User Manual for Parsortix PR1 Cell Separation System For Research Use Only Not for Clinical or Diagnostic use Version: PR1-OM-C Parsortix PR1 User Manual Dated: 13 October 2015 © ANGLE Europe 2015

PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

  • Upload
    others

  • View
    48

  • Download
    0

Embed Size (px)

Citation preview

Page 1: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

 

 

 

 

 

 

User Manual for Parsortix PR1 Cell Separation System For Research Use Only Not for Clinical or Diagnostic use

Version: PR1-OM-C Parsortix PR1 User Manual Dated: 13 October 2015

 

© ANGLE Europe 2015

 

 

 

 

 

 

Page 2: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015

PR1‐OM‐CParsortixPR1UserManual

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Page 3: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015 

PR1‐OM‐CParsortixPR1UserManual

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Page 4: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015

PR1‐OM‐CParsortixPR1UserManual

Contents    

 

1  Before you start .......................................................................................... 1 

1.1  What is the Parsortix PR1? .................................................................. 1 

1.2  Get to know the equipment ................................................................ 1 

1.3  Handling the separation cassettes ...................................................... 3 

1.4  Process workflow ................................................................................ 5 

1.5  Collecting and handling blood samples ............................................... 5 

2  How to use the Parsortix PR1 ..................................................................... 6 

2.1  Instrument protocols ........................................................................... 6 

2.2  Initial checklist ..................................................................................... 6 

2.3  Priming................................................................................................. 8 

2.4  Blood separation ................................................................................. 9 

2.5  Cell harvest ........................................................................................ 11 

2.6  Cleaning ............................................................................................. 13 

3  In‐cassette immunofluorescence staining ............................................... 14 

3.1  Background ........................................................................................ 14 

3.2  Procedure .......................................................................................... 14 

4  Maintenance ............................................................................................ 16 

5  Troubleshooting ....................................................................................... 18 

6  Assembly and installation ......................................................................... 19 

6.1  Positioning and interference ............................................................. 19 

6.2  Unpack the box .................................................................................. 19 

6.3  Attach the drip tray ........................................................................... 19 

6.4  Apply the non‐slip pads ..................................................................... 20 

6.5  Connect the cassette clamp .............................................................. 22 

6.6  Connect the reagent tubes ................................................................ 23 

6.7  Connect the harvest valve waste tube .............................................. 24 

6.8  Start the Parsortix PR1 ...................................................................... 24 

6.9  Before first use .................................................................................. 25 

7  Technical specifications ............................................................................ 26 

8  Health and safety ..................................................................................... 27 

Page 5: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015 

PR1‐OM‐CParsortixPR1UserManual

8.1  Potential hazards .............................................................................. 27 

8.1.1  Biohazards ................................................................................. 28 

8.1.2  Handling reagents and solutions ............................................... 29 

8.2  Labelling ............................................................................................ 29 

9  Technical support ...................................................................................... 32 

9.1  Contact us ......................................................................................... 32 

9.2  Consumables and spares .................................................................. 32 

9.3  Material compatibility ....................................................................... 34 

10  Disposal of electronic equipment ............................................................. 35 

11  Certificate of decontamination ................................................................. 36 

12  Declaration of conformity ......................................................................... 37 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Page 6: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015

PR1‐OM‐CParsortixPR1UserManual

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Page 7: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015 

1PR1‐OM‐CParsortixPR1UserManual

1 Beforeyoustart

1.1 WhatistheParsortixPR1?

The Parsortix  system  enables  the  separation  and  capture of  cells present  in 

blood based on their size and deformability.    It comprises an  instrument, the 

Parsortix  PR1,  and  a  series  of  disposable  separation  cassettes  that  can  be 

selected according to need.   

The equipment’s  intended use  is to filter  large, clinically relevant cells from a 

liquid sample, typically blood. Examples  include circulating tumor cells  (CTCs) 

and fetal cells.  

The key aspects of the system are:  

1. The Parsortix PR1: a laboratory benchtop instrument.  

2. The  separation  cassette:  a  disposable,  non‐sterile  unit  containing  a 

separation structure comprising a series of steps across which cells are 

forced to pass. 

3. The separation principle: cells separate on the combined basis of their 

size and deformability. 

4. Capture of  target cells:  target cells  (e.g. CTCs and  fetal  trophoblasts) 

are captured on a separation “step” inside the cassette. 

5. Protocols: the system is operated via a set of ready to use automated 

procedures. 

   

1.2 Gettoknowtheequipment

Key components of the Parsortix PR1 (Figure 1.1) are:  

1. Screen and control panel 

2. Cassette clamp (holding a separation cassette) 

3. 500 ml waste container 

4. 250 ml buffer reservoir 

5. Blood sample tube (Vacutainer or Falcon tube) 

6. 15 ml tubes (x6) for staining reagents  

WARNING: The Parsortix PR1 is for RESEARCH USE ONLY and not for use in clinical or diagnostic procedures. 

WARNING: The device is intended for use by suitably qualified and experienced personnel in laboratory environments. 

WARNING: It is essential to read the Health and Safety Information presented in Section 8 before assembling or operating the Parsortix system. 

Important information 

Older versions of the instrument may be named Parsorter PR1, but there is no difference between instruments with these alternative names, nor in their operation.  The name Parsortix PR1 will be used throughout this document. 

Page 8: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015

2 PR1‐OM‐CParsortixPR1UserManual

7. 50 ml tubes (x2)  for cleaning and priming reagents  

8. Harvesting valve 

9. Harvest waste tube 

 

Figure 1.1 Overview of key components of the Parsortix PR1. 

 

 

 

The  screen  (Figure  1.2)  displays  information  about  the  system  status  and 

prompts  for  user  input.  The  four  buttons  are  soft  keys  that  invoke  various 

functions as displayed in the screen above each button. 

 

Figure 1.2 Control panel showing the main menu on screen. 

The main menu, shown  in Figure 1.2, allows access to the protocol  library.  In 

order  to browse  the  list of available protocols use  the keys  for  left  [<<] and 

right [>>] to scroll then press [Run] to activate the selected protocol. When the 

system  is  running,  two  soft  keys will be  allocated  the  functions  [Abort]  and 

[Esc].  Pressing  and  holding  [Abort] will  stop  the  protocol  and  return  to  the 

main menu.  Following  an  [Abort]  action,  the  instrument must  be  reset  by 

running  a  cleaning  process  (Section  2.6).  Pressing  and  holding  [Esc]  will 

WARNING: Risk of eye damage. This equipment has several sections of flexible plastic tubing that are not fixed. Handle with care to avoid tubing striking eyes. 

