5
QUICK REFERENCE GUIDE: MTA7900 Medi-Therm ® III Hyper/Hypothermia System SET UP Fill the reservoir with distilled water per the Operator’s Manual. Connect power, turn Medi-Therm III ON, Medi-Therm III will be in STANDBY mode. Apply Rapr•Round ® Body Wraps to patient without closing the Velcro ® -like closures. NOTE: Blue printed label is away from patient. Vest connects to one gray hose. See Figure 1. Legs: Connect one set of clear hoses from each leg together and the other set to the second gray hose. See Figure 2. BE SURE TO CHECK AND OPEN ALL HOSE CLAMPS PRIOR TO USE. Refer to FAQ #5 to check water flow. Insert temperature probe into patient. Connect patient temperature probe to adaptor and then to the Medi-Therm III. Once the Mode is selected and the pads are filled with water, adjust and secure Velcro ® -like closures. NOTE: WHEN THERAPY ENDS, CLOSE ALL HOSE CLAMPS PRIOR TO DISCONNECTING WRAPS. RECONNECT GRAY HOSES TOGETHER. FIGURE 1 FIGURE 2 AW 101274000 Rev. B SKIN CARE Routine skin care is required to maintain skin integrity. Document the skin integrity prior to initiation of therapy per your skin care protocol. Frequency of assessment and documentation will vary depending on the individual patient. Manually turn and reposition the patient at a minimum of every two (2) hours. Check the body wrap(s) to ensure that they are not constricting the soft tissue. Readjust the Velcro ® -like closures as necessary. Every two (2) hours check that the patient’s skin is in direct contact with the body wrap(s) and note any change in the skin integrity related to: Excess moisture Color of the epidermis Skin texture (other than already noted.) Dry the skin if excess moisture exists. A circular pattern identical to the body wraps may be seen on the patient’s skin: this pattern typically lasts for the period of time the body wraps are in place and resolves after the wraps have been discontinued.

QUICK REFERENCE GUIDE: MTA7900 Medi-Therm …es242-digitalhealth-spring2011.wikispaces.com/file/view/Gaymar+III...QUICK REFERENCE GUIDE: MTA7900 Medi-Therm® III Hyper/Hypothermia

  • Upload
    hakiet

  • View
    219

  • Download
    2

Embed Size (px)

Citation preview

Page 1: QUICK REFERENCE GUIDE: MTA7900 Medi-Therm …es242-digitalhealth-spring2011.wikispaces.com/file/view/Gaymar+III...QUICK REFERENCE GUIDE: MTA7900 Medi-Therm® III Hyper/Hypothermia

QUICK REFERENCE GUIDE: MTA7900Medi-Therm® III Hyper/Hypothermia System

SET UP• Fillthereservoirwithdistilledwaterperthe

Operator’sManual.

• Connectpower,turnMedi-ThermIIION,Medi-ThermIIIwillbeinSTANDBYmode.

• ApplyRapr•Round®BodyWrapstopatientwithoutclosingtheVelcro®-likeclosures.NOTE:Blueprintedlabelisawayfrompatient.

• Vestconnectstoonegrayhose.SeeFigure1.

• Legs:Connectonesetofclearhosesfromeachlegtogetherandtheothersettothesecondgrayhose.SeeFigure2.

• BESURETOCHECKANDOPENALLHOSECLAMPSPRIORTOUSE.RefertoFAQ#5tocheckwaterflow.

• Inserttemperatureprobeintopatient.

• ConnectpatienttemperatureprobetoadaptorandthentotheMedi-ThermIII.

• OncetheModeisselectedandthepadsarefilledwithwater,adjustandsecureVelcro®-likeclosures.

NOTE:WHENTHERAPYENDS,CLOSEALLHOSECLAMPSPRIORTODISCONNECTINGWRAPS.RECONNECTGRAYHOSESTOGETHER.

FIGURE 1

FIGURE 2AW 101274000 Rev. B

SKIN CARE• Routineskincareisrequiredtomaintainskinintegrity.

Documenttheskinintegritypriortoinitiationoftherapyperyourskincareprotocol.Frequencyofassessmentanddocumentationwillvarydependingontheindividualpatient.

• Manuallyturnandrepositionthepatientataminimumofeverytwo(2)hours.Checkthebodywrap(s)toensurethattheyarenotconstrictingthesofttissue.ReadjusttheVelcro®-likeclosuresasnecessary.

