11
TBME-00524-2006-R3-preprint 1 AbstractThis study shows that even a simple Proportional- Integral-Derivative (PID) controller can be used in a clinical MRI system for real-time navigation of a ferromagnetic bead along a pre-defined trajectory. Although the PID controller has been validated in vivo in the artery of a living animal using a conventional clinical MRI platform, here the rectilinear navigation of a ferromagnetic bead is assessed experimentally along a two-dimensional (2D) path as well as the control of the bead in a pulsatile flow. The experimental results suggest the likelihood of controlling untethered microdevices or robots equipped with a ferromagnetic core inside complex pathways in the human body. Index TermsReal-time control, magnetic resonance, minimally invasive interventions, blood vessels, ferromagnetic I. INTRODUCTION M inimally invasive interventions (MII) are a modality increasingly considered by physicians as a way to reduce potential risks and injuries incurred to patients while allowing shorter recovery periods. One of these modalities, catheterism, offers an efficient way to reach remote areas. Nevertheless, being restricted by its radius of curvature and frictions on the endothelial membrane, catheters have limitations in sinuous and narrow vessels. Although the use of catheters or probes remain advantageous in many cases, new techniques aimed to alleviate Copyright (c) 2006 IEEE. Personal use of this material is permitted. However, permission to use this material for any other purposes must be obtained from the IEEE by sending an email to [email protected] . Manuscript received July 26, 2006; revised April 24, 2007. This work was supported in part by a strategic grant from the Natural Sciences and Engineering Research Council of Canada (NSERC) and in part by the Canada Research Chair (CRC) in Conception, Fabrication and Validation of Micro/Nanosystems and grants from the Canadian Funds for Innovation (CFI). Asterisk indicates corresponding author. S. Tamaz was with École Polytechnique de Montréal (EPM), Montréal, QC, H3C 3A7, Canada. R. Gourdeau and J.-B. Mathieu are with École Polytechnique de Montréal (EPM), Montréal, QC, H3C 3A7, Canada. A. Chanu is with iMetrik, Montréal, QC, H3C 3X6, Canada. *S. Martel is with École Polytechnique de Montréal (EPM), Montréal, QC H3C 3A7, Canada (e-mail: [email protected], www.nano.polymtl.ca ). these limitations are beginning to appear in the medical field. In [1] for instance, a camera capsule pill is carried naturally through the gastrointestinal pathway and acts as an untethered endoscope. Nonetheless, MII involving cardiovascular navigation of untethered devices for applications such as highly localized drug delivery for deep-seated tumor cells, thermal treatment at specific sites, diagnosis or surgeries, and conveyance of magnetic resonance imaging (MRI) contrast mediums to name but a few possibilities, may provide highly valuable tools in the future to the medical practitioners. Hence, in order to reach remote areas within the complex cardiovascular pathway, magnetic propulsion [2-9] offers an advantage at such a scale over other proposed actuation methods [10, 11] for operation in the human body. Other demonstrations of controlling the displacement of a ferromagnetic device in vitro using magnetic gradients [2, 9] for future potential uses in humans have been performed without initially considering during the tests, an imaging modality capable of localizing such a device inside the human body. Here, the displacement control of a ferromagnetic core is performed using the imaging gradient coils of a clinical MRI system in order to validate the approach under the constraints imposed by such a platform capable of providing feedback position of the device inside the human body. Considering existing constraints, such as delays imposed by MR-imaging sequences, or communication between hardware and software modules in the MRI platform, to name but only two examples, the results would reflect a more realistic view of controlled navigation of a ferromagnetic device inside the human cardiovascular network. In sight of such an application, an MRI system by comparison to other apparatus found in clinics is an ideal platform, providing not only enhanced soft tissue imaging contrasts, but also magnetic gradient propulsion, imaging, interfaces, and the control capabilities required for endovascular navigation. Reference [12] mentions the difficulty of propelling microdevices in a cardiovascular system due to the relative wide range of vessels’ diameters. Thus, the control algorithm presented here is dedicated for macro scale navigation, which focuses in conveying the device in vessels such as arteries and arterioles whereas microscale navigation is introduced in [13]. The damage that the bead will induce on the vessel’s walls, either by collision or friction, is considered to be negligible. Real-time MRI-based Control of a Ferromagnetic Core for Endovascular Navigation Samer Tamaz, Richard Gourdeau, Member, IEEE, Arnaud Chanu, Jean-Baptiste Mathieu, Student Member, IEEE-EMBS, and Sylvain Martel*, Senior Member, IEEE

Real-time MRI-based Control of a Ferromagnetic Core for ...wiki.polymtl.ca/nano/images/0/02/J-2008-MRSUB-TBME-Samer.pdf · another along a pre-determined path using MRI-based navigation

  • Upload
    others

  • View
    0

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Real-time MRI-based Control of a Ferromagnetic Core for ...wiki.polymtl.ca/nano/images/0/02/J-2008-MRSUB-TBME-Samer.pdf · another along a pre-determined path using MRI-based navigation

TBME-00524-2006-R3-preprint 1

Abstract—This study shows that even a simple Proportional-

Integral-Derivative (PID) controller can be used in a clinical MRI

system for real-time navigation of a ferromagnetic bead along a

pre-defined trajectory. Although the PID controller has been

validated in vivo in the artery of a living animal using a

conventional clinical MRI platform, here the rectilinear

navigation of a ferromagnetic bead is assessed experimentally

along a two-dimensional (2D) path as well as the control of the

bead in a pulsatile flow. The experimental results suggest the

likelihood of controlling untethered microdevices or robots

equipped with a ferromagnetic core inside complex pathways in

the human body.

Index Terms—Real-time control, magnetic resonance,

minimally invasive interventions, blood vessels, ferromagnetic

I. INTRODUCTION

Minimally invasive interventions (MII) are a modality

increasingly considered by physicians as a way to reduce

potential risks and injuries incurred to patients while allowing

shorter recovery periods. One of these modalities, catheterism,

offers an efficient way to reach remote areas. Nevertheless,

being restricted by its radius of curvature and frictions on the

endothelial membrane, catheters have limitations in sinuous

and narrow vessels.

Although the use of catheters or probes remain

advantageous in many cases, new techniques aimed to alleviate

Copyright (c) 2006 IEEE. Personal use of this material is permitted.

However, permission to use this material for any other purposes must be

obtained from the IEEE by sending an email to [email protected].