Page 9: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015 

3PR1‐OM‐CParsortixPR1UserManual

progress  the  system  to  the  next  activity  in  the  current  protocol.  The  key 

[System] is generally used for instrument servicing and in cases of operational 

problems  in  consultation  with  Angle  Technical  Support.  In  normal  use,  it 

should  not  be  operated.  If  the  system  control  is  activated  by mistake  the 

screen will  show  three  controls:  [USB],  [Shutdown] and  [Exit]. Press  [Exit]  to 

return to the main menu. 

1.3 Handlingtheseparationcassettes

Cassettes are available with different gap sizes.  Contact ANGLE for details.   

During normal operation of  the Parsortix PR1,  two  cassettes are  required: a 

separation cassette and a cleaning cassette. They are both a Parsortix GEN3 

Separation Cassette but each is assigned to a different process. The separation 

cassette  is  a  new  cassette  requested  during  the  priming  process  for  the 

subsequent separation of a blood sample. The cleaning cassette is used during 

the  intermediate  and  final  cleaning  processes  (use  a GEN3P10  or GEN3D10 

cassette for cleaning). The cleaning cassette can be reused. A cleaning cassette 

in repeated use should be changed after approximately one month.   

The cassette in use is placed inside the cassette clamp. To remove and insert a 

new cassette: 

1. Hold  the clamp upside down  to avoid dropping  the cassette. Release 

the hinged retainer and open the clamp.  

2. Remove  the  cassette  already  in  place.  Retain  cleaning  cassettes  for 

subsequent processes.  Retain separation cassettes for continuing use 

(e.g. harvest,  staining, analysis, etc.) or dispose of  them  if no  longer 

required. 

3. Insert the cassette required for the protocol in progress. The cassette 

must  be  inserted  in  the  correct  orientation.  To  achieve  this,  the 

cassette alignment slots should sit on the clamp retaining pins and the 

orientation  arrows  in  the  cassette must  coincide with  the  inlet  and 

outlet direction as labelled on the cassette tubing (Figure 1.3).  

4. Carefully close the clamp and secure with the hinged retainer. 

 

 

WARNING:  Used  cassettes  should  be  handled  and disposed of as a bio‐hazard. 

Important information 

The separation cassettes are fragile and must be handled with care. ALWAYS handle the cassette by the edges and avoid applying pressure to its surfaces. 

Useful tip 

Examining your cassette to find the side with the film will help orient the slots.  With the cassette film side up, the slots are beneath the film.   

To close the clamp, slightly pull the hinged retainer away and over the clamp.

WARNING: Risk of aerosol. Never attempt to remove a Vacutainer from its mount, or a cassette from its clamp, whilst instrument is in operation.  

Page 10: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015

4 PR1‐OM‐CParsortixPR1UserManual

 

Figure 1.3 Positioning of the separation cassette inside the clamp. Notice the arrows showing the cassette alignment slots and the corresponding retaining pins on the clamp. 

 

 

Page 11: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015 

5PR1‐OM‐CParsortixPR1UserManual

1.4 Processworkflow

The operation of the Parsortix PR1 is divided into distinct processes controlled 

by software based protocols. The workflow is summarised in Figure 1.4. 

 

Figure 1.4 Parsortix PR1 workflow. 

 

1.5 Collectingandhandlingbloodsamples

Blood  for  separation  in  the  Parsortix  PR1  must  be  drawn  into  an  EDTA 

Vacutainer to inhibit coagulation. It is strongly recommended to collect blood 

samples  into  a Becton Dickinson 10 ml Vacutainer, product number 366643 

(US)  or  367525  (UK).    The  Parsortix  PR1  is  specifically  designed  to  fit  this 

Vacutainer. 

The  Vacutainer  used  for  phlebotomy  can  be  attached  directly  to  the 

instrument or, alternatively, a fraction of the sample can be separated  into a 

second Vacutainer.    In this case, ensure that the second Vacutainer has been 

rinsed with PBS to avoid exposing the blood sample to a higher concentration 

of EDTA.  

The Parsortix PR1 also accommodates a 50 ml tube for use with larger sample 

volumes.  The  mount  fits  a  specific  50  ml  tube  (Corning  product  number 

352070).  

In order to avoid spillages, a blood tube should only contain up to 80% of  its 

total volume.   

 

In‐cassette staining Cell harvest

Short clean

Priming

Blood separation

Full clean

Important information 

Blood drawn into EDTA Vacutainers should be stored at 4°C if it is not to be immediately separated. When stored in this way blood samples can be separated successfully on the Parsortix system for up to 48 hours post draw. 

WARNING: Do not mount blood sample tubes made of glass onto the Parsortix PR1 as there is a risk of breakage and injury.

Page 12: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015

6 PR1‐OM‐CParsortixPR1UserManual

2 HowtousetheParsortixPR1

2.1 Instrumentprotocols

The series of processes performed in routine operation, summarised in Figure 

1.4,  are  implemented  by means  of  automated  protocols  embedded  in  the 

Parsortix PR1.  The processes and their respective protocols are summarised in 

Table 2.1. 

Table 2.1 Operation processes of the Parsortix PR1. 

Process  Description  Protocol name  Time (approx.) 

Priming  

Prepares the instrument and a new cassette to receive a blood sample 

PX2_P  15 min 

Separation  A blood sample is placed on the instrument and separated in the cassette 

PX2_S99F  2 h for an 8 ml sample 

Pre‐harvest flush  Flushing the instrument separation flowpath to remove residual blood cells prior to harvest 

PX2_CT2  25 min 

Harvest  Removal  of the cells captured in the cassette and collection for downstream analysis 

PX2_H  5 min 

In‐cassette staining  Immune fluorescent staining of captured cells  

PX2_S99F_stain4H  6.5 h 

Short clean  Cleans and resets the instrument 

PX2_CT  45 min 

Full clean  Cleans and resets the instrument including the staining reagent lines 

PX2_C  1 h 15 min 

 

2.2 Initialchecklist

Before running a protocol on the Parsortix PR1,  it  is  important to ensure that 

the following conditions are fulfilled: 

1. The reagents required for normal operation are in place according to 

Table 2.2. 

2. The Parsortix PR1 must have been cleaned after the last operation. 

3. An empty 50 ml Falcon tube and a cleaning cassette are in place. 

4. The reagent tubes 1‐6 are empty and the lines are clean. 

5. The harvesting valve is turned clockwise 

6. The harvest waste tube is empty.  

 

 

 

 

Important information 

First time here? Go to Section 8  for important Health and Safety advice. 

Important information 

The liquid in the waste reservoir must be less than 400 ml to avoid overfilling and system pressurisation. If necessary, empty the bottle and replace 50 ml Distel before running a protocol. 

If the waste reservoir is accidentally overfilled (see section 2.2), abort the protocol and switch off the instrument. 

The system does not provide feedback regarding air or incorrect fluids entering the instrument, nor if the waste reservoir is full. 

