Upload
others
View
4
Download
0
Embed Size (px)
Citation preview
PT/ILC Requirements -
R48 and some
implementation ideas.
(R80 also covered)
By
John Wilson
Accreditation and Metrology
Services (Pty) LTD.
Accreditation & Metrology Services -John Wilson 2017 1
What is PT ?
Accreditation & Metrology Services -John Wilson 2017 2
Unfortunately - In our case
– it is not ….
What is PT?
• PT stands for Proficiency Testing
• The Collins dictionary defines Proficiency as
an exam which test how proficient or skilledsomeone is in a particular activity, field of study, language, etc
• Similarly test (ing) (the noun) is defined as
A test is a series of questions that you must answer or actions that you must perform in order to show how much you know about a subject or how well you are able to do something.
Collins English Dictionary. Copyright © HarperCollins Publishers
Accreditation & Metrology Services -John Wilson 2017 3
What is an ILC ?
• ILC is the acronym for Inter Laboratory
Comparison.
• It is basically the same thing as PT and the
two words are interchangeable.
• Traditionally, PT has been used by Analytical
and Chemical Metrology Laboratories and ILC
by Physical Metrology Laboratories.
Accreditation & Metrology Services -John Wilson 2017 4
What is an ILC ?
• In medical laboratories, it is sometimes
known as EQA (external quality assurance).
• There is an international move to phase out
the use of ILC (words) and just use PT.
Accreditation & Metrology Services -John Wilson 2017 5
What is PT/ILC ?
• Neither PT or ILC are defined in the VIM
(Vocabulaire international de métrologie)
• More appropriately, there are definitions in
ISO/IEC 17025, ISO/IEC 17011, ISO/IEC 17043,
ILAC P9, SANAS A01, SANAS PM01, etc.
• We are today particularly interested in the
SANAS R48 and R80 documents and how they
effect the laboratory.
• R48-04 - Proficiency Testing and other comparison
programme requirements for Calibration laboratories.
• R80-03 – Proficiency Testing … for Testing & Medical
laboratories and Blood transfusion services.
Accreditation & Metrology Services -John Wilson 2017 6
What do SANAS require?
• The SANAS documents reflect largely what
the International documents require and the
requirements are similar to what most
accreditation bodies require of a laboratory.
• The R48 and R80 documents define the
detailed requirements for laboratories. (Read
them)!
• It all starts with a plan!
Accreditation & Metrology Services -John Wilson 2017 7
What do SANAS require?
• Calibration laboratories need to have a plan
that reflects the activities done in the last 5
years AND the plan for next 5 years.
• This plan must cover the listed parameters
and the ranges on the accreditation schedule
at least once in each 5 year period.
Accreditation & Metrology Services -John Wilson 2017 8
What do SANAS require?
• The Testing laboratories are required to have
a plan which covers their scope over two
accreditation cycles, one cycle in arrears and
one in the future.
• These plans differ from initial assessment PT
requirements where the lab has to prove it
can achieve the range and CMCs requested.
Accreditation & Metrology Services -John Wilson 2017 9
What do SANAS require?
• All plans must address the risk profile of the
laboratory and it’s activities. The new 17025
puts a much larger emphasis on risk analysis.
• All plans must be addressed during the
Laboratory Management meetings for
applicability, progress and performance.
• Details of what must be in the plan are listed
in the appropriate R48 or R80 documents.
Accreditation & Metrology Services -John Wilson 2017 10
What do SANAS require?
• Calibration laboratories are required to
participate in the National Audit Program and
may use these audits as part of the PT plan.
• All laboratories must preferably use PT
schemes that have independently been shown
to comply with the requirements of ISO/IEC
17043.
• All laboratories shall participate in PT
schemes for the items on their schedule
including specific tests or methods or where
specific instruments have been separately
listed. Calibration should also consider
including different methods in their plans.
Accreditation & Metrology Services -John Wilson 2017 11
What do SANAS require?
• Where no formal PT is available, Cal Labs shall
indicate other interlaboratory activities. Such an
ILC must still comply with the full R48 PT/ILC
activity and reporting requirements.
• Where no formal PT is available, Testing Labs
shall indicate suitable alternative means such as
reference materials, intralaboratory or other
comparisons. Again R80 compliance is required.