Manuscript received July 26, 2006; revised April 24, 2007. This work was

supported in part by a strategic grant from the Natural Sciences and

Engineering Research Council of Canada (NSERC) and in part by the Canada

Research Chair (CRC) in Conception, Fabrication and Validation of

Micro/Nanosystems and grants from the Canadian Funds for Innovation

(CFI). Asterisk indicates corresponding author.

S. Tamaz was with École Polytechnique de Montréal (EPM), Montréal,

QC, H3C 3A7, Canada.

R. Gourdeau and J.-B. Mathieu are with École Polytechnique de Montréal

(EPM), Montréal, QC, H3C 3A7, Canada.

A. Chanu is with iMetrik, Montréal, QC, H3C 3X6, Canada.

*S. Martel is with École Polytechnique de Montréal (EPM), Montréal, QC

H3C 3A7, Canada (e-mail: [email protected],

www.nano.polymtl.ca).

these limitations are beginning to appear in the medical field.

In [1] for instance, a camera capsule pill is carried naturally

through the gastrointestinal pathway and acts as an untethered

endoscope. Nonetheless, MII involving cardiovascular

navigation of untethered devices for applications such as

highly localized drug delivery for deep-seated tumor cells,

thermal treatment at specific sites, diagnosis or surgeries, and

conveyance of magnetic resonance imaging (MRI) contrast

mediums to name but a few possibilities, may provide highly

valuable tools in the future to the medical practitioners.

Hence, in order to reach remote areas within the complex

cardiovascular pathway, magnetic propulsion [2-9] offers an

advantage at such a scale over other proposed actuation

methods [10, 11] for operation in the human body. Other

demonstrations of controlling the displacement of a

ferromagnetic device in vitro using magnetic gradients [2, 9]

for future potential uses in humans have been performed

without initially considering during the tests, an imaging

modality capable of localizing such a device inside the human

body. Here, the displacement control of a ferromagnetic core

is performed using the imaging gradient coils of a clinical MRI

system in order to validate the approach under the constraints

imposed by such a platform capable of providing feedback

position of the device inside the human body. Considering

existing constraints, such as delays imposed by MR-imaging

sequences, or communication between hardware and software

modules in the MRI platform, to name but only two examples,

the results would reflect a more realistic view of controlled

navigation of a ferromagnetic device inside the human

cardiovascular network. In sight of such an application, an

MRI system by comparison to other apparatus found in clinics

is an ideal platform, providing not only enhanced soft tissue

imaging contrasts, but also magnetic gradient propulsion,

imaging, interfaces, and the control capabilities required for

endovascular navigation.

Reference [12] mentions the difficulty of propelling

microdevices in a cardiovascular system due to the relative

wide range of vessels’ diameters. Thus, the control algorithm

presented here is dedicated for macro scale navigation, which

focuses in conveying the device in vessels such as arteries and

arterioles whereas microscale navigation is introduced in [13].

The damage that the bead will induce on the vessel’s walls,

either by collision or friction, is considered to be negligible.

Real-time MRI-based Control of a

Ferromagnetic Core for Endovascular

Navigation

Samer Tamaz, Richard Gourdeau, Member, IEEE, Arnaud Chanu, Jean-Baptiste Mathieu, Student

Member, IEEE-EMBS, and Sylvain Martel*, Senior Member, IEEE

Page 2: Real-time MRI-based Control of a Ferromagnetic Core for ...wiki.polymtl.ca/nano/images/0/02/J-2008-MRSUB-TBME-Samer.pdf · another along a pre-determined path using MRI-based navigation

TBME-00524-2006-R3-preprint 2

Nevertheless, to insure a smooth conveyance of the bead to

destination, collisions and the risk that the bead gets trapped in

the endothelium need to be minimized and as such, an

appropriate explorative 2D control strategy is adopted based

on the PID controller. This control algorithm can be easily

adapted to 3D and tested once the MRI is equipped with

additional coils capable of generating magnetic gradients

susceptible to levitate the bead.

Controlling a device in the human cardiovascular system

proves to be quite challenging due to the non-negligible

pulsatile blood flow, whose variations in waveform amplitude

and frequency from one vessel to another, add complexity in

the conveyance of the device to the targeted area through a

pre-assigned path. In such a complex environment, proper

delay in the software architecture is essential to sustain stable

control of the device and guarantee real-time navigation.

Operating in an MRI system prevents propulsion and

tracking to be applied simultaneously therefore adding another

major constraint to such a control application. Hence, by

alternating between these two processes, we demonstrate that it

is still possible to convey the device from one location to

another along a pre-determined path using MRI-based

navigation capabilities. This path is acquired at a preliminary

stage of the intervention and fragmented into rectilinear

segments [14] which are then delimited by waypoints. Along

each segment, a PID-based feedback controller calculates the

propulsion command to be applied to the device based on its

acquired position [15, 16]. The operation modes of the

tracking, propulsion and control subsystems are scheduled by

the MRI software architecture [17,18].

This paper presents the first 2D in vitro control experiments

using a clinical MRI system.

II. METHOD

A. Software Architecture

The presented architecture allows the guidance of a

ferromagnetic microdevice through the use of a standard real-

time clinical MRI system [19] in a specially designed

phantom. In this study, a Siemens Avanto 1.5 T clinical MRI

system is used. The architecture and sequence presented here

are analyzed in terms of their impact on the controller needs

and the minimum reachable real-time feedback time.

1. Real-time MRI Environment

The propulsion and navigation of the device is made

possible through a seamless interaction with the MRI imaging

gradient coils that are able to induce a force on the

ferromagnetic core as explained in more details in Sect. II.A.3.

Such propulsion capabilities are thus intimately linked with the

MRI specifications, allowing the displacement of the device.

However, another major key aspect in the feasibility of the

software architecture is the real-time aspect of the Siemens

Avanto MRI system. Since most clinical and interventional

MRI systems are now engineered with real-time capabilities,

an application such as the MR-Sub project [20] is feasible on

most upcoming hardware systems.

The environment considered here includes two main

modules that are used during every Siemens MRI acquisition.