Page 13: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015 

7PR1‐OM‐CParsortixPR1UserManual

Table 2.2 Reagents required for normal operation 

Reagent  Detail  Location  Volume  Comment 

Phosphate buffered saline (PBS) 

10 mM sodium phosphate / 150 mM sodium chloride, pH 7.3‐7.5 

250 ml buffer reservoir  

50‐200 ml  Other PBS formulations or isotonic saline buffered with other reagents (e.g. Hepes) can be used 

Disinfectant: Either Distel or sodium hypochlorite bleach 

Undiluted solution 

500 ml waste reservoir 

50 ml  (Empty reservoir when the waste volume reaches 400 ml and replace 50 ml of disinfectant) 

50 ml enables a minimum concentration of 10% v/v as bottle fills with waste fluid.  

Alkaline cleaning reagent: Either Decon Decomatic or LabKlenz 110 

Decon Decomatic, or LabKlenz 110, at 10% v/v in distilled water 

50 ml reagent tube connected to line labelled “C” 

15‐50 ml   

Priming fluid: Ethanol 

100%  50 ml reagent tube connected to line labelled “P” 

10‐50 ml   

Note: Contact ANGLE for a list of branded reagent suppliers.   

 

   

Page 14: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015

8 PR1‐OM‐CParsortixPR1UserManual

2.3 Priming

The priming process prepares the  instrument to receive a blood sample. This 

process  involves  filling  a  new  cassette  with  buffer  solution.  The  priming 

process must follow an instrument cleaning process.  

 

To run a priming process: 

1. Ensure that the system is ready to run. Refer to the Initial Checklist in 

Section 2.2.  

2. In the instrument’s main menu screen, select the protocol PX2_P using 

the scroll arrows. 

3. Press [Run] then [Start]. 

4. When prompted, insert a new cassette.  

5. Press [OK] to continue the priming process.  

6. At prompt  “Finished P”, press  [OK]  then  [Continue]  to  return  to  the 

main menu on the screen. 

7. This process takes approximately 15 minutes to complete. 

   

Parsortix PR1 workflow 

 

In‐cassette staining

Cell harvest

Short clean

Priming

Blood separation

Full clean

IMPORTANT: A cleaning process must always be run after the last separation, harvest or staining process. If in doubt, the instrument should be cleaned again before running the priming protocol (see the description of the cleaning process in Section 2.6). 

Page 15: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015 

9PR1‐OM‐CParsortixPR1UserManual

2.4 Bloodseparation

If  blood  has  been  stored  in  a  4°C  refrigerator,  it  should  be  allowed  to 

equilibrate to room temperature by placing it on a roller mixer for a minimum 

of 20 minutes. 

After priming, run a separation process as follows: 

1. In  the main menu  screen,  select protocol PX2_S99F  and press  [Run] then [Start]. 

2. At prompt “Rinse Vacutainer” 

Pull the 50 ml Falcon tube (left in place after the previous cleaning as explained in Section 2.6) off its mount keeping the line inside the tube. 

Press [OK] to start the rinse. Collect the rinse fluid inside the Falcon tube. 

Fully remove the Falcon tube. Avoid flicking fluid off the tip of the line (see useful tip on sidebar). 

3. At prompt “Attach Vacutainer” 

Invert the Vacutainer containing the blood sample several times to fully  re‐suspend  the blood  cells,  then  immediately mount  it onto the instrument. 

A  twisting action will help  to push  the Vacutainer over  the O‐ring and into position. 

Secure  the  Vacutainer  by  pushing  it  into  the  vertical  position (Figure 2.1). 

Press [OK].  

 Figure 2.1 Blood sample Vacutainer in the secure vertical position. 

4. At prompt “Start?” 

Press [OK] to start the separation. This process takes approximately 2 hours for an 8‐10 ml sample. 

5. After  approximately  30  min,  re‐suspend  the  settled  blood  cells  by tapping  the  bottom  of  the  Vacutainer  while  it  sits  in  the  inclined position (Figure 2.2). 

 

Parsortix PR1 workflow 

 

In‐cassette staining

Cell harvest

Short clean

Priming

Blood separation

Full clean

Useful tip 

Use an alcohol damp wipe to avoid splash from the tip of the line while removing the Vacutainer.

 

Page 16: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015

10 PR1‐OM‐CParsortixPR1UserManual

 Figure 2.2 Re‐suspending the settled blood cells by tapping the Vacutainer. 

 

Re‐secure the Vacutainer by pushing it into the vertical position.  

6. At prompt “Finished S99F”, press [OK] then [Continue] to return to the main menu on the screen. 

 At the end of the separation process, the blood should have been drawn from 

the Vacutainer and passed through the cassette. The cassette will have been 

rinsed with buffer. There  should be only  trace  amounts of blood  left  in  the 

Vacutainer and  the cassette should appear clear of blood  to  the eye. At  this 

stage, the target cells will have been captured in the cassette. 

   

Page 17: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015 

11PR1‐OM‐CParsortixPR1UserManual

2.5 Cellharvest

The cell harvesting process enables pulses of  liquid to be passed through the 

cassette  in  reverse  direction.    The  liquid  displaces  the  captured  cells  and 

pushes them through the harvest line into a collection tube or multiwell plate. 

Cell harvesting can only follow directly from a blood separation process.  

In  order  to  minimise  residual  blood  cells,  it  is  strongly  recommended  to 

perform a pre‐harvest flush of the instrument.  (This process can be left out if 

warranted by operational needs, but doing  so will  considerably  increase  the 

number of  residual nucleated blood cells  in harvests.)   The pre‐harvest  flush 

procedure is:     

1. In the main menu screen, select the protocol PX2_CT2 and press [Run] 

then [Start]. 

2. At prompt  “Insert cleaning cassette” 

Open  the  clamp,  remove  the  separation  cassette, examine under microscope if intended and keep safe.  

Insert the cleaning cassette and press [OK]. 3. At prompt “Empty rgt tubes” 

Ensure reagent tubes are empty. 

Press  [OK]  to  start  the  flush  process.  This  process  takes 

approximately 25 minutes to complete. 

4. Depending on  experimental need,  the  cells  captured  in  the  cassette 

can be examined by microscope at  this point.   Otherwise,  keep  in a 

safe place until ready to re‐insert into the clamp. 

5. At prompt “Finished CT2” press [OK] then [Continue] to return to the 

main menu on the screen. 

6. Remove the cleaning cassette and reinsert the separation cassette. 

The  harvested  cells  can  be  collected  in  different  collection  vessels  such  as 

Eppendorf tubes or multiwell plates. Once the collection vessel  is ready, start 

the harvesting process: 

1. Ensure the harvest waste tube is empty (labelled 9 on Figure 1.1). 

2. In  the main menu screen, select  the protocol PX2_H and press  [Run] 

then [Start]. 

3. When prompted, rotate the harvest valve anticlockwise to the position 

HAR and press [OK]. 

4. At prompt “Start”  

Remove  the  harvest  line  from  the  harvest  waste  tube  and 

clean it with an alcohol soaked wipe. 

