- :po- Journal of Research of the Na ti ona l Bureau of Standards Vol. 58, No.4, Apr il 195 7 Re search Paper 2752 Reference Wavelengths for Calibrating Prism Spectrometers Earle K. Plyler, 1. R. Blaine, and Matthew Nowak Stan dard wa \" clengLhs for ca li bra tin g inf ra red -p rism ins tr uments nav e been meas ur ed f ro m t he a bsorpLion ba nds of polys ty rene, 1,2, 4-tr ichl or obe n zen e, a nd tolu ene. Sev er al bands of di dym ium glass h av e been ca librated for the ncar-infrared r eg ion, an d 27 r ot a- ti on a lli nes of the 3.3-micro n band of methane h av e been meas ur ed. Th e wav eleng th s fall in Lhe reg ion fr om 0. 6 to 24 microns, but the gre atest number arc in t he 3- to 6-micron region. A few ban ds of se ve ral ot h er m ate r ia ls ar e also given in t he \\'av c!eng th t a b les, and 11 fi g ur E'S of t h e Rp e ctra mark ed with the cali braLi ng values are incl ucl ed for identi fi ca t ion . 1. Introduction Two re por ts [\ JL ha ve been issued b.'Tthe Bu rea u 0 11 ca libr at ing wav eleng ths fo r 'P ri sm spect rom eters. T he present report in co rporat es th e dat a of th e earlier reports a nd gives t lt e result s of recen t cali- br at ions th at h ave b ee11 mad e. Other papers h ay e been published b." several laborat ories [ 2] and t bei )' results arc of value fo r addi tional ea libration po int s. All of the st and a rd wavelengt hs Lll at have been suppli ed arc n ot sufficient fo r calibrat ions n ee ded fo r th e ran ge of the infrared r eg ion from 0.8 to ] 50).L 01' for t he wid e var i et y of inst ruments used in tbis r eg io n, espeeia ll." witl)' t he differences in resolution obtained wit h different prism m at eri als. For ex, ampl e, the CO and CO2 absorption band s between 4 and bJ.L, are well su ited for cali br at ion bu L canH O (, be resolved into their rotati onal s tru ct UJ' e b, ' a r oc k- salt prism. Th e requirement s for the calibration of a spect rom eter using a rock-salt prism are mu ch dif- ferent from those of an instr ument us in g prisms of lit hium HUOl'id e an d of calcium flu orid e. Wh en the thermocoupJe of an ins tr ument is replaced by a PbS det ect or, more bands th at contain the ro ta - tional stru ct ur e of gases can be used for calibration. Still higher r eso lut ion can be ob ta ined by repl ac ing the pri sm in the spectrometer with a gratin g. Thi s allows the use of many bands for calibration, and if too high a pr ecision is not required, many bands th at have been measUl' ed on large gra ting instrumen ts can be used. This repor t docs not in cl ud e met hods of calibrn.- tion of hi gh-resoIu Lion grating spec trometers. Th e use of th e fringe s." stem of a Fabry -Perot inte)' , ferom eter for accur at e measur ement of line positions observed wi th grating spectrometers ha s been pr e- viously described [3]. The cali br ation poin ts li sted in th e pr ese nt paper are primarily suit ed for low .. resolu tion ins trument s. 2. Method of Measurement A lar ge grat ing spectrometer can easil." measur e sharp sp ectral lines wi th an error not greater th an 1 par t in 100,000, Thi s aCCllra cy can be obtained 1 Fi gur es in bmckets indi cat( the:litcrature referen ces a(t be end of this paper. b.\' using sta nd ard lilles of neon, argon, kry pton , or xenon, in higher orders th at overl ap th e ba nd being measUl' ed. T he standard spe ct ral lin es 01' the fringe s." stem of a Fabry-Perot int erf erometer are re cord ed simult aneously with the absorp ti on band being studied. Th e grating spectrometer a nd th e method of calibration ar e desc rib ed in reference [3 ]. Th e meaS Ul' ement of the absorpt ,ion bands of liquids and solids und er high resolu tion often reveals side ba nds 0 1' complex st ru c Lur e. Th e way in which the co mp onents a dd up to pro du ce a seemi ngl.\' sharp band on a prism in strument mu st be a ppr oximat ed by the in Lensit y and the positions of the various co mpon ents. Any ba nd in Lh e ran ge 1 to 10 J.L that is found to have an uncer ta in t \T of more than ± 0.00 3 ).L is not listed as a wavelengt li for cali bration. Anot her approach in the calibra tion of bands is to open th e sli t of the gr at ing spectrom et er 0 t hat r esolu tion equ al Lo Lhat of a prism spectr ome ter is obtained. Thi s meLhodreduces the high precision of m eaS Ul' emen t with the gra ting spec tr ometer, but wi th the wide range of resolu tion of differ ent sp ecLro- meters, the band shapes could n ot be m atc h ed fo r all instr umen ts. In order Lo avoid the d iffi culties of the low-resolution method, all the ba nds reported in t hi s work h ave b een calibrated wi th hi gh-r esolu tion spectrometers. In th e measur ement of Q-branches of ba nd s, con- sidera ble fine st ru ct ur e is obser ved. Th e Q-b ranch of th e NH3 ba nd at 2.999 ).L is shown in fi gure 1. Th ere are 30 or mor e sepa ra te lines and the band extend s from 2.996 to 3. 002 J.L; th e center on low r esol ution is estimat ed to be 2.9990 ± 0.0005 ).L . All of the bands measur ed on the grating instr um en t, app ear to be broad , and th e es timat e of the cen ter for low resolution acc ounts for the uncertaint y giv en with th e r es ults listed in th e tables. 2.1. Region From 0.6 to 2.6 Microns Although the intense infrar ed bands of molecule ar e observ ed at wavelengths greater than 2 p. , it is sometimes necessary to measur e th e t ransmission of filters a nd the absorption of combination bands in the near-infrared r eg ion . Th e atomi c sp ectral lin es ar e well suited for ca li bration in this region. Th e 195

