38
1 Reprocessing & Remanufacturing Bill Enos Global Head of Microbiology 2016 BSI Roadshow Copyright © 2016 BSI. All rights reserved.

Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

  • Upload
    others

  • View
    13

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

1

Reprocessing & Remanufacturing Bill Enos

Global Head of Microbiology

2016 BSI Roadshow

Copyright © 2016 BSI. All rights reserved.

Page 2: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 2

Agenda

1. New procedure general review

2. Examples of what is being reprocessed

3. Overview of risks and concerns

4. Regulatory References

Page 3: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 3

What Is Being Reprocessed

Spaulding Classification

• Revisited 1994

Page 4: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 4

Non-critical use: A device that does not touch the patient or touches only intact skin.

Pulse Oximeters

ECG Leads

DVT Sleeves

Blood Pressure Cuffs

Splints

Cervical Collars

Page 5: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 5

Semi-critical use: A device that comes into contact with intact mucous membranes and does not penetrate tissues.

Endoscopes Colonoscopes

Bronchoscopes

Esophageal Manometry Probes

Endotracheal Tubes Laryngoscope Blades

Page 6: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 6

Critical use: A device that is used for surgically invasive medical procedures.

External Fixation Devices Drill Bits

Saw Blades

Diagnostic Ultrasound Catheter

Diagnostic Electrophysiology Catheter Laparoscopes

Arthroscopes

Page 7: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 7

Risks and Concerns to Consider

Copyright © 2016 BSI. All rights reserved.

Page 8: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 8

Release Transit Storage

Patient Use

Decontamination Cleaning

Re-Manufacturing Performance

Testing

Packaging

Sterilization

New Device

Discard Device

Risk Management Validation Traceability throughout lifecycle Documentation

Quality Management System

Page 9: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 9

Questions from the Notified Body

• How are incoming devices inspected and

considered acceptable for use or disqualified for reprocessing?

• Devices that are single use will not have had cleaning and disinfection studies completed in the past

• What is the efficacy of the cleaning and disinfection process?

• Can the device be effectively cleaned or disinfected?

• Are the materials of construction compatible with enzymatic cleaners or disinfectants?

• How does the device or the material stand up to cleaner or disinfectant residues?

• What is the lifetime of the device? • Will the device be labelled single use? • How many times can the device be re-

manufactured? How do you ensure the maximum number is not exceeded?

• Design, Manufacturing Process, In Use – Risk Management?

• As new risks are introduced, do you have a clinical evaluation and reconsidered risk benefit analysis?

• Are you monitoring and measuring complaints/incidents versus re-manufacturing?

• What is your process to recall devices? • How do you know if the original manufacturer

has recalled devices?

Page 10: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 10

Questions from the Notified Body

• What are the original functional performance

specifications? • If you could not obtain the original specifications

can you justify that the reverse engineering will result in a safe device that performs as intended?

• Mechanical qualification: • Compression, Tension, Torsion, Deflection, Bending, Fatigue,

Joint Seal Integrity

• Electrical qualification: • Safety, isolation, continuity, consistent and unbroken

• How do you know if the original manufacturer has changed the device?

• Will you inform your Notified Body of those changes?

• Is biocompatibility impacted by new or additional

processes? • Enzymatic cleaners

• Detergents

• Disinfectants

• Disinfectant Residues

• Different sterilization method

• Repetitive cycles of the same method

• Differences in exposure levels during sterilization cycles

• If the sterilization method is the same, material compatibility needs to be reassessed due to the device being subjected to multiple sterilization cycles

• Does a change in sterilization method or addition of multiple sterilization cycles change the biocompatibility of the device? Will biocompatibility testing be repeated for all device and material types?

Page 11: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 11

Questions from the Notified Body

• Will the devices be re-sterilized using the original

sterilization method? • If not, then material compatibility with the sterilization

method has to be reassessed

• The device will be subjected to additional and/or different sterilization methods, does that change:

• Performance of the device

• Lifetime of the device

• Risk associated with the device, it’s use and identification of new hazards

• Change in clinical risk to the patient

• Are packaging validation conducted on every

device type or family? • Are transit (drop, vibration) studies conducted on

every device type or family? • Are shelf life studies conducted on every device

type or family? • Is the information supplied with each device

updated with the changes due to re-manufacturing:

• a remaining contaminate

• persistence of chemical substances used during the reprocessing process

• alterations in the performance due to the reprocessing

Page 12: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 12

Questions From the QMS Perspective

• Is the reprocessing of devices identified in the quality manual, scope of registration, or other documentation?

• Is there a procedure for the receipt and handling of product for reprocessing? How are orders for new product versus reprocessed product taken and processed?

• Are device reprocessing and associated regulatory requirements covered under management review?

• Is there a mechanism for identification and traceability of devices for reprocessing? • Are there specific labeling requirements for reprocessed devices? • Is there a mechanism to track complaints or vigilance for reprocessed product? • Is the system of device reprocessing covered under the internal audit program?

27/09/2016

Page 13: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 13

Update in the New MDR???

