DOE SSL CALiPER Report Product Test Reference: CALiPER 08-37 T-8 Replacement Lamp DOE TEST REPORT 08-37 – SUMMARY PAGE Product Category T-8 Linear Replacement Lamp Product Description Light Emitting Designs LED T8 retrofit series Date of Test(s) CALiPER 08-37-01A & B CALiPER 08-37-02 CALiPER 08-37-03 April 17, 2008 April 4, 2008 April 7, 2008 Laboratory Performing Testing Luminaire Testing Laboratory, Inc (LTL) List of Tests Performed Spectroradiometry & Goniophotometry in situ following IESNA LM-79, temperature Total Luminaire Light Output Bare Lamp CALiPER 08-37-01A CALiPER 08-37-01B 1004 lm 1028 lm Luminaire Efficacy Bare Lamp CALiPER 08-37-01A CALiPER 08-37-01B 51.7 lm/W 52.6 lm/W Product Photo Polar candela distribution curve CALiPER 08-37-03 (2 lamps, Prismatic troffer) Comment: Goniophotometry was conducted in the context of a testing series for fluorescent tube replacement lamps (mounted in a typical T8 troffer and in a typical T12 troffer). See CALiPER Round 5 Summary Report for comparisons with other replacement tubes under same in situ conditions. DOE SSL CALiPER results may not be used for commercial purposes under any circumstances; see “No Commercial Use Policy” (http://www.netl.doe.gov/ssl/comm_testing.htm ) for more information.

Product Test Reference: CALiPER 08-37 T-8 Replacement Lamp · 2016. 9. 20. · Light Emitting Designs LED T8 retrofit series. Date of Test(s) CALiPER 08-37-01A & B CALiPER 08-37-02

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

DOE SSL CALiPER Report

Product Test Reference: CALiPER 08-37 T-8 Replacement Lamp DOE TEST REPORT 08-37 – SUMMARY PAGE Product Category T-8 Linear Replacement Lamp Product Description Light Emitting Designs LED T8 retrofit seriesDate of Test(s) CALiPER 08-37-01A & B CALiPER 08-37-02 CALiPER 08-37-03

April 17, 2008 April 4, 2008 April 7, 2008

Laboratory Performing Testing Luminaire Testing Laboratory, Inc (LTL) List of Tests Performed Spectroradiometry & Goniophotometry in situ

following IESNA LM-79, temperature Total Luminaire Light Output Bare LampCALiPER 08-37-01A CALiPER 08-37-01B

1004 lm 1028 lm

Luminaire Efficacy Bare Lamp CALiPER 08-37-01A CALiPER 08-37-01B

51.7 lm/W 52.6 lm/W

Product Photo Polar candela distribution curve CALiPER 08-37-03 (2 lamps, Prismatic troffer)

Comment: Goniophotometry was conducted in the context of a testing series for fluorescent tube replacement lamps (mounted in a typical T8 troffer and in a typical T12 troffer). See CALiPER Round 5 Summary Report for comparisons with other replacement tubes under same in situ conditions.

DOE SSL CALiPER results may not be used for commercial purposes under any circumstances; see “No Commercial Use Policy” (http://www.netl.doe.gov/ssl/comm_testing.htm) for more information.

CALiPER 08-37 T8 Linear Replacement Lamp Page 2

Product Test References CALiPER Test Number 08-37 Manufacturer’s name Light Emitting Designs, LLC Product Category LED T8 Linear Replacement Product Description T8 4 Feet LED Fluorescent Light Tube

L.E.D. LLC Catalog Number LED8-PL48-120-12-300-DL Acquisition Notes Purchased direct from manufacturer

LED Lamp/Package References (Light Emitting Designs datasheet 081507_T8) LED Lamp/Package Number of LEDs 300 white LEDs Manufacturer Luminaire CCT 4000 - 4500K or 5400K (two different

mentions of CCT on product literature for this model)

Manufacturer Luminaire CRI 85 Manufacturer Luminaire Output 975 lm Manufacturer Luminaire Efficacy 70 lm/W Manufacturer Luminaire Power Rating 12W Manufacturer Luminaire Driver Description

“…can impact energy usage and costs in a major way. Absolutely no danger of mercury vapor contamination.”

