9
Research Article Stability-Indicating HPLC Determination of Trandolapril in Bulk Drug and Pharmaceutical Dosage Forms Leena A. Al-Hawash, 1,2 Ashok K. Shakya, 1 and Maher L. Saleem 1,3 1 Faculty of Pharmacy and Medical Sciences, Al-Ahliyya Amman University, P.O. Box 263, Amman 19328, Jordan 2 Arab Company for Drug Industries and Medical Appliances (ACDIMA), P.O. Box 925161, Amman 11190, Jordan 3 Faculty of Pharmacy, Middle East University, Airport Road, Amman 11831, Jordan Correspondence should be addressed to Ashok K. Shakya; [email protected] Received 30 September 2014; Revised 2 January 2015; Accepted 5 January 2015 Academic Editor: Philip M. Williams Copyright © 2015 Leena A. Al-Hawash et al. is is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. A rapid, simple, accurate, precise, economical, robust, and stability indicating reverse phase HPLC-PDA procedure has been developed and validated for the determination of trandolapril. e trandolapril was separated isocratically on Hypersil-Gold C18 column (250 mm × 4.6 mm, 5 m) with a mobile phase consisting of 50% acetonitrile and 50% water (containing 0.025% triethylamine, pH 3.0 ± 0.1), at 25 ± 2 C. Retention time of the drug was 4.6 min. e eluted compounds were monitored and identified at 210 nm. e linearity of the method was excellent ( 2 > 0.9999) over the concentration range of 1–24 g/mL; the limit of detection (LOD) and limit of quantitation (LOQ) were 0.0566 g/mL and 0.1715 g/mL, respectively. e overall precision was less than 2%. Mean recovery of trandolapril was more than 99%; no interference was found from the component present in the preparation. Stability studies indicate that the drug was stable to sunlight and UV light. e drug gives 6 different oxidative products on exposure to hydrogen peroxide. Slight degradation was observed in acidic condition. Degradation was higher in the alkaline condition compared to other conditions. e robustness of the method was studied using factorial design experiment. 1. Introduction Trandolapril (Figure 1) is a long-lasting angiotensin converting enzyme (ACE) inhibitor which was approved by the U.S. Food and Drug Administration for lowering blood pressure in doses up to 2 mg even aſter discontinuation of treatment [1]. It is also used for patient with evidence of Leſt Ventricular (LV) systolic dysfunction aſter Acute Myo- cardial Infarction (AMI). It can be given safely over a prolonged period of time [2]. Trandolapril is the inter- national nonproprietary name of (2S, 3aR, 7aS)-1-[(2S)-2- {[(2S)-1-ethoxy-1-oxo-4-phenylbutan-2-yl]amino}propan- oyl]–octahydro-1H–indole-2-carboxylic acid. Trandolapril is rapidly absorbed and metabolized to its biologically active diacid form, trandolaprilate, in liver which shows high lipophilicity compared to other ACE inhibitors. Determina- tion of trandolapril alone has been analyzed less oſten than other drugs; indeed, only few methods involving amper- ometric biosensors [3], a potentiometric enantioselective membrane electrode [4], liquid chromatography tandem mass spectrometric [5], HPLC [613], HPTLC [14, 15], UV spectroscopy [16], and capillary electrophoresis [17] have been used. Trandolapril is official in British Pharmacopoeia and it is estimated by nonaqueous titration [18]. A few stability-indicating HPLC methods [1113] have been reported, which provides variable level of degradation of trandolapril. Stability-indicating method reported by Manju Latha and Gowri Sankar [11] does not produce any degraded product in different stressed conditions, although it is well documented that trandolapril is susceptible to hydrolysis. Impurity profiles of trandolapril under stress (acidic and neutral) conditions studied by Dendeni et al. [12] require sophisticated LC-MS-MS instrument. Sahu et al. [13] have reported the validated stability-indicating method which can separate the hydrolytic degraded product of trandolapril. However none of the HPLC method reported the oxidative degraded product of trandolapril. International Conference on Harmonization (ICH) guidelines [19] require perfor- mance of stress tests on the drug substance, which can help to identify the likely degradation products. Moreover, validated Hindawi Publishing Corporation International Journal of Analytical Chemistry Volume 2015, Article ID 820517, 8 pages http://dx.doi.org/10.1155/2015/820517

Research Article Stability-Indicating HPLC Determination of ...Arab Company for Drug Industries and Medical Appliances (ACDIMA), P.O. Box , Amman , Jordan Faculty of Pharmacy, Middle

  • Upload
    others

  • View
    0

  • Download
    0

Embed Size (px)

Citation preview

  • Research ArticleStability-Indicating HPLC Determination of Trandolapril inBulk Drug and Pharmaceutical Dosage Forms

    Leena A. Al-Hawash,1,2 Ashok K. Shakya,1 and Maher L. Saleem1,3

    1Faculty of Pharmacy and Medical Sciences, Al-Ahliyya Amman University, P.O. Box 263, Amman 19328, Jordan2Arab Company for Drug Industries and Medical Appliances (ACDIMA), P.O. Box 925161, Amman 11190, Jordan3Faculty of Pharmacy, Middle East University, Airport Road, Amman 11831, Jordan

    Correspondence should be addressed to Ashok K. Shakya; [email protected]

    Received 30 September 2014; Revised 2 January 2015; Accepted 5 January 2015

    Academic Editor: Philip M. Williams

    Copyright © 2015 Leena A. Al-Hawash et al. This is an open access article distributed under the Creative Commons AttributionLicense, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properlycited.