The first module known as Integrated Development

Environment for Applications (IDEA) is the sequence module

used to design and run the developed sequences for propulsion

and tracking. The other module is the Image Calculation

Environment (ICE) and is mainly responsible for the image

reconstruction using the k-space data acquired during the

sequence. A common sequence would then excite the protons

using an RF wave and encode the frequencies and phases of

the proton volume in the k-space matrix. Each line of the k-

space would be sent to the ICE module for image construction

by inverse Fourier Transforms. Using the provided real-time

MRI capabilities, one can react on the running sequence from

the ICE module. That is, gradient amplitudes and directions,

RF wave and ADC parameters can be changed on the fly

through the use of the real-time special framework. Such real-

time changes can trigger special events in the sequence and

generate new data for the ICE module, hence creating a real-

time feedback loop between the two modules. By integrating a

real-time controller in the ICE real-time framework, it is

possible to constantly monitor the acquired data for device

tracking and apply the required gradient strength and direction

for propelling the device towards a given target.

2. System Overview

An overview of the system software architecture is depicted

in Fig. 1. It illustrates all the required components and

interactions between the modules to allow the 3D navigation

of a microdevice in complex vascular networks. As shown in

Fig. 1, three sub-modules (Positioning, RoadMap and

Gradients calculation) and one Main Agent embedded in the

ICE module were implemented to allow propulsion and

steering control of the microdevice. For endovascular

navigation, high definition angiography images for target

selection by a potential user may be required. The RoadMap

module takes care of this operation by gathering the

appropriate image of the region of interest until the user has

made the target selection. The target position is then

transmitted to the main agent for final destination coordinates.

But at present, this RoadMap module is only used to embed a

pre-selected path into the architecture at runtime. This path is

in the form of a text file containing the x, y and z positions of

the desired waypoints to be followed by the device. The

Position module is responsible for the device’s position

calculation, based on the acquired projections — rather than

standard MR 2D images that would be affected by artifacts —

through an MR-sequence based on magnetic signature

selective excitation [21-23]. These projections are correlated

with an initial position mask, constructed during the first

acquisition of the sequence. The maximum amplitude from the

correlation is used as the actual shift in position on a given

axis in the MRI referential.

These positions are then used by the Gradients calculation

Page 3: Real-time MRI-based Control of a Ferromagnetic Core for ...wiki.polymtl.ca/nano/images/0/02/J-2008-MRSUB-TBME-Samer.pdf · another along a pre-determined path using MRI-based navigation

TBME-00524-2006-R3-preprint 3

module to compute the correct gradient amplitudes for the next

waypoint in the planned path to the final target. These data are

sent to the Main Agent that acts as a manager for

communication between both sides of the framework, namely

IDEA and ICE.

The Main Agent sends the new gradient vector to the

running sequence through the real-time framework. The main

concern in this system is the time delay for a given feedback

loop. As described in more details in Sect. II.A.3, this delay is

responsible for the controller’s sample time and is directly

proportional to the complexity of the controller itself.

3. Suggested Architecture and Sequence Design

In order to evaluate the most appropriate control

environment, we developed a 3D navigation capable sequence

for 1D and 2D experiments for which the presented controller

(see Sect. II.B) is implemented. We present in Fig. 1 the

selected sequence and illustrate the three aspects relevant for

the controller’s performance; i.e. the time delay between a

position request on a given axis and the actual propulsion of

the device on the same axis denoted tdelay, the sampling time h

that represents the time between two successive position

requests, and the feedback time which stands for the minimum

allowed delay between a call to the ICE real-time controller

function and the return to the paused running sequence

denoted tfeedback. Fig. 2 illustrates the timing and event

occurrences of the sequence design for 3D control.

The Tracking module considers the first acquired position

as the origin of the Cartesian reference mark. From data in our

previous experiments [21,22] the minimal time to acquire the

xyz-coordinates of a current position is tgpos = 16 ms. The

minimal paused running sequence obtained experimentally is

tfeedback = 30 ms, which is closely related to the complexity of

the controller as noted in Sect. II.A.2.

Fig. 1 Overview of the software system architecture adapted from [17].

Fig. 2. Overview of the real-time sequence for the 3D control environments with time multiplexed positioning and propulsion phases.

Page 4: Real-time MRI-based Control of a Ferromagnetic Core for ...wiki.polymtl.ca/nano/images/0/02/J-2008-MRSUB-TBME-Samer.pdf · another along a pre-determined path using MRI-based navigation

TBME-00524-2006-R3-preprint 4

Therefore, the time left to apply the propulsion gradients is

tgprop = tfeedback - tgpos = 14 ms. Experiments are thus carried out

with h = tfeedback + tadc = 33.8 ms, and tdelay = h + tgpos + tadc =

53.7 ms as depicted in Fig. 1. A minimal tfeedback is chosen to

minimize tdelay. Therefore a duty cycle tgprop/h = 41.37 % is

obtained.

Fig. 3. MRI Cartesian coordinates referential.

B. Control

To ensure the rectilinear motion of the core along the

segments traveled by the ferromagnetic bead in 2D in the

presence of a flow, the commands applied along the x and z

axes are determined based on Tangent-Normal-Binomial

(TNB) frame and then converted to the Cartesian coordinate

system. The resulting command is directly proportional to the

magnetic force generated by the MR-imaging coils defined as

HMVF mm

(1)

where is the permeability of free space,

mV is the

magnetic volume of the ferromagnetic bead, M

its

magnetization, H

is the magnetic field and the gradient

operator.

Fig. 3 illustrates the orientation of the Cartesian system with

respect to the MRI bore. A discrete regulator makes use of a

PID controller acting along the tangent direction of the

trajectory segment and a PD controller acting along the normal

direction (represented as N

in Fig. 4). When the bead

navigates along a vessel subjected to a flow, the normal steady

state error is negligible compared to the tangential error.

Therefore, the integral term of former controller is discarded.

e

e

Ne

x

Te

x

e

x

),( zx

N

x

T

x

),( ww zx

x

re

x

z

wr

Fig. 4. Cartesian and polar systems in the horizontal plane of the MRI

system.

The Laplace transfer function of the tangent controller is

expressed as

s

KsKK

sE

sUsK TI

TDTP

T

TT

)(

)()( . (2)

Using the backward rule transformationzh

zs

1, the

discrete transfer function of the tangent controller becomes

1

1

)(

)()(

z

zhK

zh

zKK

zE

zUzK TITDTP

T

TT

. (3)

In (3), UT is the magnitude of the command along the

tangent direction of the current segment wr

, with KTP, KTD,

and KTI representing the tangent proportional, derivative, and

integral gains respectively, ET represents the tangent quadratic

distance separating the targeted position of the bead from its

current position, defined as

)()sin()()( eerT signzEzE (4)

where αe represents the angle formed by the quadratic error

vector re

and the normal quadratic error vector Ne

. The third

term on the r.h.s. of (4) acknowledges the sign of the error so

that tangent error vector Te

is always pointing towards the

waypoint.

