12
1 Nickel-containing N-doped carbon as effective electrocatalysts for the reduction of CO 2 to CO in a continuous-flow electrolyzer Nick Daems a,b*,# , Bert De Mot a,# , Daniel Choukroun a , Kevin Van Daele a,b , Chen Li c , Annick Hubin d , Sara Bals c , Jonas Hereijgers a , Tom Breugelmans a,b a Applied Electrochemistry and Catalysis, U Antwerpen, Campus Drie Eiken, Universiteitsplein 1, 2610 Wilrijk, Belgium b Separation & Conversion Technologies, Flemish Institute for Technological Research (VITO), Mol, Belgium c Electron Microscopy for Materials Research (EMAT), University of Antwerp, 2020 Antwerp, Belgium d Research Group Electrochemical and Surface Engineering, Vrije Universiteit Brussel, Pleinlaan 2, 1050 Brussels, Belgium * Corresponding author, email: [email protected] # These authors contributed equally. Physicochemical characterization techniques 1 Thermal analysis The as-prepared electrocatalysts were analyzed for their nickel content using a Q500 thermogravimetric analyzer (TA instruments). The temperature of the furnace was ramped from 40°C to 800°C at 5°C.min -1 under a flow of air (25 sccm), after which it was kept constant until the mass change was below 0.05% min -1 . All samples were loaded on platinum pans. The Ni content was calculated from the residue at 800°C (after subtracting for the blank), assuming complete oxidation of Ni to NiO. 2 XRD X-ray diffraction experiments were performed on a Huber X-ray diffractometer equipped with a G670 Guinier camera (Huber GmbH&Co, Germany) using the cu Kα1 radiation (λ = 1.5405981 Å). Each experiment had a duration of 15 min during which the image plate was scanned 3 times. In total, 4 consecutive measurements were performed and the final intensity value was the sum of all measurements. A few mg of electrocatalyst powder was dispersed in ethanol and grinded before deposition onto a 6 μm Mylar® thin-film. Afterwards, the sample was mounted onto the sample holder and measured. 3 N 2 physisorption Porosity and surface area measurements were performed on a Quantachrome Quadrasorb SI (Quantachrome instruments, Odelzhauze, Germany) automated gas adsorption system using liquid nitrogen as the adsorbate. All the samples were degassed in an AS-6 degasser under vacuum for 16 h at 200°C before analysis. The surface area was calculated using the BET method and the average pore size was deduced as the maximum of the BJH plots. Electronic Supplementary Material (ESI) for Sustainable Energy & Fuels. This journal is © The Royal Society of Chemistry 2019

reduction of CO2 to CO in a continuous-flow electrolyzer

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: reduction of CO2 to CO in a continuous-flow electrolyzer

1

Nickel-containing N-doped carbon as effective electrocatalysts for the

reduction of CO2 to CO in a continuous-flow electrolyzer

Nick Daemsa,b*,#, Bert De Mota,#, Daniel Choukrouna , Kevin Van Daelea,b , Chen Lic, Annick Hubind, Sara Balsc,

Jonas Hereijgersa, Tom Breugelmansa,b

aApplied Electrochemistry and Catalysis, U Antwerpen, Campus Drie Eiken, Universiteitsplein 1, 2610 Wilrijk,

Belgium

bSeparation & Conversion Technologies, Flemish Institute for Technological Research (VITO), Mol, Belgium

cElectron Microscopy for Materials Research (EMAT), University of Antwerp, 2020 Antwerp, Belgium

dResearch Group Electrochemical and Surface Engineering, Vrije Universiteit Brussel, Pleinlaan 2, 1050

Brussels, Belgium

*Corresponding author, email: [email protected]

#These authors contributed equally.

Physicochemical characterization techniques

1 Thermal analysis

The as-prepared electrocatalysts were analyzed for their nickel content using a Q500 thermogravimetric

analyzer (TA instruments). The temperature of the furnace was ramped from 40°C to 800°C at 5°C.min-1 under a

flow of air (25 sccm), after which it was kept constant until the mass change was below 0.05% min-1. All samples

were loaded on platinum pans. The Ni content was calculated from the residue at 800°C (after subtracting for the

blank), assuming complete oxidation of Ni to NiO.

2 XRD

X-ray diffraction experiments were performed on a Huber X-ray diffractometer equipped with a G670 Guinier

camera (Huber GmbH&Co, Germany) using the cu Kα1 radiation (λ = 1.5405981 Å). Each experiment had a

duration of 15 min during which the image plate was scanned 3 times. In total, 4 consecutive measurements were

performed and the final intensity value was the sum of all measurements. A few mg of electrocatalyst powder was

dispersed in ethanol and grinded before deposition onto a 6 µm Mylar® thin-film. Afterwards, the sample was

mounted onto the sample holder and measured.