    A rapid, simple, accurate, precise, economical, robust, and stability indicating reverse phase HPLC-PDA procedure has beendeveloped and validated for the determination of trandolapril. The trandolapril was separated isocratically on Hypersil-GoldC18 column (250mm × 4.6mm, 5 𝜇m) with a mobile phase consisting of 50% acetonitrile and 50% water (containing 0.025%triethylamine, pH 3.0 ± 0.1), at 25 ± 2∘C. Retention time of the drug was ∼4.6min. The eluted compounds were monitored andidentified at 210 nm. The linearity of the method was excellent (𝑟2 > 0.9999) over the concentration range of 1–24 𝜇g/mL; thelimit of detection (LOD) and limit of quantitation (LOQ) were 0.0566𝜇g/mL and 0.1715 𝜇g/mL, respectively. The overall precisionwas less than 2%. Mean recovery of trandolapril was more than 99%; no interference was found from the component present inthe preparation. Stability studies indicate that the drug was stable to sunlight and UV light. The drug gives 6 different oxidativeproducts on exposure to hydrogen peroxide. Slight degradation was observed in acidic condition. Degradation was higher in thealkaline condition compared to other conditions. The robustness of the method was studied using factorial design experiment.

    1. Introduction

    Trandolapril (Figure 1) is a long-lasting angiotensinconverting enzyme (ACE) inhibitor which was approved bythe U.S. Food and Drug Administration for lowering bloodpressure in doses up to 2mg even after discontinuation oftreatment [1]. It is also used for patient with evidence ofLeft Ventricular (LV) systolic dysfunction after Acute Myo-cardial Infarction (AMI). It can be given safely over aprolonged period of time [2]. Trandolapril is the inter-national nonproprietary name of (2S, 3aR, 7aS)-1-[(2S)-2-{[(2S)-1-ethoxy-1-oxo-4-phenylbutan-2-yl]amino}propan-oyl]–octahydro-1H–indole-2-carboxylic acid. Trandolapril israpidly absorbed and metabolized to its biologically activediacid form, trandolaprilate, in liver which shows highlipophilicity compared to other ACE inhibitors. Determina-tion of trandolapril alone has been analyzed less often thanother drugs; indeed, only few methods involving amper-ometric biosensors [3], a potentiometric enantioselectivemembrane electrode [4], liquid chromatography tandem

    mass spectrometric [5], HPLC [6–13], HPTLC [14, 15], UVspectroscopy [16], and capillary electrophoresis [17] havebeen used. Trandolapril is official in British Pharmacopoeiaand it is estimated by nonaqueous titration [18]. A fewstability-indicating HPLC methods [11–13] have beenreported, which provides variable level of degradation oftrandolapril. Stability-indicating method reported by ManjuLatha and Gowri Sankar [11] does not produce any degradedproduct in different stressed conditions, although it is welldocumented that trandolapril is susceptible to hydrolysis.Impurity profiles of trandolapril under stress (acidic andneutral) conditions studied by Dendeni et al. [12] requiresophisticated LC-MS-MS instrument. Sahu et al. [13] havereported the validated stability-indicating method which canseparate the hydrolytic degraded product of trandolapril.However none of the HPLC method reported the oxidativedegraded product of trandolapril. International Conferenceon Harmonization (ICH) guidelines [19] require perfor-mance of stress tests on the drug substance, which can help toidentify the likely degradation products. Moreover, validated

    Hindawi Publishing CorporationInternational Journal of Analytical ChemistryVolume 2015, Article ID 820517, 8 pageshttp://dx.doi.org/10.1155/2015/820517

  • 2 International Journal of Analytical Chemistry

    OO

    NH

    O

    NH

    H

    O

    OH

    Figure 1: Structure of trandolapril.

    stability-indicatingmethods should be applied in the stabilitystudies [20] once they have demonstrated their suitability fortheir intended purpose. Thus, stability-indicating methodshave to demonstrate that they are specific, which involvesevaluating the drug in the presence of its degradationproducts. The aim of the present work was the developmentand validation of an HPLC stability-indicating method fordetermining trandolapril in its pharmaceutical form follow-ing ICH recommendations to achieve this goal; a stress studyof the drug was performed in order to validate the stability-indicating power of the developed analytical method and toidentify the key factors that will impact the stability of thedrug product. The robustness of the developed method wasstudied using design of experiment utilising factorial design.

    2. Experimental

    2.1. Chemicals and Reagents. The reference sample of tran-dolapril (99.8%, TD0131207) was obtained as a gift fromHetero Pharmaceutical Ltd, Hyderabad, India. The marketedformulations of drug (Odrik hard gelatin capsules, strength2mg, manufactured by Abbott GmbH & Co KL, Germany)were purchased from local pharmacy. All reagents were ofanalytical grade unless stated otherwise. Reverse osmosisquality water (purified with a Milli-RO plus Milli-Q stationMillipore Corp., USA) and HPLC quality water were usedthroughout. Acetonitrile and methanol (HPLC grade) weresupplied by Panreac (Barcelona, Spain).