The time based error is the quadratic error defined as

22)()()( tzztxxte wwr

(5)

where wx and wz are the waypoint’s coordinates, and )(tx

and )(tz are the current position’s coordinate with respect to

time. The angle from (4) is defined as

ww

ww

xx

zz1tan

(6)

where wx and wz are the previous waypoint’s coordinates.

The commands along the x and z axes are derived

respectively by the following equations

)()sin()()(

)()cos()()(

zUzUzU

zUzUzU

zNTz

xNTx

(7)

where UNx and UNz represent respectively the magnitudes of

the command along the x and z coordinates which are applied

along the normal direction of the segment wr

, as described

by the following equations

).sin()()sin()()(

)cos()()cos()()(

vNNDeNNPzN

vNNDeNNPxN

zVKzEKzU

zVKzEKzU

(8)

The angles from (8) e and

v are respectively formed by

Page 5: Real-time MRI-based Control of a Ferromagnetic Core for ...wiki.polymtl.ca/nano/images/0/02/J-2008-MRSUB-TBME-Samer.pdf · another along a pre-determined path using MRI-based navigation

TBME-00524-2006-R3-preprint 5

the positive x-axis and the normal error EN and the normal

velocityNV . The angle e from (4) is defined as

.)(

)(tan 1

txx

tzz

w

w

e

(9)

A linear normalization is used to ensure that the commands

applied do not exceed the gradient limits of the x and z axes. A

saturator is applied to the integral component of the PID in

order to avoid the integral windup effect. A formal stability

analysis based on Lyapunov was undertaken but its full report

is beyond the limits of the manuscript.

C. Model

A simplified model for simulating the 3D MRI-based

navigation of a ferromagnetic spherical bead in a fluidic in

vitro environment is considered. Besides the magnetic force

exposed in Sect. II.B, the model regroups four other major

forces acting on the bead as illustrated in Fig. 5. The fluid drag

is expressed as

u

uCAuD Dtf

2

2

1 (10)

where ρf is the fluid density, u

is the relative velocity

between the bead and the fluid, u

it’s norm, At is the cross-

sectional area of the bead, and DC is the drag coefficient for a

sphere defined as

;4.0Re1

6

Re

24DC .102Re0 5

(11)

In (11), Re is the Reynolds number and it is defined as

bf du

Re (12)

where μ is the fluid viscosity, and db is the diameter of the

bead. The infinite sign in (11) indicates that the formula do not

take into account the wall effect. The weight of the bead is

defined as

gV W

(13)

with ρ underlying the mass density of the bead. The

buoyancy force is expressed as

gV b

f (14)

where V is the physical volume of the bead and g

is the

gravitational acceleration.

As explained in [20] a ferromagnetic sphere propelled on a

surface inside the MRI bore will roll along the x-axis and will

slide along the z-axis since the z-axis has the same orientation

as the DC flux density Bo (here at 1.5 T). Therefore, both types

of frictions force have been taken into account. The sliding

friction force is defined as

bWf xs

(15)

with x

is the sliding friction coefficient, which can either

be statics

or kinetic k . The rolling friction force [24] is

defined as

bWR

aff

g

r

(16)

where R is the radius of the bead, gf is the rolling friction

coefficient. In (16), a is the radius of the contact circle

between both objects (bead, static surface) defined as

.4

)(3 31

*E

rbWa (17)

In (17), the Young's modulus *E and the equivalent radius

r are defined respectively as

21

2

2

2

1

2

1

*

111

,111

RRr

E

v

E

v

E (18)

where 1v and

2v represent the Poisson's ratio of the static

and the moving object respectively, 1E and

2E their Young's

modulus and 1R and

2R their radii.

z Dz

W

FMz

FMx

x

y

b

fz

Dx

fx

Bo

Trajectory

Fig. 5. Kinematics and dynamic motion description of the bead in the

horizontal 2D plane.

III. EXPERIMENTS

Two types of experiments were carried out in order to

evaluate the performance of the controller in a 2D

environment. At first, a quiescent flow environment in a two

dimensional environment followed by a pulsatile flow along

one dimension only. All experiments used a chrome steel bead

of 1.5 mm diameter [25] and mass density ρ = 8410 kg/m3. A

value of 1.35x106 A/m (1.69 T) for the saturation

magnetization of the bead was measured with a VSM (Walker

Scientific MG-50) under an applied field Bo = 1.5 T. The

Page 6: Real-time MRI-based Control of a Ferromagnetic Core for ...wiki.polymtl.ca/nano/images/0/02/J-2008-MRSUB-TBME-Samer.pdf · another along a pre-determined path using MRI-based navigation

TBME-00524-2006-R3-preprint 6

maximum peak magnetic gradient amplitudes along the x and z

axes were set to ±40 mT/m.

The immersion of both setups in water is required in order

to acquire the position of the sphere with an MR-tracking

technique using RF excitation of the surrounding protons

excited by the bead’s magnetic field [21,22].

The 2D controller described in Sect. II.B is used to perform

the closed loop control experiments. According to [26], the

static friction coefficient between steel and

polymethylmethacrylate (PMMA) is 0.5. Taking into account

these parameters, the simulations were carried out with

MATLAB/SIMULINK whose block diagram is shown in Fig.

6. Based on the simulation, the control gains were adjusted

heuristically in order to minimize the steady state error, the

oscillations and the percentage overshoot.

Fig. 6. Block diagram of the discrete closed loop control system.

Fig. 7. 2D experimental setup in quiescent flow.

A. 2D Quiescent Flow Control Experiment

1. 2D Quiescent Flow Setup

The ferromagnetic sphere is placed on a thin rigid PMMA

plate that is mounted on three threaded nylon rods equipped

with nylon nuts which are tightened to a thick PMMA plate

forming the base of the setup. By adjusting nuts height on the

threaded rods, the horizontality of the thin plate is attained

based on an air level. A PMMA lid is placed on the thin plate

to trap the bead that allows a 3 mm vertical gap. The lid is

fastened using rubber bands as shown in Fig. 7.

Due to the significant friction forces existing between the

bead and the PMMA plate, open loop experiments were

conducted in order to evaluate the kinetic friction coefficients

along both the x and z axes individually (Fig. 8 and Fig. 9) and

simultaneously (Fig. 10). The same upward gradients along y

are applied in order to reduce the effects of friction.

