Computing Division ALBA Project Document No: Prepared by: Created: 26/Feb/2008 Page: 1 of 11 CCD-GDVC-OP-0001 Sergi Rubio Manrique Modified: 28/July/2009 Rev. No.: 2.1 Procedure to Configure an MKS Residual Gas Analyzer Abstract This document describes the steps to follow to connect an MKS Microvision-IP RGA equipment to a Vacuum laptop, using a DHCP server for IP address resolution. Contents 1 Reloading RGA Factory Default Configuration ............................................................................. 2 1.1 Configuration of DHCP ................................................................................................... 2 1.2 Loading RGA Defaults .................................................................................................... 3 1.3 Configuring a new connection in Process Eye...................................................................... 5 2 Changing the RGA IP Address ................................................................................................. 6 2.1 Using the RGA Device Manager to change the IP address of a device ...................................... 6 2.2 Using RGA Device Manager Troubleshooting to force the detection of a new device. ................... 6 3 Troubleshooting .................................................................................................................... 8 4 Appendix: Identify the device IP Address ................................................................................... 10 4.1 For a device to be connected to the tunnel network (Storage Ring and/or Front Ends) ................. 10 4.2 For a device used exclusively in the Vacuum Lab or during Installation (connected directly to one of the Vacuum laptops) ................................................................................................... 10

Procedure to Configure an MKS Residual Gas Analyzer

  • Upload
    others

  • View
    5

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Procedure to Configure an MKS Residual Gas Analyzer

ALBA Project Document No: Page: 1 of 11

CCD-GDVC-OP-0001 Rev. No.: 2.1

Computing Division

ALBA Project Document No: Prepared by: Created: 26/Feb/2008 Page: 1 of 11

CCD-GDVC-OP-0001 Sergi Rubio Manrique Modified: 28/July/2009 Rev. No.: 2.1

Procedure to Configure an MKS Residual GasAnalyzer

Abstract

This document describes the steps to follow to connect an MKS Microvision-IP RGA equipment to a Vacuumlaptop, using a DHCP server for IP address resolution.

Contents1 Reloading RGA Factory Default Configuration ............................................................................. 2

1.1 Configuration of DHCP ................................................................................................... 21.2 Loading RGA Defaults .................................................................................................... 31.3 Configuring a new connection in Process Eye...................................................................... 5

2 Changing the RGA IP Address ................................................................................................. 62.1 Using the RGA Device Manager to change the IP address of a device ...................................... 62.2 Using RGA Device Manager Troubleshooting to force the detection of a new device. ................... 6

3 Troubleshooting .................................................................................................................... 8

4 Appendix: Identify the device IP Address ................................................................................... 104.1 For a device to be connected to the tunnel network (Storage Ring and/or Front Ends) ................. 104.2 For a device used exclusively in the Vacuum Lab or during Installation (connected directly to one

of the Vacuum laptops) ................................................................................................... 10

Page 2: Procedure to Configure an MKS Residual Gas Analyzer

ALBA Project Document No: Page: 2 of 11

CCD-GDVC-OP-0001 Rev. No.: 2.1

1 Reloading RGA Factory Default Configuration• First of all, I advice to reboot the laptop before doing an RGA Scan. RGA Server installed by MKS Software

use to hung up while the laptop is hibernated.

• When login, use operatorvc user. All needed applications are available in the desktop.

1.1 Configuration of DHCPThis steps are needed for ethernet communications, if you are using serial cable you should jump tothe next section (Loading RGA Defaults).

• DHCP configuration is used to communicate with any RGA withoug modifying your CELLS laptop config-uration; so it is assumed that your laptop is configured to get an IP address using DHCP. If you are not sureabout this issue then crosscheck it first,.

• With ethernet unplug, launch the tftp32 application. The application should be available in your Desktop orin the Start Menu (Inicio)

If it is not the case, the installer can be found at:

http://www.cells.es/Intranet/Divisions/Computing/Controls/Help/Vacuum/tftpd32

• To communicate with a test/laboratory equipment the configuration should match exactly what is shown inthe image. If you have to modify a previous one don’t forget to press the SAVE button after typing the newvalues.

