PSG Basic & PSG Plus Gas Sampling Probes
Instruction Manual
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 2 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Important Instructions
Read this page before proceeding!
AGT-PSG GmbH & Co. KG designs, manufactures and tests its products to meet many national and international stand-ards. Because these products are sophisticated technical products, you MUST properly install, operate and maintain them to ensure they continue to operate within their normal specifications. The following instructions should be adhered to and integrated into your safety program when installing, using and maintaining AGT-PSG GmbH & Co. KG products.
Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life, personal injury, property damage, damage to the product - and warranty invalidation. Read all instructions prior to installing, operating, and servicing the product. If you do not understand any instructions, contact your AGT-PSG representative for clarification. Follow all warnings, cautions, and instructions marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation, and maintenance of the product. Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes. Connect all products to the proper electrical and pressure sources. To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product. When replacement parts are required, ensure that qualified people use replacement Parts specified by AGT-PSG. Unauthorized parts and procedures can affect the product’s performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY. Look-alike substitutions may result in fire, electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being per-formed by qualified persons, to prevent electrical shock and personal injury. The information contained in this document is subject to change without notice. 1st Edition 07/2016 2nd Edition 12/2019 3rd Edition 05/2021 © 2021 by AGT-PSG GmbH & Co.KG
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 3 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Introduction The instruction manual contains information about the component assembly, function, installation, operation, maintenance, return and disposal of the following PSG gas sampling probes: PSG Basic and PSG Plus incl. PSG Plus DSBP (Dual Stage Back Purge).
This manual covers these PSG gas sampling probes, information about configurations and options. But the manual does not cover installation / operation of probes in hazardous areas (explosive environments), with flammable gas components and mobile gas sampling probes (in such cases, please look at additional manuals to cover these options and applications).
Use
Gas sampling probes are used for Gas applications: continuous measurement acc. to the extractive measurement method – for Continuous Emission Monitoring (CEM) and for Process Analysis Technology (PAT). The gas sampling probes PSG Basic & PSG Plus / Plus DSBP are designed for continuous gas sampling to meet the requirements for these applications. To accomplish a constant, “clean” and unaltered sample gas PSG takes care for applicable filter material, high sample gas temperatures, elevated or low pressure, high dust concentrations / water & acid dew points, corrosive gas components, aerosols & mist and tough environmental conditions to avoid wrong gas sampling via probe. Subsequently follows a gas transfer to the different gas measuring units via (heated) sample gas lines. The probes are designed for representative sampling through rough filtering of (hot and wet) sample gases which contain different amounts of dust particles. Due to the diverse options they meet all the different requirements of varying applica-tions regarding filtering, corrosion resistance – especially taking care for high sample gas dew points (acid dew points) by heated external filters. For very high dust amounts additional an internal filter (pre-filter) or alternatively a sophis-ticated single or dual stage back purge system is offered.
Definitions
The following definitions explain the use of the terms WARNING, CAUTION and NOTE in this manual.
Indicates an imperative operational procedure or an important condition or instruction
Indicates an operational or maintenance procedure, a process, a condition, an instruction, etc. Failure to comply may result in injury, death or permanent health risk.
Indicates an operational or maintenance procedure, a process, a condition, an instruction, etc. Failure to comply may result in damage to or destruction of the instrument, or impaired performance.
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 4 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Introduction
Symbols – some are used in this manual
This symbol indicates general information relevant to the safety or health of personnel if no specialized symbol is available.
Dangerous voltages may be exposed.
Possible danger of explosion
Warning of flammable substances
Toxic substances may be present
Substances harmful to health may be present
Warning of corrosive substances
Indicates notes relating to heavy instruments
This symbol indicates general noteworthy information or tips which have no influence on the safety or health of personnel.
The instrument must be connected to ground before use.
The instrument must be disconnected from the main power supply before being serviced or repaired.
Unplug the power cable.
Follow the instructions.
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 5 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Introduction
Symbols – some are used on or inside the device
Dangerous voltages may be exposed. The instrument may only be opened and serviced by qualified per-sonnel after disconnecting the power supply.
Hot surfaces may be exposed.
Grounding connection point
Follow the instructions.
These symbols are - if used - on the device's user interface to indi-cate the unit's operational status.
EC Declaration of Conformity and CE Marking
CE Certification
The product described in this instruction manual complies with the following EU directives: EMV Instruction
The requirements of the EU directive 2014/30/EU “Electromagnetic compatibility“ are met. Low Voltage Directive
The requirement of the EU directive 2014/35/EU “Low Voltage Directive“ are met. The compliance with this EU directive has been examined according to DIN EN 61010. CE marking guideline
The guideline 93/68/EWG is followed as well as 2014/68/EU for pressure equipment. Declaration of conformity The CE Certification can be downloaded from the AGT-PSG homepage.
Scope of Delivery
Sample gas probe Flange gasket and screws - according to the order: see order acknowledgement Instruction manual Attached and enclosed accessories: see separate position in the order acknowledgement.
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 6 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Introduction
Information on Name Plates
AGT-PSG provides 2 name plates: On filter housing On terminal box The filter housing name plate provides details on the configuration of the probe, installed sample gas options. It indicates the following details:
Part No. Job No. AGT-PSG Pmax Porosity SN Pt CW CE
Filter Housing Name Plates as Example (New & former version) This name plate is located outside the sample probe filter housing. The terminal box name plate provides the following details:
AGT-PSG Company address Order No. Article No. Year of construction Probe No.
This name plate is located outside the terminal box of the sample gas probe. Please refer to this serial number if you have any questions about your sample gas probe or if you need to order spare
parts or consumables.
AGT-PSG GmbH & Co. KG Weißkirchener Straße 3 D-61449 Steinbach/Ts.
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 7 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Safety Instructions
Intended Use
The PSG gas sampling probes PSG Basic and PSG Plus or PSG Plus DSBP are intended for use as part of sample handling of a gas analyzer system using extractive measuring techniques.