    2.2. HPLC Instrumentation and Conditions. Analysis wasperformed with a Shimadzu Prominence liquid chromato-graph equipped with LC-20AD UFLC quaternary solventdelivery system, SIL-20A Prominence autosampler having auniversal loop injector of capacity 1–100𝜇L, and an SPD-M20A diode array detector monitored between 200 and350 nm and CBM-20A, Communication Bus module. Theequipment was controlled by a Windows 7 based LC-Solution version 1.25 (2009-2010) work station software.Thermo Hypersil-Gold C-18 column (250mm × 4.6mm i.d.;5 𝜇m) was used. The mobile phase was acetonitrile : water(50 : 50 v/v, containing triethylamine (TEA) 250𝜇L/L, final

    pH adjusted to 3.0with orthophosphoric acid (OPA)).Mobilephase was degassed using ultrasonic bath and sample solu-tions were filtered through 0.45𝜇m filters prior to analysis.Mobile phase flow rate was 1.0mL/min. All the analysis wascarried out at 25 ± 2∘C. Retention times, UV spectrum, andpeak purity were used to identify trandolapril.

    2.3. Preparation of Stock and Standards Solutions

    2.3.1. Stock andWorking Solutions. Trandolapril (active phar-maceutical ingredient (API), equivalent to 100mg of tran-dolapril) was weighed and transferred to 100mL calibratedvolumetric flask quantitatively. It was dissolved in acetonitrile(20mL) with the aid of sonication. The final volume wasmade up to the mark with acetonitrile : water (50 : 50% v/v)to produce stock solution (1000 𝜇g/mL). Working solutionsof trandolapril (25, 100, and 200𝜇g/mL) were prepared usingsuitable aliquots of stock or intermediate solution. All thesolutions were stored in refrigerator.

    2.3.2. Calibration Standards. Calibration standards were pre-pared freshly using the intermediate working solutions oftrandolapril. Standard solution of concentrations 1, 2, 4, 8,10, 12, 16, 20, and 24𝜇g/mL was prepared. These solutionswere analyzed immediately to avoid degradation and as perschedule.

    2.3.3. Quality Control Samples. Quality control samples atthree concentrations (4, 12, and 22 𝜇g/mL) level were pre-pared separately as low quality control (LQC), mediumquality control (MQC), and high quality control (HQC).

    2.3.4. Preparation of Sample for Assay. Average weight oftwenty capsules content (each containing 2mg trandolapril)was determined. A quantity of powder equivalent to 10mg oftrandolapril was weighed and transferred to 50mL calibratedvolumetric flask. Acetonitrile (10mL) was added to the sameflask and sonicated for 1 minute. The volume was made upto 50mL with acetonitrile : water (50 : 50 v/v) solution. Thetheoretical concentration of the stock solution of trandolaprilwas 200𝜇g/mL. The solution was filtered using 0.45 𝜇mnylon filter (Microsyringe filter). Appropriate dilutions wereprepared for analysis.

    2.4. Analytical Method Validation

    2.4.1. Linearity, Limit of Detection (LOD), and Limit of Quan-titation (LOQ). Appropriate volumes of trandolapril stockstandard solution (1000mg/mL) was diluted with mobilephase to produce concentrations of 1, 2, 4, 8, 10, 12, 16, 20,and 24𝜇g/mL. Each day two different sets of calibrationstandard were independently prepared and analyzed. Sixdifferent calibration curves were prepared on three differentdays. To define the correlation between the response andconcentration of trandolapril, the area was plotted againstconcentration of trandolapril with weighting factor 𝑥, 1/𝑥,or 1/𝑥2. The method was evaluated by determination of thecorrelation coefficient and intercept values.The linear best fit

  • International Journal of Analytical Chemistry 3

    line (weighting factor 𝑥) was used to measure the concen-tration of all samples throughout the batch. The acceptancecriterion for each back calculated concentration was less than2% from nominal values except for LOQ. Microsoft OfficeExcel 2007 was used for statistical analysis. The method wasvalidated according to ICH guidelines of the validation ofanalytical methods [19, 20]. A 5% significance level was usedfor evaluation. LOD and LOQ were determined from thecalibration function.

    2.4.2. System Suitability. System suitability parameters weretested with six replicate injections of the diluted sample ofworking standards (10 𝜇g/mL) at the start of the project and atthe end of the project.The system suitability parameters werecalculated using the internal features of LC-Solution softwareas per United States Pharmacopoeia [21]. The parameterswere retention time, peak area and height, width at halfpeak height, tailing factor, efficiency, and height equivalenttheoretical plate (HETP). System suitability was measuredon the basis of precision (RSD). The precision, as measuredby coefficient of variation was determined at each set’sparameters and it should be less than 2% at the beginning ofvalidation and at the end of validation.

