72
Dulcodes W UV System Assembly and operating instructions A0767 Please stick the nameplate here Original operating instructions Part no.985734 BA DS 035 12/12 EN Please carefully read these operating instructions before use! · Do not discard! The operator shall be liable for any damage caused by installation or operating errors! Technical changes reserved.

ProMinent UV Ingles

Embed Size (px)

Citation preview

  • Dulcodes WUV System

    Assembly and operating instructions

    A0767

    Please stick the nameplate here

    Original operating instructionsPart no.985734 BA DS 035 12/12 EN

    Please carefully read these operating instructions before use! Do not discard!The operator shall be liable for any damage caused by installation or operating errors!

    Technical changes reserved.

  • In order to make it easier to read, this document uses the maleform in grammatical structures but with an implied neutral sense. Itis aimed equally at both men and women. We kindly ask femalereaders for their understanding in this simplification of the text.

    Read the following supplementary information in its entirety!The following are highlighted separately in the document:n Enumerated lists

    Instructions Results of the instructions

    Information

    This provides important information relating to the correct operation of the system or is intended to makeyour work easier.

    Safety informationSafety information are provided with detailed descriptions of theendangering situation, see Chapter 2.1 Explanation of thesafety information on page 6

    General non-discriminatory approach

    Supplementary information

    Supplemental instructions

    2

  • Table of contents1 About this system.................................................................. 5

    1.1 Correct and Proper Use ............................................... 52 Safety chapter....................................................................... 6

    2.1 Explanation of the safety information............................ 62.2 Users' qualifications...................................................... 72.3 Dulcodes safety information......................................... 82.4 Safety Equipment ...................................................... 102.5 Information in the Event of an Emergency ................. 10

    3 Function.............................................................................. 113.1 Commissioning........................................................... 113.2 Normal mode.............................................................. 123.3 Auto rinse interval....................................................... 123.4 Switching off............................................................... 12

    4 Control................................................................................ 134.1 Display ....................................................................... 134.2 Operating status display and parameter settings....... 154.3 Trend display.............................................................. 164.4 Change Access Code ................................................ 174.5 Setting the Language ................................................. 174.6 Ballast bus activate / deactivate................................. 174.7 Set lamp current (not for Dulcodes 75 W).................. 184.8 Sensor signal display.................................................. 194.9 Calibrating the sensor................................................. 194.10 Adjusting the display range of the trend display....... 204.11 Setting the safety threshold...................................... 204.12 Setting the warning threshold................................... 214.13 Analog output sensor signal: assigning the standard

    signal........................................................................ 214.14 Activating the pump control...................................... 224.15 Setting the start rinse duration.................................. 224.16 Set maximum free rinse duration.............................. 234.17 Setting the lamp postburning duration...................... 234.18 Adjustment of off-time and auto flushing interval...... 244.19 Set minimum mains voltage...................................... 244.20 Pause function.......................................................... 254.21 Displaying/Resetting the counter.............................. 254.22 Behaviour of the system in the event of a fault......... 254.23 Alarm signal relay..................................................... 274.24 Fault switch input...................................................... 27

    5 Mounting and installation.................................................... 285.1 Radiation chamber...................................................... 295.1.1 Assembly................................................................. 295.1.2 Attach the warning label.......................................... 305.1.3 Connectors, hydraulic.............................................. 305.2 Control cabinet and control......................................... 315.2.1 Assembly................................................................. 315.2.2 Connectors, electric................................................. 315.2.3 Opening the control................................................. 325.3 Fitting the lamp protection tubes................................. 34

    Table of contents

    3

  • 5.4 Assembly and connection of the UV lamp.................. 356 Commissioning................................................................... 37

    6.1 Leak testing and ventilation of the radiation chamber 376.2 Switching on the UV system....................................... 376.3 Calibrating the UV sensor........................................... 38

    7 Maintenance....................................................................... 397.1 Cleaning the UV lamp protection tube........................ 397.1.1 Cleaning After Dismantling the UV Lamp Protection

    Tube......................................................................... 407.1.2 Cleaning with a Cleaning Solution .......................... 427.1.3 Cleaning the UV Sensor.......................................... 437.2 Changing the UV lamps.............................................. 447.3 Calibrating the UV sensor........................................... 457.4 Replacement of the filter mats on the fan and the air

    outlet filter................................................................... 468 Troubleshooting.................................................................. 479 Technical data..................................................................... 53

    9.1 Performance data....................................................... 539.2 Dimensions sheet....................................................... 569.2.1 Dimension sheet Dulcodes W 75 ............................ 569.2.2 Dimension sheet Dulcodes W 80 and 130............... 579.2.3 Dimension sheet Dulcodes W 230 .......................... 589.3 Electrical data............................................................. 61

    10 Appendix............................................................................. 6310.1 Spare parts drawings................................................ 6310.2 Spare Parts List........................................................ 6310.3 Dulcodes wiring diagram ......................................... 6510.4 EC Declaration of Conformity................................... 6610.5 Dulcodes UV system unit operating log.................... 67

    11 Index................................................................................... 68

    Table of contents

    4

  • 1 About this system The Dulcodes UV systems are used for the photochemical treatment and disinfection of:n Drinking watern Process watern Waste watern Swimming pool waterThe water to be disinfected is subjected to short-wave UV radiationduring the UV disinfection process. This so-called UV radiationquickly and reliably kills any germs.The Dulcodes UV systems are supplied fully wired. They are available in different versions, which are defined by their identity code.The performance data can be found in the data sheet enclosedwith the Dulcodes UV system.Scope of supplyn Radiation chambern Lamp with lamp protection tuben UV sensorn Control cabinet with controln Documentation

    1.1 Correct and Proper Use Correct and Proper UseThe operator is liable for damage caused by installation and operating errors! The system is intended solely for the treatment ofwater. The system may only be used in accordance withthe technical data and specifications outlined in theoperating instructions! Any other use or modification of the system is prohibited. The system may only be operated by trained andauthorised personnel! It is imperative that the information in the operatinginstructions relating to the different phases of theunit's service life is observed!

    Application

    About this system

    5

  • 2 Safety chapter 2.1 Explanation of the safety information

    These operating instructions provide information on the technicaldata and functions of the product. These operating instructions provide detailed safety information and are provided as clear step-by-step instructions.The safety information and notes are categorised according to thefollowing scheme. A number of different symbols are used todenote different situations. The symbols shown here serve only asexamples.

    DANGER!Nature and source of the dangerConsequence: Fatal or very serious injuries.Measure to be taken to avoid this dangerDanger! Denotes an immediate threatening danger. If this is

    disregarded, it will result in fatal or very seriousinjuries.

    WARNING!Nature and source of the dangerPossible consequence: Fatal or very serious injuries.Measure to be taken to avoid this dangerWarning! Denotes a possibly hazardous situation. If this is

    disregarded, it could result in fatal or very seriousinjuries.

    CAUTION!Nature and source of the dangerPossible consequence: Slight or minor injuries, material damage.Measure to be taken to avoid this dangerCaution! Denotes a possibly hazardous situation. If this is

    disregarded, it could result in slight or minor injuries. May also be used as a warning about materialdamage.

    Introduction

    Safety chapter

    6

  • NOTICE!Nature and source of the dangerDamage to the product or its surroundingsMeasure to be taken to avoid this dangerNote! Denotes a possibly damaging situation. If this is

    disregarded, the product or an object in its vicinitycould be damaged.

    Type of informationHints on use and additional informationSource of the information, additional measuresInformation! Denotes hints on use and other useful information.It does not indicate a hazardous or damaging situation.

    2.2 Users' qualifications WARNING!Danger of injury with inadequately qualified personnel!The operator of the plant / device is responsible forensuring that the qualifications are fulfilled.If inadequately qualified personnel work on the unit orloiter in the hazard zone of the unit, this could result indangers that could cause serious injuries and materialdamage. All work on the unit should therefore only be con

    ducted by qualified personnel. Unqualified personnel should be kept away from

    the hazard zone

    Training DefinitionInstructed personnel An instructed person is deemed to be a person who has been instructed and,

    if required, trained in the tasks assigned to him/her and possible dangers thatcould result from improper behaviour, as well as having been instructed in therequired protective equipment and protective measures.

    Trained user A trained user is a person who fulfils the requirements made of an instructedperson and who has also received additional training specific to the systemfrom ProMinent or another authorised distribution partner.

    Trained qualified personnel

    A qualified employee is deemed to be a person who is able to assess thetasks assigned to him and recognize possible hazards based on his/hertraining, knowledge and experience, as well as knowledge of pertinent regulations. The assessment of a person's technical training can also be based onseveral years of work in the relevant field.

    Safety chapter

    7

  • Training DefinitionElectrician Electricians are deemed to be people, who are able to complete work on elec

    trical systems and recognize and avoid possible hazards independently basedon his/her technical training and experience, as well as knowledge of pertinentstandards and regulations.Electricians should be specifically trained for the working environment inwhich the are employed and know the relevant standards and regulations.Electricians must comply with the provisions of the applicable statutory directives on accident prevention.

    Customer Service department

    Customer Service department refers to service technicians, who havereceived proven training and have been authorised by ProMinent to work onthe system.

    Note for the system operatorThe pertinent accident prevention regulations, as wellas all other generally acknowledged safety regulations,must be adhered to!

    2.3 Dulcodes safety information WARNING!UV raysPossible consequence: Serious injuriesUV radiation is harmful to the eyes and skin Only operate the UV lamp UV-C when it is fully fittedand installed Install the UV lamp into the UV system in accord

    ance with the regulations prior to commissioning

    WARNING!Live parts!Possible consequence: Fatal or very serious injuries Measure: The device must be disconnected from

    the power supply before it is opened Disconnect damaged, defective or manipulated

    devices from the power supply

    Safety chapter

    8

  • WARNING!Insufficient water treatmentPossible consequence: IllnessPlease read the technical data sheet for your UVsystem. Calibrate the UV systems correctly.Ensure that: the maximum permissible water flow rate is not

    exceeded and UV transmission does not drop below the permis

    sible level as otherwise adequate treatment cannot be

    guaranteed.