Place a collection vessel (e.g. an Eppendorf tube) beneath the 

harvest line  

Press  [OK]  to  start  the harvest. A volume of 200 l will  flow through the harvest line. 

 

Parsortix PR1 workflow 

 

In‐cassette staining

Cell harvest

Short clean

Priming

Blood separation

Full clean

Page 18: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015

12 PR1‐OM‐CParsortixPR1UserManual

5. At prompt “Further flush?” 

Press  [Yes]  or  [No]  depending  on  the  requirements  for 

downstream  analysis.  If  selected  YES,  a  further  1 ml will  be 

collected. 

6. Place  the harvest  line back  in  the harvest waste  tube, ensuring  that 

the tip does not sit in cleaning fluid.  

7. On prompt,  rotate  the harvest valve  clockwise  to  the position  “SEP” 

and press [OK]. 

8. At prompt  “Finished H”  press  [OK]  then  [Continue]  to  return  to  the 

main menu on the screen. 

   

Important information 

A further 1 ml flush may result in a higher level of white blood cells in the collection tube along with any captured target cells. For high captured cell purity applications a further flush is not recommended. 

Page 19: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015 

13PR1‐OM‐CParsortixPR1UserManual

2.6 Cleaning

The  cleaning  process  is  used  to  clean  the  flow  path  and  the  critical 

components of  the  system. At  the end of a workflow  (which, depending on 

experimental need may be the end of a separation, of an  in‐cassette staining 

process, or of a harvest),  it  is necessary  to clean  the  instrument  to reset  the 

system  leaving  it  ready  for  the  next  process.  Two  cleaning  protocols  are 

available:  

A  full  cleaning  process  (PTX2_C)  is  necessary  only  after  in‐cassette 

staining  (or after other protocols  that may use  the  reagent  tubes)  to 

clean the whole system including lines 1‐6 used for staining reagents. 

When reagent  lines are not used, a short cleaning process  (PTX2_CT) 

can be performed at the end of the harvest or, in cases where cells are 

not being harvested, at the end of the separation. This process takes 

approximately 45 minutes to complete.  

It is also recommended to perform a full clean before priming for experiments 

involving  in‐cassette staining. The  full process takes approximately 1 hour 15 

minutes.  

To run a cleaning process: 

1. In  the main menu  screen,  select  the  appropriate  cleaning  protocol 

(PX2_C  for  full clean or PX2_CT  for short clean) and press  [Run]  then 

[Start]. 

2. At prompt  “Insert cleaning cassette” 

Open the clamp and remove the separation cassette.  

Insert a cleaning cassette and press [OK]. 3. At prompt “Empty rgt tubes”  

Empty  reagent  tubes  1‐6  if  used  for  in‐cassette  staining.  Retain 

reagents if necessary. 

Press [OK] to start the clean process.  

4. At prompt “Finished C” or “Finished CT” press [OK] then [Continue] to 

return to the main menu on the screen. 

5. Empty the fluids from the harvest waste tube. 

6. In the case of a full clean process (PX2_C), discard the reagent tubes 1‐

6, and clean  the outside of  the  lines with an alcohol  soaked wipe  to 

remove  residual  cleaning  fluid. Replace  the  reagent  tubes with  new 

ones. 

7. Remove the vacutainer from its mount and discard safely.  

8. Clean  the O‐ring and  the outside of  the  line using an alcohol  soaked wipe. 

9. Place a 50 ml Falcon tube on the mount. This tube collects rinse fluid 

in the separation process and can be reused. 

10. A twisting action will help to push the Falcon tube over the O‐ring and into position. 

11. Secure the Falcon tube by pushing it into the vertical position.   

Important information 

The mount must be free of any visible blood residue. If you find it difficult to remove the blood, use a wipe soaked with a mild, dilute cleaning fluid (e.g. Distel or similar), followed by an alcohol soaked wipe.

Parsortix PR1 workflow 

 

In‐cassette staining Harvest

Short clean

Priming

Blood separation

Full clean

Page 20: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015

14 PR1‐OM‐CParsortixPR1UserManual

3 In‐cassetteimmunofluorescencestaining

3.1 Background

Blood  separation  can  be  followed  by  immunofluorescence  staining  of  the 

captured cells for analysis under a fluorescence microscope.  

The harvested cells can be stained  in‐cassette using the  instrument’s reagent 

lines.  

The automated protocol follows the procedure described below: 

1. Fixative is drawn into the cassette and incubated for 20 min. 

2. Permeabilization reagent is drawn into the cassette and incubated for 

5 min,  followed by  two additional draws of permeabilization  reagent 

and incubation for 5 min. 

3. Primary  antibodies  are  drawn  into  the  cassette  and  incubated  for  4 

hours, followed by PBS wash. 

4. Secondary  antibodies  and  DAPI  are  subsequently  drawn  into  the 

cassette and incubated for 1 hour, followed by PBS wash. 

The  Parsortix  PR1  is  designed  for  end‐user  versatility.  Antibodies  and 

concentrations should be tested and optimised by the end user before use.  

3.2 Procedure

Specifically,  it  is  strongly  recommended  that  no more  than  48  hours  have 

passed between a Full Clean and the  instrument’s priming at the start of the 

following process.  

 

To prepare for the in‐cassette staining process: 

1. Add the following to the reagent tubes using fresh clean 15 ml Falcon 

tubes: 

Line 1: 2 ml 4% formaldehyde 

Line 2: 2 ml 0.1% Triton X‐100 in PBS 

Line 3: 1 ml blocking buffer 

Line  4:  1  ml  primary  antibodies  (at  optimised  concentration  in 

blocking buffer) 

Line  5:  1 ml  secondary  antibodies  (at  optimised  concentration  in 

blocking buffer containing 1:5000 DAPI) 

INFORMATION: Prior to experiments involving in‐cassette staining, the system must be fully cleaned using the Full Clean protocol PX2_C prior to blood separation, which cleans the separation flow path and the reagent lines.  

Useful tip 

It is possible to perform the separation and the in‐cassette staining as an integrated process using protocol PX2_S99F_stain4h 

Ensure that the staining reagents are in place before starting the combined protocol. 

Parsortix PR1 workflow 

 

In‐cassette staining

Cell harvest

Short clean

Priming

Blood separation

Full clean

Page 21: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015 

15PR1‐OM‐CParsortixPR1UserManual

2. In the main menu screen, select protocol PX2_stain4h and press [Run] 

then [Start] to initialise the staining process.  

3. At prompt “Finished” press [OK] then [Continue] to return to the main 

menu on the screen. The cassette can now be removed and visualised 

using fluorescence microscopy and the appropriate channels. 

4. After  in‐cassette  staining,  a  full  cleaning  process must  be  run  using 

protocol PX2_C (Section 2.6). 

