an optimal tension value during the suturing was failed because of lack of the
information and the value selected from surgeon’s personal experience. For regarding
the suture pullout value from the various tissues, the experimental results can be
known and summarized in Table 7.2 [83]. The suture pullout value are defined as
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 158/183
141
the tension value, which causes the tissue failure. The free tail of the suture is fixed
Table 7.2: Suture pullout values.
Tissues tension(N)Fat 1.96
Muscle 12.45Skin 17.85
Fascia 36.98
with a fixture and the needle tail is attached to one of the EndoBot as shown in
Figure 7.20. The joint velocities are estimated with the washout filter and then
transformed to the Cartesian velocities, xm = J (q )q . The discrete controller was
implemented at 2ms sampling rate.
7.5.2 Experimental Evaluation of Tension Controller
This section presents the results of the implementation of the explicit force
control strategy described in the previous section. The basic tension control used
in these experiments consisted of first bringing the end-effector close to a certain
position where the suture is not in tension. The hybrid control scheme was then
started and the desired force commanded in the direction of pulling the suture. Thiscaused the manipulator to come into in tension and exert the desired tension. The
data from each control loop was captured at the sampling rate (100Hz) and stored
by the PC for off-line analysis. The force reference in these experiments is stepped
to −15N . The performance of each controller was quantified using the following
measures:
RMS force error eRMS = N
k=1(f m(k)−f r)2
N
Maximum force error e∞ = f m(k)∞
The first set of experiments was performed to see the effect of the active damping
and the resulting force trajectories are plotted in Figures 7.21–7.23. As expected
and seen in Figure 7.21, the integral force control gave rise to the zero steady state
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 159/183
142
error to the step such that the actual force converged to the desired value of -15
Newtons. However, the initial force spike was observed during the transient contact
period due to the integral action and lack of the damping. Next two figures show
the effect of adding the active damping into the system in order to reduce the initialforce spike. With same integral gain, the active damping seems very effective to
achieve the compliant contact transition. The improvement in the performance of
this controller was made by changing the active damping gain, K v, as shown in
Figure 7.23. Table 7.3 summarizes the performance of these controller. From these
Table 7.3: Comparison of force controller with active damping.
Gains eRMS N e∞ N
K i = 0.5, K v = 0 5.92 71.91K i = 0.5, K v = 1000 4.76 31.45K i = 0.5, K v = 2000 4.70 27.21
results, it is clear that the active damping seems promising method to reduce the
initial force spike at the cost of increasing the settling time. Introducing the propor-
tional gain, K p achieved the faster response at the expense of slightly increasing the
force spike as shown in Figure 7.24. Figure 7.25 shows the effect of increasing the
proportional gain, K p to the stability bound. As can be seen in plot, the bouncing
phenomenon was observed and the performance was degraded as shown in Table 7.4.
Table 7.4: Effect of the proportional gain.
Gains eRMS N e∞ NK p = 0.3, K i = 0.5, K v = 2000 3.68 31.24K p = 0.5, K i = 0.5, K v = 2000 4.84 33.19
The bouncing phenomenon was also observed as decreasing the active damp-
ing gain, K v, and the experimental results can be seen in Figures 7.26–7.28. As
predicted, reducing the active damping gives rise to lower stability bounds of the
control gain and consequently shows the tendency of being unstable. A significant
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 160/183
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 161/183
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 162/183
145
0 5 10 15 20 25 30−80
−70
−60
−50
−40
−30
−20
−10
0
10
time(s)
F o r c e ( N )
Desired ForceMeasured Force
Figure 7.25: Experimental data from PI control plus active damping withK p = 0.5, K i = 0.5, and K v = 2000.
0 5 10 15 20 25 30−80
−70
−60
−50
−40
−30
−20
−10
0
10
time(s)
F o r c e ( N )
Desired ForceMeasured Force
Figure 7.26: Experimental data from PI control plus active damping withK p = 0.2, K i = 0.75, and K v = 2000.
