Upload
lyduong
View
213
Download
0
Embed Size (px)
Citation preview
QUALITY CONTROL AND QUALITY ASSURANCE FOR MOLECULAR
INFECTIOUS DISEASE LABS
Richard L. Hodinka, Ph.D. University of South Carolina School of Medicine Greenville
Greenville Health System, Greenville, SC [email protected]
Objectives
Describe pre-analytical, analytical, and post-analytical phases of QC/QA
Discuss guidelines for test requests and specimen collection, handling and processing
Explain lab design and best lab practices
Define reagent prep and use and instrument maintenance and calibration
Summarize proficiency testing, training and competency assessment
Discuss interpreting, recording and reporting test results
Quality Control – refers to the measures that must be included during each assay run to verify that the test is working properly
Quality Assurance – defined as the overall program used to ensure that the final results being reported by the laboratory are correct
Quality Assessment
Quality Assessment (Proficiency Testing) – means to determine the quality
of the results generated by the laboratory. Quality assessment is a challenge
to the effectiveness of the QC and QA programs that are in place.
Variables Affecting Quality of Results
Educational background and training of personnel
Selection and condition of specimens
Controls used in test runs
Reagents
Equipment
Interpretation, recording and reporting of results
Landscape of Laboratory Testing
Diagnostic lab testing single highest volume activity in
healthcare
Drives 70% of decision-making across medicine
12.8 billion tests performed each year in U.S.
Large scale analysis of 1.6 million results from 46 of 50
most commonly ordered laboratory tests
On average, 30% of all tests are probably unnecessary
Overutilization and underutilization are wide spread
Molecular ID Testing
Plays major role in clinical care of patients with infectious diseases
Identify specific pathogens
Determine and optimize appropriate therapy
Monitor response to therapy
Assess prognosis
Provide disease surveillance
Accurate and timely diagnosis required
Laboratory Testing Errors
Quality Use of Microbiology Tests
Using the right test
At the right time
For the right patient
With the right clinical condition
At the right price
With the right outcome
Molecular QC/QA
One of the largest stumbling blocks for traditional Clinical Microbiology Laboratories that start performing molecular testing
Molecular QC/QA more closely resembles Clinical Chemistry QC/QA
Newer CAP Microbiology Checklists have been refined and full sections now devoted to Molecular Microbiology
Many changes in QC/QA; many are new to traditional microbiologists
Molecular QC/QA Program
Written clearly defined program that ensures quality throughout the preanalytical, analytical, and post-analytical phases of testing
Capable of identifying systems problems and opportunities for system improvement
The laboratory must be able to develop plans for corrective/preventative action based upon data from its QC/QA system
Monitoring Testing Process
Importance of Quality Control
Ensure accuracy and reproducibility of lab test results used in patient care
Ensure integrity and confidence in test results
Provide safety in management of patients
Comply with local, regional, and national regulations and lab accreditation requirements
Maintain staff morale and reputation of lab
Types of Molecular Assays
Unmodified FDA-Cleared/Approved Tests
Modified FDA-Cleared/Approved Tests
Laboratory-Developed Tests
Qualitative versus Quantitative Tests
Pre-Analytical Phase
Addresses activities that occur prior to testing
Test requests/ordering
Patient identification
Specimen collection and transport
Specimen handling and processing
Rejection of specimens
Test Requests/Ordering
Develop well-designed order entry (EMR or manual) Allow for sufficient identification of patient and
ordering physician, the test requested, and pertinent clinical information
Should include date and time of specimen collection and specimen type; information on collector
Should accompany patient’s specimen to the laboratory
Ensure that clinical indications are appropriate for test(s) requested
Specimen Collection
Specimen collection, transport, handling and processing can have a significant effect on final test results; integrity of target nucleic acid sequence must be maintained
Establish clearly defines criteria for monitoring proper specimen collection, labeling, preservation, transportation, and storage of specimens
Establish educational programs
Develop firm (but fair) rejection policy