3 N2 physisorption

Porosity and surface area measurements were performed on a Quantachrome Quadrasorb SI (Quantachrome

instruments, Odelzhauze, Germany) automated gas adsorption system using liquid nitrogen as the adsorbate. All

the samples were degassed in an AS-6 degasser under vacuum for 16 h at 200°C before analysis. The surface area

was calculated using the BET method and the average pore size was deduced as the maximum of the BJH plots.

Electronic Supplementary Material (ESI) for Sustainable Energy & Fuels.This journal is © The Royal Society of Chemistry 2019

Page 2: reduction of CO2 to CO in a continuous-flow electrolyzer

2

4 XPS

X-ray photoelectron spectroscopy (XPS) measurements were performed on a Physical Electronics PHI 1600

multi-technique system using an Al Kα (1486.6 eV) monochromatic X-ray source, operated at 15 kV and 150 W

at a base pressure of 2*10-9 Torr. To correct possible deviations caused by electric charging, the graphitic C1s

band at 284.6 eV was taken as internal standard. For the deconvolution and integration of the XPS signals the

MultiPak software was utilized. A lot of time and effort went in to selecting those parameters (FWHM, peak

position, etc.) to set to get the most accurate and reproducible results. These parameters were kept constant for the

different materials. The red dash line that is visible in the deconvoluted XPS spectra is representative for the

remaining difference between the fit and the actual XPS data.

5 Raman

Room-temperature Raman spectra were recorded on a LabRAM HR Evolution spectrometer from HORIBA

Scientific. The spectroscope was equipped with a high stability confocal microscope with XYZ motorized stage

and objectives of 10 X 50x, 100x magnification, a multichannel air cooled detector (with < 1 cm-1 spectral

resolution) and a solid state laser with a wavelength of 532 nm (Nd:YAG). Peak integration and deconvolution

was done by using the IgorPro software.

6 HAADF-STEM

The samples were dispersed in methanol and drop-casted onto transmission electron microscopy (TEM) grids.

An FEI Osiris electron microscope in the scanning TEM (STEM) mode was employed to image the Ni-N-AC-C

sample using a high-angle annular dark-field (HAADF) detector at 200 kV. A FEI titan electron microscope with

a CEOS probe corrector was employed to acquire atomic resolution images on the Ni-N-AC-B1 in STEM mode

at 300 kV. A converge angle of 22.5 mrad and a camera length of 73 mm was used for imaging. A HAADF detector

with collection angle range 79-200 mrad and a low-angle ADF detector (LAADF) with collection angle range 19-

74 mrad were both used for imaging. Electron energy loss spectroscopy (EELS) was used to identify the single

atoms on the carbon support. The settings for EELS are 0.5 eV/channel for dispersion, 29.5 mm for the camera

length, 5 mm for the EELS aperture and 100 seconds for the total acquisition time. determine the elemental

distributions.

Page 3: reduction of CO2 to CO in a continuous-flow electrolyzer

3

0 5 10 15 20 25 300.00

0.05

0.10

0.15

0.20

0.25

0.30

0.35

0.40

Ni-N-AC-B1

N-AC-pure-B

Pore width (nm)

Pore

vol

ume

(cm

³/g.n

m)

Figure S1. N2 physisorption isotherm of N-AC-pure-B and Ni-AC-B1 (top) and BJH plots of N-AC-pure-B and

Ni-AC-B1 (bottom).

Page 4: reduction of CO2 to CO in a continuous-flow electrolyzer

4

Figure S2. XRD diffractogram of the different electrocatalysts. The peaks are identified as follows: Δ is metallic

nickel and * is nickel hydroxide.

Figure S3. Deconvoluted XPS spectra of Ni-N-AC-B2: N1s (A) and Ni2p (B).

(A) (B)

Page 5: reduction of CO2 to CO in a continuous-flow electrolyzer

5

Figure S4. Deconvoluted XPS spectra of Ni-N-AC-B3: N1s (A) and Ni2p (B).

Figure S5. Deconvoluted XPS spectra of Ni-N-AC-C: N1s (A) and Ni2p (B).

Figure S6. Deconvoluted N1s XPS spectra of N-AC-pure-1 (A) and N-AC-pure-2 (B).

(A) (B)

(A) (B)

(A) (B)

Page 6: reduction of CO2 to CO in a continuous-flow electrolyzer

6

Figure S7. Full spectrum of Ni2p for Ni-N-AC-B1

Figure S8. Charging current densities plotted versus scan rate. Slopes are equal to the capacitance of the sample.

Page 7: reduction of CO2 to CO in a continuous-flow electrolyzer

7

Table S1. Capacitance values as determined by CV measurements at different scan rates and corresponding

roughness factor.

