1
A Low-Power Configuration for a Tunable Diode Laser Trace Gas Analyzer to Measure N2O & CO2 Fluxes using Flux Gradient Method Benjamin Conrad 1* , Steve Sargent 1 1 Campbell Scientific, Inc., Logan, UT, USA *corresponding author: [email protected] INTRODUCTION Campbell Scientific has been manufacturing tunable diode laser trace gas analyzers (TGAs) since 1993. These field-rugged instruments provide measurements with high accuracy, low noise, and the excellent frequency response required for eddy covariance applications. Widespread use of the TGA for measuring trace gas fluxes remains limited by the high power (~1000 W) pump required for eddy covariance measurements. The new TGA200A has recently shown excellent frequency response for eddy covariance with a relatively low power (240 W) sample pump (Somers and Sargent, 2015, Brown, et al, 2016) However, this power level is still difficult to provide without access to AC mains power. An alternative to eddy covariance, the flux-gradient method (Denmead, 2008) allows the use of a much lower-power pump. Here we present the results of a field trial to evaluate the feasibility of measuring N 2 O fluxes with a very low-power flux-gradient system. In addition, we investigate the tradeoff of system performance with the TGA temperature control disabled to further reduce power demand to less than 30 W, which can easily be supplied with solar panels. The resulting system provides the data integrity and reliability characteristic of closed-path analyzers with the low power normally associated with open-path analyzers. METHODS An N 2 O/CO 2 analyzer with low-power flux-gradient sampling system was tested at the instrumentation garden located at Campbell Scientific, Inc. in Logan Utah. The equipment included (all equipment manufactured by Campbell Scientific, Inc. except as noted): TGA100 trace gas analyzer: This legacy analyzer, manufactured in 1994, was recently upgraded with a new thermoelectrically-cooled interband cascade laser for measuring N 2 O and CO 2 . AP200 Intake Assemblies: Ambient air was sampled through two AP200 heated intake assemblies which include a heated rain cap (0.25 W), filter, 0.007” orifice to set the flow to ~250 ml min -1 , and 750 ml mixing volume. Sampling System: A standard 16-inlet sampling system was modified for use with low-power pumps. Solenoid valves switched between intakes every 30 s. 15 s were omitted for equilibration and 15 s included in the average N 2 O and CO 2 . Pumps: A prototype pump module used three low-power, double-head diaphragm pumps (Parker BTC 11S series, Parker Hannifin Corp., Hollis, NH, USA) in a four-stage configuration. The pump speeds were controlled by a CR3000 datalogger in the sampling system to pull ~180 ml min -1 sample flow through the analyzer at 50 mb and bypass the remaining flow at 400 mb. IRGASON open-path CO 2 /H 2 O eddy covariance system with CR6 datalogger running EasyFlux TM –DL datalogger program. Custom power module with AC/DC power converter (QS10.121, Puls GmbH, Munich, Germany) and a CR1000 datalogger to measure power consumption for each subsystem. The system was operated in four different configurations: Measurement precision with heaters on, sampling compressed air Measurement precision with heaters off, sampling compressed air Zero Gradient with heaters off, sampling ambient air through a tee to both intakes Flux-Gradient with heaters off, sampling ambient air