This device must not be used in medical, diagnostic or life support applications or as safety device. The implementation as a safety device is prohibited in cases where SIL
certification is required.
This instrument may be operated and serviced only by trained personnel who are famil-
iar with the function of this device and aware of the risk on an open and energized unit.
This device is not designed and should not be used in hazardous areas without addi-
tional safety measures. Use of this equipment in explosive atmospheres could result in in-jury, explosion or death. Consult factory for hazardous area equipment!
This device is not designed and should not be used in the analysis of explosive or flam-mable gases. Use of this equipment in this way could result in flash burn, injury, explosion
or death. Consult factory for applications with flammable gases!
General Safety Notice / Residual Risk
A safe and proper operation requires a proper transport, storage, unpacking and a correct installation followed by a proper set-up finally. After start of operation a careful maintenance in the requested intervals is necessary - only by qualified personal. Read these SAFETY instructions carefully prior to installation and the operating information before start-up. Follow the information and warnings given in this instruction manual! The regulations, standards and guidelines in this operator's manual are applicable acc. to EU legislation. Take care that the relevant national regulations should be followed when the device is used in other countries. If this equipment is used in a manner not specified in these instructions, safety may be impaired.
Despite incoming goods inspections, production control, routine tests and application of state-of-the-art test methods, an element of risk remains when operating a process gas sampling probe.
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 8 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Safety Instructions
Even when operated as intended and observing all applicable safety instructions some residual risks remain, including, but not limited to, the following:
• An interruption of the protective earth line, e.g. in an extension cable, may result in risk to the user.
• Live parts are accessible when operating the device with cover removed.
• The emission of gases hazardous to health may even be possible when all gas connections have been correctly made.
Avoid exposure to the dangers of these residual risks by taking particular care when in-stalling, operating, maintaining and servicing the analyzer.
Authorized Personnel
In-depth specialist knowledge is a necessity for working with a gas sampling probe.
Authorized personnel for installing, operating, servicing and maintaining the probe are instructed and trained qual-ified personnel of the operating company / the manufacturer.
It is the responsibility of the operating company to
• Train staff, observe safety regulations and follow the instruction manual.
Operators must - before starting work -
• Have been trained,
• Have read & understood all relevant sections of the instruction manual,
• and know the safety mechanisms and regulations.
To avoid personal injury and loss of property, do not install, operate, maintain or service this instrument before reading and understanding this instruction manual and receiving appropriate training.
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 9 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Safety Instructions
Electrical Safety
General Considerations
The operation and maintenance of electrical apparatus always requires the observance of several safety rules. Please observe the following safety rules when making electrical connections, performing maintenance or repair work and espe-cially when replacing components. These safety rules are for your own personal safety as well as the safety of the appa-ratus.
Electrical components of this system may only be serviced by qualified technicians. Adher-
ence to relevant local safety specifications is mandatory.
Ensure that all contacts that can be touched while performing work are either voltage free or protected from such contact.
If the probe must be opened disconnect all power supplied to the device. If the probe must remain powered in order to perform the maintenance or repair work, then this work may only be performed by qualified personnel capable of judging potential dangers.
Never disconnect ground connections within or outside of the instrument. Disconnecting the ground could result in potential safety hazards.
When replacing electrical fuses use only fuses of the specified type.
If you feel that safe operation of the instrument is no longer possible, take it out of operation. Furthermore, take steps to ensure that it cannot be put back into operation by unauthorized personnel.
Fuses may only be replaced by fuses of an identical type and with identical ratings. It is forbidden to bypass fuses or use repaired fuses.
Power Supply and Grounding
Before switching the unit on, ensure that the local mains voltage corresponds to the unit's oper-ating voltage.
A fuse should be provided at the installation site which will completely disconnect the unit in case of failure. An emergency cut-off switch should also be provided within easy reach of the instrument.
The power cord used with the unit must provide a protective earth (PE) contact.
Connect the protective earth connection (PE) before all other connections.
Working on units equipped with screw-type terminals for electrical connections may require the exposure of energized components.
Cables for external data processing must be double-insulated against mains power. If this is not possible, cables must be laid in such a way as to guarantee a clearance of at least 5 mm from power cables. This clearance must be permanently secured (e.g. with cable ties).
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 10 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Safety Instructions
Surface Temperature
This analytical sample handling device (sample gas probe) utilizes components which require high operating temperatures. These components which are accessible only after opening the cover may be designated in the diagram above.
Particular caution is required in handling the high temperature heater. Under operating condi-tions, the heater can have a temperature as high as 180 °C or even higher (up to 250 °C). These high temperatures are protected from direct contact by the thick insulation of the heat-ing device and normally present no danger. The metal components have temperatures ranging from 80° to 180°C (up to 250 °C). Servicing of the probe heater (or exchange the heater ) should only be performed by qualified personnel. Thermally insulated gloves must be worn when exchanging the heated parts.
Take care to wait long enough and let the parts cool down to avoid problems with the heating device.
Under operating conditions, the gas sampling probe temperature is controlled to approx. 180 °C (self-control) up to 250 °C (regulated). The heated components are accessible only after opening the cover.
Safe Operation not Possible
If it is apparent that safe operation is no longer possible, the device should be put out of operation and secured against
unauthorized use.
Safe operation is no longer possible:
o If the device is visibly damaged, o If the device no longer operates (indicated by alarm message), o After prolonged storage under adverse conditions, o After severe transport stresses, o If permissible temperature cannot be provided, o If PSG filter unit cannot be protected against direct heat radiation, rain, draughts & heavy soiling.
Remark: Protection of the filter unit with a protective box can be a good solution, c.f. (consult factory)!
