© IFM Quality Services Pty Ltd Page 1
This report is only for the use of participating laboratories
Proficiency Testing FINAL REPORT
15E29 Glow wire & Needle
flame test
Proficiency Testing Provider Certificate Number 3189-02.
We gratefully express our thanks to IECEE CTL WG2 for their support. In particular, thanks to Hubert Bachl for his assistance.
Revision 1: Equipment audit program status report added.
Program Coordinator: Ingrid Flemming IFM Quality Services Pty Ltd PO Box 877 Ingleburn NSW 2565 Australia Tel: +61 (0)2 9618 3311 Fax: +61 (0)2 9618 3355 Email: [email protected] Samples Dispatched: 5 August 2015 Results Due: 18 September 2015 Report Issued: 18 December 2015
© IFM Quality Services Pty Ltd Page 2
This report is only for the use of participating laboratories
Operational Summary
NUMBER OF ROUNDS AND FREQUENCY OF TESTING
Glow wire and needle testing proficiency testing programs operate approximately every 3 years.
APPLICABLE STANDARDS
• IEC 60695-11-5
• IEC 60695-2-10
SAMPLE CONTENT, MANUFACTURE AND ASSURANCE OF SAMPLE INTERGRITY
Samples A and B were intended for glow wire testing and consisted of 7 white and 7 clear plastic plaques
respectively. Samples C and D were intended for needle flame testing and consisted of 4 clear and 4 grey
plastic plaques respectively. Every unit in each sample was sourced from the same production batch.
Samples were sent to an accredited laboratory for homogeneity evaluations. Samples met the required
criteria for homogeneity.
METHOD OF DISPATCH
All samples were sent via express national/international courier.
ASSESSMENT PROCEDURE
Please refer to the assessment criteria on the following pages.
CONFIDENTIALITY
The identity of participants and any identifying information supplied by participants is treated as confidential.
Appropriate policies and procedures have been put into place to ensure this confidentiality is maintained.
Please refer to http://www.ifmqs.com.au/ifm_policy_on_confidentiality.htm for the detailed policy.
TEST INSTRUCTIONS
Instructions applicable to this testing round may be found at:
http://www.ifmqs.com.au/proficiency/instructions/15e29_Instructions.pdf
INDIVIDUAL RESULTS
Should individual test results be required, these will be provided upon request.
CORRECTIVE ACTIONS
Where corrective action is required, participants are advised to start as soon as possible. CB scheme and
accredited laboratories are expected to register any required action items in their internal quality
management system in accordance with the requirements of ISO/IEC 17025:2005.
Participants are referred to IFM’s website for policies relating to participant appeal of assessments and
IFM’s role in monitoring corrective actions (http://www.ifmqs.com.au/ifm_terms_and_conditions.htm).
IFM Quality Services will monitor the completion of corrective actions for a period of 3 months from the date
of issue of this report. After this time, IFM will refer any remaining open items for CB scheme laboratories
© IFM Quality Services Pty Ltd Page 3
This report is only for the use of participating laboratories
to the IECEE secretariat for continuation of this monitoring activity and possible issue of a General Non-
Conformity Report.
Please note that corrective actions will only be accepted where evidence is submitted which demonstrates
compliance with all items in ISO/IEC 17025:2005 §4.11 AND 4.12, including registration of the items in the
company’s quality management system, however named.
Actions will be closed by IFM only after all required items have been satisfied, namely:
♦ Indication that review of the results has occurred (17025:2005 §4.10 and 4.11.1),
♦ Corrective action registration reference/code in the laboratory’s quality management system is
provided,
♦ Determination of root cause has been conducted (17025:2005 §4.11.2). Hints and tips relating
to acceptance of root cause can be found on:
http://www.ifmqs.com.au/proficiency/references/QRD003-01_From_Direct_to_Root_Cause.pdf
♦ The selection and conduct of corrective actions has been undertaken (17025:2005 §4.11.3),
♦ The success of corrective actions has been confirmed (17025:2005 §4.11.4) and
♦ Preventive actions have been developed and implemented (17025:2005 §4.12)
In cases where voluminous data is generated, a summary in English, on Company letterhead may be
submitted together with summarized findings for the above dot-points.
♦ It is the laboratory’s responsibility to ensure they maintain (on-site) accurate records of the
conduct of their corrective actions and copies of the reports. As is the case with all non-
conformities, these reports and information should be reviewed by the laboratory’s
management, internal and external assessors.
♦ It is the responsibility of the relevant NCB and/or accreditation body to review the corrective
action reports during on-site assessments to confirm the technical relevance of any findings
and actions.
♦ As IFM’s role is only to monitor completion of the corrective action process, individual
laboratory submissions will not be held by IFM after the submitted reports have been sighted,
checked and logged.
Corrective actions are to be submitted via email to [email protected]
© IFM Quality Services Pty Ltd Page 4
This report is only for the use of participating laboratories
Available tests Assessment Criteria
Enumerative Tests at 840 °C, 900 °C and 960 °C:
• Time to Ignite
For enumerative tests, robust statistics are applied in order to determine a Z score for each result. Acceptable Z scores are |Z|<3.00.The statistically acceptable high and low was rounded to the nearest 0.5 sec conforming to clause 9 of IEC60695-1-11:2004. Whenever possible, non-numeric results are assessed by applying pass/fail assessments without issue of Z scores.
Variations to the normal assessment procedure will be detailed in the discussion.
Yes/No response at 840 °C, 900 °C and 960 °C:
• Ignition (occurrence of ignition)
• Flame persistent, when samples ignited
• Ignition of specified layer
Participants are assessed against the consensus result when 75% consensus is achieved.
Available Tests
GLOW WIRE TEST
Available tests Assessment Criteria
Enumerative Tests:
• Flame burn time
Participants’ results are grouped into results less than or greater than 30 seconds. Consensus results are used for assessment. 75% or greater consensus is preferred.
Variations to the normal assessment procedure will be detailed in the discussion.
Yes/No response:
• Ignition
• Ignition of specified layer Participants are assessed against the consensus result when 75% consensus is achieved.
NEEDLE FLAME TEST
Participation Summary
⇒ 119 organisations enrolled
Equipment Audit
⇒ 66 laboratories are required to follow up for glow wire test
⇒ 40 laboratories are required to follow up for needle flame test
Proficiency Test
⇒ 115 organisations participated
⇒ 179 sets of results were submitted
⇒ 5817 individual test results were received
⇒ 3077 results were assessed
⇒ 2740 results were not assessed
⇒ 77 results were assessed as outliers
© IFM Quality Services Pty Ltd Page 5
This report is only for the use of participating laboratories
Equipment Audit
For each question, the total number of responses requiring follow-up are shown in Table 1 (Glow Wire Test)
and Table 2 (Needle Flame Test).