Manufacturer Published Product Life 50,000 hrs

DOE SSL CALiPER results may not be used for commercial purposes under any circumstances; see “No Commercial Use Policy” (http://www.netl.doe.gov/ssl/comm_testing.htm) for more information.

CALiPER 08-37 T8 Linear Replacement Lamp Page 3

Lensed Housing (Troffer) References –for Goniophotometry 08-17-02 (Lithonia catalog and cutsheet, 4/2008) Lensed Troffer Manufacturer, Model, Description (08-37-02)

Lithonia Lighting, (T12) housing with clear prismatic lens #A12, Universal Lighting Technologies 446-SLH-TC-P ballast.

49051 – 2GT 2 40 A12 120 ES Manufacturer Published Troffer Luminaire Efficiency (08-37-02)

71% (with two F40T12, 3150 lm lamps)

Manufacturer Published Luminaire Output (08-37-02)

4492 lm (with two F40T12, 3150 lm lamps)

Parabolic Housing (Troffer) References –for Goniophotometry 08-17-03 (Lithonia catalog and cutsheet, 4/2008) Parabolic Troffer Manufacturer, Model, Description (08-37-03)

Lithonia Lighting, 105XF8 (T8) housing with parabolic semi-specular aluminum louver, Sylvania QTP2X32T8/UNV ISN-SC ballast

2PM3N G B 2 32 12LD MVOLT GEB10IS Manufacturer Published Troffer Luminaire Efficiency (08-37-03)

76% (with two F32T8/SPX35 lamps)

Manufacturer Published Luminaire Output (08-37-03)

4430 lm (with two F32T8/SPX35 lamps)

DOE SSL CALiPER results may not be used for commercial purposes under any circumstances; see “No Commercial Use Policy” (http://www.netl.doe.gov/ssl/comm_testing.htm) for more information.

CALiPER 08-37 T8 Linear Replacement Lamp Page 4

Summary of Luminaire Tests Conducted CALiPER Test Label(s) CALiPER 08-37-01A Laboratory Performing Testing LTL Testing Lab. Internal Reference #12925 Date Testing Commenced April 17, 2008 List of Tests Performed Spectroradiometry following IESNA LM-79 Comment This product requires no ballast..

CALiPER Test Label(s) CALiPER 08-37-01B Laboratory Performing Testing LTL Testing Lab. Internal Reference #12926 Date Testing Commenced April 17, 2008 List of Tests Performed Spectroradiometry following IESNA LM-79 Comment This product requires no ballast.

CALiPER Test Label(s) CALiPER 08-37-02 Laboratory Performing Testing LTL Testing Lab. Internal Reference #12927 Date Testing Commenced April 4, 2008 List of Tests Performed Goniophotometry following IESNA LM-79 Comment Testing conducted with samples 08-37 A and B

installed in Lithonia Troffer 49051 (T12) housing with clear prismatic lens, with the ballast disabled.

CALiPER Test Label(s) CALiPER 08-37-03 Laboratory Performing Testing LTL Testing Lab. Internal Reference #12928 Date Testing Commenced April 7, 2008 List of Tests Performed Goniophotometry following IESNA LM-79 Comment Testing conducted with samples 08-37 A and B

installed in Lithonia Troffer 105XF8 (T8) housing with parabolic semi-specular aluminum louver, with the ballast disabled.

DOE SSL CALiPER results may not be used for commercial purposes under any circumstances; see “No Commercial Use Policy” (http://www.netl.doe.gov/ssl/comm_testing.htm) for more information.

CALiPER 08-37 T8 Linear Replacement Lamp Page 5

Description of Luminaire Tested Product Photos

Test Configuration Description Comments regarding product geometry and mounting:

Testing in integrating sphere (08-37-01 A & B) was performed on each lamp separately using 120VAC.. Testing in goniophotometry was performed with lamps mounted in troffers representative of fixtures in which these replacements could be installed. 08-17-02 was performed with lamps A & B installed in a lensed troffer with the ballast disabled. 08-17-03 was performed with lamps A & B installed in a parabolic troffer with the ballast disabled.