• Remember to clean up any previous connections stored in the program:

Page 3: Procedure to Configure an MKS Residual Gas Analyzer

ALBA Project Document No: Page: 3 of 11

CCD-GDVC-OP-0001 Rev. No.: 2.1

WARNING: DO NOT CLOSE TDFTPD32 UNTIL YOU FINNISH YOUR RGA SCAN

1.2 Loading RGA Defaults• Unplug the device power unit and plug it again.

• Press the RESET button

• And then press PRESET button; keep it pressed during 20 seconds until the STATUS led blinks a couple oftimes.

• If the STATUS doesn’t blink, unplug/replug the device and repeat the previous steps.

• If the STATUS blink two times, then the default configuration is loaded.

• Connect the laptop using a cross-over ethernet cable (a Grey cable with Yellow connectors at both sides)

• Unplug the RGA device power unit and plug it again.

• Wait 1 minute ...

– during this time the Network Icon of the Laptop will change from this shape:

– to this one:

Page 4: Procedure to Configure an MKS Residual Gas Analyzer

ALBA Project Document No: Page: 4 of 11

CCD-GDVC-OP-0001 Rev. No.: 2.1

• To see if the connection succeds, you have to wait until the Network Icon in the laptop stops moving; therecould be several retries.

• The RGA device should appear in the RGA Device Manager panel; if it doesn’t appear you should repeatthe reset/preset/unplug/replug process.

• When closing the application you will get this menu, in which you have to select exclusive access for allinstruments:

• Now you should be able to open the device from ProcessEye, create a new connection an perform anEasyView RGA scan.

NOTE

Even if the device appears in red color in the RGA device manager, it should be accessible fromProcessEye.

But sometimes the RGA manager said that it was needed to download a new program to the device. Ifthat’s your case, DON’T DOWNLOAD ANYTHING, just do a reboot (unplug power / plug power)and it solves the problem.

Page 5: Procedure to Configure an MKS Residual Gas Analyzer

ALBA Project Document No: Page: 5 of 11

CCD-GDVC-OP-0001 Rev. No.: 2.1

1.3 Configuring a new connection in Process Eye• ProcessEye allows to add a new IP to connect manually:

• The IP to be set will be the one identified in Chapter 1 (the example shows the IP for Vacuum Lab devices)

• If the device is detected, it will appear in the following screen. If the device doesn’t appear under the IP, thenyou will have to execute the RGA Device Manager Troubleshooting.

• After you select your device you will have to close Process Eye and open it again.

• After you restart ProcessEye the device should appear in the bottom of the screen.

Page 6: Procedure to Configure an MKS Residual Gas Analyzer

ALBA Project Document No: Page: 6 of 11

CCD-GDVC-OP-0001 Rev. No.: 2.1

2 Changing the RGA IP Address• The best method for forcing communication with new/test devices is loading RGA configuration defaults.

• Use a new IP instead of defaults when you want to setup an RGA in its permanent location.

• IP’s for all permanent RGA’s are listed at the end of this document.

2.1 Using the RGA Device Manager to change the IP address of a device• You can enter a new address manually:

• For test/lab units it is recommended to use the default values: IP = 192.168.0.250; Mask = 255.255.252.0

2.2 Using RGA Device Manager Troubleshooting to force the detection of a new device.• It will be needed with Serial devices or some devices that cannot be connected even after reloading defaults.

• If you don’t detect the device, you can follow the troubleshooting

• Introduce RGA serial number

Page 7: Procedure to Configure an MKS Residual Gas Analyzer

ALBA Project Document No: Page: 7 of 11

CCD-GDVC-OP-0001 Rev. No.: 2.1

• Introduce the IP of the device. Use 192.168.0.250 for test/laboratory equipments or, if configuring the deviceto be permanently installed, the IP detailed in appendix 1.

• Check if STATUS led is blinking

• The RGA will be restarted and an error could appear when finnishing the process; don’t care.

• Launch ProcessEye and try to connect directly to the device.

• If didn’t worked, you could need to unplug/replug the ethernet cable before opening RGA Device Manageragain.

• If none of these operations work you can try:

– Detect the RGA using the RGA Device Manager Troubleshooting.

– Detect the RGA using ProcessEye->Instruments->AddConnection

– Create a trouble ticket in the Controls queue at http://rt.cells.es.

Page 8: Procedure to Configure an MKS Residual Gas Analyzer

ALBA Project Document No: Page: 8 of 11

CCD-GDVC-OP-0001 Rev. No.: 2.1

3 TroubleshootingBefore proceeding with other operations, ask yourself this questions:

Is the laptop configured to use DHCP?