Reference wavelengths for calibrating prism spectrometers · 4 and bJ.L, are well su i ted for calibration bu L canHO (, be resolved into their rotational structUJ'e b,' a rock salt

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Reference wavelengths for calibrating prism spectrometers · 4 and bJ.L, are well su i ted for calibration bu L canHO (, be resolved into their rotational structUJ'e b,' a rock salt

- :po-

Journal of Research of the Na tiona l Bureau of Standards Vol. 58, No.4, April 1957 Research Paper 2752

Reference Wavelengths for Calibrating Prism Spectrometers

Earle K. Plyler, 1. R. Blaine, and Matthew Nowak

Stan da rd wa \"clengLhs for cali b ratin g infrared-prism in str um ent s nave been measured fro m t he a bsorpLion ba nds of polysty rene, 1,2,4-tr ichlorobenzen e, and toluene. Several b an ds of didym ium glass have been ca librated for t he nca r- infra red r egion, an d 27 rota ­t ionallines of t h e 3. 3-micro n ban d of m ethan e have been m easured.

The wavelengths fall in Lhe region fr om 0. 6 to 24 m icron s, bu t t he g reatest numb er a rc in t he 3- to 6-m icron reg ion . A few bands of several other materials are a lso give n in t he \\'avc!e ngth t a bles, and 11 figurE'S of t h e Rpectra ma rked wit h t he cali braLing valu es a re incl ucled for iden t ifi ca t ion .

