Quick Start GuideLIQ-QSG-499ACL-02, Rev K

July 2017

Rosemount™ 499ACL-02

Total Chlorine Sensor

Page 2: Quick Start Guide July 2017 Rosemount 499ACL-02 · PDF fileChapter 1 Plan ... Chapter 6 Accessories ... Rosemount 499ACL-02-54 Sensor Wiring to Rosemount 1056 and 56 Transmitters Figure

Safety information

CAUTION!SENSOR/PROCESS APPLICATION COMPATIBILITY

The wetted sensor material may not be compatible with process composition and operating conditions. Application compatibility isentirely your responsibility.

CAUTION!EQUIPMENT DAMAGE

Do not exceed pressure and temperature specifications.

Pressure: 65 psig (549 kPa abs) max. Temperature: 32 to 122 °F (0 to 50 °C)

Page 3: Quick Start Guide July 2017 Rosemount 499ACL-02 · PDF fileChapter 1 Plan ... Chapter 6 Accessories ... Rosemount 499ACL-02-54 Sensor Wiring to Rosemount 1056 and 56 Transmitters Figure

Contents

Chapter 1 Plan ..................................................................................................................................11.1 Unpacking and inspection .............................................................................................................. 11.2 Product description ........................................................................................................................11.3 Specifications .................................................................................................................................1

Chapter 2 Install ...............................................................................................................................3

Chapter 3 Wire .................................................................................................................................5

Chapter 4 Calibrate ........................................................................................................................ 134.1 Zero point calibration ...................................................................................................................134.2 Full scale calibration ..................................................................................................................... 13

Chapter 5 Maintenance .................................................................................................................. 155.1 Cleaning the membrane ...............................................................................................................155.2 Replacing the electrolyte solution and membrane ....................................................................... 15

Chapter 6 Accessories .................................................................................................................... 17

Contents

Quick Start Guide i

Page 4: Quick Start Guide July 2017 Rosemount 499ACL-02 · PDF fileChapter 1 Plan ... Chapter 6 Accessories ... Rosemount 499ACL-02-54 Sensor Wiring to Rosemount 1056 and 56 Transmitters Figure

Contents

ii Rosemount 499ACL-02

Page 5: Quick Start Guide July 2017 Rosemount 499ACL-02 · PDF fileChapter 1 Plan ... Chapter 6 Accessories ... Rosemount 499ACL-02-54 Sensor Wiring to Rosemount 1056 and 56 Transmitters Figure

1 Plan

1.1 Unpacking and inspection1. Inspect the shipping container. If it is damaged, contact the shipper immediately for

instructions.

2. Save the box.

3. If there is no apparent damage, unpack the container. Be sure all items shown on thepacking list are present. If items are missing, notify Rosemount immediately.

1.2 Product description

Rosemount 499ACL-02 Sensor PartsFigure 1-1:

A. Membrane retainerB. Membrane assemblyC. O-ringD. CathodeE. Electrolyte fill plug (wrap with pipe tape)F. Pressure equalizing portG. Sensor cable (integral cable shown)

1.3 Specifications

Sensor specificationsTable 1-1:

Physical characteristics Specifications

Pressure 0 to 65 psig (101 to 549 kPa abs)

Temperature (operating) 0 to 50 °C (32 to 122 °F)

Process connection 1 in. MNPT

Plan

Quick Start Guide 1

Page 6: Quick Start Guide July 2017 Rosemount 499ACL-02 · PDF fileChapter 1 Plan ... Chapter 6 Accessories ... Rosemount 499ACL-02-54 Sensor Wiring to Rosemount 1056 and 56 Transmitters Figure

Sensor specifications (continued)Table 1-1:

Physical characteristics Specifications

Wetted parts Noryl®(1), Viton®(2), and silicone

Cathode Gold

Process connections Sensor must be used in flow cell PN 24091-01

(1) Noryl is a registered trademark of General Electric.

(2) Viton is a registered trademark of DuPont Performance Elastomers.

Plan

2 Rosemount 499ACL-02

Page 7: Quick Start Guide July 2017 Rosemount 499ACL-02 · PDF fileChapter 1 Plan ... Chapter 6 Accessories ... Rosemount 499ACL-02-54 Sensor Wiring to Rosemount 1056 and 56 Transmitters Figure

2 Install

The sensor is intended for use only in the Rosemount TCL sample conditioning system fortotal chlorine. Consult the instruction manual for the Rosemount TCL sample conditioningsystem for more information.

