36
Rad T 290 Generators

Rad T 290 Generators. Generator Components control console kVp adjust mA adjust time adjust transformer high voltage (step up) filament »low

Embed Size (px)

Citation preview

Page 1: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

Rad T 290

Generators

Page 2: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

Generator ComponentsGenerator Components

• control console kVp adjust mA adjust time adjust

• transformer high voltage (step up) filament

» low voltage (step down)

• electronics cabinet support circuitry

or mAs adjust

Page 3: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

Incoming PowerIncoming Power

• Line affects generator performance– diameter of wire– length or wire– other devices sharing branch circuit

• Resistance of power line wires can reduce generator voltage during exposure affecting power available to x-ray tube calibration

Page 4: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

Line Voltage CompensationLine Voltage

Compensation

• Incoming voltage can vary during day

• Generators need to correct for changes in line voltage power line fluctuations affect calibration

Incoming Power Line

Generator

CircuitBreaker

Page 5: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

Line

Line Voltage CompensationLine Voltage Compensation

• Compensation may be Automatic

» Almost everything today

Manual» user must make adjustment

LineCompensation

Page 6: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

LineAuto-trans-former

High Voltage

Transformer

Rectifier Circuit

TimerCircuit

+

•High voltage Transformer has fixed ratio•Autotransformer has variable ratio•Autotransformer needed to provide variable kilovoltage to tube

Autotransformer

Filament Transformer

mA regulator

Page 7: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

AutotransformerAutotransformer

Line

LineCompensation

TimerCircuit

to high voltage transformer

primary

to filament transformer

primarymA

regulator

major kV selector

minor kV selector

Autotransformer does line compensation & kVp selection

Page 8: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

Generator VoltagesGenerator Voltages

• Input line voltage single or three phase 115 - 480 Volts AC

• Autotransformer provides variable voltage to primary of high

voltage transformer

1

PowerLine

AutoTransformer

High VoltageTransformer

TimerCircuit

Page 9: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

High Voltage CircuitHigh Voltage Circuit

• Supplies high voltage for x-ray tube

• Step-up transformer primary from autotransformer secondary to rectifier circuit mA monitored at center grounded point of secondary

Auto-transformer

RectifierCircuitmA

High Voltage Transformer

Page 10: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

High Voltage TransformerHigh Voltage Transformer

• Grounded metal box

• filled with oil electrical insulator

• Function increases or decreases alternating voltage

• Also contains rectifier circuit

changes alternating current into direct current

Page 11: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

Fullwave RectifierFullwave Rectifier• Four diodes• 120 pulses/second• exposure times half of halfwave circuit

Secondary of High Voltage Transformer

Voltage applied to tube(also mA waveform)

Page 12: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

Fullwave RectifierFullwave Rectifier

+

-X

X

First Half Cycle Second Half Cycle

Voltage applied to tube(also mA waveform)

X

X

+

-

Page 13: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

Full-Wave RectificationFull-Wave Rectification

• Rectifiers Four diode “bridge” configuration used

with single phase

• both + & - half cycle of high tension transformer used efficient circuit reverses negative half cycle &

applies to x-ray tube

Applied to X-ray TubeOutput of High Tension Transformer

Tube

Page 14: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

Pulsed RadiationPulsed Radiation• single phase input power results in

pulsed radiation

• Disadvantages intensity only significant when voltage is near peak low voltage heats target and produces low-energy

photons» absorbed in tube, filter, or patient

• can contribute to dose

Applied to X-ray Tube Radiation Waveform

Page 15: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

Three-Phase GeneratorsThree-Phase Generators

• Commercial power generally delivered as 3 phase

• phases 120o apart

Single Phase Power Three Phase Power

Page 16: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

Three-Phase GeneratorsThree-Phase Generators• Rectifier circuit

Inverts negative voltage sends highest of 3 phases to x-ray tube

To X-Ray Tube

Input 3 Phase VoltageRectified

Page 17: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

Three-Phase GeneratorsThree-Phase Generators• much higher tube ratings than

single phase• more efficient than single phase

shorter exposures lower exposure

Three Phase OutputSingle Phase Power

Page 18: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

3 Generator Circuits3 Generator Circuits

• pulses number of peaks per 1/60 second (16.6 msec)