TABLE 1: SUMMARY OF RESPONSES REQUIRING FOLLOW-UP : GLOW WIRE TEST
Question Number
What is the concentration of Nickel in the glow wire loop? 1 What is the concentration of chromium in the glow wire loop? 1
The diameter of a straight section of the loop wire is 3 Has the glow wire loop been annealed? 3
Is there a feedback mechanism to maintain temperature during the test 23 When operational position of the glow wire loop is 0 The force that the loop applies to a test sample is 7
What is the maximum penetration that the loop can exert on a sample? 5 Check the tip of the glow wire loop. Are there any cracks on the tip? 6
The thermocouple placed in the pocket hole at the back of glow wire tip is type: (J, K, T etc) 1
The class of the above thermocouple type is 4 The diameter of the above thermocouple is 1
The distance of the tip of the above thermocouple to the end of the pocket hole is 0
When the test specimen is in position, what is the distance from the specimen to the front of the chamber? 4
When the test specimen is in position, what is the distance from the specimen to the left side of the chamber? 2
Perform a silver foil check. At what temperature does the silver foil melt? 6 Is the chamber a draught free location for performing the test? 9
Is the chamber vented to remove smoke and fumes? 0 If yes above, when does venting take place? 1 The amount of light falling on the specimen is 5
The resolution of the timing device used for this test is 14 How often is the glow wire tip cleaned? 11
How often is the glow wire tip checked for damage? 28 Does the check for damage include inspection for cracks? 4
Does the check for damage include inspection for contamination of thermocouple/pocket hole? 12
Is the specimen held vertically? 1
Is the glow wire tip applied centrally to the specimen? 0
Have precautions been made to protect staff from risks of burning/fire 1
Have precautions been made to protect staff from the risks of electric shock 1
Have precautions been made to protect staff from inhalation of smoke/toxic products 2
Have precautions been made to protect staff from exposure to toxic residues 4
© IFM Quality Services Pty Ltd Page 6
This report is only for the use of participating laboratories
TABLE 2: SUMMARY OF RESPONSES REQUIRING FOLLOW-UP : NEEDLE FLAME TEST
Question Number
What is the length of the burner? 0 What is the diameter of the bore? 10
What gas is supplied to the burner? 1 What is the purity of the gas supplied to the burner? 1
Is air supplied to the burner during use or calibration? 15
How long is the burner allowed to reach equilibrium before performing tests? 9 The minimum frequency of performing copper block confirmatory test is dictated by: 0
Perform the confirmatory procedure as described in Annex A of IEC60695--11-5. How long does it take for the temperature to ascend to 700 degrees C from 100 degrees C? 2
The background against which the flame is viewed is: 4
The volume inside the test chamber is 1
Is the chamber free of draught? 5
Can the chamber can be completely closed? 4
Can the chamber be vented to extract fumes? 0
If Yes, when does fume extraction occur? 2 The tolerance of the timing device used for this test is 6
© IFM Quality Services Pty Ltd Page 7
This report is only for the use of participating laboratories
SAMPLE CUTTING AND NUMBER OF TEST REPEATS
Participants were requested to report whether they cut the sample, and how many repeat tests were
conducted. The answers to these questions were not formally assessed. However, the results of these
were used in comparisons with participant’s responses to other questions.
Whether the sample was cut or whole does not appear to impact on participant results for these samples.
However, due to the small number of participants reporting cutting the sample it is difficult to observe a
trend. Figure 1 compares ignition of the specified layer during the 840 °C test for cut and whole samples.
The number of repeats performed also did not appear to have an impact on participant results for these
samples. Figure 2 compares the number of repeats performed and ignition of the specified layer at 840 °C.
Some participants reported observing different results for duplicate tests. It would be advisable to
investigate test repeatability and reproducibility where inconsistencies in ignition patterns are observed.
Discussion - Glow Wire Test
FIGURE 1
FIGURE 2
109
42
10 9
0
20
40
60
80
100
120
No Yes
Ignition of the specified layer
Num
ber o
f par
ticip
ants
Glow wire test - Ignition of the specified layer at 840 °C for dif ferent sample preparation 15E29-Sample A
Whole sample
Cut sample
3
35
6 3
15
65
16 18
0
10
20
30
40
50
60
70
0 1 2 3 or more
Number of repeats
Num
ber o
f par
ticip
ants
Glow wire test Ignition of the specif ied layer at 840 °C for dif ferent numbers of repeats 15E29-Sample A
Ignition of the specified layer
No ignition of the specified layer
© IFM Quality Services Pty Ltd Page 8
This report is only for the use of participating laboratories
IGNITION OF THE SAMPLE
Sample A
Participants unanimously agreed that Sample A ignited at all temperatures tested.
Sample B
Refer to Figure 3. The percentage of participants reporting ignition increased with increasing temperature.
Participants received pass assessments for reporting no ignition at 840 and 900 °C and ignition at 960 °C.
Participants reporting no ignition at 960 °C were not assessed as 75% consensus was not achieved.
Discussion - Glow Wire Test
FIGURE 3
312
60
0
20
40
60
80
100
840 900 960Perc
enta
ge o
f par
ticip
ants
repo
rting
igni
tion
Glow wire temperature (°C)
Glow wire test - Ignition at dif frent temperaures 15E29 Sample B
IGNITION OF THE PRESCRIBED LAYER
Sample A
Figure 4 presents the percentage of participants reporting ignition of the prescribed layer at each test
temperature. This test has not been assessed as 75% consensus was not achieved for any temperature.
Varying results may have been due to the application of differing standards. Clause 9 in IEC60695-2-11
and IEC60695-2-12 both specify the sample must be observed for an additional 30 seconds after the glow
wire has been removed. However, clause 9 in IEC60695-2-13 specifies the additional time to be 5
seconds.
The question in this program did not specify the amount of time the sample must be observed after the glow
wire had been removed.
Sample B
Participants generally agreed that the prescribed layer did not ignite at any of the temperatures tested.
© IFM Quality Services Pty Ltd Page 9
This report is only for the use of participating laboratories
Discussion - Glow Wire
OTHER TESTS
A summary of the assessments made to other glow wire tests can be found in Table 3.
