Radiation Safety While Operating Dental IOPA X-ray Unit ATOMIC ENERGY REGULATORY BOARD Issued by: (Government of India) Niyamak Bhavan, Anushaktinagar Mumbai-400094. www.aerb.gov.in This poster must be displayed at prominent location for due adherence by all radiations workers. Are you using Lead Cone in your dental – IOPA X-ray unit? Why Lead Cone is to be used in dental-IOPA X-ray unit & how to identify it? Why Plastic Cone should not be used in dental-IOPA X-ray unit & how to identify it? Lead cone is heavier than the Plastic Cone & it provides adequate radiation shielding to the Dentist, Operator & Patient. Plastic cone is light weighted as compared to lead cone & it does not provide adequate radiation shielding. Use of plastic cone in dental-IOPA unit may lead to unnecessary radiation exposure to Dentist, Operator & Patient. Minimum Lead Cone Length 60 to 75 kV: 20 cm More than 75 kV: 30 cm Check for AERB Type Approval certificate with the Supplier for the Dental X-ray Unit and ask to provide dental Lead Cone to your IOPA X-ray Unit Plastic Cone Lead Cone

Radiation Safety While Operating Dental IOPA X-ray …aerb.mindspacetech.com/images/PDF/Dental Poster.pdfRadiation Safety While Operating Dental IOPA X-ray Unit Issued by: ATOMIC ENERGY

  • Upload
    others

  • View
    5

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Radiation Safety While Operating Dental IOPA X-ray …aerb.mindspacetech.com/images/PDF/Dental Poster.pdfRadiation Safety While Operating Dental IOPA X-ray Unit Issued by: ATOMIC ENERGY

Radiation Safety While Operating Dental IOPA X-ray Unit

ATOMIC ENERGY REGULATORY BOARDIssued by:

(Government of India)Niyamak Bhavan, AnushaktinagarMumbai-400094.www.aerb.gov.in

This poster must be displayed at prominent location for due adherence

by all radiations workers.

Are you using Lead Cone in your dental – IOPA X-ray unit?

Why Lead Cone is to be used in dental-IOPA X-ray unit & how to identify it?

Why Plastic Cone should not be used in dental-IOPA X-ray unit & how to identify it?

Lead cone is heavier than the Plastic Cone & it provides adequate radiation shielding to the Dentist, Operator & Patient.

Plastic cone is light weighted as compared to lead cone & it does not provide adequate radiation shielding. Use of plastic cone in dental-IOPA unit may lead to unnecessary radiation exposure to Dentist, Operator & Patient.

Minimum Lead Cone Length60 to 75 kV: 20 cmMore than 75 kV: 30 cm

Check for AERB Type Approval certificate with the Supplier for the Dental X-ray Unit and ask to provide dental Lead Cone to your IOPA X-ray Unit

Plastic Cone

Lead Cone