RJ–11 Multiplexer Installation

Page 2: Rj 11 Multi. Insta

GE Medical Systems

GE Medical Systems: Telex 3797371P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A.(Asia, Pacific, Latin America, North America)

GE Medical Systems — Europe: Telex 261794Shortlands, Hammersmith, London W6 8BX U.K.

DAMAGE IN TRANSPORTATION

All packages should be closely examined at time of delivery. If damage is apparent,have notation “damage in shipment ” written on all copies of the freight or expressbill before delivery is accepted or “signed for” by a General Electric representative ora hospital receiving agent. Whether noted or concealed, damage MUST be reportedto the carrier immediately upon discovery, or in any event, within 14 days afterreceipt, and the contents and containers held for inspection by the carrier. A trans-portation company will not pay a claim for damage if an inspection is not requestedwithin this 14 day period.

Call Traffic and Transportation, Milwaukee, WI (414) 785–5052/8*323–5052immediately after damage is found. At this time be ready to supply name of carrier,delivery date, consignee name, freight or express bill number, item damaged andextent of damage.

Complete instructions regarding claim procedure are found in Section “S” of thePolicy & Procedure Bulletins.

3/12/92

Page 3: Rj 11 Multi. Insta

GE MEDICAL SYSTEMS LANGUAGE POLICY FOR SERVICE DOCUMENTATION

REV 0 DIRECTION 2128126

������

� THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.

� IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE OTHERTHAN ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO PROVIDETRANSLATION SERVICES.

� DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS SERVICEMANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.

� FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE SERVICEPROVIDER, OPERATOR OR PATIENT FROM ELECTRIC SHOCK, MECHANICALOR OTHER HAZARDS.

� CE MANUEL DE MAINTENANCE N’EST DISPONIBLE QU’EN ANGLAIS.

� SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANS UNE AUTRELANGUE QUE L’ANGLAIS, C’EST AU CLIENT QU’IL INCOMBE DE LE FAIRETRADUIRE.

� NE PAS TENTER D’INTERVENTION SUR LES ÉQUIPEMENTS TANT QUE LEMANUEL SERVICE N’A PAS ÉTÉ CONSULTÉ ET COMPRIS.

� LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNER CHEZ LETECHNICIEN, L’OPÉRATEUR OU LE PATIENT DES BLESSURES DUES À DESDANGERS ÉLECTRIQUES, MÉCANIQUES OU AUTRES.

� DIESES KUNDENDIENST–HANDBUCH EXISTIERT NUR IN ENGLISCHER SPRACHE.

� FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENÖTIGT,IST ES AUFGABE DES KUNDEN FÜR EINE ENTSPRECHENDE ÜBERSETZUNGZU SORGEN.

� VERSUCHEN SIE NICHT, DAS GERÄT ZU REPARIEREN, BEVOR DIESESKUNDENDIENST–HANDBUCH NICHT ZU RATE GEZOGEN UND VERSTANDENWURDE.

� WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGENDES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTENDURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGEGEFAHREN KOMMEN.

� ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS.

� SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEMS SOLICITA UN IDIOMAQUE NO SEA EL INGLÉS, ES RESPONSABILIDAD DEL CLIENTE OFRECER UNSERVICIO DE TRADUCCIÓN.

� NO SE DEBERÁ DAR SERVICIO TÉCNICO AL EQUIPO, SIN HABERCONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO.

� LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A QUE ELPROVEEDOR DE SERVICIOS, EL OPERADOR O EL PACIENTE SUFRANLESIONES PROVOCADAS POR CAUSAS ELÉCTRICAS, MECÁNICAS O DE OTRANATURALEZA.

WARNING

AVERTISSEMENT

WARNUNG

AVISO

Page 4: Rj 11 Multi. Insta

GE MEDICAL SYSTEMS LANGUAGE POLICY FOR SERVICE DOCUMENTATION

REV 0 DIRECTION 2128126

������

� ESTE MANUAL DE ASSISTÊNCIA TÉCNICA SÓ SE ENCONTRADISPONÍVEL EM INGLÊS.

� SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA, QUE NÃO AGEMS, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, É DARESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE TRADUÇÃO.

� NÃO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO ECOMPREENDIDO ESTE MANUAL DE ASSISTÊNCIA TÉCNICA.

� O NÃO CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A SEGURANÇADO TÉCNICO, OPERADOR OU PACIENTE DEVIDO A‘ CHOQUES ELÉTRICOS,MECÂNICOS OU OUTROS.

� IL PRESENTE MANUALE DI MANUTENZIONE È DISPONIBILESOLTANTO IN INGLESE.

� SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEMS RICHIEDE ILMANUALE IN UNA LINGUA DIVERSA, IL CLIENTE È TENUTO A PROVVEDEREDIRETTAMENTE ALLA TRADUZIONE.

� SI PROCEDA ALLA MANUTENZIONE DELL’APPARECCHIATURA SOLO DOPOAVER CONSULTATO IL PRESENTE MANUALE ED AVERNE COMPRESO ILCONTENUTO.

� NON TENERE CONTO DELLA PRESENTE AVVERTENZA POTREBBE FARCOMPIERE OPERAZIONI DA CUI DERIVINO LESIONI ALL’ADDETTO ALLAMANUTENZIONE, ALL’UTILIZZATORE ED AL PAZIENTE PERFOLGORAZIONE ELETTRICA, PER URTI MECCANICI OD ALTRI RISCHI.

ATENÇÃO

AVVERTENZA

Page 5: Rj 11 Multi. Insta

GE MEDICAL SYSTEMS

DIRECTION 2137747

RJ–11 MULTIPLEXER INSTALLATION

A

REV 0

LIST OF EFFECTIVE PAGES

REVISION DATE PRIMARY REASON FOR CHANGE

REV 0 June 9, 1995 Initial Release.. . . . . . . . . . . . . . . .

PAGE REV PAGE REV PAGE REV PAGE REV PAGE REV

Title Page 0. . . . . . . . . . . Damage In Trans. –. . . . Direction 2128126 0*. . . A 0. . . . . . . . . . . . . . . . . . B blank. . . . . . . . . . . . . . 1–1 to 1–9 0. . . . . . . . . . GE Logo 0. . . . . . . . . . .

* This revision number corresponds to the indicated document’s revision control system.

Page 6: Rj 11 Multi. Insta

GE MEDICAL SYSTEMS

DIRECTION 2137747

RJ–11 MULTIPLEXER INSTALLATION

Blank

Page 7: Rj 11 Multi. Insta

RJ–11 MULTIPLEXER INSTALLATIONGE MEDICAL SYSTEMS

DIRECTION 2137747

���

REV 0

SECTION 1 – RJ–11 MULTIPLEXER INSTALLATION

1–1 Overview

This document provides the GE Medical Systems usage guidelines and installation instructions for the RJ–11 Multi-plexer.

The following RJ–11 Multiplexers are available:

o 4–line, 115VAC/60Hz (46–328475P1)

o 8–line, 115VAC/60Hz (46–328475P2)

o 4–line, 220VAC/50Hz (46–328475P3)

o 8–line, 220VAC/50Hz (46–328475P4)

1–2 GE POLICY FOR RJ–11 MULTIPLEXER INSTALLATIONS

Refer to the following policies for fixed sites:

o Use only 46–328475P1–4 multiplexers. Other multiplexers are incompatible with InSite software.

o Connect only GE service equipment to RJ–11 Multiplexers in Canada or the United States. Sites outsideof Canada or the United States may also have a fax machine connected to EXTENSION (port) 1.

NoteIf you will be connecting a fax machine to a multiplexer (outside of Canada or the United States),inform the customer that InSite will be unavailable when the fax machine is using the line.

o Connect no more than a total of two Cath Lab, CT, MR, or PET systems on a single multiplexer. Theremaining systems may be X–Ray, Nuclear, Ultrasound, or other GE serviced devices.

o Connect a direct dial (incoming) line to the multiplexer. Indirect lines (that connect through a switchboardor require dialing an extension) will not work with InSite software and do not provide reliable modem con-nections.

o Never connect a telephone (for voice communications) or an answering machine to a multiplexer.

Page 8: Rj 11 Multi. Insta

RJ–11 MULTIPLEXER INSTALLATIONGE MEDICAL SYSTEMS

DIRECTION 2137747

���

REV 0

1–2 GE POLICY FOR RJ–11 MULTIPLEXER INSTALLATIONS (continued)

Notes for Canadian Installations:

The Canadian Department of Communications label identifies certified equipment. This certification means that theequipment meets certain telecommunications network protective, operational and safety requirements. The Depart-ment does not guarantee the equipment will operate to the user’s satisfaction.

Before installing this equipment, users should ensure that it is permissible to be connected to the facilities of the localtelecommunications company. The customer should be aware that compliance with the above conditions may notpervent degradation of services in some situations.

Repairs to certified equipment should be made by an authorized Canadian maintenance facility designated by thesupplier. Any repairs or alterations made by the user to this equipment, or equipment malfunctions, may give thetelecommunications company cause to request the user disconnect the equipment.