Temperature Test Response Assessment A Assessment B
840 °C
Sample Ignition Yes Pass Fail No Fail Pass
Time to Ignite ≤ 2.5 sec Not assessed
Flame Persistent Yes Pass
Not assessed No Fail
Layer Ignition Yes
Not assessed Fail
No Pass
900 °C
Sample Ignition Yes Pass Fail No Fail Pass
Time to Ignite ≤ 2.5 sec Not assessed
Flame Persistent Yes Pass
Not assessed No Fail
Layer Ignition Yes
Not assessed Fail
No Pass
960 °C
Sample Ignition Yes Pass No Fail
Time to Ignite ≤ 2.0 sec Not assessed
Flame Persistent Yes Pass
Not assessed No Fail
Layer Ignition Yes
Not assessed Fail
No Pass
Not assessed
TABLE 3: SUMMARY OF GLOW WIRE TEST ASSESSMENTS
FIGURE 4
2832 33
05
101520253035
840 900 960Perc
enta
ge o
f par
ticip
ants
repo
rting
igni
tion
Glow wire temperature (°C)
Glow wire test - Ignition of the prescribed layer at different temperature15E29-Sample A
© IFM Quality Services Pty Ltd Page 10
This report is only for the use of participating laboratories
Discussion - Needle flame
SAMPLE CUTTING AND NUMBER OF REPEATS FOR NEEDLE FLAME TESTS
Participants were requested to report whether they cut the sample, and how many repeat tests were
conducted. These questions were not formally assessed. Please refer to Figure 5 and Figure 6.
Figure 5 compares ignition of cut and un-cut samples.
Figure 6 compares ignition with the number of repeat tests conducted.
No correlations were observed.
FIGURE 5
63 67
10 7
01020304050607080
No Yes
Needle Flame Sample Ignition
Num
ber o
f par
ticip
ants
Needle f lame test - Ignition of sample for different sample preparation 15E29-Sample D
Whole sample
Cut sample
FIGURE 6
8
2926
108
34
24
5
05
10152025303540
0 1 2 3 or more
Number of repeats
Num
ber o
f par
ticip
ants
Needle f lame test - Ignition of sample for different numbers of repeats 15E29-Sample D
Ignition of the sample
No ignition of the sample
© IFM Quality Services Pty Ltd Page 11
This report is only for the use of participating laboratories
Test Response Assessment C Assessment D
Needle Flame ignition Yes Pass
Not assessed No Fail
Needle Flame Layer Ignition Yes Fail Fail No Pass Pass
Needle Flame Burn Time <30 Fail >30 Pass
Not assessed
TABLE 4: SUMMARY OF NEEDLE FLAME TEST ASSESSMENTS
NEEDLE FLAME BURN TIME
Sample C
Participants generally agreed the sample did not extinguish within 60 seconds.
Sample D
Participants generally agreed that the sample extinguished immediately after the removal of the needle
flame. However, as consensus was not reached for ignition of this sample, this test was not assessed.
OTHER TESTS
A summary of the assessments made for needle flame tests can be found in Table 4.
Discussion - Needle flame
NEEDLE FLAME IGNITION
Sample C
Participants generally agreed that the sample ignited when subjected to a needle flame test.
Sample D
There was a near equal number of participants reporting yes and no for ignition for sample D. This test was
not assessed as consensus was not achieved.
© IFM Quality Services Pty Ltd Page 12
This report is only for the use of participating laboratories
Results
GLOW WIRE TEST: IGNITION OF SAMPLE A
169
10
20406080
100120140160180
Yes(Pass)
No(Fail)
15e29 Glow Wire & Needle FlameQ09 Glow Wire 960°C Ignition
Sample A (Y/N)
Freq
uenc
y
170
00
20406080
100120140160180
Yes(Pass)
No(Fail)
15e29 Glow Wire & Needle FlameQ05 Glow Wire 900°C Ignition
Sample A (Y/N)
Freq
uenc
y
170
00
20406080
100120140160180
Yes(Pass)
No(Fail)
15e29 Glow Wire & Needle FlameQ01 Glow Wire 840°C Ignition
Sample A (Y/N)
Freq
uenc
y
© IFM Quality Services Pty Ltd Page 13
This report is only for the use of participating laboratories
Results
GLOW WIRE TEST: IGNITION OF SAMPLE B
102
68
0
20
40
60
80
100
120
Yes(Not assessed)
No(Not assessed)
15e29 Glow Wire & Needle FlameQ09 Glow Wire 960°C Ignition
Sample B (Y/N)
Freq
uenc
y
21
149
020406080
100120140160
Yes(Fail)
No(Pass)
15e29 Glow Wire & Needle FlameQ05 Glow Wire 900°C Ignition
Sample B (Y/N)
Freq
uenc
y
5
165
020406080
100120140160180
Yes(Fail)
No(Pass)
15e29 Glow Wire & Needle FlameQ01 Glow Wire 840°C Ignition
Sample B (Y/N)
Freq
uenc
y
© IFM Quality Services Pty Ltd Page 14
This report is only for the use of participating laboratories
15e29 Glow Wire & Needle FlameQ02 Glow Wire 840°C Time to Ignite
Sample A (sec)Number of Results Analysed 157Quartile 1 0.0000Quartile 3 1.0000IQR 1.0000NIQR 0.7413Median 0.2000Uncertainty of Assigned ValueNIQR/SQRT(No. of Results) 0.0592Acceptable High:Median + (3*NIQR) 2.4239Acceptable Low:Median - (3*NIQR) -2.0239Statistically Acceptable Range 4.4478
15e29 Glow Wire & Needle FlameQ06 Glow Wire 900°C Time to Ignite
Sample A (sec)Number of Results Analysed 156Quartile 1 0.0000Quartile 3 1.0000IQR 1.0000NIQR 0.7413Median 0.1650Uncertainty of Assigned ValueNIQR/SQRT(No. of Results) 0.0594Acceptable High:Median + (3*NIQR) 2.3889Acceptable Low:Median - (3*NIQR) -2.0589Statistically Acceptable Range 4.4478
Results
GLOW WIRE TEST: TIME TO IGNITION: SAMPLE A
15e29 Glow Wire & Needle FlameQ10 Glow Wire 960°C Time to Ignite
Sample A (sec)Number of Results Analysed 156Quartile 1 0.0000Quartile 3 0.8000IQR 0.8000NIQR 0.5930Median 0.1000Uncertainty of Assigned ValueNIQR/SQRT(No. of Results) 0.0475Acceptable High:Median + (3*NIQR) 1.8791Acceptable Low:Median - (3*NIQR) -1.6791Statistically Acceptable Range 3.5582
0102030405060708090
0.1
0.5
0.9
1.3
1.7
Mor
e
Freq
uenc
y
Result (sec)
15e29 Glow Wire & Needle FlameQ10 Glow Wire 960°C Time to Ignite
Sample A
0102030405060708090
0.16
5
0.66
5
1.16
5
1.66
5
2.16
5
Mor
e
Freq
uenc
y
Result (sec)
15e29 Glow Wire & Needle FlameQ06 Glow Wire 900°C Time to Ignite
Sample A
0102030405060708090
0.2
0.7
1.2
1.7
2.2
Mor
e
Freq
uenc
y
Result (sec)
15e29 Glow Wire & Needle FlameQ02 Glow Wire 840°C Time to Ignite
Sample A
© IFM Quality Services Pty Ltd Page 15
This report is only for the use of participating laboratories
Results
GLOW WIRE TEST: TIME TO IGNITION: SAMPLE B
05
101520253035404550
1.76 8.
