6
1 Asher Holland [email protected] RECON MEDICAL GEN 4 TOURNIQUET ULTRA-LOW TEMPERATURE STUDY

RECON MEDICAL GEN 4 TOURNIQUET ULTRA-LOW … · 2020. 11. 17. · Tourniquets tested post-ULTE, or -109.3 °F, for 19 hours had an average tensile strength of 2637.0 lbf. Tourniquets

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

Page 1: RECON MEDICAL GEN 4 TOURNIQUET ULTRA-LOW … · 2020. 11. 17. · Tourniquets tested post-ULTE, or -109.3 °F, for 19 hours had an average tensile strength of 2637.0 lbf. Tourniquets

1

Asher Holland [email protected]

RECON MEDICAL GEN 4

TOURNIQUET ULTRA-LOW

TEMPERATURE STUDY

Page 2: RECON MEDICAL GEN 4 TOURNIQUET ULTRA-LOW … · 2020. 11. 17. · Tourniquets tested post-ULTE, or -109.3 °F, for 19 hours had an average tensile strength of 2637.0 lbf. Tourniquets

2

THE EFFECTS OF ULTRA LOW TEMPERATURE ON RECON MEDICAL GEN 4 TRAUMA TOURNIQUET TENSILE STRENGTH

Asher Holland Project Engineer

Recon Medical, LLC

ABSTRACT In this study we analyze the effects of Ultra-Low

Temperature Environment (ULTE) on Recon Medical

Generation 4 (GEN 4) trauma tourniquet (TQ) tensile strength.

Analysis was compiled by placing 10 tourniquets in a closed

environment with 14.02 pounds of dry ice for 19 hours and

comparing to a control group at ambient temperature.

Tourniquets were individually removed from the ULTE and

immediately tensile strength tested using destructive testing

equipment aptly named Recon Ripper (see figure 1). The

resultant average ultimate tensile strength of the tourniquets

tested post-ULTE was 2637.0 pound-force (lbf). By comparing

this average tensile strength to the average ultimate tensile

strength of tourniquets stored at ambient temperature, which was

2540.0 lbf, we see a 3.8% increase in tensile strength in

tourniquets after being stored in an ULTE. This data proves

Recon Medical GEN 4 trauma tourniquets perform exceptionally

well after prolonged exposure in a controlled ULTE.

The effects of submerging tourniquets in water and then

placing in a closed environment with dry ice for 19 hours was

also considered. Tourniquets were similarly individually

removed from ULTE and immediately tensile strength tested.

Results were inconclusive as ice build-up in hook and loop

fastener caused premature failure. Fortunately, this failure mode

can be prevented by properly staging tourniquet.

INTRODUCTION The components that make up the Recon Medical GEN 4

tourniquet are comprised of a buckle made of nylon injection

molding with a 6061-T6 aluminum core, nylon webbing, nylon

hook and loop fastener, and Kevlar® stitching, all of which are

under massive tension during tourniquet application. However,

unique to Recon Medical tourniquets, this tension is supported

by a “floating”, high-strength strap design in combination with

an aluminum core buckle.

Trauma tourniquets are primarily carried on the outer

garments of military personal who are sometimes required to

operate in extreme, low temperature environments. Not only is

the temperature low, but conditions may even be harsh (i.e. snow,

ice, freezing rain). Ultra-low temperatures and harsh conditions

can have a significant impact on strength of materials as well as

ice build-up can cause tourniquet to not function properly.

Furthermore, abrasion against rocks and other hard surfaces is

likely on exterior of personal equipment. What happens to a

tourniquet when it has been carried by personal in a harsh, ULTE

for an extended period, and is then applied to a trauma patient?

Will the plastic injection molding shatter? Will the strength of

the nylon webbing be compromised and the tourniquet rip before

occlusion to the wound? Will ice build and prevent the tourniquet

from functioning? These are the questions this experiment was

designed to answer.

Figure 1: Recon Ripper

EXPERIMENTAL METHODS A total of 13 tourniquets were placed in a Yeti Tundra 75

hard cooler with 14.02 pounds of dry ice. 3 of the 13 tourniquets

were submerged in water for a total of 30 minutes immediately

prior to placement in cooler. The 3 tourniquets submerged in

water were isolated from other tourniquets in cooler. The

tourniquets were then left overnight in a climate-controlled

room. The following morning the cooler was taken to the testing

facility for tensile strength testing. The tourniquets spent at total

of 19 hours in cooler.

Prior to testing tourniquets in cooler, 3 additional

tourniquets at ambient temperature were tensile strength tested

to settle the control group, as seen in table 1.

A quantity of 10 tourniquets were then individually removed

from cooler and placed on the Recon Ripper. An infrared

thermometer was used to record the temperature of the

tourniquet immediately before beginning tensile test. The nylon

webbing between the buckle and the plate was targeted for

tourniquet temperature measurement. This portion of the

tourniquet can be seen in figure 2 below.