The rolling friction forces between steel and PMMA is

evaluated at 73.9 nN, based on equations (16), (17) and (18)

where 29.01v , 33.02v , PaE 11

1 1005.2 ,

PaE 11

2 101.3 , 5.11R mm and 2R were considered

[27].

Fig. 8. Plot of the position in x versus time of an open loop control

experiment carried out along the x-axis withxB = 15 mT/m and

yB = 30

mT/m, with two corresponding simulation plots considering s

= 0.5, and

xk= 0.1 and

xk= 0 respectively and one simulation plot considering a

73.9 nN rolling friction force.

2. 2D Quiescent Flow Results

Results of the open loop experiment along x that are shown

in Fig. 8 reveal that when correlating the simulation to the

experimental curve, the kinetic coefficient is approximated to

0.1, producing a kinetic sliding friction force of 12.8 μN.

Along z, open loop results presented in Fig. 9 disclosed a

kinetic coefficient zk= 0.4 which is close to values from the

literature [28]. Fig. 10 illustrates results carried out along the

horizontal plane. The gradients in z were set higher than the

gradients in x to ensure a diagonal motion of the bead. Once

again, upon fitting the simulated curve to the experimental

one, the closest kinetic coefficients obtained are xk= 0.1 and

zk = 0.3.

Closed loop experiments were carried out by providing

waypoints coordinates to the ICE as well as the control gains.

These were conducted along the x and z axes following the

waypoint track. Fig. 13.(A) and Fig. 13.(B) show the same

experimental data correlated with simulation data differing in

the choice of friction coefficient values. Fig. 14 illustrates the

data of an experiment where there are no intermediate

waypoints along the first segment.

In the simulation, the water density ρf and viscosity μ are

respectively set to 1 g/cm3 and 0.001002 Pa.s. Based on the

Page 7: Real-time MRI-based Control of a Ferromagnetic Core for ...wiki.polymtl.ca/nano/images/0/02/J-2008-MRSUB-TBME-Samer.pdf · another along a pre-determined path using MRI-based navigation

TBME-00524-2006-R3-preprint 7

tuning technique, the controller gains were adjusted to KTP = 2,

KTD = 0.02, KTI = 0.05, KTNP = 5 and KTND = 0.1. The anti-

windup limits were set to half the maximum gradient’s peak

magnitude.

Fig. 9. Plot of the position in z versus time of an open loop control experiment

carried out along the z-axis withzB = -26 mT/m and

yB = 30 mT/m, with

two corresponding simulation plots considering as

= 0.5, and axk= 0.4

and azk= 0.46 respectively.

Fig. 10. Plot of the position in z versus the position in x of an open loop

control experiment carried out along the horizontal plane withxB = -15

mT/m, yB = 30 mT/m and

zB = -26 mT/m, with three corresponding

simulation plots considering s

= 0.5, xk

= 0.1 and zk

= 0.3, zk

= 0.4,

zk= 0.46 respectively, t = 2.64.

The choice of coefficients obtained in the 2D open loop

experiments is confirmed from the results displayed in Fig. 10.

In Fig. 13.(A), the simulated curve is delayed with respect to

the experimental curve, whereas in Fig. 13.(B), the simulated

and experimental plots reach the same level along the

waypoint track.

Fig. 11. 1D pulsatile flow setup.

B. 1D Pulsatile Flow Control Experiment

1. 1D Pulsatile Flow Setup

A ferromagnetic sphere with identical characteristics is

placed in a rigid cylindrical tube of inner diameter of 9.82 mm.

The tube is inserted in a PMMA base that ensures its

horizontality as shown on Fig. 11. Flexible tubes are used to

connect the cardiovascular pump outlet and inlet to the rigid

tube’s ends. Handmade PMMA filters enclosing nylon meshes

are inserted between the rigid tube and the flexible tube to

avoid leakage of the fluid pump and to constrain the bead

inside the rigid tube. For the pump to function adequately, the

fluid consists of a mix of 40% of glycerol and 60% of water

which emulates the blood density. The alignment of the rigid

tube with the x-axis is ensured with the MRI laser beam. A

femoral flow, whose typical rate waveform (see Fig. 12) in

human body is provided by a dedicated programmable pump

[29].

At the beginning of the experiment, the pump would carry

the bead in the positive x-axis along the tube. Upon visual

inspection, once the pump reversed its pumping direction,

causing the bead to drift to the other side of the tube, the

sequence was launched, assigning to the controller a positive

target along the x-axis, and thus acting opposite to the flow.

Fig. 12. Human femoral rate flow waveform with a scale factor of 50.

2. 1D Pulsatile Flow Results

Both waveforms of Fig. 15 reveal periodic oscillations

Page 8: Real-time MRI-based Control of a Ferromagnetic Core for ...wiki.polymtl.ca/nano/images/0/02/J-2008-MRSUB-TBME-Samer.pdf · another along a pre-determined path using MRI-based navigation

TBME-00524-2006-R3-preprint 8

induced by the pulsatile flow close to the set point value. In the

simulation, the flow is considered to be directed along the

segment being followed by the bead which is why the

segment’s angle is passed on to the model in Fig. 6 and set to

zero. Experiments were carried out with average flow rate of 6106 and 6101.7 m

3/s which coincides with the stability

limits of the controller shown for both waveforms of Fig. 15.

The same controller gains were used as in Sect. II.A.1 except

for KTP that was increased to 4 to counteract the pulsatile flow.

IV. DISCUSSION

A. 2D Quiescent Flow Control Analysis

The kinetic coefficients obtained for z in Fig. 10 are lower

than those obtained in Fig. 9. On one side, it is most likely that

the relative high magnetic propulsion gradients applied to the

bead will induce vibrations to the phantom, hence diminishing

the effects of friction forces between the bead and the PMMA

plate. On the other side, it could be explained by the fact that

the bead’s motion is not steady as it is rolling and sliding at the

same time.