    WARNING!Drinking water disinfectionCause: Already contaminated pipes.Possible consequence: Illness.Measure: For applications with high disinfectionrequirements (e.g. drinking water disinfection) disinfectthe downstream pipework, e.g. by superchlorination,before commissioning.

    CAUTION!Depot effect in pool waterCause: Pool water disinfected with UV radiation has noprotection against contamination.Possible consequence: IllnessMeasure: When treating pool water with UV radiation,disinfection with a depot effect e.g. chlorine is necessary.

    CAUTION!Overheating of lamp and treatment chamberPossible consequence: material damage Ensure that, with the exception of when the lamp is

    warming up, the radiation chamber has a sufficientflow of water through it so that the radiationchamber cannot overheat

    Only switch on the UV system after the radiationchamber has been filled with water

    Switch the UV system off if the flow of water isinterrupted for a long time (several hours).You can automate switch-off using the pause contact on the control. The chapter entitled "Function"describes more options for preventing overheating.

    Safety chapter

    9

  • CAUTION!Unauthorised operating parameterPossible consequence: material damageEnsure that: the installation location is dry and frost-free the protection of the UV system from chemicals,

    dyes and vapours is guaranteed the ambient temperature and the radiation tem

    perature in the direct vicinity of the system may notexceed 40 C

    the maximum permissible operating pressure is notexceeded and

    there are no solid particles and no turbidity in thewater to be treated.

    if necessary, fit an appropriate treatment systemupstream of the UV system.

    2.4 Safety Equipment ATTENTION: Hazardous ultraviolet radiation

    UV-C UV-C radiation is harmful to the eyes and skin. The lamps mayonly be operated when installed. The system should be installed inaccordance with all pertinent regulations prior to commissioningthe lampsATTENTION: Danger

    Disconnect the system from the mains power supply or switchoff the main switch prior to commencing maintenance work on thesystem. Depressurise the radiation chamber prior to commencingmaintenance work.

    Disconnect the system from the mains power supply or switchoff the main switch prior to opening the cabinet.

    2.5 Information in the Event of an Emergency In the event of an emergency, switch the red-yellow main switch onthe side of the control cabinet to OFF or disconnect from the mainspower supply.

    Labels on radiation chamber

    Labels on Control cabinet

    Safety chapter

    10

  • 3 Function Shut-off valves and flushing valves do not form part ofthe scope of supplyThe shut-off valves and flushing valves do not formpart of the scope of supply of the UV system and aremade available by the site operator.

    A0917

    1

    23

    4 5 67

    Fig. 1: Typical installation schematic of a UV system1 Shut-off valve (site-supplied)2 Flushing valve (site-supplied)3 Sampling tap (site-supplied), fireproof4 Shut-off valve (site-supplied)

    5 Stainless steel radiation chamber6 UV sensor7 Control

    The water to be treated flows through the stainless steel radiationchamber and past the UV lamps. The UV radiation kills the germsand reduces substances, such as chloramines, in the swimmingpool water.The low pressure UV lamps generate a very high output level ofUV radiation, which is particularly effective for purposes of disinfection at a wavelength of 254 nm. The UV lamps are located in lampprotection tubes made of high-grade quartz with a high level of UVtransparency. The compact design of the radiation chamber andthe optimum flow of radiation as well as the integrated turbulatorsresults in evenly-distributed irradiation of the entire flow of water.A control monitors the UV system along with a UV sensor.

    3.1 Commissioning Once the Dulcodes UV-system has been switched on, the UVlamps are ignited. For UV systems on a data bus to ballasts, thebus is first activated prior to ignition. This can take several seconds, depending on the size of the UV system. Following ignition,the UV lamps need several minutes until they reach operating temperature.The UV sensor monitors the UV lamps. As soon as the UV outputhas exceeded the minimum irradiance threshold, the rinse valveopens for the start rinse.If the minimum irradiance is not exceeded within the maximum permissible warm-up time, then the rinse valve will open regardless.However, If the minimum irradiance is also not exceeded within themaximum rinse duration, then the controller switches the UV-system off and goes into fault mode.The shut-off valve is opened after the start rinse is completed. TheUV-system goes into normal mode.

    Function

    11

  • 3.2 Normal mode In normal mode, the UV-C sensor continues to monitor the UVoutput:If the UV output falls below the warning threshold: a warning isemitted.If the UV output falls below the minimum irradiance threshold: Theshut-off valve closes and the rinse valve opens. However, If theminimum irradiance threshold is also not exceeded within the maximum rinse duration, then the controller switches the UV-system offand goes into fault mode.All UV lamps are monitored to ensure that they operate optimally. Ifa UV lamp fails, the shut-off valve is closed and the controllerswitches the UV-system off and goes into fault mode.

    3.3 Auto rinse interval If auto rinse is active, the rinse valve opens for the auto rinseprocess after the maximum off-time has been reached.

    3.4 Switching off When the UV system is switched off, the shut-off valve closes andthe UV lamps are turned off. If the UV lamps require postburning,then the lamps are switched off after the lamp postburning durationhas elapsed.

    Function

    12

  • 4 Control NOTICE!With the exception of sensor calibration lamp currentcalibration, modifications to the settings should only beundertaken when the UV system is switched off.

    As the electronics and software are always subject to improvements, the version number is used as a means of identification.This should be stated with complaints. It can be called up on thedisplay.

    The Dulcodes UV systems' controls are factory-preset. For manyapplications it is therefore not necessary to change the settings.

    4.1 Display The system is provided with a graphical LCD display.

    NOTICE!START/STOP buttonHold down the STOPSTART key for at least 2 seconds.The display returns to the normal display for therespective operating mode 5 minutes after the key hasbeen pressed for the last time.

    In operating moden Display of the operating moden Warnings are indicated by flashing arrows and displaysn Faults are displayed by a flashing fault alertIn programming moden Flashing display of the numerical values and inputs that can be

    changed.

    Version

    Default settings

    Control

    13

  • STOPSTART

    2

    3

    4

    5

    8

    6

    7

    1

    XXXXXX

    Fig. 2: Display and operating unit

    Position number Key Function1 Housing2 LCD display5 UP key

    In programming mode: Raises the displayed numerical value orchanges an input

    7 BACK keyMoves back one level in the menu

    6 DOWN keyIn programming mode: Lowers the displayed numerical value orchanges an input

    8 CHANGE keyIn operating mode: Changes the display windowIn programming mode: Changes adjustable parameters

    3STOPSTART

    START/STOP buttonSwitches on and off the UV-system

    4 ENTER keyIn operating mode: Changes to programming mode or acknowledges afaultIn programming mode: Applies a set value or mode

    Control

    14

  • 4.2 Operating status display and parameter settings

    Behaviour at faults? Plant off

    A

    BA

    xxxx h Operationxxxx - time On

    Version

    Lamp current(min. x.x A)(max. x.x A)

    Bus activeBus passive

    Cal. factor =Time x:xx m

    Safety threshold

    xxx W/m 2Warning threshold

    Lamp hoursLamp turn-ons

    Commissioning Rinsing time

    Maximum off-time

    Intervall- Rinsing time

    Minimum line voltage

    Pause under pause contact

    Operat. hoursTurn-ons

    Display of sensor signal in : x

    Language

    Standard display

    Change of displayPump control

    Of f

    Lamp postburnTime xx:xx

    Maximum free rinsing

    Trend display

    FrAccess code change?

    Bus xxxx

    Analogue output

    Windowr-sensor calibration (only appears when system running)

    Display ranger Trend displayx TDay(s)

    with "pump control" = "on"

    Behavior at Faults = emerg.service

    A0748

    with "pump control" = "on"

    with "max. off-time" = "0"

    Fig. 3: Operating status display (with the UV system running)A + Back to Trend displayB = Change to Programming modeC = Change to "Change Access Code" mode

    Control

    15

  • (adjustable numbers/ settings flash) Memory display 1s B (Request access code)

    Changes are not saved Changes are savedBack to display window

    Access code incorrect

    P_PMA_DS_0016_SW

    Fig. 4: Programming manual

    NOTICE!Access codeOnce the access code has been correctly enteredonce, it is not necessary to re-enter the code for furtherprogramming processes; the flashing numbers or settings will appear directly as soon as the -key ispressed. The access code is automatically cancelled 5minutes after a key was last pressed or after a returnto the Trend or Standard display.

    4.3 Trend display NOTICE!Display calibration Each calibration of the UV sensor is documented

    by a vertical continuous line in the trend display The content of the trend display is deleted when

    the display range changes and when the operatinghour counter is reset

    1 UV sensor signal2 Warning threshold3 Safety threshold4 Switch on's/off's5 CalibrationThe trend display is used to monitor the ageing of the UV-lamps,the formation of a film coating on the lamp protection tubes orchanges to the water quality.The progress of the UV sensor signal is shown in a time frame.Horizontal lines show the safety threshold and the warningthreshold respectively. The short vertical lines show when the UVsystem is switched on. The display range of the UV sensor signallies between 0 W/m2 (or 0%) and the value which has beenassigned to the analogue output value of 20 mA. The time framecan be adjusted and guarantees an ongoing display: Once theselected time has expired, the oldest value is deleted and the newvalue is displayed.Default settingn Time frame: 100 daysn Maximum value of the UV sensor signal: 400 W/m2

    Programming manual

    Fig. 5: Trend display

    Control

    16

  • 4.4 Change Access Code CAUTION!Change access codeDefault setting "5000" Note down the access code!