    2.4.3. Precision. Precision was measured using triplicatedetermination of quality control samples of 4 𝜇g/mL (LQC),12 𝜇g/mL (MQC), and 22𝜇g/mL (HQC) of trandolapril, onthree different occasions (0, 3, and 6 h) and different days.The precision (RSD) of the method was determined asintraday precision (repeatability) and intermediate precision.The intermediate precisionwas estimated from theRSDof theanalysis of the samples prepared at the same concentrationbut on 3 different days at different concentration levels, whileintraday precision was calculated by analyzing the sameconcentration during the same day at different time.

    2.4.4. Accuracy. Accuracy (as percentage recovery) was mea-sured using replicate sample of trandolapril prepared usingcapsule matrix. Different samples (𝑛 = 3, at each level of80%, 100%, and 120%) were prepared using capsules content(2mg as 100%) and adding known quantity of trandolapril (at80%–120% level). From these fortified samples, appropriatesample solutions were prepared and analyzed and the totalamount recovered was calculated. Accuracy was calculatedby comparing with true value. The concentrations wereback-calculated by regression equation 𝑦 = 21225𝑥 + 1303(weighting factor 𝑥).

    2.4.5. Robustness. It is a measure of reproducibility of testresults under normal, expected, operational condition fromanalyst to analyst. The robustness of the method was eval-uated on the basis of precision, as measured by percentcoefficient of variation (% CV or RSD), determined as eachconcentration level was required not to exceed 2%. Designof experiments (DOE) was used to study robustness of themethod. A 24 factorial design was used to test the robustnessof chromatographic separation. The experimental design isuseful for this kind of study as it facilitates the investigation of

    several parameters by reducing the number of experiments.Acetonitrile content of the mobile phase, volume of peakmodifier, pH, and flow rate were investigated. Upper andlower limits are shown in Table 5. The experiments were runrandomly and the selected responses were retention time(𝑇𝑟), tailing factor (𝑇

    𝑓), and area count.

    2.4.6. Stability Studies. Stress study like oxidative stress,alkaline stress, acidic stress, exposure to sunlight, and UVlight (254 nm) were carried out using trandolapril rawmaterial. Chromatograms were recorded in order to studythe specificity of the method. The chromatograms of thesamples were compared with those of control samples thatwere freshly prepared from the stock standard solution andwithout stress. All samples were analyzed in triplicate. Thepeak purity was checked using the tools of the LC-Solutionsoftware. This assessment was based on the comparison ofspectra recorded during the elution of the peak. UV spectraand peak purity were used to assess purity of trandolapril.

    (1) Oxidative Stress. Trandolapril (5mg) was weighed accu-rately and transferred to 50mL volumetric flask. 5mL of30% hydrogen peroxide was added to it. It was stirred forone hour, and then the contents were diluted to 50mL withmobile phase. Replicate solutions of concentration 20 𝜇g/mLwere prepared and chromatograms of these solutions wererecorded and compared with the chromatograms obtainedfrom the fresh solution of trandolapril having the sameconcentration and the chromatogram of the blank (solutioncontaining only hydrogen peroxide).

    (2) Effect of Acid and Alkaline Media. Trandolapril (5mg)was weighed accurately and transferred to 50mL volumetricflask. It was shaken for one hour with 5mL of either 0.1Mhydrochloric acid (HCl) or 0.1M sodium hydroxide (NaOH).After one hour the content was diluted to 50mL withmobile phase. Replicate solutions of concentration 20 𝜇g/mLwere prepared and chromatograms of these solutions wererecorded and compared with the chromatograms obtainedfrom the fresh solution of trandolapril having the sameconcentration and the chromatogram of the blank.

    (3) Effect of UV Light or Sunlight. Trandolapril API (100mg)was placed in an open watch glass and exposed to eitherUV-irradiation (∼100W/m2) or direct sunlight for two hourswith occasionally shifting the content using stainless steelspatula. After two hours 10mg of trandolapril was weighedand transferred to 10mL volumetric flask. It was dissolved inmobile phase. 1mL of the prepared solution was transferredto a 50mL volumetric flask and diluted to the mark withmobile phase. Chromatograms were recorded and comparedwith the chromatogram of unexposed API.

    (4) Stock Stability.Thestability of stock solutionwas evaluatedat zero time and stored in the refrigerator (2–8∘C). Sampleswere prepared and analysed at days 0, 7, 14, and 21.

  • 4 International Journal of Analytical Chemistry

    (min)

    (mAU

    )

    0.0 2.5 5.0 7.5 10.0

    0

    25

    50

    PDA multi 1

    1 PDA multi 1/210nm 4nm

    4.272/trandolapril

    Figure 2: Representative chromatogram showing signal of tran-dolapril in the selected mobile phase.

    3. Results and Discussion

    3.1. Analytical Method Development. TheHPLC method wasdeveloped as a stability-indicatingmethod to determine tran-dolapril in the presence of the possible degradation products(trandolaprilate) of the drug. Therefore, a retention timebetween 4 and 5min was chosen for the drug since it allowedboth a rapid determination of the drug, which is importantfor routine analysis, and a complete drug separation. Duringdevelopment step various mobile phases of water :methanolor water : acetonitrile (35 : 65, 40 : 60, 45 : 55, 50 : 50, 55 : 45,60 : 40, and 65 : 35 v/v, with or without peak modifier, pH 3,5, or 7) were tried and the responses were recorded. On thebasis of responses and chromatographic parameters studieda mobile phase of acetonitrile: water (50 : 50 v/v, containing0.25mL/L TEA, final pH adjusted to 3.0 with OPA) wasselected as suitable mobile phase, which can separate thedrug and degraded products. Under these conditions thedrug was eluted at ∼4.3min at ambient temperature (25 ±2∘C). The absorption maximum of the drug (𝜆 210 nm) wasselected for detection, as there was no interference fromexcipients present in drug. Figure 2 depicts the representativechromatogram obtained with the present method.