0 5 10 15 20 25 30−80
−70
−60
−50
−40
−30
−20
−10
0
10
time(s)
F o r c e ( N )
Desired ForceMeasured Force
Figure 7.27: Experimental data from PI control plus active damping withK p = 0.2, K i = 0.75, and K v = 1000.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 163/183
146
0 5 10 15 20 25 30−80
−70
−60
−50
−40
−30
−20
−10
0
10
time(s)
F o r c e ( N )
Desired ForceMeasured Force
Figure 7.28: Experimental data from PI control plus active damping withK p = 0.2, K i = 0.75, and K v = 500.
0 5 10 15 20 25 30−80
−70
−60
−50
−40
−30
−20
−10
0
10
time(s)
F o r c e ( N )
Desired Force
Measured Force
Figure 7.29: Experimental data from PI control plus active damping withK p = 0.2, K i = 0.3, K v = 2000, and the modified force referencetrajectory with F des = −(f + 5)N.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 164/183
CHAPTER 8
CONCLUSIONS
The goal of this thesis is to develop a robotic system for laparoscopic suturing task.
This chapter concludes the thesis by providing a summary of results obtained in the
preceding chapters and suggestion of areas for future research.
8.1 Summary
The focus of this thesis is supervisory autonomous robotic system for MIS su-
turing. A new surgical robotic system built in-house for minimally invasive surgery
is presented. Based on the analytical model, parameter identification is carried out
with various input signals. For the robotic suturing using the EndoBot, we carried
out the analysis on the suturing task and developed and implemented robotic su-
turing algorithms. For knot placement, the sliding condition based on a two-point
contact model of a knotted suture is developed. In order to guarantee the safe op-
eration in the suturing task in which dynamics of the system cannot be completely
predicted over the entire range of operating due to the non-deterministic nature of
behavior of environments, a human sharing supervisory controller is implementedwith the corresponding state transition diagram for suturing task. For autonomously
evolution of suturing task, an energy based output feedback controller and an opti-
mal state feedback controller with the Kalman filter based on the globally linearized
system are implemented. When human operators interact with robot systems, the
augmentation plays a key role in order to enhance the human’s capability. When a
surgeon and a robot share the different control aspect, shared control proposed in
this thesis can augment the human’s capability by imposing the artificial constraints.
Finally, a base force/torque sensor method is presented for tension measurement and
a hybrid force/position strategy is implemented to effectively regulate the tension
while achieving the desired motion profile.
147
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 165/183
148
8.2 Future Work
This session describes several possible areas of future research.
Manipulator Enhancement Medical palpation plays an important role in both
diagnosis as well as therapy. The ability to sense the tactile information can
enhance the performance of robustness on the suturing tasks [86, 87]. A
major extension of this research related to palpation can be to develop the
MIS instruments with tactile sensors. With tactile sensors, a closed-loop feed-
back control for regulating the gripping force can be implemented. Clearly, it
can be useful in handling soft tissues and securely grasping the suture. For
performing the cooperative tasks with two the EndoBots, calibration between
two local frames is required. At the present, kinematic calibration is a tediousand time consuming procedure. It is desirable to develop a more automatic
calibration method. One possibility is to equip with supporting arms with
position sensors.
Vision Feedback Vision-based control can enhance the performance of the sutur-
ing task in autonomous operations. The primary advantage of vision sensors
is their ability to provide information on suture trajectory. From the prelim-
inary experiments on the suturing task with the EndoBot system, graspingthe suture tail is a difficult and challenging task. Since it is not possible to
predict the trajectory of the suture tail, integrating the vision sensor in the
feedback loop can enhance the robustness of the suturing algorithms. Calibra-
tion with two cameras to get 3D depth information and visibility due to blood
can be technical challenge issues. Vision sensors can be used to generate the
local map so as to provide an active guidance in manual mode. With active
deformable models generated from vision information, surgeons can teach the
niche for surgical procedures such as knot tying and cutting.
Quantification of Knot Quality Following directly from this research, it will be
important to find the values of the optimal tension for securing the knot.
The optimal tension values can be different to the strength of tissues and the
tensile strength of the sutures, but no publication was reported. Therefore,
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 166/183
149
it is necessary to obtain the optimal tension values through the experiments
with in vivo tissues. Another thrust of the research on suturing will be to
quantitatively evaluate the knot quality. It might be necessary to construct
the evaluation criteria such as the size of the knot and the magnitude of knotbreakage force.