    Parameters can only be set once the correctaccess code has been entered

    The default access code does not guarantee anyprotection from unauthorised changes

    New Access codexxxx

    Memory display 1 s(Request Access code)

    Changes are saved

    Confirm Access codexxxx

    Access code incorrect Changes are not saved

    C

    Back to Change Access Code request P_PMA_DS_0017_SW

    Fig. 6: Change Access CodeTo protect against unauthorised changes of the settings, thesystem controller has an access code for the programming mode.It can be freely selected by the operator. The programming modeis still disabled after a change of the access code. It is only enabled when the new access code is entered.

    4.5 Setting the Language It is possible to select between different languages.

    4.6 Ballast bus activate / deactivate For UV systems equipped with ballasts via a bus interface, switch the bus to active . The UV lamps cannotbe ignited if the bus is passive.The ballast of UV systems of type Dulcodes 75W doesnot have a bus interface. With these system types thebus must be set to passive .

    Language :

    German

    Fig. 7: Setting the Language

    Control

    17

  • The ballasts for the high-flux UV lamps (e.g. 80-W lamp, 130-Wlamp, 230-W lamp) are equipped with a bus interface. This businterface enables the UV lamps to be ignited and switched off. Inaddition, they are monitored for optimal operation. Moreover, thelamp current can be changed. When the bus is active and once theUV system is switched on, the number of ballasts is shown. Keys[UP] and [DOWN] can be used to call up the serial numbers of theindividual ballasts.

    For UV systems equipped with a bus interface ballast, active , otherwise passive (Dulcodes 75 W).

    4.7 Set lamp current (not for Dulcodes 75 W) CAUTION!Deterioration of UV output / premature wearCause: Incorrectly set lamp current.Consequence: Poor disinfection performance of theUV system. Any operation of the UV lamp with currents outside of the permissible range can cause theUV system to fail prematurely. Excessive lamp currentcauses the UV lamps to overheat, causing the UVoutput to deteriorate. Insufficient lamp current leads toa significant reduction in the UV output.Measure: Set the lamp current to a reliable valve.Please refer to the enclosed data sheet for the permissible range for lamp current and rated current.

    The UV lamps should be operated with the default current, a deviation only makes sense and is only permitted in a few exceptional cases. Please refer to theenclosed data sheet for the permissible range for lampcurrent and rated current. For UV systems, which arealso suitable for hot water disinfection, two default currents are specified in the data sheet; for hot and coldapplications.

    Bus passive

    Fig. 8: Bus is passive

    Bus active 10 lamps

    EVG codeEVG#001

    Fig. 9: Bus is activeDefault setting

    Control

    18

  • The lamp current can be freely selected within a certain range forballasts which are equipped with a bus interface. This means thatthe UV lamps can be adapted to specific operational conditions.The lamp current can also be adjusted in operating mode andflushing mode. The permissible ranges for the lamp current are notmonitored when the UV system is switched off. If a lamp current isset outside of the permissible ranges, then the Lamp current faultindication is shown once the UV system is switched on and the bushas initialised.

    In the event that the UV-lamps indicate a slight reduction in UVoutput for water temperatures < 8 C, then the lamp current can beincreased slightly by 0.1 to 0.3 A above the rated current for UVoutput.

    The lamp current can be increased by 0.2 to 0.4 A above the ratedcurrent for UV output towards the end of the lamp service life inorder to increase the UV output slightly.

    2 ampere

    4.8 Sensor signal display The UV sensor monitors the UV lamps.Reduction in the sensor signal can be caused by the following:n Coating formed on the lamp protection tubesn Significant deterioration of the UV transmission in the watern Reduction of the UV output of the UV lamps due to lamp

    ageing.It is possible to select between an absolute value in W/m2 and arelative display of the sensor signal in % .

    W/m2

    4.9 Calibrating the sensor The UV sensor is pre-calibrated at the factory and does not have tobe calibrated again.

    Lamp current = 2 A

    A0769

    Fig. 10: Set lamp current (not for Dulcodes 75 W)Water temperatures < 8 C

    End of service life

    Default setting:

    Display of sensor signal

    A0750

    Fig. 11: Sensor signal displayDefault setting

    Time

    Cal.factor =

    Fig. 12: Calibrating the sensor

    Control

    19

  • 4.10 Adjusting the display range of the trend display The recording time of the sensor signal for the trend display can beadjusted. The value (in days) is interpreted as the time frame andthus guarantees a continuous display: Once the selected time hasexpired, the oldest value is deleted and the new value is displayed.

    100 days

    4.11 Setting the safety threshold WARNING!UV lamp replacementPossible consequence: Illness Check and possibly reset the safety and warning

    threshold when the UV lamp is replaced! Only a correctly set safety threshold will guar

    antee adequate UV radiation

    Reliable and safe water treatment can no longer be guaranteed ifthe UV output falls so low that the UV sensor signal falls below thesafety threshold. In this case a shut-off valve, should one be fitted,will close. When the signal falls below the safety threshold on thedisplay, this is shown by two flashing arrows.A signal device can be connected to the [SAFETY THRESHOLD]signal relay of the control. The relay is closed when the signal fallsbelow the safety threshold.

    NOTICE!The safety threshold must be below the warningthreshold. It is not possible to set it above the warningthreshold.

    1. Switch on the UV system with the STOPSTART button2. Wait until the UV lamp has reached its full capacity, i.e. until

    the UV sensor signal is stable3. Read the UV intensity displayed and note it down4. Switch off the UV system with the STOPSTART button5. Set the safety threshold to 50 % of the displayed UV intensity

    Example: Measured UV intensity: 100 W/m2therefore follows:safety threshold = 100 W/m2 0.50 = 50.0 W/m2

    6. Now set the warning threshold

    Display range Trend display100 day(s)

    Fig. 13: Adjusting the display range ofthe trend displayDefault setting:

    Safetythreshhold

    A0770

    Fig. 14: Setting the safety threshold

    Setting the safety threshold

    Control

    20

  • 4.12 Setting the warning threshold WARNING!Poor disinfectionCause: Insufficient UV system radiation power.Consequence: Poor disinfection performance of theUV system.Check and possibly reset the safety and warningthresholds when lamps are replaced.

    The warning threshold must be above the safetythreshold.It is not possible to set it below the safety threshold.

    The system issues a warning should the UV output drop so far thatthe sensor signal falls below the warning threshold. To prevent thesignal falling below the safety threshold, the UV lamp protectiontubes should be cleaned, the UV lamps should be replaced or thewater quality should be improved by means of appropriate watertreatment. When the signal falls below the warning threshold, thisis indicated on the display by a flashing arrow.A signal device can be connected to the[WARNING THRESHOLD] signal relay of the control. The relay isclosed when the signal falls below the warning threshold.

    Requirements:n Stable UV intensity for the UV lamps has been notedn The safety threshold has been set

    Set the warning threshold to 110 % of the set safetythreshold

    Example: Set safety threshold: 50 W/m2 therefore follows:warning threshold = 50 W/m2 1.10 = 55.0 W/m2

    The safety and warning thresholds have been set, the UVsystem is now ready for operation. Switch on the UVsystem with the STOPSTART button.

    4.13 Analog output sensor signal: assigning the standard signal The sensor signal assigned to the 20 mA is simultaneously the maximum value of the trend display. Adjustthis sensor signal value to 125 % of the maximumvalue so that the trend display can never overflow .

    The signal from the UV sensor can also be recorded for documentation purposes using a recorder. To do so, connect the recorder tothe standard output of the control.

    Warning thresholdA0771

    Fig. 15: Setting the warning threshold

    Analogue output 0 W/m2 = 0 mA400 W/m2 = 20 mA

    A0751

    Fig. 16: Analog output sensor signal

    Control

    21

  • It is possible to choose from a 0 to 20 mA and 4 to 20 mA standardsignal:n 0 or 4 mA corresponds to the sensor signal 0 W/m2n 20 mA can be assigned to any value

    0 W/m2 = 0 mA400 W/m2 = 20 mA (dependent on the settings)

    4.14 Activating the pump control CAUTION!Damage to the UV lamps and radiation chamberCause: The UV-system may only be operated whenthe radiation chamber is fully filled with water. With anempty or only partially filled radiation chamber there isa risk of damaging the UV lamps and radiationchamber.Measure: It must therefore be ensured that the radiation chamber cannot run empty when the pump isswitched off.

    As the feed pump generally has a certain post-rinseduration, the lamp postburning duration should be setto at least one minute. This ensures that water conveyed during the post-rinse period is disinfected.

    The pump control system must be activated in order to control thefeed pump with the pump relay.The pump relay releases when the UV system is switched off andremains released with pump control off , even when the UVsystem is running.If the UV system is switched on with Pump control on , then theUV system will exit warning mode only when the warning thresholdis exceeded. The pump relay operates. If the UV system isswitched off or the UV system switches over to pause mode, thepump relay will release.If the minimum irradiance is undershot during operation or a UVlamp fails, then the pump relay releases and the UV systemswitches over to fault mode.

    Off

    4.15 Setting the start rinse duration In most cases, a start rinse duration of 1 min is sufficient. If there is no start rinse duration specified, thenthe start rinse duration will be set to 0 min .

    Default setting

    Pump control

    OnFig. 17: Activating the pump control

    Default setting:

    Control

    22

  • The start rinse duration is intended to ensure that only optimallytreated water can flow through to the consumer. As soon as the UVlamps reach operating temperature after ignition and the sensorsignal has exceeded the minimum irradiance level, the automaticrinse valve opens (if fitted) for the start rinse process. Only thendoes the shut-off valve open.

    1 min

    4.16 Set maximum free rinse duration The free rinsing process is primarily used for the disinfection ofdrinking water. In doing so, a maximum free rinse duration of over10 hours is often used. After periods of heavy rainfall it is possiblethat the UV transmission will decrease, for example, if the waterhas poor natural filtering characteristics. The sensor signal undershoots the minimum irradiance value. The shut-off valve closesand the rinse valve opens.Once the water quality has improved, the UV system switchesback over to normal mode. However, If the sensor signal fails toexceed the minimum irradiance threshold within the maximum freerinse duration, then the UV system switches over to fault mode.

