Planmeca ProMax pre install guide Page 1 of 7 August, 2008 ProMax PRE INSTALLATION INSTRUCTIONS AND DIMENSIONS INPUT / OUTPUT OUTLETS All outlets must be within 3ft from the back of the column and 12”-48” from the floor. 1. Power: 110V, 20A dedicated circuit. 2. Remote exposure switch (See page 6 for wiring diagram): 2”x4” electrical outlet boxes at ProMax and remote exposure switch location. ½ ” conduit between the 2"x4" electrical boxes required. All parts supplied for exposure switch connection. 3. Pan/ceph film units w/ Admark: 1½ ” conduit to the darkroom, 30ft. cable supplied. 4. ProMax Ethernet interface: Standard Ethernet RJ45 port to the network HUB/SWITCH. Wall preparation: The pan or pan/ceph creates pull-out force of 1216 pounds. Wall bracket has 4 pre drilled holes, 16” and 12” on center. Either ones can be used. If metal studs are in wall, use a mounting board to expand the load over several studs to meet the pull - out requirement. If needed, sandwich the wall on both sides. YOU MUST USE EITHER 1. Top wall bracket and floor screws OR 2. Two wall brackets. Only one wall bracket is supplied with each system. If planning for 2 wall brackets, you must add part# 10001073 to your order. 48” – 50” 8-16” 16” Version 17

ProMax Preinstall Instructions

Embed Size (px)

Citation preview

Page 1: ProMax Preinstall Instructions

Planmeca ProMax pre install guide

Page 1 of 7 August, 2008

ProMax PRE INSTALLATION INSTRUCTIONS AND DIMENSIONS

INPUT / OUTPUT OUTLETS All outlets must be within 3ft from the back of the column and 12”-48” from the �oor.

1. Power: 110V, 20A dedicated circuit. 2. Remote exposure switch (See page 6 for wiring diagram):

2”x4” electrical outlet boxes at ProMax and remote exposure switch location. ½ ” conduit between the 2"x4" electrical boxes required.

All parts supplied for exposure switch connection. 3. Pan/ceph �lm units w/ Admark: 1½ ” conduit to the darkroom, 30ft. cable supplied. 4. ProMax Ethernet interface: Standard Ethernet RJ45 port to the network HUB/SWITCH.

Wall preparation: The pan or pan/ceph creates pull-out force of 1216 pounds. Wall bracket has 4 pre drilled holes, 16” and 12” on center. Either ones can be used. If metal studs are in wall, use a mounting board to expand the load over several studs to meet the pull -out requirement. If needed, sandwich the wall on both sides.

YOU MUST USE EITHER 1. Top wall bracket and �oor screws OR 2. Two wall brackets.

Only one wall bracket is supplied with each system. If planning for 2 wall brackets, you must add part# 10001073 to your order.

48”

– 50

8-16

16”

Version 17

Page 2: ProMax Preinstall Instructions

Planmeca ProMax pre install guide Version 17 Page 2 of 7

August, 2008

ProMax Panoramic

60 ½

” – 9

5 ¾

38 ½

” – 7

4”

chin

rest

48” –

50”

8”-1

6”

22” (9”)

27 ½ ”

64” (

49 ½

”)

39” (33 ½ ”) 37

” re

st p

ositi

on

(ent

ry1)

61” (42 ½ ”)

27 ½ ” 42” (39”) 64” (48”)

Digital Film

Recommended operating dimensions (actual product dimensions in parenthesis)

Top

limits

can

be

set a

t the

inst

all s

ite.

22” (9”)

Page 3: ProMax Preinstall Instructions

Planmeca ProMax pre install guide Version 17 Page 3 of 7

August, 2008

Promax Film Ceph

60 ½

” – 9

5 ¾

51” –

86

½ ”

46” (44 ½ ”)

27 ½”

28”

42” (39”)

38” –

73

½ ”

ear s

uppo

rts

32 ½

“ –

67 ¾

38 ½

” – 7

4”

chin

rest

37”

rest

pos

ition

(e

ntry

1)

48” –

50”

8-

16”

88” (83 ½ ”)

Top

limits

can

be

set a

t the

inst

all s

ite

Recommended operating dimensio ns (actual product dimensions in parenthesis)

64” (

49 ½

”)

Page 4: ProMax Preinstall Instructions

Planmeca ProMax pre install guide Version 17 Page 4 of 7

August, 2008

ProMax Digital ceph

38” –

73

½”

ear s

uppo

rts

33”

– 68

½”