Guidelines for Specimens
Clearly defined method(s) of collection of specimens from all sources
Sample type and quantity of specimen
Preparation of collection site
Timing of collection during disease
Specimen quality/adequacy
Collection and transport devices
Identification of specimens
Appropriate transport conditions
Storage time, temperatures, and conditions
Specimens for RNA Detection
Specimens for RNA detection require special collection and handling conditions
RNA may be degraded by RNases
Use method that stabilizes RNA until specimen can be processed and stored appropriately
Analytical Phase
Written Procedure Manual Nucleic Acid Extraction and Storage Contamination Control
Laboratory Design Laboratory Practices
Testing Controls Test Performance Equipment Maintenance Personnel Competency Proficiency Testing Accreditation
Procedure Manual
Written instructions that contain sufficient detail that qualified laboratory personnel can perform tests consistently and accurately
Must include: Principle of Test Calibration
Clinical Significance Reference Ranges
Specimen Requirements Calculations
Reagents Needed Result Reporting
Procedural Steps Test Limitations
Quality Control References
Yearly review by Director or designee; review all new procedures and revisions prior to implementation
Electronic or paper procedures acceptable
Nucleic Acid Isolation
Adequate procedures for release and isolation of target nucleic acid
Keep process simple and minimize number of manipulations
Maintain integrity of target while inactivating or removing inhibitory or interfering substances
Yield, purity, intactness can be measured Ideally, isolated nucleic acid should be tested
immediately; if delay, store under appropriate conditions until testing can be completed
Nucleic Acid Isolation/Preparation
Contamination Control
During specimen collection and transport
During initial specimen handling and
processing by lab
During nucleic acid extraction
During amplification
During product detection
Contamination of reagents
Laboratory Design
Molecular space should be divided into at least 3 separate work areas: Area 1 – Reagent preparation area/room
Area 2 – An area/room for specimen processing/nucleic acid extraction
Area 3 – An area/room for amplification and detection
*Area 4 – For preparation of controls, calibrators, standards
Laboratory Practice 1
Dedicated supplies and reagents for each work area
Use plugged aerosol resistant pipette tips
Use closed systems or methods to control product carryover (e.g., uracil-N-glycosylase (UNG) to inactivate amplified product)
Use appropriate positive and negative controls
Meticulous cleaning is a must!
Laboratory Practice 2
Follow unidirectional flow from pre-PCR to post-PCR
Pulse-spin reagents, samples, controls; Add reagents to vials before samples
Use gloves and protective clothing which are dedicated to each area
Decontaminate pipettes and instruments and wipe work surfaces with 10% bleach; rinse with 70% ethanol
Meticulous cleaning is a must!
Laboratory Practice 3
Carefully open and close all tubes to minimize aerosolization of contents
Keep all non-essential tubes closed during sample addition
Order of preparation and loading of samples for PCR should be actual clinical specimens first, followed by positive and then negative controls
Hoods and/or glove boxes (with UV light) are often necessary for nucleic acid isolation and set up of amplification reactions
Meticulous cleaning is a must!
General Rules of Thumb
Traffic can only go from clean to dirty area
Objects can only go from clean to dirty area
Reagents and supplies needed in a clean area need to be prepared and stored in a clean area
Equipment used in clean area cannot be moved and used in dirty area
Personal safety equipment cannot move with you from dirty to clean area
Meticulous cleaning is a must!
General Rules of Thumb
Protocol books, worksheets, sample tubes, test data and results cannot move from dirty area to clean area
Can employ color coded lab coats, pipettes, safety glasses, racks, etc. to monitor traffic
Have respect for RNases if working with RNA
Meticulous cleaning is a must!
Reagents & Solutions
All test reagents and controls should be stored properly and in a manner which minimizes contamination and degradation
Prepare and divide into single-use aliquots Store in area that is separate from specimen
preparation and post amplification Do not store in frost-free freezers Dedicated equipment and supplies should be used Reagents can be premixed into single-use, master
mixes Meticulous cleaning is a must!