at two heights: 2.6 and 1.8 m The open-path EC system provided CO 2 flux estimates that were used without further corrections. The TGA concentrations were averaged over 30 minutes, and the difference between the intakes was calculated. The eddy diffusivity (k-term) was calculated as the ratio of the EC flux and the CO 2 gradient. This k-term was then used to calculate an N 2 O flux: = − where K g is the eddy diffusivity term, and is the vertical gradient of the gas. RESULTS CONCLUSIONS ACKNOWLEDGEMENTS The authors would like to thank Paul Fluckiger for technical assistance. RESULTS REFERENCES Figure 5. 30 min averages of TGA, pump, sampling system, IRGASON, heater power and total power for same conditions as Figure 2. Figure 2. Ambient and TGA enclosure temperatures for a) compressed air with heaters on b) compressed air with heaters off, c) ambient air with heaters off, and d) gradient with heaters off. Figure 3. 30 min CO 2 and N 2 O means for same conditions as Figure 2. Note: compressed air tank was changed between a and b data collection. Denmead, O.T. , 2008. Approaches to measuring fluxes of methane and nitrous oxide between landscapes and the atmosphere. Plant Soil 309, 5-24 Brown, S.E., S. Sargent, J. Somers, and C. Wagner-/riddle (2016) A Long-Term Field Trial of a Lower- powered Eddy Covariance System for Measuring Nitrous Oxide Fluxes, presented at the 32 nd conference and Agricultrual and Forest Meteorology Somers, J. and S. Sargent, 2015, A novel low-power, high-performance, zero-maintenance closed- path trace gas eddy-covariance system with no water vapor dilution or spectroscopic corrections, poster presented at 2015 Fall Meeting of the American Geophysical Union. Figure 1: TGA sampling system, AP200 Intakes (in zero-gradient arrangement) and low-power sample and prototype pump module. Figure 6. 30 min N 2 O and CO 2 fluxes for two separate time periods: a) corresponds to a 12 mm rain event on 9/14/2016-9/15/2016 b) corresponds to a warm breezy day on 9/15/2016. The 30-minute N 2 O gradient precision was 0.055 ppb with the heaters on, and increased to 0.068 ppb with the heaters off (Figure 4a-b). The 30-minute CO 2 gradient precision was 7 ppb with the heaters on, and increased to 17 ppb with the heaters off (Figure 4a-b). The N 2 O and CO 2 gradient bias was much smaller than the precision with heaters on or off, sampling either compressed air or ambient air (Figure 4a-c). The average N 2 O and CO 2 gradients were much larger than the precision (Figure 4a-d). Total system power with the TGA heaters on was ~100 W. This is dominated by the heater power, which depends on the temperature set point and ambient temperature (Figure 5a). Total system power with the TGA heaters off was <30 W (<25 W without the IRGASON) (Figure 5b-d). The flux-gradient method can measure N 2 O fluxes at a power level which can easily be provided with solar panels. Repeat the test with a new TGA200A. Its smaller sample cell volume (200 ml) would allow faster. switching to further reduce the effect of low-frequency noise. Test the system for multi-site gradient measurements. Run the system on solar panels. Ambient Air Heaters Off 12:00:00 16:00:00 20:00:00 00:00:00 04:00:00 08:00:00 Temperature (C) 0 10 20 30 40 Ambient TGA Temp 1 TGA Temp 2 c Compressed Air Heaters Off 0:00:00 4:00:00 8:00:00 12:00:00 16:00:00 20:00:00 0:00:00 4:00:00 8:00:00 N2O (ppm) 0.32 0.33 0.34 0.35 0.36 0.37 0.38 CO2 (ppm) 360 380 400 420 440 460 480 500 N2O CO2 b Ambient Air Heaters Off 11:00:00 15:00:00 19:00:00 23:00:00 03:00:00 07:00:00 N2O (ppm) 0.32 0.33 0.34 0.35 0.36 0.37 0.38 CO2 (ppm) 360 380 400 420 440 460 480 500 N2O CO2 c Compressed Air Heaters On 19:00:00 23:00:00 03:00:00 07:00:00 N2O (ppm) 0.32 0.33 0.34 0.35 0.36 0.37 0.38 CO2 (ppm) 360 380 400 420 440 460 480 500 N2O CO2 a Gradient Heaters Off 15:00:00 19:00:00 23:00:00 3:00:00 7:00:00 11:00:00 N2O Gradient (ppm) 0.0000 0.0005 0.0010 0.0015 0.0020 CO2 Gradient (ppm) 0 2 4 6 8 10 12 14 16 18 N2O Gradient CO2 Gradient N2O Gradient Mean 3.37 E-4 ppm STDEV 4.26 E-4 ppm CO2 Gradient Mean 2.756 ppm STDEV 3.498 ppm d Figure 4. 30 min CO 2 and N 2 O Upper – Lower Intake for same conditions as Figure 2. FUTURE WORK N2O Gradient Mean 8.64E-6 ppm STDEV 6.83E-5 ppm CO2 Gradient Mean -0.00291 ppm STDEV 0.0168 ppm a Compressed Air Heaters Off 0:00:00 4:00:00 8:00:00 12:00:00 16:00:00 20:00:00 0:00:00 4:00:00 8:00:00 N2O Gradient (ppm) 0.0000 0.0005 0.0010 0.0015 0.0020 CO2 Gradient (ppm) -0.3 -0.2 -0.1 0.0 0.1 0.2 N2O Gradient CO2 Gradient b Compressed Air Heaters On 19:00:00 23:00:00 03:00:00 07:00:00 Temp (C) 0 10 20 30 40 Ambient TGA Temp 1 TGA Temp 2 a Gradient Heater Off 15:00:00 19:00:00 23:00:00 03:00:00 07:00:00 11:00:00 Temperature (C) 0 10 20 30 40 Ambient TGA Temp 1 TGA Temp 2 d Gradient Heater Off 15:00:00 19:00:00 23:00:00 3:00:00 7:00:00 11:00:00 N2O (ppm) 0.32 0.33 0.34 0.35 0.36 0.37 0.38 CO2 (ppm) 360 380 400 420 440 460 480 500 Date vs N2O Date vs CO2 d Power Consumption Gradient Heaters Off 14:00:00 18:00:00 22:00:00 02:00:00 06:00:00 10:00:00 Average Power (W) 0 30 60 90 120 TGA Pump Sampling IRGASON Heater Date vs Total d Power Consumption Compressed Air Heaters On 19:00:00 23:00:00 03:00:00 07:00:00 Average Power (W) 0 30 60 90 120 TGA Pump Sampling IRGASON Heater Date vs Total a Compressed Air Heaters On 19:00:00 23:00:00 03:00:00 07:00:00 N2O Gradient (ppm) 0.0000 0.0005 0.0010 0.0015 0.0020 CO2 Gradient (ppm) -0.3 -0.2 -0.1 0.0 0.1 0.2 N2O Gradient CO2 Gradient N2O Gradient Mean 1.41 E-5 ppm STDEV 5.46E-5 ppm CO2 Gradient Mean -0.00231 ppm STDEV 0.00688 ppm a Ambient Air Heaters Off 11:00:00 15:00:00 19:00:00 23:00:00 03:00:00 07:00:00 N2O Gradient (ppm) 0.0000 0.0005 0.0010 0.0015 0.0020 CO2 Gradient (ppm) -0.3 -0.2 -0.1 0.0 0.1 0.2 N2O Gradient CO2 Gradient N2O Gradient Mean -3.26E-6 ppm STDEV 6.62E-5 ppm CO2 Gradient Mean -0.0108 ppm STDEV 0.0545 ppm c Power Consumption Ambient Air Heaters Off 11:00:00 15:00:00 19:00:00 23:00:00 03:00:00 07:00:00 Average Power (W) 0 30 60 90 120 TGA Pump Sampling IRGASON Heater Date vs Total c Power Consumption Compressed Air Heater Off 00:00:00 04:00:00 08:00:00 12:00:00 16:00:00 20:00:00 00:00:00 04:00:00 08:00:00 Average Power (W) 0 30 60 90 120 TGA Pump Sampling IRGASON Heater Date vs Total b Compressed Air Heaters Off 00:00:00 04:00:00 08:00:00 12:00:00 16:00:00 20:00:00 00:00:00 04:00:00 08:00:00 Temperature (C) 0 10 20 30 40 Ambient TGA Temp 1 TGA Temp 2 b 2016 Ameriflux PI Meeting 30 Min CO2 and N2O flux 15:00:00 19:00:00 23:00:00 03:00:00 07:00:00 N2O Flux (ng m-2 s-1) -100 -50 0 50 100 150 CO2 Flux (mg m-2 s-1) -0.6 -0.4 -0.2 0.0 0.2 0.4 0.6 0.8 1.0 Date vs N2O Flux Date vs CO2 Flux a 30 min CO2 and N2O Flux 11:00:00 15:00:00 19:00:00 23:00:00 N2O Flux (ng m-2 s-1) -100 -50 0 50 100 150 CO2 Flux (mg m-2 s-1) -0.6 -0.4 -0.2 0.0 0.2 0.4 0.6 0.8 1.0 N2O Flux CO2 Flux a