14.2
4
20.4
8
26.7
2
32.9
6
Mor
e
Freq
uenc
y
Result (sec)
15e29 Glow Wire & Needle FlameQ10 Glow Wire 960°C Time to Ignite
Sample B
Time to ignition was only assessed when the consensus indicated sample ignition.
Consensus was achieved for 840 °C (no ignition) and 900 °C (no ignition).
Consensus was not achieved for ignition of the sample at 960 °C. Therefore, the time to ignition was not
assessed. Results for time to ignite at 960 °C are shown below.
© IFM Quality Services Pty Ltd Page 16
This report is only for the use of participating laboratories
Results
GLOW WIRE TEST: PERSISTENCE OF FLAME: SAMPLE A
170
00
20406080
100120140160180
Yes(Pass)
No(Fail)
15e29 Glow Wire & Needle FlameQ11 Glow Wire 960°C Flame Persistent
Sample A (Y/N)
Freq
uenc
y
170
00
20406080
100120140160180
Yes(Pass)
No(Fail)
15e29 Glow Wire & Needle FlameQ07 Glow Wire 900°C Flame Persistent
Sample A (Y/N)
Freq
uenc
y
168
1 10
20406080
100120140160180
Yes(Pass)
No(Fail)
Not Assessed
15e29 Glow Wire & Needle FlameQ03 Glow Wire 840°C Flame Persistent
Sample A (Y/N)
Freq
uenc
y
© IFM Quality Services Pty Ltd Page 17
This report is only for the use of participating laboratories
Results
GLOW WIRE TEST: PERSISTENCE OF FLAME: SAMPLE B
6
105
0
20
40
60
80
100
120
Yes(Not assessed)
No(Not assessed)
15e29 Glow Wire & Needle FlameQ11 Glow Wire 960°C Flame Persistent
Sample B (Y/N)
Freq
uenc
y
Persistence of flaming was only assessed when the consensus indicated sample ignition.
Consensus was achieved for 840 °C (no ignition) and 900 °C (no ignition).
Consensus was not achieved for ignition of the sample at 960 °C. Therefore, the persistence of flaming
was not assessed. Results for persistence of flaming at 960 °C are shown below.
© IFM Quality Services Pty Ltd Page 18
This report is only for the use of participating laboratories
Results
GLOW WIRE TEST: IGNITION OF SPECIFIED LAYER SAMPLE A
56
114
0
20
40
60
80
100
120
Yes(Not assessed)
No(Not assessed)
15e29 Glow Wire & Needle FlameQ12 Glow Wire 960°C Layer Ignition
Sample A (Y/N)
Freq
uenc
y
54
116
020406080
100120140
Yes(Not assessed)
No(Not assessed)
15e29 Glow Wire & Needle FlameQ09 Glow Wire 900°C Layer Ignition
Sample A (Y/N)
Freq
uenc
y
48
122
020406080
100120140
Yes(Not assessed)
No(Not assessed)
15e29 Glow Wire & Needle FlameQ04 Glow Wire 840°C Layer Ignition
Sample A (Y/N)
Freq
uenc
y
© IFM Quality Services Pty Ltd Page 19
This report is only for the use of participating laboratories
Results
GLOW WIRE TEST: IGNITION OF SPECIFIED LAYER SAMPLE B
1
161
020406080
100120140160180
Yes(Fail)
No(Pass)
15e29 Glow Wire & Needle FlameQ12 Glow Wire 960°C Layer Ignition
Sample B (Y/N)
Freq
uenc
y
0
155
020406080
100120140160180
Yes(Fail)
No(Pass)
15e29 Glow Wire & Needle FlameQ09 Glow Wire 900°C Layer Ignition
Sample B (Y/N)
Freq
uenc
y
0
155
020406080
100120140160180
Yes(Fail)
No(Pass)
15e29 Glow Wire & Needle FlameQ04 Glow Wire 840°C Layer Ignition
Sample B (Y/N)
Freq
uenc
y
© IFM Quality Services Pty Ltd Page 20
This report is only for the use of participating laboratories
Results
NEEDLE FLAME TEST: IGNITION OF SAMPLE
77 79
0102030405060708090
Yes(Not assessed)
No(Not assessed)
15e29 Glow Wire & Needle FlameQ13 Needle Flame Ignition
Sample D (Y/N)
Freq
uenc
y
152
20
20406080
100120140160
Yes(Pass)
No(Fail)
15e29 Glow Wire & Needle FlameQ13 Needle Flame Ignition
Sample C (Y/N)
Freq
uenc
y
© IFM Quality Services Pty Ltd Page 21
This report is only for the use of participating laboratories
Results
NEEDLE FLAME TEST: IGNITION OF SPECIFIED LAYER
153
00
20406080
100120140160180
Yes(Pass)
No(Fail)
15e29 Glow Wire & Needle FlameQ14 Needle Flame Layer Ignition
Sample D (Y/N)
Freq
uenc
y
19
135
020406080
100120140160
Yes(Fail)
No(Pass)
15e29 Glow Wire & Needle FlameQ14 Needle Flame Layer Ignition
Sample C (Y/N)
Freq
uenc
y
© IFM Quality Services Pty Ltd Page 22
This report is only for the use of participating laboratories
Results
NEEDLE FLAME TEST: BURN TIME
56
7
0
10
20
30
40
50
60
<30 sec(Not assessed)
>30 sec(Not assessed)
15e29 Glow Wire & Needle FlameQ15 Needle Flame Burn Time
Sample D (sec)
Freq
uenc
y
3
150
020406080
100120140160
<30 sec(Fail)
>30 sec(Pass)
15e29 Glow Wire & Needle FlameQ15 Needle Flame Burn Time
Sample C (sec)
Freq
uenc
y
© IFM Quality Services Pty Ltd Page 23
This report is only for the use of participating laboratories
ôEnd of Reportò
Lab Code Reason Withdrawn / Comment Due DateAction Alert
Glow Wire and Needle Flame - 15e29 (201505)
Electrical Program Status Report
1 CBTL - Certification Body Testing Laboratory
2 CBTL - Certification Body Testing Laboratory
3 CBTL - Certification Body Testing Laboratory
4 CBTL - Certification Body Testing Laboratory
5 Private
6 CBTL - Certification Body Testing Laboratory
7 CBTL - Certification Body Testing Laboratory
8 CBTL - Certification Body Testing Laboratory
9 CBTL - Certification Body Testing Laboratory
10 CBTL - Certification Body Testing Laboratory
11 CBTL - Certification Body Testing Laboratory
12 CBTL - Certification Body Testing Laboratory
13 CBTL - Certification Body Testing Laboratory
14 Private
15 CBTL - Certification Body Testing Laboratory
Y Y16 CBTL - Certification Body Testing Laboratory
O 18/12/2015 Part01: 1 result assessed as outlying.