Did you delete all previous connections in tftp before proceeding?

Did you restarted the RGA after configuring tftpd?

unplug and plug again the power unit

If the ProcessEye applications hungs or none of these procedures worked, try this:

• Open StartMenu -> MKS RGA Software ->Tools -> AdministrativeSetup

Page 9: Procedure to Configure an MKS Residual Gas Analyzer

ALBA Project Document No: Page: 9 of 11

CCD-GDVC-OP-0001 Rev. No.: 2.1

• Enter Yes/Yes/Next/Next ... Select COM1 as serial line ... Search RGA’s on any Network ... Yes/Next/Finish

• Unplug the ethernet and open ProcessEye.

• Go to Instruments -> Connections, select each IP and click RemoveConnection

• Plug the ethernet and try again to connect, and if fails:

– Reload RGA default configurations.

– Unplug/replug the RGA; run RGA Device Manager

– If fails, Run RGA troubleshooting.

– If fails, ProcessEye->Instruments->Connections->AddConnection

– If fails, Controls ticket at https://rt.cells.es

Page 10: Procedure to Configure an MKS Residual Gas Analyzer

ALBA Project Document No: Page: 10 of 11

CCD-GDVC-OP-0001 Rev. No.: 2.1

4 Appendix: Identify the device IP Address

Information to configure permanently a device for the tunnel.

4.1 For a device to be connected to the tunnel network (Storage Ring and/or Front Ends)Search the Equipment Name of the RGA to be installed in the following list of devices, equipments are identifiedby system (SR/FE) and sector (S01, S02, ...):

Equipment Name Hostname IP address

SR-VC-RGA-S01-01 dvcsrrga01 10.0.11.1

SR-VC-RGA-S02-01 dvcsrrga02 10.0.12.1

SR-VC-RGA-S03-01 dvcsrrga03 10.0.12.2

SR-VC-RGA-S04-01 dvcsrrga04 10.0.14.1

SR-VC-RGA-S05-01 dvcsrrga05 10.0.15.1

SR-VC-RGA-S06-01 dvcsrrga06 10.0.16.1

SR-VC-RGA-S07-01 dvcsrrga07 10.0.17.1

SR-VC-RGA-S08-01 dvcsrrga08 10.0.18.1

SR-VC-RGA-S09-01 dvcsrrga09 10.0.19.1

SR-VC-RGA-S10-01 dvcsrrga10 10.0.20.1

SR-VC-RGA-S11-01 dvcsrrga11 10.0.21.1

SR-VC-RGA-S12-01 dvcsrrga12 10.0.22.1

SR-VC-RGA-S13-01 dvcsrrga13 10.0.23.1

SR-VC-RGA-S14-01 dvcsrrga14 10.0.24.1

SR-VC-RGA-S15-01 dvcsrrga15 10.0.24.2

SR-VC-RGA-S16-01 dvcsrrga16 10.0.25.1

• Take note of the MAC Address (something like 02:80:AD:20:FB:6) that is written in the electronic con-troller of the RGA.

• Take note of the IP Address that must be configured for the device.

• Send an email to [email protected] specifying the Hostname , MAC Address and IP Address of the RGAthat is going to be configured.

• Follow the procedure detailed in CHAPTER 2 to configure the device.

• Put a label in the equipment containing both Equipment Name and Hostname and also the TUNEL key-word.

4.2 For a device used exclusively in the Vacuum Lab or during Installation (connecteddirectly to one of the Vacuum laptops)

If the device has not been labeled yet:

• Take note of the MAC Address (something like 02:80:AD:20:FB:6) that is written in the electronic con-troller of the RGA.

• Send an email to [email protected] specifying the MAC Address and asking for a Hostname and IP, spec-ifying that this equipment is going to be used exclusively in the Vacuum Lab.

Page 11: Procedure to Configure an MKS Residual Gas Analyzer

ALBA Project Document No: Page: 11 of 11

CCD-GDVC-OP-0001 Rev. No.: 2.1

• Follow the procedure detailed in CHAPTER 2 to configure the device.

• Put a label in the equipment containing TEMPORARY and also the VC LAB keyword.

• The IP Address to be configured until Systems will provide a new one will be:

IP Address: 192.168.0.250MASK: 255.255.252.0