    3.2. Method Validation. The method was validated withrespect to parameters including linearity, LOQ, LOD, preci-sion, accuracy, specificity, robustness, system suitability, andstability.

    3.2.1. Linearity. Different calibration curves (𝑛 = 6) wereconstructed for trandolapril was linear over the concentrationrange of 1–24𝜇g/mL. Peak area of trandolapril was plot-ted versus trandolapril concentration and linear regressionwas performed using LC-Solution software and MicrosoftOffice Excel 2007. Different calibration curves were pre-pared on different days. The mean regression equation fortrandolapril was found to be 𝑦 = 21121 (±157.1)𝑥 +1125 (±511.6)(weighting factor 𝑥, Table 1). The regressioncoefficient was 0.9999 or higher. The linearity range reportedin other methods ranged between 4 and 150𝜇g/mL [6–13].

    3.2.2. LOD and LOQ. The LOD and LOQ values were 0.0566and 0.1716 𝜇g/mL calculated using calibration curve as per

    Table 1: Linearity of the present method.

    Conc. 𝜇g/mL Mean area∗ SD∗ RSD1 22,044 364.7 1.652 41,605 532.0 1.284 86,309 1194.8 1.388 172,070 1699.5 0.9910 212,583 2283.2 1.0712 255,392 2189.3 0.8616 336,841 3371.1 1.0020 424,233 3825.2 0.9024 507,842 3534.7 0.70Slope 21,121.5 157.1Intercept 1,125.3 511.6𝑟 0.99996∗Mean and SD of six determinations.

    Table 2: Accuracy of the method.

    Amounttaken (mg)

    Amount added % Recovery(Mean ± SD) (𝑛 = 3) % CV% (mg)

    2 80 1.6 99.8 ± 0.29 0.292 100 2.0 99.3 ± 0.25 0.252 120 2.4 100.2 ± 0.24 0.24

    Table 3: Precision study of the proposed method.

    Concentration(𝜇g/mL)

    Intraday precision Interday precisionConc. foundMean ± SD % CV

    Conc. foundMean ± SD % CV

    4 3.937 ± 0.068 1.74 4.008 ± 0.039 0.9812 11.949 ± 0.088 0.74 12.011 ± 0.108 0.9022 21.853 ± 0.102 0.47 21.731 ± 0.151 0.70

    ICH guideline. The LOD and LOQ reported by Sahu et al.[13] were 0.1 and 0.8 𝜇g/mL, based on signal to noise ratiomethod.

    3.2.3. Accuracy and Precision. The accuracy and precision ofthe analytical method were established across its linear rangeas indicated in the guideline. As shown from the data inTable 2, excellent recoveries (99.3 to 100.2%) were obtainedat different added concentration level. The results obtainedfor the intraday and interday precision of the method wereexpressed as RSD values. As shown in the table, the intradayand interday RSD was < 2.0% for all concentrations tested indifferent situations studied (Table 3).

    3.2.4. Specificity. Specificity of the method was assessedby comparing the chromatograms obtained from capsulecontent and drug standards.The retention times of drug fromstandard solutions and from capsule content were identicaland no coeluting peaks from the diluents were observed,indicating specificmethod for quantitative estimation of drugin the commercial formulation.

  • International Journal of Analytical Chemistry 5

    Table 4: System suitability parameters for trandolapril.

    SN 𝑇𝑟(min) Area HeightConc.

    (𝜇g/mL) Accuracy %Tailingfactor

    Theoreticalplate

    USPwidth HETP

    Mean 4.28 210594 32575 10.06 100.6 1.12 8326 0.190 30.03SD 0.003 737.05 118.93 0.03 0.35 0.00 92.54 0.00 0.33RSD 0.07 0.35 0.37 0.35 0.35 0.12 1.11 0.55 1.11Values represent mean and SD of six determinations.

    Table 5: Factorial design data for robustness of chromatographic separation.

    (a)

    Selected parameters and their variations −1 (lower limit) +1 (upper limit)Acetonitrile in mobile phase (%) (𝐴) 45 55Peak modifier (concentration of TEA) (𝐵) 200 300Final pH of the mobile phase (𝐶) 2.75 3.25Flow rate (mL/min) (𝐷) 0.9 1.1

    (b)

    Exp. number Run order Factors Responses𝐴 𝐵 𝐶 𝐷 𝑇

    𝑟

    𝑇𝑓

    Area1 13 45 200 2.75 0.9 5.2 1.12 2119452 14 55 200 2.75 0.9 4.3 1.09 2084103 8 45 300 2.75 0.9 5.1 1.08 2095994 10 55 300 2.75 0.9 4.2 1.06 2084775 3 45 200 3.25 0.9 5.1 1.12 2081046 2 55 200 3.25 0.9 4.4 1.09 2096487 1 45 300 3.25 0.9 5.1 1.07 2010458 7 55 300 3.25 0.9 4.8 1.11 2081549 12 45 200 2.75 1.1 4.7 1.10 20845010 11 55 200 2.75 1.1 4.1 1.07 20764911 5 45 300 2.75 1.1 4.5 1.07 20847812 4 55 300 2.75 1.1 4.2 1.10 20845113 9 45 200 3.25 1.1 4.6 1.10 21045014 16 55 200 3.25 1.1 4.05 1.06 20714515 15 45 300 3.25 1.1 4.6 1.10 20945716 6 55 300 3.25 1.1 4.1 1.07 207124𝑇𝑟= retention time of drug, 𝑇

    𝑓= tailing factor for drug, and area count.