17 CBTL - Certification Body Testing Laboratory
Y Y18 CBTL - Certification Body Testing Laboratory
O 18/12/2015 Part01: 2 results assessed as outlying.
19 CBTL - Certification Body Testing Laboratory
20 CBTL - Certification Body Testing Laboratory
21 CTF - Customer’s Testing Facility
22 Private
23 CBTL - Certification Body Testing Laboratory
24 CBTL - Certification Body Testing Laboratory
25 CBTL - Certification Body Testing Laboratory
26 CBTL - Certification Body Testing Laboratory
27 CBTL - Certification Body Testing Laboratory
28 CBTL - Certification Body Testing Laboratory
29 CBTL - Certification Body Testing Laboratory
30 CBTL - Certification Body Testing Laboratory
31 CBTL - Certification Body Testing Laboratory
32 CBTL - Certification Body Testing Laboratory
33 CBTL - Certification Body Testing Laboratory
34 CBTL - Certification Body Testing Laboratory
35 CTF - Customer’s Testing Facility
Y Y36 CTF - Customer’s Testing Facility
O 18/12/2015 Part01: 1 result assessed as outlying.
37 CBTL - Certification Body Testing Laboratory
38 CBTL - Certification Body Testing Laboratory
18/12/2015 Page 1 of 5Produced by IFM Quality Services Pty Ltd
WWW.IFMQS.COM.AU - Ph: +61 2 9618 3311
Lab Code Reason Withdrawn / Comment Due DateAction Alert
Glow Wire and Needle Flame - 15e29 (201505)
Electrical Program Status Report
39 CBTL - Certification Body Testing Laboratory
40 CBTL - Certification Body Testing Laboratory
41 CBTL - Certification Body Testing Laboratory
Y Y42 CBTL - Certification Body Testing Laboratory
O 18/12/2015 Part01: 1 result assessed as outlying.
43 CBTL - Certification Body Testing Laboratory
44 CBTL - Certification Body Testing Laboratory
Y Y45 CBTL - Certification Body Testing Laboratory
O 18/12/2015 Part02: 2 results assessed as outlying.O 18/12/2015 Part01: 1 result assessed as outlying.
Y Y46 CTF - Customer’s Testing Facility
O 18/12/2015 Part02: 1 result assessed as outlying.O 18/12/2015 Part01: 1 result assessed as outlying.
47 CBTL - Certification Body Testing Laboratory
48 Private
49 CBTL - Certification Body Testing Laboratory
50 CTF - Customer’s Testing Facility
Y Y51 CTF - Customer’s Testing Facility
O 18/12/2015 Part01: 1 result assessed as outlying.
52 CBTL - Certification Body Testing Laboratory
53 CBTL - Certification Body Testing Laboratory
55 CBTL - Certification Body Testing Laboratory
Y Y56 CTF - Customer’s Testing Facility
O 18/12/2015 Part01: 2 results assessed as outlying.
58 CBTL - Certification Body Testing Laboratory
59 CBTL - Certification Body Testing Laboratory
Y Y60 CBTL - Certification Body Testing Laboratory
O 18/12/2015 Part01: 1 result assessed as outlying.
Y Y61 Private
O 18/12/2015 Part01: 1 result assessed as outlying.O 18/12/2015 Part02: 2 results assessed as outlying.
Y Y62 CBTL - Certification Body Testing Laboratory
O 18/12/2015 Part02: 1 result assessed as outlying.O 18/12/2015 Part01: 1 result assessed as outlying.
Y Y63 CBTL - Certification Body Testing Laboratory
O 18/12/2015 Part01: 2 results assessed as outlying.O 18/12/2015 Part02: 2 results assessed as outlying.
Y Y64 CBTL - Certification Body Testing Laboratory
O 18/12/2015 Part01: 1 result assessed as outlying.
65 CBTL - Certification Body Testing Laboratory
66 CBTL - Certification Body Testing Laboratory
67 Private
Y Y68 CBTL - Certification Body Testing Laboratory
O 18/12/2015 Part01: 1 result assessed as outlying.
69 CBTL - Certification Body Testing Laboratory
18/12/2015 Page 2 of 5Produced by IFM Quality Services Pty Ltd
WWW.IFMQS.COM.AU - Ph: +61 2 9618 3311
Lab Code Reason Withdrawn / Comment Due DateAction Alert
Glow Wire and Needle Flame - 15e29 (201505)
Electrical Program Status Report
Y Y70 CBTL - Certification Body Testing Laboratory
O 18/12/2015 Part01: 2 results assessed as outlying.
Y Y71 Private
O 18/12/2015 Part02: 1 result assessed as outlying.O 18/12/2015 Part01: 1 result assessed as outlying.
Y Y72 CBTL - Certification Body Testing Laboratory
O 18/12/2015 Part02: 1 result assessed as outlying.O 18/12/2015 Part01: 1 result assessed as outlying.