Printing - Campbell Sci · 2016-10-21 · Printing: This poster is 48” wide by 36” high. It’s designed to be ... USA) in a four-stage configuration. The pump speeds were controlled

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Printing - Campbell Sci · 2016-10-21 · Printing: This poster is 48” wide by 36” high. It’s designed to be ... USA) in a four-stage configuration. The pump speeds were controlled

Printing:This poster is 48” wide by 36” high. It’s designed to be printed on a largeformat printer.

Customizing the Content:The placeholders in this formatted for you. placeholders to add text, or click an icon to add a table, chart, SmartArt graphic, picture or multimedia file.

Ttext, just click the Bullets button on the Home tab.

If you need more placeholders for titles, a copy of what you need and drag it into place. PowerPoint’s Smart Guides will help you align it with everything else.

Want to use your own pictures instead of ours? No problem! Just pictureMaintain the proportion of pictures as you resize by dragging a corner.

A Low-Power Configuration for a Tunable Diode Laser Trace Gas Analyzer to Measure N2O & CO2 Fluxes using Flux Gradient Method

Benjamin Conrad1*, Steve Sargent1

1 Campbell Scientific, Inc., Logan, UT, USA*corresponding author: [email protected]

INTRODUCTIONCampbell Scientific has been manufacturing tunable diode laser trace gas analyzers (TGAs) since 1993. These field-rugged instruments provide measurements with high accuracy, low noise, and the excellent frequency response required for eddy covariance applications. Widespread use of the TGA for measuring trace gas fluxes remains limited by the high power (~1000 W) pump required for eddy covariance measurements. The new TGA200A has recently shown excellent frequency response for eddy covariance with a relatively low power (240 W) sample pump (Somers and Sargent, 2015, Brown, et al, 2016) However, this power level is still difficult to provide without access to AC mains power.

An alternative to eddy covariance, the flux-gradient method (Denmead, 2008) allows the use of a much lower-power pump. Here we present the results of a field trial to evaluate the feasibility of measuring N2O fluxes with a very low-power flux-gradient system. In addition, we investigate the tradeoff of system performance with the TGA temperature control disabled to further reduce power demand to less than 30 W, which can easily be supplied with solar panels. The resulting system provides the data integrity and reliability characteristic of closed-path analyzers with the low power normally associated with open-path analyzers.

METHODSAn N2O/CO2 analyzer with low-power flux-gradient sampling system was tested at the instrumentation garden located at Campbell Scientific, Inc. in Logan Utah. The equipment included (all equipment manufactured by Campbell Scientific, Inc. except as noted):

• TGA100 trace gas analyzer: This legacy analyzer, manufactured in 1994, was recently upgraded with a new thermoelectrically-cooled interband cascade laser for measuring N2O and CO2.

• AP200 Intake Assemblies: Ambient air was sampled through two AP200 heated intake assemblies which include a heated rain cap (0.25 W), filter, 0.007” orifice to set the flow to ~250 ml min-1, and 750 ml mixing volume.

• Sampling System: A standard 16-inlet sampling system was modified for use with low-power pumps. Solenoid valves switched between intakes every 30 s. 15 s were omitted for equilibration and 15 s included in the average N2O and CO2.