    1 min

    4.17 Setting the lamp postburning duration In large-scale UV systems for drinking water disinfection, it cantake some time for the shut-off valve to close in some cases, or forthe water flow to be stopped by another means. In this case, thelamp postburning system prevents insufficiently disinfected waterfrom reaching the consumer while the UV system is being shutdown. In most cases, a start lamp postburning duration of 1 minuteis sufficient.

    Off

    Start rinse duration

    1:00 min:secFig. 18: Setting the start rinse durationDefault setting:

    Maximumfree rinse duration

    00:01 h:minFig. 19: Set maximum free rinse duration

    Default setting:

    Lamp postburning Off

    Fig. 20: Setting the lamp postburningdurationDefault setting:

    Control

    23

  • 4.18 Adjustment of off-time and auto flushing interval The off-time flushing system is used primarily for internal drinkingwater supply systems. One must always assume that there will beprolonged periods without water removal in UV systems withinternal drinking water supplies. In such cases, it is advantageousto make use of the off-time flushing system in order to avoid impermissible warming and irradiance of the water.

    A flow detector or temperature switch is connected to the switchinput [FLOW TEMPERATURE] of the control. The contact closesupon exceeding of a minimum flow or the maximum temperature.The flushing valve is opened for the flushing interval, provided thatno water has been drawn during the maximum off-time.However, if the [FLOW TEMPERATURE] switch input on the control does not have a flow detector or temperature switch connectedto it, in other words, the switch input is open, then the flushingvalve opens after the maximum off-time for the auto flushinginterval (periodic flushing).In most cases, the maximum off-time is set to 5 hours. An autoflushing interval of 1 minute is generally sufficient.If the maximum off-time is set to 00:00, then no auto rinse processis undertaken.

    00:00 min1 min

    4.19 Set minimum mains voltage Modifications to the minimum mains voltage may onlybe carried out in agreement with the manufacturer.

    Monitoring the mains voltage prevents uncontrolled failure of theUV system and the UV lamps due to insufficient mains voltage.If the mains voltage drops to the minimum value, then the controlswitches over to undervoltage mode and the UV system is shutdown.When the mains voltage exceeds the permissible minimum valueagain, then the UV system starts up again automatically.

    180 V

    Maximum off-time 00:00 h:min

    Fig. 21: Setting the off-time

    Auto rinse interval 00:00 min:sec

    Fig. 22: Setting the auto flushinginterval

    Default setting:

    Minimum line voltage 180 V

    Fig. 23: Set minimum mains voltageDefault setting:

    Control

    24

  • 4.20 Pause function The UV-system can be switched on and off by opening and shutting an external contact that is connected to the Pause input of thecontroller.It is possible to select whether the UV system starts up with anopen or closed Pause contact.

    Pause with pause contact closed.(UV-system starts up when the pause contact is open).

    4.21 Displaying/Resetting the counter The [OPERATING HOURS] and [SWITCH ONS] counters cannotbe reset.

    The [LAMP HOURS] and [LAMP SWITCH ONS] can be reset.

    4.22 Behaviour of the system in the event of a fault Normally the UV system is switched off in the event of a fault. However, for certain applications it can be a good idea to continueoperating the system at reduced capacity (emerg.service).

    Pause on Pause contact closed

    Fig. 24: Pause functionDefault setting:

    Operating hours 400 hTurn-ons 25

    Fig. 25: OPERATING HOURS /SWITCH ONS

    Lamp hours 400 hlamp turn-ons 25

    Fig. 26: LAMP HOURS / LAMPSWITCH ONS

    Behaviour at Fault?Plant off

    A0752

    Fig. 27: Behaviour of the system inthe event of a fault

    Control

    25

  • CAUTION!Reduced disinfection efficiency The disinfection efficiency of the UV system is

    severely reduced in emerg.service mode Emerg.service mode is not permissible for applica

    tions where high demands are made of the disinfection efficiency, for example, drinking water disinfection or similar applications

    A significantly reduced system output must beexpected for applications where emerg.servicemode is permissible for the UV system

    In emerg.service mode, a possibly equipped shut-off valve opens immediately once the STOPSTART button hasbeen pressed, and not after the safety thresholdhas been exceeded.The system does not monitor whether the UVlamps are warmed up or whether sufficient UVoutput is available. It may be necessary to manually interrupt the water flow for 5 to 10 minutesonce emerg.service mode has been activated(manual shut-off valve, manual deactivation of thefeed pump, etc.)

    The operator must check prior to enabling theemerg.service mode in the programming menu thatno risk is posed to personnel and property as aresult of the significantly reduced system efficiency

    ProMaqua provides no warranty and accepts noliability or claims for damages resulting from operation of the UV system in emerg.service mode.

    A special code is required for reprogramming the system's faultbehaviour which can only be obtained on request from ProMaqua.Once the special access code has been entered, it is possible toselect emerg.service mode for the system's behaviour in the eventof a fault.If emerg.service mode has been selected for fault events, then thesystem switches over to fault mode as previously in the event oflamp failure or if the safety threshold is undershot after completionof the UV system's free rinse duration. Pressing the buttoncauses the UV system to switch over to emerg.service mode. Inother words, the fault is not to be acknowledged with the button.Emerg.service mode can also be triggered by means of closinggate input flow/temperature . Bridging the input causes automaticswitch-over into emerg.service mode.The UV lamps are ignited in emerg.service mode, the shut-offvalve is opened and the pump relay is activated. However, the faultindicating relay will remain dropped and lamp failure and sensorsignals are no longer monitored. The signal relay[WARNING AND SAFETY THRESHOLD] undershot remainsdropped

    No auto rinse interval is possible for UV systemswhere emerg.service mode is enabled. The programming windows [max. off-time] as well as[Auto rinse interval] are permanently hidden.

    Control

    26

  • Due to the fact that in emerg.service mode the pump relay is stillactivated with pump control [Off], the pump relay can be used toindicate emerg.service mode on UV systems with the pump controlspecification [Off]. For UV systems with the specification pumpcontrol [On], emerg.service mode can be detected by the fact thatthe pump relay continues to be activated despite a dropped faultindicating relay.

    4.23 Alarm signal relay A signal device can be connected to the ALARM signal relay of thecontroller.The relay drops out if there is a fault/malfunction or in the event ofa power failure.

    4.24 Fault switch input CAUTION!remove bridgePossibility of faulty operationWhen a fault signalling device is connected, removethe jumper as otherwise no fault will be reported.

    NOTICE!remove bridgePossibility of faulty operationThe [FAULT] switch input is bridged when the systemis delivered. If the jumper is removed, without connecting up a fault signalling device, then the control willgo into fault mode and the UV system can then nolonger be operated.

    An external fault signalling device, such as a flow rate monitor, canbe connected to the [FAULT] switch input.

    Control

    27

  • 5 Mounting and installation n User qualification, mechanical installation: trained qualified per

    sonnel, see Chapter 2.2 Users' qualifications on page 7n User qualification, electrical installation: Electrical technician,

    see Chapter 2.2 Users' qualifications on page 7

    CAUTION!Unauthorised operating parameterPossible consequence: Material damage.Ensure that: the installation location is dry and frost-free the protection of the UV system from chemicals,

    dyes and vapours is guaranteed the ambient temperature and the radiation tem

    perature in the direct vicinity of the system may notexceed 40 C

    the maximum permissible operating pressure is notexceeded and

    there are no solid particles and no turbidity in thewater to be treated. if necessary, fit a suitable filter prior to the UV

    system.

    NOTICE!Switching on and offPossibility of increased wear to UV lampOperate the UV system in such a way that the frequentswitching on and off of the UV lamp is avoided.

    Mounting and installation

    28

  • 5.1 Radiation chamber The design of the radiation chamber can be found in the encloseddata sheet:n Vertical wall installation designn Horizontal designn Free-standing design

    A0772

    1.2. 3. 4.5.

    6. 7. 8.9. 10.

    11.

    12.13.

    14.

    15.16.

    17.

    Fig. 28: Construction of the radiation chamber1. Lamp cover2. O-ring3. Teflon ring4. Connector plug5. O-ring6. Water outlet7. Fastenings8. Radiation chamber9. Lamp protection tube

    10. UV lamp11. Water supply12. Water drain screw with O-ring13. O-ring14. UV sensor15. Vent screw with O-ring16. UV lamp protection tube holder17. Thumb nut

    5.1.1 Assembly NOTICE!Permissible installation typesIt is only permissible to install the system in accordance with the enclosed data sheet.Leave adequate room for maintenance work. Therequired room can be found on the enclosed datasheet.

    Vertical wall installationn Fasten the radiation chamber vertically to the wall with suitable

    mounting fixturesHorizontal installationn Fasten the radiation chamber horizontally to the wall or appro

    priate frame with suitable mounting fixtures.

    Mounting and installation

    29

  • Free-standing installationn Position the radiation chamber on the floor.

    5.1.2 Attach the warning label The supplied self-adhesive warning label should beattached to the radiation chamber so that it is clearlyvisible.

    5.1.3 Connectors, hydraulic WARNING!Automatic shut-off valveIf the water flow is not interrupted in the event of asystem fault by means of automatic deactivation of thefeed pump, then an automatic shut-off valve is to befitted downstream of the radiation chamber. This is tobe connected to the shut-off valve output of the control. The shut-off valve must close autonomously, inorder that the water flow is interrupted in the event ofpower failure.

