3
Project concepts The project concept was initially developed for CT image guidance, with the following workflow: 1) The patient is installed on the CT scanner table, and images are acquired for planning purpose. 2) The robot base is fixed to the patient and the patient can be prepared. 3) The robot is attached to the base part, ideally outside the scanner tunnel. 4) The patient is positioned again into the scanner to perform robot registration, robot automatic positioning and then the intervention, using force feedback teleoperation. The radiologist can change the needle orientation, either to adjust the needle path or compensate for needle bending. Fig. 1. Procedure workflow. From left to right: stages 1) to 4). This workflow was based on the use of a body-mounted system, to have natural physiological motion compensation at the surface of the patient. Such a system is sensitive to body attachment and gravity effects. To limit these drawbacks, partial compensation of the system gravity can be considered using a supporting system, in order to make a consensus between the slave robot supporting and the patient’s natural breathing. Initial solutions Based on the previous workflow, we proposed several designs for the slave system. First, we made systematic mechanical architecture synthesis to solve the problem in all its generality, and provide the system with 2T2R (2 translations, 2 rotations) mobility. This led to an original architecture (Fig. 2a), which is a contribution to needle manipulation development [MMT2014]. Despite a clear interest for this type of parallel architecture to perform the needle manipulation, several difficult shortcomings were underlined: the presence of isolated singularities within the robot workspace for rotations greater than 45 deg. w.r.t the normal to the patient's skin; the difficulty to integrate four actuators and their transmissions given the required torque, even after optimization; the positioning of the actuators outside of the CT imaging plane. It then appeared that the positioning of the needle at the entry point, which justified the two translations in the 2T2R system, could possibly be simplified. A possible solution was to make coarse positioning of the system base at the entry point, using table motions under the laser beam of the CT scanner. As a consequence, the needle orientation was possible with 2R mobility only. This system was completely designed (Fig. 2c), including a needle manipulation device (NMD) for grasping, insertion, and force measurement. The software for the CT table control was also implemented, in cooperation with Siemens. Fig. 2. From left to right, successive designs of the system: a) prototype of 2T2R architecture; b) first design based on a 2R architecture;

Project concepts - Équipe AVR

  • Upload
    others

  • View
    5

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Project concepts - Équipe AVR

ProjectconceptsTheprojectconceptwasinitiallydevelopedforCTimageguidance,withthefollowingworkflow:1)ThepatientisinstalledontheCTscannertable,andimagesareacquiredforplanningpurpose.2)Therobotbaseisfixedtothepatientandthepatientcanbeprepared.3)Therobotisattachedtothebasepart,ideallyoutsidethescannertunnel.4) The patient is positioned again into the scanner to perform robot registration, robot automaticpositioningandthentheintervention,usingforcefeedbackteleoperation.Theradiologistcanchangetheneedleorientation,eithertoadjusttheneedlepathorcompensateforneedlebending.

Fig.1.Procedureworkflow.Fromlefttoright:stages1)to4).

Thisworkflowwas based on the use of a body-mounted system, to have natural physiologicalmotioncompensation at the surface of the patient. Such a system is sensitive to body attachment and gravityeffects.To limit thesedrawbacks,partial compensationof thesystemgravitycanbeconsideredusingasupporting system, inorder tomakea consensusbetween the slave robot supportingand thepatient’snaturalbreathing.Initialsolutions

Based on the previous workflow, we proposed several designs for the slave system. First, we madesystematicmechanical architecture synthesis to solve theproblem in all its generality, andprovide thesystem with 2T2R (2 translations, 2 rotations) mobility. This led to an original architecture (Fig. 2a),whichisacontributiontoneedlemanipulationdevelopment[MMT2014].Despiteaclearinterestforthistype of parallel architecture to perform the needle manipulation, several difficult shortcomings wereunderlined:thepresenceofisolatedsingularitieswithintherobotworkspaceforrotationsgreaterthan45deg.w.r.tthenormaltothepatient'sskin;thedifficultytointegratefouractuatorsandtheirtransmissionsgiven the required torque, even after optimization; the positioning of the actuators outside of the CTimagingplane.Itthenappearedthatthepositioningoftheneedleattheentrypoint,whichjustifiedthetwotranslationsinthe2T2Rsystem,couldpossiblybesimplified.Apossiblesolutionwastomakecoarsepositioningofthesystem base at the entry point, using table motions under the laser beam of the CT scanner. As aconsequence, the needle orientation was possible with 2R mobility only. This system was completelydesigned (Fig. 2c), including a needle manipulation device (NMD) for grasping, insertion, and forcemeasurement.ThesoftwarefortheCTtablecontrolwasalsoimplemented,incooperationwithSiemens.