Sample Capacitance (mF/cm²) Roughness factor

N-AC-pure-A 42.2 2.11

N-AC-pure-B 49.4 2.45

Ni-N-AC-A 21.7 1.09

Ni-N-AC-B1 21.7 1.07

Ni-N-AC-B2 21.9 1.08

Ni-N-AC-B3 20.2 1.00

Ni-N-AC-C 32.8 1.62

Figure S9. Linear sweep voltammetry plots of Ni-N-AC-B1 in the absence and presence of CO2 in the 0.5M

KHCO3 electrolyte.

Page 8: reduction of CO2 to CO in a continuous-flow electrolyzer

8

-1.20 -1.10 -1.00 -0.90 -0.80 -0.70 -0.60 -0.50 -0.40 -0.30-25.0

-20.0

-15.0

-10.0

-5.0

0.0

N-AC-pure-A

N-AC-pure-B

Ni-N-AC-A

Ni-N-AC-B1

Ni-N-AC-B2

Ni-N-AC-B3

Ni-N-AC-C

E vs. RHE (V)

Cur

rent

Den

sity

j,no

rm (m

A c

m-2

)

Figure S10. Plots of normalized current density vs. potential measured in CO2 saturated 0.1M KHCO3.

-1.20 -1.10 -1.00 -0.90 -0.80 -0.70 -0.60 -0.50 -0.40 -0.300

10

20

30

40

50

60

70

80

90

100

N-AC-pure-A

N-AC-pure-B

Ni-N-AC-A

Ni-N-AC-B1

Ni-N-AC-B2

Ni-N-AC-B3

Ni-N-AC-C

E vs. RHE (V)

Fara

daic

eff

icie

ncy

H2

(%)

Figure S11. Faradaic efficiencies to H2.

Page 9: reduction of CO2 to CO in a continuous-flow electrolyzer

9

Figure S12. GC chromatograms for Ni-N-AC-B1 at different potentials in the H-cell.

Page 10: reduction of CO2 to CO in a continuous-flow electrolyzer

10

y = 22424xR² = 1

0

5000

10000

15000

20000

25000

0 0.2 0.4 0.6 0.8 1

volp

pm

Area (mV.min)

CO calibration curve

y = 1,009,885xR² = 1

0

5000

10000

15000

20000

25000

0 0.005 0.01 0.015 0.02 0.025

volp

pm

Area (mV.min)

H2 calibration curve

Figure S13. Calibration curves for CO (left) and H2 (right) used to convert the detected peak areas of CO and H2

into concentrations.

The calibration curves in figure S13 were determined in the same setup that was used for the measurement

(however in absence of an applied potential and a catalyst) utilizing standardized gas mixtures. The GC area is

plotted with respect to the available gas concentrations in the certified gas mixtures.

Table S2. Tafel slope as determined from the jCO-η curves.

Sample Tafel slope (mV dec-1)

N-AC-pure-A 360

N-AC-pure-B 267

Ni-N-AC-A 200

Ni-N-AC-B1 185

Ni-N-AC-B2 218

Ni-N-AC-B3 185

Ni-N-AC-C 287

Page 11: reduction of CO2 to CO in a continuous-flow electrolyzer

11

-1 -0.9 -0.8 -0.7 -0.6 -0.5 -0.4 -0.3-20.0

-18.0

-16.0

-14.0

-12.0

-10.0

-8.0

-6.0

-4.0

-2.0

0.0

Ni-N-AC-B1 0.1M KHCO3

Ni-N-AC-B1 0.5M KHCO3

Ni-N-AC-B1 0.5M KHCO3;200 µg loading

η vs. RHE (V)

Cur

rent

Den

sity

jCO

,nor

m (m

A c

m-2

)

-1 -0.9 -0.8 -0.7 -0.6 -0.5 -0.4 -0.30

10

20

30

40

50

60

70

80

90

100

Ni-N-AC-B1 0.1M KHCO3

Ni-N-AC-B1 0.5M KHCO3

Ni-N-AC-B1 0.5M KHCO3;200 µg loading

η vs. RHE (V)

Fara

daic

eff

icie

ncy

CO

(%)

Figure S14. (A) Normalized partial current densities to CO and (B) Faradaic efficiencies to CO measured after

25 min of chrono-amperometry experiments performed at set potentials using different electrolyte concentrations

(0.1 vs. 0.5 M) and different catalyst loadings (785 µg.cm-2 vs. 200 µg.cm-2).

(A)

(B)

Page 12: reduction of CO2 to CO in a continuous-flow electrolyzer

12

0 500 1000 1500 2000 2500 3000 3500-4.5-4.4-4.3-4.2-4.1

-4-3.9-3.8-3.7-3.6

IrO2

Pt

Time (s)

Ece

ll (V

)

Figure S15. Comparison between potential decay at -50 mA.cm-2 for two different anodes in the flow-cell setup

(Fumatech® membrane, anolyte: 2M KOH, katholyte: 1M KHCO3).