51 ¼

” –

86 ½

60 ½

” –

95 ¾

38 ½

” – 7

4”

chin

rest

46” (44 ½”)

30”

27 ½”

37”

rest

pos

ition

(e

ntry

1)

39” (33 ½”)

85” (78”)

Recommended operating dimensions (actual product dimensions in parenthesis)

Top

lim

its c

an b

e se

t at t

he in

stal

l site

64” (

49 ½

”)

Page 5: ProMax Preinstall Instructions

Planmeca ProMax pre install guide Version 17 Page 5 of 7

August, 2008

Other install information: Each �lm based machine comes with one Autoprint Pan Cassette ( pan unit ) OR Standard Pan Cassette plus Ceph Cassette for Admark use ( pan/ceph unit ) Extra cassettes Standard panoramic, order number w/ Lanex screens: Autoprint panoramic, order number 10001288

,cirtemolahpeC Film recommended: pan: Agfa Dentus Ortholux 6”x12”

”01x”8 xulohtrO sutneD afgA :hpec Bite sticks: Bite sticks are autoclavable

08240001 rebmun redro ,kap-01,skcits etib artxE 1 rebmun redro :sdor 2 /w kap-01 ,setalp etib omoT 0005520

2nd wall bracket: order number 10001073 (optional) Handheld exposure order number 10004040 switch w/ wall mount (optional) Darkroom: Light tight, GBX2 �lter on safe light, 15W frosted maximum.

.mlif nepo morf ”84 naht resolc oN

Computer specs: 512 MB of RAM for digital Win 2000, XP

PCAnywhere connection in Imaging Station and Image server computers Hard drive: 2x 40GB for average size practice. Please consult with Planmeca certi�ed computer provider for your computer needs. Requirements vary greatly based on usage.

For more technical information, please contact Planmeca Customer Care (630) 529-230 0 or visit www.PlanmecaUSA.com

50024021

order number 50024015

Page 6: ProMax Preinstall Instructions

Page 6 of 7 Planmeca ProMax pre install guide

August, 2008 Version 17

Page 7: ProMax Preinstall Instructions

Ethernet ReadyThe new, built-in Ethernetconnection for Planmeca

Digital X-rays o�ers directnetwork connectivity.

You can literally plug theEthernet connection

directly into your network.

Note: An UninterruptedPower Supply (UPS) isrecommended forthe X-Ray.

Image AcquisitionWorkstation (2D)• Dimaxis Software acquires the

Panoramic, Tomographicand Cephalometric images (2D).

• Dimaxis Imaging Software is usedto enhance, �lter, measure,diagnose, archive and exportthe images.

Note: If using Vista, only VistaBusiness or Vista Ultimate(32 bit)

Server• Stores and archives all data

to MS SQL or SOLID.The data server should havelarge hard drive disk space,hardware redundancy and alogical backup system.

Note:

Additional WorkstationFor Panoramic,Tomographic andCephalometric images (2D)• Dimaxis Software

PlanmecaDigital X-Ray(ProMax Shown)

Local Area Network ConnectionRequired for sharing printer or addingadditional workstations.

Dimaxis Imaging Softwarefor 2D X-Ray Imaging

OptionalCrossover

OptionalPrinter

CPU Speci�cationsDigital ImageWorkstationFor Panoramic, Bitewing, Tomographicand Cephalometric 2D images• Pentium 4 or Higher• 60 GB or better HD• 2 GB or more RAM• WINDOWS XP / Server 2003 OS /

Vista (32 bit OS Only)• Ethernet 10/100 Mbit• 24-32 bit graphic card• monitor 1024 x 768 res. min.• High Speed Internet, DSL or better

CPU Speci�cationsAdditional WorkstationsFor Panoramic,Tomographic and Cephalometricimages• Refer to Digital Image

Acquisition Workstation (2D)

Database ServerSpeci�cations• Pentium 4 or Higher• 2 Hard Drives, 160 GB each min.• 1 GB or more RAM• WINDOWS XP / Server 2003 OS /

Vista (32 bit OS Only)

• Ethernet 10/100 Mbit

P L A N M E C AX -R a y Im a g in g S y s t e m s

Digital X-ray Imaging Hardware Diagramsand System Requirements

NOTE:Dimaxis is NOT compatiblewith 64Bit OS Systems

OS is requiredand requires Dimaxis 4.3.Xor newer.

forPlanmeca Digital Panoramicand Dixi X-Ray Systems

If using Vista, only VistaBusiness or Vista Ultimate(32 bit) OS is requiredand requires Dimaxis 4.3.Xor newer.