• Pumps: A prototype pump module used three low-power, double-head diaphragm pumps (Parker BTC 11S series, Parker Hannifin Corp., Hollis, NH, USA) in a four-stage configuration. The pump speeds were controlled by a CR3000 datalogger in the sampling system to pull ~180 ml min-1 sample flow through the analyzer at 50 mb and bypass the remaining flow at 400 mb.

• IRGASON open-path CO2/H2O eddy covariance system with CR6 datalogger running EasyFluxTM –DL datalogger program.

• Custom power module with AC/DC power converter (QS10.121, Puls GmbH, Munich, Germany) and a CR1000 datalogger to measure power consumption for each subsystem.

The system was operated in four different configurations:

• Measurement precision with heaters on, sampling compressed air

• Measurement precision with heaters off, sampling compressed air

• Zero Gradient with heaters off, sampling ambient air through a tee to both intakes

• Flux-Gradient with heaters off, sampling ambient air at two heights: 2.6 and 1.8 m

The open-path EC system provided CO2 flux estimates that were used without further corrections. The TGA concentrations were averaged over 30 minutes, and the difference between the intakes was calculated. The eddy diffusivity (k-term) was calculated as the ratio of the EC flux and the CO2

gradient. This k-term was then used to calculate an N2O flux:

𝐹𝑔 = −𝐾𝑔 ∗𝜕𝜌𝑔

𝜕𝑧

where Kg is the eddy diffusivity term, and 𝜕𝜌𝑔

𝜕𝑧is the vertical gradient of the gas.

RESULTS

CONCLUSIONS

ACKNOWLEDGEMENTSThe authors would like to thank Paul Fluckiger for technical assistance.

RESULTS

REFERENCES

Figure 5. 30 min averages of TGA, pump, sampling system, IRGASON, heater power and total power for same conditions as Figure 2.

Figure 2. Ambient and TGA enclosure temperatures for a) compressed air with heaters on b) compressed air with heaters off, c) ambient air with heaters off, and d) gradient with heaters off.

Figure 3. 30 min CO2and N2O means for same conditions as Figure 2.Note: compressed air tank was changed between a and b data collection.

Denmead, O.T. , 2008. Approaches to measuring fluxes of methane and nitrous oxide between landscapes and the atmosphere. Plant Soil 309, 5-24

Brown, S.E., S. Sargent, J. Somers, and C. Wagner-/riddle (2016) A Long-Term Field Trial of a Lower-powered Eddy Covariance System for Measuring Nitrous Oxide Fluxes, presented at the 32nd

conference and Agricultrual and Forest Meteorology

Somers, J. and S. Sargent, 2015, A novel low-power, high-performance, zero-maintenance closed-path trace gas eddy-covariance system with no water vapor dilution or spectroscopic corrections, poster presented at 2015 Fall Meeting of the American Geophysical Union.

Figure 1: TGA sampling system, AP200 Intakes (in zero-gradient arrangement) and low-power sample and prototype pump module.

Figure 6. 30 min N2O and CO2 fluxes for two separate time periods: a) corresponds to a 12 mm rain event on 9/14/2016-9/15/2016 b) corresponds to a warm breezy day on 9/15/2016.

• The 30-minute N2O gradient precision was 0.055 ppb with the heaters on, and increased to 0.068 ppb with the heaters off (Figure 4a-b).

• The 30-minute CO2 gradient precision was 7 ppb with the heaters on, and increased to 17 ppb with the heaters off (Figure 4a-b).

• The N2O and CO2 gradient bias was much smaller than the precision with heaters on or off, sampling either compressed air or ambient air (Figure 4a-c).

• The average N2O and CO2 gradients were much larger than the precision (Figure 4a-d).

• Total system power with the TGA heaters on was ~100 W. This is dominated by the heater power, which depends on the temperature set point and ambient temperature (Figure 5a).

• Total system power with the TGA heaters off was <30 W (<25 W without the IRGASON) (Figure 5b-d).