73 CBTL - Certification Body Testing Laboratory
Y Y74 CBTL - Certification Body Testing Laboratory
O 18/12/2015 Part02: 1 result assessed as outlying.O 18/12/2015 Part01: 2 results assessed as outlying.
75 CBTL - Certification Body Testing Laboratory
76 CBTL - Certification Body Testing Laboratory
Y Y77 CBTL - Certification Body Testing Laboratory
O 18/12/2015 Part01: 1 result assessed as outlying.
Y Y78 CBTL - Certification Body Testing Laboratory
O 18/12/2015 Part01: 1 result assessed as outlying.
79 CBTL - Certification Body Testing Laboratory
80 CBTL - Certification Body Testing Laboratory
81 Private
82 CBTL - Certification Body Testing Laboratory
83 CBTL - Certification Body Testing Laboratory
84 Private
Y Y85 CBTL - Certification Body Testing Laboratory
O 18/12/2015 Part01: 1 result assessed as outlying.O 18/12/2015 Part02: 1 result assessed as outlying.
86 CBTL - Certification Body Testing Laboratory
Y Y87 CBTL - Certification Body Testing Laboratory
O 18/12/2015 Part02: 1 result assessed as outlying.O 18/12/2015 Part01: 1 result assessed as outlying.
88 Private
89 CBTL - Certification Body Testing Laboratory
Y Y90 CTF - Customer’s Testing Facility
O 18/12/2015 Part01: 3 results assessed as outlying.O 18/12/2015 Part02: 3 results assessed as outlying.
91 CBTL - Certification Body Testing Laboratory
92 CBTL - Certification Body Testing Laboratory
93 Private
94 CBTL - Certification Body Testing Laboratory
95 CBTL - Certification Body Testing Laboratory
96 CTF - Customer’s Testing Facility
97 Private
98 CBTL - Certification Body Testing Laboratory
99 CBTL - Certification Body Testing Laboratory
18/12/2015 Page 3 of 5Produced by IFM Quality Services Pty Ltd
WWW.IFMQS.COM.AU - Ph: +61 2 9618 3311
Lab Code Reason Withdrawn / Comment Due DateAction Alert
Glow Wire and Needle Flame - 15e29 (201505)
Electrical Program Status Report
100 CBTL - Certification Body Testing Laboratory
101 Private
Y Y102 Private
O 18/12/2015 Part01: 1 result assessed as outlying.O 18/12/2015 Part02: 3 results assessed as outlying.
Y Y103 CBTL - Certification Body Testing Laboratory
O 18/12/2015 Part01: 1 result assessed as outlying.
Y Y104 CBTL - Certification Body Testing Laboratory
O 18/12/2015 Part01: 1 result assessed as outlying.O 18/12/2015 Part02: 1 result assessed as outlying.
105 Private
Y Y106 Private
O 18/12/2015 Part01: 1 result assessed as outlying.O 18/12/2015 Part02: 1 result assessed as outlying.
Y Y107 CBTL - Certification Body Testing Laboratory
O 18/12/2015 Part01: 3 results assessed as outlying.O 18/12/2015 Part02: 3 results assessed as outlying.
108 CBTL - Certification Body Testing Laboratory
109 CBTL - Certification Body Testing Laboratory
110 Private
Y Y111 ACTL - Associated CBTL
O 18/12/2015 Part02: 2 results assessed as outlying.O 18/12/2015 Part01: 1 result assessed as outlying.
Y Y112 CBTL - Certification Body Testing Laboratory
O 18/12/2015 Part01: 3 results assessed as outlying.O 18/12/2015 Part02: 3 results assessed as outlying.
113 CBTL - Certification Body Testing Laboratory
114 Private
115 Private
116 CBTL - Certification Body Testing Laboratory
117 CBTL - Certification Body Testing Laboratory
Y Y118 CBTL - Certification Body Testing Laboratory
O 18/12/2015 Part01: 1 result assessed as outlying.
119 Private
120 CBTL - Certification Body Testing Laboratory
121 CBTL - Certification Body Testing Laboratory
18/12/2015 Page 4 of 5Produced by IFM Quality Services Pty Ltd
WWW.IFMQS.COM.AU - Ph: +61 2 9618 3311
Lab Code Reason Withdrawn / Comment Due DateAction Alert
Glow Wire and Needle Flame - 15e29 (201505)
Electrical Program Status Report
>> Report Indicator Columns and Action History Codes
- "Y" when a submitted result was determined to be an outlier requiring corrective action(s) or additional information is required to clarify a result. Action history will then follow.- "Y" when corrective or other action(s) have not been completed- Date by which corrective or other action(s) are required to be completed. This date is not set until the FINAL report has been issued.
X - Reminder sent to laboratoryC - Action Closed (Completed)F - Overdue action referred to IEC-EE
Action History Codes:
O - Action Opened by IFMN - Notification of action sent to laboratoryA - Acknowledgement received from laboratoryR - Response received from laboratory
Column:
Action
AlertDue Date
18/12/2015 Page 5 of 5Produced by IFM Quality Services Pty Ltd
WWW.IFMQS.COM.AU - Ph: +61 2 9618 3311
Lab Code Reason Withdrawn / Comment Due DateAction Alert
Glow Wire and Needle Flame (Audit) - 15e29 Audit (201505)
Electrical Program Status Report
1 CBTL - Certification Body Testing Laboratory
2 CBTL - Certification Body Testing Laboratory
3 CBTL - Certification Body Testing Laboratory
4 CBTL - Certification Body Testing Laboratory
Y Y5 Private
O 13/08/2015 audit not received
Y6 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 10/09/2015 CAR info received and accepted.
7 CBTL - Certification Body Testing Laboratory
8 CBTL - Certification Body Testing Laboratory
Y9 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 26/08/2015 Car info received and accepted. Question misunderstood, equipment audit revised.
Y Y10 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow Wire
Y11 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 31/08/2015 CAR info received and accepted.C 31/08/2015 CAR info received.
Y12 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 19/08/2015 CAR info received and accepted.
Y13 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 7/09/2015 CAR info received and accepted.
Y14 Private
O 12/08/2015 Equipment Audit Outlier - Glow WireC 18/08/2015 Laboratory has advised previous version of standard still in use. Lab has advised inspection
and checking of glow wire will be revised and engineers retrained.This has been accepted.
Y15 CBTL - Certification Body Testing Laboratory
A 25/06/2015 Lab has advised difficulty in entering decimal places in audit form and the maximum penetration of the loop can be up to 12mm.