    CAUTION!Overheating of lamp and treatment chamberPossible consequence: material damage Ensure that, with the exception of when the lamp is

    warming up, the radiation chamber has a sufficientflow of water through it so that the radiationchamber cannot overheat

    Only switch on the UV system after the radiationchamber has been filled with water

    Switch the UV system off if the flow of water isinterrupted for a long time (several hours). Install the flushing valve (if necessary in combi

    nation with a flow detector or temperature monitor) and set an appropriate flushing routing onthe control.

    NOTICE!Execute the hydraulic Connectors on the radiationchamber in compliance with the applicable generalguidelines and local installation regulations.Use UV-resistant materials for the hydraulic connector.If PVC is used, it is highly likely that the PVC will discolour in the area of the connection. In unfavourablecircumstances the material may become brittle andcrack.

    Mounting and installation

    30

  • Provide valves upstream and downstream of the radiation chamber to shut off the radiation chamber formaintenance work.Provide suitable fireproof valves upstream and downstream of the radiation chamber to allow for microbiological sampling.

    5.2 Control cabinet and control 5.2.1 Assembly

    The connecting cable for the lamp and the sensorcable may not be lengthened.

    The switch cabinet or the mounting panel with controller andpower supply unit should be fitted to the wall or a suitableframe in such a way that the UV lamps and the UV sensorscan be connected to the cables provided.

    5.2.2 Connectors, electric WARNING!Mains voltage on protective low voltagePossible consequence: Fatal or very serious injuries.Measure: If connecting the protective low voltage(SELV) to one of the X4 terminals, the X4 terminalsmust not be connected to mains voltage.

    NOTICE!General notes on the electrical installation The electrical installation must be carried out by an

    authorised electrical technician Ensure a continuous power supply by means of a

    suitable fault current protection switch Connect a protective earth conductor to the radia

    tion chamber.

    The connecting cable for the lamp and the sensorcable may not be lengthened.

    Mounting and installation

    31

  • 5.2.3 Opening the control WARNING!Electrical installationCause: Danger from electrical voltage.Possible consequence: Fatal or very serious injuries.Measure: Before opening the control, ensure that control is switched off (in a zero-volts state).

    Opening of the control is only necessary, if the controlis not fitted in the control cabinet.

    5.2.3.1 Opening the device

    Fig. 29: Opening the device1. Loosen the 4 captive screws (1).2. Lift the upper section of the device from the lower section (2).

    A wide flathead screwdriver may be of assistance.3. Insert the upper section with both guide rails into the lower

    section (3 and 4) (parked position)

    Mounting and installation

    32

  • 5.2.3.2 Electrical Installation when wall mounted (Dulcodes 75 W)

    NOTICE!Threaded holesUsing a suitable tool( approx. 4 mm), punch out thethreaded holes according to the number of cables. Punch aids are provided to punch out the threaded

    holes

    A0014

    Fig. 30: Punching out the threaded holes

    1. Screw connection M20 x 1.5 2. Thrust collar M203. Thrust collar M20 4. Dummy washer M20

    1. Remove cable sheathing over a sufficient length2. Fit screw connection (1), thrust collar (2) and seal (3) onto

    cable3. Insert cable and fittings into the threaded hole4. Align the cable and push in until enough cable is in the con

    trol housing5. Screw in screw connection and tighten firmly6. Shorten cable wires to the precise overall length and strip off

    approx. 8 mm insulation7. Fit cable end sleeves to the wires.8. Connect the wires to the terminals in accordance with the

    electrical terminal diagram

    Mounting and installation

    33

  • Punched out threaded holes can be resealed with the M20 dummywashers (4) provided.The M12 x 1.5 screw connections and brass lock nuts are used forthe 4 openings in the front row.

    Fig. 31: M12x1.5 screw connections

    5. Screw connection M12 x 1.5 6. Lock nut M12x1.51. Fit lock nut M12x1.5 (6) on the inside2. Fit screw connection M12x1.5 (5) from the outside and

    tighten firmly

    5.3 Fitting the lamp protection tubes Check the UV lamp protection tubes for damagebefore fitting.Damaged UV lamp protection tubes may not be fitted.

    1. Loosen the attachment screws of the UV lamp protectiontube holders using the supplied Allen key

    2. Remove the holders of the UV lamp protection tube from theradiation chamber by screwing out

    3. Carefully push the O-ring provided approx. 40 mm over theopen end of the UV lamp protection tube.

    4. Feed the UV lamp protection tube into the radiation chamber.

    Mounting and installation

    34

  • 5.Ensure that the lamp protection tube is sittingcorrectly.The lamp protection tube may not project outmore than 30 mm and may not be offset at anangle.Installing the lamp protection tube holders at anangle can cause the UV lamp protection tubes tobe damaged.

    Check whether the Teflon ring on the lamp protection tubeholder is lying in the groove provided. If it is not, insert theTeflon ring supplied into the UV lamp protection tube holderuntil it pops into the designated groove.

    6. Push the UV lamp protection tube holder over the UV lampprotection tube and screw into the radiation chamber; tightenby hand

    7. Tighten the UV lamp protection tube holder with the suppliedsickle spanner; only minimal force is required.

    5.4 Assembly and connection of the UV lamp WARNING!Electrical installationCause: Danger from electrical voltage.Possible consequence: Fatal or very serious injuries.Measure: Prior to assembly and connection of the UVlamp, switch off the master switch or pull the mainsplug out of the socket.

    WARNING!UV-C radiationCause: Danger due to UV-C UV radiation.Consequence: UV radiation is harmful to the eyes andskin.Measure: Only operate the UV lamps when they areproperly installed.

    NOTICE!Premature UV lamp failureNever touch the glass of the UV lamp with bare hands.Fingerprints burn into the glass and can result in earlyfailure. Clean off fingerprints from the lamp with a clothmoistened with alcohol before installing.

    Mounting and installation

    35

  • The UV lamp of the UV system should be installed inaccordance with all pertinent regulations prior to commissioning.Do not modify the fitted lamp connection cable withoutauthorisation.Do not modify the gap between the plug and the lampcover. Otherwise, it can not be guaranteed that the UVlamp lies against the closed end of the UV lamp protection tube. This is a prerequisite for safe disinfection.

    1. Lay the O-ring into the UV lamp protection tube holdergroove provided.

    2. Insert the UV lamp into the UV lamp protection tube andallow it to project out approx. 100 mm

    3.The connector plug can only be plugged on incertain positions, if necessary turn through 90before trying again.

    Insert the connector plug onto the UV lamp4. Insert the UV lamp fully into the UV lamp protection tube5. Secure the lamp cover on the UV lamp protection tube holder

    using the knurled nut; apply only minimal force to the knurlednut

    Mounting and installation

    36

  • 6 Commissioning n User qualification, commissioning: trained user, see

    Chapter 2.2 Users' qualifications on page 7

    WARNING!Contaminated pipework / tubingCause: The following pipework / tubing, etc. could becontaminated.Possible consequence: Serious illnesses.Measure: The pipework / tubing requires disinfectionafter the commissioning process (e.g. by superchlorination).

    6.1 Leak testing and ventilation of the radiation chamber 1. Open the air vent on the radiation chamber2. Slowly open the shut-off valve upstream of the radiation

    chamber3. Fill the radiation chamber until water emerges from the bleed

    valve4. Close the bleed valve5. Check that the radiation chamber is not leaking6. Open the shut-off valve downstream from the radiation

    chamber (only necessary with a manual shut-off valve)

    6.2 Switching on the UV system NOTICE!Only switch on the UV system after the radiationchamber has been filled with water.

    1. Switch on the main switch or connect to the mains powersupply

    2. Check the parameters in programming mode and change ifnecessary

    NOTICE!Ensure that the UV lamps are operated at the ratedcurrent.

    3. Switch on the UV system using the [START/STOP] key. Inorder to do so, the [START/STOP] key must be held down forat least 2 seconds

    4. Should the controller go into [PAUSE] mode, activate thePause contact Once the UV lamp has ignited, it will take several minutesuntil the full UV output has been reached.

    Commissioning

    37

  • 6.3 Calibrating the UV sensor Raw water with very low UV-transmission propertiesIf raw water with very low UV-transmission propertiesis used, it may occur that the safety threshold is notexceeded. In this case, you must adapt and adjust thesafety threshold and the warning threshold duringstartup of the UV system.

    All new UV lamps need a burn-in time of 100 to 200hours. For this reason, the safety threshold andwarning threshold should be checked approx. 200operating hours after commissioning.

    The UV sensor is pre-calibrated at the factory and does not have tobe calibrated again.

    Commissioning

    38

  • 7 Maintenance WARNING!Background information about maintenance The UV lamps should be replaced at the latest

    after their maximum permissible service life. Otherwise, the operating safety of the UV system cannotbe guaranteed

    The maximum permissible service life is 10,000operating hours, unless otherwise stated in theenclosed data sheet

    Before any maintenance work, disconnect thesystem from the mains power supply or switch offthe main switch

    Depressurise the radiation chamber before commencing any maintenance work

    UV radiation is harmful to the eyes and skin Only operate the UV lamp when it is correctly fitted The UV system should be installed in accordance

    with all pertinent regulations prior to commissioningthe UV lamps

    Dirty filter mats on the fan and air outlet filter canlead to the control cabinet overheating andbecoming damaged

    Carry out the following maintenance work at regular intervals:n Replace the UV lamp at the end of its useful service lifen Clean and check the lamp protection tube (normally only nec

    essary when replacing the lamp)n Replace the filter mats on the fan and the air outlet filter on the

    control cabinet (normally annually)An operating log should be kept as a record. A form is enclosed inthe appendix for this purpose.

    7.1 Cleaning the UV lamp protection tube Deposits of, for example iron, manganese or limescale, can formon the lamp protection tubes during operation. As these depositsabsorb UV radiation, they should be removed at regular intervals.

    Clean the tube at the very latest when the sensorsignal falls below the warning threshold, without thisbeing based on other causes, such as ageing of thelamp or serious deterioration of the UV transmission.