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 11 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Technical Description
Main Common Features – PSG Basic, PSG Plus and Plus DSBP
Very large active outer filter surface leads to extreme low maintenance
o Dust concentration and corresponding maintenance:
< 100 mg/m3 requires maintenance any 2 years
< 1 g/m3 requires maintenance twice a year
< 3 g/m3 requires maintenance any 3 months
Different sample gas temperatures and corresponding tubes
o < 600 °C with Tube SS 1.4571
o < 900 °C with Tube SS 1.4893 or Hastelloy C4 (high resistant)
o < 1300 °C with Tube Kanthal APM
Temperature controlled filter heating up to 250 °C
o Temperature alarm contact should be 20 °C below setting
Self-regulated filter heating – fixed to 160 °C
o Temperature alarm contact included at 140 °C
Corrosion resistant probe leads to long lifetime being realized with
o Housing: SS 1.4571, Fittings: SS 316, Filter: SiC, Gaskets: FPM/ FFKM
Filter and gaskets material – standard and options
o Silicon carbide (SiC) standard filter - resistant ceramic version
o Borosilicate glass fiber filter – option if no acidic components available
o Gaskets FPM as standard (< 200 °C) - high resistant FFKM (< 250 °C)
Calibration (test) gas & single back purge connection as standard included
Filter change without tools and sample line dismounting
Compact protective housing for outdoor installation
o Weather protection box enables operation under rough conditions
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 12 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Technical Description
Performance/ Special Functions – Probe PSG Basic
Continuous gas sampling for low & medium dust loaded gases o Up to 3 g/m3
High dust load with options pre-filter PF or single stage back purge
o Up to 10 g/m3
Quick and easy installation and mounting
Extreme compact design allows installation with small space
Easy maintenance due to space-saving bracket design
Performance/ Special Functions – Probe PSG Plus
Single or dual stage back purge for high and very high dust concentrations
Dual stage back purge
o up to 40 g/m3
o Up to 280 g/m3 – under certain conditions, c.f.
Comfortable and quick filter changes without tools
8-hole flange for variable mounting
Upgradeable with diverse options
Special additional Function – Probe PSG Plus DSBP
Dual stage back purge via valve block & accumulator on board
Dual stage back purge
o up to 280 g/m3
2 Coax solenoid valves direct mounted at the probe
2 l pressure accumulator mounted direct at the probe
Additional insulation for ambient temperatures up to -30 °C
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 13 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Technical Description
Technical Data – Probe PSG Basic, PSG Plus and PSG Plus DSBP
Probe PSG Basic and Common Part of PSG Plus & Plus DSBP
Process Gas Sampling Conditions Low & Medium Dust
Pressure pabs: 50…6000 hPa (mbar (abs))
Sample Gas Temperature: Up to 600 °C Tube: SS 1.4571
Sample Gas Temperature: Up to 900 °C Standard Tube: High resistant Tube:
SS 1.4893 Hastelloy C4
Sample Gas Temperature: Up to 1300 °C Tube: Kanthal APM
Ambient temperature 250 W heater: -30 to +50 °C 350 W heater: -50 to +50 °C
Flow 30…1500 l/h, referred to 1013 hPa and 0 °C
Pressure drop Approx. 0.6 hPa at 100 l/h
Standard Basic with standard fil-ter unit
Dust concentration: Maintenance:
< 100 mg/m³ Any 2 years
< 1 g/m³ Twice a year
< 3 g/m³ Any 3 months
High Dust Content 1-stage back purge Pre-Filter PF > 3 up to < 10 g/m³ Option
Connections Sample gas G1/4" f (DIN ISO 228/1)
Test gas (stand.) / tubing (option) G1/4" f (DIN ISO 228/1) / 6 mm Tube
Filter Heating
Type Content
Heating sleeve incl. PT100
230/ 115 VAC, 50 - 60 Hz 250 W
230/ 115 VAC, 50 - 60 Hz 350 W
Ring heater self-regulated 230/ 115 VAC, 48 - 62 Hz 2 x 100 W
Isolation
Removable insulation jacket, heating sleeve only
Additional insulation protective housing, heating sleeve only, for ambient temperature:
-30 to +60 °C
Temperature, self-regulating Standard: 160 °C Alarm: 140 °C
Temperature, regulated
Up to 250 °C; Acid dew point and salification to be evaluated
Alarm: 20 °C below setting
Temperature control
PID-controller ST49 incl. solid state relays for DIN-rail-mounting
Heating sleeve only
With controller in connection box, heating sleeve only
Temperature sensor PT100 (only heating sleeve)
Filter Properties
Filter with Surface of 212 cm2 Ceramic, silicon carbide (SiC) Standard
Glass fiber: if no acidic components to be measured Special
Porosity SiC ceramic: 0.3 µm Borosilicate glass fiber: 0.1 μm
Tightness 10-4 hPa l/s
Dead volume ca. 280 ml
Dimensions 50/20 x 135 mm
Protective Housing
Dimensions 250 x 205 x 270 mm (L x B x T)
Material Stainless steel SS 304
Ambient temperature -20 °C to +60 °C; Option: -30 °C to +60 °C Add. measures
Weight Approx.14 kg Complete probe
For heating with thermostat control With adapted connection box
Without protective housing With adapted connection box
Without protective housing, for heating with thermostat control
With adapted connection box
Protection class connection box IP67 EN 60529
Mounting Flange DN65, PN6, 4-hole, form B according to DIN 2527
Installation angle 10° to 35° inclination to horizontal position Recommended
Materials in Contact with Sample Gas
Housing, gas connections / flange Stainless steel SS 1.4571 // Fitting: stainless steel SS 316
Gaskets FPM as standard: Up to 200 °C
FFKM as corrosion resistant version: Up to 250 °C
Filter material Silicon Carbide (SiC) Standard
Borosilicate (glass fiber) Special Pressure reduction valve at probe outlet