• The flux-gradient method can measure N2O fluxes at a power level which can easily be provided with solar panels.

• Repeat the test with a new TGA200A. Its smaller sample cell volume (200 ml) would allow faster. switching to further reduce the effect of low-frequency noise.

• Test the system for multi-site gradient measurements.• Run the system on solar panels.

Ambient Air Heaters Off

12:

00:0

0

16:

00:0

0

20:

00:0

0

00:

00:0

0

04:

00:0

0

08:

00:0

0

Te

mp

era

ture

(C

)

0

10

20

30

40

Ambient

TGA Temp 1

TGA Temp 2

c

Compressed Air Heaters Off

0:0

0:00

4:0

0:00

8:0

0:00

12:

00:0

0

16:

00:0

0

20:

00:0

0

0:0

0:00

4:0

0:00

8:0

0:00

N2

O (

pp

m)

0.32

0.33

0.34

0.35

0.36

0.37

0.38

CO

2 (

pp

m)

360

380

400

420

440

460

480

500

N2O

CO2

b

Ambient Air Heaters Off

11:

00:0

0

15:

00:0

0

19:

00:0

0

23:

00:0

0

03:

00:0

0

07:

00:0

0

N2

O (

pp

m)

0.32

0.33

0.34

0.35

0.36

0.37

0.38

CO

2 (

pp

m)

360

380

400

420

440

460

480

500

N2O

CO2

c

Compressed Air Heaters On

19:

00:0

0

23:

00:0

0

03:

00:0

0

07:

00:0

0

N2

O (

pp

m)

0.32

0.33

0.34

0.35

0.36

0.37

0.38

CO

2 (

pp

m)

360

380

400

420

440

460

480

500

N2O

CO2

a

Gradient Heaters Off

15:

00:0

0

19:

00:0

0

23:

00:0

0

3:0

0:00

7:0

0:00

11:

00:0

0

N2

O G

rad

ient (p

pm

)

0.0000

0.0005

0.0010

0.0015

0.0020

CO

2 G

rad

ient (p

pm

)

0

2

4

6

8

10

12

14

16

18

N2O Gradient

CO2 Gradient

N2O GradientMean 3.37 E-4 ppmSTDEV 4.26 E-4 ppm

CO2 GradientMean 2.756 ppmSTDEV 3.498 ppm

d

Figure 4. 30 min CO2and N2O Upper – Lower Intake for same conditions as Figure 2.

FUTURE WORK

N2O GradientMean 8.64E-6 ppmSTDEV 6.83E-5 ppm

CO2 GradientMean -0.00291ppm STDEV 0.0168 ppm

a

Compressed Air Heaters Off

0:0

0:00

4:0

0:00

8:0

0:00

12:

00:0

0

16:

00:0

0

20:

00:0

0

0:0

0:00

4:0

0:00

8:0

0:00

N2

O G

rad

ient (p

pm

)

0.0000

0.0005

0.0010

0.0015

0.0020

CO

2 G

rad

ient (p

pm

)

-0.3

-0.2

-0.1

0.0

0.1

0.2

N2O Gradient

CO2 Gradient

b

Compressed Air Heaters On

19:

00:0

0

23:

00:0

0

03:

00:0

0

07:

00:0

0

Te

mp

(C

)

0

10

20

30

40

Ambient

TGA Temp 1

TGA Temp 2

a

Gradient Heater Off

15:

00:0

0

19:

00:0

0

23:

00:0

0

03:

00:0

0

07:

00:0

0

11:

00:0

0

Te

mp

era

ture

(C

)

0

10

20

30

40

Ambient

TGA Temp 1

TGA Temp 2

d

Gradient Heater Off

15:

00:0

0

19:

00:0

0

23:

00:0

0

3:0

0:00

7:0

0:00

11:

00:0

0

N2

O (

pp

m)

0.32

0.33

0.34

0.35

0.36

0.37

0.38

CO

2 (

pp

m)