Page 3: RECON MEDICAL GEN 4 TOURNIQUET ULTRA-LOW … · 2020. 11. 17. · Tourniquets tested post-ULTE, or -109.3 °F, for 19 hours had an average tensile strength of 2637.0 lbf. Tourniquets

3

Each individual tourniquet temperature was recorded in

Table 2. The tensile strength test procedure was then executed,

and the ultimate tensile strength results were recorded in Table

2. See appendix, figure 7 for visual of typical failure.

Lastly, the 3 tourniquets that were submerged in water for

30 minutes were tested. Prior to tensile strength testing, the 3

tourniquets had to be forcibly removed from dry ice. Regardless

of extensive ice build-up (see appendix, figure 4), tourniquets

were successfully engaged, and tensile strength tested, and

results can be seen in Table 3.

Figure 2: Temperature Assessment Zone

DATA Table 1 presents tensile strength test data of tourniquets at

ambient temperature. Table 2 presents tensile strength test data

of tourniquets after prolonged ULTE exposure. Table 3 presents

tensile strength test data from tourniquets submerged in water

and then placed in ULTE. Equation 1 was used to determine

overall performance change when comparing ambient to ULT

tourniquet tensile strength

Table 1. Tensile Strength at Ambient Temperature

Sample Ambient

Temp (°F) TQ Temp

(°F) Ultimate Tensile Strength (LBF)

1 75.5 76.2 2570

2 75.5 76.8 2695

3 75.5 74.2 2355

AVG 75.5 75.7 2540.0

Table 2. Tensile Strength Post-ULTE

Sample Ambient

Temp (°F) TQ Temp

(°F) Ultimate Tensile Strength (LBF)

1 75.5 20.2 2860

2 75.5 18.2 2660

3 75.5 18.5 2705

4 75.5 14.4 2700

5 75.5 11.0 2720

6 75.5 9.9 2750

7 75.5 23.2 2460

8 75.5 14.6 2410

9 75.5 22.7 2745

10 75.5 14.6 2360

AVG 75.5 16.7 2637.0

Table 3. Tensile Strength Post-water soak and Post-ULTE

Sample Break Force (lbf)

1 2060* 2 2195* 3 1510*

* Failure due to ice build-up

Eq. 1 – Difference in strength by temperature

𝑆𝑡𝑟𝑒𝑛𝑔𝑡ℎ 𝑐ℎ𝑎𝑛𝑔𝑒 =𝑈𝐿𝑇 𝑎𝑣𝑔 − 𝐴𝑚𝑏𝑖𝑒𝑛𝑡 𝑎𝑣𝑔

𝐴𝑚𝑏𝑖𝑒𝑛𝑡 𝑎𝑣𝑔𝑥100

% 𝐼𝑛𝑐𝑟𝑒𝑎𝑠𝑒 =2637.0 − 2540.0

2540.0𝑥100

% 𝐼𝑛𝑐𝑟𝑒𝑎𝑠𝑒 = 3.8%

RESULTS

Control tourniquets tested at ambient air temperature had an

average tensile strength of 2540.0 lbf.

Tourniquets tested post-ULTE, or -109.3 °F, for 19 hours

had an average tensile strength of 2637.0 lbf.

Tourniquets submerged in water and then stored in ULTE

for 19 hours returned inconclusive results due to ice build-up.

As seen in Eq. 1, tourniquets revealed a 3.8% increase in

ultimate tensile strength after being exposed to ULTE.

A special condition was imposed on 1 tourniquet from the

set of 3 submerged in water. After removal from dry ice and prior

to tensile strength testing, a large impact force was repeatedly

imposed on tourniquet using a metal rod. The purpose was to

simulate abrasion from being worn on external personal

equipment and determine whether abrasion would have an effect

on tensile strength. The results were inconclusive as tourniquet

failed tensile strength test prematurely due to ice build-up.

DISCUSSION

Dry ice is the solid form of carbon dioxide. Carbon dioxide

naturally occurs in our atmosphere as a gas, but it can be

solidified into hard, white blocks with an extremely low surface

temperature. Rather than melting due to heat, dry ice slowly

reverts to a gaseous carbon dioxide form through a process called

sublimation. A block of dry ice maintains a surface temperature

of -109.3 °F, or -78.5 ℃. Although the surface temperature for

dry ice far exceeds any expected tourniquet operating

environment, it was not feasible to invest in a laboratory freezer

for this experiment. Therefore, dry ice was chosen as the agent

for achieving an ULTE.

Page 4: RECON MEDICAL GEN 4 TOURNIQUET ULTRA-LOW … · 2020. 11. 17. · Tourniquets tested post-ULTE, or -109.3 °F, for 19 hours had an average tensile strength of 2637.0 lbf. Tourniquets

4

Several variables are introduced when using dry ice such as

having no control over the rate at which the dry ice sublimates.

There is also no way to ensure uniform cooling of the

tourniquets. Recording the temperature using an infrared

thermometer immediately after removal from cooler and before

tensile testing was an effort to gain greater insight into the

effectiveness of the dry ice. As seen in Table 2, column “TQ

Temp”, the average tourniquet temperature was 16.7 °F. This is

significantly warmer than the actual temperature of dry ice and

is likely due to varying thermal conductivity, or rate at which

heat is transferred, in tourniquet components.

