5
@ Tel. 351 21 27 40606 Fax351 212'740897 Medium: Air. Calibration due: 1 8,02-2014 Soc. Com. Equipamentos de Controlo, Lda. Instrumentation - Automation - Control Rua ConceigAo Sameiro Antunes, 268 2805-122 ALM^D^ Portugal GERTilF[@ATE Ref.: M/T "GENMAR ST. NIKOLAS" Dale: 1810212013 Serviced by: Carlos Augusto We hereby cediry that the SUIVIfIER/I4 Vacuum Gauge, for Si Pressure calibrator, Range: 0 to -760mmHg, has been calibration checked against standard, with the following results: Metrology requirements of Bourdon type pressure gauge: Standard no. EN837-1 Subdivision minimum: 20 mmHg (resolution) ENFM Press. Gauge Cal. Ref. Press. Gauge -50 mmHg -48,9 mmHg -100 mmHg -98,3 mmHe -200 mmHg -198,1mmH9 -300 mmHg -297,6 mmHg -400 m mHg -500 mmHg -397,8 mmHg -497,O mmHg -600 mmHg -603,5 mmHg -700 mmHg -710,2 mmHg - The readings are within the manufacturer's specified accuracy. Reference Press. Gauge: KELLER, LEXI/ serial no. 2593 Uncertainty: 0.01% FS Traceability: Laborat6rio Nacional de Metrologia certificate no 512j2103971 Room Temperature = 22,1oC ,/. O A,\At)\ a- ,/,1:./ .4. (t^/,/ ./ A-Pt ngrneer

Pressure Calibrators

  • Upload
    sng1881

  • View
    33

  • Download
    0

Embed Size (px)

DESCRIPTION

Marine Pressure Calibrators

Citation preview

Page 1: Pressure Calibrators

@Tel. 351 21 27 40606Fax351 212'740897

Medium: Air.

Calibration due: 1 8,02-2014

Soc. Com. Equipamentos de Controlo, Lda.Instrumentation - Automation - Control

Rua ConceigAo Sameiro Antunes, 2682805-122 ALM^D^ Portugal

GERTilF[@ATE

Ref.: M/T "GENMAR ST. NIKOLAS"Dale: 1810212013Serviced by: Carlos Augusto

We hereby cediry that the SUIVIfIER/I4 Vacuum Gauge, for Si Pressure calibrator,Range: 0 to -760mmHg, has been calibration checked against standard, with thefollowing results:

Metrology requirements of Bourdon type pressure gauge:Standard no. EN837-1Subdivision minimum: 20 mmHg (resolution)

ENFM Press. Gauge Cal. Ref. Press. Gauge

-50 mmHg -48,9 mmHg

-100 mmHg -98,3 mmHe

-200 mmHg -198,1mmH9

-300 mmHg -297,6 mmHg

-400 mmHg

-500 mmHg

-397,8 mmHg

-497,O mmHg

-600 mmHg -603,5 mmHg

-700 mmHg -710,2 mmHg

- The readings are within the manufacturer's specified accuracy.

Reference Press. Gauge: KELLER, LEXI/ serial no. 2593Uncertainty: 0.01% FS

Traceability: Laborat6rio Nacional de Metrologia certificate no 512j2103971

Room Temperature = 22,1oC

,/. O A,\At)\

a- ,/,1:./ .4.(t^/,/ ./ A-Pt

ngrneer

sng
Stamp
sng
Rectangle
sng
Rectangle
sng
Stamp
Page 2: Pressure Calibrators

lEil

IEIIEI EII@ilEIEilE @ilE IE!ilE!IEU t@I@IE!IE!il84 EItEilEl

ControliS

Tel. 351 21 27 40606Fax35l 21 2740897

Soc. Com. Equipamentos de Controlo, Lda.Instrumentation - Automation - Control

Rua Conceig6o Sameiro Antunes, 26E2805-122 ALMADA

FIGATE

Portugal

Cert. no.Ref.: M/T "GENMAR ST. NIKOLAS"Dale: 18102120'13Serviced by: Carlos Augusto

We hereby certify that the Si, Hydraulic Hand Held Pressure Calibrator.Type HTPI - Range: 0 to 700bar.Press. gauge: Sika model: LEO2, serial no- 021837Range: 0 to 700 bar / Resolution: 0,1 bar / Accuracy: 10,1% Full Scale.lnstrument has been checked and calibrated with the following results:

SIKA Press. Gauge Ret. Press. Gauge

5,0 bar 5,0 bar

25,0 bat 25,0 bar

50,0 bar 50,0 bar

100,0 bar 100,0 bar

200,0 bar 200,1 bat

300,0 bar 300,1 bar

400,0 bar 400,2 bar

500,0 bar 500,3 bar

600,0 bar 600,3 bar

700,0 bar 700,3 bar

Accuracy of readings is within manufacturer's specification-Medium: OilReference Gauge: KELLER LEX10 serial no. 4968'1

Uncertainty: 0.05% FST.aceability; Laborat6rio Nacional MetroCalibration due: 1a.O2.2O1 4

no.213.121651384

tEil

EEEEEEEEEEEEEEEElGit!=l

EE@

Eq6E6EEE6

lEil

EllGil

lEil

El

Elt=lErl

ffiffiEit

qE@

EEEEEEEEEEEEEEEtsl

tEu Ell EU ElilEIElilEt EIIIE lrEiltElltElrElrE EIIE @t @IE!IE!

sng
Stamp
sng
Rectangle
sng
Rectangle
sng
Stamp
Page 3: Pressure Calibrators

ControliS\'*-/

Soc. Com. Equipamentos de Controlo, Lda.