 

   

Page 22: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015

16 PR1‐OM‐CParsortixPR1UserManual

4 Maintenance

 

Following every separation and harvest or in‐cassette staining: 

Run a cleaning process to minimise the risk of cross contamination 

between experiments and the build‐up of cellular or plasma residue in 

the system’s flowpath. 

Check the level of fluid in the waste bottle. If more than 400 ml, empty 

the container and add 50 ml Distel disinfectant. 

Check the level of fluid in the buffer reservoir and add more if 

necessary.  

On a daily basis, ensure that the instrument is clean and undamaged: 

Check that all the external pipework, unions and other fittings are not 

kinked or damaged. 

Check that the tubing connecting the instrument to the clamp is not 

kinked or blocked.   

Check that the waste and buffer reservoir connectors are tight at both 

the instrument bulkhead and at the respective reservoir lids. 

Check that the O‐rings on the Vacutainer mount are in place, clean and 

in good condition.  (Replacement O‐rings are available from ANGLE.)   

Check that the clamp is clean.   

Check that the filters on the buffer and waste reservoirs are in place 

and are not contaminated or damaged. Replace if necessary. 

Clean the external cover of the instrument with a weak detergent or 

70% ethanol or isopropanol followed immediately by a water damp 

towel.  

If necessary, remove and clean the grey housing surrounding the 

Vacutainer. 

Contact ANGLE Technical Support if kinks, blocks or other damage are 

identified.   

On a 3‐monthly basis: 

Replace the filters on the buffer and waste reservoirs (see Section 9.2 

for part numbers). 

Replace the two O‐rings on the Vacutainer mount.   

   

Important information: 

Contact ANGLE Technical 

Support when damage to 

any component of the 

instrument is found or 

suspected. 

Contact details can be 

found in Section 9.1. 

CAUTION: The routine maintenance plan must be complied with at all times to ensure results are not adversely affected by contamination or sub‐optimal system performance. 

Page 23: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015 

17PR1‐OM‐CParsortixPR1UserManual

On an annual basis and in conjunction with ANGLE Technical Support: 

Replace the internal filters and precision syringe components. 

Calibrate and check the integrity of the internal pneumatic system. 

Check the integrity of the internal fluidic system. 

Replace the moving components in accordance with the instrument 

service schedule. 

   

Page 24: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015

18 PR1‐OM‐CParsortixPR1UserManual

5 TroubleshootingThe  following  flowchart  describes  the  recommended  actions  to  address 

problems during the separation procedure.  

Contact Angle Technical Support if further advice is required. 

 

Figure 5.1 Actions for addressing common problems during the separation procedure. 

   

Important information: 

Biohazard: Clean leaks 

safely following local 

Health and Safety 

procedures. Relevant 

advice can be found in 

Section 8.1.1.  

Page 25: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015 

19PR1‐OM‐CParsortixPR1UserManual

6 AssemblyandinstallationCarefully follow the instructions outlined in this Section to assemble and install 

the Parsortix PR1. 

6.1 Positioningandinterference

The  Parsortix  PR1  device  complies  with  the  emission  and  immunity 

requirements  described  in  the  IEC  61326  standard  and  as  set  out  in  the 

declaration of conformity in Section 12 of these instructions. 

This device complies with Part 15 of the FCC Rules. Operation is subject to the 

condition that this device does not cause harmful interference and this device 

must accept any interference received, including interference that may cause 

undesired operation. 

Do not use this device in close proximity to sources of strong electromagnetic 

radiation  (e.g. unshielded  intentional RF sources), as these can  interfere with 

the proper operation. 

6.2 Unpackthebox

Remove all  the  components  from  the  shipping  carton and verify  that all  the 

items listed in the enclosed contents list are present and not damaged. Do not 

dispose of the packaging until all the components are removed and accounted 

for.  Some  components  (e.g.  spare parts) may not be used  immediately  and 

should be stored safely for future use.  

Place  the  instrument  in  the  selected  final  location  on  the  benchtop.   Once 

assembled, do not move the equipment whilst the reagent bottles and tubes 

are connected as these are not held securely in place and may fall and break. 

6.3 Attachthedriptray

The  drip  tray  surrounds  the  encasing  for  the  blood  sample  tube  and  helps 

contain spillages or leaks.  

Attach the drip tray over the sample mount as shown in Figure 6.1. It is held in 

place by a magnetic catch. 

WARNING: The electromagnetic environment  that  this device is to be located within should be evaluated prior to operation of the device.  

Page 26: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015

20 PR1‐OM‐CParsortixPR1UserManual

 

Figure 6.1 Positioning of the drip tray. 

6.4 Applythenon‐slippads

Adhere the non‐slip pads onto the surface of the reservoir bottles bay on the 

left (large pad) and the reagent rack bay on the right (small pad). First, identify 

the correct position and then peel the backing paper from the pads and apply 

them (Figure 6.2). 

   

Figure 6.2 Placement of the non‐slip pads on the Parsortix PR1: large pad on the left side bay and small pad on the right side bay. 

The waste and buffer reservoirs are Duran type glass bottles and are provided 

with caps pre‐fitted with tubing, connectors and PTFE filters. 

Identify each reservoir bottle and its cap according to the information outlined 

in Table 6.1. 

   

Page 27: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015 

21PR1‐OM‐CParsortixPR1UserManual

Table 6.1 Identification of bottles and pre‐fitted caps. 

Reservoir  Bottle volume 

Tubing in cap 

Inlet tube label 

Connection port in the Parsortix PR1 

Waste  500 ml  Two lines  W  Left and middle ports Buffer  250 ml  Single line  B  Right port 

 

 

Figure 6.3 Waste (back) and buffer (front) bottles with pre‐fitted caps. 

Attach  the  caps  to  the  bottles  (Figure  6.3)  and  connect  the  tubing  to  the 

Parsortix PR1 (Figure 6.4): 

Buffer  reservoir:  carefully  remove  the  red  protection  cap  from  the 

right connection port on the front of the  instrument. Manually screw 

the plastic tubing connector into the hole. Place the bottle on the right 

of the bottle bay. 

Waste reservoir: carefully remove the red protection cap from the left 

and middle connection port on the front of the  instrument. Manually 

screw the plastic tubing connectors into the holes. Place the bottle on 

the left of the bottle bay. 

   

Figure 6.4 Connection of the buffer reservoir (left image) and waste bottle (right image). 

Important information: 

The connection assembly 

comprises a yellow ferrule, 

a metal sleeve and a 

plastic locking nut. The 

yellow ferrule forms a seal 

as the plastic nut is 

screwed into the 

connection port until 

resistance is felt then one 

further quarter turn . 