Surgical Simulator The other trust to surgical applications can be to develop a
virtual reality simulator for the suturing tasks in minimally invasive surgery
[88, 89, 90]. A dynamic suture modeling is required to train surgeons for the
suturing tasks [85].
Surgical Robot System for the Beating Heart Beating heart bypass surgery
can eliminate many of the complications associated with stopping the heart
and using a heart-lung machine [92]. Beating heart bypass surgery requires
surgeons to stitch together vessels on the beating heart. Motion synchroniza-
tion is the key technical difficulty [91].
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 167/183
LITERATURE CITED
[1] H. Paul, B.D. Mittelstadt, W.L. Bargar, B.L. Musits, R.H. Taylor, P.Kazanzides, J.F. Zuhars, and B. Williamson, “A Surgical robot for total hipreplacement surgery,” in Proceedings of the 1992 IEEE International Conference on Robotics and Automation , vol. 1, pp 606-611, Nice, France,1992.
[2] T.C. Kienzle, S.D. Stulberg, M. Peshkin, A. Quaid, and C-H. Wu,“Anintegrated CAD-robotics system for total knee replacement surgery,” inProceedings of the 1993 IEEE International Conference on Robotics and Automation , vol. 1, pages 889-894, Atlanta, GA, 1993.
[3] M. Fadda, T. Wang, M. Marcacci, S. Martelli, G. Marcenaro, M. Nanetti, C.Paggetti, A. Visani, and S. Zaffagnini, “Computer-assisted knee arthroplastyat Rizzoli institutes,” In Proceedings of the International Symposium on Medical Robotics and Computer Assisted Surgery , pages 26-30, 1994.
[4] Ho, S.C., R.D. Hibberd, and B.L. Davies, “Robot assisted knee surgery,”IEEE Engineering in Medicine and Biology Magazine Sp. Issue on Robotics in Surgery , May/June, pages 292-300, 1995.
[5] R.E. Ellis, “From scans to sutures: robotics methods for computer-enhancedsurgery,” in 9th International Symposium of Robotics Research , pp203-210,
Snowbird, Utah, 1999.
[6] I. W. Hunter, T. D. Doukoglou, S. R. Lafontaine, P. G. Charette, L. A. Jones,M. A. Sagar, G. D. Mallinson, and P. J. Hunter, “A teleoperatedmicrosurgical robot and associated virtual environment for eye surgery,”Presence , vol. 2, pages 265-280, 1993.
[7] Y. S. Kwoh, J. Hou, E. A. Jonkeere, and S. Hayati, “A robot with improvedabsolute positioning accuracy for CT guided stereotactic brain surgery,” IEEE Transactions on Biomedical Engineering , 35(2):153-160, 1988.
[8] B.Davies, S.Starkie, S.J. Harris, E. Agterhuis, V.Paul, and L.M. Auer, “
Neurobot: a special-purpose robot for neurosurgery,” in Proceedings of the 2000 IEEE International Conference on Robotics and Automation , vol. 4,pages 4104-4109, San Francisco, CA, 2000.
[9] A. Faraz and S. Payandeh, “A Robotic Case Study: Optimal design forlaparoscopic positioning stands,” in Proceedings of the 1997 IEEE International Conference on Robotics and Automation , pages 1553-1560,Albuquerque, NM, 1997.
150
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 168/183
151
[10] J. Funda, B. Eldridge, K. Gruben, S.Gomory, and R. Taylor,“Comparison of two manipulator designs for laparoscopic surgery,” IEEE Engineering in Medicine and Biology Magazine , Vol.14, No.3, pages 289-291, 1994.,Telemanipulator and Telepresence Technologies , SPIE, Vol.2351, pages172-183, 1994.
[11] A.J. Madhani and J.K Salisbury, “Wrist mechanism for surgical instrumentfor performing minimally invasive surgery with enhanced dexterity andsensitivity,” U.S. Patent No. 5,797,900, 1997.
