Upload
others
View
0
Download
0
Embed Size (px)
Citation preview
Page ‹#›
Oct 2007 Adapted from Tom Lewellen, PhD
Robert MiyaokaDivision of Nuclear MedicineUniversity of Washington Medical CenterSeattle, Washington USA
An introduction to PET physics
Oct 2007 Adapted from Tom Lewellen, PhD
PET - the basic steps to make an image
Decoding Detector corrections
PrimaryDetection
Coin.processing
Databinning
Datacorrections
Imagereconstruction
Oct 2007 Adapted from Tom Lewellen, PhD
NN
N P
PP P
P N NN
NN
N P
PP P
P P NN
ν
e-e+
Proton decays toneutron in nucleus -
Positron combines withelectron and annihilates
γ
γ
Two anti-parallel 511 keVphotons produced
Unstable parentnucleus
positron and anti-neutrinoemitted
PET basics - positron annihilation
Oct 2007 Adapted from Tom Lewellen, PhD
Spatial Resolution(positron range)
18F 11C 15Omax eng (MeV) 0.633 0.959 1.734most probable eng (MeV) 0.203 0.326 0.696
combined effects of tomoresolution and positron range
*Palmer and Brownell, IEEE Trans Med Img vol.11(3), 1992
Oct 2007 Adapted from Tom Lewellen, PhD
Spatial Resolution(photon non-collinearity)
Photon non-collinearity:Range of angle is + 0.25degrees, centered at 180degrees
R180ο = 0.0022 X D
~2.2 mm for 1 m diameterPET detector system
Oct 2007 Adapted from Tom Lewellen, PhD
What do we use for detectors?
Page ‹#›
Oct 2007 Adapted from Tom Lewellen, PhD
Some details about scintillators
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1
NaI:Tl BGO LSO GSO:Ce
Scintillator
To
tal lin
. attn
. (cm
-1
)
0
20
40
60
80
100
120
NaI:Tl BGO LSO GSO:Ce
Scintillator
Relative Lig
ht Y
ield
Ideally - want dense,fast, bright, cheap
So far, we can notget all at once
Detection efficiency Energy resolution, etc.
Oct 2007 Adapted from Tom Lewellen, PhD
• 3 cm thick bismuth germanateoxide - BGO
µ = 0.91 cm -1
Interaction probability = 0.93Coincidence probability = 0.87
• 2.5 cm thick Sodium Iodide -NaI(Tl)
µ = 0.34 cm -1
Interaction probability = 0.57Coincidence probability = 0.33
No energy discrimination
BGO versus NaI(Tl) - a simplistic comparison
Dedicated BGORing system
Dual headCoincidenceGamma camera
Oct 2007 Adapted from Tom Lewellen, PhD
Two basic detector design approaches
• Driven by scintillator characteristics
• Continuous detectors - no longer a significant player in dedicated PET
• Pixilated detectors - discrete crystals orsaw cut “slabs”
Oct 2007 Adapted from Tom Lewellen, PhD
Large pixilated detectors - evolutionfrom large NaI(Tl) based PET systems.
1 2 3 4trigger
Current system (Philips) uses 4x6x20 mm GSOcrystals mounted to a light guide and the arrayof PMTs - decoding is done essentially thesame way as was done for the NaI version
Oct 2007 Adapted from Tom Lewellen, PhD
An example of a BGO Detector Block
Reflective material
Discrete BGO crystals4 mm transverse8 mm axial30 mm radial
Plastic light shield
2 dual photocathode PMTs
Alternative to saw cuts - discrete crystals with surface treatments/ coupling compound difference to produceequivalent light sharing.