    3.2.5. System Suitability. System suitability parameters werestudied with six replicates standard solution of the drug andthe calculated parameters are within the acceptance criteria.The tailing factor, the number of theoretical plates, andHETPwere in the acceptable limits (RSD less than 2%). The systemsuitability results are shown in Table 4.

    3.2.6. Robustness. Robustness of the methods was illustratedby getting the resolution factor and tailing factor, whenmobile phase acetonitrile content (±5%), pH (±0.25 units),peak modifier (±0.05mL/L), and flow rate (±0.1mL/min)were deliberately varied. It was studied using factorial designexperiment using Design Expert software version 8.0 (StatEase Inc, USA).The deliberate changes in the method do notaffect the retention time, tailing factor, and area count fordrug significantly. The scaled and centered coefficient plots

    for the above responses revealed that different parametersdid not affect responses significantly, so that the developedmethod was considered rugged and robust. Results arepresented in Figure 3 and Table 5.

    3.2.7. Stability Studies. The prepared stock and sampleswere stable up to 21 days when stored in refrigerator (2–8∘C) and did not produce degraded compounds duringexperimental conditions. The peak purity was 0.985 or moreduring the validation studies. On exposure to hydrogenperoxide (30%), trandolapril produces six major degradationproducts having retention time 3.69, 4.79, 5.53, 5.75, 10.85,and 14.74min. The percentage of unoxidized trandolaprilwas 40.9% (Figure 4(a), Table 6). Figure 4(b) representsthe chromatogram, peak purity, and UV spectra of thefreshly prepared sample. After exposure to 0.1M NaOH

  • 6 International Journal of Analytical Chemistry

    0.15

    0.1

    0.05

    0

    −0.05

    −0.1

    −0.15

    −0.2

    −0.25

    −0.3

    −0.35

    −0.4

    A B C D

    A∗B

    A∗C

    B∗C

    B∗D

    C∗D

    A∗D

    (a)

    0.02

    0.015

    0.01

    0.005

    0

    −0.005

    −0.01

    −0.015

    A B C D A∗B

    A∗C

    A∗D

    B∗C

    B∗D

    C∗D

    (b)A B C D

    A∗D

    B∗C

    C∗D

    1500

    1000

    500

    0

    −500

    −1000

    −1500

    A∗B

    B∗D

    A∗C

    (c)

    Figure 3: Scaled and centered coefficient of variation (%) of (a) retention time (𝑇𝑟

    ), (b) tailing factor of drug (𝑇𝑓

    ), and (c) area count.

    trandolapril gives 3 degradation product with retention time2.943 (relative percentage 54.12%), 3.177 (2.22%), and 3.478(43.66%).This indicates that alkaline conditions facilitate theconversion of trandolapril to different degraded compounds(Figure 4(c)). Results also indicate trandolapril degradesafter exposure to 0.1M HCl and forms degradation producthaving retention time 3.920 and 6.129min. The percentageof undegraded trandolapril was 99.6% (Figure 4(d)). Nodegradation products were produced on exposure to the UVlight or sunlight, which indicates that trandolapril has highstability under these stressed conditions.

    LC-MS-MS identification of different degraded productand impurity profile of trandolapril under acidic and neutralconditions has been reported [12]. Sahu et al. [13] studiedthe hydrolytic decomposition of trandolapril (at a drug con-centration of 2mg/mL) under different conditions (acidic,alkaline, or neutral) at 80∘C. The degradation of trandolaprilwas 50 and 65% under acidic and alkaline conditions,respectively. None of these studies (HPLC) reported theoxidative degradation of trandolapril. Our results reveal thatthe trandolapril is also susceptible to the oxidation. Vikas etal. [15] reported the two oxidative products of trandolaprilusing developed and validated HPTLC method.

    3.2.8. Assay. The proposed method was applied to the deter-mination trandolapril in capsule formulations. The results ofthese assays yielded 99.2% (RSD = 0.89%) of label claimed.

    Table 6: Stability data under different stressed conditions.

    Stress conditions % TrandolaprilremainedRelative percentage ofdegraded products

    Oxidative stress(30% H2O2)

    40.9 (peak #2)13.3 (peak #1), 3.38 (3),

    10.05 (4), 26.65 (5), 3.35 (6),and 8.05 (7)

    Acidic(0.1 N HCl) 99.6 (peak #2)

    0.2 (peak #1) and 0.2(peak 3)

    Alkaline(0.1 N NaOH) 0.0

    54.12 (peak #1), 2.22 (2),and 43.66 (3)

    Ultraviolet light(2 hours, 80W) 100.0 0.0

    Direct sunlight 100.0 0.0Aqueous stability(after 21 days) 99.5 ± 0.1 0.0

    Table 7: Assay of marketed pharmaceutical formulation and API.