Variation occurred also in individual tourniquet temperature

readings, again seen in Table 2, column “TQ Temp.” This

variation is due to several limiting factors in infrared

thermometer operation. The accuracy and resolution of the

Etekcity Lasergrip 1080 infrared thermometer is ±2% and 0.1°F,

respectively. An infrared thermometer accuracy is affected by

distance from surface being measured and consideration was

given to this variable but measuring from the exact same distance

was not feasible in this experiment. Surface emissivity is also

important to consider in no-contact temperature readings and the

emissivity of the nylon webbing is not known. Further research

to determine the emissivity could contribute to a more repeatable

and accurate temperature reading.

The Yeti cooler used to contain the 14.02 pounds of dry ice

and tourniquets is not a true closed environment. The tourniquets

were positioned on the bottom of the cooler beneath the dry ice

per dry ice manufacturer recommendation. Although Yeti coolers

are high grade insulators, they are not able to prevent warm

ambient air from entering ULTE. The performance specifications

for the Yeti cooler used are not available from the manufacturer.

This could have also caused variation in tourniquet temperature.

The 19 hours tourniquets spent in ULTE may not have been

an adequate time duration for tourniquets to reach a uniform

temperature. Further considering the material properties of each

component of the tourniquet in future experiments may give a

more accurate time duration necessary for tourniquets to reach a

uniform temperature.

The time each tourniquet is in transition from cooler to

tensile strength test varies due to experiment methods. The

ambient air was an average of 75.5°F. The rate at which the

tourniquets absorb heat and return to ambient temperature is not

known but assumed to be a very fast rate given the temperature

differential between that of dry ice and ambient air. Performing

the ultimate tensile strength testing in a climate-controlled

environment and limiting time to transfer would reduce the

opportunity for heat absorption in the sample.

Secondary testing involving tourniquets submerged in water

was also completed. As mentioned previously, tourniquets were

submerged for 30 minutes and then placed in cooler with dry ice.

Upon removal the temperature was not measured prior to tensile

strength testing as temperature and its effect on tensile strength

was not the focus. Rather the effects of water and potential for

ice build-up and its effect on tensile strength were of interest.

As you can see in Table 3, the tensile strength of the

tourniquet was negatively impacted by submerging in water. The

failure mode was abnormal and in fact due to extensive ice build-

up in the single-component hook and loop material on the

tourniquet strap (see appendix, figure 3). This ice build-up

prevented the hook and loop from properly securing and

therefore delaminated at the break force listed in table 3 (see

appendix, figures 5 and 6). Fortunately, this failure mode can be

prevented by staging tourniquet before stowing on person. By

adhering terminating end of tourniquet to itself, water intrusion

and furthermore ice build-up is effectively reduced. Further

research is necessary to confirm properly prepped tourniquets

submerged in water would exhibit a tensile strength comparable

to results in tables 1 and 2.

Lastly, the single tourniquet repeatedly impacted with a

metal bar visually maintained structural integrity significantly

well (see appendix, figure 8) and did not fail tensile strength test

due to damage incurred from impact forces. Instead, tourniquet

failed due to ice build-up. This is not surprising as primary

weight bearing components of tourniquet, the 1” wide nylon

webbing and the aluminum buckle core, are internal components

and are shielded from external factors. Therefore, it is reasonable

to assume abrasion against rocks and other hard surfaces pose

little risk to tourniquet tensile strength.

CONCLUSION The Recon Medical GEN 4 trauma tourniquet can withstand

up to 2200 pound-force, a 150% increase in strength over its

predecessor, the Generation 3. Significant improvements were

made with the GEN 4 including an aluminum buckle design - a

6061-T6 core with nylon overlay. Although the force required to

occlude blood flow varies greatly from person to person and

location of tourniquet application, there is little doubt that the

GEN 4 tourniquet has the capacity to sustain the force required

to stop the bleed and save lives – and now this same life-saving

technology can safely be recommended for use in harsh, extreme

cold and artic operational environments. This study confirms the

GEN 4 tourniquet performs exceptionally well at very low

temperatures, and by correctly staging the GEN 4 tourniquet you

can rest assured no ice build-up will prevent tourniquet from

operational success.

Page 5: RECON MEDICAL GEN 4 TOURNIQUET ULTRA-LOW … · 2020. 11. 17. · Tourniquets tested post-ULTE, or -109.3 °F, for 19 hours had an average tensile strength of 2637.0 lbf. Tourniquets

5

APPENDIX

Figure 3: Hook and loop fastener ice build-up Figure 4: Extensive ice build-up upon removal

from cooler

Figure 5: Premature failure due to ice build-up

on single-component hook and loop fastener

Figure 6: Premature failure due to ice build-up

on single-component hook and loop fastener

Page 6: RECON MEDICAL GEN 4 TOURNIQUET ULTRA-LOW … · 2020. 11. 17. · Tourniquets tested post-ULTE, or -109.3 °F, for 19 hours had an average tensile strength of 2637.0 lbf. Tourniquets

6

Figure 8: Impact force tourniquet prior to tensile

strength testing Figure 7: Typical Failure