to prevent pressure peaks during back purge, 6 mm O.D. tube connection
Option
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 14 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Technical Description
Probe PSG Plus & PSG Plus DSBP
Process Gas Sampling Conditions
Pressure pabs // Sample Gas Temperature // Ambient temperature // Flow // Pressure drop
See: PSG Basic
Low & Medium Dust Content
PSG Plus: with standard unit filter
Dust concentration: Maintenance:
< 100 mg/m³ Any 2 years
< 1 g/m³ Twice a year
< 3 g/m³ Any 3 months
High Dust Content with single stage back purging
> 3 up to < 10 g/m³
Very High Dust Content with dual stage back purging
> 3 up to < 40 g/m3
Highest Dust Contents with dual stage back purging
> 40 up to 280 g/m3 PSG Plus DSBP
Connections
Sample gas // Calibration (test) gas (standard) / Tubing (option) See: PSG Basic
Back purge (standard)
Tubing (option)
2 x G3/8" f (DIN ISO 228/1)
Single stage 12 mm tube
Dual stage 12 mm tube
Filter Heating Type // Isolation See: PSG Basic
Filter Properties Filter // Porosity // Tightness // Dead volume // Dimensions See: PSG Basic
Protective Housing
Dimensions 330 x 205 x 270 mm (L x B x T)
Material // Ambient temperature // Weight // For heating with thermostat control // Protection class connection box
See: PSG Basic
Mounting Flange DN65, PN6, 8-hole, form B according to DIN 2527
Installation angle See: PSG Basic
Materials in Contact with Sample Gas
Housing, flange, gas connections Gaskets
See: PSG Basic
PSG Plus DSBP Options
2 Coax solenoid valves mounted at the probe & wired to connection box, voltage: 24 VDC
Additional insulation of protective housing
for ambient temperatures up to -30 °C
Pressure accumulator 2 l, material stainless steel, mounted at probe
connected with 12 mm tube at coax valves, max. operating pressure: 7 bar, pressure air connection: 12 mm tube
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 15 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Technical Description
Options for PSG Basic and/or PSG Plus and Plus DSBP
Performance – Gas Sampling Tubes PSG ST
Reliable Gas Sampling for precise representative measurement
For dust concentrations up to 3 g/m3
o without additional options
Temperature resistant tubes up to 1300 °C
o Up to 600 °C or 900 °C also selectable
Sampling length definable
Various tube materials available
o SS 1.4571
o SS 1.4893
o Hastelloy C4
o Kanthal APM
o Corrosion resistant options to be selected
Solutions for nearly all low, medium and high dust applications
o Additional options for very and extreme high dust contents
Sample mounting
o R3/4“-male thread fittings
Technical Data – Tubes PSG ST
Tube Material Max. Temp. [°C] Length* L [mm] Max. Length** [mm]
D [mm]
Part No.
Stainless steel SS 1.4571 600
1000 2500 27
80060022
Extension stainless steel SS 1.4571 100 80060645
Hastelloy C4 (very corrosion resistant) 900
1000 2500 25
08060438
Extension Hastelloy C4 100 80060646
Stainless steel SS 1.4893 900
1000 2500 27
80060388
Extension stainless steel SS 1.4893 100 80060647
Kanthal APM (corrosion resistant) 1300
1000 1500 27
80060660
Extension Kanthal APM 100 80060661
*Standard length/ standard extension: Other arbitrary intermediate length maybe selected – c.f. **Self-supporting, increased lengths may be selected (if supported at the sampling place due to appropriate measures like e.g. traverses) – c.f.
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 16 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Technical Description
Performance – Gas Sampling Pre-Filter PSG PF
For dust Concentrations up to 10 g/m3 without back purge
o Requires no dry purge gas at site
o Solution for dust concentrations > 3 g/m3 up to 10 g/m3
Low differential pressure even at high flow rates
Temperature resistant up to 900 °C
Stainless steel or corrosion resistant Hastelloy sinter metal filter
Sample length selectable with extension tube
Service life prolongation and protection against abrasion by deflector
Dust precipitation direct in the process
Large active surface
Long lifetime
Quick and simple mounting
Technical Data – Pre-Filter PSG PF
*Standard length/ standard extension: Other arbitrary intermediate length maybe selected – c.f.
Pre-Filter PSG PF Max. Temp. [°C] Length L [mm] Connection Porosity [µm]
D [mm]
Part No.
Stainless steel SS 1.4571 600 220
G3/4“a 2 45 80060492
520 50 08060572
Hastelloy C4 (corrosion resistant)
900 220
G3/4“a 2 45 80060664
520 50 80060665
Extension tube*
Stainless steel SS 1.4571 600
1000 R3/4” a – G3/4” i - 26
80060526
Extension stainless steel 100 80060645
Hastelloy C4 900
1000 R3/4” a – G3/4” i - 26
80060666
Extension Hastelloy C4 100 80060646
Deflector
Stainless steel SS 1.4571 600 220
clamping plate - - 80060493
520 80060632
Hastelloy C4 (corrosion resistant)
900 220
clamping plate - - 80060667
520 80060668
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 17 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Technical Description
Performance – Heated Gas Sampling Tube PSG HT
Homogeneous heating over entire tube length
o For low dew point sample gas – to avoid condensation
No cold spots
Corrosion resistant out of SS316 / 316L
Controlled heated up to 180 °C - for high acid dew points
In arbitrary lengths up to 2 m available
Internal tube diameter 12-20 mm available
With cable for electrical connection via connection box of PSG Basic & Plus
With thermocouple Fe-CuNi
Protection class IP68
Extensions with unheated tube or pre-filter:
o Tube end with G3/4“ female thread
Low dead volume
Technical Data – Tube PSG HT
Part No.