1. Introduction

Two repor ts [\ JL have been issued b.'T t he Bureau 0 11

calibrat ing wavelengths for 'Pri sm spectrometers. T he pr esent repor t inco rporates the data of the earlier reports and gives t lte results of recen t cali­brations that have bee11 made. Other p apers haye been published b." several laboratories [2] and t bei)' resul ts ar c of value for addi t ional ealibration points . All of the s tandard wavelengths Lllat have been supplied arc not sufficient for calibrations needed for the range of the infrared region from 0.8 to ] 50).L 01' for t he wid e variety of instrumen ts used in tbis region , espeeiall." witl)' t he differences in resolution ob ta ined with different prism materi als. For ex, ample, the CO and CO2 absorp tion bands between 4 and bJ.L, are well su i ted for calibration bu L canHO (, be resolved in to th eir rotational structUJ'e b,' a rock­sal t prism. The requ irements for the calibration of a spectrometer using a rock-salt prism are much dif­feren t from t hose of an ins trumen t using prisms of li thium HUOl'ide and of calcium flu oride. When the thermocoupJe of an instrumen t is r eplaced by a PbS detector, more bands that conta in the ro ta­t ional s tru cture of gases can be used for calibra tion. Still higher resolu t ion can be ob tained by replac ing the prism in the sp ect rometer with a grating. This usuall~- allows the use of many bands for calibration , and if too high a precision is no t required, many bands that have been measUl'ed on large gra ting instrumen ts can be used.

This repor t docs no t include me thods of calibrn.­tion of high-resoIu Lion gra ting spectrometers. The use of the fringe s." stem of a Fabry-Pero t in te)' , ferometer for accur ate measuremen t of line positions observed wi th grating spectrometers has been pre­viously describ ed [3].

The cal ibra tion poin ts lis ted in the present paper are primarily suited for low .. resolu tion instruments .

2 . Method of Measurement

A large grating spectrometer can easil." m easure sharp spectral lines wi th an error not greater than 1 par t in 100,000, This aCCllracy can be obtained

1 Figures in bmckets indicat( the:litcrature references a(tbe end of th is paper .

b.\' using s tandard lilles of neon, argon , krypton, or xenon , in higher orders that overlap the band being measUl'ed. T he standard spectral lines 01' the fringe s." s tem of a Fabry-Pero t interferometer are recorded simultaneously with th e absorp ti on band being s tudi ed . The gra t ing spectr ometer and the method of calibration are described in reference [3].

Th e meaSUl'ement of the absorpt,ion bands of liquids an d solids under high resolu tion often reveals side bands 0 1' complex strucLure. The way in which t he componen ts add up to produce a seemi ngl.\' sharp ba nd on a prism instrument must b e approximated by the in Lensity and the positions of t he various components. An y band in Lhe range 1 to 10 J.L t hat is found to have an un certa in t \T of more than ± 0.003 ).L is no t lis ted as a wavelengtli for calibratio n.

Another approac h in th e calibra tion of bands is to open the sli t of the grating spectrometer 0 that r esolu tion equal Lo Lhat of a prism spectrometer is ob tained. This meLhodreduces the h igh precision of meaSUl'emen t wi t h the grating spectrometer , bu t wi th the wide range of resolu tion of different specLro­meters, the band shapes could not be match ed for all instrumen ts. I n order Lo avoid the difficul ties of the low-resolution method, all t he bands r epor ted in this work have b een calibra ted wi th high-resolu tion spectrometers.

In the measuremen t of Q-branches of bands, con­siderable fine structure is observed. The Q-branch of the NH3 band at 2.999 ).L is shown in figure 1. There ar e 30 or more separate lines and the band extends from 2.996 to 3.002 J.L; the cen ter on low r esolution is estimated to be 2.9990 ± 0.0005 ).L . All of the bands measured on t he gra ting instrumen t, appear to be broad, and the es timate of the cen ter for low resolution accounts for t he un certainty given with the results listed in the tables.