 

Page 28: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015

22 PR1‐OM‐CParsortixPR1UserManual

6.5 Connectthecassetteclamp

The clamp is the metal case that securely holds the separation cassette during 

operation of  the Parsortix PR1  (Figure 6.5).  It  is  supplied with  its connection 

tubes  already  attached.  The  inlet  and outlet  tubes  are  labelled  “I”  and  “O”, 

respectively. The cassette connects to the back of the harvest valve mounted 

on the right side of the instrument (Figure 6.6). 

  

Figure 6.5 Cassette clamp (1) connected to the harvest valve (2). 

  Figure 6.6 Rear view of the harvest valve. 

 

To connect the clamp, carefully remove the red protection covers on the two 

unused ports  at  the back of  the harvest  valve  and  connect  each  line  in  the 

correct port according to the information in Table 6.2. 

 Table 6.2 Identification of connection ports in the harvest valve. 

Line  Port number 

Position (as viewed from rear of valve) 

Inlet “I”  1  Left hand (9 pm)  

Outlet “O”  4  Right hand (3 pm) 

inlet outlet

Page 29: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015 

23PR1‐OM‐CParsortixPR1UserManual

6.6 Connectthereagenttubes

The  reagent  tubes  hold  the  fluids  required  for  priming  and  cleaning  the 

system, as well as the reagents for in‐cassette cell staining.  

The cluster of tubing lines emerging from the right hand side of the instrument 

enables the system to dispense reagents. The lines are tagged with a coloured 

label according to the diagram in Figure 6.7. The lines connect to 15 ml or 50 

ml plastic tubes held in an 8‐space metal rack placed on the right side bay.  

 

Figure 6.7 Reagent lines identification labels. 

To connect  the  lines, remove  the end cap  from  the  line and carefully slide  it 

through the luer cap of a tube (Figure 6.8). A small amount of force might be 

necessary but take care not to kink the line. The line should reach just short of 

the bottom of the tube when the cap is screwed on. Refer to the table below 

to match the  lines with the correct reagent tube. Once the  line  is connected, 

place the tube in the 8‐space metal rack.  

Table 6.3 Identification of reagent tubes and lines. 

Line  Quantity  Tube size  Use 

1‐6 (coloured)  6  15 ml  Staining reagents P  1  50 ml  Priming fluid C  1  50 ml  Cleaning fluid 

 

   

Figure 6.8 Connection of lines to reagent tubes. 

  

Page 30: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015

24 PR1‐OM‐CParsortixPR1UserManual

6.7 Connecttheharvestvalvewastetube

The waste  fluid produced during  the cleaning process  is collected  in a 50 ml 

plastic tube held in a single rack behind the harvest valve. 

The harvest valve has an attached  line with a  free end. To connect  this  line, 

remove the black end cap and carefully slide it through the luer cap of a 50 ml 

tube (Figure 6.9). Place the tube on the single rack. 

 

Figure 6.9 Connection of the harvest waste line to the waste tube. 

6.8 StarttheParsortixPR1

Connect the power cord to the Parsortix PR1 and to the mains outlet. Turn on 

the  equipment  using  the  power  switch  on  the  left  of  the  instrument.  The 

internal  instrumentation will be heard moving as the start‐up sequence takes 

place. The display screen will report the progress before presenting the main 

menu with the instruction [Select Protocol] (Figure 6.10). Now the system has 

successfully booted and the equipment is ready for use. 

 

Figure 6.10 Control panel showing the main menu on screen. 

Important information 

Do not connect external 

equipment to the 

instrument, with the 

exception of USB drives. 

The sockets on the left 

hand side are for 

manufacturers use only. 

Page 31: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015 

25PR1‐OM‐CParsortixPR1UserManual

6.9 Beforefirstuse

After assembly and before the first use, a new Parsortix PR1 must undergo one 

priming  and  two  full  cleaning  processes  to  clean,  rinse  out  and  reset  the 

instrument. Refer to Section 2 for instructions on how to run the instrument. 

Page 32: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015

26 PR1‐OM‐CParsortixPR1UserManual

7 Technicalspecifications 

Supply voltage  100‐240 V AC 50/60Hz input 24 V DC 4.16 A output 

Fuse rating  3.15 A Fast 20 mm 

Operating conditions  18°C to 25°C  95‐105kPa 

Storage conditions  0°C to 40°C   95‐105kPa 

Maximum operating altitude  2000 m 

 

The Parsortix PR1 has been tested  in accordance with the following electrical safety standards: 

IEC 61010‐1:2010, EN 61010‐1:2010 

IEC 61010‐2‐081:2001+A1:2003, EN 61010‐2‐081:2002 

UL61010‐1:2012 

CAN / CSA C22.2 No. 61010‐1‐12  

A Declaration of Conformity for EU markets can be found in Section 12 of this document. 

 

The Parsortix PR1 is certified under certificate number: U8 13 08 85079 001 as compliant with the relevant standards in the US and Canada. 

 

 

Page 33: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015 

27PR1‐OM‐CParsortixPR1UserManual

8 HealthandsafetyThe Parsortix cell separation system  is designed for use as part of biomedical 

research. It must: 

Be  situated  and  operated  only  in  facilities  with  the  specialised 

infrastructure  and  general  equipment  required  for  biomedical 

research or clinical laboratory activity. 

Be operated by personnel who have relevant training, knowledge and 

experience  in  biomedical  research  and/or  clinical  laboratory 

operations,  including  equipment  and materials  required  to  perform 

laboratory tasks correctly and safely. 

Be  used  in  association  with  and  within  existing  management  and 

technical  processes  concerning  correct  and  safe  use  of  chemicals, 

biological  materials  and  other  hazards  commonly  encountered  in 

biomedical research and clinical operations. 

It  is not normally necessary to use the equipment within hazard containment 

equipment;  however,  users  should  consider  its  use  as  part  of  their  risk 

assessments,  taking  into  account  the  specific  reagents  and  other materials 

being used. 

8.1 Potentialhazards

Table 8.1 Risks associated to the operation of the Parsortix PR1. 

Risk  Description  Actions to minimise risk 

Biohazards 

  

Exposure to blood and other biohazards can result in infection 

Exposure during phlebotomy and sample handling: establish appropriate operating procedures, use containment and personal protective equipment and consider vaccinations. Exposure during equipment operation: adhere to the instructions in Section 2.4 for mounting and removing sample tubes and wear personal protective equipment.  Clean spillages as described in this section. Handling of waste: ensure that the waste reservoir is prepared as instructed in Section 2.2 and that the waste material produced is handled and disposed of appropriately. Wear personal protective equipment.  

WARNING:  The  use  of  the  Parsortix  cell  separation  system involves two main potential hazards as described in Table 8.1. 

Page 34: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015

28 PR1‐OM‐CParsortixPR1UserManual

Chemical exposure 

Exposure to chemicals that may be caustic, noxious, toxic, volatile or flammable 

Establish operation procedures for risk assessment, storage, handling, deployment and disposal of chemical reagents. Wear personal protective equipment. 