[12] M.C. Cavusoglu, F. Tendick, M. Cohn, and S.S. Sastry, “A laparoscopictelesurgical workstation,” IEEE Transactions on Robotics and Automation ,Vol.15, No. 4, pages 728-739, 1999.
[13] A. Faraz and S. Payandeh, Synthesis and workspace study of endoscopic extenders with flexible stems , ERL Technical Report.
[14] A. Faraz, S. Payandeh, and A. Nagy, “Issues and design concepts inendoscopic extenders,” in Proceedings of 1995 IFAC Man/Machine Systems ,pages 231-236,1995.
[15] C. Reboulet and S. Durand-Leguay, “Optimal design of redundant parallelmechanism for endoscopic surgery,” in Proceedings of the 1999 IEEE/RSJ International Conference on Intelligent Robots and Systems , Vol. 2, pages1432-1437, 1999.
[16] C. Scott-Conner, The Sages Manual: Fundamentals of laparoscopy and GI endoscopy , Springer, 1998.
[17] Z. Szabo and A. Cuschieri, Tissue approximation in endoscopic surgery , ISISMedical Media, 1995.
[18] C. Bernard, H. Kang, S.K. Singh, and J.T. Wen, “Robotic system forcollaborative control in minimally invasive surgery,” Industrial Robot , 26(6),Nov/Dec, pages 476-484, 1999.
[19] A.Melzer, M.O. Schurr, M. Lirich, B. Klemm, D.Stockel, and G.Buess,“Future trends in endoscopic suturing,” End. Surg., pages 78-82, 1994.
[20] A.J. Madhani, G. Niemeyer, and J.K. Salisbury, “The black falcon: a
teleoperated surgical instrument for minimally invasive surgery,” inProceedings of the 1998 IEEE/RSJ International Conference on Intelligent Robots and Systems , Vol. 2, pages 936-944, 1998.
[21] C. Cao, C. MacKenzie, S. Payandeh,“Task and motion analyses in endoscopicsurgery,” in ASME IMECE Conference Proceesings: 5th Annual Symposium on Haptic Interfaces for Virtual Environment and Teleoperator Systems ,pages 583-590, Atlanta, GA, 1996.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 169/183
152
[22] J. Funda, R. Taylor, K. Gruben, and D. La Rose, “Optimal motion control forteleoperated surgical robots,” in SPIE Symposium on Optical Tools for Manufacturing and Advanced Automation , Vol. 2057, pages 211-222, Boston,MA, 1993.
[23] J. Funda, R. Taylor, B. Eldridge, S. Gomory, and K. Gruden, “Constrainedcartesian motion control for teleoperated surgical robots,” IEEE Transactions on Robotics and Automation , Vol.12, No. 3, pages 453-465, 1996.
[24] Y. Wang, D.R. Uecker, K.P. Laby, J. Wilson, S. Jordan, and J. Wright,“Method and apparatus for performing minimally invasive cardiacprocedures,” U.S. Patent No. 5,762,458, 1998.
[25] M. Mitsuishi, T. Watanabe, H. Nakanishi, T. Hori, H. Watanabe, and B.Kramer, “A tele-microsurgery system across the internet with a fixedviewpoint/operation-point,” in Proceeding of the 1995 IEEE/RSJ
International Conference on Intelligent Robots and Systems , Vol. 2, pages178-185, 1995.
[26] S. Charles, H. Das, T. Olm, C. Boswell, G. Rodrigues, R. Steele, and D.Istrate, “Dexterity-enhanced telerobotic microsurgery,” in International Conference on Advanced Robotics , Monterey, CA, 1997.
[27] F. Tendick and M.C. Cavusoglu,“Human-machine interfaces for minimallyinvasive surgery,” in Proceedings of the IEEE International Conference on Engineering Medical Biology Society , Chicago, IL, 1997.
[28] F. Tendick, M.S. Downes, M.C. Cavusoglu, and L.W. Way, “Development of virtual environments for training skills and reducing errors in laparoscopicsurgery,”in Proceedings of the SPIE International Symposium on Biological Optics (BIOS’98), San Jose, CA, 1998.
[29] H. Kazerooni and J. Guo, “Human extenders,” Journal of Dynamics Systems,Measurement and Control , Vol. 115, pages 281-290, 1993.