power line cycle

• windings 3 primary coils (one for each phase) 3 or 6 secondary

» with 6 secondaries, 2 secondary coils induced per primary

Three Phase Output

Page 19: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

Three Phase TransformingThree Phase Transforming

• 3 coils can be hooked up in 2 ways

Delta Wye

Page 20: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

3-phase generator3-phase generator

• Primary windings generally delta

• Secondary windings may be delta or wye

Primary

Secondary

Page 21: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

3 Phase Generator3 Phase Generator

• 6-Pulse Twelve Rectifier 1 delta primary 2 wye secondaries

» 6 secondary windings

• two diodes per winding

13.5% ripple

Three Phase Output

Ripple

Primary

SecondarySecondary

Page 22: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

3 Phase Generator3 Phase Generator• 12-Pulse Twelve Rectifier

1 delta primary 2 secondaries, 1 wye, 1 secondary

» 30o phase difference between secondaries

» 6 secondary windings

• 2 diodes per winding

3.5% ripple

Three Phase Output

Ripple

Primary

SecondarySecondary

Page 23: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

LineAuto-trans-former

High Voltage

Transformer

Rectifier Circuit

TimerCircuit

+

•Circuitry for mA selection•Adjusts mA on the fly during exposure.

mA regulator

Filament Transformer

mA regulator

Page 24: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

LineAuto-trans-former

High Voltage

Transformer

Rectifier Circuit

TimerCircuit

+

Steps down AC voltage from Autotransformer & mA selector to smaller AC voltage required by filament (8-12 volts typical)

Filament Transformer

Filament Transformer

mA selector

Page 25: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

mA selectionmA selection

Line

LineCompensation

to filament transformer

primarymAstabilizer

•Allows selection from available discrete mA stations.•Applies correct voltage to primary of filament transformer.

10 mA

25 mA

50 mA100 mA

200 mA

300 mA

400 mA

Page 26: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

mA Stabilization During Exposure

mA Stabilization During Exposure

• On first trigger mA regulator supplies anticipated voltage to filament

transformer primary

• mA monitored during exposure

• Corrections made to filament voltage during exposure as necessary if mA low, filament voltage boosted if mA high, filament voltage lowered

Page 27: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

1 vs. 3 Generators1 vs. 3 Generators

• Typical home & small business power

• inexpensive• transformer

windings 1 primary coil 1 secondary coil

Industrial power expensive transformer windings

• 3 primary coils• one for each phase

• 6 secondary coils» 2 secondary coils

induced per primary)

1 3

Page 28: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

1 vs. 3 Generators1 vs. 3 Generators

• 100% ripple• 8 ms minimum exp.

Time 1/120th second

• lower output intensity

• puts less heat in tube for same technique

4-13% ripple• higher average kVp

• slightly less patient exposure

<=1 ms minimum exp. time

higher output intensity puts more heat in tube

1 3

Page 29: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

Exposure Time ControlExposure Time Control

• electronic, measuring» time (crystal)

» power line pulses

• automatic (phototimingphototiming) terminates exposure based on radiation received by

receptor

Page 30: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

Phototiming GeometryPhototiming Geometry• entrance type

detector in front of film detector must be essentially

invisible

• exit type detector behind film obsolete except for

mammography» detector visible because of high

contrast image

Exit typeSensor

Grid

FilmEntrance typeSensor

Page 31: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

Phototiming Radiation DetectorsPhototiming Radiation Detectors

• ionization chambers

• solid-state detectors

Page 32: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

Ionization ChambersIonization Chambers

• Almost always entrance type

• Notes thin parallel aluminum plates

are electrodes» voltage applied between plates

» collect ions produced by radiation in air between electrodes

collected ions produce electric current

+-

+Photon

-

Page 33: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

Solid State DetectorsSolid State Detectors• PN semiconductor junction

generates current when struck by radiation

• small

• fast response

• little beam attenuation

Photon Electric Current

Page 34: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

Phototiming FieldsPhototiming Fields• 1, 2, or 3

• fields may be selected individually or in combination

• proper positioning critical

Page 35: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

Phototiming NotesPhototiming Notes

• must be calibrated for particular film-screen system

• some generators allow selection from several preset film/screen combinations

Page 36: Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low

Phototiming NotesPhototiming Notes

• phototimer must correct for rate response kVp response of

» film/screen system» phototiming sensor

Higher kVp beam more penetrating» Less attenuated by phototimer detector

• safety exposure limited to 600 mAs if phototimer

does not terminate exposure (2000 mAs for < 50 kV)