Te1.35121 2740606Fax351 21 2740897

Instrumentation - Automation - Control

Rua Conceiqdo Sameiro Antunes, 26E2805-i22 ALMADA Portugal

GERTIFIGATEcert. no. Sddi4li,3.Ref.: M/T "GENMAR ST. NIKOLAS"Dale: 1810212013Serviced by: Carlos Augusto

We hereby ceftiry that lhe ENFM Pres-sure Gauge,calibntof, Range: 0 to 40 Bar, has been calibrationthe lollowing results:

Metrology requirements of Bourdon type pressure gauge:Standard no. EN837-'lSubdivision minimum: 1 bar (resolution)

ENFM Press. Gauge Cal. Ref. Press. Gauge

2 Bar 2,781 Bat

5 Bar 5,161 Bar

10 Bar 10,008 Bar

15 Bar 14,950 Bar

20 Ber 19,920 Bar

25 Bar 24,894 Bar

30 Bar 29,877 Bar

40 Bar 39,853 Bar

- The readings are within the manufacturels specified accuracy.

Reference Press. Gauge: KELLER, LEXI/ serial no. 2593Uncertainty: 0.01% FS

Traceability: Laborat6rio Nacional de Metrologja certificate no. 512.12103971

Room Temperature = 22,1oC

Medium: Air.

Calibration due: 14,02,2014

model @/ for Si Pressurechecked against standard, with

,,,i-at're+\+, "4

neer

sng
Rectangle
sng
Stamp
sng
Rectangle
sng
Stamp
Page 4: Pressure Calibrators

ControliS\\--lSoc. Com. Equipamentos de Controlo, Lda,

Tel.351 21 2740606Fax35121 274O897

Cert. no. 3370/13Ref.: M/T "GENMAR ST. NIKOLAS"Dale. 1910212013Serviced by: Carlos Augusto

We hereby certify that the MS Pressure Gauge forref- (1208-5), Range: 0 to 3,5 MPa, seial no, 6013531,against standard, with the following results:

Metrology requirements of Bourdon type pressure gauge:Standard no. EN837-1Class Precision: Kl. '1,5

Subdivision minimum: 0,1 MPa (resolution)

Instrumentation - Automation - Control

Rua Conceigao Sameiro Antunes, 26E2805-122 ALMADA Portugal

Boiler Drum Sfeam Press.has been calibration checked

NKS Press. Gauge Cal. Ref. Press. Gauge

0,2 MPA 0,202 MPa

0,5 MPA 0,502 MPa

1,0 MPA 1,002 MPa

1,5 MPA 1,513 MPa

2,0 MPA 2,014 MPa

2,5 MPA 2,514 MPa

3,0 MPA 3,014 MPa

4,0 MPA 4,015 MPa

- The readings are within the manufacturer's specified accuracy.

Reference Press. Gauge: KELLER, LEXI/ serial no. 2593, Uncertainty: 0.01% FS

Traceabil;ty: Laborat6rio Nacional de Metrologia certifjcate no.512.12103971Room Temperature = 17,2'CMedium: Air.

Calibration due: 19.O2.2O1 4

Servlcd

sng
Rectangle
sng
Stamp
sng
Rectangle
sng
Stamp
Page 5: Pressure Calibrators

Soc. Com. Equipamentos de Controlo, Lda.Instrumentation - Automation - Control

Rua Conceigdo Sameiro Antunes, 26E2805-122 ALMADA Portugal

/,:\ T= r:\-:t n T= n /:\ ^

E f-I{:t=tlt | |t:lllt:/A\ | ]=\:./L-]UU U ULI U\:./l-r-!] LJ LJ

Cert. no. 337{113Ref.: M/T .,GENMAR ST. NIKOLAS"Date: 1910212013Serviced by: Carlos Augusto

We hereby certify that the /VKS Pr€-ssur€ Gauge forref. (1208-P), Range: 0 to 3,5 MPa, seial no. 601479,against standard, with the following results:

Metrology requirements of Bourdon type pressure gauge:Standard no. EN837-1Class Precisioni Kl. 1,5Subdivision minimum: 0,'1 MPa (resolution)

Tel.35l 21 2740606Fax351 21 2740897

Boiler Drum Sfeam Press.has been calibration checked

NKS Press. Gauge Cal. Ref. Press. Gauge

0,2 MPA 0,204 MPa

0,5 MPA 0,504 MPa

1,0 MPA 1,004 MPa

1,5 MPA 1,505 MPa

2,0 MPA 2.005 MPa

2,5 MPA 2,515 MPa

3,0 MPA 3,016 MPa

4,0 MPA 4.016 MPa

- The readings are within the manufacturer's specified accuracy.

Reference Press. Gauge: KELLER, LEXl/ serial no. 2593, Uncertainty: 0.01% FS

Traceability: Laboratorio Nacional de lvletrologiatqrtificate no. 512j2103971Room Temperature = 17,2oC

Medium: Air.Calibration due: 19.02.201 4

T

Contr

sng
Rectangle
sng
Stamp
sng
Rectangle
sng
Stamp