[30] T.B. Sheridan, Telerobotics, automations, and human supervisory control ,Cambridge, MIT Press, 1992.
[31] P.G. Backes and K.S. Tso, “UMI: an interactive supervisory and shared
control system for telerobotics,” in Proceedings of the 1990 IEEE International Conference on Robotics and Automation , pages 1096-1101,Cincinnati, OH, 1990.
[32] S.A. Hayati and S.T. Venkataraman, “Bilevel shared control forteleoperators,” U.S. Patent No. 5,086,982, 1992.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 170/183
153
[33] A. Douglas and Y. Xu, “Real-time shared control system for spacetelerobotics,” Journal of Intelligent and Robotic Systems: Theory and Applications , vol. 13, pages 247-262, 1995.
[34] R. Kumar, P. Berkelmen, P. Gupta, A. Barnes, P. Jensen, L.L. Whitcomb,
and R.H. Taylor, “Preliminary experiments in cooperative human/robot forcecontrol for robot assisted microsurgical manipulation,” in Proceedings of the 2000 IEEE International Conference on Robotics and Automation , 2000.
[35] R. Kumar, P. Jensen, and R.H. Taylor, “Experiments with a steady handrobot in constrained compliant motion and path following,” In Proceedings of IEEE RO-MAN99 , pages 92-97, Pisa, Italy, 1999.
[36] R. Kumar, T.M. Goradia, A.C. Barnes, P. Jensen, L.L. Whitcomb, D.Stoianovici, L.M. Auer, and R.H. Taylor, “Performance of roboticaugmentation in microsurgery-scale motions,” In Proceedings of Medical
Image Computing and Computer Assisted Intervention , pages 1108-115,Cambridge,UK, 1999.
[37] G.J. Hamlin, and A.C. Sanderson, “A novel concentric multilink spherical joint with parallel robotics applications,” in Proceedings of the 1994 IEEE International Conference on Robotics and Automation , pages 1267-1272, 1994.
[38] R.M. Murray, Z. Li, and S.S. Sastry, A mathematical introduction to roboti manipulation , CRC Press, Boca Raton, FL, 1994.
[39] B. Armstrong, “Friction: Experimental determination, modelling andcompensation,” in Proceedings of the 1988 International Conference on Robotics and Automation , pages 1422-1427, 1988.
[40] B. Armstrong-Helouvry, Control of machine with friction , Kluwer Academic,1991.
[41] P. Lischinsky, C. Canudas-de-Wit, and G. Morel, “Friction compensation foran industrial hydraulic robot,” IEEE Control Systems Magazine , vol. 19, no.1, pages 25-32, 1999.
[42] S.C.P Gomes and J.P. Chretien, “Dynamic modeling and frictioncompensated control of a robot manipulator joint,” in Proceedings of the 1992
International Conference on Robotics and Automation , pages 1429-1435, 1992.
[43] G. Morel and S. Dubowsky, “The precise control of manipulators with jointfrcition: a base force/torque sensor method,” in Proceedings of the 1996 IEEE International Conference on Robotics and Automation , pages 360-365, 1996.
[44] M. Meggiolaro, P. Jaffe, and S. Dubowsky, “Achieving fine absolutepositioning accuracy in large powerful manipulators,” in Proceedings of the
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 171/183
154
1999 IEEE International Conference on Robotics and Automation , pages2819-2824, 1999.
[45] J.T. Wen and D.S. Bayard, “A new class of control laws for roboticmanipulators, Part I: Non-adaptive case,” International Journal of Control ,
47(5), pages 1361-1385, 1988.
[46] J.T. Wen and S.H. Murphy, “Position and force control of robot arms,” IEEE Transactions on Automatic Cotnrol , 36(3), pages 365-374, 1991.
[47] C. Canudas-de-Wit, B. Siciliano, and G. Bastin, Theory of robot control ,Springer, 1996.
[48] M.R. Schroeder, “Synthesis of low-peak-factor signals and binary sequences of low autocoorelation,” IEEE Transactions on Information Theory , 16, pages85-89, 1970.
[49] K. Koztowski, Modeling and identification in robotics , Springer, 1998.