The oscillations clearly observed in the experimental plots

with intermediate waypoints (Fig. 13) are mainly due to the

large tdelay. Such oscillations appear as the bead approach a

waypoint. It is noted that the oscillations are smaller in

amplitude (Fig. 14) even when the bead is getting closer to

waypoint ( 6 , 6 )w wx cm z cm because it is carried by the

momentum of the high gradients. Despite the friction’s aspect,

correlation does exist between experimental results and

simulation, which validates the chosen model. In the absence

of friction, the normal control main purpose is to act when the

bead changes its course. As it can be noticed in the figures

related to the 2D closed loop control, the bead doesn’t drift

away even when turning with an angle close to 90º. The

derivative element of the normal controller contributes in

canceling the normal velocity vector along the current

segment, which is in fact the tangent velocity vector along the

previous segment, whereas the proportional element acts in

reducing the normal error. The normal PD controller is

therefore useful to avoid the bead being propelled in unwanted

junction at the bifurcation of blood vessels. Another pertinent

observation found in these figures is the significant drift of the

bead along the x direction to the detriment of the z direction as

the bead leaves the current waypoint and set its course towards

the next waypoint along the first segment. Since the first

segment has a slope of 45º, the command signals calculated by

the controller must have equal magnitudes. But the fact that

the frictions along the z-axis are much more significant than

those along the x-axis the controller is unable to compensate

rapidly with a relatively significant tdelay.

A.

B.

Fig. 13. Plot of the position in z versus the position in x of a closed loop

control experiment carried out on the horizontal plane withyB = 30 mT/m

and mT/m 40 maxmax zx BB , along a waypoint track whose xz-

coordinates are [(2,2) ; (4,4) ; (6,6) ; (5,8.8)] cm, with a corresponding

simulation plot considering s

= 0.5, and xk= 0.1,

zk= 0.4 for (A) and

xk = 0.05, zk = 0.3 for (B) respectively, t = 13.5 s.

When moving from a waypoint to another, the bead starts

first by traveling along the x direction and then along the z

direction due to the fact that the friction force is larger in

magnitude along the z direction than in the x direction.

Moreover, when departing from a waypoint, the normal

controller command is not high enough when compared to the

tangent controller command to bring back the bead along the

Page 9: Real-time MRI-based Control of a Ferromagnetic Core for ...wiki.polymtl.ca/nano/images/0/02/J-2008-MRSUB-TBME-Samer.pdf · another along a pre-determined path using MRI-based navigation

TBME-00524-2006-R3-preprint 9

track. Besides, based on simulation, if normal PD controller

gains are increased in order to reach a higher command, the

bead will oscillate even more.

The comparison of the experimental results observed along

the first segment between Fig. 13 and Fig. 14 reveals the

advantage of the intermediate waypoints as they prevent the

bead from drifting away from the trajectory.

It is noted that once the MRI system will be equipped with

coils capable of generating high amplitude gradients, the

friction forces won’t be a major issue. Nevertheless, even if the

levitation of the bead is unfeasible for the moment, the friction

that would exist between the bead and endothelial membrane

of the blood vessels would most probably be of viscous type

and hence negligible compared to sliding friction, as a thin

layer of blood would be separating the bead from the vessel’s

wall at all time.

B. 1D Pulsatile Flow Control Analysis

The compliance effect due to the flexible tubes [30] and the

handmade filters bring a strong alteration to the signal. The

programmed waveform — where the femoral waveform is

chosen — signal gets significantly attenuated in amplitude

when it reaches the tube since these have an effect comparative

to low pass filters. Nevertheless, as mentioned in Sect. II.B,

since the pump is adjusted empirically to meet the

controllability limits, the fluid drag force becomes significant

enough to evaluate the controller performance. Based on the

results obtained in Fig. 15, it might be possible, without the

need of measuring the actual flow waveform signal to which

the bead is subdued, to attenuate or remove these oscillations

by applying a feedforward control. This would be performed

first by measuring and registering an early oscillation cycle

based on the first two cycles. Then a command signal whose

waveform is proportional to one oscillation cycle subtracted to

the steady state error is then applied with a 180º phase shift,

i.e. opposite in sign. In order to meet this phase shift,

synchronization between the command signal and the pulsatile

flow is necessary and established mainly by taking into

account tdelay and the duty cycle. For a more reliable approach

destined for in vivo navigation, synchronization could be

attained by the use of a pulse oximeter [15]. It is to be noted

that the feedforward control would find utility mainly for final

waypoints rather than intermediate waypoints as there would

be a necessity to perform the intervention with a maximum

stability.

C. General Analysis

The control experiments conducted in vitro mimic in vivo

behavior. Nevertheless, major aspects need to be taken into

account in order to perform MII efficiently. First and foremost,

the nature of waypoints (intermediates and finals) need in

some cases, to be compensated for motion artifacts such as the

patient breathing and movements using registration techniques

as discussed in [22]. Second, the 2D controller needs to be

tuned in order to prevent the system from exiting the region of

attraction as determined by the formal stability analysis, due to

the variations of physiological parameters such as the ratio of

the diameter of the bead over the vessel’s diameter, blood

viscosity and density. Still, early in vivo experiments without

the improvements of these aspects proved the acceptability of

the current control algorithm [31].

Fig. 14. Plot of the position in z versus the position in x of a closed loop

control experiment carried out on the horizontal plane withyB =30 mT/m

and mT/m 40 maxmax zx BB , along a waypoint track whose xz-

coordinates are [(2,2) ; (4,6)] cm, with a corresponding simulation plot

considering s

= 0.5, and xk

= 0.05 and zk

= 0.3 respectively, t = 20.2 s.

Fig. 15. Plot of the position in x versus time of a closed loop control

experiment carried out withyB = 40 mT/m and

maxxB = 40 mT/m, along

the x-axis with a waveform whose scale factor is adjusted respectively to 50

(with an offset of 3 cm along the x-axis) for waveform (a) and to 60 for

waveform (b) for a common set point wx = 6 cm.

V. CONCLUSION

In this paper, we consider the use of an MRI system for

cardiovascular navigation. Through two complementary

experiments, the PID controller has shown that 2D real-time

control is feasible despite some limitations. As being part of

the trajectory planning, a velocity profile can be envisioned

instead of a positional profile. Due to the existence of a large

range of variation in physiological and physical parameters,

although the initial version of the controller has proven to be

Page 10: Real-time MRI-based Control of a Ferromagnetic Core for ...wiki.polymtl.ca/nano/images/0/02/J-2008-MRSUB-TBME-Samer.pdf · another along a pre-determined path using MRI-based navigation

TBME-00524-2006-R3-preprint 10

effective during experiments conducted in the carotid artery of

a living swine [31], an adaptive control algorithm would be

susceptible to improve the robustness to those variations.

Subsequent to such improvements, the system may eventually

be applied to other areas of the biomedical field such as

precise guide wire or needle [32] steering into deformable

tissues which would ensure safer insertions and precise

placements. The technique may also prove to be applicable

within blood vessels as small as capillaries for applications

such as the direct delivery of therapeutic agents to tumor cells.