2 .1. Region From 0.6 to 2 .6 Microns

Although the in tense infrared bands of molecule are observed at wavelengths grea ter than 2 p. , it is sometimes necessary to measure the transm ission of filt ers and the absorp tion of combination bands in the near-infrared r egion . The atomic spectral lines are well sui ted for calibration in this region . The

195

Page 2: Reference wavelengths for calibrating prism spectrometers · 4 and bJ.L, are well su i ted for calibration bu L canHO (, be resolved into their rotational structUJ'e b,' a rock salt

- .. =- -WAVELENGTH, I-L

z o l­n. 0: o en CD

2.97

1\

2.9 8

3360

2.99 3.00

3350 3340 WAVE NUMBER S ,em- I

F I GCRE 1. Absorption spectmm of NH3 in the Q-branch of the 2. 999-micron band with hi gh 1·esolution. The spect rum w as obser ved with a meIer cell a t 5 mm (Hg) pressure.

spectr a of m ercury, neon, argon, kr~-pton , and xenon are rich in near-infrared lines, som e of which have been m easured and listed in another publica­t ion [3].

The atomic spectr al lines are observed in emission, bu t it is usually easier to use calibrating wavelengths in absorption on most infrared spectrometers.

Several bands of polystyr ene, 1,2,4-trichloroben­zene, didymium glass, and carbon disulfid e have b een m easured from 0.6 to 2.6 fJ., and th e wavelengths are listed in table 1. In addition to wavelength s, the wave numbers in vacuum, the state, t he thick­ness, and t he material are tabulated. For spec-

TABLE 1. Calibration wavelengths for p1·ism in slr1l1nenis from 0.6 to 2.6 }J.

" . a velength W ave 'rhiek· Ji. (air) number Slale ness Suhsta nce

(vacuum ) ----------------

cm- 1 mm O. 684 ± 0. 001 14616 Solid ...... 6 Did ymium glass. . 740 ± .00l 13510 ._ . do _ . . . _. n Do.

a.743 ± .002 13455 ... do __ ... . 6 Do . .748 ±. 001 13365 ... do ....... 6 Do . . 808 ± .00l 12373 . _. do_ ._ ... 6 Do .

. 880 ± .001 11361 .. . do __ . ... 6 Do . 1. 067 ±. 002 9369 ... do ...... 6 Do . 1. 220 ±. 002 8195 . _. do _ .. . .. 6 Do . 1. 517 ± .003 6590 ... do _._ .. . 6 Do . 1. 681 ±. 002 5947 Film ______ 0.6 P olystyrene.

1. 918 ± . 003 5212 Solid ...... 6 Didymium glass. 2.144 ±. OOI 4663 Film .. . ... 0.6 P olystyren e.

b 2. 170 ±. 002 4607 ... do _. _._ . . 6 Do. 2. 187 ± . 001 4571 ... do .. . ... .6 Do . 2.222 ±. 002 4499 Liquid . ... 5 Carbon disnlfide.

1. 6606 ± . 0002 6020.3 ... do . . _ ... 0.5 l,2,4-rrrichlorobenzenc . 2. 1526 ± .0002 4644.3 . _. do _. _._. .5 Do . 2.3126 ± .0002 4322.9 . _. do _._ ... . 5 Do . 2. 4030 ± .0002 4160.3 . _. do _ . . . _. .5 Do . Z.4374 ± . 0002 4101. 6 _ . . do_ . . _ .. .5 Do.

2.494 ± .002 4009 . . . do __ . . _. . 5 Do . 2.543 ± . 002 3931 _ . . do _._ . . . .5 Do.

• Average value of t he maxima for 0.740 and 0.748 Ji.. b Principal band obsen·ed with prism spectrometer.

trometers that are able to r esolve the rotational structure of the CO band at 2.34 J.i-, there are provided about 50 lines from 2.3 to 2.5 fJ. t hat arc known to a high accuracy [4]. Table 1 is essentially the same as that given previously [1].

The observed spectrum of didymium glass, 1 ,2,4-trichloro b enzen e, and polystyrene with 10'" resolut ion are shown in figures 2, 3, and 4.

2 .2. Region From 2 .6 to 22 Microns

The region from 2 .6 to 15 fJ. is employed for analysis of many compounds. As there are many automatic instrumen ts in use that give the wave­length readings directly, it is important to haye a

z '2 ...