Reagents
Test new lots for purity, functionality, concentration, and contamination before use
Compare performance of new reagents to ones currently in use
Test new reagents against reference materials; monitor quality with use
Keep meticulous QC records
QC for Molecular ID Assays
Qualitative
Controls (match matrix) Amplification: Positive for each
analyte; Negative (reagent blank or specimen) and Sensitivity
Every run, establish range values, and monitor
Internal (inhibition) for extraction and amplification
Calibration verification (if use cutoff to distinguish positive/negative) – weak positive
Every 6 mo; after any major system component change and QC failure
Quantitative
Controls (match matrix) Amplification: Positives at ≥2
levels and Negative
Every run, establish range values, and monitor
Internal (inhibition) for extraction and amplification
Calibration verification (verify cutoff)
Analytical measurement range verification (low, mid, high) Every 6 mo; after any major
system component change and QC failure
Adapted from EM Burd 2010 Clin. Microbiol. Rev. 23:550-576
Daily QC
QC of Multiplex Tests
Ideally, include a positive control for each of the nucleic acid targets included in test for each run
Often impractical if number of targets is large
Multi-target control or rotate single target controls
Example of Multiplex Controls
ANALYTE M211 M212
Adenovirus Positive Negative
Coronavirus 229E Negative Positive
Coronavirus HKU1 Negative Positive
Coronavirus NL63 Negative Positive
Coronavirus OC43 Negative Positive
Human Metapneumovirus Positive Negative
Human Rhinovirus/ Enterovirus Entero 1 Positive Negative
Entero 2 Positive Negative
HRV1 Positive Negative
HRV2 Positive Negative
HRV3 Positive Negative
HRV4 Positive Negative
Influenza A H1-2009
FluA-H1-
2009
Positive Negative
Influenza A H3 FluA-H1-pan Positive Positive
FluA-H3 Positive Negative
FluA-pan1 Negative Positive
FluA-pan2 Positive Negative
Influenza B Negative Positive
Parainfluenza Virus 1 Positive Negative
Parainfluenza Virus 2 Negative Positive
Parainfluenza Virus 3 Negative Positive
Parainfluenza Virus 4 Positive Negative
Respiratory Syncytial Virus Negative Positive
Bordetella pertussis Negative Negative
Chlamydophila pneumoniae Negative Negative
Mycoplasma pneumoniae Negative Negative
M211 and M212 Main Molecular
Controls run periodically and with each
new lot. No need for negative control.
Assay contains internal RNA process
and second stage DNA PCR controls.
FilmArray RP Panel M210 (Maine Molecular Controls)
QC Acceptance Limits & Statistics
Inhibition Assessment
Distribution of QC Results
W. Greg Miller. 2011. Chapter 10, Quality Control in Henry’s 22nd ed.,
Elsevier, Philadelphia, PA
Types of Errors
Random Systematic Random & Systematic
BLUE = True Value Average
GREEN = Sample Value Average
Westgard QC Evaluation Rules
Used to define specific performance limits for a particular assay
Can detect both random and systematic errors Warning rules and mandatory rules have been
applied over the years Violation of warning rules triggers a review of
test protocol, reagent performance and equipment calibration
Violation of mandatory rules triggers rejection
Levey-Jennings Plots
https://www.westgard.com
Warning Rules
Mandatory Rules
Westgard Sigma Multirule QC
Primers/Probes - LDTs
Commercial Reference Materials
Quality control material may be difficult to find for less common pathogens
Number of companies now provide reference material for a growing list of pathogens Acrometrix
Advanced Biotechnologies
American Type Culture Collection (ATCC)
Main Molecular Quality Controls, Inc.