Process Gas Sampling Conditions PSG HT
Pressure pabs = 50…600 kPa
Temperature max. 250 °C
Flow 30…1500 l/h, referred to 100 kPa and 0 °C
Max. dust concentration 3 g/m³
Connections Extension tube / pre-filter G3/4" i (DIN ISO 228/1) optional 80060674
Heating
Type Fixed resistance heater
Temperature 180 °C
Temperature alarm 150 °C (adjusted at optional temperature controller)
Temperature control (optional)
PID-controller ST49 incl. solid state relay for DIN-rail-mounting 230 V, 50/60 Hz
50078850
PID-controller ST49 incl. solid state relay for DIN-rail-mounting 115 V, 50/60 Hz
50078851
Temperature sensor Fe-CuNi
Design
Standard lengths (others on request)
0.5 m 80060670
1 m 80060671
1.5 m 80060672
2.0 m 80060673
Material Stainless steel SS316 / 316L
Ambient temperature -20 °C to +180 °C
Weight approx. 9 kg (1 m)
Diameter external: 48.3 mm internal: 12 mm (up to 20 mm possible)
Dead volume 113 ml/m (12 mm internal diameter)
Protection class IP68 EN 60529
Electrics
Heat capacity 130 W/m
Voltage 230 VAC, 50…60 Hz (115 VAC on request)
Electrical connection 3 m cable (3 x 1.5 mm2 and 2 x thermocouple)
Electrical standard EN 61010, EN 60519-1
Mounting
Flange Stud bolts on both sides M12 x 40 mm corresponding to DN65 PN6 form B according to DIN 2527
Material Flange gasket Klingersil DN65 PN6, 4 x nut M12, 4 x spring washers and washers
Installation angle 10 ° - 35 ° inclination to horizontal position (rec-ommended)
Materials in con-tact with sample gas
Tube, flange, connections Stainless steel SS316 / 316L
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 18 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Technical Description
Performance – Pressure Reduction Valve PSG PR
Option for single stage back purge at PSG Basic
Option for single and dual stage back purge at PSG Plus & Plus DSBP
Safe protection of analyzer and gas conditioning components against damage
Reliable reduction of pressure peaks at sample gas outlet while back purging
o Reducing 6 bar (70000hPaabs) for successful back purge
o Protection of sensitive Analyzer components against high pressure
Reduction to max. 800 mbar at outlet
Increase of back purge efficiency
Dual function with connection for heated sample line
Dismountable for simple cleaning
Quick and simple mounting
Self-cleaning effect
Corrosion resistant
Temperature resistant
Compact design
The pressure reduction valve PSG PR can be used only without Pre-Filter PSG PF
Technical Data – Pressure Reduction Valve PSG PF
PSG PR
Part No. 80060195 80060675
Max. pressure sample gas outlet PSG Plus mbar 800
Closing pressure at 180 °C hPa abs. > 1.800
May. flow rate Nl/h 1000
Max. pressure bar (hPa abs.)
6 (70000)
Connections 6 mm – ¼“ RT m 8 mm – ¼“ RT m
Max. media temperature °C +250
Materials of media wetted parts SS 1.4571
Ambient temperature °C -40…+320
Design Data
Length mm 37.9 38.7
Weight kg approx. 0.04
Mounting threaded joint
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 19 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Technical Description
Dimensions and Descriptions
Heated Gas Sampling Probe PSG Basic
Dimensions in mm
PSG Basic Filter
PSG Basic with Protection Box, Back Purge and Calibration (Test) Gas Inlet
PSG Basic with Back Purge and Calibration Gas in-let & Sample Gas Outlet
PSG Basic with Connecting Flange (Test) and Sample Gas In-let
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 20 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Technical Description
Heating sleeve incl. PT100 Ring heater self-regulating
PID-controller ST49 & solid-state relays (25A) with heat sink for DIN-rail-mounting, heating sleeve only
Removable insulation jacket
Common Options for PSG Basic, PSG Plus and Plus DSBP
PSG Basic Dimensions PSG Basic with Protection Box, Back Purge Inlet and Calibration (Test) Gas Inlet
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 21 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Technical Description
Heated Gas Sampling Probe PSG Plus
PSG Plus Filter
PSG Plus with Protection Box: Back Purge Inlets via Filter Element & Chamber and Calibration (Test) Gas Inlet
PSG Plus with Flange and Sample Gas Inlet Back Purge Inlets via Filter & Chamber and Calibration (Test) Gas Inlet
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 22 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Technical Description
PSG Plus with back purge inlets via filter element & chamber and calibration gas inlet
PSG Plus w/o protective housing
PSG Plus lid design PSG Plus filter change w/o tools
PSG Plus Comfortable Filter Change
PSG Plus with protection box
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 23 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Technical Description
Heated Gas Sampling Probe PSG Plus DSBP
PSG Plus DSBP - Standard
Dimensions in mm
PSG Plus DSBP with Protection Box and 2 Coax Solenoid Valves
2 Coax Solenoid Valves
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 24 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Technical Description
PSG Plus DSBP - Options
PSG Plus DSBP with 2 Coax Valves and a 2 l Pressure Accumulator
Addition al DSBP Insulation
DSBP Coax solenoid valve for back purge DSBP 2 l Pressure Accumulator
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 25 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Technical Description
Gas Sampling Tubes PSG ST
Dimensions in mm
Gas Sampling Pre-Filter PSG PF
Length L: 220 / 520 mm (also Deflector)
Extension tube L: Standard Length = 1000 mm / Extension = 100 mm
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 26 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Technical Description
Heated Gas Sampling Tube PSG HT
Dimensions in mm
The four Standard Length Variations of Heated Probe Type
Length 1 500 mm
Length 2 1000 mm
Length 3 1500 mm
Length 4 2000 mm
Other Lengths on request – c.f.
Pressure Reduction Valve PSG PR
Dimensions in mm
PSG PR is available for tubing with 6 mm O.D. or 8 mm O.D.
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 27 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Installation
Unpacking
Danger of breakage - only for special PSG probe tube! Only a special silicon carbide tube is fragile (c.f.).
Step Action
1 Unpack the tube, the filter unit and, if applicable, the temperature controller.
2 Make sure that any accompanying accessories do not get lost.
3 Check the contents of the delivery to ensure that is complete by comparing the actual goods with the dispatch note
Keep the packing material for possible future transport. If damage has occurred during transport due to improper handling, please submit a damage report to the transport institution within seven days.
Probe Tube Installation in General
Only the silicon carbide tube (special version – on request) must be installed vertically.
Corrosion risk! The heated gas sampling tube PSG HT must be put into operation immediately after as-sembly and installation; otherwise there is a risk of corrosion!