~ o

WAVELENGTH_

F IGUR E 2 . Absorption sp ectrum of didymium glass, 6 milli­meters thicle, in the Tegion from 0.6 to 2 microns.

196

Page 3: Reference wavelengths for calibrating prism spectrometers · 4 and bJ.L, are well su i ted for calibration bu L canHO (, be resolved into their rotational structUJ'e b,' a rock salt

-- - .. -

~ I

FIGUR E 3. N em'-inJrared spectrum of 1 ,2,4-tTichlorobel1 ze ne fOT a cell thickness of 0.5 milLimeteT.

T z o I­a '" -' ... '" o

::l.. 0 ... 0 "! ,..,

::l.. ::l.. ::l.. ::l..::l.. ,..,

'" '" 0 '" '" ,.., 0

~ ,.., ,.., ,.; .,; ,..;

f-- WAVELENGTH

F I GUR E 5. I nJj'ared abs01'ption bands of polystyrene in the region of 3.4 m ic1'on s for a fiLm about 25 microns thick , as obs erved with a lithium ./lu01·ide p1·ism .

r i few substa nces t ha t can be used Lo check t he cal i­brat ion of the instrumen t. A numb er of absorpt ion bands have been measured for Lhi s r egion. The largest number of bands have bee n selected from the spectrum of polystyrene, toluene, and 1,2,4-t ri­cblorobenzene.

I 6

(

I­a w -' u.. w o

WAVELENGTH -+-FIGURE 4. N ear-inf m red spectrum of polystyrene for a thick­

ness of 0.6 millimeter.

Pol~-sL}Tene fllms are easily made, and Llle spec­trum is su itable 1'0 1' a check of the instrument . Several of the ba nds have been l'emeasured, and the wavelengt hs cbeck closcl? wi th the previous values [1], and an estimate of Lhe accuracy of Lhe wave­lengths has been given.

Figure 5 shows the bands in t he 3.4-1l region of pol.vstyr ene as measured wi th a LiF prism.

Fig ure 6 is a record of the spectrum from 2 to 15 Il , m eas ured wiLh a rock-saIL prism . "\Then Lhe

IOO r-----------------------------------------------------------------------------------------------,

80

~

z 60 Q <f)

'!'. :;, <f)

~ 4 0 0: I-

20

0

FIGURE 6.

__ 3.507

3.3026 ~

3.422 ./'

2 3 4 5

1 1

6.238~

6.692~ 1

6 7 8 WAVELENGTH , f1-

9 10 II 12 13 14 15

I nJmred absorption spectrum, from 2 to 15 microns, of a 50-micron film of polystyrene measured with a rock-salt prism instrument.

197

Page 4: Reference wavelengths for calibrating prism spectrometers · 4 and bJ.L, are well su i ted for calibration bu L canHO (, be resolved into their rotational structUJ'e b,' a rock salt

100 ,---------------------------------------------------------------------------------,

80

Z 60 o u; til

~ til

:i 4 0 <r >-

20 / 3.2908

o L-----~2----~3~----~4----~5~----~6-------'7------~8 -------'9------1~0----~II------1~2------:':13:--~

WAVELENGTH, iJ-

F I GURE 7. Infrared spectr1l1n of toluene from 2 to 15 11licrans,~as:abserved with a rock-salt prism. Cell thickness wa s 0.05 mm.

100

8 0

Z 60 o iii til

~ :Q 4 0 <1 <r >-

20

4 5 6 7 8 9 10 II 12 13 14 15 WAVELENGTH, f-L

FIGURE 8, Spectrwn of 1 ,2,4-trichlorobenzene f rom 2 to 15 microns, as observed with a rock-salt prism. Cell was 0.025 rum, and the insert spectrum from 3. 5 to 5 IJ. was measured with a cell of 0.2 mm .

surfaces of the film are parallel a number of inter­ference fringes appear in the spectrum. The fringes are of low intensity and usually do not change the wavelength of the absorption bands .

