OraQuick ADVANCE Rapid HIV Testing : AT WELLONE Oraquick Rapid HIV Testing can be processed using a blood specimen or oral specimen- At WellOne, we use only blood for specimen processing The test is not approved for persons under age 13 Testing should be offered to all persons age 13 to 64 routinely Click here for more information on rapid HIV testing Click here to go back Click here to move on
Citation preview
RAPID HIV TESTINGOraQuick
ADVANCE HIV 1/2
WellOne Primary
Medical and Dental Care for Medical Clinical Staff
Click here to
move on
OraQuick ADVANCE Rapid HIV Testing : AT WELLONE
Oraquick Rapid HIV Testing can be processed using a blood specimen or oral specimen- At WellOne, we use only blood for specimen processing
The test is not approved for persons under age 13Testing should be offered to all persons age 13 to 64 routinely
Click here for more
information on rapid HIV
testingClick here to go back
Click here to move on
TRUE or FALSE: OraQuick Advance Rapid HIV testing is not approved for
persons under age 16 True
False
Click here to select this
answerClick here to select this
answer
Click here to go back
The correct answer is FALSE
The test is not approved or persons
under age 13Click here
to move onClick here to go back
The OraQuick® ADVANCE Rapid HIV-1/2 Antibody Test kit consists of:•A divided pouch containing a test device and a developer solution vial•Reusable test stands•Collection loops•Subject information pamphlet•Package insert.
Click here to move on
Click here to go back
Oraquick Rapid HIV Testing : Storage & StabilityTest Kits
At WellOne, kits are stored in the laboratory unrefrigerated
The laboratory room temperature must be monitored daily. If the temperature is out of acceptable storage range (35-80º F), notify your supervisor immediately
ControlsControls are stored refrigerated at 2-80 C (35-460 F) Do not use controls past the expiration date printed on the outer carton. Open control vials only when you are performing tests.Recap and store the vials in their original container at 2-80 C after use. Date the controls with the date opened and “do not use after” date.Opened vials expire 8 weeks after opening. The temperature of the biological fridge is checked daily. Action is to be taken if results are out of range. Do not use controls if the reagent appears visually cloudy or discolored. Practice standard precautions when handling controls.
Click here to move on
Click here to go back
TRUE or FALSE: The OraQuick control solution is good for 8 weeks after
receiving the solution True
False
Click here to select this
answerClick here to select this
answer
Click here to go back
The correct answer is FALSE
The control solution is stable until the expiration date when
stored in the refrigerator at appropriate temperatures. After opening, the control
solution expires in 8 weeks from the date first opened.
Click here to move on
Click here to go back
The laboratory room temperature and refrigerator temperature must be checked and documented by the MA:
Weekly
Monthly
Daily
Click here to select this
answer
Click here to select this
answer
Click here to select this
answer
Click here to go back
The correct answer is….
DAILY
Click here to move on
Click here to go back
HIV TESTING: EXTERNAL CONTROLSExternal Controls (three total) including and HIV 1 positive, HIV 2positive, and HIV 1/2 negative control will be run and recorded onQA form: • With each new lot of test packs received prior to placing them
in service• If there is a change in the conditions of testing (e.g. new
location, lighting, temperature, etc.)• If the temperature of the test storage area falls outside of 2-
270C (35-800F)• If the temperature of the testing area falls outside of 15-270C
(59-800F)• Monthly (May change depending on testing volume)• Whenever two consecutive invalid test results are obtained on
a person being tested.
Click here to see the control
test package insert
Click here to move on
Click here to go back
The OraQuick ADVANCE® controls should be run in each of the following situations except:
When a new lot ofOraQuick ADVANCE® kits arereceived.
If there is a change inconditions of the storagearea.
By each operator at thestart of a shift.
Click here to select this
answer
Click here to select this
answer
Click here to select this
answerClick here to go back
The correct answer is……….
By each operator at the start of a shift.
External control testing must be performed when a newlot of OraQuick ADVANCE® kits are received and if there isa change in conditions of the storage area.
Click here to move on
Click here to go back
WARNING
Conducting this test exposes the operator to potentially infectious material. Standard precautions, including glove use are required for this procedure.
HIV controls are potentially infectious and should be handled in the same manner as a patient specimen, using all appropriate precautions.
Click here to move on
Click here to go back
Testing Procedure
– Identify patient using two patient identifiers (name and DOB)– Explain procedure to patient in culturally appropriate, age appropriate
terms. – Be sure the patient has received an OraQuick Advance Subject Testing
Information Pamphlet.– Gather all necessary equipment and supplies prior to start of the
procedure.– Inspect all equipment and supplies for damage, general integrity and
expiration date- do not use any damaged or expired materials– Perform hand hygiene before and after patient contact; before and
after donning gloves.
Click here to move on
Click here to go back
Open the divided pouch. Using the notched corners, tear the top of each end of the divided pouch containing the test device and developer solution vial. Remove the developer solution bottle. To prevent contamination, leave the Test Device in the divided pouch until needed.
Flat Pad: Do NOT Touch
Click here to move on
Click here to go back
Don gloves. Remove the Developer Solution Vial from the Divided Pouch.Firmly holding the Developer Solution Vial, carefully uncap the vial by gently rocking the cap back and forth. Slide the uncapped Developer Solution Vial into the top of the slot in the angled Reusable Test Stand.
Place the Reusable Test Stand on a flat, level surface. Use only the stand provided.
Be sure the vial is completely seated in the stand. DO NOT Force the
vial into the stand from the front of the slot, as
splashing may occur.