O 12/08/2015 Equipment Audit Outlier - Glow WireC 7/10/2015 CAR info received and accepted.
16 CBTL - Certification Body Testing Laboratory
Y17 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 4/09/2015 CAR info received and accepted. Lab advised to retain records of corrective actions for their
assessor at their next on site assessment.
18 CBTL - Certification Body Testing Laboratory
Y19 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 17/08/2015 CAR info received. Lab has advised Type N thermocouple used as it is more correct than type K.
20 CBTL - Certification Body Testing Laboratory
21 CTF - Customer’s Testing Facility
A 1/07/2015 Needle flame only in scope.
Y22 Private
O 12/08/2015 Equipment Audit Outlier - Glow Wire
18/12/2015 Page 1 of 7Produced by IFM Quality Services Pty Ltd
WWW.IFMQS.COM.AU - Ph: +61 2 9618 3311
Lab Code Reason Withdrawn / Comment Due DateAction Alert
Glow Wire and Needle Flame (Audit) - 15e29 Audit (201505)
Electrical Program Status Report
C 16/09/2015 Lab has advised equipment issues corrected.
23 CBTL - Certification Body Testing Laboratory
24 CBTL - Certification Body Testing Laboratory
Y25 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 7/09/2015 CAR info received and accepted.
26 CBTL - Certification Body Testing Laboratory
27 CBTL - Certification Body Testing Laboratory
Y28 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 21/09/2015 CAR info received and accepted. Lab has advised all items addressed.
Y29 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireA 31/08/2015 Lab has advised (CAR# 153915130) opened.C 7/10/2015 CAR info received and accepted.
30 CBTL - Certification Body Testing Laboratory
31 CBTL - Certification Body Testing Laboratory
32 CBTL - Certification Body Testing Laboratory
Y33 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 16/09/2015 CAR info received and accepted.
Y Y34 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow Wire
Y35 CTF - Customer’s Testing Facility
O 12/08/2015 Equipment Audit Outlier - Glow WireA 17/08/2015 CAR for glow wire test received and accepted.C 26/08/2015 Car info for NFT received and accepted.
36 CTF - Customer’s Testing Facility
Y37 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 16/09/2015 Car info received and accepted.
Y38 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 26/08/2015 CAR received and accepted. Insufficient knowledge of standard. Training provided.
39 CBTL - Certification Body Testing Laboratory
40 CBTL - Certification Body Testing Laboratory
Y41 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 22/10/2015 Air was interpreted to mean gas. Yes was entered for is the tip cracked as question was not
read propely.
Y42 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 17/09/2015 Glow wire now check before test instead of before testing session.
43 CBTL - Certification Body Testing Laboratory
Y44 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 26/11/2015 Question misunderstood and outlier due to typo.
Y45 CBTL - Certification Body Testing Laboratory
18/12/2015 Page 2 of 7Produced by IFM Quality Services Pty Ltd
WWW.IFMQS.COM.AU - Ph: +61 2 9618 3311
Lab Code Reason Withdrawn / Comment Due DateAction Alert
Glow Wire and Needle Flame (Audit) - 15e29 Audit (201505)
Electrical Program Status Report
O 12/08/2015 Equipment Audit Outlier - Glow WireC 17/09/2015 Lab has advised equipment issued addressed.
46 CTF - Customer’s Testing Facility
A 25/06/2015 Lab has advised they only perform glow wire test.
47 CBTL - Certification Body Testing Laboratory
Y Y48 Private
O 13/08/2015 audit not received
49 CBTL - Certification Body Testing Laboratory
50 CTF - Customer’s Testing Facility
Y51 CTF - Customer’s Testing Facility
O 12/08/2015 Equipment Audit Outlier - Glow WireC 14/09/2015 CAR info received and accepted.
Y52 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 18/08/2015 CAR info received. All outliers addressed. Accepted.
Y53 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 14/09/2015 CAR info received and accepted.
YProgram Enrolment Cancelled
54 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 15/09/2015 laboratory stated they do not do GW or needle flame tests for custoers therefore can cancel
corrective actions.
Y Y55 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow Wire
56 CTF - Customer’s Testing Facility
Y58 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireA 27/08/2015 Missing info submitted and accepted. IFM has requested CAR info for outliers.C 28/08/2015 Revised equipment audit submitted. Accepted.
Y59 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 27/08/2015 Lab has advised corrective actions completed on 20/8/15
Y60 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 2/09/2015 CAR received and accepted.
Y61 Private
O 12/08/2015 Equipment Audit Outlier - Glow WireC 2/09/2015 CAR info received and accepted. Lab has advised all corrective actions made according to
iso/iec17025 and procedures of their labroatory.
Y62 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 7/09/2015 CAR info received and accepted.
Y63 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 8/12/2015 Typo in audit entry. In future audit will be checked by technical manager.
Y64 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 19/08/2015 information received. Lab misunderstood question. Glow wire loop is inspected and cleaned
before each test.
18/12/2015 Page 3 of 7Produced by IFM Quality Services Pty Ltd
WWW.IFMQS.COM.AU - Ph: +61 2 9618 3311
Lab Code Reason Withdrawn / Comment Due DateAction Alert
Glow Wire and Needle Flame (Audit) - 15e29 Audit (201505)
Electrical Program Status Report
65 CBTL - Certification Body Testing Laboratory
Y66 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireA 19/08/2015 CAR info received. Accepted for GWT. IFM has requested review of info for NFT.C 19/08/2015 Revised CAR received and accepted.
Y67 Private
O 12/08/2015 Equipment Audit Outlier - Glow WireC 10/09/2015 Car info received and accepted.
Y68 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 26/08/2015 CAR info received. Accepted.
Y69 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 21/09/2015 CAR info received and accepted.
Y70 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 5/11/2015 CAR info received and accepted.
Y71 Private
O 12/08/2015 Equipment Audit Outlier - Glow WireC 2/09/2015 CAR info received and accepted.
Y72 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 31/08/2015 CAR info received and accepted. Error due to lack of understanding.
73 CBTL - Certification Body Testing Laboratory
74 CBTL - Certification Body Testing Laboratory
Y75 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 27/08/2015 CAR Info received. Audit questions misunderstood. Revised audit form submitted.
Y76 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Needle FlameC 17/08/2015 Lab has advised outlier due to typo. CAR info accepted.
Y77 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 21/09/2015 Incorrect translation lead to incorrect answer being submitted.