Probe Tube and Filter
Possible combinations of the probe tubes and filters units:
Probe Tube Type Filter Unit Types Type of Assembly
Unheated PSG Basic/Plus Bolted
Heated PSG Basic/Plus Flange: DN65 PN6
Step Action
1 Install the customer-supplied wall tube with assembly flange into the wall of the flue gas duct or chimney according to the planning documents.
2 Use the green seal from accessories pack to seal the space between the probe tube and the wall tube.
3 Insert the pre-assembled gas sampling system in the wall tube.
4 Screw the assembly flange to the flange of the filter unit.
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 28 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Installation
Installation of PSG Basic, PSG Plus and Plus DSBP
Wall Tube Mounting of PSG Probe with Filter Unit
The following table states the recommended mounting angle of the PSG filter unit:
Mounting angle 10˚ 15˚ 20˚ 25˚ 30˚ 35˚
xmin mm 229 248 268 287 307 324
Connect the sample gas line to gas port B at the sample gas connection of the filter unit by means of a clamp ring screw fitting (not in the scope of supply and delivery).
Possible leakage! When tightening the screw connection, relieve the pressure on the gas port, e.g. by holding back with a suitable spanner. Otherwise there is a danger that the gas port is twisted and becomes un-tight or breaks off.
If probe back purging is available, connect the compressed air lines to gas port C (Single stage back purge) If test gas is to be fed in at the probe, connect the test gas line to gas port B. If the PSG Basic filter unit is operated at ambient temperatures below -20 °C, additional heating must be provided.
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 29 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Installation
Gas Connections
PSG Basic Gas Connections
A Sample gas outlet, G1/4” B Calibration (test) gas inlet, G1/4” C Back purge (single stage) gas inlet, G3/8”
A
C
B
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 30 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Installation
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 31 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Installation
PSG Plus and Plus DSBP Gas Connections
A Sample gas outlet, G1/4” B Calibration (Test) gas inlet, G1/4” C Back purge gas inlet (chamber & tube), G3/8” D Back purge gas inlet (filter), G3/8”
D
A B
C
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 32 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Installation
PSG Plus Gas Connections
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 33 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Installation
PSG Plus DSBP Gas Connections
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 34 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Installation
Gas and Electrical Connections PSG Basic and PSG Plus
PSG Basic
A Fitting, sample gas connection for tubing: 6 mm / 1/4” B Plug, G1/4”, calibration (test gas) for tubing 6 mm / 1/4” 1 Terminal box 2 2 pcs. cable gland M16 (Pt100 and reserve) 3 3 pcs. cable gland M20 (230 V / 115 V) 4 Cable gland M16 (230 V / 115 V) 5 Cable gland M16 (Pt100) 6 Bulkhead fitting, calibration / test gas connection for tubing: 6 mm / 1/4” 7 Mounting clamp 8 Bulkhead fitting, instrument air connection for tubing: 12 mm / 1/2” 9 Bulkhead fitting, sample gas outlet connection for tubing: 6 mm / 1/4”
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 35 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Installation
PSG Plus
A Bulkhead fitting, calibration / test gas connection for tubing: 6 mm / 1/4” B Bulkhead fitting, instrument air via filter for tubing: 12 mm / 1/2” C Bulkhead fitting, back purge via filter chamber & probe tube for tubing: 12 mm / 1/2” D Bulkhead fitting, sample gas outlet connection for tubing: 6 mm / 1/4” 1 Calibration / test gas tube, VA 1.4571 / 316, 6 x 1 mm 2 Back purge tube, CU, 12 x 1 mm 3 Terminal box, IP66 4 2 pcs. cable gland M16 (Pt100 and reserve) 5 3 pcs. cable gland M20 (230 V / 115 V) 6 Back purge tube, CU, 12 x 1 mm 7 Mounting clamp 8 Cable gland M16 (230 V / 115 V) 9 Cable gland M16 (Pt100)
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 36 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Installation
Installation of PSG Basic Components
Filter unit Filter
Heating Sleeve Insulation
Terminal Protective Box
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 37 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Installation
PSG Basic Filter Unit Components
1 Flange clamp 3” 2 Filtering Unit 3 Probe housing 4 Stainless steel ring 5 Upper filter holder 6 O-ring seal 45 x 3 mm 7 O-ring seal 59 x 4 mm 8 O-ring seal 50 x 4 mm 9 Filter stone 10 O-ring seal 45 x 3 mm 11 Lower filter holder
1
2
3
11
10
9
8
7
6
5
4
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 38 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Installation
PSG Plus & Plus DSBP Filter Unit Components
1 T-handle 2 Bridge 3 Detaching disk 4 Locking screw 5 Upper filter holder 6 Flange 7 O-ring seals 8 Filter stone 9 Bridge holding device 10 Probe housing 11 Lower filter holder
1
2
3 4
5
6
10
9
7
8
7
11
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 39 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Operation & Maintenance
Operating Instructions
This Operating Instruction contains the information that is required to install, put into operation, operate and maintain
probe tubes (unheated and heated) and filter unit types (PSG Basic and PSG Plus) safely and as specified. The Operating Instruction also contains information about the operation of the temperature controller. The complete Oper-ator's Manual of the temperature controller is a constituent of the scope of supply and delivery. If the information in this operator's manual does not cover a particular situation, PSG Service will be pleased to supply additional information as required.
Quick Guide for the Installation
We recommend strongly to read the complete instruction manual to get the complete in-
formation about the product! Just in case of a Short Form Manual (Quick Guide) you can use this short form – but take in mind: Not all important messages are included in the Quick Guide!
Step Action
1 Unpack the supplied equipment (see chapter “Installation”,section “Unpacking”)
2 Install the probe tube, see sections “Probe Tube Installation in General” and “Probe Tube and Filter”.
Install gas sampling tube PSG ST (unheated type)
Install gas sampling pre-filter PSG PF
Install heated gas sampling tube PSG HT
3 Mounting of the filter unit, see sections “Probe Tube and Filter” and “Installation of PSG Basic, PSG Plus and Plus DSBP”.