[50] C.H. An, C.G. Atkeson, and J.M. Hollerbach, “Estimation of inertialparameters of rigid body links of manipulators,” in Proceedings of the 24th IEEE Conference on Decision and Control , pages 990-995, 1985.
[51] B. Armstrong, “On finding exciting trajectories for identification experimentsinvolving systems with nonlinear dynamics,” International Journal of Robotics Research , Vol.8, pages 28-48, 1989.
[52] M. Gautier and W. Khalil, “Exciting trajectories for the identification of base
inertial parameters of robots,” International Journal of Robotics Research ,Vol.11, pages 362-375, 1992.
[53] M. Gautier, “Optimal motion planning for robot’s inertial parametersidentification,” in Proceedings of the 31th IEEE Conference on Decision and Control , pages 70-73, 1992.
[54] C. Presse and M. Gautier, “ New criteria of exciting trajectories for robotidentification,” in Proceedings of the 1993 IEEE Inernatinal Conference on Robotics and Automation , pages 907-912, 1993.
[55] M. Gautier, “A comparison of filtered models for dynamic identification of
robots,” in Proceedings of the 35th IEEE Conference on Decision and Control ,pages 875-880, 1996.
[56] L.S. Wilfinger, J.T. Wen, and S. Murphy, “Integral force control withrobustness enhancement,” IEEE Control System Magazine , 14(1), pages31-40, 1994.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 172/183
155
[57] M. Takegaki and S. Arimoto, “A new feedback method for dynamic control of manipulators,” ASME J. Dynamic Systems, Measurement and Control , 103,pages 119-125, 1981.
[58] S. Arimoto and F. Miyazaki, “Stability and robustness of PID feedback conrol
for robot manipulators of sensory capability,” In Proceeding of the 1st International Symposium of Robotics Research , pages 783-799, 1983.
[59] J. Luh, M. Walker, and R. Paul, “Resolved acceleration control of mechanicalmanipulators,” IEEE Transactions on Automatic Control , 25, pages 468-474,1980.
[60] R. Paul, Robot manipulators: Mathematics, programming, and control ,Cambridge, MIT Press, 1981.
[61] A. Bryson and Y.C. Ho, Applied Optimal Control , Halsted, 1975.
[62] E. Freund, “Fast nonlinear control iwth arbitrary pole-placement forindustrial robots and manipulators,” International Journal of Robotics Research , 1, pages 65-78, 1982.
[63] K. Kreutz, “On manipulator control by exact linearization,” IEEE Transactions on Automatic Control , 34, pages 763-767, 1989.
[64] G.L. Luo, G.N. Saridis, and C.Z. Wang, “A dual-mode control for roboticmanipulators,” In Proceeding of IEEE conference on Decision Control , pages409-414, 1986.
[65] R. Johansson, “Quadratic optimization of motion coordination and control,”IEEE Transactions on Automatic Control , 35(11), pages 1197-1208, 1990.
[66] J.T.Wen, “A unified perspective on robot: The energy Lyapunov functionapproach,” International Journal of Adaptive Control and Signal Processing ,4(6), pages 487-500, 1990.
[67] F. Lin and R.D. Brandt, “An optimal control approach to robust control of robot manipulators,” IEEE Transactions on Robotics and Automation , 14(1),pages 69-77, 1998.
[68] C. Abdallah, D. Dawson, P. Dorato, and M. Jamshidi, “Survey of robust
control for rigid robots,” IEEE Control Systems Magazine , 11(2), pages 24-30,1991.
[69] H.G. Sage, M.F. De Mathelin, and E. Ostertag, “Robust control of robotmanipulators: a survey,” International Journal of Control , 72(16), pages1498-1522, 1999.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 173/183
156
[70] H. Berghuis, P. Lohnberg, and H. Nijmeijer, “Tracking control of robots usingonly position measurements,” in Proceeding of Conference on IEEE Decision and Control , pages 1039-1040, 1991.
[71] B. Bona and M. Indri, “Analysis and implementation of observers for robotic
manipulators,” in Proceeding of the IEEE International Conference on Robotics and Automation , pages 3006-3011, 1998.