• Everytwo(2)hourscheckthatthepatient’sskinisindirectcontactwiththebodywrap(s)andnoteanychangeintheskinintegrityrelatedto:

Excessmoisture

Coloroftheepidermis

Skintexture(otherthanalreadynoted.)

• Drytheskinifexcessmoistureexists.

• Acircularpatternidenticaltothebodywrapsmaybeseenonthepatient’sskin:thispatterntypicallylastsfortheperiodoftimethebodywrapsareinplaceandresolvesafterthewrapshavebeendiscontinued.

Page 2: QUICK REFERENCE GUIDE: MTA7900 Medi-Therm …es242-digitalhealth-spring2011.wikispaces.com/file/view/Gaymar+III...QUICK REFERENCE GUIDE: MTA7900 Medi-Therm® III Hyper/Hypothermia

AUTO MODES - COOLING

PresstheAUTObuttonuntiltheLEDlightindicatesthedesiredmode.

AUTORAPIDModecoolsthepatientasquicklyaspossible.SeeFigure3.

AUTOMODERATEModelimitsthewatertemperatureto15°C/27°Fbelowpatienttemperature.SeeFigure4.

AUTOGRADUAL Modelimitsthewatertemperatureto10°C/18°Fbelowpatienttemperature.SeeFigure5.

Select°C/°Ftemperaturescale

AdjusttheSetPointtothedesiredpatienttemperaturesettingperyourprotocol:____________°C/°Ffor___________hours

Assesspatientandreadjustwrapsifneeded.

NOTE:SetPointTemperatures(°C)areforexampleonly.PleaseusetheappropriateSet/Targettemperatureperyourprotocol.Patienttemperaturesshownareforillustrationpurposes.

AW 101274001 Rev. B

Auto Moderate Limited to 15°C below patient temperature

Auto Rapid Cools as rapidly as possible

Auto Gradual Limited to 10°C below patient temperature

QUICK REFERENCE GUIDE: MTA7900Medi-Therm® III Hyper/Hypothermia System

FIGURE 3 FIGURE 4 FIGURE 5

Page 3: QUICK REFERENCE GUIDE: MTA7900 Medi-Therm …es242-digitalhealth-spring2011.wikispaces.com/file/view/Gaymar+III...QUICK REFERENCE GUIDE: MTA7900 Medi-Therm® III Hyper/Hypothermia

NOTE:WhenusingAUTO MODERATEorAUTO GRADUALwarming,changingthemodeorSetPointduringtherapymayaffectthedurationofthetherapy.

NOTE:SetPointTemperatures(°C)areforexampleonly.PleaseusetheappropriateSet/Targettemperatureperyourprotocol.Patienttemperaturesshownareforillustrationpurposes. AW 101274002 Rev. B

Select°C/°Ftemperaturescale

AdjusttheSetPointtothedesiredpatienttemperaturesettingperyourprotocol:____________°C/°Ffor____________hours

Assesspatientandreadjustwrapsifneeded.

PresstheAUTObuttonuntiltheLEDlightindicatesthedesiredmode.

AUTORAPIDModewarmsthepatientasquicklyaspossible.SeeFigure6.

AUTO MODERATEMode:Rateofwarmingisapprox.0.33°C/houror1.0°Cevery3hours.SeeFigure7.

AUTO GRADUALMode:Rateofwarmingisapprox.0.17°C/houror1.0°Cevery6hours.SeeFigure8.

AUTO MODES - WARMING

FIGURE 6

Auto Rapid Warms as rapidly as possible

RateofwarmingasdirectedbyprotocolsrequiresuserinterventiontoselectSetPointatappropriateintervals.

RATE CONTROLLED WARMING MODES

FIGURE 7 FIGURE 8

Auto Gradual (0.17°C/hr or 1.0°C every 6 hours)

Auto Moderate (0.33°C/hr or 1.0°C every 3 hours)

QUICK REFERENCE GUIDE: MTA7900Medi-Therm® III Hyper/Hypothermia System

Page 4: QUICK REFERENCE GUIDE: MTA7900 Medi-Therm …es242-digitalhealth-spring2011.wikispaces.com/file/view/Gaymar+III...QUICK REFERENCE GUIDE: MTA7900 Medi-Therm® III Hyper/Hypothermia

FREQUENTLY ASKED QUESTIONS1. Why does the WATER temperature display continually fluctuate

during maintenance and warming in Automatic Mode?TheunitisdesignedtousethewatertowarmortocoolthepatientbasedontheSETPOINT.WhenthePATIENTtemperatureishigherthantheSETPOINT(see①inthefigureatright),theunitwillautomaticallylowertheWATERtemperaturetobringthePATIENTbacktotheSETPOINT.Inthesamemanner,whenthePATIENTtemperatureislowerthantheSETPOINT(see②inthefigureatright)theunitwillautomaticallywarmthewatertoincreasethePATIENTtemperature.