Oct 2007 Adapted from Tom Lewellen, PhD
Tran
sver
se Axial
Block Crystal Decoding
A
B
C
D
A,B,C,D = PMTs
Example for a 6x6array of crystals
E = A + B + C + D
Axial = Z = C + D( ) E
Transverse = X = B + D( ) E
Page ‹#›
Oct 2007 Adapted from Tom Lewellen, PhD
Example Crystal Map
Oct 2007 Adapted from Tom Lewellen, PhD
Block Energy
Peaks fordifferent crystals
at differentpositions
Window centerand width
adjusted for eachcrystal
Oct 2007 Adapted from Tom Lewellen, PhD
Spatial Resolution(discrete crystal detector system)
intrinsic(~0.5 w)
w
photon penetration(DOI)
scatter withindetector
statisticaluncertainty(positioning)
dense, good photoelectric cross-section, high light output
Oct 2007 Adapted from Tom Lewellen, PhD
A few things to note about positron decay
1. Range of positron ( < 1 mm for F-18)
2. Non-collinearity - the gamma rays are not exactly at 180 degrees, the FWHM of the angular spread is ~ 0.3 degrees
3. Resolution effects, like gamma cameras, generally add in quadrature:
FWHM = FWHMdet
2+ FWHMcolin
2+ FWHMrange
2
Oct 2007 Adapted from Tom Lewellen, PhD
Channel 1
Channel 2
SummedChannel
γ
γ
coincidence events
“Electronic” collimation => high sensitivity
PET basics - coincidence detection
Oct 2007 Adapted from Tom Lewellen, PhD
Data is binned into a sinogram
Data Acquisition
Image ReconstructionActivity
Distribution(single slice) Projection Data “Sinogram”
“distance”, d
θ
Y
Xθ
d
Page ‹#›
Oct 2007 Adapted from Tom Lewellen, PhD
Data is binned into a sinogram
neg pos
Distance from center (bin)-125 0 125
0 90 179
Angle
-125 0 125
-125 0 125-125 0 125
-125 0 125-125 0 125
negpos
negpos
negpos
negpos
Oct 2007 Adapted from Tom Lewellen, PhD
True
Scatter
Randoms
Types of coincidence events
Oct 2007 Adapted from Tom Lewellen, PhD
Coincidence detection
# events
time between detector response
Randoms and count rate
Prompt window
Delayed window
Oct 2007 Adapted from Tom Lewellen, PhD
Randoms and count rate - II
• We can also measure the detector singles rates
– R = 2tS1S2 Where t = resolving time, S is the detector singles rates
– Need to some other corrections - particularly accounting for energywindowing and proper dead time corrections.
– Pro= less noise in the correction.
– Con = is getting the corrections right and accounting for any rapidchanges in counting rate during acquisition (works best in list mode ifthe activity is rapidly changing during the scan)
Oct 2007 Adapted from Tom Lewellen, PhD
Scintillator speed and randoms
Slow Fast
Randoms ~ constant versus time
In this example, #trues = constant
“fast” has ~ 1/2 the randomsOct 2007 Adapted from Tom Lewellen, PhD
Count Rate 3D (LLD = 375keV)
0
100
200
300
400
500
600
0 0.5 1 1.5
Activity Conc. (uCi/cc)
Count
rate
(k
cps)
NECR (2R)
T
R
S
Randoms go as activity squared!
For full ring systems using BGO, LSO, or GSOrandoms limit the maximum usable count rate.
Page ‹#›
Oct 2007 Adapted from Tom Lewellen, PhD
So, we can correct for randomsand we have a handle on countrate limitations.
Another major effect is that ofphoton attenuation in the patient
How do we correct for that?