Click here to move on
Click here to go back
Testing ProcedureFollow finger stick procedureUse the Specimen Collection Loop and collect the blood on the rounded end of the loop, making sure the loop is completely filled with blood. Immediately immerse the blood-filled Specimen Collection Loop in the developer solution inside the Developer Solution Vial. Use the Specimen Collection Loop to stir the specimen in the developer solution. Remove the Specimen Collection Loop from the Developer Solution Vial and discard the used loop in a biohazard waste container. Examine the solution in the Developer Solution Vial to ensure that it appears pink, indicating that the blood specimen was properly introduced. If the developer solution is not pink after adding the specimen, discard the Developer Solution Vial as infectious waste, open a new Divided Pouch, and collect a new specimen.
Click here to move on
Click here to go back
Testing ProcedureRemove the Test Device from the Divided Pouch without touching the flat pad. Insert the Test Device, flat pad first, into the Developer Solution Vial containing the specimen. Be sure that the result window faces forward and the flat pad touches the bottom of the Developer Solution Vial.
Label the specimen with the patient’s initials and DOBDO NOT cover the two holes in the back of the Test Device after placing it into the developer solution. Doing so may cause an invalid result
Click here to move on
Click here to go back
Begin Timing and Documenting SET THE TIMER
FOR 20 MINUTES
Document on the HIV testing log:
Date of testing Patient name
DOBTime Test was StartedTemperature of room where testing is being
doneClick here
to move onClick here to go back
Read and Document Results
RESULTS MUST BE INTERPRETED
NO EARLIER THAN 20
MINUTES and NO LATER THAN 40 MINUTES after introducing the
specimen into the developer solution.
CORRECT TIMING OF
INTERPRETATION IS CRITICAL
Click here to move on
Click here to go back
10; 2020; 6020; 40
OraQuick ADVANCE® results must be interpreted at no less than _______minutes and no more than _______minutesafter introducing the specimen into the developer solution.
Click here to select this
answerClick here to select this
answerClick here to select this
answer
Click here to go back
10; 2020; 6020; 40
The Correct Answer Is……………..
Click here to select this
answerClick here to select this
answerClick here to select this
answer
RESULTS MUST BE INTERPRETED NO EARLIER THAN 20 MINUTES and NO LATER THAN 40 MINUTES after introducing the specimen into the developer solution.
Click here to move on
Click here to go back
Read and Document Results
After recording results, dispose of the test vial and test device in the biohazard waste container.
Click here to move on
Click here to go back
Rapid HIV Testing: Internal Control ResultAn internal control is built into the test unit.
A line appears at the “C” (control) , indicating a valid test
This is read as “control is positive”
Click here to move on
Click here to go back
Rapid HIV Testing: Internal Control ResultAn internal control is built into the test unit.
If the control line does not appear, the test is not valid and must be repeatedThe lack of a control line may mean that an insufficient amount of blood has been added to the test device or the test device is not working
Click here to move on
Click here to go back
Rapid HIV Testing: Negative (Nonreactive) Test Result This is a negative test with
no line appearing at the “T” on the testing unit and a positive control line appearing at the “C”
Click here to move on
Click here to go back
Rapid HIV Testing: Negative Test Result
A Non-Reactive result does not preclude the possibility of exposure to HIV or infection with HIV.An antibody response to recent exposure may take several months to reach detectable levels. Rule out a history of exposure to HIV within 3 months. Recommend a retest for clients with a recent exposure.
Click here to move on
Click here to go back
Rapid HIV Testing: Positive (Reactive) Test ResultVV
The intensity of the test line does not correlate with disease activity/severity.All OraQuick reactive (preliminary positive) results must be followed up with either a Western Blot (WB) for confirmation.Whenever the OraQuick® test result is reactive, the Health Care Provider must explain the results to the patient and provide the patient with a confirmatory testing lab slip to be performed at the lab. A follow up appointment must be scheduled to review the results with the patient.The health care provider must indicate on the specimen requisition form that the specimen is from an individual who had a reactive OraQuick rapid test result. Click here
to move onClick here to go back
TRUE or FALSE - The intensity of the color of the line next to the “T” zone that appears during a test signifies
the amount of HIV antibodies in the specimen. True
False
Click here to select this
answerClick here to select this
answer
Click here to go back
The correct answer is FALSE
The intensity of the test line does not correlate with disease activity/severity.
Click here to move on
Click here to go back
TRUE or FALSE: A Western Blot must be ordered for preliminary positive results
True
False
Click here to select this
answerClick here to select this
answer
Click here to go back
The correct answer is TRUE• A Western Blot (not an ELISA) must be
ordered as confirmatory testing.• Be sure to indicate on the specimen
requisition form that the specimen is from an individual who had a reactive OraQuick rapid test result.
Click here to move on
Click here to go back
Rapid HIV Testing: Invalid Test ResultsNo line is present in the area adjacent to either the “C” or “T” triangle (figure 1)A line appears opposite the “T” triangle but not the “C” triangle (figure 2)A red background in the result window makes it difficult to read the results after 20 minutes (figure 3)A line appears, but not opposite the “C” (or “T”) triangle – misalignment
(figure 4)
Click here to move on
Click here to go back
Which of the following is an example of an invalid test result using an OraQuick ADVANCE® test kit?
No line is present in the areanext to either the “C” or “T”triangle.
A line appears opposite the “T” triangle butnot the “C” triangle.
The pinkish red background in the resultwindow does not disappear and makes itdifficult to read the result after 20 minutes.
All of the above.
Click here to select this
answerClick here to select this
answerClick
here to select this
answerClick here to select this
answer
Click here to go back
The Correct Answer is……
All of the above
Click here to move on
Click here to go back
You have reached the end of this program…..