Y78 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 7/10/2015 Correct answers submitted. Most outliers were due to a misunderstanding of the questions.
79 CBTL - Certification Body Testing Laboratory
Y80 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 2/09/2015 CAR info received and accepted. Engineer did not fully understand settings of test apparatus.
Y81 Private
O 12/08/2015 Equipment Audit Outlier - Glow WireC 17/09/2015 Misinterpreted question regarding feedback mechanism.
Y82 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 26/08/2015 Car info received. Laboratory misunderstood question. Accepted.
83 CBTL - Certification Body Testing Laboratory
84 Private
18/12/2015 Page 4 of 7Produced by IFM Quality Services Pty Ltd
WWW.IFMQS.COM.AU - Ph: +61 2 9618 3311
Lab Code Reason Withdrawn / Comment Due DateAction Alert
Glow Wire and Needle Flame (Audit) - 15e29 Audit (201505)
Electrical Program Status Report
Y Y85 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow Wire
86 CBTL - Certification Body Testing Laboratory
87 CBTL - Certification Body Testing Laboratory
Y88 Private
O 12/08/2015 Equipment Audit Outlier - Glow WireC 27/08/2015 CAR info received and accepted. Lab has made changes to ensure compliance with
requirements of standard.
Y89 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 21/09/2015 CAR info received and accepted. Engineer not sufficiently trained.
Y Y90 CTF - Customer’s Testing Facility
O 12/08/2015 Equipment Audit Outlier - Glow Wire
Y91 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 31/08/2015 Loop force corrected. Training provided re checking for tip damage.
Y92 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 17/08/2015 CAR info received. Lab has advised maximum penetration reported for penetration of loop wire
but equipment set to 7mm for GWT
93 Private
Y94 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 14/09/2015 CAR info received and accepted.
95 CBTL - Certification Body Testing Laboratory
96 CTF - Customer’s Testing Facility
Y97 Private
O 12/08/2015 Equipment Audit Outlier - Glow WireA 16/09/2015 CAR info received for feedback mechanism. Accepted.C 17/09/2015 Lab has corrected all equipment issues. Accepted.
98 CBTL - Certification Body Testing Laboratory
Y99 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 17/08/2015 Car Info received. Lab has advised no feedback mechanism installed in test rig.
Y100 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 26/08/2015 CAR info received. Lab has advised new thermal coupler to be installed.
101 Private
Y Y102 Private
O 12/08/2015 Equipment Audit Outlier - Glow Wire
Y103 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 31/08/2015 CAR info received and accepted.
Y104 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 21/09/2015 Equipment recheck and found to be compliant.
Y Y105 Private
O 12/08/2015 Equipment Audit Outlier - Glow Wire
18/12/2015 Page 5 of 7Produced by IFM Quality Services Pty Ltd
WWW.IFMQS.COM.AU - Ph: +61 2 9618 3311
Lab Code Reason Withdrawn / Comment Due DateAction Alert
Glow Wire and Needle Flame (Audit) - 15e29 Audit (201505)
Electrical Program Status Report
Y Y106 Private
O 12/08/2015 Equipment Audit Outlier - Glow Wire
Y107 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 26/08/2015 CAR info received and accepted.
108 CBTL - Certification Body Testing Laboratory
109 CBTL - Certification Body Testing Laboratory
A 29/06/2015 Line 37 - the wording is confusing. The laboratory assumes that based on detail the depth of insertion of the thermocouple shall be reported.
A 29/06/2015 Line 91 - in general the confirmatory procedure is performed yearly and also after the gas supply is changed. This option is not provided in the form.
Y110 Private
O 12/08/2015 Equipment Audit Outlier - Glow WireC 26/08/2015 CAR info received and accepted.
Y111 ACTL - Associated CBTL
O 12/08/2015 Equipment Audit Outlier - Glow WireC 17/08/2015 CAR info received and accepted.
Y112 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 23/09/2015 Lab has advised corrective actions completed.
Y113 CBTL - Certification Body Testing Laboratory
R 11/08/2015 Lab has advised copper block currently unavailable.S 11/08/2015 Awaiting receipt of audit result once copper block becomes available.O 12/08/2015 Equipment Audit Outlier - Needle FlameO 12/08/2015 Equipment Audit Outlier - Glow WireA 18/08/2015 Car info received. Lab has confirmed feedback mechanism deactivated during testing.A 27/08/2015 Revised equipment audit received with missing information regarding copper block. IFM has
requested further info regarding resolution of timing device.C 17/09/2015 Revised equipment audit received. All non-conformities have been addressed.
Y114 Private
O 12/08/2015 Equipment Audit Outlier - Glow WireC 20/08/2015 CAR info received. Lab has confirmed tolerance of timing device.
115 Private
116 CBTL - Certification Body Testing Laboratory
A 25/06/2015 Lab has advised, only perform Needle flame test and that It was not possible to insert the purity of the gas supplied to the burner with 99,5 %.
Y117 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 16/09/2015 CAR info received and accepted.
Y118 CBTL - Certification Body Testing Laboratory
O 12/08/2015 Equipment Audit Outlier - Glow WireC 21/09/2015 CAR info received and accepted.
Y119 CBTL - Certification Body Testing Laboratory
O 19/10/2015 Equipment Audit Outlier - Glow WireC 19/10/2015 Outliers due to typo's.
18/12/2015 Page 6 of 7Produced by IFM Quality Services Pty Ltd
WWW.IFMQS.COM.AU - Ph: +61 2 9618 3311
Lab Code Reason Withdrawn / Comment Due DateAction Alert
Glow Wire and Needle Flame (Audit) - 15e29 Audit (201505)
Electrical Program Status Report
>> Report Indicator Columns and Action History Codes
- "Y" when a submitted result was determined to be an outlier requiring corrective action(s) or additional information is required to clarify a result. Action history will then follow.- "Y" when corrective or other action(s) have not been completed- Date by which corrective or other action(s) are required to be completed. This date is not set until the FINAL report has been issued.
X - Reminder sent to laboratoryC - Action Closed (Completed)F - Overdue action referred to IEC-EE
Action History Codes:
O - Action Opened by IFMN - Notification of action sent to laboratoryA - Acknowledgement received from laboratoryR - Response received from laboratory
Column:
Action
AlertDue Date
18/12/2015 Page 7 of 7Produced by IFM Quality Services Pty Ltd
WWW.IFMQS.COM.AU - Ph: +61 2 9618 3311