• It doesn’t look like we’re going to get clear direction in the new MDR • It may be a “regulate as we go” approach with the commission adding

strength to requirements over time

•MHRA has a guidance document - BSI has the responsibility • Key point – reprocessed devices must be stay within the same location (similar to the German model)

• All technical documentation shall be reviewed, no sampling •Member states make the call

27/09/2016

Page 14: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 14

Existing Global Requirements

Copyright © 2016 BSI. All rights reserved.

Page 15: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 15

• Third-party or hospital reprocessors must comply with the same requirements that apply to original equipment manufacturers, including submitting documents for pre-market notification or approval, registration of reprocessing firms and listing all products on the FDA website.

• Reprocessors must also submit adverse event reports, track devices whose failure could have serious outcomes, correcting or removing from the market unsafe devices.

• Requirements for single use devices are that the label must contain “single use” and the instruction for use must contain “reuse is prohibited.”

• A reprocessor who reprocesses a medical device originally intended for single use only becomes the manufacturer of the reprocessed device, and as such is required to apply the relevant conformity assessment procedure for the product.

Page 16: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 16

• As of 2015 Health Canada will issue medical device licences to commercial reprocessors who reprocess SUDs for healthcare facilities.

• All RSUD’s are to be licensed by September 1, 2016.

• Reprocessors will be considered the ‘manufacturer’ as defined in the MDR.

• Reprocessors of all risk classes of device will be required to maintain objective evidence to demonstrate that their RSUD’s meet the S&E requirements of the MDR.

• Reprocessing of single use medical devices is not currently regulated in the European Union and different national legislations regulate this practice throughout Europe.

• Some countries allow the reprocessing of single use medical devices and have developed guidelines (e.g. Germany), while some countries prohibit it (e.g. France) and some Member States do not have any specific regulations on this aspect.

Page 17: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 17 17 Copyright © 2013 BSI. All rights reserved.

Article 2 Definitions

Page 18: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 18

Article 2

• ‘single-use device’ means a device that is intended to be used on an individual patient during a single procedure

• ‘reprocessing’ means the process carried out on a used device in order to allow its safe reuse including cleaning, disinfection, sterilisation and related procedures, as well as testing and restoration of the technical and functional safety of the used device;

• ‘placing on the market’ means the first making available of a device, other than an investigational device, on the Union market

• ‘putting into service’ means the stage at which a device, other than an investigational device, has been made available to the final user as being ready for use on the Union market for the first time for its intended purpose

Page 19: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 19

Article 2

• ‘manufacturer’ means the natural or legal person who manufactures or fully refurbishes a device or has a device designed, manufactured or fully refurbished, and markets that device under his name or trademark, regardless of whether these operations are carried out by that person himself or on his behalf by a third party.

• For the purposes of the definition of manufacturer, fully refurbishing is defined as the complete rebuilding of a device already placed on the market or put into service, or the making of a new device from used devices, to bring it in conformity with this Regulation, combined with the assignment of a new lifetime to the refurbished device;

• ‘health institution’ means an organisation whose primary purpose is the care or treatment of patients or the promotion of public health;

Page 20: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 20

Article 2

Reprocess Remanufacture Fully Refurbish

Page 21: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 21 21 Copyright © 2013 BSI. All rights reserved.

Article 4

Page 22: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 22

Article 4

• 4a. With the exception of the relevant general safety and performance requirements set out in Annex I, the requirements of this Regulation shall not apply to devices, manufactured and used only within health institutions, provided that the following conditions are met:

• (aa) the device is not transferred to another legal entity, • (a) manufacture and use of the devices occur under appropriate quality

management systems, • (b) the health institution establishes in its documentation that it has given due

consideration as to whether the target patient group’s specific needs cannot be met by an equivalent device available on the market,

• (c) the health institution provides information on an annual basis on the use of such devices to their competent authority, which shall include a justification of their manufacturing, modification and use,

Page 23: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 23

Article 4

• (d) the health institution draws up a declaration, that it shall make publicly available, including: • the name and address of the manufacturing health institution; • the details necessary to identify the devices; • a declaration that the devices meet the general safety and performance

requirements set out in Annex I of this Regulation and, where applicable, information on which requirements are not fully met with reasoned justification,

• (da) the health institution draws up documentation, allowing an understanding of the manufacturing facility, the manufacturing process, the design and performance data of the devices, including the intended purpose, sufficiently detailed to enable the competent authority to ascertain that the general safety and performance requirements set out in Annex I of this Regulation are met;

Page 24: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 24

Article 4

• (e) the health institution takes all necessary measures to ensure that all devices are manufactured in accordance with the documentation referred to in the previous subsubparagraph, and

• (f) the health institution reviews experience gained from clinical use of the devices and takes all necessary corrective actions.

• Member States may require that the health institutions submit to the competent authority any further relevant information about such devices which have been manufactured and used on their territory. Member States shall retain the right to restrict the manufacture and use of any specific type of such devices and shall be permitted access to inspect the activities of the health institutions.

• These provisions do not apply to devices which are manufactured on an industrial scale.

Page 25: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 25 25 Copyright © 2013 BSI. All rights reserved.