DOE SSL CALiPER results may not be used for commercial purposes under any circumstances; see “No Commercial Use Policy” (http://www.netl.doe.gov/ssl/comm_testing.htm) for more information.

CALiPER 08-37 T8 Linear Replacement Lamp Page 6

Measured Photometric Quantities -- Test Results: Light Output & Efficacy Light Output & Efficacy Value

Test identifier

Measured Radiant Flux (milliwatts)

Measured Power

(Watts)

Total Measured Light Output

(lumens)

Calculated Luminaire Efficacy

(lumens/Watt)

CALiPER 08-37-01A 3277 mW 19.4 W 1004 lm 51.7 lm/W CALiPER 08-37-01B 3362 mW 19.6 W 1028 lm 52.6 lm/W

Measured Photometric Quantities -- Test Results: Color Metrics Color Quantities

Chromaticity Coordinates

Chromaticity Coordinates

Correlated Color Temperature

(CCT)

Test identifier

x y u’ v’ CALiPER 08-37-01A 0.2958 0.3147 0.1913 0.4580 7727 CALiPER 08-37-01B 0.2955 0.3145 0.1912 0.4578 7751

Spectral Power Distribution Curves CALiPER 08-37-01A

05

10

15202530

3540

350 400 450 500 550 600 650 700 750 800 850Wavelength (nm)

Spec

tral

Flu

x (m

W/n

m)

CALiPER 08-37-01B

05

10

15202530

3540

350 400 450 500 550 600 650 700 750 800 850Wavelength (nm)

Spec

tral

Flu

x (m

W/n

m)

DOE SSL CALiPER results may not be used for commercial purposes under any circumstances; see “No Commercial Use Policy” (http://www.netl.doe.gov/ssl/comm_testing.htm) for more information.

CALiPER 08-37 T8 Linear Replacement Lamp Page 7

DOE SSL CALiPER results may not be used for commercial purposes under any circumstances; see “No Commercial Use Policy” (http://www.netl.doe.gov/ssl/comm_testing.htm) for more information.

Chromaticity Diagram CALiPER 08-37-01A CALiPER 08-37-01B

CALiPER 08-37 T8 Linear Replacement Lamp Page 8

DOE SSL CALiPER results may not be used for commercial purposes under any circumstances; see “No Commercial Use Policy” (http://www.netl.doe.gov/ssl/comm_testing.htm) for more information.

Color Quality CRI values are reported below with the reminder that, in certain cases, a light source may be acceptable, and even preferred, by users for given applications even though its CRI value is relatively low. Readers are urged to be aware of the complexities of assessing color quality and aware of the limitations of CRI with regard to SSL technologies.1,2 Alternative metrics are under development. In the meantime, qualitative visual assessment by human observers may provide additional insight regarding the suitability of color quality of a luminaire for a given application.

Test identifier Color Rendering Index (CRI)

CALiPER 08-37-01A 75.9 CALiPER 08-37-01B 76.1

Relative Spectral Power Distribution Overlays

0.0

0.5

1.0

350 400 450 500 550 600 650 700 750 800

Wavelength (nm)

Rel

ativ

e (o

ut o

f 1)

LED CALiPER 08-37A7727K

LED CALiPER 07-56A3485KLED CALiPER 08-17A15044KLED CALiPER 08-19A2952KFluorescent CALiPER08-28A 3932K

1 Protzman, J. Brent and Kevin W. Houser. October 2006. LEDs for General Illumination: The State of the Science. Leukos. Vol. 3, No. 2, pp. 121-142. 2 Narendran N, Deng L. 2002. Color rendering properties of LED light sources. Proc. of SPIE: Solid State Lighting II.

CALiPER 08-37 T8 Linear Replacement Lamp Page 9

DOE SSL CALiPER results may not be used for commercial purposes under any circumstances; see “No Commercial Use Policy” (http://www.netl.doe.gov/ssl/comm_testing.htm) for more information.