In such cases, because factors such as inertia, buoyancy and

gravity are not dominant for devices navigating in the

microvasculature, a 3D controller can be envisioned. But

because at such a scale other forces dominate and since the

blood is no more homogeneous, the controller would need to

be adapted accordingly.

ACKNOWLEDGMENT

The authors would like to thank particularly O. Felfoul, and

E. Aboussouan for their active involvement in the experiments

as well as their suggestions, Dr. D. Ménard for providing the

facilities in order to measure the magnetisation of the

ferromagnetic beads, L.-P. Carignan for his help in operating

the VSM, and R. Ouellet from the Montreal Heart Institute for

providing the cardiovascular pump.

REFERENCES

[1] L. Given Imaging, "Given Imaging Home Page," 2006/06/30

http://www.givenimaging.com

[2] S. Guo, J. Sawamoto, and Q. Pan, "A novel type of microrobot for

biomedical application," presented at 2005 IEEE/RSJ International

Conference on Intelligent Robots and Systems, 2005.

[3] K. Ishiyama, M. Sendoh, and K. I. Arai, "Magnetic micromachines for

medical applications," Journal of Magnetism and Magnetic Materials,

vol. 242-245, pp. 41-46, 2002.

[4] M. B. Khamesee, N. Kato, Y. Nomura, and T. Nakamura, "Design and

control of a microrobotic system using magnetic levitation,"

IEEE/ASME Transactions on Mechatronics, vol. 7, pp. 1-14, 2002.

[5] M. B. Khamesee, N. Kato, Y. Nomura, and T. Nakamura,"Performance

improvement of a magnetically levitated microrobot using an adaptive

control," presented at International Conference on MEMS, NANO and

Smart Systems, 2003.

[6] R. G. McNeil, R. C. Ritter, B. Wang, M. A. Lawson, G. T. Gillies, K. G.

Wika, E. G. Quate, M. A. Howard, III, and M. S. Grady, "Functional

design features and initial performance characteristics of a magnetic-

implant guidance system for stereotactic neurosurgery," IEEE

Transactions on Biomedical Engineering, vol. 42, pp. 793-801, 1995.

[7] R. C. Ritter, M. S. Grady, M. A. Howard, III, and G. T. Gillies,

"Magnetic stereotaxis: Computer-assisted, image-guided remote

movement of implants in the brain," Innovation et Technologie en

Biologie et Medecine, vol. 13, pp. 437-449, 1992.

[8] K. B. Yesin, P. Exner, K. Vollmers, and B. J. Nelson, "Design and

control of in-vivo magnetic microrobots," presented at 8th International

Conference on Medical Image Computing and Computer-Assisted

Intervention - MICCAI 2005.

[9] K. B. Yesin, K. Vollmers, and B. J. Nelson, "Modeling and control of

untethered biomicrorobots in a fluidic environment using

electromagnetic fields," International Journal of Robotics Research,

vol. 25, pp. 527-536, 2006.

[10] G. Kosa, M. Shoham, and M. Zaaroor, "Propulsion of a Swimming

Micro Medical Robot," presented at the 20th International Conference

on Robotics and Automation, pp. 1327-1331, Barcelona, April 18-22,

2005.

[11] G. Kosa, M. Shoham, and M. Zaaroor, "Analysis of a Swimming Micro

Robot," presented at The First IEEE/RAS-EMBS International

Conference on Biomedical Robotics and Biomechatronics, pp. 130-134,

Pisa, February 20-22, 2006.

[12] J. B. Mathieu, G. Beaudoin, and S. Martel, "Method of propulsion of a

ferromagnetic core in the cardiovascular system through magnetic

gradients generated by an MRI system," IEEE Transactions on

Biomedical Engineering, vol. 53, pp. 292-299, 2006.

[13] J. B. Mathieu and S. Martel, "Magnetic steering of iron oxide

microparticles using propulsion gradient coils in MRI," presented at

28th IEEE-EMBS Annual International Conference of the Engineering

in Medicine and Biology Society, pp. 472-475, New York, Aug 30-Sept.

3, 2006.

[14] W. Sabra, M. Khouzam, A. Chanu, and S. Martel, "Use of 3D Potential

Field and an Enhanced Breadth-first Search Algorithms for the Path

Planning of Microdevices Propelled in the Cardiovascular System,"

presented at 27th Annual International Conference of the Engineering in

Medicine and Biology Society, Shangai, China, Sept 1-4, 2005.

[15] S. Tamaz and S. Martel, "Impact of the MRI-based Navigation System

Constraints on the Step Response Using a PID Controller," presented at

27th IEEE-EMBS Annual International Conference of the Engineering

in Medicine and Biology Society, pp. 5073-5076, Shangai, China, Sept

1-4, 2005.

[16] S. Tamaz and S. Martel, "Bidimensional MRI-based navigation system

using a PID controller," presented at 28th IEEE-EMBS Annual

International Conference of the Engineering in Medicine and Biology

Society, pp. 1746-1749, New York, Aug 30 – Sept. 3, 2006.

[17] A. Chanu, S. Martel, and G. Beaudoin, "Real-time Magnetic Resonance

Gradient-based Propulsion of a Wireless Microdevice Using Pre-

Acquired Roadmap and Dedicated Software Architecture," presented at

27th IEEE-EMBS Annual International Conference of the Engineering

in Medicine and Biology Society, pp. 5190-5193, Shangai, China, Sept

1-4, 2005.

[18] A. Chanu, E. Aboussouan, S. Tamaz, and S. Martel, "Sequence design

and software environment for real-time navigation of a wireless

ferromagnetic device using MRI system and single echo 3D tracking,"

presented at 28th IEEE-EMBS Annual International Conference of the

Engineering in Medicine and Biology Society, pp. 1746-1749, New

York, Aug 30-Sept. 3, 2006.

[19] A. G. Siemens, "Magnetom Avanto," 2006/06/30

http://www.medical.siemens.com/webapp/wcs/stores/servlet/ProductDis

play?productId=75336&storeId=10001&langId=-1&catalogId=-

1&catTree=100001,12786,12754&level=0&view=clinical 2005.

[20] J.-B. Mathieu, S. Martel, L. H. Yahia, G. Soulez, and G. Beaudoin,

"Preliminary investigation of the feasibility of magnetic propulsion for

future microdevices in blood vessels," Bio-Medical Materials and

Engineering, vol. 15, pp. 367-374, 2005.