    An annual clean of the lamp protection tubes when replacing theUV lamp suffices for many UV systems. For UV systems used forwaste water disinfection, then cleaning may sometimes be necessary every one to two months. All lamp protection tubes must becleaned in systems with multiple UV lamps. The UV lamp protection tubes can be cleaned manually when dismantled or can becleaned by filling the radiation chamber with a cleaning solution.Acids, such as diluted phosphoric acid, acetic acid or diluted nitricacid, are particularly suitable for cleaning.

    Maintenance

    39

  • WARNING!Safety data sheet for the cleaning agent selectedCause: Danger due to cleaning agent.Possible consequence: Damage to the health. Property damage.Measures: Observe the safety data sheet for thecleaning agent selected. Wear protective clothingwhen cleaning (protective eye wear, protective gloves).Do not use corrosive acids or acids that could causestress cracks, such as hydrochloric acid. Ensure thatno cleaning solution penetrates the lamp protectiontube. Ensure, when cleaning UV systems that nocleaning solution enters the pipework.Dispose of the waste cleaning solution in accordancewith the pertinent guidelines and regulations.

    7.1.1 Cleaning After Dismantling the UV Lamp Protection Tube WARNING!UV raysPossible consequence: Serious injuries.UV radiation is harmful to the eyes and skin Only operate the UV lamps UV-C when they are properly installed Install the UV lamp into the UV system in accord

    ance with the regulations prior to commissioning

    WARNING!Live parts!Possible consequence: Fatal or very serious injuries Measure: The device must be disconnected from

    the power supply before it is opened Disconnect damaged, defective or manipulated

    devices from the power supply

    CAUTION!Fingerprints on the UV lampPossible consequence: Early failure of the UV lamp Only touch the glass of the UV lamp with cotton

    gloves Fingerprints or impurities burn into the glass and

    can result in early failure For this reason always clean the lamp thoroughly

    with a cloth moistened with alcohol before installing it

    Then wipe the UV lamp with a soft cloth

    Maintenance

    40

  • Cleaning the UV SensorEvery time you clean the UV lamp protection tube, alsoclean the UV sensor

    1. Close the shut-off valves upstream and downstream of theradiation chamber

    2. Switch off the UV system with the STOPSTART button3. Switch off the main switch or disconnect from the mains

    power supply4. Drain the radiation chamber5. Open the water drain screw and the vent screw, drain the

    radiation chamber6. Undo the UV lamp protection tube holder by hand7. Screw out the lamp cover far enough to permit disconnection

    of the connecting cable plug from the UV lamp8. Fully remove the UV lamp and lay it to one side9. Loosen the UV lamp protection tube holder using the sup

    plied sickle spanner10. Remove the holder of the UV lamp protection tube from the

    radiation chamber by screwing out11. Pull out the lamp protection tube12. Remove the O-ring from the lamp protection tube13. Wash the lamp protection tube with cleaning solution or

    immerse it in cleaning solution until the film has beenremoved without leaving a trace

    14. Rinse the lamp protection tube with clean water and dry thoroughly with a soft cloth

    15. Carefully push the lamp protection tube into the radiationchamber until it reaches its stop position

    16. Push a new O-ring onto the end of the lamp protection tube -the sealing surfaces of the O-ring must be smooth and clean

    17.CAUTION!Check the lamp protection tube for damagebefore fitting A damaged lamp protection tube may not be

    refitted Ensure that the lamp protection tube is sitting

    correctly The lamp protection tube may not project

    out more than 40 mm and may not beoffset at an angle

    Feed the UV lamp protection tube into the radiation chamber.18. Push the UV lamp protection tube holder over the UV lamp

    protection tube and screw into the radiation chamber; tightenby hand

    19. Tighten the UV lamp protection tube holder with the suppliedsickle spanner; only minimal force is required.

    20. Assembly and connection of the UV lamp, see Chapter 5.3Fitting the lamp protection tubes on page 34

    Maintenance

    41

  • 21. Screw in the water drain screw and tighten; very little effort isneeded for this

    22. Slowly open the shut-off valve upstream of the radiationchamber

    23. Fill the radiation chamber until water emerges from the ventscrew

    24. Close the air vent screw and tighten; very little effort isneeded for this

    25. Open the shut-off valve downstream from the radiationchamber (only necessary with a manual shut-off valve)

    26. Check that the radiation chamber is not leaking27. Switch on the main switch or connect to the mains power

    supply The UV-system is again ready for operation.

    7.1.2 Cleaning with a Cleaning Solution NOTICE!Handling the cleaning solution It is also recommended in case of radiation cham

    bers regularly cleaned by filling with a cleaning solution that the water drain connector and the airvent plug be replaced by appropriate valves

    With larger radiation chambers, it is recommendedthat they are filled through the water drain openingusing an appropriate acid-resistant pump

    If the radiation chamber is filled with a pump, it isalso useful to circulate the cleaning solutionthrough the air vent opening This shortens the cleaning time and achieves a

    better result If the cleaning solution is collected in a suitable

    container and stored, it can be reused severaltimes

    1. Switch off the UV system using the STOPSTART key2. Switch off the main switch or disconnect from the mains

    power supply3. Close the shut-off valves upstream and downstream of the

    radiation chamber4. Empty the radiation chamber5. Screw in the water drain screw again and tighten; very little

    effort is needed for this6. Fill the radiation chamber with the cleaning solution through

    the vent opening Allow the cleaning solution to work for at least 20 minutes

    7. Open and remove the water drain screw8. Empty the radiation chamber and dispose of the cleaning sol

    ution in accordance with the pertinent regulations9. Flush the radiation chamber thoroughly with clean water until

    all the remains of cleaning solution have been removed

    Cleaning the lamp protection tubes byfilling the radiation chamber with acleaning solution:

    Maintenance

    42

  • 10. Screw in the water drain screw and tighten; very little effort isneeded for this

    11. Slowly open the shut-off valve upstream of the radiationchamber

    12. Fill the radiation chamber until water emerges from the ventplug

    13. Close the air vent plug and tighten; very little effort is neededfor this

    14. Open the shut-off valve downstream of the radiation chamber(only necessary with a manual shut-off valve) Check that the radiation chamber is not leaking

    15. Switch on the main switch or connect to the mains powersupply The UV-system is again ready for operation.

    7.1.3 Cleaning the UV Sensor 1. Loosen the sensor connection cable from the UV sensor2. Turn the UV sensor out of the radiation chamber3. Clean the quartz window with a cloth that has been saturated

    with cleaning solution until the coating has been removedwithout leaving a trace

    4. Rinse the quartz window with clean water and dry with a softcloth

    5. Examine the O-ring for damage and replace any damagedseals

    6. Screw in the UV sensor again and tighten; very little effort isneeded for this

    7. Connect the sensor connection cable to the UV sensor8. Slowly open the shut-off valve upstream of the radiation

    chamber9. Fill the radiation chamber until water emerges from the vent

    screw10. Close the air vent screw and tighten; very little effort is

    needed for this11. Open the shut-off valve downstream from the radiation

    chamber (only necessary with a manual shut-off valve) Check that the radiation chamber is not leaking

    12. Switch on the main switch or connect to the mains powersupply The UV-system is again ready for operation.

    Maintenance

    43

  • 7.2 Changing the UV lamps The UV lamps should be replaced by new lamps at the very latest:n when the sensor signal approaches the minimum irradiance

    threshold without this being based on other causes, such asthe formation of a coating on the lamp protection tubes or aserious deterioration of the UV-transmission;

    n when the operating life of the lamp is approaching the maximum service life of the lamp or has exceeded it

    WARNING!UV raysPossible consequence: Serious injuriesUV radiation is harmful to the eyes and skin Only operate the UV lamp UV-C when it is fully fittedand installed Install the UV lamp into the UV system in accord

    ance with the regulations prior to commissioning

    NOTICE!Never touch the glass of the UV lamp with bare hands.Fingerprints burn into the glass and can result in earlyfailure. Clean off fingerprints from the lamp with a clothmoistened with alcohol before installing.

    The UV lamp of the UV system should be installed inaccordance with all pertinent regulations prior to commissioning.Do not modify the fitted lamp connection cable withoutauthorisation.Do not modify the gap between the plug and the lampcover. Otherwise, it can not be guaranteed that the UVlamp lies against the closed end of the UV lamp protection tube. This is a prerequisite for safe disinfection.

    Note regarding multiple UV lamp systems: When UV lamps are being exchanged at the endof the maximum lamp service life, all of the UVlamps should be exchanged for new ones. When UV lamps are being exchanged due to lampwear, all of the UV lamps should be exchanged fornew ones. When exchanging faulty UV lamps, only new UVlamps may be used. In the event that only the faulty UV lamp is beingreplaced, then the new UV lamp should always beinstalled in the UV lamp protective tube which isthe furthest away from the UV sensor measuringposition. In the event of lamp failure towards the end of themaximum lamp service life, all of the UV lampsshould be exchanged for new ones.

    Maintenance

    44

  • Clean the lamp protection tubes each time a lamp isreplaced.Dispose of the used UV lamps in accordance with theapplicable guidelines and directives. Usually these canbe disposed of together with used fluorescent tubes.

    1. Switch off the UV system using the [START/STOP] key2. Switch off the master switch or disconnect from the mains

    power supply3. Undo the UV lamp protection tube holder by hand4. Screw out the lamp cover far enough to permit disconnection

    of the connecting cable plug from the UV lamp5. Fully remove the UV lamp and lay it to one side6. Check whether the O-ring on the UV lamp protection tube

    holder is lying in the groove provided and is not damaged.Damaged seals are too be replaced.