Test Configuration Description: In Situ Testing in Goniophotometer Lensed Housing Schematic of product and/or mounting configuration for “in situ” testing in goniophotometer

CALiPER 08-37-02

CALiPER 08-37-02

Comments

Testing conducted with samples 08-37 A and B installed in Lithonia Troffer 49051 (T12) housing with clear prismatic lens, with ballast disabled.

Temperature Measurement CALiPER 08-37-02 Temperature Temperature 08-37A 40.8°C Temperature 08-37B 41.7°C

CALiPER 08-37 T8 Linear Replacement Lamp Page 10

Light Output & Efficacy Values “In Situ” (2 lamps in Lensed Housing) Test identifier

Measured Power

(Watts)

Total Measured Light Output

(lumens)

Calculated Luminaire Efficacy

(lumens/Watt)

CALiPER 08-37-02 39.5 W 1693 lm 42.8 lm/W Beam & Intensity “In Situ” (2 lamps in Lensed Housing) Test identifier: CALiPER 08-37-02 0° Measured Center Beam Candle Power (Candelas) 862 cd

ZONAL LUMEN SUMMARY

ZONE LUMENS %FIXTURE 0-30 636 37.57 0-40 990 58.47 0-60 1490 88.00 0-90 1693 100.00

Intensity Distribution Plots “In Situ” (2 lamps in Lensed Housing) Polar candela distribution curve CALiPER 08-37-02

DOE SSL CALiPER results may not be used for commercial purposes under any circumstances; see “No Commercial Use Policy” (http://www.netl.doe.gov/ssl/comm_testing.htm) for more information.

CALiPER 08-37 T8 Linear Replacement Lamp Page 11

Isofootcandle plot Mounting height 5 feet “In Situ” (2 lamps in Lensed Housing) CALiPER 08-37-02

LTL REPORT NUMBER 12927-ISOFOOTCANDLE VALUES ARE BASED ON A MOUNTING HEIGHT OF 5', WITH THE LUMINAIRE LOCATED AT 0,0. ISOFOOTCANDLE VALUES ARE CALCULATED FROM THE DIRECT CONTRIBUTION FROM THE LUMINAIRE ONLY. WALL, CEILING, AND FLOOR CONTRIBUTION IS NOT INCLUDED.

DOE SSL CALiPER results may not be used for commercial purposes under any circumstances; see “No Commercial Use Policy” (http://www.netl.doe.gov/ssl/comm_testing.htm) for more information.

CALiPER 08-37 T8 Linear Replacement Lamp Page 12

DOE SSL CALiPER results may not be used for commercial purposes under any circumstances; see “No Commercial Use Policy” (http://www.netl.doe.gov/ssl/comm_testing.htm) for more information.

Test Configuration Description: In Situ Testing in Goniophotometer Parabolic Housing Schematic of product and/or mounting configuration for “in situ” testing in goniophotometer

CALiPER 08-37-03

CALiPER 08-37-03

Comments

Testing conducted with samples 08-37 A and B installed in Lithonia 2’ x 4’ Troffer (nominal T8) housing with parabolic semi-specular aluminum (12 cells) louver, no-air handling, with ballast disabled.

Temperature Measurement CALiPER 08-37-03 Temperature Temperature 08-37A 41.0°C Temperature 08-37B 39.7°C

CALiPER 08-37 T8 Linear Replacement Lamp Page 13

Light Output & Efficacy Values “In Situ” (2 lamps in Parabolic Housing) Test identifier

Measured Power

(Watts)

Total Measured Light Output

(lumens)

Calculated Luminaire Efficacy

(lumens/Watt)

CALiPER 08-37-03 39.5 W 1711 lm 43.4 lm/W Beam & Intensity “In Situ” (2 lamps in Parabolic Housing) Test identifier: CALiPER 08-37-03 Measured Center Beam Candle Power (Candelas) 785 cd

ZONAL LUMEN SUMMARY

ZONE LUMENS %FIXTURE 0-30 582 34.02 0-40 930 54.35 0-60 1590 92.94 0-90 1711 100.00

Intensity Distribution Plots “In Situ” (2 lamps in Parabolic Housing) Polar candela distribution curve CALiPER 08-37-03

DOE SSL CALiPER results may not be used for commercial purposes under any circumstances; see “No Commercial Use Policy” (http://www.netl.doe.gov/ssl/comm_testing.htm) for more information.