 

8.1.1 BiohazardsThe use of  the Parsortix cell separation system may  involve any or all of  the 

following biohazards: 

blood and its risk of pathogenic infection 

other body fluids (e.g. urine) and their risk of pathogenic infection 

primary or cultured cells that may be hazardous to health if they enter 

a person’s tissues 

primary or cultured cells that may themselves be hazardous to health 

if they enter a person’s tissues 

biochemicals that may be hazardous to health if they contact or enter 

a person’s tissues. 

Handling bio‐hazardous materials, particularly blood for the Parsortix system, 

presents the risk of microbial and viral infection through accidental injury with 

contaminated sharps or by exposure  to blood, or blood aerosol, without  the 

use of protective equipment. 

The  operation  of  the  Parsortix  system  requires  handling  blood  for  two 

separate aspects: 

phlebotomy and subsequent sample preparation 

mounting a blood sample on the instrument during operation 

Phlebotomy  and  subsequent  manipulation  of  blood  samples  must  involve 

procedures,  containment  equipment  and  protective  clothing  for  protection 

against  blood  and  infection.  Such  measures  are  likely  to  form  part  of 

mandatory policies  imposed by  institutions  in which Parsortix equipment will 

be  used.  Vaccination  should  be  considered  as  a  protection measure  for  all 

personnel whose work involves handling or exposure to blood. 

The Parsortix PR1 and disposable separation cassettes have been designed to 

minimise risk of  inadvertent exposure  to blood. The design  includes a closed 

flow  path  between  blood  sample  tubes  and  final  outflow  into  waste  fluid 

receptacles. However, mounting on  and  removing blood  sample  tubes  from 

the equipment  involves  them being opened and a  small  risk of blood egress 

and exposure from residue present on an instrument component. 

Some  simple  actions  negate  the  minor  risk  of  exposure  involved  in  the 

operation of the Parsortix system: 

Page 35: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015 

29PR1‐OM‐CParsortixPR1UserManual

Follow  the advice  for mounting and  removing  sample  tubes  (Section 

2.4). 

Wearing  standard  protective  equipment  including  gloves  and  a  face 

mask. 

In  case  of  a  leak  during  operation  (e.g.  in  the  cassette‐clamp 

assembly), or  if bio‐hazardous material  is  accidentally  spilled on  the 

instrument,  it must  first  be mopped  away  using  a  paper  towel  or 

similar, then cleaned further using 5% Decon Decomatic or equivalent. 

Alternatively, if there is particular concern about the biohazard caused 

by  a  spill,  after  the  material  is  mopped  away  the  area  can  be 

decontaminated and  cleaned using 10% Hydrogen Peroxide  followed 

by 5% Decon Decomatic or equivalent. 

Hazardous waste  produced  through  instrument  operations must  be 

handled,  stored  and  disposed  of  with  full  regard  to  appropriate 

laboratory safety and environmental regulations.  

It  is not normally necessary  to operate  the  instrument within a containment 

cabinet. However, the equipment can be used within containment equipment 

if this is preferred, or if the application or situation in which the equipment is 

being  deployed  is  considered  to  involve  particular  risks  of  infection.  A 

horizontal flow cabinet with an air flow that enters from behind the user, and 

that allows open access at  the  front  is particularly convenient, especially  if a 

microscope is to be used to observe separations. 

8.1.2 HandlingreagentsandsolutionsAll chemicals and biochemicals required for use with the equipment are those 

in  common use  in biomedical  research. Though  several are hazardous,  their 

safe  use  is  possible  through  policies  and  procedures  commonly  applied  in 

laboratories  for  risk assessment,  storage, handling, deployment and disposal 

of chemicals and biological reagents. 

8.2 Labelling

The  Parsortix  PR1  instrument  carries  the  following  hazard,  warning  and 

operational labels: 

SYMBOL  

MEANING   

 

READ INSTRUCTIONS  Read the instruction manual for description of principles of operation and details of potential hazards 

 

BIOHAZARD  This instrument contains potential biohazards 

Page 36: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015

30 PR1‐OM‐CParsortixPR1UserManual

 

CE MARK  Indicates compliance with a range of European Directives as set out in the accompanying Certificate of Conformity 

 

MANUFACTURER  Name and address of the instrument manufacturer  

 

CAUTION  Consult the instructions for use for important warning or cautionary information. 

 

WEEE  Do not dispose of in domestic waste. 

 

LOT NUMBER  Manufacturer’s lot number for traceability of components 

 

SERIAL NUMBER  Manufacturer’s serial number for instrument identification 

 

SINGLE USE ONLY  The components must not be used more than once to avoid the risk of cross‐contamination 

  

DO NOT USE IF PACKAGING IS DAMAGED 

Do not use the labelled component if the packaging is already open in order to avoid the risk of contamination 

  

KEEP AWAY FROM SUNLIGHT 

Parts labelled are sensitive to sunlight and should be stored away from sunlight 

 

FRAGILE  Indicates that the contents of a package or container are fragile and should be handled with care 

 10 

CONTAIN SUFFICIENT FOR 10 TESTS 

Indicates the number (10 as displayed) of tests (or related consumable items) supplied 

 

 

DIRECT CURRENT  Indicates direct current supply to the equipment. 

Page 37: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015 

31PR1‐OM‐CParsortixPR1UserManual

 

 

FUSE  Location, value and characteristics of the external system fuse. 

 

THIS WAY UP  Preferred packaging orientation for storage and shipping. 

 

KEEP DRY  Packaging should be kept dry at all times during storage and shipping. 

 

 

HARVEST/SEPARATE  A switchable valve to select cell separation or cell harvesting mode. 

 

NOTIFIED BODY SYMBOL 

The product has been safety tested by TÜV for conformity to IEC 61010‐1 or equivalent standards for markets including EU, USA and Canada. 

Page 38: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015

32 PR1‐OM‐CParsortixPR1UserManual

9 Technicalsupport

9.1 Contactus

Manufacturer and authorised manufacturer’s representative in the EU:.  

ANGLE Europe Ltd 3 Frederick Sanger Road The Surrey Research Park Guildford, Surrey GU2 7YD United Kingdom Tel: +44 (0) 1483 685830 Email: eu‐[email protected] www.angleplc.com 

 Authorised manufacturer’s representative in the USA:  

ANGLE North America Inc University City Science Center  3711 Market Street  Floor 8 Philadelphia PA 19104 Tel: +1 215 966 6240 Email: us‐[email protected] www.angleplc.com  

9.2 Consumablesandspares

The  following  consumables  are  available  from  ANGLE,  or  in  the  case  of 

standard  laboratory  components,  a  range  of  scientific  equipment  and 

consumables providers. 

