18

PowerPoint-presentatie Resuscitaire pp.pdfdr. Chris R. Mol, BME, NORTEC, 2015 Resuscitaire Use: Approach / priorities © Resuscitaire

  • Upload
    others

  • View
    18

  • Download
    0

Embed Size (px)

Citation preview

Page 1: PowerPoint-presentatie  Resuscitaire pp.pdfdr. Chris R. Mol, BME, NORTEC, 2015 Resuscitaire Use: Approach / priorities © Resuscitaire
Page 2: PowerPoint-presentatie  Resuscitaire pp.pdfdr. Chris R. Mol, BME, NORTEC, 2015 Resuscitaire Use: Approach / priorities © Resuscitaire
Page 3: PowerPoint-presentatie  Resuscitaire pp.pdfdr. Chris R. Mol, BME, NORTEC, 2015 Resuscitaire Use: Approach / priorities © Resuscitaire
Page 4: PowerPoint-presentatie  Resuscitaire pp.pdfdr. Chris R. Mol, BME, NORTEC, 2015 Resuscitaire Use: Approach / priorities © Resuscitaire
Page 5: PowerPoint-presentatie  Resuscitaire pp.pdfdr. Chris R. Mol, BME, NORTEC, 2015 Resuscitaire Use: Approach / priorities © Resuscitaire
Page 6: PowerPoint-presentatie  Resuscitaire pp.pdfdr. Chris R. Mol, BME, NORTEC, 2015 Resuscitaire Use: Approach / priorities © Resuscitaire
Page 7: PowerPoint-presentatie  Resuscitaire pp.pdfdr. Chris R. Mol, BME, NORTEC, 2015 Resuscitaire Use: Approach / priorities © Resuscitaire

dr.ChrisR.Mol,BME,NORTEC,2015

Resuscitaire Use: Approach / priorities

© Resuscitaire

Page 8: PowerPoint-presentatie  Resuscitaire pp.pdfdr. Chris R. Mol, BME, NORTEC, 2015 Resuscitaire Use: Approach / priorities © Resuscitaire
Page 9: PowerPoint-presentatie  Resuscitaire pp.pdfdr. Chris R. Mol, BME, NORTEC, 2015 Resuscitaire Use: Approach / priorities © Resuscitaire

dr.ChrisR.Mol,BME,NORTEC,2015

Apgar score and Apgar timer

© Resuscitaire

Thetestisgenerallydoneatone and five minutes after birth,andmayberepeatedlaterifthescoreisandremainslow.

• Apgarscores7andabovearegenerallynormal,• Apgarscores4to6fairlylow,• Apgarscores3andbelowaregenerallyregardedascriticallylow.

AnApgar Timer willsoundabuzzerforthedurationof3secondsatintervalsof1, 5, and 10 minutes,therebysupportingtheusertotaketheApgarscoresatpreciseintervals.

Alowscoreontheone-minutetestmayshowthattheneonaterequiresmedicalattentionbutdoesnotnecessarilyindicatealong-termproblem,particularlyifthescoreimprovesatthefive-minutetest.AnApgarscorethatremainsbelow3atlatertimes—suchas10,15,or30minutes—mayindicatelonger-termneurologicaldamage.

NotthattheApgartest'spurposeistodeterminequicklywhetheranew-bornneedsimmediatemedicalcare.Itisnotdesignedtopredictlongtermhealthissues.

Page 10: PowerPoint-presentatie  Resuscitaire pp.pdfdr. Chris R. Mol, BME, NORTEC, 2015 Resuscitaire Use: Approach / priorities © Resuscitaire
Page 11: PowerPoint-presentatie  Resuscitaire pp.pdfdr. Chris R. Mol, BME, NORTEC, 2015 Resuscitaire Use: Approach / priorities © Resuscitaire

dr.ChrisR.Mol,BME,NORTEC,2015

Components (selection)

© Resuscitaire

TheBassinetsideandfrontpanelsmaybefoldeddowntopermitmaximumaccesstotheinfant.Themattressmaybetiltedupintherearata5or10degreeangle.Openingsareprovidedoneachsideofthebassinetfortheinsertionofan(optional)X-rayCassetteTray.

TheControllercontrolsallfunctions,includingtheuserinterfacecontrol,thetimingofthepre-heating,alarms,theApgarTimer,theSkinTemperatureprobe,thermostat,etc.etc.

