Release Notes - CT Dose Profile Analyzer
RTI Electronics ABR&D Department
=============================================CT Dose Profile Analyzer v4.0C, 2010-10-01build 4.0.0.15DLL v2.4.0.8345Improvements and Corrections=============================================- Database is updated- RTICmLib.dll is updated
CT Dose Profile Analyzer v4.0B, 2010-03-15build 4.0.0.15DLL v2.4.0.7914Improvements and Corrections=============================================
- RTICmLib.dll - updated
CT Dose Profile Analyzer v4.0A, 2009-05-28build 4.0.0.15DLL v2.4.0.7481=============================================For complete installation instructions seechapter 2 of the CT Dose Profile Analyzer manual.
Improvements and Corrections=============================================
- Measure/New is not supported in this version.Use Open example insted.
- Main support of the new CT Dose profiler that is an enhanced version of the old CTSD-16 .- Support of four detectors instead of only one. Two is dedicated for CT dose profile measurement, two is of general purposeand applicable specially when measured with the dose template only
- Detector ID display window now indicates detector that is used:0=CTSD-16, 1=CT Dose Profiler, 2=R100B, 3=Piranha Dose Probe, N/A =Not supporteddetectors- Auto selection of energy and field width corrections for CT applications for the CT Dose Profiler and CTSD-16.- New Measurement parameter in the CTDI template based on CT IEC standard:Geometric efficiency = Integral of collimation, NT divided with the total integral of the dose profile in percent
Only defined without phantom in the beam (RQR beam qualities)- Field width corrections for measurement without phantom in the CTDI template for narrow beam down to 3 mm.- Updated CT Database 3.0 20090630 support new k- values and support for CT DoseProfiler- Now Show vertical scrollbar and to support uPC with limited vertical resolution of 600 pixels.- Allow user to resize (maximize) the program window- New updated manual including more information about detector specifications and references of CT articles.
7/28/2019 Release Notes - CT Dose Profile Analyser
2/3
- Fixed bug where Barracuda and Piranha sometimes crashes in the settings screen.
- Bigger font size so the user easier can read the "System info / Instructions "- Delay and window was wrongly displayed when selecting "time "mode in the setting screen.- Upper X-axis name for CTDI template is now called Time(s) instead of Z-axis(s)- Program renamed to CT Dose Profile Analyzer instead of CT Dose Profile Analyser
CT Dose Profile Analyser v3.0A, 2008-06-30DLL v2.3.0.7300=============================================For complete installation instructions seechapter 2 of the CT Dose Profile Analyser manual.
Improvements and Corrections=============================================- Changed name from CT-SD16 to CT Dose Profile Analyser- Old measurements can be updated with just a double-click in the new CT Databas
e and resaved- Different Result field for Central point method, Five point method, Free in Air and Dose- New settings layout- Support measuring time up to 160s- New Measure and Report menu- Simplifier QA with one point method- Support of CT Database- Timed mode preferred method- Continuous method supported for Dose- CTDI double x-axis scale: time and length- Support of DLP, CTDIvol, CTDIw according to CT panel readings- Support of energy corrections and k-factors from CT Database.
- Support for automatically identification of Barracuda or Piranha- Setting screens simplified using tabs- Support of Power management- Dose template improved with support of two cursors and calculation between cursors- noise filter for graph data- Support of examples for Siemens, GE, Toshiba and Philips- Fixing error screen when File open was aborted without selecting file- Open example now navigate to the example directory- CT theses and publishing material from Mayo clinics now support our data.- Red stars marks needed setting values and green stars marks values set by theDatabase- Support of New Measurement without loading example
- Report menu supports print and html of Front panel- Export data and graph to excel compatible txt file for both dose and CTDI template- Report generator- New installer on the RTI CD- Support of VISTA- User can configure the CT Database- Support for new improved symmetrical detector- Support for manual Z-axis movement
7/28/2019 Release Notes - CT Dose Profile Analyser
3/3
=============================================CT-SD16 version 2.0B 2008-02-12DLL version 2.1.0.7136=============================================For complete installation instructions seechapter 2 of the CT-SD16 manual.
The following bugs/issues are fixed/improved:
- DLL updated to support new firmware.
- Correction of USB communication issue.
- Support for CT-SD16 version 2.
Known Limitations:
- When measuring CTDIvol in the peripheral holesof the phantom it is recommended to use a lowpitch value around 0.5 to be sure to get a
sufficient dose contribution from the CT beam.For measurement in the central hole the pitchvalue is not critical since the CT-SD16detector is symmetrically placed in the middleof the phantom.
- Conversion of detector data to the new detectorsystem is not possible using CT-SD16.Please see the CT-SD16 section in the document
Step-by-step guide of how to update yourBarracuda/Piranha system found on www.rti.se.
=============================================
CT-SD16 version 2.0B 2007-07-06DLL version 2.0.1.6837=============================================
For complete installation instructions seechapter 2 of the CT-SD16 manual.
The following bugs/issues are fixed/improved:
- DLL updated to support new firmware.
- Some improvements to the CT-SD16 setup program.
- Fix for issue where measurement in Timed modedid not return any data.
- The program for sending support information isupdated to better assist you in resolving issues.
- Support for Detector information stored in themeter and the detector.