CALiPER 08-37 T8 Linear Replacement Lamp Page 14

Isofootcandle plot Mounting height 5 feet “In Situ” (2 lamps in Parabolic Housing) CALiPER 08-37-03

LTL REPORT NUMBER 12928-ISOFOOTCANDLE VALUES ARE BASED ON A MOUNTING HEIGHT OF 5', WITH THE LUMINAIRE LOCATED AT 0,0. ISOFOOTCANDLE VALUES ARE CALCULATED FROM THE DIRECT CONTRIBUTION FROM THE LUMINAIRE ONLY. WALL, CEILING, AND FLOOR CONTRIBUTION IS NOT INCLUDED.

DOE SSL CALiPER results may not be used for commercial purposes under any circumstances; see “No Commercial Use Policy” (http://www.netl.doe.gov/ssl/comm_testing.htm) for more information.

CALiPER 08-37 T8 Linear Replacement Lamp Page 15

Appendix A: Description of Test Conditions, Test Configuration, Test Procedures Photometric equipment used

CALiPER 08-37-01A&B Spectroradiometer

Integrating sphere with spectroradiometer

CALiPER 08-37-02 Goniophotometer

DATA WAS ACQUIRED USING THE CALIBRATED PHOTODETECTOR METHOD OF ABSOLUTE PHOTOMETRY. A UDT MODEL #211 PHOTODETECTOR AND UDT MODEL #S370 OPTOMETER COMBINATION WERE USED AS A STANDARD. A SPECTRAL MISMATCH CORRECTION FACTOR WAS EMPLOYED BASED ON THE SPECTRAL RESPONSIVITY OF THE PHOTODETECTOR AND THE SPECTRAL POWER DISTRIBUTION OF THE TEST SUBJECT.

THIS TEST WAS CONDUCTED USING RELATIVE PHOTOMETRY TECHNIQUES ACCORDING TO STANDARD IESNA PROCEDURES. THE USER MUST THEREFORE USE CAUTION IN THE FOLLOWING SITUATIONS: 1) THIS TEST WAS PERFORMED USING A SPECIFIC BALLAST/LAMP COMBINATION. EXTRAPOLATION OF THESE DATA FOR OTHER BALLAST/LAMP COMBINATIONS MAY PRODUCE ERRONEOUS RESULTS. 2) ACCORDING TO IESNA PROCEDURES, THE BALLAST(S) AND LAMP(S) ARE PRESUMED TO PRODUCE 100% OF RATED OUTPUT. AN APPROPRIATE BALLAST FACTOR MUST BE APPLIED TO THE LUMEN OUTPUT RATINGS AND LUMINOUS INTENSITY VALUES GIVEN. 3) THIS TEST WAS CONDUCTED IN A CONTROLLED LABORATORY ENVIRONMENT WHERE THE AMBIENT TEMPERATURE WAS HELD AT 25°C ±1°C. FIELD PERFORMANCE MAY DIFFER PARTICULARLY IN REGARDS TO CHANGE IN LUMINOUS OUTPUT AS A RESULT OF DIFFERENCE IN AMBIENT TEMPERATURE AND METHOD OF MOUNTING THE LUMINAIRE.

DOE SSL CALiPER results may not be used for commercial purposes under any circumstances; see “No Commercial Use Policy” (http://www.netl.doe.gov/ssl/comm_testing.htm) for more information.

CALiPER 08-37 T8 Linear Replacement Lamp Page 16

DOE SSL CALiPER results may not be used for commercial purposes under any circumstances; see “No Commercial Use Policy” (http://www.netl.doe.gov/ssl/comm_testing.htm) for more information.