Oct 2007 Adapted from Tom Lewellen, PhD
Direct coincidence measurement with rotatingsource
Singles measurements
Combined PET/CT scanner
Options in collecting and processingtransmission scan data
Oct 2007 Adapted from Tom Lewellen, PhD
Transmission rod sourcesrotate around the patient
Acquisition is a high energy CT scan
Data allows accurate attenuation correction
Transmission scanning in PET
Oct 2007 Adapted from Tom Lewellen, PhD
Effects of Attenuation in PET
P1= e
! µ x( )dx0
"x
#
P2= e
! µ x( )dx"x
a
#
PC= P
1P2= e
! µ x( )dx0
a
"
Oct 2007 Adapted from Tom Lewellen, PhD
Coincidence + gating rejects mostscatter
Problem - count rate limited bydetector closest to rod source
Count rate limited by thefar detector
Problems - scatter; hardwaremodifications to tomograph
Gated coincidence transmission Singles transmission
Solution - use image segmentation on attenuation data setsOct 2007 Adapted from Tom Lewellen, PhD
T+E - Transmission + emission(transmission after dose is injected)
Essential for clinical scanning with conventionalPET scanner
standard technique (without CT scanner)
Collect emission data
Collect transmission data
Use emission data to subtract emission contamination from transmission data
Either interleave T and E data or takeall E data first and then T data
Page ‹#›
Oct 2007 Adapted from Tom Lewellen, PhD
Segmented Attenuation Correction(SAC)
Smooth andreconstruct
0
1000
2000
3000
4000
5000
6000
7000
8000
0 0.02 0.04 0.06 0.08 0.1 0.12 0.14 0.16
Series1
0
5000
10000
15000
20000
25000
0 20 40 60 80 100 120 140
Series2
Histogramµ values
Extractclusters
Classifyimage Forward
project
Oct 2007 Adapted from Tom Lewellen, PhD
Images courtesy of CTI and GEMS web sites.
PET/CT
Oct 2007 Adapted from Tom Lewellen, PhD
1. Scout scan(5-20 sec)
CT PET
4. Whole-body PET(6-40 min)
CT PET
PET/CT scan protocol
3. Helical CT (1-2min)
CT PET
2. Selectionof scanregion (1-2min)
Scout scan image
Oct 2007 Adapted from Tom Lewellen, PhD
CT-based Attenuation Correction• The mass-attenuation coefficient (µ/ρ) is remarkably similar for all non-bone
materials since Compton scatter dominates for these materials. Bone has ahigher photoelectric absorption cross-section due to presence of calcium
Oct 2007 Adapted from Tom Lewellen, PhD
CT-based Attenuation CorrectionBi-linear scaling methods apply different scale factors forbone and non-bone materials
air-watermixture
water-bonemixture
air soft tissue dense bone
Oct 2007 Adapted from Tom Lewellen, PhD
Effect of Contrast Agent on CT to PETScaling
The presence of Iodine confounds the scaling process as Iodinecannot be differentiated from bone by CT number alone.
Curve thatshould be usedfor contrastagent
Page ‹#›
Oct 2007 Adapted from Tom Lewellen, PhD
What we have so far:
Basics of positron decay
The kinds of coincidence events
An overview of randoms and count rate issues
Several options for attenuation correction
What are the options in basic tomograph designs?
Oct 2007 Adapted from Tom Lewellen, PhD
Some test questions
Oct 2007 Adapted from Tom Lewellen, PhD
D68. Positron cameras detect:
A. Positrons of the same energy in coincidence.B. Positrons and electrons in coincidence.C. Photons of different energies in coincidence.D. Annihilation photons in coincidence.E. Annihilation photons in anticoincidence.
Oct 2007 Adapted from Tom Lewellen, PhD
D69. Spatial resolution of PET systems is determined by:
A. Detector size.B. The ring diameter of the system.C. The detector material.D. Energy of the positron emitter in use.E. All of the above.
Oct 2007 Adapted from Tom Lewellen, PhD
D77. The major limitation on the resolution of an FDG scan on a modernwhole body PET scanner is:
A. Range of the positron.B. Image matrix size.C. The physical size of the individual detectors.D. The non-collinearity between the annihilation photons.E. Attenuation correction.
Why? Camera res ~ 5 mm, positron range depends onisotope but for 18F is < 1 mmNon-collinearity (100 cm diameter) ~3.0 mm.Physical size of detector (~4 mm ) -> ~2.0 mm
Reconstruction filter -> ~10 mm
Oct 2007 Adapted from Tom Lewellen, PhD
D85. All of the following are true statements about PET scanning,except:
A. Radioisotopes are cyclotron produced.B. Positrons are not detected directly.C. Coincident detection at 180° is required.D. Images are generally axial tomograms.E. The detector photopeak is centered at 1.02 MeV.