• The lack of commercially available PTs is not
considered a valid reason for non participation.
• Where PTs are not available or practical, suitable
alternatives may be proposed by the laboratory to
SANAS for consideration.
Accreditation & Metrology Services -John Wilson 2017 12
What do SANAS require?
• PT reports shall be clear and comprehensive
and have the information required by the R48
or R80 documents. (Identification (people and
items), responsibilities, evaluation methods
and conclusions are key).
• All laboratories shall review their own
performance and investigate all measurement
results. If some fail to meet the minimum
acceptance criteria then document the root
cause analysis and all corrective and
preventative actions taken. (i.e. raise a non
conformance in the quality system).
Accreditation & Metrology Services -John Wilson 2017 13
Common problems found
• Assessors have found the following problems
during assessments:-
• Failure to have a comprehensive plan.
• Failure to keep the plan up to date.
• Failure to adhere to the plan.
• Not using commercially accredited schemes
when they are available. (False economy).
• Participation with another lab and each lab
draws up it’s own report with different
conclusions. (i.e. no protocol done before ILC)
• Inappropriate choice of artefact and/or
reference value/material.
Accreditation & Metrology Services -John Wilson 2017 14
Common problems found
• Trying to pass off a PT for one device as
being adequate as competence to calibrate
another (Drying oven PT given for Autoclave).
• Failing to prepare a measurement protocol
before commencing the PT/ILC. (This means
different labs measure different points and
report differently so evaluation is not
possible).
• The report does not cover the required items
listed in the SANAS documents.
Accreditation & Metrology Services -John Wilson 2017 15
Common problems found
• Inadequate information in the report to
validate the findings. (e.g.. Results, En value
but no uncertainties recorded).
• No evaluation of the PT and it’s adequacy.
• One word reporting. (e.g. Conclusion –
“Acceptable” (even when results do not
overlap).
Accreditation & Metrology Services -John Wilson 2017 16
Considering Alternatives
• There is no benefit from evaluating a device
which has also been the source of
traceability.
• Intralaboratory comparisons only evaluates
one metrologist against another, usually the
procedure, equipment, traceability is exactly
the same. It does not confirm that the
procedure, equipment or traceability is valid.
Accreditation & Metrology Services -John Wilson 2017 17
Considering Alternatives
• To do a good ILC, it is not a matter of two or
more labs doing an intercomparison and then
expecting the assessor to do the evaluation
and decision making regarding who passes
and who fails. Things to consider are:-
• Decide on an artefact and check if it will be
adequate for the required accuracy or UoM
• Decide how the reference value will be
established.
• How will drift be monitored.
• Who will write an acceptable protocol.
Accreditation & Metrology Services -John Wilson 2017 18
Considering Alternatives
• Define how evaluation will be done. (En, Z)
• Who will do the evaluation.
• Who will write the report to the required
standard defined in the documents.
• The organising lab cannot usually use the ILC
to demonstrate it’s own performance,
especially if it provides the reference value
and the reference UoM.
• How will confidentiality be taken care of?
Accreditation & Metrology Services -John Wilson 2017 19
Conclusion
• PT/ILCs are definitely an integral part of an
accredited laboratory’s life.
• They should add value and confidence to the
laboratory (not just satisfy the assessor).
• They are a wonderful training tool for
evaluating new metrologists progress and
provides a source of CPD points for the
metrologist concerned.
Accreditation & Metrology Services -John Wilson 2017 20
Conclusion
• Be creative and innovative when planning the
PT schedule so that it is interesting and value
adding.
Accreditation & Metrology Services -John Wilson 2017 21
Conclusion
• Thanks to Neville Tayler from SANAS for his
valuable information.
• Thanks to Steve Sidney from the NLA for his
contribution based on running the NLA PT/ILC
accredited schemes.
• Thanks to all the assessors who diligently
work with the laboratories during
assessments to make sure that the PT/ILC
plans are reasonable and within the correct
risk profile for the laboratory
• Thanks to you for listening.
Accreditation & Metrology Services -John Wilson 2017 22
PT/ILCs – here to stay!
Accreditation & Metrology Services -John Wilson 2017 23
John WilsonCell : +27 83 652 0770eMail : [email protected]