Mounting the PSG Basic filter unit (see section “PSG Basic Gas Connections”)
Mounting the PSG Plus filter unit (see section “PSG Plus and Plus DSBP Gas Connections”)
4 Electrical connections (see section “ Gas and Electrical connections PSG Basic and PSG Plus”)
Electrical connections for heated gas sampling tube PSG HT
Electrical Connections for the PSG Basic & Plus Filter Units
Electrical Connections for the Temperature Controller After you have carried out these steps, the probe tube and the filter unit are operative.
Please refer to the planning documents for the installation. Please also refer to the technical data (see chapter "Technical Description", see section “Technical Data”). The wall tube must be installed at the sampling point before the probe is installed!
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 40 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Operation & Maintenance
Inspection and Maintenance - PSG Basic and PSG Plus
Danger of burns!
The work described in this chapter requires specialist knowledge. As a result, it may only be carried out by persons who are qualified and specially trained. The metal parts may have high temperatures as a result of operation.
Health hazard! Dependent on the substances which it comes into contact with during operation, the filter element could be contaminated with toxic or corrosive substances. Always wear suitable
protective clothing for cleaning work. Filter Unit Cleaning interval - PSG Basic and PSG Plus:
The operating life of filter elements depends on the operating conditions. If necessary, you can remove it so that you can make a visual inspection and clean mechanically the element from impurities or exchange it. If the filter stone is obviously damaged, you must replace it with a new one. When this is the case is also depending on the operating conditions. This is to be identified during operation.
PSG Basic - Filter Unit Maintenance
Filter Unit Dismounting
Step Action
1 Loosen the flange clamp (1) with the help of the wing nut and remove it
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 41 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Operation & Maintenance Step Action
2 Pull the ring to remove the complete filter unit (2) using the stainless-steel ring (4) from the housing
If the filter unit - due to heavy pollution, corrosion, etc. - cannot be pulled out, so the dual function of the clamp is used to press the filter element:
Insert the bottom edge of the clamp (1) into the conical groove
Screw the clamp (1) to the end
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 42 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Operation & Maintenance
Step Action
3 Accordingly, the filter unit (2) is slightly raised and can be pulled out using the ring (4)
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 43 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Operation & Maintenance
Filter Stone Replacement
Step Action
1 Unscrew the lower filter holder (11) with help of a web
2 Remove the filter stone
3 Replace O-ring seals (6,10)
4 Insert the new filter stone (or clean the old one)
5 Screw the lower filter holder (11)
Filter Unit Installation
Step Action
1 Replace upper O-ring seals (7, 8)
2 Insert the filter unit (2) into the housing (3)
3 Insert manually the flange clamp (1) and tighten by hand
Toxic substances may be present due to being in contact with sample gas containing acidic and / or alkaline compo-nents.
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 44 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Operation & Maintenance
PSG Plus - Filter Unit Maintenance
Filter Unit Dismounting
Never undo or tighten up the hexagon screws 5! They have been adjusted at the factory, so that the detaching disk 3 can be easily moved.
Step Action
1 Turn the T-handle 1 of the filter removal device 1–3 in a counter-clockwise direction. This pulls the filter element 8 out of the casing 10 via the detaching disk 3.
Toxic substances may be present due to being in contact with sample gas containing acidic and / or
alkaline components.
2 Turn bridge 2 until it can be pulled off the bridge holding device 9 through the elongated holes
3 Pull out the filter element 8 with bridge 2 and detaching disk 3.
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 45 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Operation & Maintenance
Filter Stone Replacement
To carry out this work, you require:
A new filter stone (with new O-rings from the accessory set)
Step Action
1 Unscrew the lower filter holder (11) with help of a web
2 Remove the filter stone
3 Replace O-ring seals (7)
4 Insert the new filter stone (or clean the old one)
5 Screw the lower filter holder (11)
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 46 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Automated Filter Cleaning – Back Purge
PSG Gas Sampling Pre-Filter PF
The gas sampling pre-filter PSG PF is installed direct in the process. This way dust precipitation happens direct in the chimney and the large additional active filter surface is increasing total filter surface. A self-cleaning effect direct in the gas path creates an automatic filter cleaning thus extending the maintenance time. For up to 10 g/m3 dust content this is an alternative to back purge which requires dry, oil-free and greaseless compressed air.
Filter Element Back Purge with Compressed Air
Purpose of Back Purging
Automated filter cleaning can be provided to maintain and/ or extend the operability of the filter unit and for very high dust contents it is mandatory. This is carried out by back purging with pulsed compressed air. The dirty particles are detached from the filter surface by the pulsed compressed air and blown into the process area. For up to 10 g/m3 dust content this is an alternative for a pre-filter or a solution to achieve longer maintenance intervals. For > 10 g/m3 a single back purge is required and for > 40 g/m3 a dual back purge is mandatory and offered with PSG Plus only. The following are considered in this chapter:
The relevant connections for the respective filter unit.
The recommended program runs for the pulsed compressed air.
Requirements for the Compressed Air
The compressed air for back purge needs to fulfil the following requirements:
Dry: with a dew point of < -20 °C
Clean: oil-free & greaseless instrument air
Pressure: 4 to 6 bar (70,000 hPaabs)
Cleaning Intervals
Select the cleaning and maintenance intervals being suitable to your individual operating conditions. Irrespective of your operational routine intervals, it is necessary to clean the filter when it becomes blocked by particles of dust, dirt or corrosion. These circumstances become especially apparent when there is a pressure drop at the sample gas outlet.
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 47 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Circuit Diagram – Gas Sampling Probes
Circuit Diagram - PSG Basic
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 48 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Circuit Diagram – Gas Sampling Probes
Circuit Diagram – PSG Basic with Thermostat Control
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 49 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Circuit Diagram – Gas Sampling Probes
Circuit Diagram - PSG Plus
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 50 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Circuit Diagram – Gas Sampling Probes
Circuit Diagram - PSG Plus with Thermostat Control
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 51 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Troubleshooting To locate and eliminate the cause of a temperature failure, proceed as follows:
Stop the measuring gas transport.