    7. Insert the new UV lamp into the UV lamp protection tube andallow it to project by approx. 100 mm

    8. Plug the connector plug onto the plug9. Secure the lamp cover on the UV lamp protection tube holder

    using the knurled nut, apply only minimal force to the knurlednut

    10. Switch on the main switch or connect to the mains powersupply

    1.Adjusting the warning and safety thresholdsAfter fitting a new UV lamp and cleaning of theUV system, the warning and safety thresholdsmust be checked and if necessary adjusted, see Chapter 4.12 Setting the warning thresholdon page 21 and Chapter 4.11 Setting thesafety threshold on page 20.

    With the system switched off, use the [CHANGE] key to display the lamp hours and lamp turn-ons

    2. Confirm with the [ENTER] key - [Request Access Code] willappear on the display

    3. Enter the access code and confirm with the ENTER key -[Reset] will appear in the display4. Confirm with the [ENTER] key - the display will now be reset

    7.3 Calibrating the UV sensor All new UV lamps need a burn-in time of 100 to 200hours. For this reason, the safety threshold andwarning threshold should be checked approx. 200operating hours after exchanging the lamp.

    Reset UV lamp hours and UV lampturn ons

    Maintenance

    45

  • The UV sensor is pre-calibrated at the factory and does not have tobe calibrated again.

    7.4 Replacement of the filter mats on the fan and the air outlet filter CAUTION!Soiled filter matsCause: Dirty filter mats on the fan and air outlet filtercan lead to the control cabinet overheating andbecoming damagedPossible consequence: Material damage.Measure: Replace the filter mats on the fan and the airoutlet filter at least once per year. The filter matsshould be replaced at shorter intervals in unfavourableambient conditions.

    1. Switch off the UV system using the [START/STOP] key2. Switch off the main switch3. Remove the cover of the fan. To do so, insert your fingers

    into the recesses on the bottom of the cover and remove thecover

    4. Remove the dirty filter mat and insert a new filter mat with thewhite side facing inwards (control cabinet side)

    5. Replace the filter mat on the air outlet filter as describedabove

    6. Switch on the main switch Switch on the UV system.

    Maintenance

    46

  • 8 Troubleshooting n User qualification, troubleshooting: trained user, see

    Chapter 2.2 Users' qualifications on page 7

    NOTICE!Troubleshooting on the open switch cabinet and thereplacement of components may only be done by anauthorised electrical engineer.

    Message: Flashing downward arrow

    Message: Flashing upward arrow The remaining free rinsing duration is displayed (instead of seconds, 2 squares flash).

    Fault Minimum irradiance level undershot (after maximumfree rinse duration has expired)

    Fault alert: UV sensor

    Acknowledge fault alert with the [ENTER] key

    Transition to emerg.service mode with display [CHANGE]key

    55. 0 W/m210 h Operation20 On/Off

    A0761

    Fig. 32: Warning threshold transgressed

    40.0 W/m2 F-rinseTime 00:11

    A0762

    Fig. 33: Safety threshold transgressed

    Fault

    UV sensorA0726

    Fig. 34: Fault

    Troubleshooting

    47

  • Emerg.service: Safety threshold transgressed

    System continues to run in emerg.service mode

    Acknowledge fault alert with the [ENTER] key.

    Fault description Cause RemedyWarning threshold undershot / safetythreshold undershot

    Coating formed on the lamp protectiontubes

    Clean the lamp protectiontube

    Insufficient UV-transmission in thewater to be treated

    Improve water quality

    UV lamp at the end of its UV emissionservice life

    Replace with new UVlamps

    Incorrect lamp current Operate with rated current

    UV sensor not correctly calibrated Recalibrate the UVsensor

    1. Acknowledge fault alert with the [ENTER] key2. Transition to emerg.service mode with display [CHANGE]

    key

    em.serv.

    UV sensorA0763

    Fig. 35: Emerg.service

    Faultxx lamp faulty Lamp #xx

    A0765

    Fig. 36: Fault: Lamp failureFault indication: xx lamp faulty

    Troubleshooting

    48

  • System continues to run in emerg.service mode

    The number of failed UV lamps is indicated.In the event that more than one UV lamp has failed,the failed UV lamp can be called up using the [Up] and[Down]-keys.

    Acknowledge fault alert with the [ENTER] key.

    Fault description Cause RemedyFault: Lamp failure UV lamp xx defective Replace with new UV lamps

    Incorrect lamp current Operate with rated currentBallast faulty Replace ballast

    Fault alert: Other fault

    Acknowledge fault alert with the [ENTER] key.

    Fault description Cause RemedyOther faults External fault signal device triggered Eliminate cause of external fault

    No external fault signal device connected and the contacts at the fault input are not bridged.

    Jumper contacts at fault input

    em.serv.xx lamp faulty Lamp #xx

    A0764

    Fig. 37: Emerg.service: Lamp failure

    Fault Misc. fault

    A0737

    Fig. 38: Other faults

    Troubleshooting

    49

  • Fault alert: Voltage supply too low

    Fault description Cause RemedyVoltage supply too low The supply voltage is or was lower than the minimum per

    missible supply voltageCheck supplyvoltage

    Fault alert: Bus fault

    Fault description Cause RemedyBus fault Bus connection interrupted Restore bus connection

    Insufficient supply voltage toone or more ballasts

    Check the supply voltage and micro fuse in the ballast(only an electrician is authorised to do so)

    Ballast faulty Replace the ballast (only an electrician is authorised todo so)

    Fault alert: Memory error

    Fault description Cause RemedyMemory error During self-testing, the control has

    detected an error in the memoryReplace the control (only an electrician isauthorised to do so)

    Faultline voltage too low

    A0766

    Fig. 39: Line voltage too low

    Fault Bus fault

    A0739

    Fig. 40: Bus fault

    Fault Memory fault

    A0740

    Fig. 41: Memory error

    Troubleshooting

    50

  • Fault alert: Default setting

    Fault description Cause RemedyDefault setting During self-testing, the control has

    detected an error in the memoryReplace the control (only an electrician isauthorised to do so)

    Message: Lamp current

    Fault description Cause RemedyLamp current fault Lamp current configured outside of per

    missible rangeConfigure the lamp current within permissible range (see datasheet)

    n User qualification, bus fault: Qualified electrician, see Chapter 2.2 Users' qualifications on page 7

    Bus with ring topologyAs the data bus is configured in a ring topology, defective ballasts or an interrupted data connection meanthat the subsequent ballasts can no longer communicate.

    Function and fault indications on ballastsThe three red LEDs on the ballasts serve for checkingfunctionality and fault finding. When the supply voltageis applied, all three LEDs light up for approx. 1 second.

    Fault Basic Setting

    A0741

    Fig. 42: Default setting

    Fault Lamp current

    A0731

    Fig. 43: Lamp current fault

    Function and fault indications on ballasts

    Troubleshooting

    51

  • LED "supply voltage" (designation on board "Power")n on

    Supply voltage to ballast is sufficientn off

    despite the fact that the master switch is on and the mainsplug is inserted

    Check supply voltage (only electrical technician) Check micro fuse

    LED "lamp" (designation on board "Fault")n flashes approx. 1 to 15 sec

    = lamp electrodes are preheated prior to ignitionn on

    UV lamp fails to light up. If the supply voltage is switchedoff and on again, the LED goes out and will only light upagain after a further attempt to ignite

    Check UV lamp

    LED "Bus" (designation on board "Tx")n lights up every 0.1 to 3 sec

    The ballast is in communication with the controller.n fails to light up

    despite the UV system being switched on: The bus connection to the ballast is interrupted or the bal

    last is faulty

    Troubleshooting

    52

  • 9 Technical data This technical data supplements the enclosed datasheet.If in doubt, the information on the data sheet applies.

    9.1 Performance data Type Quantity

    UV lampUV lampPerformance(Watt)

    ConnectorPerformance(Watt)

    Max. flow rate(98%/cm transm.,400 J/m2)(m3/h)

    Nominal widthConnector

    1x75W 1 75 90 5.7 G 1 1/4 "1x80W 1 80 100 5.4 G 1 1/4 "1x130W 1 130 150 8.7 G 2 "1x230W 1 230 250 20.0 DN 652x230W 2 2x230 500 64.0 DN 1253x230W 3 3x230 750 117.0 DN 1504x230W 4 4x230 1000 184.0 DN 2005x230W 5 5x230 1200 228.0 DN 2006x230W 6 6x230 1400 273.0 DN 2007x230W 7 7x230 1700 369.0 DN 2508x230W 8 8x230 1900 418.0 DN 2509x230W 9 9x230 2100 467.0 DN 25010x230W 10 10x230 2400 514.0 DN 25011x230W 11 11x230W 2600 561.0 DN 25012x230W 12 12x230 2800 600.0 DN 250

    Technical data

    53

  • Type mm

    Radiation chamber lengthmm

    Min. free space forexchanging lamps(mm)

    Empty weightOperating weight(kg)

    1x75W 140 1115 910 12/271x80W 114 630 600 8/141x130W 114 940 900 10/201x230W 140 1,468 1400 24/462x230W 220 1,640 1500 41/963x230W 273 1,665 1500 53/1384x230W 324 1,690 1600 65/1505x230W 324 1,690 1600 70/1906x230W 406 1,790 1600 75/2007x230W 406 1,920 1600 115/3108x230W 406 1,920 1600 115/3109x230W 406 1,920 1600 130/32010x230W 406 1,920 1600 130/32011x230W 406 1,920 1600 130/32012x230W 406 1,920 1600 130/320

    Technical data

    54

  • Permissible operating temperatures:n Water temperature: 5 ... 70 C; 5 ... 30 C for Dulcodes 75Wn Ambient temperature: 5 ... 40 CRequirements of the water to be treated:n Maximum operating pressure: 10 bar*

    * The permissible operating pressure is possibly lower withsystems that have a non-standard design, see the UVsystem data sheet

    n No corrosive or abrasive properties, chloride content < 250ppm

    n No tendency for sedimentationn Iron: 50 g/ln Manganese 20 g/ln Turbidity: 0.3 FNU

    NOTICE!Chemical cleaning processesWhere UV systems are cleaned using chemicalcleaning processes (CIP), you must observe the following: do not use any corrosive cleaning chemicals, e.g.