Install

Quick Start Guide 3

Page 8: Quick Start Guide July 2017 Rosemount 499ACL-02 · PDF fileChapter 1 Plan ... Chapter 6 Accessories ... Rosemount 499ACL-02-54 Sensor Wiring to Rosemount 1056 and 56 Transmitters Figure

Install

4 Rosemount 499ACL-02

Page 9: Quick Start Guide July 2017 Rosemount 499ACL-02 · PDF fileChapter 1 Plan ... Chapter 6 Accessories ... Rosemount 499ACL-02-54 Sensor Wiring to Rosemount 1056 and 56 Transmitters Figure

3 Wire

NOTICE

For additional wiring information on this product, including sensor combinations not shownhere, please refer to the Liquid Transmitter Wiring Diagrams.

Rosemount 499ACL-02-54 Sensor Wiring to Rosemount 1056 and 56Transmitters

Figure 3-1:

Wire

Quick Start Guide 5

Page 10: Quick Start Guide July 2017 Rosemount 499ACL-02 · PDF fileChapter 1 Plan ... Chapter 6 Accessories ... Rosemount 499ACL-02-54 Sensor Wiring to Rosemount 1056 and 56 Transmitters Figure

Rosemount 499ACL-02-54-60/499ACL-02-54-VP Sensor Wiring toRosemount 1056 and 56 Transmitters

Figure 3-2:

Wire

6 Rosemount 499ACL-02

Page 11: Quick Start Guide July 2017 Rosemount 499ACL-02 · PDF fileChapter 1 Plan ... Chapter 6 Accessories ... Rosemount 499ACL-02-54 Sensor Wiring to Rosemount 1056 and 56 Transmitters Figure

Rosemount 499ACL-02-54 Sensor Wiring to Rosemount 5081 transmitterFigure 3-3:

Wire

Quick Start Guide 7

Page 12: Quick Start Guide July 2017 Rosemount 499ACL-02 · PDF fileChapter 1 Plan ... Chapter 6 Accessories ... Rosemount 499ACL-02-54 Sensor Wiring to Rosemount 1056 and 56 Transmitters Figure

Rosemount 499ACL-02-54-60/499ACL-02-54-VP Sensor Wiring toRosemount 5081 Transmitter

Figure 3-4:

Wire

8 Rosemount 499ACL-02

Page 13: Quick Start Guide July 2017 Rosemount 499ACL-02 · PDF fileChapter 1 Plan ... Chapter 6 Accessories ... Rosemount 499ACL-02-54 Sensor Wiring to Rosemount 1056 and 56 Transmitters Figure

Rosemount 499ACL-02-54 Sensor Wiring to Rosemount 1066 TransmitterFigure 3-5:

Wire

Quick Start Guide 9

Page 14: Quick Start Guide July 2017 Rosemount 499ACL-02 · PDF fileChapter 1 Plan ... Chapter 6 Accessories ... Rosemount 499ACL-02-54 Sensor Wiring to Rosemount 1056 and 56 Transmitters Figure

Rosemount 499ACL-02-54-60/499ACL-02-54-VP Sensor Wiring toRosemount 1066 Transmitter

Figure 3-6:

Rosemount 499ACL-02-54-VP Sensor Pin-out Diagram (Top View ofConnector End of Sensor)

Figure 3-7:

Wire

10 Rosemount 499ACL-02

Page 15: Quick Start Guide July 2017 Rosemount 499ACL-02 · PDF fileChapter 1 Plan ... Chapter 6 Accessories ... Rosemount 499ACL-02-54 Sensor Wiring to Rosemount 1056 and 56 Transmitters Figure

When making a connection through a junction box (PN 23550-00), wire point-to-point.

NOTICE

Use a wire nut and pigtail (included) when connecting several wires to the same terminal.