Verification that test conditions comply with LM-79

CALiPER 08-37-01A Measured value & units

Measured ambient temperature 25°C

Measured voltage during testing 120 V

Measured current during testing 0.3071 A

Measured power during testing 19.42 W

Calculated power factor during testing 0.53

CALiPER 08-37-01B Measured value & units

Measured ambient temperature 25°C

Measured voltage during testing 120 V

Measured current during testing 0.3071 A

Measured power during testing 19.56 W

Calculated power factor during testing 0.53

CALiPER 08-37-02 (2 lamps) Measured value & units

Measured ambient temperature 25°C

Measured voltage during testing 120 V

Measured current during testing 0.6144 A

Measured power during testing 39.52 W

Calculated power factor during testing 0.54

CALiPER 08-37-03 (2 lamps) Measured value & units

Measured ambient temperature 25°C

Measured voltage during testing 120 V

Measured current during testing 0.6125 A

Measured power during testing 39.46 W

Calculated power factor during testing 0.54

CALiPER 08-37 T8 Linear Replacement Lamp Page 17

Tabulated Spectral Power Distribution (CALiPER 08-37-01A)

Wavelength (nm)

Spectral Flux (mW/nm)

Wavelength (nm)

Spectral Flux (mW/nm)

350 0.1036 610 10.5590 360 0.0876 620 8.8788 370 0.1088 630 7.3293 380 0.1421 640 5.9191 390 0.1488 650 4.7146 400 0.1788 660 3.7288 410 0.4472 670 2.9278 420 2.0684 680 2.2805 430 7.4385 690 1.7465 440 18.8010 700 1.3381 450 34.6990 710 1.0211 460 25.5220 720 0.7588 470 13.4750 730 0.6379 480 7.7004 740 0.4371 490 6.2148 750 0.3687 500 7.8657 760 0.3179 510 11.2470 770 0.2630 520 14.3440 780 0.1779 530 15.9920 790 0.1992 540 16.7560 800 0.1478 550 16.7360 810 0.0548 560 16.5170 820 0.1039 570 15.8840 830 0.0481 580 14.9030 840 0.1149 590 13.7570 850 0.2783 600 12.2370

DOE SSL CALiPER results may not be used for commercial purposes under any circumstances; see “No Commercial Use Policy” (http://www.netl.doe.gov/ssl/comm_testing.htm) for more information.

CALiPER 08-37 T8 Linear Replacement Lamp Page 18

(CALiPER 08-37-01B)

Wavelength (nm)

Spectral Flux (mW/nm)

Wavelength (nm)

Spectral Flux (mW/nm)

350 0.1075 610 10.8040 360 0.1103 620 9.0994 370 0.1292 630 7.5017 380 0.1475 640 6.0574 390 0.1389 650 4.8227 400 0.1678 660 3.8057 410 0.4605 670 3.0019 420 2.1104 680 2.3262 430 7.6049 690 1.7687 440 18.7750 700 1.3395 450 35.0520 710 1.0782 460 27.0280 720 0.8168 470 14.1710 730 0.6272 480 8.0435 740 0.4914 490 6.4283 750 0.4251 500 8.0669 760 0.2802 510 11.5000 770 0.2729 520 14.6690 780 0.2349 530 16.4000 790 0.1378 540 17.1690 800 0.1576 550 17.1490 810 0.1122 560 16.9110 820 0.0823 570 16.2650 830 0.2170 580 15.2680 840 0.2385 590 14.0570 850 0.2888 600 12.5430

DOE SSL CALiPER results may not be used for commercial purposes under any circumstances; see “No Commercial Use Policy” (http://www.netl.doe.gov/ssl/comm_testing.htm) for more information.