TheBlended Gas supply unit mixesairandoxygen,tobeusedduringventilation.

TheSuction machine, bottle and filter areneededtoe.g.removefluidsfromnew-bornairways.

Page 12: PowerPoint-presentatie  Resuscitaire pp.pdfdr. Chris R. Mol, BME, NORTEC, 2015 Resuscitaire Use: Approach / priorities © Resuscitaire

dr.ChrisR.Mol,BME,NORTEC,2015© Resuscitaire

Ventilator and Gas supply

control(example)

Page 13: PowerPoint-presentatie  Resuscitaire pp.pdfdr. Chris R. Mol, BME, NORTEC, 2015 Resuscitaire Use: Approach / priorities © Resuscitaire

dr.ChrisR.Mol,BME,NORTEC,2015

Alarms

© Resuscitaire

High Temperature Alarm:whenskintemperatureexceeds39.0OCtheheaterisautomaticallyturnedoffandanalarmisgiven.

Probe fail Alarm:iftheSkinTemperatureProbefails(short-circuitedoropen).

Apgar Timer:thetimerwillshowelapsedminutesandsecondsandtheaudiblealarmwillsoundatthe1,5and10minuteApgartimeintervals.

Baby Temperature Alarm:whenthetemperaturesensedbytheSkinTemperatureProbeis1OCaboveor1OCbelowtheselectedSetTemperatureDisplaysetting.

Power Fail Alarm:ifthePowerisinterruptedwhilethecontrollerison.

System Fail Alarm:ifaninternalmalfunctionisdetected;anerrorcodemaybedisplayed.

Blender Differential Bypass Alarm:willsoundwhenthepressuredifferentialbetweenO2andAirsuppliesexceeds20psi.Whenthisoccurs,theblenderwillcontinuetosupplywhichevergasisatthehigherpressure:either100%airor100%Oxygen.

Page 14: PowerPoint-presentatie  Resuscitaire pp.pdfdr. Chris R. Mol, BME, NORTEC, 2015 Resuscitaire Use: Approach / priorities © Resuscitaire

dr.ChrisR.Mol,BME,NORTEC,2015

Input / Output

© Resuscitaire

Inputs• air• oxygen• water• power

Outputs• air/oxygenunderpressureforventilation• heatforwarming• light• suction• alarms

Page 15: PowerPoint-presentatie  Resuscitaire pp.pdfdr. Chris R. Mol, BME, NORTEC, 2015 Resuscitaire Use: Approach / priorities © Resuscitaire

dr.ChrisR.Mol,BME,NORTEC,2015

Trouble shooting: temperature measurement

© Resuscitaire

Thetemperature probe isafrequentproblem.Thetechnologyintheprobeissimple;typicallyjustathermistor.However,determiningtheexacttypeandresistanceofthermistorrequiresdatafromthemanufacturer.Ifthisdatacanbeobtained,thanareplacementprobecanbemanufacturedfromastandardthermistorandtheproperconnector.Withoutthemanufacturer’sdata,itcanbeimpossibletoconstructtheprobe.

Apossiblesolutiontoamissingthermistorprobeistoplaceaknown,fixedresistorinitsplace,essentiallyforcingthedevicetooperateinmanual mode.Apotentiometercanbeusedtodeterminethevalueofresistancethatisrequiredtoforcethelightstocomeon.

Thermocouple probes aresometimesusedonwarmingunits,inwhichcasearesistancewillnotsubstitutefortheprobe.Avoltagesourcemightwork,however.Usingapotentiometer,abatteryandafixedresistor,avoltagedividercansometimesbecreatedandadjusteduntilthewarmingunitcomeson.Sincethedeviceisnowinmanualmode,careshouldbeusedtoinsurethatthepatient’stemperatureiscarefullymonitored.

Page 16: PowerPoint-presentatie  Resuscitaire pp.pdfdr. Chris R. Mol, BME, NORTEC, 2015 Resuscitaire Use: Approach / priorities © Resuscitaire

dr.ChrisR.Mol,BME,NORTEC,2015

Trouble shooting

© Resuscitaire

Thesametypesoffaultsoccurasonanaestheticmachine:• gasleaks,• rotametersstuck,• suctionnotworking,

thesecanbedealtwithinthesamewayasdescribedinthosesections.

Thesystemneedstobecheckedforelectricallysafetyatleastonceayear.