Article 15 Single-use devices and their reprocessing

Page 26: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 26

Article 15

1. Reprocessing and further use of single-use devices may only take place where permitted by national law and only in accordance with this article.

Any natural or legal person who reprocesses a single-use device to make it suitable for further use within the Union shall be considered to be the manufacturer of the reprocessed device and shall assume the obligations incumbent on manufacturers laid down in this Regulation.

Page 27: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 27

Article 15

1. a) By way of derogation from paragraph 1, as regards single-use devices that are reprocessed and used within a health institution, Member States may decide not to apply all rules relating to manufacturers' obligations laid down in this Regulation provided that they ensure that:

• the safety and performance of the reprocessed device is equivalent to that of the original device and the requirements in Article 4(4a) (a), (c), (d) and (e), are complied with;

Page 28: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 28

Article 15

b) the reprocessing is performed according to common specifications detailing the requirements:

• on risk management, including the analysis of the construction and material, related properties of the device (reverse engineering) and procedures to detect changes in the design of the original product as well as of its planned application after reprocessing,

• on validation of procedures for the entire process, including cleaning steps, • on product release and performance testing, and • on the quality management system.

• Member States shall notify the Commission and the other Member States of the

national provisions, introduced pursuant to this paragraph and the grounds for introducing them. The Commission shall keep the information publicly available.

Page 29: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 29

Article 15

1. b) Member States may choose to apply provisions referred to in paragraph 1a also as regards single-use devices that are reprocessed by an external reprocessor at the request of a health institution, provided that the reprocessed device in its entirety is returned to that health institution and the reprocessor complies with the requirements referred to in paragraph 1a (a) and (b).

Page 30: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 30

Article 15

1. c) The Commission shall adopt the necessary Common Specifications referred to in paragraph 1a(b) by the date of application of this regulation.

In case Common Specifications are not adopted by the date of application of this regulation, reprocessing shall be performed according to relevant harmonized standards and national provisions that ensure compliance with the requirements outlined in paragraph 1a(b). The compliance with the common specifications or, in the absence of common specifications, the relevant harmonized standards and the national provisions, shall be certified by a notified body.

Page 31: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 31

Standards

• Perinorm – 1,500,000 records

• British Standards On Line – 50,000

• Compliance Navigator – checks for changes daily

• Cleaning

• Disinfection

• Sterilisation

• Packaging

• Shelf life and Transit Testing

• Biological Evaluation

• Information Supplied

• Quality Management Systems

• Risk Management

• Clinical Investigation

• Performance

Page 32: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 32

Article 15

2. Only single-use devices that have been placed on the Union market in accordance with this Regulation, or prior to [date of application of this Regulation] in accordance with Directive 93/42/EEC may be reprocessed.

3. Only reprocessing of single-use devices that is considered safe according to the latest scientific evidence may be carried out.

Page 33: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 33

Article 15

4. The Commission, by means of implementing acts, shall establish and regularly update a list of categories or groups of single-use devices which cannot be reprocessed safely and therefore may under no circumstances be reprocessed. Those implementing acts shall be adopted in accordance with the examination procedure referred to in Article 88(3).

http://ec.europa.eu/growth/sectors/medical-devices/index_en.htm

Page 34: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 34

Devices that cannot be remanufactured:

• Stents

• Devices with batteries, pressure cartridges or filters

• Sutures

• Devices containing latex

• Gowns and drapes

• Devices with bioactive coatings (e.g. heparin, collagen)

• Devices with absorbable components

• Devices with medications, betadine or hemostatic agents

• Angiographic catheters

• Devices with hydrophilic coatings

• Devices requiring calibration

Page 35: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 35

Article 15

5. The name and address of the legal or natural person referred to in paragraph 1 and the other relevant information in accordance with Section 19 of Annex I shall be indicated on the label and, where applicable, in the instructions for use of the reprocessed device.

The name and address of the manufacturer of the original single-use device shall no longer appear on the label, but shall be mentioned in the instructions for use of the reprocessed device.

Page 36: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 36

Annex I – Section 19.2 Section 19.3

• The following particulars shall appear on the device or where not practical or appropriate on the trade packaging:

a) … s) …

• The instructions for use shall contain the following particulars:

a) … u) … + name and address of original manufacturer Changes: • a remaining contaminate • persistence of chemical

substances used during the reprocessing process

• alterations in the performance due to the reprocessing

Page 37: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 37

Article 15

6. A Member State that permits reprocessing of single-use devices may maintain or introduce stricter national provisions restricting or prohibiting, within its territory, the following:

a) the reprocessing of single-use devices and the transfer of single-use devices to another Member State with a view to their reprocessing;

b) the making available or further use of reprocessed single-use devices.

• Member States shall notify the Commission and the other Member States of the national provisions. The Commission shall keep the information publicly available.

Page 38: Reprocessing & Remanufacturing - BSI Group Reso… · instruction for use must contain “reuse is prohibited.” • A reprocessor who reprocesses a medical device originally

Copyright © 2013 BSI. All rights reserved. 38

Questions & Answers

Copyright © 2016 BSI. All rights reserved.