[72] N. Hogan, “Impednace control: An approach to manipulation. PartsI-III,”ASME Journal of Dynamic Systems, Measurement, and Control , 107, pages1-24, 1985.
[73] M. Spong and M. Vidyasagor, Robot dynamics and control , John Wiley andSon, New York, 1989.
[74] B. Waibel and H. Kazerooni, “Theory and experiments on the stability of
robot compliance control,” IEEE Transactions on Robotics and Automation ,7(1), pages 95-104, 1991
[75] M. Raibert and J. Craig, “Hybrid position/force control of manipulators,”ASME Journal of Dynamic Systems, Measurement, and Control , 103, pages126-133, 1981.
[76] T. Yoshikawa, “Dynamics hybrid position/force control of robotmanipulators- description of hand constraints and calculation of joints drivingforces,” in Proceeding of the IEEE International Conference on Robotics and Automation , pages 1393-1398, 1986.
[77] O. Khatib, “A unified approach for motion and force control of robotmanipulators: The operational space formulation,” IEEE Journal of Robotics and Automation , pages 43-53, 1987.
[78] T. Yoshikawa, “Force control of robot manipulators,” in Proceeding of the IEEE International Conference on Robotics and Automation , pages 220-226,2000.
[79] S. Chiaverini and L. Sciavicco, “The parallel approach to force/positioncontrol of robotic manipulators,” IEEE Transactions on Robotics and Automation , 9(4), pages 361-373, 1993.
[80] L.S. Wilfinger, J.T. Wen, and S. Murphy, “Integral force control withrobustness enhancement,” in Proceeding of the IEEE International Conference on Robotics and Automation , pages 100-106, 1994.
[81] N. Mandal and S. Payandeh, “Force control strategies for compliant and stiff contact: An experimental study,” in IEEE International Conference on Systems, Man, and Cybernetics , 2 , pages 1285-1290, 1994.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 174/183
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 175/183
APPENDIX A
INVERSE KINEMATICS USING SUBPROBLEMS
Since the EndoBot has a simple kinematic configuration, the inverse kinematics can
be solved with conventional algebraic or geometric method. But for consistency
purpose, the exponentials formula is introduced to solve the inverse kinematics.
The exponentials formula can solve the inverse kinematics problem with the Paden-
Kahan subproblems whose solutions are known in [38] and has a geometric meaning.
The inverse kinematics can be solved from the following steps:
Step 1 (solve for the translational distance, q 4) From the forward kinematicsderivation above, a given position vector from the base frame to tool frame
p0T can has the following form:
p0T = eh2q2eh1q1eh3q3 p34 (A.1)
where, p34 = q 4h3. Since the position vector p34 lies on the axis of h3, it yields
p0T = eh2q2eh1q1 p34. (A.2)
Taking the norm, it gives the translational distance, q 4:
p0T =eh2q2eh1q1 p34
= p34 = q 4. (A.3)
Step 2 (solve for first two angles, q 1 and q 2) Since q 4 is known and h1 and h2
intersect, (A.2) can be solved using Subproblem 2 (Rotation about two sub-
sequent axes) and it gives q 1 and q 2 from (A.2). Due to the multiple solutionsproperty of Subproblem 2, two possible solution might be existed, but it can
be easily determined from the working range consideration.
Step 3 (solve for the roll angle, q 3) The joint angle of the axis 3, q 3, is simply
the roll angle φ.
158
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 176/183
APPENDIX B
LAGRANGIAN FOR A ROBOT MANIPULATOR
B.1 Kinetic Energy of a Robot Manipulator
To calculate the kinetic energy of a robot manipulator, consider the ith link
in Figure B.1. Define the coordinate frame, Oi, attached to the ith joint. Let hi
Figure B.1: Kinetic energy of a rigid body.