    Drug/formulation Present method BP [18]% Assay RSD % Assay RSD

    Capsule 99.2 0.89 98.3 0.95API 98.9 0.1 98.2 1.2Student’s 𝑡-test indicates no significant difference (𝑃 > 0.05).

    Low value of precision indicates that the method can be usedprecisely for the estimation of drug in formulations (Table 7).

  • International Journal of Analytical Chemistry 7

    (min)(m

    AU)

    0.0 2.5 5.0 7.5 10.0 12.5 15.0 17.5 20.0

    0

    25

    50

    75

    100PDA multi 1

    1 PDA multi 1/210nm 4nm

    3.693

    /4.275

    /tran

    dola

    pril

    4.799

    /5.529

    /5.746

    /

    10.853

    /

    14.738

    /

    (a)

    Chromatogram

    Contour plot

    (𝜇V

    )

    0

    10000

    20000

    3.693/

    4.275

    /tran

    dola

    pril

    4.799/

    5.529/

    5.746/

    (min)0.0 2.5 5.0 7.5 10.0

    1 PDA multi 1/210nm 4nm

    Purity

    (min)4.44.34.24.1

    0.0

    0.5

    1.0

    0

    5

    10

    Spectrum

    (nm)

    (mAU

    )0

    5

    10

    200 210 220 230 240 250 260 270 280 290

    227252 275222 269234

    200

    220

    240

    260

    280

    (nm

    )

    0 2 4 6 8

    (min)

    0

    100

    (b)

    (mAU

    )

    0

    25

    50

    75

    100PDA multi 1

    (min)0.0 2.5 5.0 7.5 10.0 12.5 15.0 17.5 20.0

    1 PDA multi 1/210nm 4nm

    2.943

    /13.177

    /23.478

    /34.200

    /tran

    dola

    pril

    (c)

    (mAU

    )

    0

    10

    20

    30PDA multi 1

    (min)0.0 2.5 5.0 7.5 10.0 12.5 15.0 17.5 20.0

    1 PDA multi 1/210nm 4nm

    3.926

    / 4.275

    /tran

    dola

    pril

    6.129

    /

    (d)

    Figure 4: Typical HPLC chromatogram of trandolapril exposed to (a) oxidative stress, (b) enlarged view of chromatogram a showing contourplot, chromatogram, UV spectra, and peak purity, (c) alkaline stress, and (d) acidic stress condition showing degraded product.

  • 8 International Journal of Analytical Chemistry

    4. Conclusion

    A validated HPLC method has been developed for determi-nation of trandolapril in formulations.The proposedmethodis simple, economical, accurate, precise, specific, robust, andstability-indicating. Robustness of chromatographic methodwas studied using design of experiments indicating robustand ruggedmethod of analysis can be easily and convenientlyadopted for the routine analysis of trandolapril in pharma-ceutical dosage form and bulk drug.

    Conflict of Interests

    The authors declare that there is no conflict of interestsregarding the publication of this paper.

    Acknowledgments

    The authors wish to thank the Dean of Faculty of Pharmacyand Medical Sciences and the Dean of Higher Education andResearch, Al-Ahliyya Amman University, Amman, Jordan,for providing them with necessary facilities.

    References

    [1] D. R. P. Guay, “Trandolapril: a newer angiotensin-convertingenzyme inhibitor,” Clinical Therapeutics, vol. 25, no. 3, pp. 713–775, 2003.

    [2] L. Køber, C. Torp-Pedersen, J. E. Carlsen et al., “A clinical trialof the angiotensin-converting-enzyme inhibitor trandolaprilin patients with left ventricular dysfunction after myocardialinfarction,” The New England Journal of Medicine, vol. 333, no.25, pp. 1670–1676, 1995.

    [3] R.-I. Stefan, J. F. van Staden, and H. Y. Aboul-Enein, “Detec-tion of S-enantiomers of cilazapril, pentopril and trandolaprilusing a potentiometric, enantioselective membrane electrode,”Electroanalysis, vol. 11, no. 3, pp. 192–194, 1999.

    [4] R.-I. Stefan, J. F. van Staden, and H. Y. Aboul-Enein, “Analysisof chiral drugs with enantioselective biosensors. An overview,”Electroanalysis, vol. 11, no. 16, pp. 1233–1235, 1999.

    [5] C. Pistos, M. Koutsopoulou, and I. Panderi, “Liquid chro-matographic tandem mass spectrometric determination oftrandolapril in human plasma,” Analytica Chimica Acta, vol.540, no. 2, pp. 375–382, 2005.

    [6] A. Gumieniczek and H. Hopkala, “High-performance liquidchromatographic assay of trandolapril in capsules,” Acta Polo-niae Pharmaceutica—Drug Research, vol. 57, no. 4, pp. 253–255,2000.