Users should ensure for their own protection that the electrical ground connections of the power utility, telephone lines,and internal metallic water pipe system, if present, are connected together. This precaution may be particularly impor-tant in rural areas.

Caution: Users should not attempt to make such connections themselves, but should contact the appropriate electri-cal inspection authority, or electrician, as appropriate.

Authorized Canadian service by Signatel, Markham Ontario L3R8VT, 416–477–9977.

Page 9: Rj 11 Multi. Insta

RJ–11 MULTIPLEXER INSTALLATIONGE MEDICAL SYSTEMS

DIRECTION 2137747

���

REV 0

1–3 INSTALLING THE RJ–11 MULTIPLEXER

Use the following procedures to install an RJ–11 Multiplexer.

1–3–1 Before You Begin

Before you begin, you need the following:

o The site must have a direct dial line to connect to the multiplexer. A line that connects through a switchboard or requires dialing an extension is incompatible with InSite software.

o The site’s electrical ground connections of the power utility, telephone lines, and internal metallic waterpipe system (if any) must be connected together. If these grounds are not connected, contact the site’smaintenance department to have the situation corrected.

o A small Phillips screwdriver is needed to access the switches (if you need to change them).

o You need the RJ–11 Multiplexer.

1–3–2 Switch 1 Setting

The switch settings on the RJ–11 Multiplexer can enable the multiplexer to have a fax machine attached to EXTEN-SION 1. The factory default settings are for no fax connection. Therefore if you are not connecting a fax machine tothe multiplexer (and this is a new installation), skip to Section 1–3–3, Multiplexer Connections.

Use the appropriate procedure to set the switches on the RJ–11 Multiplexer.

NoteDo not connect a fax machine to the multiplexer on sites in Canada or the United States.

On a 4–line multiplexer use the following steps to set switch 1.

1. Remove the two Phillips screws on the bottom of the multiplexer. Keep these screws for later use.

2. Turn the multiplexer over, and remove the top cover of the multiplexer case.

3. If you are connecting a fax machine to EXTENSION 1, push switch 1 (on the S1 switches) down toward thenumeral 1. See Illustration 1–1.

If you are not connecting a fax machine to the multiplexer, push switch 1 down toward OPEN. See Illustration1–1.

4. Slide on the top cover of the multiplexer case. If the sides do not meet exactly, rotate the top by 180 degrees.

5. Replace the two Phillips screws on the bottom of the multiplexer. Go to Section 1–3–3, Multiplexer Connections.

Page 10: Rj 11 Multi. Insta

RJ–11 MULTIPLEXER INSTALLATIONGE MEDICAL SYSTEMS

DIRECTION 2137747

���

REV 0

1–3–2 Switch 1 Setting (continued)

46–328475P146–328475P3

(TOP VIEW)

SWITCH 1

Switches on 4–line RJ–11 multiplexerIllustration 1–1

On an 8–line multiplexer use the following steps to set the switches.

1. Loosen the two Phillips screws from the switch plate on the bottom of the multiplexer. See Illustration 1–2.

SWITCHPLATE

(BOTTOM VIEW)

46–328475P2 AND 46–328475P4

SWITCH PLATE ON 8–LINE RJ–11 MULTIPLEXERIllustration 1–2

Page 11: Rj 11 Multi. Insta

RJ–11 MULTIPLEXER INSTALLATIONGE MEDICAL SYSTEMS

DIRECTION 2137747

���

REV 0

1–3–2 Switch 1 Setting (continued)

2. Remove the switch plate.

3. If you are connecting a fax machine to EXTENSION 1, push switch 1 toward ON. See Illustration 1–3. If you are not connecting a fax machine to the multiplexer, push switch 1 toward the numeral 1. See Illustration1–3.

4. Replace the switch plate and tighten the screws. Go to Section 1–3–3, Multiplexer Connections.

SWITCHES

ON

1 2 3 4 5 6 7 8

WITHOUTFAX

WITHFAX

Switches on 8–line rj–11 multiplexerIllustration 1–3

Page 12: Rj 11 Multi. Insta

RJ–11 MULTIPLEXER INSTALLATIONGE MEDICAL SYSTEMS

DIRECTION 2137747

���

REV 0 (7 June 1995)

1–3–3 Multiplexer Connections

1. Place the RJ–11 Multiplexer in a secure, dry area (also outside the 10 Gauss line for MR systems). The 8–lineunits may be rack–mounted in the customer’s phone or utility room using the included brackets.

2. Connect each modem (and fax machine, if any) to its designated extension. See Illustrations 1–4 and 1–5 forconnection guidelines.