Seracare Life Sciences
Zeptometrix
International Standards
WHO, NIST (US) and NIBSC (UK) have been active in establishing IS for viruses
HIV-2, HIV-2, HAV, HBV, HCV, Parvovirus, HPV-16, HPV-18, CMV, and EBV
Primary (reference) standards are assigned quantity values accepted as “gold” standards
Provide means to normalized inter-laboratory and method to method variability
Secondary standards are larger pools of material that are quantified by direct comparison to the primary standards
Calibrators should be traceable to IS if available
Equipment & Instruments
Ensure equipment and instrument function
Establish system for monitoring
Detection of drift, instability, or malfunction before problem affects test results
Regular preventive maintenance and cleaning
If multiple instruments, functions must be check against each other every 6 months or after repair
Keep meticulous maintenance, service and repair records
Establish service contracts with vendor
Quality Assurance/Monitoring
A continuous process
Documents that a test which has already been verified/validated is repeatedly giving the expected results as test is performed over time
Confirms that test continues to perform according to laboratory’s requirements and its intended use
Quality Monitoring Process
Personnel competency assessment Meticulous record keeping Vigilance in laboratory Quality control monitoring Quality improvement Internal and external proficiency testing Correlation with clinical findings; interacting with
healthcare providers Trend analysis Storage of specimens for follow-up
testing/evaluations Integral part of a labs QA program
Quality Control Monitoring
Have system in place to continuously monitor performance of positive and negative controls, standards, and calibrators
Perform wipe testing – can reinforce need to disinfect the environment daily
Monitor prevalence rates of specific viral diseases
Training & Competency
Establish a complete program of employee training, verification, and competency
Verify personnel competency at least semiannually during first year and at least annually thereafter
Reverification necessary if change in method or instrumentation
Assay Training
Determine magnitude of training and skill level required
Each operator must demonstrate proficiency with the method; may require certification
Should include understanding of: Sample collection, handling and storage
Reagent handling and storage
Proper test protocol
How to interpret and report results
QC/QA for the system
Competency Assessment
Direct observation of routine patient test performance, including specimen handling, processing, testing, instrument maintenance, and function tests
Monitor test result recording and reporting
Review worksheets, QC records, PT results, and PM records
Wet testing with proficiency samples
Assessment of problem-solving skills
Proficiency Testing/EQA
Challenge for PT providers to keep up with pace of new test development and new technology in molecular diagnostics
Many laboratories need combinations of commercially-available and alternative proficiency testing programs
CAP Proficiency Tests
Alternative performance assessment
Integrate into the workload
Rotated among people at the bench
Corrective action on failed PT surveys
Proficiency Testing/EQA
General principles and rules of PT should be applied
PT samples should challenge the diagnostic scope of the test Range of genetic diversity (QUAL) and extremes of
measurement ranges (QUANT)
Samples should be relevant to current test targets and compatible with available technologies
Samples should free of “matrix” effects and reflect the range of specimens types tested
Source of nucleic acid can have a significant effect on participant performance
Alternative PT Assessment
Used when no commercial PT products available
May also be used to challenge pre-analytical and post-analytical processes not addressed by commercially available PT
Example - Quality Control for Molecular Diagnostics (QCMD)
Participate in ungraded/educational PT
Split sample analysis with other laboratories
Establish in-house PT Program
Use regional pools (can be purchased)
Clinical validation by chart review
Internal or External Split-Sample PT
Minimum of 3-5 samples per testing event
At least 2-3 testing events at equal intervals per year
Use full range of positive and negative specimens in panel
Samples should represent clinical specimens
Use of in-house specimens or manufactured reference materials
Corrective Actions
Mistakes will occur; avoid assigning blame
Clearly document what has happened
Thoroughly investigate, take immediate corrective action, institute preventive measures
Errors often result of larger process issues with reagents, personnel, equipment, workflow
Ultimate goal: DO NO HARM
Post-Analytical Phase
Laboratory Test Reports
Timeliness of Reporting
Correction of Errors
Patient Confidentiality
Recording & Reporting Results
Laboratory test reports should be clear, concise, accurate, and fully interpretive
Verify all results before reporting as final Monitor and report specimen (in)adequacy Final report should include a summary of test method,
result(s) and interpretation if appropriate Gene target (biological false positives and negatives) Amplification method Lower limit of detection ASR disclaimer if LDT Quantitative (AMR, LLQ vs. LLD, log10 conversion, and action
thresholds) Name and location of laboratory
Good practice to have reports reviewed by 2nd party
Interpretation & Reporting of Results
Significance of results must be evaluated with respect to agent, specimen site, and clinical situation
Interpretive criteria must include provisions for equivocal results
Reporting of results should include as much information as necessary for clinician to properly interpret the results
Interact with healthcare providers
Quality Management: Turnaround Times
Provide test results in a timely manner to have greatest impact on patient care and management
Have system to monitor TATs Are TATs appropriate for the intended
purpose of the test? Consultation with medical staff Batch testing versus on-demand Examples: HSV and enteroviruses in CSF;
respiratory viruses; viral load tests; MRSA and C. difficile
Quality Management: Statistics
Additional Items
Correction of Errors – Laboratory should have a system in place for timely review and correction of clerical and analytical errors
Confidentiality – Testing records and reports should be maintained in a manner that preserves patient privacy and confidentiality
Total Quality Management
Comprehensive organizational approach
Focus on continually improving quality and efficiency with which lab operates
Includes not only QC and QA, but other technical and administrative considerations that may indirectly influence quality and efficiency