Wire

Quick Start Guide 11

Page 16: Quick Start Guide July 2017 Rosemount 499ACL-02 · PDF fileChapter 1 Plan ... Chapter 6 Accessories ... Rosemount 499ACL-02-54 Sensor Wiring to Rosemount 1056 and 56 Transmitters Figure

Wire

12 Rosemount 499ACL-02

Page 17: Quick Start Guide July 2017 Rosemount 499ACL-02 · PDF fileChapter 1 Plan ... Chapter 6 Accessories ... Rosemount 499ACL-02-54 Sensor Wiring to Rosemount 1056 and 56 Transmitters Figure

4 Calibrate

4.1 Zero point calibrationThe sample conditioning system converts total chlorine into iodine, which the Rosemount499ACL-02 sensor measures. Even in the absence of iodine, the Rosemount 499ACL-02sensor generates a small signal called the zero current. Failure to correct for the zerocurrent can produce a bias, particularly if the total chlorine concentration is small (<0.2ppm). Zero the sensor when it is first placed in service and every time the fill solution ischanged.To zero the sensor:

Procedure

1. Pour a cup of deionized or bottled water.

2. Place the sensor in the water.

3. Wait until the sensor current has reached a stable low value (at least two hours).

4. Follow the transmitter prompts for zeroing the sensor.

The zero current should be between -10 and +50 nA. For more information, refer to theRosemount TCL Manual.

4.2 Full scale calibrationBecause stable dilute chlorine standards are not available, the sensor must be calibratedagainst the results of a laboratory test run on a grab sample of the process liquid.

1. Place the sensor in the flow cell inside the TCL.

2. Start the sample and reagent flow.

3. Adjust the concentration so that it is near the upper end of the operating range.

4. Wait for the readings to stabilize.

5. Follow the transmitter prompts to complete the calibration.

6. After calibration, go to the Diagnostics menu and check the sensitivity.

The sensitivity should be between 900 and 1,200 nA/ppm. For more information, refer tothe Rosemount TCL manual.

Calibrate

Quick Start Guide 13

Page 18: Quick Start Guide July 2017 Rosemount 499ACL-02 · PDF fileChapter 1 Plan ... Chapter 6 Accessories ... Rosemount 499ACL-02-54 Sensor Wiring to Rosemount 1056 and 56 Transmitters Figure

Calibrate

14 Rosemount 499ACL-02

Page 19: Quick Start Guide July 2017 Rosemount 499ACL-02 · PDF fileChapter 1 Plan ... Chapter 6 Accessories ... Rosemount 499ACL-02-54 Sensor Wiring to Rosemount 1056 and 56 Transmitters Figure

5 Maintenance

When used in clean water, the total chlorine sensor requires little maintenance. Generally,the sensor needs maintenance when the response becomes sluggish or noisy or whenreadings drift following calibration. Maintenance frequency is best determined byexperience. Sensors used in dirty water require more frequent maintenance andcalibration. However, if experience shows that the sensor is holding calibration and notdrifting appreciably between calibration intervals, the maintenance interval can beextended.

WARNING!

PRESSURIZED SPRAY INJURYBefore removing the sensor, be absolutely certain that the process pressure is reduced to0 psig and the process temperature is lowered to a safe level!

5.1 Cleaning the membraneKeep the membrane clean. Clean the membrane with water sprayed from a wash bottle. Use a soft tissue to gently wipe the membrane.

5.2 Replacing the electrolyte solution andmembrane

WARNING!

HARMFUL SUBSTANCEFill solution may cause irritation. May be harmful if swallowed. Read and follow manual.

Procedure

1. Unscrew the membrane retainer.

2. Remove the membrane assembly and O-ring.

See Figure 1-1.

3. Hold the sensor over a container with the cathode pointing down.

4. Remove the fill plug.

5. Allow the electrolyte solution to drain out.

6. Remove the old pipe tape from the plug.

7. Wrap the plug with one or two turns of pipe tape..

8. Prepare a new membrane.

Maintenance

Quick Start Guide 15

Page 20: Quick Start Guide July 2017 Rosemount 499ACL-02 · PDF fileChapter 1 Plan ... Chapter 6 Accessories ... Rosemount 499ACL-02-54 Sensor Wiring to Rosemount 1056 and 56 Transmitters Figure

a. Hold the membrane assembly with the cup formed by the membrane andmembrane holder pointing up.

b. Fill the cup with electrolyte solution.

9. Hold the sensor at about a 45° angle with the cathode end pointing up.

10. Add electrolyte solution through the fill hole until the liquid overflows.

11. Tap the sensor near the threads to release trapped air bubbles.

12. Add more electrolyte solution if necessary.

13. Place the fill plug in the electrolyte port and begin screwing it in.

14. After several threads have engaged, rotate the sensor so that the cathode ispointing up and continue tightening the fill plug.

Do not overtighten.