Fig.2.Fromlefttoright,successivedesignsofthesystem:a)prototypeof2T2Rarchitecture;b)firstdesignbasedona2Rarchitecture;

Page 2: Project concepts - Équipe AVR

c)finaldesignwithlasersintering/polymermaterials,includingsolutionsforneedlegraspingandinsertion,andforCTregistration.However, discussions with the IHU Scientific Board in 2015 led to conclude that: i) the wholesystemwouldbe toocomplex froman industrialperspective; ii) thewholesystemwouldbe tooexpensive.Then,wedecidednottofabricatethewholesystembutrathertofocusremainingtimeandfundsonthedesignandassessmentoftheNMD,whichistheoriginal(andpatented)partofthesystem.Needlemanipulationdevices,andlastyeardevelopments Fig. 2c is a CADviewof the last generationof 2R solution (July2015),which includes theNMDwhoseprinciple is basedon successive grasp and releaseof theneedle, combinedwithneedle insertionwhengrasped. The NMD [BIOROB2016] includes a cable actuated needle-grasping device, a force sensor[IEEEMech2016]and a translation mechanism actuated by piezoelectric motors. During last year, theremaining funds were entirely dedicated to its development, with the perspective that in thefuture this tool should be associated to differentmedical robots.Wespentmostof the remainingfundstohireanengineerforsixmonths,withtwoobjectives:1)tofabricateandimplementtheinitiallyproposedNMD;2) toproposeand implementanalternativetoolwiththesametechnologybutatmuchlowercost.ThesetwoNMDhavesimilarprinciple,basedonthreesubsystems:aneedleinsertiondevice,aforcesensor,andaneedle-graspingdevice.ThefirstNMD[Biorob2016]wasdesignedinordertomeetthefollowingrequirements:

• minimalsizeinordertobeutilizedwithinaCT-scannertunnel;• forcemeasurementalongtheneedle,upto10N(axial);• graspingofneedleswithdifferentdiameters,self-centering;• possibilitytograspandreleasetheneedle;• possibilitytomanuallyremovetheneedleatanymoment;• constructioncompatiblewiththeimagingmodalities,inparticularaslittlemetalaspossible.

Theresultingsystemwasimplementedusingadvancedrapidprototypingtechniques(Fig.3aand3b).Theinsertiondeviceincludesacustomforcesensorconstructedwithpolymermaterialsandstraingaugestoensuremedicalimagingtransparency[IEEETMech2016].Asthissystemwashighlyoptimizedbecauseofitsexpected integrationwithin thewholesystem, itsvolume isquite limited. Theactuator forgraspingwereplacedawayfromtheCTplane.ForneedleinsertionverysmallpiezoelectricactuatorswereselectedwithagoodcompatibilitywithCTimaging(Fig.3cand3d).

Fig.3.Fromlefttoright:a)needlegraspingdevicedetails;b)prototypeofneedlemanipulationdevice;c)CTimagesoftheprototype;

d)volumicrenderingofthetestbedobtainedfromastandardabdominalspiralacquisition.Inparallel,anewversionoftheNMDhasbeendeveloped.Thesystemisthoughttobeassociatedtothefluoroscopy Artis Zeego system. It is designed to serve as a tool mounted onto the end effector of amedical robot. ThisNMD is compatiblewith theKukaLBR IIWA that canbeused in an automatic or acollaborativeway.TherewasanothergoodreasonforthechoiceoftheKukarobot:Siemenshadsucha

Page 3: Project concepts - Équipe AVR

robotat IHU,andused itwith theSiemensArtisZeego inother researchprojects.However, thoughweexpected to use it in the proteCT project, itwas finally not possible. Fortunately, in themeantime,weapplied and obtained an IDEX funding from Strasbourg University for two Kuka IIWA (they are beinginstalledwhilewewritethisreport).The new NMD includes the same force sensor and needle-grasping device, but its needle insertionmechanism isdifferent.The constraintson themetalparts and thebulkare less severewith theZeegothanwithaCT.Asaresult,thetransmissionmechanismandtheactuatorshavebeenchanged,andlowercostcomponentshavebeenselected.Inthisversion,low-levelcontrolisalsoimplementedusinglowcostsystems(anArduinoboardconnectedtoaRaspberrysystem).Asimplegraphicalinterfaceisavailableonatactiledisplay,inordertoallowtheelementarymotionsofneedlegrasping,releaseandinsertion.Thevaluesmeasured by the force sensor are also displayed to the user. This low cost control architecture,however,isnotcompatiblewithforcefeedbackteleoperation.Thissystemhasbeentestedinvivo(Fig.4bto4d)usingapassivesupportingarm.WehavetestedtheprincipleoftargetingwiththeSiemensiGuidesoftware (planning in the images, and targeting using the system laser and the Zeego alignment). Ofcourse, theuseofeither thecollaborativemanipulationmodeof theKukarobot,or itspositioningaftertheautomaticregistration,willmaketargetingmucheasier.

Fig.4.DesignanduseofthenewlowcostNMD,fromlefttoright:a)CADviewofthesystemwithsideandtopcoversremoved;b)toolinitsversioncompatiblewiththeKukaIIWA(heremountedonapassivearm)duringexperimentsintheArtisZeegoroom;c)experimentsonananesthetizedswine:insertionintotheliverusingSiemensiGuidesoftware;d)3Dreconstructedimagesofthe

needleinsertionexperiment,whichconsistedintargetingthetipofapre-insertedneedle.