360

380

400

420

440

460

480

500

Date vs N2O

Date vs CO2

d

Power Consumption Gradient Heaters Off

14:

00:0

0

18:

00:0

0

22:

00:0

0

02:

00:0

0

06:

00:0

0

10:

00:0

0

Ave

rage P

ow

er

(W)

0

30

60

90

120

TGA

Pump

Sampling

IRGASON

Heater

Date vs Total

d

Power Consumption Compressed Air Heaters On

19:

00:0

0

23:

00:0

0

03:

00:0

0

07:

00:0

0

Ave

rage P

ow

er

(W)

0

30

60

90

120

TGA

Pump

Sampling

IRGASON

Heater

Date vs Total

a

Compressed Air Heaters On

19:

00:0

0

23:

00:0

0

03:

00:0

0

07:

00:0

0

N2

O G

rad

ien

t (p

pm

)

0.0000

0.0005

0.0010

0.0015

0.0020

CO

2 G

rad

ien

t (p

pm

)

-0.3

-0.2

-0.1

0.0

0.1

0.2

N2O Gradient

CO2 Gradient

N2O GradientMean 1.41 E-5 ppmSTDEV 5.46E-5 ppmCO2 GradientMean -0.00231 ppm STDEV0.00688 ppm

a

Ambient Air Heaters Off

11:

00:0

0

15:

00:0

0

19:

00:0

0

23:

00:0

0

03:

00:0

0

07:

00:0

0

N2

O G

rad

ien

t (p

pm

)

0.0000

0.0005

0.0010

0.0015

0.0020

CO

2 G

rad

ien

t (p

pm

)

-0.3

-0.2

-0.1

0.0

0.1

0.2

N2O Gradient

CO2 Gradient

N2O GradientMean -3.26E-6 ppmSTDEV 6.62E-5 ppmCO2 GradientMean -0.0108 ppmSTDEV0.0545 ppm

c

Power Consumption Ambient Air Heaters Off

11:

00:0

0

15:

00:0

0

19:

00:0

0

23:

00:0

0

03:

00:0

0

07:

00:0

0

Ave

rage P

ow

er

(W)

0

30

60

90

120

TGA

Pump

Sampling

IRGASON

Heater

Date vs Total

c

Power Consumption Compressed Air Heater Off

00:

00:0

0

04:

00:0

0

08:

00:0

0

12:

00:0

0

16:

00:0

0

20:

00:0

0

00:

00:0

0

04:

00:0

0

08:

00:0

0

Ave

rage P

ow

er

(W)

0

30

60

90

120

TGA

Pump

Sampling

IRGASON

Heater

Date vs Total

b

Compressed Air Heaters Off

00:

00:0

0

04:

00:0

0

08:

00:0

0

12:

00:0

0

16:

00:0

0

20:

00:0

0

00:

00:0

0

04:

00:0

0

08:

00:0

0

Te

mp

era

ture

(C

)

0

10

20

30

40

Ambient

TGA Temp 1

TGA Temp 2

b

2016 AmerifluxPI Meeting

30 Min CO2 and N2O flux

15:

00:0

0

19:

00:0

0

23:

00:0

0

03:

00:0

0

07:

00:0

0

N2O

Flu

x (n

g m

-2 s

-1)

-100

-50

0

50

100

150

CO

2 F

lux

(mg m

-2 s

-1)

-0.6

-0.4

-0.2

0.0

0.2

0.4

0.6

0.8

1.0

Date vs N2O Flux

Date vs CO2 Flux

a

30 min CO2 and N2O Flux

11:

00:0

0

15:

00:0

0

19:

00:0

0

23:

00:0

0

N2O

Flu

x (

ng m

-2 s

-1)

-100

-50

0

50

100

150

CO

2 F

lux (

mg m

-2 s

-1)

-0.6

-0.4

-0.2

0.0

0.2

0.4

0.6

0.8

1.0

N2O Flux

CO2 Flux

a