100.------------------------------------,

Figure 7 represents the spectrum of toluene. Some of the bands that have been measured fall close to the calibrated bands of polystyrene and of 1,2,4-trichlorobenzene (fig . 8) . Anyone of these substances would be sufficient for checking any changes that may occur in a spectrometer with a wavelength scale.

80

;,!1 .60

Z o til til

~ 4 0 til Z <1 <r f-

20

(~

The absorption spectrum of Corning fused quartz has a strong band at 2.7 11-. This band has been measured for a calibration point in this region , and the center occurs at 2.735 11-. The absorption spec­trum is shown for a thickness of 0.7 mm in figure 9. With a thinner sample the band is not as broad and is less absorbing. This band is produced by occluded water and it is not present in all types of fused quartz.

o L-__ ~ ____ ~ ____ ___' ______ ~ ____ ~==~

Other materials with absorption bands, listed in table 2, arc carbon dioxide, carbon disulfide, methyl-

FIGURE g,

198

2 3 WAVELENGTH ,f-L

4

Band at 2.735 microns 1:n Corning fused sil ica for O.7-millimeter thickness.

,J

\ I

I

Page 5: Reference wavelengths for calibrating prism spectrometers · 4 and bJ.L, are well su i ted for calibration bu L canHO (, be resolved into their rotational structUJ'e b,' a rock salt

TABU:; 2. Calibrat'ing wavelength s from 3 to 24 J.1.

W avclength (a ir)

Wave n um-

ber State (vac-uum)

p. cm-1

D escription Substa nce

2.735 ± 0.0002 3655.3 Solid __ __ 0. 7 mm thick____ Quartz . 2.9990 ± . 0005 3333.5 Gas _____ 1-m cell , 5 mm NIh

(H g). 3.3101 3020.2 __ _ do ____ 5.0-cm cell , 10 M ethanc.

em (H g). 3.2204 3104.4 Solid ____ 25-/' film _____ ___ Polysty rene. 3.2439 ±. 0002 3081. 9 __ _ do _________ do __________ Do.

3. 2668 ±. 0002 3.3026 ±. OOO I 3.3303 ±. 0001 3.422 ± .00l 3. 5070 ±. 0005

5. 138 ±. 005 5. 343 ±. 003 5.549 ±. 002 6.238 6.692

3060.3 ___ do ___ ______ do _____ ____ _ 3027. 1 ___ do _____ ____ do _________ _ 3001. 9 ___ do _________ do _________ _ 2921.5 ___ do ______ ___ d o ___ ______ _ 2860. 7 ___ do _________ do ___ ______ _

1945.8 ___ do ____ 5O-/'fi lm ___ ____ _ 1871. 1 ___ do ______ ___ do _________ _ 1801. 6 ___ do _________ do _________ _ 1602.7 ___ do ______ ___ do ________ _ _ 1493.9 ___ do _________ do _________ _

8.662 llM.2 __ _ do ________ _ do _________ _ 9.724 1028. 1 ___ do _________ do _________ _

11. 035 906.0 ___ do _________ do _ 3.2393 ± .OOO6 3086.2 Liquid __ 0.05-mm eeIL __ _ 3.2908 ± .0003 3038.0 ___ do _______ __ do _________ _

Do. Do. Do. Do. Do.

Do. Do. Do. Do. Do.

Do. Do. Do.

Tolucne. D o.

3.4220 ±. 0009 2921. 5 ___ do _______ __ do __________ Do. 3.4840 ±. 0004 2869.5 ___ do _________ do ____ ______ Do. 3. 6566 ± .0002 2734.0 ___ clo _________ do __________ Do. 5. 14i2 ±. 0007 1942. 3 ___ do _________ do ________ __ Do. 5.3818 ±. 0006 1857.6 ___ do _________ do __________ Do.