NoteConnect no more than a total of two Cath Lab, CT, MR, or PET systems on a single multiplexer, and donot connect a fax machine to the multiplexer on sites in Canada or the United States.

MULTIPLEXER

EXTENSION 1

EXTENSION 2

EXTENSION 3

EXTENSION 4

LINE

5X MR SYSTEM MODEM FORINSITE

EDM MAGNET MODEM FORMAGNET MONITORINGMONITORING

CT SYSTEM MODEM

INCOMING LINE

X–RAY SYSTEM MODEM

EXAMPLE: 4–LINE MULTIPLEXER EXTENSION CONNECTIONS Without a fax machine

Illustration 1–4

MULTIPLEXER

EXTENSION 1

EXTENSION 2

EXTENSION 3

EXTENSION 4

LINE

5X MR SYSTEM MODEM FORINSITE

EDM MAGNET MODEM FORMAGNET MONITORINGMONITORING

PET SYSTEM MODEM

INCOMING LINE

FAX MACHINE

Example: 4–LINE MULTIPLEXER EXTENSION CONNECTIONS With a fax machine

Illustration 1–5

Page 13: Rj 11 Multi. Insta

RJ–11 MULTIPLEXER INSTALLATIONGE MEDICAL SYSTEMS

DIRECTION 2137747

���

REV 0

1–3–3 Multiplexer Connections (continued)

3. Write down the exact configuration on the extension identification sticker (see Illustration 1–6) that came with theRJ–11 Multiplexer. Print the system ID of each room on the line for the appropriate extension.

Attach this sticker to the top of the multiplexer.

Ext. System System ID1234

extension identification sticker for a 4–LINE rj–11 multiplexerIllustration 1–6

4. Connect the incoming phone line to the LINE jack on the multiplexer.

NoteDo not connect anything to the CONTROL jack on the multiplexer.

5. Verify that the power supply has the correct voltage rating (115VAC/60Hz or 220VAC/50Hz) for the site.

6. Connect the circular connector of the power supply to the multiplexer, and plug the power supply line cord into areceptacle that meets the multiplexer’s power and grounding requirements. Refer to Section 1–3–6, Specifica-tions.

Page 14: Rj 11 Multi. Insta

RJ–11 MULTIPLEXER INSTALLATIONGE MEDICAL SYSTEMS

DIRECTION 2137747

���

REV 0 (7 June 1995)

1–3–4 Installation Checkout

Call the OnLine Center (OLC) at 800–321–7937 to complete the multiplexer installation.

CAUTION

Failure to follow this check–out procedure will cause InSite to be unable to connect to your system,which will adversely affect customer satisfaction.

Talk to the support group for EACH modality that has a system connected to the RJ–11 Multiplexer. Each OLC engi-neer needs the following information:

o The System ID for each room

o The extension jack to which each system is connected

NoteIf a system was not connected previously to InSite, you need to perform the standard InSite check–out procedure.

The OLC engineers will update each phone number database, and then attempt to manually connect to each systemto test that the installation and the database are both functioning properly.

1–3–5 Indications

The RJ–11 multiplexers have the following indicators available on the front panel to aid in troubleshooting:

INDICATOR CONDITION MEANINGPOWER ON Power on

LINE FLASHING Ring on lineON Call in program

EXTENSIONS ON Extension off hook

Page 15: Rj 11 Multi. Insta

RJ–11 MULTIPLEXER INSTALLATIONGE MEDICAL SYSTEMS

DIRECTION 2137747

���

REV 0

1–3–6 Specifications

The RJ–11 Multiplexers (46–328475P1–4) have the following specifications:

o Power

� UL and CSA Wall Mount Power Supply, 115 VAC, 60 Hz, 45W

� UL, CSA and VDE Wall mount Power Supply, 100 – 240 VAC, 50 – 60 Hz, 60W

o Physical

4 line 8 line

� Height – 2.5 inches 1.75 inches

� Width – 8 inches 16.5 inches

� Length (deep) – 6.5 inches 7.9 inches

� Weight – 2.5 pounds 6 pounds

o Connectors

� The LINE and all EXTENSIONS have RJ–11 modular connectors.

� POWER has a 5–pin DIN jack to wall mounted supply.

o Telephone Line Interface

� FCC Registration is DNM64S–15658–VP–E

� Ringer Equivalence is 0.3B

� FCC Part 15 Sub B was tested and certifed by CCL

� Canadian Models have Certification CD–4, # DOC 1794 4702 A, Family Approval for CD–8 Certificate # 8978, Load Number; 19

� Insertion Loss is less than 3 dBm.