Table 9.1 Consumables and spares 

Item  Supplier and part number 

PR1 user manual and user manual CD  Parsortix PR1‐IFU Ten disposable Parsortix 10 μm cell separation cassettes 

Parsortix GEN3P10 or Parsortix GEN3D10 

Parsortix cell separation cassette clamp assembly with inlet and outlet lines attached, each line with assembly for connection to instrument 

Parsortix GEN3Clamp  

Two spare nut + ring + ferrule + gasket sets for clamp 

Parsortix GEN3Gasket 

One flushnut tool (for use in adjustment of clamp fittings) 

Parsortix GEN3Flushnut  

Page 39: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015 

33PR1‐OM‐CParsortixPR1UserManual

 Two replacement O‐rings for blood sample connection 

Parsortix O‐rings 

One three port bottle lid with filter and port plug fitted (for buffer reservoir) 

Parsortix BuffLid  

One three port bottle lid with filter fitted (for waste reservoir)  

Parsortix WasteLid  

One power supply with UK and US mains leads 

Parsortix PSU 

Two reagent tube racks  

Parsortix Rack 

Six 15ml reagent tube caps with ferrules fitted  

Parsortix 15mlLid 

Three 50ml reagent tube caps with ferrules fitted  

Parsortix 50mlLid 

Two rubber non‐slip mats  Parsortix Mats EDTA blood container / phlebotomy tubes BD 10mL Vacutainer™ Plastic Blood Collection Tubes with K2EDTA  

BD 366643 (US, Canada)  BD 367525 (Europe, Latin America, Asia Pacific, Australia, New Zealand, Canada) 

250 ml Glass laboratory bottle (Buffer reservoir) 

Fisher Scientific FB33145 or  Camlab Part: 1146888 

500 ml Glass laboratory bottle (Waste reservoir) 

Fisher Scientific BTF‐730‐090F or Camlab Part: 1146890 

15 ml Falcon conical reagent tube  BD or Corning 352097 50ml Falcon conical reagent tube  BD or Corning 352070 Plastic nut for bulkhead union fittings  Upchurch LT‐115 ¼‐28 Super 

Flangeless PEEK Short Nut Plastic ring and ferrule for bulkhead union fittings 

Upchurch P‐259X Super Flangeless Ferrule 

Waste reservoir bottle filter  PTFE Disposable Syringe Filter; 0.20 μm; 15 mm dia; nonsterile; Female Luer inlet, male Luer outlet. Cole Parmer WZ‐29550‐08 or equivalent 

Buffer reservoir bottle filter  Nylon Disposable Syringe Filter; 0.22 μm; 13 mm dia; nonsterile; Female Luer inlet, male Luer outlet. Cole Parmer WZ‐81054‐20 or equivalent 

 

 

 

 

 

Page 40: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015

34 PR1‐OM‐CParsortixPR1UserManual

9.3 Materialcompatibility

The Parsortix PR1 has been assessed for compatibility with a range of known 

aggressive chemicals and reagents as listed below. 

Table 2 List of compatible chemicals and reagents. 

Chemical/Reagent 

Sodium chloride in aqueous solution to a maximum concentration of 0.9% w/v 100% ethanol Hydrogen peroxide in aqueous solution to a maximum concentration of 10% v/v Triton X100 detergent in aqueous solution to a maximum concentration of 10% v/v Tween 20 detergent in aqueous solution to a maximum concentration of 5% v/v Sodium dodecyl sulphate (sodium lauryl sulphate) detergent in aqueous solution to a maximum concentration of 1% w/v Contrad 100 detergent in aqueous solution to a maximum concentration of 10% v/v Decon 90 detergent in aqueous solution to a maximum concentration of 2% v/v Bleach (sodium hypochlorite) in aqueous solution to a maximum concentration of 10%w/v Coomassie blue dye as commercial ready to use reagent Haematoxylin and Eosin dye combination as commercial ready to use reagent Wright’s reagent dye as commercial ready to use reagent Reagent grade Acetone (typically ≥99.5% purity) Reagent grade Butanol (typically ≥99.5% purity) Anhydrous glacial acetic acid 0.1M sodium hydroxide 

 

For assessment of  the  instrument compatibility with  chemicals and  reagents 

that  are  not  on  this  list,  please  contact  ANGLE  Technical  Support  prior  to 

deployment. 

   

Page 41: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015 

35PR1‐OM‐CParsortixPR1UserManual

10 DisposalofelectronicequipmentIt is important to understand and follow all laws regarding the safe and proper 

disposal of electrical instrumentation.  

Disposing of  this product  correctly will  save  valuable  resources  and prevent 

any potential negative effects on human health and  the environment which 

could otherwise arise from inappropriate waste handling. If you are unsure of 

your national requirements with respect to disposal please contact your  local 

authority or ANGLE Technical Support for further information. 

The  symbol  of  a  crossed‐out  wheeled  bin  on  the 

product means  that  the  device must  not  be  disposed 

via the general waste collection system.  

In  the  European  Union,  this  symbol  is  required  in 

accordance  with  the  Waste  Electrical  and  Electronic 

Equipment  (WEEE)  Directive  2012/19/EC  of  the 

European  Union  on  all  devices  put  on  the  European 

market after 13/August/2005. 

Please contact an ANGLE Technical Support  for appropriate decontamination 

information  and  details  on  the  take  back  program which will  facilitate  the 

proper  collection,  treatment,  recovery,  recycling,  and  safe  disposal  of  the 

device. 

   

Page 42: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015

36 PR1‐OM‐CParsortixPR1UserManual

11 Certificateofdecontamination 

An  electronic  copy  of  this  certificate  is  available  from  ANGLE  Technical 

Support.  ANGLE  will  not  accept  returned  equipment  without  a  valid  RMA 

number or a fully completed certificate. 

Page 43: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015 

37PR1‐OM‐CParsortixPR1UserManual

12 Declarationofconformity 

 

 

Page 44: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation
Page 45: PR1-OM-C Parsortix PR1 User Manual - ANGLE PLC · PR1‐OM‐C Parsortix PR1 User Manual 1 1 Before you start 1.1 What is the Parsortix PR1? The Parsortix system enables the separation

 

©ANGLEEurope2015 

PR1‐OM‐CParsortixPR1UserManual

DO NOT PRINT THIS PAGE Document Control – Information for Users Document Status: FINAL | Dated: 13 October 2015

Document Revision History

Approved by  Signature  Date 

Dr Shane Booth 

(Technical)   29 September 2015 

Martin Cooke 

(Regulatory) 

 

29 September 2015 

Michael O’Brien 

(Commercial, EU)  

13 October 2015 

Peggy Robinson 

(Commercial US)  

09‐Oct‐2015 

Version Issue Date Revision Author

OM - C1 1 28 August 2015 First draft for review Akemi Nogiwa

OM-C2 1 9 September 2015 Revision after circulation for comment

Shane Booth

OM-C3 1 23 September 2015

Revision from SB, MC and AN Akemi Nogiwa

OM-C 1 13 October 2015 Final approved version SCB