15. Place a new O-ring in the groove around the cathode post.

16. Cover the holes at the base of the cathode stem with several drops of electrolytesolution.

17. Insert a small blunt probe, like a toothpick with the end cut off, through the pressureequalizing port.

See Figure 1-1.

CAUTION!EQUIPMENT DAMAGE

Do not use a sharp probe. It will puncture the bladder and destroy the sensor.

18. Gently press the probe against the bladder several times to force liquid through theholes at the base of the cathode stem. Keep pressing the bladder until no air bubblescan be seen leaving the holes. Be sure the holes remain covered with electrolytesolution.

19. Place a drop of electrolyte solution on the cathode; then place the membraneassembly over the cathode.

20. Screw the membrane retainer in place.

The sensor may require several hours operating at the polarizing voltage toequilibrate after the electrolyte solution has been replenished.

Maintenance

16 Rosemount 499ACL-02

Page 21: Quick Start Guide July 2017 Rosemount 499ACL-02 · PDF fileChapter 1 Plan ... Chapter 6 Accessories ... Rosemount 499ACL-02-54 Sensor Wiring to Rosemount 1056 and 56 Transmitters Figure

6 Accessories

Part # Description

33523-00 Electrolyte fill plug

9550094 O-ring, Viton 2-014

33521-00 Membrane retainer

23501-02 Total chlorine membrane assembly: includes one membrane assembly andone O-ring

23502-02 Total chlorine membrane kit: includes three membrane assemblies and three O-rings

9210438 Total chlorine sensor fill solution, 4 oz (120 mL)

Accessories

Quick Start Guide 17

Page 22: Quick Start Guide July 2017 Rosemount 499ACL-02 · PDF fileChapter 1 Plan ... Chapter 6 Accessories ... Rosemount 499ACL-02-54 Sensor Wiring to Rosemount 1056 and 56 Transmitters Figure

Accessories

18 Rosemount 499ACL-02

Page 23: Quick Start Guide July 2017 Rosemount 499ACL-02 · PDF fileChapter 1 Plan ... Chapter 6 Accessories ... Rosemount 499ACL-02-54 Sensor Wiring to Rosemount 1056 and 56 Transmitters Figure

Accessories

Quick Start Guide 19

Page 24: Quick Start Guide July 2017 Rosemount 499ACL-02 · PDF fileChapter 1 Plan ... Chapter 6 Accessories ... Rosemount 499ACL-02-54 Sensor Wiring to Rosemount 1056 and 56 Transmitters Figure

LIQ-QSG-499ACL-02

Rev K

2017

Global HeadquartersEmerson Automation Solutions6021 Innovation BlvdShakopee, MN 55379, USA

+1 800 999 9307 or +1 952 906 8888

F +1 952 949 7001

[email protected]/RosemountLiquidAnalysisNORTH AMERICAEmerson Automation Solutions8200 Market BlvdChanhassen, MN 55317

Toll Free +1 800 999 9307

F +1 952 949 7001

[email protected]/RosemountLiquidAnalysis

EUROPEEmerson Automation SolutionsNeuhofstrasse 19a P.O. Box 1046CH-6340 BaarSwitzerland

T + 41 (0) 41 768 6111

F + 41 (0) 41 768 6300

[email protected]/RosemountLiquidAnalysis

MIDDLE EAST AND AFRICAEmerson Automation SolutionsEmerson FZEJebel Ali Free ZoneDubai, United Arab Emirates, P.O. Box 17033

T +971 4 811 8100

F +971 4 886 5465

[email protected]/RosemountLiquidAnalysis

ASIA-PACIFICEmerson Automation Solutions1 Pandan CrescentSingapore 128461Singapore

T +65 777 8211

F +65 777 0947

[email protected]/RosemountLiquidAnalysis

Linkedin.com/company/Emerson-Automation-Solutions

twitter.com/rosemount_news

Facebook.com/Rosemount

youtube.com/RosemountMeasurement

google.com/+RosemountMeasurement

AnalyticExpert.com

©2017 Emerson Automation Solutions. All rights reserved.

The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is amark of one of the Emerson family of companies. All other marks are the property of theirrespective owners. The contents of this publication are presented for informationpurposes only, and, while effort has been made to ensure their accuracy, they are not tobe construed as warranties or guarantees, expressed or implied, regarding the products orservices described herein or their use or applicability. All sales are governed by our termsand conditions, which are available on request. We reserve the right to modify or improvethe designs or specifications of our products at any time without notice.