    HCl (hydrochloric acid) the maximum temperature of the CIP process must

    not exceed the following values: 90 C for alkaline cleaning media 50 C for acidic cleaning media

    Switch the UV system off during the CIP process

    Technical data

    55

  • 9.2 Dimensions sheet 9.2.1 Dimension sheet Dulcodes W 75

    Mains cable length approx. 0.3 m

    Sensor cable length approx. 5 m

    UV lamp cable approx. 1.3m

    View

    low

    A0836

    Fig. 44: Dimension sheet Dulcodes W 75* Fittings are not contained in the scope of supply

    Type L1mm

    L2mm

    L3mm

    L4mm

    L5mm

    LAmm

    LFmm

    LSmm

    DKmm

    ThreadDIN 2999

    VolumeLitre

    75W 250 935 1050 1115 1180 2150 124 605 139.7 Rp 1 1/4 15

    Technical data

    56

  • 9.2.2 Dimension sheet Dulcodes W 80 and 130

    UV lamp cable length approx. 2m

    Sensor cable length approx. 5m = 80W/130W

    low

    A0837

    Fig. 45: Dimension sheet Dulcodes W 80 and 130* Fittings are not contained in the scope of supply

    Type L1mm

    L2mm

    L3mm

    L4mm

    LAmm

    LMmm

    R"

    W 80 85 425 630 570 1230 96 1 1/4W 130 100 715 940 875 1840 98 2

    Technical data

    57

  • 9.2.3 Dimension sheet Dulcodes W 230

    FlangeUV lamp cable length

    low

    Flange

    Sensor cable length approx. 5m

    A0838

    Fig. 46: Dimension sheet Dulcodes W 230 single UV lamp system* Fittings are not contained in the scope of supply

    The radiation chamber can also be fitted with the flange aligned upwards

    Single UV lamp system

    Technical data

    58

  • UV lamp cable length

    low

    Flange

    Sensor cable length

    low

    A0839

    Fig. 47: Dimension sheet Dulcodes W 230 two or three UV lamp system* Fittings are not contained in the scope of supply

    The radiation chamber can also be fitted with the flange aligned upwards

    TypeW

    L1mm

    L2mm

    L3mm

    L4mm

    LAmm

    LFmm

    Hmm

    DFmm

    DKmm

    FlangeDIN 2576 PN 10

    2x230 300 1030 1640 1523 3140 190 193 250 219.1 DN 125x139,73x230 300 1030 1665 1532 3155 250 220 285 273 DN 150x168,3

    Two or three UV lamp system

    Technical data

    59

  • UV lamp cable length

    low

    Flange

    Flange

    Sensor cable length

    low

    A0840

    Fig. 48: Dimension sheet Dulcodes W 230 multiple UV lamp systemFittings are not contained in the scope of supplyThe radiation chamber can also be fitted with the flange aligned upwards

    TypeW

    L1mm

    L2mm

    L3mm

    L4mm

    L5mm

    L6mm

    L7mm

    4x230/32 270 1000 1690 1531 175 230 10805x230/32 270 1000 1690 1531 175 230 10806x230/32 270 1000 1790 1590 175 230 10807x230/40 340 1050 1920 1720 215 300 11308x230/40 340 1050 1920 1720 215 300 11309x230/40 340 1050 1920 1720 215 300 113010x230/40 340 1050 1920 1720 215 300 113011x230/40 340 1050 1920 1720 215 300 113012x230/40 340 1050 1920 1720 215 300 1130

    Multiple UV lamp system

    Technical data

    60

  • TypeW

    LAmm

    LFmm

    DFmm

    DKmm

    FlangeDIN 2576 PN 10

    4x230/32 3290 276 340 323.9 DN 200x219.15x230/32 3290 276 340 323.9 DN 200x219.16x230/32 3390 300 395 323.9 DN 250x2737x230/40 3520 350 395 406.4 DN 250x2738x230/40 3520 350 395 406.4 DN 250x2739x230/40 3520 350 395 406.4 DN 250x27310x230/40 3520 350 395 406.4 DN 250x27311x230/40 3520 350 395 406.4 DN 250x27312x230/40 3520 350 395 406.4 DN 250x273

    9.3 Electrical data The fuses are located in fuse boxes with a bayonetlock, on the right of the terminal box of the control.

    Radiation chamber:Lamp type Lamp current:HighFlux 130/230 W normal: 2.0 A

    Maximum: 2.3 AMinimum: 2.0 A

    Long Life 75 W not adjustable

    ControlFuse Site of installation Part no.230 V AC,50/60 Hz

    Upper fuse(self-supply by the control)

    712048

    230 V AC,50/60 Hz

    Lower fuse(switched mains outputs XR1-XR3)

    712033

    Technical data

    61

  • Nominal voltage230 V AC 10 %50/60 Hz

    Contact inputs (-X3:1 ... -X3.6):for contacts or switching transistors:Open circuit voltage: 5V 0.5 VInput resistance: 10 kOhm

    Voltage outputs (-X1:1 ... -X1.10):Type of contact: NOCLoad capacity: 250 V AC / 3 A / 100 VAWith inductive loads provide RC protection circuits!

    Relay outputs (-X4:1 ... -X4.6):Type of contact: NOCLoad capacity: 250 V AC / 3 A / 100 VAWith inductive loads provide RC protection circuits!

    Alarm relay (-X4:7 ... -X4.9):Type of contact: Changeover contactLoad capacity: 250 V AC / 3 A / 100 VAWith inductive loads provide RC protection circuits!

    Standard signal output mA (-X3:7 ... -X3.8):0/4...20 mA, potential-freeMaximum load: 600 Ohm

    Control cabinet

    Technical data

    62

  • 10 Appendix 10.1 Spare parts drawings

    A0773

    1. 2.3. 4. 5. 6. 7.

    8.9.

    10.

    12.13.

    11.

    Fig. 49: Spare parts drawings

    10.2 Spare Parts List No. Description Part no. Replacement interval each1 UV lamp cable 3.5 m for UV systems from 2002

    year of manufacture1017867 on request 1

    1 UV lamp cable 7.5 m for UV systems prior to2002 year of manufacture

    1024826 on request 1

    2 O-ring UV lamp protection tube 31.75-3.53EPDM/P

    1004920 after every opening ofthe radiation chamber

    1

    3 Teflon Ring 36.9/30.7 x 3 PTFE white 1002751 on request 14 O-ring UV lamp protection tube 31.75-3.53

    EPDM/P1004920 after every opening of

    the radiation chamber1

    5 UV lamp protection tube 40 x 2 x 1000 mm Qfor 75 W UV systems

    1020845 1

    5 UV lamp protection tube 32 x 2 x 600 mm Q for80 W UV systems

    1002467 on request 1

    5 UV lamp protection tube 32 x 2 x 900 mm Q for130 W UV systems

    1002468 on request 1

    5 UV lamp protection tube 32 x 2 x 1400 mm Qfor 1 x 230 W UV systems

    1002469 on request 1

    5 UV lamp protection tube 32 x 2 x 1500 mm Qfor 2-5 x 230 W UV systems

    1002470 on request 1

    5 UV lamp protection tube 32 x 2 x 1600 mm Qfor 6-12 x 230 W UV systems

    1002471 on request 1

    6 UV lamp 75 W Long Life 1020911 10 -14,000 h 16 UV lamp 80 W High Flux 1002485 8 -10,000 h 16 UV lamp 130 W High Flux 1002486 8 -10,000 h 1

    Appendix

    63

  • No. Description Part no. Replacement interval each6 UV lamp 230 W High Flux 1002487 8 -10,000 h 1-127 O-ring screwed plug G 1/4 " 10.82-1.78 EPDM/P 1002754 on request 18 Screwed plug DIN 910 G 1/4 /71.4571 1002752 on request 19 O-ring for UV sensor 22.00-3.00 EPDM/P 1002175 on request 110 UVC-U-Sensor P/D/W/R G 3/4 " 71.4571 1028115 on request 110 UV conversion kit from 80 W conversion from I to

    U sensor incl. UV sensor, sensor cable and compact control

    1029846 on request 1

    11 2m sensor cable for UV sensor 1029262 on request 111 5 m sensor cable for UV sensor 1021041 on request 111 10 m sensor cable for UV sensor 1028063 on request 112 O-ring screwed plug G 1/4 " 10.82-1.78 EPDM/P 1002754 on request 113 Screwed plug DIN 910 G 1/4 /71.4571 1002752 on request 1

    Components in the control cabinetDescription Part no. Replacement interval eachBallast for Dulcodes 75 W 1029051 on request 1Ballast for Dulcodes 80 W, 130 W, 230 W 731210 on request 1-12Replacement filter mat for the control cabinet ventilation

    1004212 on request 2

    Appendix

    64

  • 10.3 Dulcodes wiring diagram WARNING!Mains voltage on protective low voltagePossible consequence: Fatal or very serious injuries.Measure: If connecting the protective low voltage(SELV) to one of the X4 terminals, the X4 terminalsmust not be connected to mains voltage.

    9PE

    Relay outputsDigital inputs

    SHUT-OFF VALVE

    OUTPUT 230V

    FLUSHING VALVE

    FEED PUMP

    WARNING THRESHOLDRELAY

    SAFETY THRESHOLDRELAY

    ALARM RELAY

    FLOW / TEMPERATURE

    PAUSE

    FAULT

    INTERFACE

    DEFAULT OUTPUT

    INPUT

    XR7:2

    UV SENSOR

    Defaultsignal

    output(only internal)

    Voltage outputs

    Fan(optional)

    A0774

    Fig. 50: Dulcodes wiring diagram