Prices and specifications subject to change without notice. PM-909-02 Rev. B 1 RF CALIBRATION AND MEASUREMENT PRODUCTS MET/CAL ® PMX Series The PM Series calibrator simplifies the tedious and complex process of RF power sensor calibration. The goal is to realize consistent, cost effective and traceable calibrations. However, the manual approach is very demanding of even the most experienced technician. A successful calibration involves: Setting instruments Keeping track of standards Computing mismatch (Γ) Computing calibration factors (k) Computing total uncertainties Programming EPROM sensors Generating reports and labels The PM Series automates and standardizes this process while providing compatibility with a wide variety of instruments and power sensors. The PM Series is built upon the 1830A metrology grade RF Power Meter. This is the only RF Power Meter on the market that is compatible with all known types of thermistor sensors including TEGAM, Agilent, Weinschel, Hughes and Millitech. A new line of compatible RF power thermistor transfer and reference standards have also been developed that provide flexibility and expandability as your RF calibration needs grow. These standards are based on the same time proven method used by NIST, PTB, NIM and other national metrology organizations around the world. PM Series Microwave Power Calibration System • Supports Sensors from most major manufactures up to 50 GHz • Faster than direct compare method • Lowest total uncertainty • National Metrology Institute class thermistor reference standard

RF CALIBRATION AND MEASUREMENT PRODUCTS …...with an advanced set of procedures for the major manufacturers of microwave power sensors. Fully auditable and ... • Fully automated

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

  • Prices and specifications subject to change without notice.PM-909-02 Rev. B 1

    R F C A L I B R A T I O N A N D M E A S U R E M E N T P R O D U C T S M E T / C A L ®

    PMX Series

    The PM Series calibrator simplifies the tedious and complex process of RF power sensor calibration. The goal is to realize consistent, cost effective and traceable calibrations. However, the manual approach is very demanding of even the most experienced technician.

    A successful calibration involves: • Settinginstruments • Keepingtrackofstandards • Computingmismatch(Γ) • Computingcalibration factors(k) • Computingtotaluncertainties • ProgrammingEPROMsensors • Generatingreportsandlabels

    The PM Series automates and standardizes this process while providing compatibility with a wide variety of instruments and power sensors.

    The PM Series is built upon the 1830A metrology grade RF Power Meter. This is the only RF Power Meter on the market that is compatiblewith all known types of thermistorsensors including TEGAM, Agilent,Weinschel, Hughes and Millitech.

    A new line of compatible RF power thermistor transfer and reference standards have also been developed that provide flexibility and expandability as your RF calibration needs grow. These standards are based on the same time proven method used by NIST, PTB, NIM and other national metrology organizations around the world.

    PM Series Microwave Power Calibration System

    • SupportsSensorsfrommostmajor manufactures up to50GHz

    • Fasterthandirect compare method

    • Lowesttotaluncertainty

    • NationalMetrologyInstituteclass thermistor reference standard

  • 2

    R F C A L I B R A T I O N A N D M E A S U R E M E N T P R O D U C T S M E T / C A L ®

    PMX Series

    ExpandtheworkloadofMET/CAL® Calibration Software to microwave power sensors

    Correctly calibrating an RF power sensor is an involved process that requires numerous complex calculations of calibration factor, mismatch correction and uncertainty. MET/CAL® Software now supports this complex process with an advanced set of procedures for the major manufacturers of microwave power sensors. Fully auditable and transparent calculations are accomplished with an Excel®spreadsheetmakingiteasytodeploybyorganizationswhoare ISO 17025 accredited. Combined with MET/CAL® Software’s industry leading features of test and measurement asset trackingand reportgeneration, it is thepreferredoption for thosewhoneed toautomatemicrowavesensorcalibration.

    • FullyautomatedRFpowersensorcalibration

    • VNAsupportforautomatedS11parametermeasurements

    • UploadanddownloadEPROMdataofmanypopularAgilent,Anritsu and Rohde & Schwarz power sensors

    • Supportmostmajornetworksanalyzersandsignalgenerators

    • Trackassetinformationincludingcalibrationandmaintenancehistory and status, traceability, users, customers, and location.

    • PreservecalibrationhistoryintheMET/TEAM®orMET/TRACK® databases

    • DynamicUncertaintyCalculations

    • Analyzeandreportassetinformation;producecustomizedprintedcertificatesandreports.

    • Makedataavailabletoothercorporatesystems.

    • ImportassetandcalibrationdataintoMET/CAL® Software

    • HelpmeettherequirementsofqualitystandardslikeISO9000,ISO/IEC17025,NRC10CFR,ANSIZ540.3,andothers.