5.549 ± .002 21. 52 3.2434 ±. 0003

1801. 6 ___ do _________ do _________ _ 464 .6 __ _ do _________ do _________ _

3082.4 ___ do ____ 0.025-mm eeIL __

3.5564 ±.0002 2811.1 ___ do _________ do ____ _____ _ 3.7067 ± .0002 2697. 1 __ _ do _________ do _ ____ ___ _

5. 3034 ±. 0025 17. '10 18.10 21. 80 22.76

20.56

2. 7144 4.866 9.672 4. 258

13.883 14.98

7.268

11.475 11. 862

1885. 1 __ _ do _________ do _________ _ 574 .6 ___ do _________ do _________ _ 550. 5 ___ do _________ do ___ ______ _ 458.6 ___ do _________ do _________ _ 439.3 ___ do _______ __ do _________ _

486.3 __ _ do ____ 0.05-mm cell (sat. soln. in OS,).

3683. 0 Vapol".. _ 5.0-em eeIL ___ _ _ 2OM.5 ___ do _________ do _________ _ 1033.6 ___ do _________ do ________ _ 2347. 9 Gas _____ Atmosphcric ___ _

720. 1 ___ do _________ do _________ _ 667.4 ___ do _________ do _________ _

1375.5 Liquid __ 0.05-mm ceIL __ _

871. 2 ___ do ______ __ _ do _________ _ 842.8 ___ do _____ ____ do _________ _

Do. Do.

1,2,4.-TricllloJ'o-benzene.

Do. D o.

D o. D o. Do. D o. Do.

1,2,3-Trich loro­benzene ,

Metha nol. Do. Do.

Oarboll dioxide .

Do. Do.

~reth y leyclo­hexane.

Do. Do.

23. 85 419. 2 Va pol'. _ Atmospheric ____ ,Yater.

cyclohexane, and methanol. The wavelengths of the bands of methanol have not been r emeasured but are taken from the work of Borden and Barker [5].

The spectrum of 1,2,4-trichlorobenzene has b een measured 111 the regLOn from 14 to 24 J.i- , figure 10. The absorption bands at 17.40, 18. 16 , and 2l.80 J.i­were measured on a grating spectrometer . The broad band at 22 .76 J.i- was measured on a pnsm spectrometer and I S probably accurate to ± 0.02 J.i- .

The absorption spectra of many gases have been measured and would be useful for calibration. If the prIsm spectrometer cannot resolve the spectra in to the rotational lines, they should not be used. Th e atmospheric-water-vapor lines ill the vibra­tional rotational band are no t properly resolved with prism instruments, but the pure rotation spectrum from 20 to 150 J.i- is well sui ted for calibration. These lines have been measured by Randall , D ennison, Ginsburg, and vveber [6], and their ar ticle lists the observed wavelengths_

The spectrum of methane at 3.3 J.i- contains many rotational lines of t he P- and R-branches, which are spaced at intervals of about 10 cm - 1. This band was recently mea ured wi th high resolu tion, and many of the lines were found to have several components. The spectrum was then recorded with a resolution of 0.3 cm- 1, and the center of the componen ts of a line was estimated by comparmg the results with the high-resolu t ion measurements . A pho tograph of the low-resolution trace IS shown m figure 11 , and t he positions of lines are gLven m table 3. Where partial resolu tion IS obtained m some lines of th e P-branch, the wavelength and wave number are listed as applying to the strongest componen t.

The absorption lines observed bet'l'leen the meth­ane lines are part of the 2V2 band of atmospheric watcr vapor. The values of the wave numbers of the lines reported in this table are probably accurate to ± 0.05 em- I, but a more precise table of wave Ilumbers, including all the components of the lines, will be listed in another paper.

Hydrogen chloride and hydrogen bromide also have many rotational lines in their spectra that fall

700 600 WAVE NUMBERS, em-I

500 450 100

80 ~

z ~ 60 Ul Ul

::E Ul

40 z <l 0:: t-

2 0

0 14

I

, , , ,

15 16 17 Ii 19 20 21 22 23 24 WAVELENGTH , fL

FIG URE 10. Sp ectn t1n of 1 ,2,4-trichl01'obenzene from 14 to 24 m icrons , as observed with a EBr prism.