WhenthePATIENTreachesSETPOINT,itisnormalfortheWATERtemperaturetofluctuatesignificantlyaboveorbelowthePATIENTtemperatureasthemachinerespondscontinuouslytothePATIENTtemperaturevariation.

2. Why is my patient warming too quickly or too slowly in the Auto Gradual or Auto Moderate Modes?

TheMedi-ThermIIIunitisdesignedtocontrolthePATIENTtemperature±1.0°Cofthewarmingratethroughoutthetherapy.ItwillautomaticallyincreasethePATIENTtemperatureataspecificrateof0.33°C(MODERATE)or0.17°C(GRADUAL)perhour.Deviationsfromthiscouldbecausedbythefollowingconditions:

a.Patient-specificfactors,includingfever,shivering,seizures,etc.

b.Interruptionofthetherapy,e.g.adjustingtheSETPOINTduringtherapyorswitchingtootherModesintherapy.Whenthetherapyisinterrupted,theMedi-ThermIIIwillautomaticallyresetthetherapy,usingthecurrentPATIENTtemperature,andthismayimpactthedurationofthetherapy.Toavoidchangesinthedurationofthetherapy,do not changetheModeortheSETPOINTuntilthetherapyiscompleted.

(Questions continued on next page)

①②

AW 101274003 Rev. B

Graph is for illustration purposes only.

QUICK REFERENCE GUIDE: MTA7900Medi-Therm® III Hyper/Hypothermia System

Page 5: QUICK REFERENCE GUIDE: MTA7900 Medi-Therm …es242-digitalhealth-spring2011.wikispaces.com/file/view/Gaymar+III...QUICK REFERENCE GUIDE: MTA7900 Medi-Therm® III Hyper/Hypothermia

QUICK REFERENCE GUIDE: MTA7900Medi-Therm® III Hyper/Hypothermia System

Gaymar Industries, Inc.10CentreDriveOrchardPark,NY14127-2295TollFree1.800.828.7341Fax1.800.993.7890International+1716.662.8636InternationalFax+1716.662.0730www.gaymar.com

AW 101274004 Rev. B

P/N: 101274000 Rev. B 06/10

3. What should I be concerned about if the WATER temperature drops more than 10°C (18°F) below the PATIENT temperature and SET POINT for more than 30 minutes during maintenance or warming?

AsustaineddropinWATERtemperatureindicatesthatthemachineisworkinghardtomaintainthePATIENTtemperatureattheSETPOINT.Thereareanumberofreasonswhythiscouldoccur,someofwhichmayhavemajorclinicalimplications.Conditionssuchasshivering,feverandseizuresmaygeneratealargeamountofheatandtheMedi-ThermIIIwilllowertheWATERtemperatureinanattempttomaintainPATIENTtemperatureatSETPOINT.IncreasedheatproductionmaynotmanifestasaPATIENTtemperatureincreaseifthemachinecanovercometheheatburdenbyloweringtheWATERtemperature.

4. Can I defibrillate my patient during temperature therapy?IfdefibrillationisrequiredduringtemperaturetherapyitisnotnecessarytoremovetheRapr•RoundVest.SimplyloosentheVelcro®-likeclosureandopenthevesttoexposethechest.Defibrillationcannotbeperformedthroughthewrap.Defibrillationpadsthatadheretotheskinmayremainontheskinunderthechestwraptoperformdefibrillationwithoutremovingthechestwrap.

5. How do I check the flow through the body wraps?(SeeFigure9):Withtheunitrunning,disconnectonegrayhosesetfromtheMedi-ThermIII.Iftheflowisnotobstructed,thealarmwillnotsound.Ifthereisanobstructiontotheflowthroughtheotherhose,thealarmwillsound.Youshouldthencheckforclosedclamps,and/orrepositionthebodywrapattachedtotheconnectedhoseuntilthealarmsilences.Reconnectthehosetothemachine.Repeattheprocedurewiththesecondhoseset.

For additional information, please review the MTA7900 Operator’s Manual.

FREQUENTLY ASKED QUESTIONS continued

FIGURE 9