14
Presentation of Presentation of the the ClearPEM-Sonic ClearPEM-Sonic mammography mammography scanner scanner ClearPEM-Sonic ClearPEM-Sonic

Presentation of the ClearPEM-Sonic mammography scanner ClearPEM-Sonic

Embed Size (px)

Citation preview

Presentation of the Presentation of the ClearPEM-Sonic ClearPEM-Sonic mammography mammography scannerscanner

ClearPEM-SonicClearPEM-Sonic

2

Breast cancer:Breast cancer:

The most frequent cancer in women

One in nine women has breast cancer in her life

X-ray mammographyX-ray mammography

Standard in breast cancer diagnosis

Based in structural tissue changesBased in structural tissue changes

High rate of false-positiveHigh rate of false-positive

Less reliable in dense breasts Less reliable in dense breasts false negatives

Breast cancerBreast cancer

3

PET mammography

X-ray mammography

PET mammographyPET mammography

PET and breast cancerPET and breast cancer•PET/FDG has good sensitivity in

breast cancer diagnosis

Radioactive tracerRadioactive tracer•FDG : glucose associated to

positron emitter F-18•Glucose has strong affinity to

cancer cells

PEM ScannerPEM Scanner•Positron Emission MammographyPositron Emission Mammography•PET equipment dedicated to breast PET equipment dedicated to breast

and axila examsand axila exams•Good spatial resolution ( ~ mm)Good spatial resolution ( ~ mm)•Good sensitivityGood sensitivity

4

New technology based in fast crystals (LSO) and avalanche photodiodes (APD) associated to low-noise integrated electronics

The new electronics has the highest integration density available worldwide: compact detectors with DoI capability

Scanner with excellent spatial resolution (1.3 mm)

Integration with ultra-sound probe with elastography from company Supersonic Imagine.

ClearPEM-Sonic technologyClearPEM-Sonic technology

5

ClearPEM scannerClearPEM scanner

6

Detector and Front End ElectronicsDetector and Front End Electronics

LYSO:Ce 2x2x20 mm3

4x8 matrix (BaSO4 walls)S8550

16 Supermodules

32 Frond End Electronic Boards

Hamamatsu

Supermodules Integration Detector modules

7

1. Water Cooling Plates 4. On-Detector Cabling

2. Supermodules 5. Service Board (LV distribution; HV regulation; Control & Monitoring)

3. Patch Panel 6. HV Matrix (HV distribution)

1

45

6

12

3

4

Detector Heads IntegrationDetector Heads Integration

8

Service Rack

Workstation

DSS – Detector Safety System

DAE – Data Acquisition Electronics

HV Power Supply (500V)

LV Power Supply (3V2, 5V)

Service Manager

Acquisition Manager / Data Storage

Transformer

Data Acquisition and ServicesData Acquisition and Services

9

Mechanical Integration and AssemblingMechanical Integration and Assembling

10

ClearPEM scannerClearPEM scanner

11

Spatial resolutionSpatial resolution

Resolution transaxial: 1.3 mm

DoI capability allows sharp imaging

With DoI

5 mm

1 mm

Without DoI

3.0 mm

2.5 mm2.0 mm

1.5 mm

1.2 mm

12

Animal imagingAnimal imaging

tumorbladder

3D OSEM6 it. – 4 sub.

2D OSEM3 it. – 4 sub.

FDG scanF-18 scan

After injection

1h after

Bone scan

13

First breast imagesFirst breast images

Na-22 source superimposed on breast data

PET mammography consortiumPET mammography consortium

Funded by

14

R&D consortium of research institutes collaborating with PETsys