CALiPER 08-37 T8 Linear Replacement Lamp Page 19

APPENDIX B: IES Summary Data 08-37-02 (2 Lamps, T12 Housing, Prismatic Lens)

CALiPER 08-37-02

IES HEADER ENTRIES

LTL NUMBER: 12927 DATE: 04-04-2008 CATALOG NUMBER: CALIPER 08-31 WITH 08-37A AND 08-37B LUMINAIRE: FORMED WHITE ENAMEL STEEL HOUSING, CLEAR PRISMATIC PLASTIC

LENS. LAMP: TWO LED REPLACEMENT LINEAR LAMPS ELECTRICAL VALUES: 120.0VAC, 0.6144A, 39.52W LUMINAIRE EFFICACY: 42.8 LUMENS/WATT MOUNTING: RECESSED NOTE: THIS TEST WAS PERFORMED USING THE CALIBRATED PHOTODETECTOR

METHOD OF ABSOLUTE PHOTOMETRY. VERTICAL TEST DATA WAS ACQUIRED IN ½ DEGREE INCREMENTS.

LUMINAIRE CANDELA DISTRIBUTION

LATERAL ANGLE VER ANG 0.0 22.5 45.0 67.5 90.0

ZONAL LUMENSUMS

0 862 862 862 862 862 5 850 854 852 855 853 81 15 807 811 808 810 805 227 25 715 720 715 715 712 328 35 568 570 569 571 568 354 45 385 388 390 391 391 300 55 222 222 222 222 218 200 65 111 106 87 104 111 107 75 80 65 53 65 80 67 85 26 24 23 27 32 28 90 2 0 1 2 3

ZONAL LUMEN SUMMARY DATA IN MAJOR ZONES

ZONE LUMENS %FIXT 0-30 636 37.57 0-40 990 58.47 0-60 1490 88.00 0-90 1693 100.00 90-180 0 0.00 0-180 1693 100.00

TOTAL LUMEN OUTPUT: 1693 CIE CLASSIFICATION TYPE DIRECT LUMINAIRE SPACING CRITERION 0-DEG 1.18 LUMINAIRE SPACING CRITERION 90-DEG 1.18 DIMENSION NORMAL TO 0-DEG PLANE 21.125" DIMENSION PARALLEL 0-DEG PLANE 45.375"

DOE SSL CALiPER results may not be used for commercial purposes under any circumstances; see “No Commercial Use Policy” (http://www.netl.doe.gov/ssl/comm_testing.htm) for more information.

CALiPER 08-37 T8 Linear Replacement Lamp Page 20

DOE SSL CALiPER results may not be used for commercial purposes under any circumstances; see “No Commercial Use Policy” (http://www.netl.doe.gov/ssl/comm_testing.htm) for more information.

5 DEG LUMEN VALUES

ZONE LUMENS 0-5 20 5-10 60 10-15 97 15-20 130 20-25 155 25-30 172 30-35 179 35-40 175 40-45 161 45-50 139 50-55 114 55-60 86 60-65 62 65-70 46 70-75 37 75-80 30 80-85 21 85-90 8

LUMINAIRE AVERAGE BRIGHTNESS DATA TABLE

LUMINANCE IN CANDELA PER SQUARE METER

LATERAL ANGLE VER ANG 0-DEG 45-DEG 90-DEG 0 1395 1395 1395 45 880 892 893 55 627 627 614 65 425 333 425 75 498 333 498 85 474 418 585

5 DEGREE LUMINAIRE CANDELA DISTRIBUTION

LATERAL ANGLE VER ANG 0.0 22.5 45.0 67.5 90.0 0 862 862 862 862 862 5 850 854 852 855 853 10 837 840 837 840 835 15 807 811 808 810 805 20 765 769 765 768 765 25 715 720 715 715 712 30 648 650 648 649 646 35 568 570 569 571 568 40 481 482 480 482 482 45 385 388 390 391 391 50 294 301 306 303 300 55 222 222 222 222 218 60 156 154 146 152 153 65 111 106 87 104 111 70 92 80 62 80 92 75 80 65 53 65 80 80 56 46 45 47 54 85 26 24 23 27 32 90 2 0 1 2 3

Product Test Reference: CALiPER 08-37 T8 Replacement Lamp Page 21

APPENDIX B: IES Summary Data 08-37-03 (2 Lamps T8 Parabolic Louvered Troffer)