[21] O. Felfoul, J. B. Mathieu, and S. Martel, "Magnetic field mapping by

selective equipotential excitation," presented at 28th Annual

International Conference of the Engineering in Medicine and Biology

Society, pp. 3775-3778, New York, Aug 30-Sept. 3, 2006.

[22] E. Aboussouan and S. Martel, "High precision absolute positioning of

medical instruments in MRI systems," presented at 28th Annual

International Conference of the Engineering in Medicine and Biology

Society, 743-746, New York, Aug 30-Sept. 3, 2006.

[23] C. H. Cunningham, T. Arai, P. C. Yang, M. V. McConnell, J. M. Pauly,

and S. M. Conolly, "Positive contrast magnetic resonance imaging of

cells labeled with magnetic nanoparticles," Magnetic Resonance in

Medicine, vol. 53, pp. 999-1005, 2005.

[24] F. Trochu and H. Yelle, Mechanical power transmission Canada

(Transmission de puissance mécanique : cours MEC3330): École

Polytechnique of Montréal, 2000.

[25] C. Salem Specialty Ball, "Chrome Steel ball," 2006/06/30

http://www.salemball.com/chrome.htm

[26] Automation Creations Inc, "Matweb Material Property Data,"

2006/06/30 http://www.matweb.com/

[27] Plastics Design Library Staff, "Fatigue and Tribological Properties of

Plastics and Elastomers," William Andrew Publishing/Plastics Design

Library, 1995.

Page 11: Real-time MRI-based Control of a Ferromagnetic Core for ...wiki.polymtl.ca/nano/images/0/02/J-2008-MRSUB-TBME-Samer.pdf · another along a pre-determined path using MRI-based navigation

TBME-00524-2006-R3-preprint 11

[28] R. P. Steijn, "Friction and Wear of Plastics," Metals Engineering

Quarterly, vol. 7, pp. 9-21, 1967.

[29] Shelley Medical Imaging Technologies, "CardioFlow 1000 MR

Computer-controlled physiological flow system," 2006/06/30

http://www.simutec.com/Media/pumps/Brochure%20-

%20Shelley%20CardioFlow%201000%20MR.pdf.

[30] W. R. Milnor, Hemodynamics. Baltimore/London: Williams and

Wilkins, 1982.

[31] S. Martel, J. B. Mathieu, O. Felfoul, E. Aboussouan, S. Tamaz, P.

Pouponneau, L. H. Yahia, G. Beaudoin, G. Soulez, and M.

Mankiewiecz, "Automatic navigation of an untethered device in the

artery of a living animal using a conventional clinical magnetic

resonance imaging system," Applied Physics Letters, vol. 90, 2007.

[32] S. P. DiMaio and S. E. Salcudean, "Needle steering and motion

planning in soft tissues," IEEE Transactions on Biomedical

Engineering, vol. 52, pp. 965-974, 2005.

[33] H. K. Khalil, Nonlinear systems, 3rd ed. Upper Saddle River, NJ:

Prentice Hall, 2002.

Samer Tamaz received the B.E. degree in

electrical engineering in 2002 from the

Lebanese American University (LAU) and the

M.S. degree in biomedical engineering from

École Polytechnique de Montréal (EPM),

Montreal, QC, Canada, in 2006. He is

currently working as a biomedical engineer at

the healthcare center and community service

of Laval, Qc, Canada.

Richard Gourdeau (S'88-M'91) received is

B.Sc.A and M.Sc. in mechanical engineering

from Laval University (1986, 1988) and his

Ph.D. from Carleton University (1991). He

joined the Department of Mathematics and

Industrial Engineering at the École

Polytechnique de Montréal in 1991 as

assistant professor and then moved to the

Department of Electrical Engineering in 1999

where he is professor since 2005. He is

teaching robotics and automatic control. His

research interests include robotic

manipulators control, control of mechanical

systems and digital control.

Arnaud Chanu received the B.Sc. degree in

computer engineering in 2003 and the

M.S. degree in biomedical engineering from

École Polytechnique de Montréal (EPM),

Montréal, QC, Canada, in 2006. He is

currently working as an embedded system

engineer at iMetrik. His work interests are the

development of embedded system software

architectures and computer hardware design.

Jean-Baptiste Mathieu received the

B.Sc. degree in mechanical engineering in

2002 and the M.S. degree in biomedical

engineering from École Polytechnique de

Montréal (EPM), Montréal, QC, Canada,

in 2004. He is currently with the

NanoRobotics Laboratory in the

department of Computer Engineering at

EPM. His current research interests are in

the development of medical applications of

microdevices propelled in the blood vessels

by magnetic gradients generated by

magnetic resonance imaging (MRI)

systems.

Sylvain Martel (S’95-M’96- SM’07) was

born in Québec City, Canada in 1959. He

received the B.E. degree from the

Université du Québec (UQTR), and the

M.Eng. and Ph.D. degrees from McGill

University, Montréal, Canada in 1989 and

1997 respectively, all in electrical

engineering. Following postdoctoral

studies at MIT, he was Research Scientist

at the BioInstrumentation Laboratory,

Department of Mechanical Engineering at

MIT. He is currently Associate Professor in

the Department of Computer Engineering

and the Institute of Biomedical

Engineering at École Polytechnique de Montréal (EPM), Montréal, Canada,

and Director of the NanoRobotics Laboratory at EPM. Dr. Martel holds the

Canada Research Chair (CRC) in Conception, Fabrication and Validation of

Micro/Nanosystems. He has over 130 refereed publications and participates

in many international committees and organizations Presently, Dr. Martel

leads a multidisciplinary team involved in research and development of new

platforms and nano-factories based on a fleet of scientific instruments

configured as autonomous miniature robots capable of high throughput

autonomous operations at the molecular scale, minimally invasive tools based

on microdevices propelled in the blood vessels by magnetic gradients

generated by Magnetic Resonance Imaging (MRI) systems, Micro-Electro-

Mechanical System (MEMS) and System-on-Chip (SoC)-based miniature

instrumented robots, development if new MEMS/NEMS based on the

integration of bacteria as biological components, and many other related

projects. Dr. Martel’s main expertise is in the field of nanorobotics, micro-

and nano-systems, and the development of novel instrumented platforms

including advanced micromechatronics systems and a variety of related

support technologies. He has also a vast experience in electronics, computer

engineering, and also worked extensively in biomedical and mechanical

engineering.