    26 GHz System 50 GHz System

  • 3

    R F C A L I B R A T I O N A N D M E A S U R E M E N T P R O D U C T S M E T / C A L ®

    PMX Series

    Excel Based Calculation Engine

  • 4

    R F C A L I B R A T I O N A N D M E A S U R E M E N T P R O D U C T S M E T / C A L ®

    PMX Series

    Signal Generatorsi Network Analyzersii Microwave Standards

    100 KHz to 18 GHz AnritsuMG3692CFluke96270AGigatronics2520BKeysightN5173B-520R&S SMF 100A

    AnritsuMS4642BR&SZVASeriesKeysightN5232Aiii

    HP 8510Ciii

    1830A with F1130B

    100 KHz to 26.5 GHz AnritsuMG3693CFluke96270AGigatronics2526BKeysightN5173B-532R&S SMF 100A

    AnritsuMS4644BR&SZVASeriesKeysightE5063A,N5242Aiii

    HP 8510Ciii

    1830A with F1130B1830A with F1135B

    100 KHz to 50 GHz AnritsuMG3695CGigatronics2550BKeysightE8257D-550

    AnritsuMS4645BR&SZVASeriesKeysightE5063A,N5245Aiii

    HP 8510Ciii

    1830A with F1130B1830A with 2510A

    i Any SPCI compliant signal generator can be utilized, the procedures allow for use up to two signal sources per procedure, including function generators. However, this list consists of the specifications of both power and frequency requirements.

    ii VNAcalibrationkitsconsistofOpen,Short,LowBandLoad,andSlidingLoad.iii KeysightPNASeriesandHP8510Cdonotoperatedownto100kHz.

    Uncertainty Calculation for a Power Sensor Calibration

    Example calibration: Frequency:18GHz Power: 1 mW

    Input Component Value of Input Uncertainty of Input Value

    Uncertainty Contribution to k of Input

    F1130B calibration factor .8851 mW/mW .01 mW/mW .01 mW/mW

    F1130B Rho .042V/V .03V/V .00114mW/mW

    F1130B Phi -41.8degrees 8 degrees .00111 mW/mW

    1830A Power Reading 1.0000 mW .07% of rdg .00059mW/mW

    DUT Rho .111V/V .03V/V .00014mW/mW

    DUT Phi 120 degrees 4degrees .00055 mW/mW

    DUT Power Reading 1.0050 mW .5% of rdg .00431mW/mW

    Repeatability .002 mW/mW .002 mW/mW

    RSS Uncertainty .0113 mW/mW

  • 5

    R F C A L I B R A T I O N A N D M E A S U R E M E N T P R O D U C T S M E T / C A L ®

    PMX Series

    V/V

    mW

    /mW

    MHz

    Typical Equivalent Source Match

    MHz

    Typical Cal Factors

    2510A

    F1135B

    V/V

    MHz

    Typical Equivalent Source Match

    MHz

    mW

    /mW

    Typical Cal Factors

    MHz

    Typical Cal Factors

    mW

    /mW

    MHz

    V/V

    Typical Equivalent Source Match

    F1130B

    Performance Graphs

  • 6

    R F C A L I B R A T I O N A N D M E A S U R E M E N T P R O D U C T S M E T / C A L ®

    PMX Series

    PM Series Supported SensorsSENSOR FREQUENCY RANGE POWER RANGE EEPROM

    ReadEEPROM

    Write

    ANRITSUMA2442A 10MHzto18GHz -67dBmto20dBm Yes YesMA2442D 10MHzto18GHz -67dBmto20dBm Yes YesMA2445A 10MHzto50GHz -67dBmto20dBm Yes YesMA2472A 10MHzto18GHz -70dBmto20dBm Yes YesMA2472D 10MHzto18GHz -70dBmto20dBm Yes YesMA2475A 10MHzto50GHz -70dBmto20dBm Yes YesMA2475D 10MHzto50GHz -70dBmto20dBm Yes YesMA24002A 10MHzto18GHz -30dBmto20dBm Yes YesMA24005A 10MHzto50GHz -30dBmto20dBm Yes YesMA2445D 10MHzto50GHz -67dBmto20dBm Yes Yes

    KEYSIGHT8481A 10MHzto18GHz -30dBmto20dBm NA NA8482A 100kHzto4.2GHz -30dBmto20dBm NA NA8485A 50MHzto26.5GHz -30dBmto20dBm NA NA8487A 50MHzto50GHz -30dBmto20dBm NA NAE4412A 10MHzto18GHz -70dBmto20dBm Yes YesE4413A 50MHzto26.5GHz -70dBmto20dBm Yes YesE9300A 10MHzto18GHz -60dBmto20dBm Yes YesE9301A 10MHzto6GHz -60dBmto20dBm Yes YesN1921A 50MHzto18GHz -35dBmto20dBm Yes NoN1922A 50MHzto40GHz -35dBmto20dBm Yes NoU2000A 10MHzto18GHz -60dBmto20dBm Yes NoU2001A Modified10MHzto6GHz -60dBmto25dBm Yes NoU2002A 50MHzto24GHz -60dBmto20dBm Yes No

    ROHDE & SCHWARZNRP-Z51 DCto18GHz -30dBmto20dBm Yes YesNRP-Z52 DCto27GHz -35dBmto20dBm Yes YesNRP-Z56 DCto50GHz -35dBmto20dBm Yes Yes

    TEGAMF1130 100kHzto18GHz -20dBmto13dBm NA NAF1135 10MHzto26.5GHz -20dBmto13dBm NA NA2510A 10MHzto50GHz -20dBmto14dBm NA NA