199 j

Page 6: Reference wavelengths for calibrating prism spectrometers · 4 and bJ.L, are well su i ted for calibration bu L canHO (, be resolved into their rotational structUJ'e b,' a rock salt

a w .. hi

3.448 WAVELENGTH, f-L

• 3.333 3.226

z o I-0.. ~ o (f)

CD «

2900 3000

Q

3 100 WAVE NUMBERS ,em- I

FIGURE 11. The 3.3-micron band of methane uith medium resolution. Celllengtb, 5 cm; pressure, 10 em (H g).

TABLE 3. Rotational lines oj methane in the 3.3-p. band

Wave· W ave Line len!!;th number

(air) (vacuum ) ---------

" cm-1

p" 3.45309 2895. 18 PIl 3. 43937 2906.72 PIO 3.42801 2916.36 p , 3.41571 2926.86 p , 3.40352 2937.34

p , 3.39131 2947.92 p , 3.37952 2958.20 p , 3.36760 2968.67 P, 3.35592 2979.00 p , 3.34552 2988.27

p , 3. 33343 2999.10 R o 3.30070 3028.84 R, 3. 29012 3038.58 R , 3.27969 3048.25 R , 3.26945 3057.79

R , 3.25932 3067.30 R , 3.24932 3076.74 R , 3.23955 3086.02 R , 3.22992 3095. 22 Rs 3.22041 3104.36

R , 3.21103 3113.42 RIO 3.20174 3122.46 Rll 3.19261 3131. 39 R1 2 3. 18365 3140.20 R 13 3.17480 3148.95

R" 3.16610 3157.61 R15 3. 15751 3166.20

in the region from 3 to 5 jJ.. These gases have recently been measured by Thompson and his co-workers [7] on a grating spectrometer, and the positions of the lines are listed in his papers.

The absorption bands listed in tables 1 and 2 are

selected for the calibration of spectrometers of low or medium resolution. The calibration of high­resolution instruments would require atomic-spectral lines or the rotational lines of the bands of gases.

Additional calibrating wavelengths can be ob­tained by u-sing the bands of methane and ammonia, which almost completely covcr the range from 6 to 12 jJ.. The 'wavelengths for these lines have not been listed because no high. precision measurements of these spectra are available, but the current measure­ments in the literature are sufficiently accurate for the calibration of prism spectrometers.

3. References [1] Earle K. Plyler an d C. Wilber P eters, J. Research NBS

45, 462 (1950) RP2159. Nicolo Acquista and Earle K. Plyler, J . R esearch NBS 49, 13 (1952) RP2338.

[2] R. A. Oetjen, Chao-Lou Kao, add H. M. Randall, Rev. Sci. Instr. 13, 515 (1 942). A. R. Downie, M. C. Magoon, Thomas Purcell , and Bryce Crawford, Jr. , J. Opt. Soc. Am. 43, 941 (1953). L. H. Jones, J . Chem. Phys. 24, 1250 (1956).

[3] Earle K. Plyler, L. R. Blaine, and Eugene D . Tidwell, J. Research NBS 55, 279 (1955) RP2630.

[4] Earle K. Plyler, W. S. Benedict, and Shirleigh Silverman, J. Chem. Phys. 20, 175 (1952).

[5] Avis Borden and E. F. Barker, J. Chem. Phys. 6, 553 (1938).

[6] H . M. Randall , D. M. Dennison, N. Ginsburg, and L. R. Weber, Phys. R ev. 52, 160 (1937) .

[7] H . W. Thompson, R. L. Williams, and H. J . Callomon, Spectrochim Acta 5, 313 (1952). I . M. Mills, H . W. Thompson, and R. L. Willi ams, Proc. Roy. Soc. [A] 218, 29 (1953) .

WASHINGTON, November 15 , 1956. ~