    [7] A. Gumieniczek and H. Hopkala, “Development and validationof a liquid chromatographic method for the determinationof trandolapril and verapamil in capsules,” Journal of LiquidChromatography and Related Technologies, vol. 24, no. 3, pp.393–400, 2001.

    [8] C. Rambabu, G. Ramu, A. Biksham Babu, and S. Venkata Rao,“Optimized and validated reverse phase HPLC method forthe determination of trandolapril in bulk and formulations,”Rasayan Journal of Chemistry, vol. 3, no. 4, pp. 777–782, 2010.

    [9] I. Cendrowska, K. Bańkowski, and J. Iskra-Jopa, “A studyon the stereochemical purity of trandolapril and octahydro-1H-indole-2-carboxylic acid by HPLC method,” Acta PoloniaePharmaceutica—Drug Research, vol. 60, no. 2, pp. 141–144, 2003.

    [10] E. R. Dawud and A. K. Shakya, “HPLC-PDA analysis ofACE-inhibitors, hydrochlorothiazide and indapamide utilizingdesign of experiments,” Arabian Journal of Chemistry, 2014.

    [11] Y. B.Manju Latha andD.Gowri Sankar, “Stability indicatingRP-HPLC method for determination of Trandolapril (TP) in pureand pharmaceutical formulation,” International Journal of DrugDevelopment and Research, vol. 5, no. 4, pp. 98–105, 2013.

    [12] M. Dendeni, N. Cimetiere, A. Amrane, and N. B. Hamida,“Impurity profiling of trandolapril under stress testing: struc-ture elucidation of by-products and development of degrada-tion pathway,” International Journal of Pharmaceutics, vol. 438,no. 1-2, pp. 61–70, 2012.

    [13] K. Sahu, C. Karthikeyan, N. S. H. N. Moorthy, and P. Trivedi,“A validated UPLC method used for the determination of tran-dolapril and its degradation products as per ICH guidelines,”Current Pharmaceutical Analysis, vol. 7, no. 3, pp. 182–188, 2011.

    [14] D. Kowalczuk, “Simultaneous high-performance thin-layerchromatography densitometric assay of trandolapril and ver-apamil in the combination preparation,” Journal of AOACInternational, vol. 88, no. 5, pp. 1525–1529, 2005.

    [15] R. J. R. Vikas, L. Sathiyanarayanan, and S. S. Yadav, “Stabilityindicating HPTLC method for trandolapril estimation in thebulk drug and tablet dosage form,” Indian Journal of Pharma-ceutical Education andResearch, vol. 44, no. 4, pp. 341–344, 2010.

    [16] S. K. Dubey, S. Kumar, R. J. Mudakavi, S. Deshpande, and A. K.Jain, “Development and validation of UV-spectrophotometricmethod for determination of Enalapril maleate,” InternationalJournal of Advances in Pharmaceutical Sciences, vol. 1, no. 4, pp.375–380, 2010.

    [17] M.-E. Capella-Peiró, M. Font-Rubert, L. Àlvarez-Rodŕıguez, J.Esteve-Romero, A. Durgbanshi, and D. Bose, “Optimizationusing a factorial design for the separation of trandolapriland verapamil by capillary electrophoresis,” Journal of LiquidChromatography and Related Technologies, vol. 30, no. 20, pp.2975–2988, 2007.

    [18] British Pharmacopoeia, 2008, http://www.sinoapi.com/Phar-macopoeia/pharmacopoeia-bp2009.asp?cas=87679-37-6.

    [19] IFPMA, “Stability ICH testing of new drug substances andproducts Q1A (R2),” in Proceedings of the International Confer-ence on Harmonization, IFPMA, Geneva, Switzerland, 2003.

    [20] IFPMA, “ICH validation of analytical procedures: text andmethodology Q2 (R1),” in Proceedings of the InternationalConference on Harmonization, IFPMA, Geneva, Switzerland,2005.

    [21] United States Pharmacopoeia, “United states pharmacopeialconvention,” Tech. Rep. USP 27, NF 22, United States Pharma-copoeia, Rockville, Md, USA, 2004.

  • Submit your manuscripts athttp://www.hindawi.com

    Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

    Inorganic ChemistryInternational Journal of

    Hindawi Publishing Corporation http://www.hindawi.com Volume 2014

    International Journal ofPhotoenergy

    Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

    Carbohydrate Chemistry

    International Journal of

    Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

    Journal of

    Chemistry

    Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

    Advances in

    Physical Chemistry

    Hindawi Publishing Corporationhttp://www.hindawi.com

    Analytical Methods in Chemistry

    Journal of

    Volume 2014

    Bioinorganic Chemistry and ApplicationsHindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

    SpectroscopyInternational Journal of

    Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

    The Scientific World JournalHindawi Publishing Corporation http://www.hindawi.com Volume 2014

    Medicinal ChemistryInternational Journal of

    Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

    Chromatography Research International

    Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

    Applied ChemistryJournal of

    Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

    Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

    Theoretical ChemistryJournal of

    Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

    Journal of

    Spectroscopy

    Analytical ChemistryInternational Journal of

    Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

    Journal of

    Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

    Quantum Chemistry

    Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

    Organic Chemistry International

    ElectrochemistryInternational Journal of

    Hindawi Publishing Corporation http://www.hindawi.com Volume 2014

    Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

    CatalystsJournal of