NMR excitation (detecting NMR)
• An ensemble of spins (of a single type and I = 1/2) generates an average magnetization, Mo, upon interaction with an external magnetic field, Bo:
• This magnetization is proportional to the population difference of spins in the low and high energy levels, and that it is precessing at a frequency o, the Larmor frequency of the particular observed spin at a particular Bo
• So far, nothing happened. We need to do something to the system to observe any kind of signal. What we do is take it away from this condition and observe how it goes back to equilibrium. This means affecting the populations...
y
x
z
Bo
Mo
z
x
i
B1 = C * cos (ot)
B1
Transmitter coil (y)
yBo
NMR excitation (continued)
• We need the system to absorb energy. The energy source is an oscillating electromagnetic radiation generated by an alternating current:
NMR excitation (continued)
• How is that something that has a linear variation can be thought as circular field? A linear variation in y is the linear combination of two counter-rotating circular fields:
+=
+=
+=
•Only the one vector that rotates at o(in the same direction
of the precession of Mo) interacts with the bulk magnetization
= +
RF Coil: Transmitting B1 Field
The effect of the tiny B1 isto cause M to spiral awayfrom the direction of thestatic B0 field
B110–4 Tesla If B1 frequency is not close toresonance, B1 has no effect
RF pulse
B1 field perpendicular to B0 Mxy
Mz
Classical Description
• Observe NMR Signal Need to perturb system from equilibrium.
B1 field (radio frequency pulse) with Bo/2 frequency Net magnetization (Mo) now precesses about Bo and B1
MX and MY are non-zero Mx and MY rotate at Larmor frequency System absorbs energy with transitions between aligned and unaligned states
Precession about B1stops when B1 is turned off
Polarization distribution after a pulse
Transverse magnetization
Polarization distribution at thermal equilibrium
PULSED MAGNETIC FIELDS
z
x y
90° Pulse
= 2 B1
• Once the pulse is turned off, the spins resume their precessional motion.
• The individual spins precess on their individual cones.• On a macroscopic scale, the bulk magnetization moment also
precesses, rotating in the xy-plane, perpendicular to the main magnetic field.
Pulse Generator & Receiver Systema) RF pulse width determines band-width of excitation
Null, no effect
Invert signal, 180o pulse
Maximum effect at
1
z
x y
Larmor precession in the rotating frame
Laboratory frame
Bo
z
x' y'
Rotating frame at 0
=-B=0 B=0!!
0 = -B0
z
x' y'
B1 is the only magnetic field “experienced” by the spin system
Larmor precession in the rotating frame
Bo
z
x' y'
Rotating frame at 0
z
x' y'
B1 is the only magnetic field “experienced” by the spin system
=-B=0 B=0!!
This is the key to how the very weak RF field can affect the magnetization in the presence of the much stronger B0 field. In the rotating frame this field along the z axis appears to shrink, and under the right conditions can become small enough that the RF field is dominant.
A Mechanical Analogy: A Swingset
• A person sitting on a swing at rest is “aligned” with externally imposed force field (gravity)
• To get the person up high, you could simply supply enough force to overcome gravity and lift him (and the swing) up– Analogous to forcing M over by turning on a huge static B1
• The other way is to push back and forth with a tiny force, synchronously with the natural oscillations of the swing– Analogous to using the tiny RF B1 to slowly flip M over
Larmor precession in the rotating frame
z
x' y'
Rotating frame at rf = 0
z
x' y'
=-B=0 B=0!!
Rotating frame at rf ≠ 0
=-B= 0 -rf
0 is in the MHz orderis in the kHz order
Laboratory and rotating framesrf rf rf
0 Hz3600 Hz -3600 Hz
1H 10s 90o pulse
cycle in 10 us cycle in 40 us1=1/21/4*10-5=Hz1=*2 rad/sec=3600*2 rad/sec
=arctan(1/°
The effective field
90°
-80°
80°
1H 10s 90o pulseB1 of 25 kHz
0,988
0,99
0,992
0,994
0,996
0,998
1
-3000 -2000 -1000 0 1000 2000 3000
Mxy/Mz0
The effective field
13C 12s 90o pulseB1 of 20.8 kHz
Mxy/Mz0
0,8
0,85
0,9
0,95
1
-100 -50 0 50 100
ppm @ 150 MHz
-40
-30
-20
-10
0
10
20
30
40
-100 -50 0 50 100
Maximum affect: width of excitation:±1/(4PW) Hz
The longer the pulseThe more selective it is
Selective pulse
1H 1ms 90o pulse ±250 Hz ±0.4 ppm at 600 MHzUsed for pulses exciting only H2O magnetization
Fourier pairs
To get a perfectly rectangular excitation profile we will need an infinitely long sinc pulse which is, naturally, impractical, so the waveform has to be truncated
Homonuclear decoupling
Main problems:• Selective irradiation is
not always feasible• Bloch-Siegert shift:
Heteronuclear Decoupling
• The frequency difference 2-1 lies in the MHz range. The second field B2 is produced by a separate transmitter.
• A curly bracket is used to indicate which nucleus is being decoupled during the experiment, e.g. 13C1H indicates a 13C spectrum acquired with 1H decoupling.
Heteronuclear DecouplingBroadband decouplingUsing the spin flip experiment (MLEV, Malcolm Levitt):
x
y
x
y
J / 2
J / 2
x
y
°
J / 2
J / 2
x
y
Heteronuclear DecouplingA train of pulses will be susceptible to the errors due to pulse length inaccuracy, since they will accumulate. The non-ideal pulse can be replaced with a composite pulse.