be the unit vector of ith rotational direction, ci represent the center of mass for ith
link, pC i denote the position vector of the ith center of mass from the ith coordinate
frame expressed in Oi, I ci be the inertia tensor of the ith link expressed in the frame
attached at the center of mass, and mi be the mass of ith
link.The kinetic energy of the ith link is
T i = 1
2
ωi
vi
T I i miˆ pci
−miˆ pci miI
M i
ωi
vi
, (B.1)
where I i = I ci+mi( pci( p
ci)T − pci2I ) denotes the inertia tensor of the ith link expressed
in Oi frame, wi is the instantaneous angular velocity and vi the linear velocity of the
ith link with respect to the instantaneous ith coordinate frame. M i is the generalized
inertia matrix for the ith link. The total kinetic energy of the manipulator with nth
joints is
T =ni=1
T i. (B.2)
159
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 177/183
160
To represent the kinetic energy with the generalized coordinates, the instantaneous
angular velocity and the linear velocity can be written by
ωi
vii
= J i(q )q, (B.3)
where J i(q ) is the partial Jacobian:
J i(q ) =
h1 · · · hi−1 hi 0 · · · 0
hi p1i · · · hi−1 pi−1,i 0 0 · · · 0
, (B.4)
where all vectors are represented in ith frame. Then, the total kinetic energy of the
manipulator can be given by
T = 1
2
ni=1
q T J T i (q )M iJ i(q )q
= 1
2 q T (
ni=1
J T i (q )M iJ i(q ))
M (q)
q. (B.5)
It gives
T =
1
2 q T
M (q )q, (B.6)
where M (q ) ∈ n×n is the symmetric, positive definite mass inertia matrix.
B.2 Potential Energy of a Robot Manipulator
The potential energy of the ith link is
P i = migT ( p0i + pci)0, (B.7)
where ( p0i + pci)0 represents the height of the ith center of mass in the base frame
and g = [0 0 − 9.8]m/s2 is the gravity vector. Then the total potential energy is
P =ni=1
migT ( p0i + pci)o. (B.8)
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 178/183
APPENDIX C
LEAST SQUARE METHOD
The equation of motion of a robot manipulator can be given with m measurements
at given time instants t1,...,tm
τ = φθ + ε, (C.1)
where φ ∈ m×n is the matrix in which each row represents regressor at each time,
τ ∈ m is a vector with τ (tk), and ε ∈ m is a vector with ε(tk).
The least square solution can be interpreted as a minimum error norm solution
through the orthogonal mapping of vector spaces as shown in Figure C.1. A matrix
Figure C.1: Orthogonal decomposition.
φ takes a vector θ into the column space, R(φ), but due to the noise and unmodeled
dynamics, τ is outside of the column space of φ. Consequently, there exists no
solution on θ. Let θ be a fictitious solution and p be a mapped vector through
p = φθ. Then, p is in the column space of φ and the error ε can be drawn as
shown in Figure C.2. From the orthogonal decomposition diagram, it always has a
161
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 179/183
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 180/183
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 181/183
164
gramian is uniformly bounded and positive definite:
αaI ≤ tf t0
LT o Lodτ =
tf t0
ΦT C T C Φdτ ≤ α2I. (D.7)
In discrete-time invariant system, the observability gramian can be given by
Go = OT O, (D.8)
where O represents the observability matrix. This discrete-time observability gramian
can be interpreted as the persistent matrix in system identification. The system can
be said to be observable if only if the observability gramian is nonsingular in control
system and the unknown parameter can be identified if only if the persistent matrix
is nonsingular.
8/10/2019 robotic surgery- Kang Thesis
http://slidepdf.com/reader/full/robotic-surgery-kang-thesis 182/183
APPENDIX E
DIFFEOMORPHISM
Diffeomorphism A map between manifolds is called as a homeomorphism if it is
bijective (one-to-one and onto), continuous, and has a continuous inverse. A
homoemorphism which is differentialble and has a differentiable inverse is a
diffeomorphism.
Bijective A transformation which is one-to-one and onto is called as bijection. In
other words, it allows every state to be accessed (onto) and has precisely one
pre-image for each state (one-to-one).
• A function f from A to B is called one-to-one (or injection) if ∀x, y ∈ A
(f (x) = f (y)) → x = y.
Figure E.1: One-to-one and not onto
• A function f from A to B is called onto (or surjection) if (∀y ∈ B ∃x ∈ A)
f (x) = y.
Figure E.2: Onto not one-to-one
165