Check whether a gas sampling probe overtemperature condition has been detected. If that is the case, first
o Press the RESET key of the temperature regulator at the gas sampling probe (connection box)
Check electrical supply!
Allow extra time for the system to heat up.
Check all fuses and all electrical contacts of the gas sampling probe, including those of the temperature sensor and temperature regulator, along with the relay contact.
Check the heater elements.
Check all of the mentioned components items for proper operation.
Check all gas lines including filter element for presence of condensate, and clean or replace as required. After elimination of any failure and after gas sampling probe has reached its temperature, the measurement can go on.
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 52 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Return, Disposal & Lifetime – CE & ISO Certificate
Return of Material
Danger of burns!
The work described in this chapter requires specialist knowledge. As a result, it may only be carried out by persons who are qualified and specially trained. The metal parts may have high temperatures as a result of operation.
Electric shock! You must disconnect the voltage for the current circuit, in which the electrical equipment is integrated, when putting the device out of service.
Otherwise there is a danger of electric shock.
Health hazard! Dependent on the substances which it comes into contact with during operation, the filter element could be contaminated with toxic or corrosive substances. Always wear suitable
protective clothing for cleaning work.
Procedure for Return of Equipment
Step Action
1 Clean the filter (see Chapter 5 "OPERATION & MAINTENANCE", see section 5.2.) before return of the equip-ment.
2 Remove the gas sampling system from the assembly flange of the bushing tube and hermetically seal the as-sembly flange with a blind flange.
3 Before switching off the heater by disconnecting the power supply to the gas sampling system: Take care for sufficient purge (at minimum 10 minutes) of the probe to remove any toxic or aggressive gas components - especially to avoid condensation and for protection of personnel.
4 Hermetically seal the connections of the gas sampling system.
If you want to send a device for maintenance or repair to AGT-PSG GmbH & Co. KG, please follow the instructions in the forms - Return Forms and Declaration of Contamination. Please fill in these forms completely. Send the device along with the form DDP (Delivery Duty Paid – AGT-PSG will ac-cept deliveries only if freight and customs to our company are fully paid - see address).
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 53 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Return, Disposal & Lifetime – CE & ISO Certificate
Proper Disposal
Depending on the service conditions, the filter stone could be contaminated with dangerous substances. Depending on the contamination, the filter stone must be disposed in accordance with good professional practice and the currently applicable local regulations.
End of Lifetime
If the product has reached the end of lifetime, proceed as follows: Secure a return authorization from AGT-PSG Sales or from your Representative before returning any equipment. Equip-
ment must be returned with complete identification in accordance with AGT-PSG instructions. Otherwise we cannot accept it. In no event AGT-PSG will be responsible for equipment without proper authorization and identification.
Packing for return
In case of a proper authorization:
Original Packing not Available
If the original packing material is no longer available, wrap the device in bubble foil or corrugated cardboard. When shipping overseas, also heat-seal the device air-tight in 0.2 mm thick polyethylene, including a desiccant (e.g. silica gel). The amount of desiccant used should be adequate for the package volume and the probable shipping time (at least 3 months).
Packing in an Adequate Box
Pack the device in an adequately large box lined with shock absorbent material (e.g. foam material). The thickness of the cushioning material should be adequate for the weight of the device and the mode of shipping. The box should also be lined with a double layer of bitumen paper for overseas shipping.
Marking of the Box
Mark the box "Fragile! Handle with care!“
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 54 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Return, Disposal & Lifetime – CE & ISO Certificate
Packaging and Shipping
If you want to send a device for maintenance or repair to AGT-PSG, please note the following information on packaging and shipping:
Please use the original packaging of the device. If this is not available, use another stable packaging in which the device is well protected from the weather and mechanical damage. Handle with care. Heavy blows could cause irreparable damage.
AGT-PSG will only accept deliveries that are not physically damaged. Ensure a transport in perfect upright posi-tion. Turning it upside down some parts could be irreparably damaged. It is best to tie down the individual package on a pallet. Do not stack. Mark your package with appropriate icons:
Fill in this form and the subsequent declaration of decontamination completely. Send the device along with the forms DPP (Delivery Duty Paid – AGT-PSG will only accept deliveries if freight and customs are fully paid) to: AGT-PSG GmbH & Co.KG Weißkirchener Straße 3 D-61449 Steinbach/Ts. Germany
Inform AGT-PSG by email ([email protected]) about the shipment of the cooler and the approximate delivery date. Please attach scans of the filled in documents.
Sender-/Return-Delivery-Adress and Contact Partner
Sender: Return-delivery to:
____________________________________ ____________________________________ ____________________________________ ____________________________________ ____________________________________ ____________________________________ ____________________________________ ____________________________________
same address as sender
Contact (name, email, phone):
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 55 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
Return, Disposal & Lifetime – CE & ISO Certificate
Declaration of Contamination Status
Legal regulations prescribe to fill in and sign the declaration of contamination status and send it back. This information is used to protect our employees. Please attach the declaration to the packing. Otherwise your repair cannot be pro-cessed.
Device / Articlenumber / Serialnumber:
Reason for return:
I hereby declare that the device specified above has been properly cleaned an decontaminated and that there are no
risks present when dealing with this device. In other cases, please describe the hazards in detail:
Explosives Acute toxidity Flammable Oxidizing
Envoronmental hazard Corrosive Irritant toxidity Health hazard
Liquid Solid Powdery Gaseous
Please include the current material safety data sheet of the hazardous material!
Company, Contact Location, Date, Stamp, Signature
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 56 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021
AGT-PSG EC Declaration of Conformity – CE Marking
Instruction Manual Gas Sampling Probes
AGT-PSG GmbH & Co.KG Errors and omissions expected
Page 57 Original – Instruction Manual Gas Sampling Probes Rev. 3/2021