CALiPER 08-37-03

IES HEADER ENTRIES

LTL NUMBER: 12928 DATE: 04-07-2008 CATALOG NUMBER: CALIPER 08-32 WITH 08-37A AND 08-37B LUMINAIRE: FORMED WHITE ENAMEL STEEL HOUSING, 12 CELL, 2-5/8" DEEP,FORMED

SEMI-SPECULAR ALUMINUM LOUVER. LAMP: TWO LED REPLACEMENT LINEAR LAMPS ELECTRICAL VALUES: 120.0VAC, 0.6125A, 39.46W LUMINAIRE EFFICACY: 43.4 LUMENS/WATT MOUNTING: RECESSED NOTE: THIS TEST WAS PERFORMED USING THE CALIBRATED PHOTODETECTOR

METHOD OF ABSOLUTE PHOTOMETRY. VERTICAL TEST DATA WAS ACQUIRED IN ½ DEGREE INCREMENTS.

LUMINAIRE CANDELA DISTRIBUTION

LATERAL ANGLE VER ANG 0.0 22.5 45.0 67.5 90.0

ZONAL LUMENSUMS

0 785 785 785 785 785 5 771 779 778 782 780 74 15 727 734 734 734 740 207 25 651 650 651 661 667 302 35 542 542 549 570 587 348 45 410 413 439 490 531 352 55 264 283 354 445 453 308 65 107 145 142 44 38 104 75 14 14 10 12 11 15 85 2 2 2 2 2 2 90 0 0 0 0 0

ZONAL LUMEN SUMMARY DATA IN MAJOR ZONES

ZONE LUMENS %FIXT 0-30 582 34.02 0-40 930 54.35 0-60 1590 92.94 0-90 1711 100.00 90-180 0 0.00 0-180 1711 100.00

TOTAL LUMEN OUTPUT: 1711 CIE CLASSIFICATION TYPE DIRECT LUMINAIRE SPACING CRITERION 0-DEG 1.19 LUMINAIRE SPACING CRITERION 90-DEG 1.25 DIMENSION NORMAL TO 0-DEG PLANE 20.750" DIMENSION PARALLEL 0-DEG PLANE 44.875"

DOE SSL CALiPER results may not be used for commercial purposes under any circumstances; see “No Commercial Use Policy” (http://www.netl.doe.gov/ssl/comm_testing.htm) for more information.

Product Test Reference: CALiPER 08-37 T8 Replacement Lamp Page 22

DOE SSL CALiPER results may not be used for commercial purposes under any circumstances; see “No Commercial Use Policy” (http://www.netl.doe.gov/ssl/comm_testing.htm) for more information.

5 DEG LUMEN VALUES

ZONE LUMENS 0-5 19 5-10 55 10-15 89 15-20 118 20-25 142 25-30 160 30-35 171 35-40 176 40-45 177 45-50 175 50-55 171 55-60 137 60-65 72 65-70 31 70-75 11 75-80 4 80-85 2 85-90 0

LUMINAIRE AVERAGE BRIGHTNESS DATA TABLE

LUMINANCE IN CANDELA PER SQUARE METER

LATERAL ANGLE VER ANG 0-DEG 45-DEG 90-DEG 0 1307 1307 1307 45 966 1034 1249 55 767 1027 1313 65 423 561 148 75 92 66 72 85 35 29 38

5 DEGREE LUMINAIRE CANDELA DISTRIBUTION

LATERAL ANGLE VER ANG 0.0 22.5 45.0 67.5 90.0 0 785 785 785 785 785 5 771 779 778 782 780 10 758 764 758 762 764 15 727 734 734 734 740 20 694 699 694 698 703 25 651 650 651 661 667 30 600 604 597 616 629 35 542 542 549 570 587 40 479 480 497 522 548 45 410 413 439 490 531 50 339 346 391 487 548 55 264 283 354 445 453 60 185 217 293 208 98 65 107 145 142 44 38 70 40 59 25 23 21 75 14 14 10 12 11 80 5 5 4 5 5 85 2 2 2 2 2 90 0 0 0 0 0